Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K190256

    Validate with FDA (Live)

    Date Cleared
    2019-03-08

    (29 days)

    Product Code
    Regulation Number
    882.4725
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RuloTM Radiofrequency Lesion Probe is an injection probe which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made.

    Device Description

    The Rulo™ Radiofrequency Lesion Probe (Rulo Probe) creates radiofrequency (RF) lesions on nerves and can also create percutaneous nerve blocks by delivering anesthetic to the site. It is a single-use, surgically invasive device. It is inserted with the use of a flexible introducer accessory to contact a nerve in an area of the spine for less than 15 minutes to treat pain. The RF energy creates a lesion on nerve tissue that conducts pain signals. The tissue's ability to conduct electrical signals is disrupted by the lesion. The targeted nerve can be localized either by using RF electrostimulation through the cannula or by injecting contrast medium through the cannula with the use of the fluid injection port and using radiography concomitantly. The nerve may be blocked by injecting local anesthetic solution or making an RF lesion. The device is used by a trained physician in a clinical setting. The device is constructed with a steel cannula and uses a variety of polymers for the insulation, hub, cable, connector, cap and fluid injection port.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Epimed Rulo™ Radiofrequency Lesion Probe. It focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed clinical study for novel performance claims. As such, many of the requested data points (e.g., sample size for test/training sets, number of experts for ground truth, MRMC studies, standalone algorithm performance) are not applicable or available in this type of submission.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and device performance are primarily focused on engineering performance and biocompatibility, demonstrating that the device functions as intended and is safe for use, and is comparable to the predicate device. Specific quantitative performance targets typically seen in AI/diagnostic device studies are not present here because this is a physical medical device.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Mechanical/PhysicalConical fittings (Luer taper) according to ISO 594-1 and ISO 594-2Device meets the requirements of ISO 594-1 and ISO 594-2.
    Sterility, integrity of epidural catheters and introducer needles (BS 6196)Device adheres to applicable sections of BS 6196.
    Performance of hypodermic needles (ISO 7864)Device meets the requirements of ISO 7864.
    Electrical SafetyBasic safety and essential performance of medical electrical equipment (ANSI/AAMI/EN 60601-1)Device complies with applicable sections of ANSI/AAMI/EN 60601-1.
    BiocompatibilityCytotoxicity (ISO 10993-5)Passed/Met necessary requirements.
    Sensitization (ISO 10993-10)Passed/Met necessary requirements.
    Irritation or Intracutaneous Reactivity (ISO 10993-10)Passed/Met necessary requirements.
    Systemic Toxicity (ISO 10993-11)Passed/Met necessary requirements.
    Functional EquivalenceCreation of RF lesions on nerves.Demonstrated comparable performance to the predicate device in creating RF lesions.
    Percutaneous nerve blocks by delivering anesthetic.Demonstrated comparable performance to the predicate device in delivering anesthetic.
    Localization of nerve by electrostimulation or contrast medium with radiography.Device supports these localization methods.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified for individual tests. The nonclinical testing involves laboratory and bench testing of device components and the complete device. This is not a study that uses patient data.
    • Data Provenance: Not applicable in the context of patient data. The nonclinical test data is generated in a laboratory setting or through engineering assessments.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This is not applicable as the submission reports on physical and chemical performance tests, not a diagnostic or AI algorithm that requires expert-established ground truth from images or patient data.

    4. Adjudication Method for the Test Set

    • Not applicable for the types of nonclinical performance and biocompatibility tests performed. These tests have defined pass/fail criteria based on standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • Not applicable. This is a physical medical device, not an AI or diagnostic imaging device that involves human readers or an AI algorithm.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This is a physical medical device and does not involve an algorithm.

    7. The Type of Ground Truth Used

    • For performance tests: The ground truth is defined by the specifications in the referenced international and national standards (e.g., ISO, BS, ANSI/AAMI/EN).
    • For biocompatibility: The ground truth is established by the specified tests and evaluation criteria outlined in ISO 10993 series.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133316

    Validate with FDA (Live)

    Date Cleared
    2014-05-28

    (212 days)

    Product Code
    Regulation Number
    868.5120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spirol® Block Continuous Peripheral Nerve Block Catheter permits placement of catheters next to nerve plexuses for continuous nerve block anesthesia or analgesia techniques including upper extremity, lower extremity, abdominal and paravertebral locations for periods not exceeding 72 hours.

    Device Description

    Epimed's Spirol® Block Continuous Peripheral Nerve Block Catheter has the following characteristics:

    • · 19 Ga. & 20 Ga., 14" to 24" in length.
    • · Offered in open and closed end distal tip configurations.
    • · Offered in styletted and non-styletted versions.
    • · Internal radiopaque, echogenic coiled reinforced wire to allow visibility under fluoroscopy and/or ultrasound.
    • · Catheters to be provided in sterile kit/set style configurations or as a standalone replacement catheter.

    The catheter is a single-use, disposable device consisting of a polymer coated stainless steel spring. The catheter is available in various lengths, as well as open-ended or closed-ended configurations; the closed-ended models having three eyes (injection ports) at the distal end. The Spirol® Block is also available in styletted or non-styletted versions, as well as 19 Ga. or 20 Ga. models.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called the "Spirol® Block Continuous Peripheral Nerve Block Catheter." This type of document focuses on demonstrating substantial equivalence to existing predicate devices rather than proving a device meets specific acceptance criteria through a comparative effectiveness study or standalone performance.

    As such, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, and MRMC studies is not directly applicable to this document because it describes a different type of device approval process.

    Here's why and what information can be extracted:

    • No Acceptance Criteria or Performance Metrics: The 510(k) process for this type of device does not typically involve defining specific numerical "acceptance criteria" and then demonstrating the device "meets" them in the way clinical trials for new drugs or AI algorithms might. Instead, the focus is on showing the new device functions similarly and is as safe and effective as a legally marketed predicate device.
    • No "Study" as such for comparative efficacy: There isn't a "study that proves the device meets acceptance criteria" in the sense of a clinical trial or a performance study with human readers improvement. The "study" here refers to non-clinical testing to confirm technological characteristics and biocompatibility.
    • No AI Component: This device is a physical catheter, not an AI or algorithm-based device. Therefore, questions related to AI-specific metrics (standalone performance, MRMC studies, human reader improvement with AI) are not relevant.

    However, I can extract information about the non-clinical testing performed and the basis for substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Demonstrated Equivalence to Predicate)Reported Device Performance (as demonstrated by testing)
    Technological Characteristics:
    - Single-use, disposable- Polymer coated stainless steel spring (identical to P1)
    - Polymer coated stainless steel spring- Available in various lengths (identical to P1)
    - Available in various lengths- Open-ended or closed-ended configurations (identical to P1)
    - Open-ended or closed-ended configurations- Three eyes (injection ports) at distal end for closed-ended (identical to P1)
    - Three eyes (injection ports) at distal end- Styletted or non-styletted versions (identical to P1)
    - Styletted or non-styletted versions- 19 Ga. or 20 Ga. models (identical to P1)
    - 19 Ga. or 20 Ga. models- Insertion by trained physician with introducer needle (identical to P1)
    - Catheter placement remains after needle withdrawal- Connects to standard Luer-lock accessories (identical to P1)
    - Conveys anesthetic agents
    Functional/Material Performance:
    - Resists kinking, collapsing, shearing- Polymer coated stainless steel spring proven to resist (bench testing)
    - Withstands minimum acceptable tensile force- Proven to withstand (bench testing)
    - Materials compatible with PNB drugs- Tested for compatibility (bench testing)
    - Injection ports remain patent- Designed to remain patent (bench testing)
    - Deliver injectants at sufficient flow rate- Designed to deliver (bench testing)
    - Catheter and connector compatibility- Tested and proven compatible (bench testing)
    Biocompatibility (per ISO 10993):
    - Cytotoxicity- Meets necessary biocompatibility requirements
    - Sensitization- Meets necessary biocompatibility requirements
    - Irritation or Intracutaneous Reactivity- Meets necessary biocompatibility requirements
    - Acute Systemic Toxicity (catheter only)- Meets necessary biocompatibility requirements
    - Sub-acute / Sub-chronic Toxicity (catheter only)- Meets necessary biocompatibility requirements
    - Genotoxicity (catheter only)- Meets necessary biocompatibility requirements
    - Implantation (catheter only)- Meets necessary biocompatibility requirements
    Non-Clinical Testing:
    - Dimensional- Performed, results demonstrate comparability
    - Substance compatibility- Performed, results demonstrate comparability
    - Kink testing- Performed, results demonstrate comparability
    - Stiffness- Performed, results demonstrate comparability
    - Tensile strength- Performed, results demonstrate comparability
    - Connector attachment- Performed, results demonstrate comparability
    - Flow test- Performed, results demonstrate comparability
    - Resistance to leakage during aspiration or vacuum- Performed, results demonstrate comparability
    - Resistance to leakage under pressure- Performed, results demonstrate comparability
    - Corrosion resistance- Performed, results demonstrate comparability
    - Skive testing- Performed, results demonstrate comparability

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any of the non-clinical tests. Medical device bench testing often uses small, representative samples, but specific numbers are not provided in this summary.
    • Data Provenance: The tests were performed internally by Epimed International, Inc. or by external labs for specific ISO 10993 testing. No country of origin for data is specified beyond the company's location in the USA. All testing described is pre-clinical/bench testing, not human clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable. "Ground truth" in the context of expert consensus is typically for diagnostic interpretation tasks (e.g., radiology). For a physical medical device, "ground truth" refers to established engineering specifications, material standards (e.g., ISO 10993), and regulatory requirements. These were assessed through the non-clinical testing listed. While engineers and quality assurance personnel ("experts") would design and oversee these tests, their number and specific qualifications for "establishing ground truth" are not detailed.

    4. Adjudication method for the test set:

    • Not applicable. This concept applies to expert review of clinical cases. The tests described are objective, physical, and chemical tests against defined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical catheter, not an AI or imaging interpretation device, and no human reader studies were performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical catheter, not an algorithm.

    7. The type of ground truth used:

    • For technological characteristics and functional performance: Engineering specifications, design inputs, and established performance of the predicate devices.
    • For biocompatibility: ISO 10993 series of standards.
    • For non-clinical testing (dimensional, strength, flow, etc.): Industry standards, internal specifications, and performance data from predicate devices.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not an AI/machine learning device. The design validation and non-clinical testing confirm the device meets its intended specifications.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.

    In summary, the document demonstrates substantial equivalence through:

    • Comparing the Intended Use and Indications For Use to Predicate Device #2 (K122027).
    • Comparing the design, manufacture, and material composition to Predicate Device #1 (K981329).
    • Presenting results from a series of non-clinical bench and laboratory performance tests (dimensional, substance compatibility, kink testing, stiffness, tensile strength, connector attachment, flow test, resistance to leakage, corrosion resistance, and skive testing) to show comparable performance to the predicate devices.
    • Confirming biocompatibility according to ISO 10993 standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K053318

    Validate with FDA (Live)

    Date Cleared
    2006-02-17

    (79 days)

    Product Code
    Regulation Number
    868.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RX Epidural Needle is intended for the epidural placement, directly or through an epidural catheter, of anesthetic agents to elicit regional anesthesia.

    Device Description

    The RX Epidural Needle consists of a stainless steel cannula with a ground beveled distal tip. A plastic hub is molded onto the proximal end of the cannula. A stylet is also provided with the device which consists of a custom trimmed shaft and a molded plastic hub. The RX Epidural Needle will be provided as a sterile, single use, disposable device. The RX Epidural Needle will be available in a variety of lengths and gauges.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the "RX Epidural Needle" and the FDA's response letter. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with specific acceptance criteria and detailed performance data on a pre-defined test set.

    Therefore, many of the requested categories for describing acceptance criteria and study details cannot be filled from the given text. This type of submission relies on the established safety and effectiveness of a predicate device, not on a new clinical or performance study with defined acceptance criteria for a novel device.

    Here's a breakdown of what can and cannot be answered based on the provided document:


    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided from the text. The document does not describe specific acceptance criteria (e.g., in terms of sensitivity, specificity, accuracy, or other performance metrics) because it's a 510(k) submission for substantial equivalence. It compares the device to a predicate based on "similar physical and technical characteristics" rather than new performance data against numerical targets.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Cannot be provided from the text. No specific test set or associated data provenance is mentioned as this is a substantial equivalence submission, not a de novo study with a test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Cannot be provided from the text. Ground truth establishment is not described as there is no specific "test set" and corresponding study in the provided text.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Cannot be provided from the text. No adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable / Cannot be provided from the text. This device is an epidermal needle, not an AI-assisted diagnostic tool. Therefore, an MRMC study related to AI assistance would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable / Cannot be provided from the text. This device is an epidermal needle, not an algorithm, so standalone performance in that context is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable / Cannot be provided from the text. No ground truth is described.

    8. The sample size for the training set

    • Not applicable / Cannot be provided from the text. No training set is mentioned.

    9. How the ground truth for the training set was established

    • Not applicable / Cannot be provided from the text. No training set or ground truth establishment is mentioned.

    Summary based on the document:

    The provided document describes a 510(k) summary for a medical device. The core of a 510(k) submission is to demonstrate substantial equivalence to a predicate, legally marketed device. This means the device is as safe and effective as a legally marketed device and does not require a new Premarket Approval (PMA) application.

    In this specific case, the RX Epidural Needle is deemed substantially equivalent to the Epimed International Epidural Needle (K030562). The "study" (or rather, the basis for approval) is this comparison.

    • Acceptance Criteria: Implicitly, the acceptance criterion for a 510(k) is demonstrating that the new device has the same intended use, similar technological characteristics, and raises no new questions of safety or effectiveness compared to the predicate device.
    • Reported Device Performance: The document states, "The RX Epidural Needle has similar physical and technical characteristics to the Epimed International Epidural Needle marketed under K030562." This is the reported "performance" in the context of this type of submission. Specific numerical performance metrics were not required or provided in the summary.
    • Study Design: This was a comparison study based on device characteristics, not a clinical trial or performance study with a test set of data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K051171

    Validate with FDA (Live)

    Date Cleared
    2005-08-26

    (112 days)

    Product Code
    Regulation Number
    868.5140
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stingray Epidural Catheter Connector is intended to provide various anesthetic and fluid administration devices with a single, common access point to an epidural catheter for delivery of anesthetics. The connector is used in conjunction with an epidural catheter for continuous administration of anesthetic agents.

    Device Description

    The Stingray Epidural Catheter Connector consists of two molded plastic body Sections that mate in a twist and lock motion. Between the two Sections is a molded bushing that compresses and grips an epidural catheter.

    AI/ML Overview

    This document, a 510(k) summary for the Stingray Epidural Catheter Connector, does not contain the detailed information required to describe the acceptance criteria and the study that proves the device meets those criteria.

    The purpose of a 510(k) summary is to demonstrate substantial equivalence to a predicate device, not necessarily to present a de novo study with acceptance criteria and performance data in the format requested. While it mentions "The testing performed and comparison to the predicate device demonstrates that the Stingray Epidural Catheter is safe and effective," it does not elaborate on what that testing entailed, its methodology, or its results.

    Therefore, I cannot provide the requested information for the following points based on the provided text:

    1. A table of acceptance criteria and the reported device performance: Not present. The document states the device is "safe and effective" and "substantially equivalent" but provides no specific performance metrics or acceptance thresholds.
    2. Sample size used for the test set and the data provenance: Not present. No information about specific test sets, their size, or origin (country, retrospective/prospective) is provided.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/present. This type of information is typically for studies involving human interpretation (e.g., image analysis), which is not the primary focus of this device's 510(k) submission. No expert review for ground truth is mentioned.
    4. Adjudication method: Not applicable/present.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a physical connector, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not explicitly stated. The "testing performed" likely referred to bench testing or engineering evaluations to ensure physical and functional equivalence to the predicate, rather than ground truth in the context of diagnostic performance.
    8. The sample size for the training set: Not applicable. This device is not an AI/machine learning model that would require a training set.
    9. How the ground truth for the training set was established: Not applicable.

    What the document does provide:

    • Device Name: Stingray Epidural Catheter Connector
    • Intended Use: A connection device to provide various anesthetic and fluid administration to an access point to an epidural catheter for delivery of anesthetics, used in conjunction with an epidural catheter for continuous administration of anesthetic agents.
    • Predicate Device: B. Braun Medical Perifix® Catheter Connector (K032144)
    • Conclusion: The device is "safe and effective and is substantially equivalent to the predicate device" based on "testing performed and comparison to the predicate device."

    To obtain the detailed information requested, one would typically need to review the full 510(k) submission, which would include detailed test reports, protocols, and results. This summary, by its nature, is an abbreviated version.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051860

    Validate with FDA (Live)

    Date Cleared
    2005-08-22

    (45 days)

    Product Code
    Regulation Number
    868.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flexible Introducer Cannula is intended for the epidural placement, directly The Flexible Introducer cannuid is Inventour of the opents to elicit regional anesthesia.

    Device Description

    The Flexible Introducer Cannula consists of a plastic cannula with a molded plastic hub. The device is also packaged with a stainless steel introducer needle which is removed after the device has been placed.

    AI/ML Overview

    This document is a 510(k) summary for a medical device called the "Flexible Introducer Cannula." It details the device's characteristics, its intended use, and its comparison to a predicate device to establish substantial equivalence.

    Based on the provided text, there is no information available about acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy) for a study involving the device's performance, nor a study proving such criteria are met.

    The document is a regulatory submission focused on establishing substantial equivalence for a physical medical device (anesthesia conduction needle), not a software or AI/ML-driven device that would typically have performance metrics and acceptance criteria for classification or detection tasks.

    Here's a breakdown of why this information is not present and what is discussed instead:

    • Acceptance Criteria and Reported Device Performance: This device is a physical instrument. The "acceptance criteria" here are related to physical and technical characteristics and safety/effectiveness as compared to a predicate device, rather than statistical performance metrics for a diagnostic or AI system. The document states:

      • "The Flexible Introducer Cannula has similar physical and technical characteristics to the predicate device and a similar intended use to the predicate device."
      • "The testing performed and comparison to the predicate device demonstrates that the Flexible Introducer Cannula is safe and effective and is substantially equivalent to the predicate device."
      • No specific quantitative performance metrics (e.g., sensitivity, specificity, accuracy) are reported.
    • Sample size for the test set and data provenance: Not applicable. There is no mention of a "test set" in the context of diagnostic performance data. The "testing performed" would likely refer to engineering validation and verification tests (e.g., materials testing, sterility, functionality) and comparison of physical attributes to the predicate.

    • Number of experts and their qualifications for ground truth: Not applicable. Ground truth as typically defined for AI/ML performance studies is not relevant here. The "ground truth" for this device would be its physical conformance to specifications and its biological safety, established through standard engineering and biocompatibility testing.

    • Adjudication method: Not applicable.

    • MRMC comparative effectiveness study: Not applicable. There is no mention of human readers or AI assistance.

    • Standalone performance study: Not applicable in the context of AI/ML. All the documentation describes the device's standalone characteristics and function.

    • Type of ground truth used: For a physical medical device like this, "ground truth" means demonstrating the device meets design specifications, material properties, biocompatibility, and functional requirements. This is typically established through engineering tests, material science analysis, and preclinical studies where relevant. The summary provided does not detail these specific tests, but rather states that "testing was performed."

    • Sample size for the training set: Not applicable. This is not an AI/ML device.

    • How ground truth for the training set was established: Not applicable.

    In summary, the provided document is a 510(k) summary for a traditional medical device (anesthesia conduction needle). It establishes substantial equivalence to a predicate device through comparison of physical and technical characteristics and intended use, rather than through performance metrics typically associated with AI/ML or diagnostic devices. Therefore, the specific questions regarding acceptance criteria, study details, and ground truth establishment in the context of AI/ML or diagnostic performance are not addressed by this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043467

    Validate with FDA (Live)

    Date Cleared
    2005-03-31

    (106 days)

    Product Code
    Regulation Number
    868.5150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flexible Spinal Needle is intended for the administration of anesthetic agents to provide regional anesthesia.

    Device Description

    The Flexible Spinal Needle consists of a coated flexible stainless steel cannula with a molded plastic hub. A stylet is also provided with the device which consists of a stainless steel cannula and a molded plastic hub. The Flexible Spinal Needle will be provided as a sterile, single use, disposable device and will be available in a variety of lengths and gauges.

    AI/ML Overview

    This document is a 510(k) summary and FDA clearance letter for a Flexible Spinal Needle. It describes the device and its intended use, and compares it to predicate devices. However, the document does not contain the detailed information necessary to answer the questions about acceptance criteria and a study proving the device meets those criteria.

    Here's why and what information is missing:

    • Type of device: This is a medical device (a spinal needle), not a software or AI-driven diagnostic device. The questions about "algorithm only without human-in-the loop performance," "effect size of how much human readers improve with AI," and "sample size for the training set" are specific to AI/ML or diagnostic software and are not applicable to a physical medical instrument like this.
    • Study Design: The document states "Bench Testing performed on the Flexible Spinal Needle to compare performance characteristics to the predicate device(s) confirmed that the performance of the Flexible Spinal Needle is similar to the predicate device(s)." This is a very high-level summary and doesn't provide the specifics requested in your prompt.

    Therefore, most of the requested information cannot be extracted from the provided text.

    Information that can be inferred or stated as not available:

    1. Table of acceptance criteria and reported device performance: Not provided. The document generally states "performance... is similar to the predicate device(s)," but no specific criteria or performance metrics are listed.
    2. Sample size used for the test set and data provenance:
      • Sample size: Not specified. "Bench Testing" is mentioned.
      • Data provenance: Not specified. Assumed to be lab/bench test data, not patient data in the typical sense requested by "country of origin" or "retrospective/prospective."
    3. Number of experts and qualifications: Not applicable/Provided. Bench testing doesn't typically involve human experts in this context.
    4. Adjudication method: Not applicable/Provided.
    5. Multi Reader Multi Case (MRMC) comparative effectiveness study: Not conducted/applicable as this is not an AI/ML diagnostic system.
    6. Standalone (algorithm only) performance: Not applicable. This is a physical device.
    7. Type of ground truth used: Not explicitly stated for bench testing, but likely derived from engineering specifications and predicate device performance standards.
    8. Sample size for the training set: Not applicable. There is no "training set" for a physical device undergoing bench testing for 510(k) clearance in this manner.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document describes a 510(k) clearance for a physical medical device (spinal needle) based on substantial equivalence to predicate devices through bench testing. It does not provide the detailed study design, acceptance criteria, or performance metrics typically found in documentation for AI/ML or diagnostic software evaluations, as implied by most of your detailed questions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041021

    Validate with FDA (Live)

    Date Cleared
    2004-09-16

    (149 days)

    Product Code
    Regulation Number
    882.4725
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Epimed RF Introduction Cannula is intended for use in radiofrequency (RF) heat lesion procedures for the relief of pain.

    Device Description

    The Epimed RF Introduction Cannula consists of a coated stainless steel cannula with a stainless steel stylet and molded plastic hub. The cannula will be available in various lengths, gauges and tip configurations.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "Epimed RF Introduction Cannula," and as such, it does not detail acceptance criteria and a study proving device performance in the way a clinical trial or AI device study would. Instead, it demonstrates substantial equivalence to predicate devices. Therefore, many of the requested categories related to AI device studies (e.g., sample size for test/training sets, expert ground truth, MRMC studies, standalone performance) are not applicable to this type of regulatory submission.

    Here's an analysis based on the provided text, focusing on what is available:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Safety and EffectivenessDemonstrated by "Bench Testing performed on the Epimed RF Introduction cannula to compare performance characteristics to the predicate device(s) confirmed that the performance of the RF Introduction Cannula is similar to the predicate device(s)." This similarity to legally marketed predicate devices (K980430, unknown K# for Cotop) implies meeting existing safety and effectiveness standards for RF cannulae.
    Similar Physical and Technical CharacteristicsThe device "has similar physical and technical characteristics to the predicate device(s)."
    Similar Intended UseThe device "has a similar intended use to the predicate device(s)," which is "for use in radiofrequency (RF) heat lesion procedures for the relief of pain."

    2. Sample size used for the test set and the data provenance

    • Not Applicable. This document describes a bench testing comparison to predicate devices, not a clinical study with a "test set" in the context of AI or clinical efficacy. The "test set" here refers to the Epimed RF Introduction Cannula itself, which was subjected to bench tests.
    • Data Provenance: The data comes from "Bench Testing performed on the Epimed RF Introduction cannula." No information on country of origin or retrospective/prospective nature is given beyond being bench tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. Ground truth in the context of expert consensus or clinical outcomes is not relevant for this type of regulatory submission, which relies on bench testing comparisons for substantial equivalence. The "ground truth" for the bench testing would be the measured physical and technical characteristics of the device and its predicate.

    4. Adjudication method for the test set

    • Not Applicable. No explicit adjudication method is mentioned. The comparison seems to be direct measurements of physical/technical characteristics against those of predicate devices.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, not done. This is not an AI device, and therefore, an MRMC study comparing human readers with and without AI assistance is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device (cannula), not an algorithm or AI system.

    7. The type of ground truth used

    • For the non-clinical data, the ground truth was the measured physical and technical characteristics of the Epimed RF Introduction Cannula and the referenced predicate devices, assessed through bench testing.

    8. The sample size for the training set

    • Not Applicable. This is not an AI device with a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI device with a training set.

    Summary of the Study (Bench Testing):

    The study performed was Non-Clinical Bench Testing. This testing aimed to compare the performance characteristics of the Epimed RF Introduction Cannula to its predicate devices. The conclusion from this bench testing was that the performance of the Epimed RF Introduction Cannula is "similar to the predicate device(s)," leading to the determination of substantial equivalence in terms of safety and effectiveness. The specific parameters measured during these bench tests are not detailed in this summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041843

    Validate with FDA (Live)

    Date Cleared
    2004-08-24

    (47 days)

    Product Code
    Regulation Number
    868.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blunt Nerve Block Needle is intended for the administration of anesthetic agents to provide regional anesthesia.

    Device Description

    The Blunt Nerve Block Needle consists of a stainless steel cannula with an atraumatic distal tip. A plastic hub is molded onto the proximal end of the cannula. A stylet is also provided with the device which consists of a stainless steel wire shaft and a molded plastic hub. The Blunt Nerve Block Needle will be provided as a sterile, single use, disposable device and will be available in a variety of lengths and gauges.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study.

    Crucially, this document is a 510(k) summary for a "Blunt Nerve Block Needle." Medical devices of this type typically do not have performance acceptance criteria in the same way as AI/software devices. Therefore, many of the requested elements for an AI product's study will not be present. Instead, the submission focuses on demonstrating substantial equivalence to a predicate device.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Physical Characteristics: Similar to predicate Pajunk Sprotte Needles.The Blunt Nerve Block Needle has similar physical and technical characteristics to the Pajunk Sprotte Needles.
    Intended Use: For the administration of anesthetic agents to provide regional anesthesia. Not for use with an Epidural Catheter.The Blunt Nerve Block Needle is intended for the administration of anesthetic agents to provide regional anesthesia. (Matches intended use of predicate according to the submission.)
    Safety and Effectiveness: Demonstrated through substantial equivalence to predicate.Conclusion: The comparison to the predicate device demonstrates that the Blunt Nerve Block Needle is safe and effective and is substantially equivalent to the predicate device(s).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable. This submission is for a physical medical device (a needle), not a diagnostic AI or imaging device that would typically have a "test set" of data. The submission relies on a comparison to a predicate device.
    • Data Provenance: Not applicable. The "study" isn't a data-driven performance study in the AI sense, but rather a regulatory comparison.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Number of Experts: Not applicable.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done: No. This is not an AI/software device, so an MRMC study is not relevant or included.
    • Effect Size of Human Reader Improvement with AI: Not applicable.

    6. Standalone (Algorithm Only) Performance Study

    • Standalone Study Done: No. This is a physical device, not an algorithm.

    7. Type of Ground Truth Used

    • Type of Ground Truth: For this submission, the "ground truth" equivalent is the established safety and effectiveness of the predicate devices (Pajunk Sprotte Needles). The Epimed needle is deemed safe and effective because it is substantially equivalent to these already-approved devices, which are presumed to have established their "truth" through prior regulatory pathways and clinical use.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. There is no concept of a "training set" for this type of medical device submission.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth was Established: Not applicable.

    Summary of the "Study" (or lack thereof, for AI purposes):

    The submission for the Epimed International, Inc. Blunt Nerve Block Needle is a 510(k) Premarket Notification based on substantial equivalence to predicate devices. It does not involve a study with acceptance criteria, test sets, or ground truth in the way an AI/software medical device would.

    Instead, the "study" is a non-clinical comparison that asserts:

    • The device has similar physical and technical characteristics to the Pajunk Sprotte Needles (K911260, K911221, K911202).
    • The device has the same intended use as the predicate devices: "For the administration of anesthetic agents to provide regional anesthesia. Not for use with an Epidural Catheter."
    • No bench testing of performance characteristics was conducted as the submission relies solely on this similarity and intended use to the predicates.

    The conclusion is that because it is sufficiently similar to an already legally marketed device with a known safety and effectiveness profile, the new device is also considered safe and effective, and therefore, substantially equivalent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023140

    Validate with FDA (Live)

    Date Cleared
    2003-08-15

    (329 days)

    Product Code
    Regulation Number
    868.5120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For administration of anesthetic agents into the epidural space to provide continuous epidural or caudal anesthesia for up to 72 hours.

    Device Description

    The Versa-Kath Epidural Catheter consists of a stainless steel spring lumen with a dual coating. The inner coating is Polyester tube and the outer coating is Fluorinated Ethylene Polypropylene tube. A stylet is also provided with the device which consists of a stainless steel wire and a molded plastic hub. The Versa-Kath will be provided as a sterile, single use, disposable device. The Versa-Kath will be a 21 gauge catheter available in various lengths.

    AI/ML Overview

    Here's an analysis of the provided document regarding the acceptance criteria and study for the Versa-Kath Epidural Catheter:

    The document does not describe an AI/ML medical device. It pertains to a physical medical device, the Versa-Kath Epidural Catheter, which is anesthesia conduction catheter. Therefore, many of the typical questions for AI/ML device studies (like sample sizes for test/training sets, ground truth establishment methods, expert qualifications, adjudication, MRMC studies, standalone performance, etc.) are not applicable in this context.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical (bench) testing.

    Here's the information that can be extracted relevant to acceptance criteria and the study, as far as it applies to a non-AI/ML device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    Tensile StrengthSimilar to predicate device (Racz Epidural Catheter K954584)Confirmed similar to predicate device
    FlowSimilar to predicate device (Racz Epidural Catheter K954584)Confirmed similar to predicate device
    Deflection (Stiffness)Conformance to BS 6196Confirmed conformance to BS 6196 and similar to predicate device
    Fluid Pressure (Catheter/Adapter Connection)Similar to predicate device (Racz Epidural Catheter K954584)Confirmed similar to predicate device
    Luminal IntegritySimilar to predicate device (Racz Epidural Catheter K954584)Confirmed similar to predicate device

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the bench testing. It refers to "Bench Testing performed," implying that a sufficient number of samples were tested to support the conclusions. As this is bench testing of a physical device, data provenance (country of origin, retrospective/prospective) is not relevant in the same way it would be for clinical data or AI data. The tests were performed in a lab setting by the manufacturer, Epimed International, Inc.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    N/A. This is a physical device. "Ground truth" in this context would be objective physical measurements from testing, not expert consensus on medical images or clinical data.

    4. Adjudication Method for the Test Set

    N/A. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert labeling or diagnoses, which is not applicable here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is not an AI device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This is not an AI device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through objective engineering measurements and adherence to recognized standards (like BS 6196) during bench testing, and comparison to the characteristics of a legally marketed predicate device.

    8. The Sample Size for the Training Set

    N/A. This is not an AI device. There is no concept of a "training set" for this type of device.

    9. How the Ground Truth for the Training Set was Established

    N/A. This is not an AI device.


    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The study was a non-clinical bench testing evaluation designed to demonstrate that the Versa-Kath Epidural Catheter has substantially equivalent physical and technical characteristics and performance to its identified predicate device, the Racz Epidural Catheter (K954584).

    Key aspects of the study:

    • Type of Study: Bench testing (non-clinical).
    • Tests Performed:
      • Tensile strength
      • Flow
      • Deflection (stiffness)
      • Fluid Pressure of the catheter/adapter connection
      • Luminal integrity
    • Methodology: The Versa-Kath Epidural Catheter was tested and its performance characteristics were directly compared to those of the predicate device. For deflection (stiffness), conformance to standard BS 6196 was also evaluated.
    • Conclusion: The manufacturer concluded that the testing confirmed the Versa-Kath's performance is "similar to that of the device [predicate]" and demonstrated substantial equivalence, leading to its determination as "safe and effective."

    The FDA's decision to clear the device (K023140) signifies their concurrence with the manufacturer's conclusion that the non-clinical data supported the substantial equivalence claim.

    Ask a Question

    Ask a specific question about this device

    K Number
    K030562

    Validate with FDA (Live)

    Date Cleared
    2003-04-03

    (41 days)

    Product Code
    Regulation Number
    868.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Epidural Needle is intended for the epidural placement, directly or through an epidural catheter, of anesthetic agents to elicit regional anesthesia.

    Device Description

    The Tuohy Epidural Needle consists of a stainless steel cannula with a ground beveled distal tip. A plastic hub is molded onto the proximal end of the cannula. A stylet is also provided with the device which consists of a stainless steel wire shaft and a molded plastic hub. The Tuohy Epidural Needle will be provided as a sterile, single use, disposable device. The Epidural Needle will be available in a variety of lengths and gauges.

    AI/ML Overview

    The provided document, K030562 pertaining to the Tuohy Epidural Needle, does not contain information related to acceptance criteria or a study proving device performance in the context of an AI/algorithm-driven medical device.

    This document is a 510(k) summary for a traditional medical device (an epidural needle) and focuses on establishing substantial equivalence to a predicate device. Therefore, none of the requested information regarding AI/algorithm performance, ground truth, sample sizes for training/test sets, expert adjudication methods, or MRMC studies is present or applicable.

    Here's a breakdown of why the information is not present based on the provided text:

    • Device Type: The device is a "Tuohy Epidural Needle," a physical medical instrument, not an AI or algorithm.
    • Regulatory Pathway: The submission is a 510(k) for substantial equivalence, which primarily involves comparing the new device to a legally marketed predicate device, rather than demonstrating de novo performance against specific criteria through a clinical or algorithmic study.
    • Non-Clinical Data Section: The document explicitly states: "Due to the fact that this product is purchased by Epimed from Manan Medical and is identical to the predicate device, bench testing to compare performance characteristics was not conducted." This confirms that no specific performance studies were performed for this particular submission.
    • Conclusion: The conclusion focuses solely on substantial equivalence to the predicate device, not on meeting predefined performance criteria through a study.

    Therefore, I cannot provide the requested information from the given text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2