Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K190942

    Validate with FDA (Live)

    Date Cleared
    2020-01-07

    (272 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    "Disposable Powder Free Nitrile Examination Glove, Pink Color" and "Disposable Powder Free Nitrile Examination Glove, Black Color" are patient examination gloves made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    This document is a 510(k) summary for a disposable powder-free nitrile examination glove, and as such, it focuses on demonstrating substantial equivalence to a predicate device based on bench testing. It does not describe an AI medical device, and therefore, many of the requested elements for AI/ML evaluation (such as sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone performance of an algorithm) are not applicable.

    However, I can extract the acceptance criteria and reported device performance from the provided text for the physical characteristics of the glove.

    1. Table of acceptance criteria and the reported device performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM D6319-10 (Reapproved 2015)
    Length≥230mmComplies with ASTM D6319-10 (230mm min.)
    Width (X-Small)70 ± 10 mmComplies with ASTM D6319-10 (70 ± 10 mm)
    Width (Small)80 ± 10 mmComplies with ASTM D6319-10 (80 ± 10 mm)
    Width (Medium)95 ± 10 mmComplies with ASTM D6319-10 (95 ± 10 mm)
    Width (Large)110 ± 10 mmComplies with ASTM D6319-10 (110 ± 10 mm)
    Width (X-Large)120 ± 10 mmComplies with ASTM D6319-10 (120 ± 10 mm)
    Thickness (Finger tip)≥0.05mmComplies with ASTM D6319-10 (0.05mm min.)
    Thickness (Palm)≥0.05mmComplies with ASTM D6319-10 (0.05mm min.)
    Physical PropertiesASTM D6319-10 (Reapproved 2015)
    Tensile strength (Before aging)≥14MPa14 MPa, min.
    Tensile strength (After aging)≥14MPa14 MPa, min.
    Elongated rate (Before aging)≥500%500% min.
    Elongated rate (After aging)≥400%400% min.
    Freedom from pinholesPassed Standard Acceptance Criteria (21 CFR 800.20, ASTM D6319-10, ASTM D5151-06)In accordance with ASTM D6319-10 and ASTM D5151-06, G-1, AQL 2.5
    Powder Residual< 2 mg/glove (ASTM D6319-10, D6124-06)< 2mg per glove (Complies with ASTM D6319-10)
    BiocompatibilityISO 10993-10, ISO 10993-5 criteria
    Primary Skin IrritationNot a primary skin irritantPasses (Not a primary skin irritant)
    Dermal sensitizationNot a primary skin sensitizerPasses (Not a primary skin sensitizer)
    In vitro cytotoxicityNot cytotoxicPasses (Not cytotoxic)

    2. Sample size used for the test set and the data provenance
    The document refers to "bench testing" and compliance with ASTM standards and ISO standards. These standards typically involve specific sample sizes and testing methodologies defined within the standard itself. The document does not explicitly state the exact number of units tested for each specific criterion. Data provenance is not specified beyond "bench testing" implies laboratory testing of the physical product.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This is a physical device (nitrile examination glove) and not an AI/ML medical device where expert ground truth for interpretation would be required. The "ground truth" here is the pass/fail result against established material and performance standards.

    4. Adjudication method for the test set
    Not applicable, as this is bench testing of physical properties against established standards, not an AI output requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI medical device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. This is not an AI medical device.

    7. The type of ground truth used
    The "ground truth" for the device's performance is based on established national and international standards for medical gloves, specifically:

    • ASTM D6319-10 (Reapproved 2015) - Standard specification for Nitrile Examination Gloves
    • 21 CFR 800.20 - Freedom from holes
    • ASTM D5151-06 (Reapproved 2015) - Test method for detection of holes in medical gloves
    • ASTM D6124-06 (Reaffirmation 2011) - Test method for residual powder on medical gloves
    • ISO 10993-10: Third Edition 2010-08-01 - Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
    • ISO 10993-5: Third Edition 2009-06 - Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity

    8. The sample size for the training set
    Not applicable. This is not an AI medical device.

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190736

    Validate with FDA (Live)

    Date Cleared
    2019-12-23

    (277 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs is a patient examination glove made from nitrile compound. non-sterile (as per 21 CFR 880.6250. Class I). The patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA, evaluating a medical device for substantial equivalence to a legally marketed predicate device. The information provided heavily focuses on bench testing and comparison to standards, rather than clinical studies with human participants or AI-driven performance metrics typically found in software as a medical device (SaMD) clearances.

    Therefore, many of the requested details, such as the use of a test set with ground truth established by experts, MRMC studies, or training set details for an AI model, are not applicable to this document. The "device" in this context is a physical product (nitrile examination gloves) and not an AI/ML software.

    The acceptance criteria and performance are related to the physical and chemical properties of the gloves, not AI performance.

    Here's an attempt to answer the questions based only on the provided text, noting where information is not applicable (N/A) or not present.

    Acceptance Criteria and Device Performance for Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are established by adherence to various ASTM standards and 21 CFR regulations for physical and chemical properties, and specifically ASTM D6978-05 for chemotherapy drug permeation. The reported performance is the device's measured characteristics against these standards.

    CharacteristicAcceptance Criteria (Standard)Reported Device PerformanceComparison
    DimensionASTM D6319-10 (Reapproved 2015)
    Length≥230mmComplies with ASTM D6319-10 (230mm min.)Met
    Width (X Small)70 ± 10 mmComplies with ASTM D6319-10 (70±10 mm)Met
    Width (Small)80 ± 10 mmComplies with ASTM D6319-10 (80±10 mm)Met
    Width (Medium)95 ± 10 mmComplies with ASTM D6319-10 (95±10 mm)Met
    Width (Large)110 ± 10 mmComplies with ASTM D6319-10 (110±10 mm)Met
    Width (X Large)120 ± 10 mmComplies with ASTM D6319-10 (120±10 mm)Met
    Thickness - Finger tip≥0.05mmComplies with ASTM D6319-10 (0.05mm min.)Met
    Thickness - Palm≥0.05mmComplies with ASTM D6319-10 (0.05mm min.)Met
    Physical PropertiesASTM D6319-10 (Reapproved 2015)
    Tensile strength (Before aging)≥14MPa≥14MPaMet
    Tensile strength (After aging)≥14MPa≥14MPaMet
    Elongated rate (Before aging)≥500%≥500%Met
    Elongated rate (After aging)≥400%≥400%Met
    Freedom from pinholes21 CFR 800.20, ASTM D6319-10 (Reapproved 2015), Test method: ASTM D5151-06 (Reapproved 2015)Passed Standard Acceptance Criteria (AQL 2.5)Met
    Powder ResidualASTM D6319-10 (Reapproved 2015), Test method: D6124-06 (Reaffirmation 2011)< 2 mg/gloveMet
    BiocompatibilityISO 10993-10: Third Edition 2010-08-01, ISO 10993-5: Third Edition 2009-06
    Primary Skin IrritationPasses"Not a primary skin irritant"Met
    Dermal SensitizationPasses"Not a primary skin sensitizer"Met
    In vitro cytotoxicityPasses"Not cytotoxic"Met
    Chemotherapy PermeationASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs (Max testing time 240 minutes)Listed below for selected drugs, with others being > 240 minutes.Met (except for stated exceptions)
    Carmustine (BCNU) (3.3 mg/ml)N/A (Comparative, less than 240 mins expected)59.4 minutesN/A
    Thiotepa (10.0 mg/ml)N/A (Comparative, less than 240 mins expected)118.5 minutesN/A
    Most other 34 chemotherapy drugs> 240 minutes> 240 minutesMet

    Note: For Carmustine and Thiotepa, the "acceptance criteria" is not explicitly defined as a threshold, but rather it is noted that these drugs have "extremely low permeation time," and the observed times are reported. The comparison is against the predicate device's performance for these specific drugs, which the proposed device improved upon (59.4 vs 12.4 for Carmustine, 118.5 vs 24.4 for Thiotepa).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify the exact sample sizes for each specific test (e.g., how many gloves were tested for pinholes, or how many samples for each chemotherapy drug). It references adherence to ASTM standards, which would dictate appropriate sample sizes for these types of bench tests.
    • Data Provenance: The tests were conducted to evaluate the qualities of the "Disposable Powder Free Nitrile Examination Glove, Blue Color" manufactured by Ever Growth (Vietnam) Co. Ltd. (Vietnam). These are bench tests on manufactured products, not data from human subjects or collected retrospectively/prospectively in a clinical setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • N/A. The "ground truth" for the performance of these gloves is established by objective, standardized bench testing methods (e.g., ASTM standards for material properties, chemical permeation tests). There is no "expert consensus" or human adjudication involved in determining these physical and chemical properties.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. No adjudication method is applicable as these are objective bench test measurements, not subjective human interpretations that require reconciliation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This device is a physical medical glove, not an AI-powered diagnostic or assistive technology. No MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    • N/A. This device is a physical medical glove. "Standalone performance" in the context of an algorithm is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the device's performance is determined by objective measurements obtained through standardized bench testing methods (e.g., ASTM International standards for physical properties, chemical permeation analysis). For example, a "pinhole" is objectively detected using a water leak test, and "tensile strength" is measured mechanically.

    8. The sample size for the training set

    • N/A. This is a physical device, not an AI/ML model for which a "training set" would be used.

    9. How the ground truth for the training set was established

    • N/A. Not applicable, as there is no training set for a physical product.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, White Color and Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Orange Color are a patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for disposable powder-free nitrile examination gloves, tested for use with chemotherapy drugs. This type of filing is for demonstrating substantial equivalence to a predicate device, not for proving the performance of an AI/ML powered device.

    Therefore, the requested information regarding acceptance criteria and studies for an AI-powered device, including details like sample sizes for test and training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance, is not applicable to this document.

    The document focuses on the physical and chemical properties of the gloves and their resistance to chemotherapy drug permeation, comparing them to established standards and a predicate glove. The "device" in this context is a physical medical glove, not an AI/ML diagnostic or assistive tool.

    However, I can extract the acceptance criteria and reported performance for the glove from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the medical gloves are primarily based on ASTM standards (ASTM D6319-10 updated 2015 for Nitrile Examination Gloves and ASTM D6978-05 for chemotherapy drug permeation) and 21 CFR 800.20 for freedom from pinholes.

    Table: Acceptance Criteria and Reported Device Performance for Disposable Powder Free Nitrile Examination Gloves

    CharacteristicAcceptance CriteriaReported Device Performance (White Color Glove)Reported Device Performance (Orange Color Glove)
    Dimensions:Standard: ASTM D6319-10 (Reapproved 2015)Complies with ASTM D6319-10Complies with ASTM D6319-10
    Length$\ge$ 230 mm$\ge$ 230 mm$\ge$ 230 mm
    Palm Width (X Small)70 $\pm$ 10 mm70 $\pm$ 10 mm70 $\pm$ 10 mm
    Palm Width (Small)80 $\pm$ 10 mm80 $\pm$ 10 mm80 $\pm$ 10 mm
    Palm Width (Medium)95 $\pm$ 10 mm95 $\pm$ 10 mm95 $\pm$ 10 mm
    Palm Width (Large)110 $\pm$ 10 mm110 $\pm$ 10 mm110 $\pm$ 10 mm
    Palm Width (X Large)120 $\pm$ 10 mm120 $\pm$ 10 mm120 $\pm$ 10 mm
    Thickness (Fingertip)$\ge$ 0.05 mm$\ge$ 0.05 mm (minimum, implied by "Complies with ASTM D6319-10")$\ge$ 0.05 mm (minimum, implied by "Complies with ASTM D6319-10")
    Thickness (Palm)$\ge$ 0.05 mm$\ge$ 0.05 mm (minimum, implied by "Complies with ASTM D6319-10")$\ge$ 0.05 mm (minimum, implied by "Complies with ASTM D6319-10")
    Physical Properties:Standard: ASTM D6319-10 (Reapproved 2015)
    Tensile Strength (Before aging)$\ge$ 14 MPa$\ge$ 14 MPa$\ge$ 14 MPa
    Tensile Strength (After aging)$\ge$ 14 MPa$\ge$ 14 MPa$\ge$ 14 MPa
    Elongation (Before aging)$\ge$ 500%$\ge$ 500%$\ge$ 500%
    Elongation (After aging)$\ge$ 400%$\ge$ 400%$\ge$ 400%
    Freedom from PinholeStandard: 21 CFR 800.20 & ASTM D6319-10 (AQL 2.5)Passed Standard Acceptance Criteria (in accordance with ASTM D5151-06)Passed Standard Acceptance Criteria (in accordance with ASTM D5151-06)
    Powder ResidualStandard: ASTM D6319-10 ($\le$ 2 mg/glove)$\le$ 2 mg/glove (in accordance with D6124-06)$\le$ 2 mg/glove (in accordance with D6124-06)
    BiocompatibilityStandards: ISO 10993-10, ISO 10993-5Primary Skin Irritation: Passes (not a primary skin irritant)Primary Skin Irritation: Passes (not a primary skin irritant)
    Dermal Sensitization: Passes (not a primary skin sensitizer)Dermal Sensitization: Passes (not a primary skin sensitizer)
    In vitro cytotoxicity: Passes (not cytotoxic)In vitro cytotoxicity: Passes (not cytotoxic)
    Chemotherapy Drugs Permeation (ASTM D6978-05)Minimum Breakthrough Detection Time (Min)Minimum Breakthrough Detection Time (Min)
    Carmustine (BCNU)Not specified an "acceptance criteria" but reported11.831.6
    ThiotepaNot specified an "acceptance criteria" but reported16.972.5
    Cyclophosphamide (Cytoxan)Not specified an "acceptance criteria" but reported>240>240
    CisplatinNot specified an "acceptance criteria" but reported>240>240
    Doxorubicin HClNot specified an "acceptance criteria" but reported>240>240
    FluorouracilNot specified an "acceptance criteria" but reported>240>240
    Dacarbazine (DTIC)Not specified an "acceptance criteria" but reported>240>240
    Etoposide (Toposar)Not specified an "acceptance criteria" but reported>240>240
    Paclitaxel (Taxol)Not specified an "acceptance criteria" but reported>240>240

    Note: For chemotherapy drugs like Carmustine and Thiotepa, the document explicitly states warnings due to "extremely low permeation time," indicating these drugs do not meet a desired high breakthrough time, hence the warning not to use them with these gloves for prolonged periods. The "acceptance criteria" for these specific drugs seem to be met by simply reporting the actual breakthrough time, coupled with a warning. For the other drugs, >240 minutes indicates they exceeded the maximum testing time, which is generally desired for chemotherapy-resistant gloves.

    Non-Applicable Information for AI/ML Devices:

    The following points are tailored for AI/ML device evaluations and are not relevant to the provided document about medical gloves:

    1. Sample size used for the test set and the data provenance: Not applicable. Tests were bench-top physical/chemical tests on glove samples, not a data-driven AI model.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth was established by laboratory measurements against ASTM standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Lab testing results are objective measurements.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-powered device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. Ground truth for the glove was defined by internationally recognized standards (ASTM, ISO) for physical properties and chemical resistance.
    7. The sample size for the training set: Not applicable. There is no AI model or training set.
    8. How the ground truth for the training set was established: Not applicable. There is no AI model or training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1