Found 9 results

510(k) Data Aggregation

    K Number
    K222460

    Validate with FDA (Live)

    Date Cleared
    2023-05-12

    (270 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intraoperative Ultrasound Probe Cover can be used to minimize contamination between the probe during ultrasound scanning procedures for both intact skin and compromised tissue. The cover allows use of the transducer in scanning and needle guided procedures for body surface, Endo cavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer.

    Device Description

    Intraoperative Ultrasound Probe Cover can be used to minimize contamination between the patient and the probe during ultrasound scanning procedures for both intact skin and compromised tissue. The cover allows use of the transducer in scanning and needle quided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer.

    The probe cover is designed for use in both invasive and non-invasive procedures. It comes in a sterile form only. The sterile cover made used in invasive procedures where it will come into direct contact with compromised tissue. The probe cover may be used with any multiple different probes including, but not limited to, an Endoscope re-usable, any rectal, vaginal, trans-urethral, gastro-entero probes, or pulmonary probe, In-vitro Fertilization instruments, biopsy probes, fiberoptic probes, catheter probes, any probe that is used during a surgical procedure, re-usable vascular (visualization) probes, and re-usable central nervous system (CNS) or cerebral probes.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an "Intraoperative Ultrasound Probe Cover." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness through a comparative effectiveness study. Therefore, the device's acceptance criteria are primarily related to its physical properties, sterility, and performance characteristics consistent with its intended use as a protective barrier, rather than diagnostic accuracy or direct patient outcomes. The study mentioned is a non-clinical performance testing study.

    Here's a breakdown of the information requested based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Test/PropertyReported Device Performance
    Barrier PropertiesViral penetrationMet acceptance criteria
    Water resistance/hydrostatic pressureMet acceptance criteria
    Water resistance/impact penetrationMet acceptance criteria
    Synthetic blood penetrationMet acceptance criteria
    Material PropertiesTensile strengthMet acceptance criteria
    ElongationMet acceptance criteria
    LintingMet acceptance criteria
    SterilitySterility (via EO Gas, ISO 11135-1:2014)Met acceptance criteria (SAL 10-6)
    Ethylene oxide residualsMet acceptance criteria
    Bacterial endotoxinsMet acceptance criteria
    BiocompatibilityBiocompatibilityMet acceptance criteria
    AcousticsAcoustics performanceMet acceptance criteria
    Packaging & Shelf LifePackagingTesting completed (results applicable from predicate)
    Shelf lifeTesting completed (results applicable from predicate)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes for each non-clinical test conducted. It generally states that "Non-clinical performance testing was conducted." The provenance of the data (country of origin, retrospective/prospective) is also not specified, beyond being part of the 510(k) submission from "Exact Medical Manufacturing" in Lancaster, NY.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of non-clinical performance testing for a medical device cover. The "ground truth" for these tests would be established by standardized testing methods and their associated industry-recognized specifications (e.g., ISO standards for sterility, ASTM standards for material properties), not by expert human interpretation of diagnostic images or clinical data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human interpretation or adjudication. The results are based on objective physical, chemical, and microbiological testing methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is an "Intraoperative Ultrasound Probe Cover," a physical barrier, not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance study was not done. This device is not an algorithm or software; it is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests were the established acceptance criteria for each specific test method, derived from recognized industry standards (e.g., ISO for sterility, unspecified standards for material properties, viral penetration, etc.). These criteria define what constitutes an acceptable performance level for the device's physical and barrier properties.

    8. The sample size for the training set

    Not applicable. This device is not an AI/machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211969

    Validate with FDA (Live)

    Date Cleared
    2021-08-13

    (50 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Ultrasonic Transducer Cover can be used to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures for external intact skin. This may help with easier cleaning and disinfection of the probe. This product is not intended for use of probes with endocavity or intraoperative use. The Exact Medical Manufacturing Ultrasonic Transducer Cover are also sold as bulk non-sterile, single use items, to repackager/ relabeler establishments for further packaging and ethylene oxide sterilization.

    Device Description

    A 0.05mm thin, 50 GSM (Grams per Square Meter), translucent high strength polyurethane film tube shape, nominally 48" x 6" with heat sealed distal end to be applied over a transducer probe to provide a Transducer Cover that can be used to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures for external intact skin. This may help with easier cleaning and disinfection of the probe. This product is not intended for use of probes with endocavity or intraoperative use. The Exact Medical Manufacturing Ultrasonic Transducer Cover are also sold as bulk non-sterile, single use items, to repackager/relabeler/kit packers establishments for further packaging and ethylene oxide sterilization.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of the EMM Ultrasonic Transducer Cover, which is a medical device.

    1. A table of acceptance criteria and the reported device performance:

    Test Performed & Description of testAcceptance CriteriaReported Device Performance / Result SummaryConclusion
    Water Resistance: Impact Penetration Test, Level 3Blotter weight gain of no more than 1.0 gAccept, 32 samples ≤ 1.0g (all samples = 0.0 g)Ability of a protective product to resist the penetration of liquids and liquid-borne microorganisms according to AAMI PB70, Level 3. Substantially equivalent barrier property compared to the predicate device.
    Water Resistance: Hydrostatic Pressure Level 3Hydrostatic resistance of at least 50 cmAccept, 32 samples > 50cm (all samples ≥ 180cm)Ability of a protective product to resist the penetration of liquids and liquid-borne microorganisms according to AAMI PB70, Level 3. Substantially equivalent barrier property compared to the predicate device.
    Flammability of Clothing Textiles (16CFR1610)Burn Time ≥ 3.5 secondsAccept, 5 samples DNI (Did Not Ignite)Able to meet and claim flammability Class 1 rating in accordance with 16CFR1610. Exceeds predicate device based upon predicate device has not been tested for this attribute.
    Tensile properties — Part 3: Test conditions for films and sheets (ISO 527-3)Tensile strength minimum 40N in machine direction, minimum 34N in cross direction. Elongation 108 % minimum in machine direction, 119% minimum in cross direction.Accept, all samples passed acceptance criteria in accordance with statistical sampling plan per ISO2859-1 (otherwise, material is rejected).Specification confirmation of tensile and elongation properties of plastic films or sheets less than 1 mm thick to assure compliant device barrier performance. Substantially equivalent tensile and elongation property compared to material specification.
    Lint and other particles generation in the dry state (ISO 9073-10)IPM =/≤ 3.5 (Index for Particle Matter)Accept, IPM 1.3 (outside), IPM 0.72 (inside)Material does not significantly generate lint or particles that may shed during an ultrasonic scan procedure. Exceeds predicate device based upon predicate device has not been tested for this attribute.
    Acoustic Output Measurement Standard for Diagnostic Ultrasound (NEMA UD 2-2004)Negligible effect on the transmission of ultrasound signal and does not degrade signal or image.Accept, Speed of sound = 1698 m/sec, Acoustic Attenuation = 0.51 dB, Acoustic Impedance = 1.63 MRayl.Based on the acoustic testing results, it was concluded that EMM's transducer/probe cover has negligible effect on the transmission of ultrasound signal and that EMM's transducer/probe cover does not degrade signal or image. Exceeds predicate device based upon predicate device has not been tested for this attribute.
    Compatibility of ultrasonic coupling gel with polyurethane (ASTM D543-20)(Acceptance Criteria Not Defined)Accept, Color N=9 Pass, Dimension N=9 Pass, Weight N=9 Pass, Coordinates N=9 Pass, Mechanical Properties N=9 Pass.Confirmation that device polyurethane material is not negatively affected by the acoustic coupling gel during an ultrasonic scanning procedure. Exceeds predicate device based upon predicate device does not provide compatibility evaluation of acoustic coupling gel with their device.
    Ethylene Oxide Residuals (ISO 10993-7:2008)Tolerable contact limit (TCL) for EO at 10ug/cm2 and ECH at 5mg/cm2Accept, Actual results for EO Residuals at Time 0 hours = 8.67ug/cm2. Actual results for ECH residuals at Time 0 hours = Not detected (within the limits of detection of 3.7ug/g).Confirmation of maximum allowable residues for ethylene oxide (EO) within the submission device sterilized with EO and to ensure that the levels of residual EO, ethylene chlorohydrin (ECH), and ethylene glycol (EG) pose a minimal risk to the patient in normal product use. Substantially equivalent to predicate device based upon acceptable testing results that comply with testing standard.
    In Vitro Cytotoxicity Test (ISO 10993-5;2009)Viability is reduced <70% of the blank indicates cytotoxic potential. The Viab.% of 100% extract of the test article is the final result.100% Test article extract 76.5% viability. No Test article extracts are <70% viability.Study conclusion, "Under the conditions of this study, the test article extract did not show potential toxicity to L929 cells."
    Closed patch Sensitization Study in Guinea Pigs (ISO 10993-10:2010)No significant evidence of causing skin sensitization in the guinea pig.No skin sensitization reaction was found in the Guinea pigs using extracts of the test article. The positive rate of sensitization was 0%. Dermal Reactions Challenge 24 Hour Score Ten test Guinea pigs, 0 (zero) reactions. 48 Hour Score Ten test Guinea pigs, 0 (zero) reactions.Under the conditions of the study, the test article extracts showed no significant evidence of causing skin sensitization in the Guinea pigs. Substantially equivalent to the predicate device.
    Intracutaneous Reactivity Test (ISO 10993-10:2010)The requirements of the test are met if the final test article score is 1.0 or less.All animals survived and no abnormal signs were observed. The skin reaction of polar and non-polar extract on testing side did not exceed that on the control side. Final test article score was calculated to be 0. Summary Analysis Test Results of Dermal Observations in Test Article Extract SC 24 hr = 0, 48 hr = 0, 72 hr = 0. Extract SO 24 hr = 0, 48 hr = 0, 72 hr = 0.The test result showed that the polar and non-polar extract of the final test sample score is less 1.0, the requirements of the test are met. Substantially equivalent to the predicate device.

    2. Sample sized used for the test set and the data provenance:

    • Water Resistance (Impact Penetration & Hydrostatic Pressure): N=32 for each test, with C=0 (meaning zero failures allowed). The sample size exceeds the reference standard requirements (N=32, C=3 for AAMI PB70 AQL of 4%/RQL of 20%). The provenance is a laboratory test of the Final Finished Device (both heat seal and TPU film).
    • Flammability: 5 replicates (Polyurethane Film). The provenance is a laboratory test of the material.
    • Tensile and Elongation: Sample size per statistical sampling plan ISO2859-1, based upon lot population quantity (Polyurethane Film). The provenance is a laboratory test of the material.
    • Linting: N=10 (Polyurethane film). The provenance is a laboratory test of the material.
    • Acoustic Output: 3 samples for Speed of Sound, 3 samples for Acoustic Attenuation, 3 samples for Acoustic Impedance (Final Finished Device). The provenance is a laboratory test of the device.
    • Compatibility with Coupling Gel: N=45 (Polyurethane Probe Cover Samples). Tested by the manufacturer of the coupling gel. This is a laboratory test.
    • Ethylene Oxide Residuals: N=15 X 3 consecutive sterilization batches = 45 samples (Polyurethane film). The provenance is a laboratory test on sterilized material.
    • In Vitro Cytotoxicity Test: N=4 test article extracts (Probe Cover polyurethane, EO Exposed). The provenance is a laboratory test.
    • Closed patch Sensitization Study in Guinea Pigs: Ten test Guinea pigs. The provenance is a biological study (animal study).
    • Intracutaneous Reactivity Test: Not explicitly stated, but implies multiple animals (rabbits) based on the "all animals survived" phrase. The provenance is a biological study (animal study).

    The data provenance is from bench testing and biological studies (animal studies) performed as part of the device submission. There is no indication of country of origin of the data beyond being part of a US FDA submission, nor is there information on whether the studies were retrospective or prospective, though bench and animal studies are typically prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This device is not an AI/ML device that requires human expert interpretation to establish a ground truth. The ground truth for this device is based on objective measurements against established industry standards and biological endpoints. Therefore, no human experts were used to establish a ground truth in the context of diagnostic interpretation.

    4. Adjudication method for the test set:

    Not applicable. This device does not involve human interpretation or subjective assessments that would require an adjudication method like 2+1 or 3+1. The tests are objective and quantitative.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-assisted diagnostic device, so no MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm-based device.

    7. The type of ground truth used:

    The ground truth used for the various tests includes:

    • Objective measurement against established standards: E.g., blotter weight gain for water resistance, burn time for flammability, tensile strength values for mechanical properties, IPM for linting, speed of sound and acoustic attenuation for acoustic performance, and residual limits for EO residuals.
    • Biological endpoints: For biocompatibility tests (cytotoxicity, sensitization, intracutaneous reactivity), the ground truth is based on the observed biological response (e.g., cell viability, skin reactions) against defined criteria.

    8. The sample size for the training set:

    Not applicable. This device does not involve an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121919

    Validate with FDA (Live)

    Date Cleared
    2013-02-14

    (227 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Surgical Drape AAMI Level 4 are sterile or non-sterile single use devices made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The EMM Surgical Drapes AAMI Level 4 are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization according to ISO 11135-1:2007

    Device Description

    Exact Medical Manufacturing Surgical Drape AAMI Level 4 are sterile or non-sterile single use devices made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.
    Exact Medical Manufacturing Surgical Drape with AAMI Liquid Barrier Level The base sheet fabric is a three layer composite comprised of polypropylene spunbond and a co-extruded PE film. Layers are thermally bonded together producing a single layer with various basis weights. The outer layers are non woven fabric, the inner layer is AAMI PB:70 Level 4 capable PE film.

    AI/ML Overview

    The provided document K121919 describes a surgical drape, not an AI/ML powered medical device. Therefore, many of the requested categories for AI/ML devices (e.g., sample sizes for test/training sets, number of experts, adjudication methods, MRMC studies, standalone performance) are not applicable.

    However, I can extract the acceptance criteria and performance data for this physical medical device based on the information provided.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Predicate Device K102666)Reported Device Performance (Exact Medical Manufacturing Surgical Drape with AAMI Liquid Barrier Level IV)
    Barrier PerformanceAAMI PB70: Level 4AAMI PB70: Level 4
    BiocompatibilityCytotoxicity, Primary Skin Irritation, Sensitization testsISO 10993-5:2009 Cytotoxicity, ISO 10993-10:2002 Skin Irritation, ISO 10993-10:2002 Sensitization
    Tear Strength (ASTM 5587-08)Unknown/Equivalent to Predicate19.4 lbs & 9.7 lbs
    Breaking Strength (ASTM D5034-09)Unknown/Equivalent to Predicate24.4 lbs & 15.9 lbs
    Flammability (16CFR1610:2007)Class 1Class 1
    Linting (ISO 9073-10:2003)Unknown/Equivalent to PredicateReported as 'Substantially Equivalent' to predicate, implying it met predicate's (unknown) criteria
    SterilitySterile via EO GasSterile (via EO Gas) ISO 11135-1:2007

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a "Summary of Testing" described in the 510(k) submission. It states:

    "Exact Medical Manufacturing Surgical Drape with AAMI Liquid Barrier Level IV is substantially equivalent and meets the same acceptance criteria as the predicate device/Drape in K102666 Non-clinical performance testing includes: Biocompatibility (cytotoxicity, irritation, sensitization) in compliance with the methods of ISO 10993, Barrier properties- AAMI PB:70 Level 4, tensile, tear strength, flammability, linting and sterility. All results of the testing met acceptance criteria."

    Regarding AI/ML Specific Information (Categories 2-9):

    As this is a physical surgical drape, the following categories are not applicable and no information is provided in the document:

    1. Sample size used for the test set and the data provenance: Not applicable for a physical device undergoing non-clinical performance testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Material properties and performance are measured against standards, not expert consensus.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For a physical device, the "ground truth" or reference for acceptance is established by recognized international and national standards (e.g., AAMI PB70, ISO 10993, ASTM, 16CFR1610).
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120045

    Validate with FDA (Live)

    Date Cleared
    2012-06-12

    (158 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Surgical Gown with AAMI Liquid Barrier Level IV are sterile or nonsterile single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Sterile surgical gowns are to be sold directly to users after E.O sterlization and validated to ISO 11135-1:2007.

    The Exact Medical Manufacturing Surgical Gowns with AAMI Level IV Liquid Barner are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization according to ISO 11135-1:2007

    Device Description

    Exact Medical Manufacturing Surgical Gown with AAMI Liquid Barrier Level IV are sterile or non-sterile single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
    Exact Medical Manufacturing Surgical Gown with AAMI Liquid Barrier Level IV is comprised of a 3 ply laminate non woven material - outer layers are nonwoven fabric - inner layer AAMI PB:70 Level 4 capable film. The gowns consist of 100% polyester cuffs sewn to the end of the sleeves using nylon thread. The gowns have a manual closure system.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Exact Medical Manufacturing Surgical Gown with AAMI Liquid Barrier Level IV. This document focuses on the regulatory submission and does NOT involve an AI/ML algorithm or a study with human readers, deep learning, or image analysis. Therefore, many of the requested categories are not applicable.

    Here's an analysis of the provided information:

    Acceptance Criteria and Reported Device Performance

    The device is a surgical gown, and its performance is evaluated against established standards for medical textiles, particularly AAMI PB:70 Level 4 for barrier properties.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Standard/Test)Reported Device Performance
    Material PerformanceAAMI PB:70 Level 4 (Liquid Barrier)Met acceptance criteria
    BiocompatibilityISO 10993 (Cytotoxicity, Irritation, Sensitization)Met acceptance criteria
    Tensile StrengthNot explicitly stated (standard)Met acceptance criteria
    Tear StrengthNot explicitly stated (standard)Met acceptance criteria
    FlammabilityNot explicitly stated (standard)Met acceptance criteria
    LintingNot explicitly stated (standard)Met acceptance criteria
    SterilityNot explicitly stated (standard)Met acceptance criteria

    Study Information (Not Applicable for this Device Type)

    The provided document describes a device for regulatory approval (510(k) submission) for a physical medical device (surgical gown), not a diagnostic AI/ML algorithm or software. Therefore, the following points are not applicable.

    1. Sample size used for the test set and the data provenance: Not applicable. Tests are conducted on physical samples of the gown material, not a "test set" of data or patient cases.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Material testing standards define performance, not expert ground truth on cases.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. "Ground truth" here refers to the measured physical properties of the material against established international standards.
    7. The sample size for the training set: Not applicable. There is no training set for a physical device like a surgical gown.
    8. How the ground truth for the training set was established: Not applicable.

    Explanation of "Not Applicable" for AI/ML-specific questions:

    The 510(k) summary is for a physical medical product (surgical gown), not a software or AI/ML algorithm. The performance evaluation revolves around bench testing of the physical material's properties (e.g., barrier effectiveness, strength, biocompatibility) against established industry standards (like AAMI PB:70 Level 4 and ISO 10993). Concepts such as "test set," "training set," "data provenance," "expert ground truth," "adjudication," "multi-reader multi-case studies," or "human-in-the-loop performance" are specific to the validation of diagnostic algorithms or AI-powered systems, which are not relevant to this product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111535

    Validate with FDA (Live)

    Date Cleared
    2011-08-26

    (85 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Surgical Gown - Spunlace w/PE Layer are sterile single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

    The Exact Medical Manufacturing Surgical Gown - Sounlace w/PE Laver are also sold as built nonsterile, single use items, to repackagerrelabeler establishments for further packaging and ethylene oxide sterilization.

    Device Description

    Exact Medical Manufacturing Surgical Gown-SPL are single use, disposable surgical gowns using in the OR as a protective covering, for the operating room staff, from the transfer of microorganisms, body fluids and particulates. Exact Medical Manufacturing Surgical Gowns- Spunlace w/PE Layer are comprised of disposable nonwoven (Spunlace), 100% polyester cuffs, Velcro neck closure, reinforced areas fabric or polyethylene

    AI/ML Overview

    This document describes the 510(k) submission for Exact Medical Manufacturing's Surgical Gowns – Spunlace w/PE Layer. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of an AI-based device. Therefore, many of the requested categories related to AI device evaluation (e.g., AI integration, multi-reader multi-case studies, training set details) are not applicable.

    Here's an analysis of the provided text in relation to your request about acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implicit from "met acceptance criteria")Reported Device Performance
    Barrier PropertiesEquivalent to the predicate device (Medline (Proxima) Surgical Gowns K964142) as per relevant standards."All results of the testing met acceptance criteria."
    Tensile StrengthEquivalent to the predicate device (Medline (Proxima) Surgical Gowns K964142) as per relevant standards."All results of the testing met acceptance criteria."
    Tear StrengthEquivalent to the predicate device (Medline (Proxima) Surgical Gowns K964142) as per relevant standards."All results of the testing met acceptance criteria."
    FlammabilityEquivalent to the predicate device (Medline (Proxima) Surgical Gowns K964142) as per relevant standards. (Typically per 16 CFR Part 1610 for apparel textiles.)"All results of the testing met acceptance criteria."
    LintingEquivalent to the predicate device (Medline (Proxima) Surgical Gowns K964142) as per relevant standards."All results of the testing met acceptance criteria."
    SterilityAchieves sterility through Ethylene Oxide (ETO) process meeting AAMI 11135-1:2007 (Sterilization of Healthcare Products) with specific parameters: - Pre-Conditioning: Temperature = 42°C, RH = 60%, Dwell Time = 6 hours - EO Cycle: Temperature = 54°C, RH >= 30%, EO Concentration = 730mg/m3, Dwell Time = 9 hours - Aeration: Temperature = ambient, Time >= 72 hours - EO Residuals: EO <= 4mg max., ECH <= 9mg max."All results of the testing met acceptance criteria." (Specific to sterility, this would mean the ETO process outlined achieves the defined sterility assurance level and residual limits.)
    BiocompatibilityMeets the methods of ISO 10993 for Cytotoxicity, Irritation, and Sensitization."All results of the testing met acceptance criteria."

    Study Proving Device Meets Acceptance Criteria:

    The document describes "Non-clinical performance testing" that was conducted to demonstrate substantial equivalence to the predicate device (Medline (Proxima) Surgical Gowns 510(k)964142).

    • Study Design: This was a series of in-vitro and laboratory tests designed to evaluate the physical properties, barrier performance, biocompatibility, and sterilization effectiveness of the new surgical gown. It's a non-clinical, comparative study against established standards and the characteristics of the predicate device.
    • Conclusion: "All results of the testing met acceptance criteria." and the gowns are deemed "substantially equivalent in all specifications and performance compared to the predicate device... and do not present any differences that could raise concerns related safety and effectiveness."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated for each test. The wording "All results of the testing met acceptance criteria" suggests that a sufficient number of samples were tested for each performance characteristic to demonstrate compliance with the relevant standards.
    • Data Provenance: Not specified. It's reasonable to assume the testing was conducted in a laboratory setting, likely in the US, given the submission to the FDA. The data would be prospective in nature for these performance tests (i.e., new tests performed on newly manufactured devices or materials).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is not an AI/diagnostic device that requires expert ground truth for image interpretation or similar tasks. The "ground truth" here is defined by established international and national performance standards (e.g., ISO 10993 for biocompatibility, AAMI 11135-1:2007 for sterilization, and likely ASTM or AAMI standards for physical properties and barrier performance).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As this is not an AI/diagnostic device, there is no expert adjudication process for results. Performance is assessed against quantitative and qualitative standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted diagnostic or interpretation device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device (surgical gown), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the performance tests consists of established industry, national, and international standards for medical device performance, material properties, biocompatibility, and sterilization. For example:
      • Physicochemical tests: Standards like ASTM for tensile/tear strength, AAMI for barrier properties.
      • Biocompatibility: ISO 10993 series.
      • Sterilization: AAMI 11135-1:2007.
      • The predicate device's performance also serves as a comparator for demonstrating substantial equivalence.

    8. The sample size for the training set

    • Not Applicable. This is not an AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI device that requires a training set or its associated ground truth establishment.
    Ask a Question

    Ask a specific question about this device

    K Number
    K101598

    Validate with FDA (Live)

    Date Cleared
    2010-10-04

    (118 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Surgical Drape SMS w/PE Sides are sterile or non-sterile single use devices made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.

    The Exact Medical Manufacturing Surgical Drape Surgical Drape SMS w/PE Sides are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization

    Device Description

    Exact Medical Manufacturing Surgical Drapes Surgical Drape SMS w/PE Sides are sterile or non-sterile single use devices made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.

    Exact Medical Manufacturing Surgical Drapes SMS w/PE Sides are comprised of a single layer of SMS (spunbond/meltblown/spunbond polypropylene), PE Layer, 3M Medical Adhesive Tape.

    AI/ML Overview

    The provided text is for a 510(k) Summary for a medical device called "EMM Surgical Drape SMS w/PE Sides." This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving performance against specific acceptance criteria for a novel AI or diagnostic device.

    Therefore, many of the requested items related to AI/diagnostic device studies (like sample size for test/training sets, expert qualifications, adjudication methods, MRMC studies, effect size of AI assistance, standalone performance, and how ground truth was established) are not applicable to this type of submission.

    However, I can extract the acceptance criteria and the summary of testing performed to demonstrate that the device met these criteria, in comparison to a predicate device.


    Acceptance Criteria and Reported Device Performance

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from Predicate Device/Industry Standards)Reported Device Performance (EMM Surgical Drape SMS w/PE Sides)
    Indications for Use: Protective patient covering to isolate a site of surgical incision from microbial and other contamination.Substantially Equivalent (Same indications for use)
    Classification: KKX, Surgical Drapes, 21CFR878.4370, Class IISubstantially Equivalent (Same classification)
    Materials & Construction: SMS, absorbent reinforcement, 3M medical grade adhesiveSubstantially Equivalent (SMS, Polyethylene, Absorbent Reinforcement, 3M medical grade adhesive)
    Barrier properties (AAMI PB70:2003 /(R)2009): Hydrostatic head = 30.5 cm (Predicate)Hydrostatic Head =/> 50 cm (meets AAMI PB70 Level 3)
    Sterility: Non-sterile (Predicate)Sterile (via EO Gas) ISO 11135-1:2007 (also available non-sterile)
    Biocompatibility (ISO 10993-5:1999, ISO 10993-10:2002): Cytotoxicity, Irritation, Sensitization PASSSubstantially Equivalent (Cytotoxicity, Irritation, Sensitization tests PASS)
    Tear Strength (ASTM D5587-08): Md=1.83 lbs, Cd= 3.24 lbs (Predicate)Substantially Equivalent (Md and Cd within general industry tolerance +/- 20%)
    Tensile Strength (ASTM D5034-09): Md=13.8 lbs, Cd=18.8 lbs (Predicate)Substantially Equivalent (Md and Cd within general industry tolerance +/- 20%)
    Flammability (16CFR1610:2010): Class 1 (Predicate)Substantially Equivalent (Class 1 - PASS)
    Lint and other Particles (ISO 9073-10:2003): No test specified for predicateTest performed (Not Applicable for direct comparison against predicate criteria as predicate had no test)
    Sterile Packaging: Not Applicable (Predicate)Chevron peel pouch (coated paper, PET12/PE40 film construction), individual CSR internal wrap

    Note: The reported device performance often states "Substantially Equivalent" indicating that the EMM Surgical Drape met or exceeded the performance characteristics of the predicate device, or adhered to recognized standards. For Barrier Properties, the EMM device reports a higher hydrostatic head, meeting a higher level (Level 3) compared to the stated predicate value, which implies it exceeds the predicate's performance.


    Study Details (as applicable for this type of submission)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable in this context. This submission is for a physical medical device (surgical drape) and relies on laboratory-based performance testing rather than clinical data or a diagnostic test set with patient data. The "test set" here refers to samples of the surgical drape material. The document does not specify sample sizes for each physical test, beyond stating that "All results of the testing met acceptance criteria." Data provenance (country of origin, retrospective/prospective) is not mentioned or relevant for these types of material property tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. Ground truth, in the sense of expert review for diagnostic accuracy, is not relevant here. The "ground truth" for material properties is established by defined physical and chemical testing standards (e.g., ISO, ASTM, AATCC) and the results obtained from those standardized tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are typically for subjective expert review of medical images or clinical cases. The testing for the surgical drape involves objective physical property measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a physical medical device, not an AI or diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an AI or diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for this device's performance is established by objective measurements against recognized industry standards and regulatory requirements for material properties and biocompatibility. This includes compliance with standards such as ISO 10993 (Biocompatibility), AAMI PB:70 (Barrier properties), ASTM D5587 (Tear Strength), ASTM D5034 (Tensile Strength), 16CFR1610 (Flammability), and ISO 9073-10 (Linting). The predicate device's established performance also serves as a benchmark for "substantial equivalence."

    8. The sample size for the training set

    • Not Applicable. There is no "training set" in the context of physical medical device testing for substantial equivalence.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set is used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K101688

    Validate with FDA (Live)

    Date Cleared
    2010-09-20

    (96 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Surgical Drape - Spunlace w/PE Sides are sterile or non-sterile single use devices made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.

    The Exact Medical Manufacturing Surgical Drape - Spunlace w/PE Sides are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization

    Device Description

    Exact Medical Manufacturing Surgical Drape -Spunlace w/PE Sides are sterile or non-sterile single use devices made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.

    Exact Medical Manufacturing Surgical Drapes-Spunlace w/PE Sides are comprised of Spunlace, Polyethylene, 3M Medical Adhesive Tape.

    AI/ML Overview

    The document provided describes the acceptance criteria and the study that proves the device meets the acceptance criteria for the "Exact Medical Manufacturing Surgical Drape-Spunlace w/PE Sides, Model 13-004".

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaReported Device Performance/Result
    Water Resistance: Impact Penetration Test (AATCC 42:2007)<1.0 gm Blotter water weight gain (Ref: AAMI PB70:2003/(R)2009, Level 3)Pass
    Water Resistance: Hydrostatic Pressure Test (AATCC 127:2008)=/> 50 cm hydrostatic resistance (Ref: AAMI PB70:2003/(R)2009, Level 3)Pass
    Biocompatibility: In Vitro Cytotoxicity (ISO 10993-5:1999)Evidence of cell lysis or toxicity < 2Pass
    Biocompatibility: Skin Irritation (ISO 10993-10:2002)No (0) edema or erythema observedPass
    Biocompatibility: Skin Sensitization (ISO 10993-10:2002)No evidence of causing delayed dermal contact sensitizationPass
    Flammability of Clothing Textiles (16CFR1610:2010)Class 1 =/> 3.5 sec. average flame spreadPass (Class 1)
    Tearing Strength of Fabrics by Trapezoid Procedure (ASTM D5587-08)Acceptance criteria not established in recognized standard. Exceeds predicate performance.Pass
    Breaking Strength and Elongation of Textile Fabrics (Grab Test) (ASTM D5034-09)Acceptance criteria not established in recognized standard. Exceeds predicate performance.Pass
    Sterilization of health care products - Ethylene oxide (ISO 11135-1:2007)SAL of > $10^{-6}$Pass
    Ethylene Oxide Sterilization Residuals (ISO 10993-7:2008)Average daily dose of EO/ECH shall not exceed 4mg/9mgPass
    Lint and other particles generation in the dry state (ISO 9073-10:2003)Acceptance criteria not established in the recognized standard.Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not explicitly state the sample size for each test. However, for each test, a single "Finished Good Lot Number" (0980APA2) of the device (Model No. 13-004 Sterile) was used as the "Test Article". This suggests that samples from this specific lot were subjected to all listed tests.
    • Data Provenance: The tests were conducted across several accredited laboratories:
      • Nelson Labs, Utah, USA
      • LexaMed, Ohio, USA
      • LexaMed/NAMSA, Ohio, USA
      • SCDC, Shanghai, CN
      • GOALS Sterilization Co., Jiaxing, CN
        The data appears to be prospective as it involves pre-market testing of the device against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This type of information is not applicable to this document. The tests performed are objective, physical, and chemical performance tests for a medical device (surgical drape), not diagnostic tests requiring expert interpretation of results or ground truth derived from expert consensus. The "ground truth" is established by the specified industry standards and their defined acceptance criteria.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable for the type of testing described. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, where multiple experts might disagree on an outcome. For these performance tests, the results either objectively meet or do not meet the predefined numerical or qualitative criteria of the standard.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a surgical drape, not an AI-powered diagnostic tool, and therefore, MRMC comparative effectiveness studies involving human readers and AI assistance are irrelevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a physical surgical drape, not an algorithm, so the concept of standalone algorithm performance does not apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests of the surgical drape is based on recognized industry standards and their predefined acceptance criteria. This includes standards like AATCC, ASTM, ISO, and 16CFR, which specify objective measurements and thresholds for properties such as water resistance, biocompatibility, flammability, and tensile strength.

    8. The sample size for the training set

    This is not applicable. The device is a physical product, not a machine learning model, so there is no concept of a "training set" in its evaluation.

    9. How the ground truth for the training set was established

    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101689

    Validate with FDA (Live)

    Date Cleared
    2010-09-20

    (96 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Equipment Cover - Polyethylene are sterile single use devices made of natural or synthetic materials intended to be used as a protective equipment covering, such as to isolate equipment from microbial and other contamination.

    The Exact Medical Manufacturing Equipment Covers-Polyethylene are also sold as bulk nonsterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization.

    Device Description

    Exact Medical Manufacturing Equipment Cover- Polyethylene are single use, disposable equipment cover used in the OR as a protective covering, for the operating equipment, from the transfer of microorganisms, body fluids and particulates. Exact Medical Manufacturing Equipment Covers - Polyethylene are comprised Polyethylene with absorbent polypropylene. The Exact Medical Manufacturing Equipment Covers-Polyethylene are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called the "EMM Equipment Cover - Polyethylene." This type of document is for regulatory clearance and focuses on demonstrating substantial equivalence to a predicate device, rather than a clinical study evaluating the device's diagnostic or therapeutic performance. As such, many of the typical elements requested for an AI/diagnostic device study (like sample sizes for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, training sets, and ground truth establishment for training data) are not applicable or present in this context.

    However, I can extract the acceptance criteria and reported device performance from the provided "Non-Clinical Testing Summary."

    1. Table of acceptance criteria and the reported device performance:

    TestAcceptance CriteriaReported Device Performance
    Water Resistance: Impact Penetration Test, Level 3<1.0 gm Blotter water weight gainPass
    Water Resistance: Hydrostatic Pressure Test, Level 3=/> 50 cm hydrostatic resistancePass
    Flammability of Clothing Textiles - Class 1Class 1 =/> 3.5 sec. average flame spreadPass
    Tearing Strength of Fabrics by Trapezoid ProcedureAcceptance criteria not established in recognized standard. Tear Strength for Md and Cd within general industry tolerance of +/- 20%Pass
    Breaking Strength and Elongation of Textile Fabrics (Grab Test)Acceptance criteria not established in recognized standard. Tensile Strength for Md and Cd within general industry tolerance of +/- 20%Pass
    Sterilization of health care products - Ethylene oxide - Part 1SAL of > 10-6Pass
    Lint and other particles generation in the dry stateAcceptance criteria not established in the recognized standardPass

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: The document does not specify the sample size for each test. It refers to a "Finished Good Lot Number: 0980APA3" for multiple tests, implying that samples were drawn from this lot.
    • Data Provenance: The tests were conducted at "Nelson Labs, Utah, USA" and "SCDC, Shanghai, CN, LexaMed, Ohio, USA." This indicates the tests were conducted in the USA and China, presumably using newly manufactured equipment covers. These are non-clinical (laboratory/performance) tests, not clinical data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The device is an equipment cover, and the "ground truth" is established by laboratory performance standards and measurements, not by expert interpretation of clinical data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable. The tests are based on objective physical and chemical properties measured according to established standards, not on expert adjudication of ambiguous cases.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is an equipment cover and not an AI-powered diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is an equipment cover and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is defined by adherence to recognized international and national standards for material properties, barrier performance, flammability, and sterility. These are objective, measurable criteria, not subjective interpretations.

    8. The sample size for the training set:

    This information is not applicable. This is a manufactured product undergoing non-clinical performance testing, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established:

    This information is not applicable. As there is no training set, there is no ground truth establishment process for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101593

    Validate with FDA (Live)

    Date Cleared
    2010-09-03

    (87 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Exact Medical Manufacturing Surgical Gown SMS w/PE Layer are sterile or nonsterile single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The Exact Medical Manufacturing Surgical Gown SMS w/PE Layer are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and ethylene oxide sterilization

    Device Description

    Exact Medical Manufacturing Surgical Gown SMS w/PE Layer has the same design, material and performance characteristics of the predicate device.

    AI/ML Overview

    The Exact Medical Manufacturing Surgical Gown SMS w/PE Layer met the acceptance criteria by demonstrating substantial equivalence to its predicate device (K023117) through a series of non-clinical performance tests.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Material Properties:
    DesignSame as predicate device (K023117)
    MaterialSame as predicate device (K023117)
    Performance CharacteristicsSame as predicate device (K023117)
    Biocompatibility (ISO 10993):
    CytotoxicityMet acceptance criteria
    IrritationMet acceptance criteria
    SensitizationMet acceptance criteria
    Barrier Properties (AAMI PB:70):
    Level 3 (Fluid resistance, microbial barrier, particulate matter)Met acceptance criteria
    Physical Strength:
    Tensile StrengthMet acceptance criteria
    Tear StrengthMet acceptance criteria
    Other Performance:
    FlammabilityMet acceptance criteria
    LintingMet acceptance criteria
    Sterility (for sterile versions/post-sterilization)Met acceptance criteria

    All results of the testing met acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document mentions "All results of the testing met acceptance criteria," implying that sufficient samples were tested for each criterion according to standard testing protocols for medical devices.
    • Data Provenance: The data is generated from non-clinical performance testing conducted by Exact Medical Manufacturing, Inc. It is internal testing performed for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is a non-clinical device (surgical gown) and does not involve human interpretation of data where expert consensus would be required to establish ground truth such as in diagnostic imaging. The "ground truth" is established by adherence to recognized national and international standards (e.g., ISO 10993, AAMI PB:70).

    4. Adjudication method for the test set:

    • Not applicable. As noted above, this involves objective, quantifiable laboratory tests against established standards, not interpretation or adjudication by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a surgical gown, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a surgical gown, not an algorithm.

    7. The type of ground truth used:

    • The ground truth is based on accepted national and international standards and specifications for surgical gowns and medical device materials, including:
      • ISO 10993 for biocompatibility.
      • AAMI PB:70 for barrier properties (Level 3).
      • Other specific test methods for tensile strength, tear strength, flammability, linting, and sterility.

    8. The sample size for the training set:

    • Not applicable. This is a physical medical device, not a machine learning model, so there is no concept of a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable, for the reason stated above.