510(k) Data Aggregation

    K Number
    K230111

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-26

    (129 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CORE 500 Digital Stethoscope is intended to be used by clinicians to electronically amplify, filter, and transfer body sounds and three lead electrocardiogram (ECG) waveforms. The CORE 500 Digital Stethoscope also displays ECG waveforms and heart rate on the display and accompanying mobile application (when prescribed or used under the care of a clinician).

    The data offered by the device is only significant when used in conjunction with clinician evaluation as well as consideration of other relevant patient data.

    Device Description

    CORE 500 Digital Stethoscope (CORE 500) is an electronic stethoscope with integrated electrodes for electrocardiogram (ECG). The device consists of a chestpiece, detachable earpiece (Eko Earpiece) and a mobile application (Eko App) and is intended as a digital auscultation tool on patients requiring physical assessment by the health care providers. CORE 500 provides the ability to amplify, filter, and transfer body sounds with the chestpiece diaphragm, and three lead ECG through electrodes integrated around the chestpiece.

    CORE 500 features three auscultation modes for better auscultation experience by filtering acoustic data and enhancing the primary frequency range of particular body sounds: Cardiac Mode for heart sounds. Pulmonary Mode for lung sounds, and Wide Band Mode for general auscultation. CORE 500 also detects and computes the heart rate in real-time based on the phonocardiogram (PCG) data.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided FDA 510(k) summary for the Eko CORE 500 Digital Stethoscope:

    Note: The provided document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It does not detail a clinical study with specific acceptance criteria related to diagnostic performance involving human subjects and ground truth established by experts. The "acceptance criteria" discussed below are based on the non-clinical performance data provided.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for an electronic stethoscope, the "acceptance criteria" are derived from the performance data provided to show equivalence and functionality. The document doesn't explicitly list pass/fail criteria with numerical thresholds in the same way a clinical trial might, but it states that "the CORE 500 Digital Stethoscope complies with" or "demonstrated compliance" with various standards and performance benchmarks.

    Acceptance Criterion TypeDescription of Criterion (Implicit)Reported Device Performance
    BiocompatibilityDevice materials in contact with the body must be biocompatible.Complies with ISO 10993-1:2018. The evaluation report concluded that the device is biocompatible.
    Electrical SafetyDevice must meet electrical safety standards.Complies with IEC 60601-1.
    EMC (Electromagnetic Compatibility)Device must meet electromagnetic compatibility standards.Complies with IEC 60601-1-2.
    Software Verification & ValidationSoftware must be verified and validated.Verified and validated according to FDA guidance.
    Bench Testing (General Performance)Differences between the subject and predicate devices do not raise new questions of safety and effectiveness.Rigorous bench testing conducted to demonstrate product performance.
    Audio PerformanceAcoustic performance (amplification, filtering) must be adequate for intended use.Testing conducted to verify audio performance. (Specific metrics not detailed in summary)
    Electrical & Mechanical FunctionElectrical and mechanical functions (e.g., buttons, display, connectivity) must operate as intended.Testing conducted to verify electrical and mechanical function. (Specific metrics not detailed in summary)
    Heart Rate MeasurementHeart rate detection must be accurate based on PCG data.Testing conducted to verify heart rate measurement. (Specific metrics not detailed in summary)
    ECG Frequency RangeWhile different from predicate (0.1-250 Hz vs 0.15-200 Hz), the wider range should not raise new safety/effectiveness questions.Deemed acceptable as it "does not raise different questions of safety and effectiveness."
    Number of ECG ElectrodesWhile different from predicate (3 dry electrodes vs 2 dry electrodes), the change should not raise new safety/effectiveness questions.Deemed acceptable as it "does not raise different questions of safety and effectiveness."
    Hardware InterfaceWhile different from predicate (additional display, capacitive touch), the added features should not raise new safety/effectiveness questions.Deemed acceptable as the "additional interfaces do not raise different questions of safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical bench testing rather than a clinical study with a "test set" of patient data. Therefore, there is no patient sample size or provenance information in the sense of a clinical trial (e.g., country of origin, retrospective/prospective). The testing involved physical devices and simulated or controlled environments to assess performance properties.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    As this was non-clinical bench testing, no medical experts were explicitly used to establish "ground truth" for a patient test set. The tests focused on objective electrical, mechanical, and software performance criteria verified against technical standards and internal specifications, not diagnostic accuracy in a clinical context.


    4. Adjudication Method for the Test Set

    Since there was no patient test set requiring expert interpretation or diagnosis, there was no adjudication method (like 2+1 or 3+1) used.


    5. If a Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document describes non-clinical performance data for substantial equivalence, not a study assessing how human readers improve with or without AI assistance.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The document pertains to the CORE 500 Digital Stethoscope hardware, which amplifies, filters, and transfers body sounds and ECG waveforms, and displays ECG waveforms and heart rate. While the device processes signals, it's a diagnostic tool, and the indications for use explicitly state: "The data offered by the device is only significant when used in conjunction with clinician evaluation as well as consideration of other relevant patient data." This indicates that the device is intended for human-in-the-loop use. Therefore, a standalone algorithm-only performance assessment in a diagnostic context was not the focus of this submission. The "heart rate detection" is a standalone function of the device, but its diagnostic interpretation is with a clinician.


    7. The Type of Ground Truth Used

    For the non-clinical tests described:

    • Biocompatibility: Ground truth is established by adherence to ISO 10993-1:2018 standards and laboratory testing results.
    • Electrical Safety & EMC: Ground truth is established by compliance with IEC 60601-1 and IEC 60601-1-2 standards.
    • Software V&V: Ground truth is established by meeting FDA Guidance for Premarket Submissions for Software and internal software requirements.
    • Bench Testing (Audio, Electrical/Mechanical, Heart Rate): Ground truth is based on engineering specifications, established physical principles, and comparison to calibrated reference instruments/signals.

    8. The Sample Size for the Training Set

    The document describes premarket notification for a hardware device (digital stethoscope) with integrated capabilities. It does not mention machine learning or AI algorithms requiring a "training set" in the context of diagnostic interpretation (e.g., for automated murmur detection or arrhythmia classification). While heart rate detection is mentioned, the details of its underlying algorithm training are not provided. No specific "training set" size is part of this 510(k) summary.


    9. How the Ground Truth for the Training Set Was Established

    Since no explicit "training set" for a diagnostic AI algorithm is described as part of this submission, the method for establishing its ground truth is not applicable here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213794

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-06-29

    (205 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Eko Murmur Analysis Software (EMAS) is intended to provide decision support to clinicians in their evaluation of patients' heart sounds. The software analyzes heart sounds and phonocardiograms (and ECG signals, when available). The software will automatically detect murmurs that may be present, and the murmur timing and character, including S1, S2, innocent heart murmurs, structural heart murmurs, and the absence of a heart murmur.

    The Eko Murmur Analysis Software is not intended as a sole means of diagnosis and is for use in environments where health care is provided by clinicians. The interpretations of heart sounds offered by the software are meant only to provide decision support to the clinician, who may use the result in conjunction with their own evaluation and clinical judgment. The interpretations are not diagnoses. The Eko Murmur Analysis Software is intended for use on pediatric and adult patients.

    Device Description

    Eko Murmur Analysis Software (EMAS) is a cloud-based service that allows users to upload heart sound/phonocardiogram (PCG) and optional electrocardiogram (ECG) data via an application programming interface (API) for analysis. The software uses signal processing (such as waveform filtering), as well as algorithms derived from machine learning, to analyze the acquired data and generate clinical decision support output for clinicians. EMAS is designed to evaluate data derived by the company's two previously cleared devices, the Eko DUO (K170874) and Eko CORE (K151319, K200776). The heart sound data from those devices can be transmitted to the Eko Cloud using either the Eko mobile application or thirdparty applications that use a software development kit (SDK). The EMAS algorithm analyzes the heart sound data and outputs a JSON file with the algorithm results, which is passed down to the requesting application and displayed by the requesting application to the user in the humanreadable format.

    The analysis will assess the signal quality of the phonocardiogram; detect heart murmurs and classify them as innocent or structural; determine the timing of S1 and S2 heart sounds; and distinguish between systolic and diastolic heart murmurs. As an integral part of a physical assessment, clinicians' interpretations of EMAS' output can help them rule in or out different pathological conditions in a patient.

    The EMAS consists of the following algorithm components:

    • Signal Quality Detection Algorithm:
      This pre-processing algorithm accepts as input the PCG sound from the API controller (e.g., a mobile smartphone application). The algorithm is used to classify PCG recordings based on their signal quality as good or poor.

    • Heart Sound Timing Algorithm:
      This algorithm detects the presence and timing of specific heart sounds including S1, S2, the systole region, and the diastole region.

    • Murmur Detection & Classification Algorithm: This algorithm is used to identify and classify heart sounds as having "No Murmur", an "Innocent Murmur" (i.e., not pathologic), or a "Structural Murmur" (i.e., pathologic).

    • Murmur Timing Algorithm:

    This algorithm is used to identify in which regions of the heart cycle (systole vs diastole) a heart murmur occurs if either an "Innocent Murmur" or "Structural Murmur" is identified by the Murmur Detection and Classification Algorithm.

    AI/ML Overview

    Here's an analysis of the Eko Murmur Analysis Software (EMAS) acceptance criteria and the study proving its performance, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance (EMAS)
    Murmur ClassificationLower bound of 95% CI for Sensitivity > 75.0% (compared to primary predicate's lower bound of 72.9%)Sensitivity: 85.6% (95% CI: 82.6 - 88.7)
    Lower bound of 95% CI for Specificity > 75.0% (compared to primary predicate's lower bound of 74.9%)Specificity: 84.4% (95% CI: 81.3 - 87.5)
    S1 DetectionNot explicitly stated as a separate acceptance criterion with a numerical threshold, but expected to demonstrate substantially equivalent performance to predicates.Sensitivity: 96.2% (95% CI: 94.9 - 97.4)
    PPV: 97.1% (95% CI: 96.3 - 98.0)
    S2 DetectionNot explicitly stated as a separate acceptance criterion with a numerical threshold, but expected to demonstrate substantially equivalent performance to predicates.Sensitivity: 92.3% (95% CI: 90.3 - 94.3)
    PPV: 94.3% (95% CI: 93.4 - 95.1)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Size: The document does not explicitly state a separate "test set" size. However, it indicates that the clinical validation used a database of 2,380 unique heart sound recordings from 615 unique subjects.
      • Of these, "recordings identified as being good signal by the expert cardiologists" (meaning suitable for analysis) included:
        • 45.8% (approx. 1090 recordings) with a confirmed structural murmur.
        • 54.2% (approx. 1290 recordings) with confirmed no murmur or innocent murmur.
      • For heart sound timing, 299 heart sound recordings were annotated.
    • Data Provenance: Retrospective analysis on a proprietary database. The country of origin is not specified, but the applicant (Eko Devices, Inc.) is based in Oakland, California, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: "Multiple cardiologists" were used. The exact number is not specified.
    • Qualifications of Experts: "Cardiologists." No further details on their years of experience or specific subspecialties are provided.

    4. Adjudication Method for the Test Set

    • Recordings were "annotated by multiple cardiologists."
    • There's no explicit mention of an adjudication method like 2+1 or 3+1. However, the ground truth for murmur classification was obtained via "pairing cardiologist annotations with gold standard echocardiogram," suggesting that the echocardiogram served as the definitive ground truth reference alongside expert opinion.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. The study focuses on the standalone performance of the EMAS algorithm against a ground truth. There is no information provided about human readers improving with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone study was done. The reported performance metrics (Sensitivity, Specificity, PPV) are directly attributed to the "EMAS algorithm testing" and represent the algorithm's performance against the established ground truth. The device is intended as "decision support" and "not intended as a sole means of diagnosis," indicating it operates standalone and then informs a clinician.

    7. The Type of Ground Truth Used

    • For Murmur Classification: Ground truth was established by pairing cardiologist annotations with gold standard echocardiogram.
    • For S1/S2 Timing: Ground truth was established via expert cardiologist annotations.

    8. The Sample Size for the Training Set

    • The document explicitly states: "No study subjects included in the training datasets were included in the test database." However, it does not provide the sample size for the training set. It only mentions that the algorithms were validated using "retrospective analysis on a proprietary database."

    9. How the Ground Truth for the Training Set Was Established

    • The document does not explicitly describe how the ground truth for the training set was established. It only refers to a "proprietary database" used for training and then tested on a separate, distinct set of subjects. Assuming a consistent approach, it's likely similar methods (expert annotations, potentially with echocardiogram correlation) were used, but this is not stated.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200776

    Validate with FDA (Live)

    Device Name
    Eko CORE
    Manufacturer
    Date Cleared
    2020-04-06

    (12 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.

    Device Description

    The Eko CORE (OTC) is a digital stethoscope device designed for use by health care professionals as well as lay users in clinical or non-clinical environments. It enables physicians transition between traditional and digital auscultation. It can electronically amplify, filter and transfer sounds to the accompanying mobile application for storage and sharing or can transmit the data for telemedicine use. It also enables regular users to listen to their body sounds (lungs, heart, arteries, veins, Gastrointestinal tract, etc.) and record and share it with their physicians.

    lt consists of two primary components: 1) The CORE: an electronic attachment for an analog stethoscope and 2) Eko App, and application that pairs with the CORE attachment.

    • . CORE is used to record audio in the stethoscope, convert it to digital data points, and transmit data to a mobile device via Bluetooth". It includes a volume adjustment button, an analog-to-digital power switch, and an LED light indicator.
    • . The app captures audio data from the CORE and provides data visualization, secure data storage, audio playback, and sharing features.

    These features enable a healthcare professional to monitor patients, seek out second opinions from a specialist or use the device for telemedicine use.

    AI/ML Overview

    The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins). It is intended for use on pediatric and adult patients by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list "acceptance criteria" in a quantitative format for specific performance metrics (e.g., sensitivity, specificity for disease detection). Instead, the performance evaluation focuses on demonstrating that the Eko CORE is substantially equivalent to its predicate devices by showing it functions as intended and adheres to safety standards.

    Acceptance Criteria (Implied by Substantial Equivalence Basis)Reported Device Performance
    Device functions as intended for its Indications for Use (amplification, filtering, transmission of auscultation sound data, use by professional and lay users).Functional testing demonstrated the device functioned as intended in all instances. This includes the ability to attach to a standard analog stethoscope, featuring both digital and analog auscultation modes, connecting via Bluetooth to the Eko App for visualization, recording, and transfer of data, and including on/off power buttons and volume adjustment controls.
    Device meets electrical safety standards.Electrical safety testing was successfully conducted, complying with IEC 60601-1.
    Device meets EMC (electromagnetic compatibility) safety standards.EMC testing was successfully conducted, complying with IEC 60601-1-2.
    Device meets wireless coexistence standards.Wireless coexistence testing was successfully conducted.
    Device is safe and effective for over-the-counter (OTC) use by lay users (as this is a new indication compared to the RX predicate).Human factors testing for OTC use yielded passing results, indicating that lay users can safely and effectively operate the device.
    Device maintains similar technological characteristics and performance to predicate devices (Eko Electronic Stethoscope System (K151319) and Tyto Stethoscope (K181612)).The document states "The Eko CORE is a modified version of the previously-cleared predicate device and has very similar technological characteristics. Minor changes have been made, but the devices have the same overall operational and technological characteristics...Both devices have the same frequency range and have the same maximum sound level." Performance data demonstrated the Eko CORE is as safe and effective as the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "human factors testing," which is a type of test set, but does not specify the sample size used for this testing. It also does not directly state the country of origin or whether the data was retrospective or prospective for any of the performance tests. The context suggests that the testing was conducted by Eko Devices, Inc. likely in the US where it's seeking FDA clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not specify the number or qualifications of experts used to establish ground truth. For human factors testing, the "ground truth" would typically relate to the usability and comprehension of the device by representative users. For the functional and safety testing, the "ground truth" is adherence to technical specifications and regulatory standards.

    4. Adjudication Method for the Test Set:

    The document does not describe any adjudication method like 2+1 or 3+1. The performance data section refers to "bench verification," "software verification and validation," "electrical safety, EMC and wireless coexistence testing," and "human factors testing." These types of tests typically involve objective measurements against predefined standards rather than expert adjudication of qualitative outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance:

    The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. The Eko CORE is an electronic stethoscope intended for amplification, filtering, and transmission of sounds, not an AI-powered diagnostic tool in the sense of automatically interpreting sounds or assisting human readers with an AI model. Therefore, this type of study is not relevant to the described device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    The Eko CORE is a hardware device (an electronic stethoscope attachment) that works with an app. Its primary functions are amplification, filtering, and transmission of sound data. It's not described as having an "algorithm only" component that would perform diagnostic interpretations independently. The "performance data" described (bench, software, electrical, EMC, wireless, human factors) are for the device's functional integrity and safety, not for a standalone diagnostic algorithm. Therefore, a standalone performance study in the context of an "algorithm only" is not applicable or described.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance claims appears to be based on:

    • Technical Specifications and Regulatory Standards: For bench verification, software V&V, electrical safety (IEC 60601-1), and EMC safety (IEC 60601-1-2).
    • User Performance and Comprehension: For human factors testing related to OTC use, assessing if the device can be used safely and effectively by lay users.

    8. The Sample Size for the Training Set:

    The document does not mention any training set. This device is an electronic stethoscope, not an AI/ML diagnostic algorithm that would typically require a training set of data for model development. The performance testing described focuses on the device's functional and safety characteristics.

    9. How the Ground Truth for the Training Set Was Established:

    Since there is no mention of a training set or an AI/ML algorithm requiring one, this information is not applicable or provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170874

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-19

    (56 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Eko Model E5 System is intended to be used by healthcare professionals to electronically amplify, filter, and transfer body sounds and single-channel electrocardiogram (ECG) waveforms. The Eko Model E5 System also displays ECG waveforms and phonocardiogram waveforms on the accompanying mobile application for storage and sharing (when prescribed or used under the care of a physician). It can be used to record heart sounds and cardiac murmurs, bruits, respiratory sounds, and abdominal sounds during physical examination in normal patients or those with suspected diseases of the cardiac, vascular, pulmonary, or abdominal organ systems. The device can be used on adults and pediatrics.

    The data offered by the device is only significant when used in conjunction with physician over read as well as consideration of other relevant patient data.

    The device should not be used on infants weighing less than 10kg.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA regarding the Eko Model E5 System. It primarily outlines the regulatory approval of the device and its intended use. It does not contain information about specific acceptance criteria for device performance, the results of a study proving those criteria were met, or details about the study design (sample size, expert involvement, ground truth establishment, MRMC study, etc.).

    Therefore, I cannot extract the requested information to describe the acceptance criteria and the study that proves the device meets them from the provided text.

    To answer your request, I would need a different document, such as a summary of the clinical or performance testing conducted for the Eko Model E5 System, which would typically be included in the full 510(k) submission or a dedicated study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151319

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-08-28

    (102 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Eko Electronic Stethoscope System is intended to be used as a part of a patient by healthcare professionals for diagnostic decision support in clinical settings. Eko is intended for use on pediatric and adult patients. It can electronically amplify, filter and transfer sounds to the accompanying mobile application for storage and sharing. It can used to record heart sounds and cardiac murmurs, bruits, respiratory sounds during physical examination in normal patients or those with suspected diseases of the cardiac, vascular, pulmonary or abdominal organ systems.

    Device Description

    The Eko Electronic Stethoscope System is intended to be used as a part of a patient by healthcare professionals for diagnostic decision support in clinical settings. Eko is intended for use on pediatric and adult patients. It can electronically amplify, filter and transfer sounds to the accompanying mobile application for storage and sharing. It can used to record heart sounds and cardiac murmurs, bruits, respiratory sounds during physical examination in normal patients or those with suspected diseases of the cardiac, vascular, pulmonary or abdominal organ systems.

    AI/ML Overview

    This document is a 510(k) clearance letter for the Eko Electronic Stethoscope System. It confirms that the device is substantially equivalent to legally marketed predicate devices. However, the letter and the "Indications for Use" statement do not contain information about acceptance criteria or a study proving that the device meets such criteria.

    The provided text describes the device's intended use and FDA regulatory information but does not include:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size for the test set or data provenance.
    3. Number of experts or their qualifications for establishing ground truth.
    4. Adjudication method.
    5. Information about a multi-reader multi-case (MRMC) comparative effectiveness study, including effect size.
    6. Whether a standalone (algorithm only) performance study was done.
    7. The type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    Therefore,Based on the provided document, I cannot answer your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria. The document is an FDA 510(k) clearance letter and an "Indications for Use" statement, neither of which includes the detailed performance study information you've requested.