510(k) Data Aggregation

    K Number
    K252686

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-21

    (87 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EFFORTMED CERVICAL PEEK CAGES are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. CERVICAL PEEK CAGES facilitate intervertebral body fusion in the cervical spine and are placed via the anterior approach and packed with autograft/autologous bone graft. Patients should have received at least six (6) months of prior non-operative treatment. EFFORTMED CERVICAL PEEK CAGES are to be used with supplemental fixation.

    The EFFORTMED LUMBAR PEEK CAGES are indicated for intervertebral body fusion at one or two contiguous levels in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) of lumbar spine with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have received at least six (6) months of prior non-operative treatment. The devices are designed to be used with supplemental fixation and autograft/autologous bone graft to facilitate fusion for each spinal region.

    The EFFORTMED ORIZABA CERVICAL CORPECTOMY CAGE is intended for use in skeletally mature patients in the cervical spine (C2-T1) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.

    The EFFORTMED COTOPAXI LUMBAR CORPECTOMY CAGE is intended for use in skeletally mature patients in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.

    The ORIZABA and COTOPAXI CORPECTOMY CAGES are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.

    ORIZABA and COTOPAXI CORPECTOMY CAGES are intended to be used with supplemental spinal fixation systems cleared for use in the cervical, thoracic, and/or lumbar spine. The use of bone grafting material is optional.

    Device Description

    The EffortMed spinal interbody cage system includes cervical intervertebral body fusion (IBFD), transforaminal lumbar interbody fusion (TLIF), and posterior lateral interbody fusion (PLIF) devices. The cervical cage models are the EffortMed Pacaya Cervical PEEK Cage and Masaya Augmented Cervical PEEK cage. The lumbar cage models are the EffortMed TLIF PEEK Cage, Effortmed TLIF PEEK Cage 4° Angled, Effortmed PLIF PEEK Cage, EffortMed PLIF PEEK Cage 5° Angled, and EffortMed Expandable PLIF PEEK Cage. The system was designed to restore height and lordotic angle in the spine. The main role of the cages is to help maintain the cleared disc space stable and intact, until a healthy bony fusion occurs between the adjoining vertebrae. To help achieve this, the inner chamber of the cage body is filled with bone graft before implantation.

    The system also includes cervical and corpectomy cages (identified as used in cervical spine (C2-T1) and in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to, tumors, fractures, and infections. The corpectomy cage models are the EffortMed Orizaba Cervical Corpectomy cage and EffortMed Cotopaxi Lumbar Corpectomy cage.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243946

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-01-15

    (26 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EffortMed Spinal Fixation System is intended for use in the noncervical spine. When used as a posterior, noncervical pedicle and non- pedicle fixation system, the EffortMed Spinal Fixation System is intended to provide additional support during fusion using autograft in skeletally mature patients in the following acute and chronic instabilities or deformities:

    • Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).

    • · Spondylolisthesis.
    • · Trauma (i.e. fracture or dislocation)
    • · Spinal stenosis
    • · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
    • · Tumor
    • · Pseudoarthrosis; and failed previous fusion

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EffortMed Spinal Fixation System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis. Additionally, the EffortMed Spinal Fixation System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis,fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach.

    The EffortMed Spinal Fixation System is intended to be used with autograft and/or allograft.

    Device Description

    EffortMed Spinal Fixation System is a top-loading, multiple component, posterior spinal fixation system consisting of polyaxial pedicle screws, monoaxial pedicle screws, cannulated and spondylolisthesis screws, rods (Straight and curved) , connectors , hooks and setscrews. It allows surgeons and functions to build a spinal implant construction to stabilize and promote spinal fusion. The System components are supplied non-sterile, for single use and fabricated from titanium alloy (Ti6Al4V-ELI) that conforms to ASTM F136. Various sizes of these components are available.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the U.S. FDA for the "EffortMed Spinal Fixation System." This document establishes the substantial equivalence of a new medical device to a legally marketed predicate device.

    Crucially, this document is for a spinal fixation system (implants), not a software-based medical device or AI system. Therefore, the concepts of "acceptance criteria" and "study that proves the device meets the acceptance criteria" as they apply to software/AI performance (e.g., accuracy, sensitivity, specificity studies with ground truth established by experts) are not applicable here.

    Instead, for a spinal fixation system, "acceptance criteria" relate to biocompatibility, mechanical properties, and structural integrity. The "study that proves the device meets the acceptance criteria" refers to non-clinical (bench) testing demonstrating that the device meets established standards for these properties, ensuring it is as safe and effective as the predicate device.

    Here's how to interpret the information provided in the context of this specific device:

    1. Acceptance Criteria and Reported Device Performance (as applicable to a physical implant):

    The acceptance criteria for a spinal fixation system are typically defined by recognized consensus standards (e.g., ASTM standards) that evaluate the mechanical and functional performance of the device. The reported "performance" for such a device is not a single numerical output like an AI's accuracy, but rather the successful demonstration of meeting these standards.

    Acceptance Criteria (based on common standards for spinal implants)Reported Device Performance (based on provided text)
    ASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model (evaluates static and fatigue properties of spinal implant constructs)Tested: "ASTM F 1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model" - Implies successful completion, demonstrating comparable static and fatigue properties to the predicate.
    ASTM F1798: Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis ImplantsTested: "ASTM F 1798: Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants" - Implies successful completion, demonstrating comparable performance of connectors and subassemblies.
    Axial gripping capacity (of pedicle screws)Tested: "Axial gripping capacity" - Implies successful completion, demonstrating adequate pullout strength.
    Axial torque gripping capacity (of pedicle screws)Tested: "Axial torque gripping capacity" - Implies successful completion, demonstrating resistance to loosening under torsional forces.
    Flexion-extension moment (of the construct)Tested: "Flexion-extension moment" - Implies successful completion, demonstrating stability under bending forces.
    ASTM F543: Test Methods for Metallic Medical Bone Screws (covering various properties of bone screws)Tested: "ASTM F543: Test Methods for Metallic Medical Bone Screws" - Implies successful completion, demonstrating compliance with general bone screw standards.
    Annex A3-Test Method for Determining the Axial Pullout Strength of Medical Bone Screws.Tested: "Annex A3-Test Method for Determining the Axial Pullout Strength of Medical Bone Screws." - Implies successful completion of specific pullout strength testing.
    Comparable shear failure force (following Chapman, J.R., et al principles)Tested: "Determining the comparable shear failure force following the equation of Chapman, J.R., et al" - Implies successful completion, demonstrating adequate shear strength.
    Tulip-dissociation testing on worst-case screw constructTested: "Tulip-dissociation testing on the worst-case bottom-loading screw construct at neutral and maximum screw angulations to evaluate the interconnection strength" - Implies successful completion, demonstrating the integrity of the screw head/rod interface.
    Biocompatibility (Material safety)Stated: "fabricated from titanium alloy (Ti6Al4V-ELI) that conforms to ASTM F136." - Implies material safety and biocompatibility due to compliance with a recognized material standard.

    2. Sample Size for Test Set and Data Provenance:

    This information is not applicable in the AI/software sense. For a physical medical device like this, the "test set" refers to the physical units and materials subjected to bench testing. The "sample size" would relate to the number of components or constructs tested to meet statistical significance for the engineering tests. This detail is not typically included in a 510(k) summary, but rather in the full test reports. The "data provenance" (country of origin, retrospective/prospective) is also not relevant here, as it's not clinical data or real-world patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not applicable. Ground truth, in the context of an AI/software device, refers to a definitive diagnosis or finding established by human experts. For a physical implant, "ground truth" is defined by the objective, measurable outcomes of standardized engineering tests, not by expert consensus on images or patient data.

    4. Adjudication Method for Test Set:

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements among human readers/experts in establishing ground truth for diagnostic or AI performance studies. This is not relevant for bench testing of a physical implant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, a MRMC study was not done. MRMC studies are used to evaluate the impact of a medical device (often AI) on human reader performance, typically in diagnostic tasks. This device is a physical implant, not a diagnostic tool where human readers assess cases. The document explicitly states: "Clinical data: Not applicable."

    6. If a Standalone Performance Study Was Done:

    No, a standalone performance study (in the AI sense) was not done. A "standalone" study for AI refers to the algorithm's performance without human intervention. For this implant, the "performance" is its mechanical and material properties, which were evaluated via the listed ASTM standard tests. These are inherent properties of the device itself and don't involve "human-in-the-loop" interaction in the way an AI diagnostic tool would.

    7. Type of Ground Truth Used:

    The "ground truth" for this device's performance is established by objective, quantitative measurements obtained from standardized engineering and material tests (e.g., tensile strength, fatigue life, pullout strength, etc.) as defined by the ASTM standards listed. There is no expert consensus, pathology, or outcomes data used as "ground truth" for proving the device meets its acceptance criteria in the context of this 510(k). Its substantial equivalence is based on meeting these objective engineering standards and being comparable to the predicate device.

    8. Sample Size for the Training Set:

    This information is not applicable. A "training set" is used for machine learning models. This device is a physical implant and does not use machine learning.

    9. How Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no training set for a physical implant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240935

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-17

    (42 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARENAL Anterior Cervical Plate System is intended for fixation and temporary stabilization of the anterior cervical spine (C2-T1) during the development of cervical spine fusions with the following indications:

    • Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    • Trauma (including fractures)
    • Tumors
    • Deformities or curvatures (including kyphosis, lordosis, or scoliosis)
    • Pseudarthrosis
    • Spondylolisthesis
    • Spinal stenosis
    Device Description

    The ARENAL Anterior Cervical Plate System is interior screw fixation of the plate to the cervical spine. The fixation construct consists of a cervical plate that is attached to the vertebral spine with self-tapping or self-drilling bone screws using an anterior approach. Plates are available in a variety of lengths, addressing multiple levels of fixation.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Arenal Anterior Cervical Plate System. It focuses on demonstrating substantial equivalence to predicate devices, primarily through non-clinical testing. It does not contain information about a study proving device performance against acceptance criteria in the context of AI/software or human readers, as the device is a spinal implant.

    Therefore, many of the requested categories are not applicable to the information provided.

    However, I can extract information related to the performance of the device based on mechanical testing.

    Here's the relevant information derived from the provided text for the mechanical testing of the device:

    1. A table of acceptance criteria and the reported device performance

    Test TypeAcceptance Criteria (from FDA Spinal Plating Systems guidance document)Reported Device Performance (Arenal Anterior Cervical Plate System)
    Static compressionNot explicitly stated, but implied as a numerical threshold in the guidance document.Static Compression Bending Yield was in exceptance criteria (likely means acceptance criteria) listed in FDA Spinal Plating Systems guidance document.
    Static compressionNot explicitly stated, but implied as a numerical threshold in the guidance document.Static Compression Bending Stiffness was in exceptance criteria (likely means acceptance criteria) listed in FDA Spinal Plating Systems guidance document.
    Dynamic compressionNot explicitly stated, but implied as a numerical threshold for runout load @ 5M cycles in the guidance document.Dynamic Compression Bending Runout Load @ 5M cycles was in excess of the acceptance criteria listed in FDA Spinal Plating Systems guidance document. (This indicates better performance than criteria)
    Static torsionNot explicitly stated, but implied as a numerical threshold in the guidance document.Static Torsion Yield was in exceptance criteria (likely means acceptance criteria) listed in FDA Spinal Plating Systems guidance document.
    Static torsionNot explicitly stated, but implied as a numerical threshold in the guidance document.Static Torsion Stiffness was in exceptance criteria (likely means acceptance criteria) listed in FDA Spinal Plating Systems guidance document.

    Study that proves the device meets the acceptance criteria:

    The study referenced is a series of non-clinical mechanical tests performed according to ASTM standards for spinal implant constructs. These tests are:

    • Static compression per ASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model
    • Dynamic compression per ASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model
    • Static torsion per ASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the sample size (number of devices tested) for each test.
    • Data Provenance: The tests were performed as part of the 510(k) submission process for a medical device (Arenal Anterior Cervical Plate System). The origin is not specified but is implicitly from testing conducted to support a US FDA submission. It is prospective testing, specifically designed to demonstrate safety and effectiveness for regulatory clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. The ground truth for mechanical testing is based on physical measurements and engineering principles, not expert interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Mechanical test results are quantitative and do not require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical spinal implant, not an AI or software-based diagnostic tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This device is a physical spinal implant, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for these tests are the defined physical properties and performance characteristics as outlined in the ASTM F1717 standard and the FDA Spinal Plating Systems guidance document. This is based on engineering standards and regulatory requirements for spinal implants.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/machine learning device.