510(k) Data Aggregation

    K Number
    K233118

    Validate with FDA (Live)

    Date Cleared
    2025-01-03

    (464 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The radiofrequency energy delivery components of the DUET-V(ESK-3261DV) and Accessories are indicated for use in: - Fractional Tips (FRF Tips): Fractional Tip is intended for use in dermatological procedures requiring ablation and resurfacing of the skin. - Thermal Tips (TRF Tips): Dermatologic and general surgical procedures for electrocoagulation

    Device Description

    DUET-V (Model: ESK-3261DV) is a High Frequency Therapy Device system using 4 MHz high frequency electrical current. Single-use, disposable Treatment Tips attach to the handpiece that comes into contact with the patient during treatment procedures. Treatment tips are classified as direct skin contact devices of limited duration (<24 hours). The RF handpiece has 2 types of fractional tips and 3 types of thermal tips. The tips are disposable and are provided in a non-sterile state. Mono-polar electrodes TRF TIP-S, TRF TIP-R, and TRF TIP-E are used with a grounding pad. Bipolar electrodes FRF TIP-64 and FRF TIP-100, unlike mono-polar electrodes, do not have a ground pad and current flows between electrodes on the contact surface. The tip has three modes: Shot mode, Auto Shot mode, and Continuous mode - Shot Mode this mode delivers the output value set by the user once - Auto Shot Mode In this mode, the set output value delivered continuously according . to the interval time set by the user. (only TRF-TIP-S) - Continuous Mode - this mode continuously delivers RF current until stopped by the user.

    AI/ML Overview

    The provided text describes the regulatory clearance for the DUET-V (Model: ESK-3261DV) electrosurgical device and its accessories. It outlines performance testing, but does not contain information about a study comparing the device's performance against human readers (MRMC study) or a standalone (algorithm only) performance study, nor does it detail the ground truth establishment for a test set in a clinical evaluation context.

    Based on the provided information, here's a summary of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily focused on the device's technical specifications, safety, and functional performance.

    ItemAcceptance CriteriaReported Device Performance
    BiocompatibilityNo risks from biological hazards associated with materials of manufacture and final device.Tests conducted according to ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-10:2020, ISO 10993-23:2021, and ISO 10993-11:2017 to ensure compliance.
    SoftwareVerification and validation according to FDA Guidance "Content of Premarket Submission for Software Contained in Medical Devices" and IEC 62304: 2006 + A1:2015. Cybersecurity compliance to FDA guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions".Firmware was verified and validated according to the specified FDA guidance and IEC 62304. Cybersecurity complied with FDA guidance.
    Electrical Safety and Electromagnetic CompatibilityControl electrical and high temperature hazards; compliance with AAMI/ANSI ES 60601-1: 2005+A1: 2012 and IEC 60601-1-2:2020.Risks controlled by design, protection method, and information according to AAMI/ANSI ES 60601-1. Electromagnetic compatibility verified according to IEC 60601-1-2.
    ReprocessingCleaning and disinfection established and validated according to FDA 'Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling'.Cleaning and disinfection for re-usable parts are established and validated in accordance with FDA guidance.
    Bench Performance - Frequency TestNot exceed the standard value of 4MHz.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Output WaveformConfirm the waveform shape as sine wave.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Power Variation1) Input voltage fluctuates within the range of ±10% of 220V, 2) No abnormalities in each part.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Output Voltage and CurrentBe within ±20% of standard value.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Output ON/OFF TimeBe within ±10% of standard value.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Max. Output Voltage and CurrentBe within ±20% of standard value when voltage connect a 500 Ω load to the output terminal, set the maximum output.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Accuracy of Temperature ControlBe within ±1°C of set value.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Automatic Stop function when temperature riseAutomatically stop at the set temperature.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Switch InterlockBe normally operated of switch interlock when control the foot or hand-piece switch select.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    Bench Performance - Output function upon skin contactOutput comes out when two or more of the four skin contact of FRF tip.Not explicitly stated as "met," but implied by "Performance Testing - Bench."
    System Testing - Thermal Effects on TissueCompliance with FDA guidance 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery' requirements.Test performed according to FDA guidance; results not detailed but described as demonstrating "substantially equivalent performance."
    System Testing - Temperature MonitoringCompliance with FDA guidance 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery' requirements.Test performed according to FDA guidance; results not detailed but described as demonstrating "substantially equivalent performance."
    System Testing - Accuracy of Output control setting (monopolar)Acceptance criteria ±20%.Conducted according to IEC 60601-2-2 201.12.1.101; results not detailed but described as demonstrating "substantially equivalent performance."
    System Testing - Accuracy of Output control setting (bipolar)Acceptance criteria ±20%.Conducted according to IEC 60601-2-2 201.12.1.101; results not detailed but described as demonstrating "substantially equivalent performance."
    System Testing - Monotonicity of Output control setting (monopolar)Acceptance criteria ±20%.Conducted according to IEC 60601-2-2 201.12.1.102; results not detailed but described as demonstrating "substantially equivalent performance."
    System Testing - Monotonicity of Output control setting (bipolar)Acceptance criteria ±20%.Conducted according to IEC 60601-2-2 201.12.1.102; results not detailed but described as demonstrating "substantially equivalent performance."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of a clinical performance evaluation (e.g., patient images or data). The performance testing described is focused on bench testing, electrical safety, biocompatibility, and software validation. Therefore, there is no information provided on sample size or data provenance for a clinical test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    As there is no clinical test set described, there is no information on the number or qualifications of experts for establishing ground truth. The evaluations are technical and regulatory compliance tests.

    4. Adjudication Method for the Test Set

    Again, given the nature of the evaluation (technical and regulatory compliance), there is no mention of an adjudication method for a clinical test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, nor is there any AI component described that would require such a study. The device is an electrosurgical unit, not an AI-powered diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    There is no mention of a standalone algorithm performance study, as the device described is an electrosurgical unit and not an AI algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the described technical and safety performance tests, the "ground truth" is established by engineering standards, regulatory guidelines (e.g., ISO, IEC, FDA guidances), and pre-defined acceptable ranges for physical and electrical parameters. For example, for "Accuracy of Output control setting," the ground truth is the ±20% acceptance criteria of the standard value. The "Thermal Effects on Tissue" and "Temperature Monitoring" tests are performed against the requirements of specific FDA guidances.

    8. The Sample Size for the Training Set

    The concept of a "training set" is relevant for machine learning algorithms. Since this device is an electrosurgical unit and not an AI-driven system, there is no training set described.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set, there is no information on how its "ground truth" was established.