510(k) Data Aggregation

    K Number
    K230337

    Validate with FDA (Live)

    Device Name
    EarliPoint
    Date Cleared
    2023-06-29

    (142 days)

    Product Code
    Regulation Number
    882.1491
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EarliPoint System is indicated for use in specialized developmental disabilities centers as a tool to aid clinicians in the diagnosis and assessment of ASD patients ages 16 months through 30 months.

    Device Description

    The device is a more compact version of the predicate device but otherwise has similar functions and features. The system uses an eye tracker to capture the patient's looking behavior while viewing a series of videos. The system then remotely analyzes the looking behavior data using software and outputs a diagnosis of the patient's ASD status and assesses the symptoms associated with ASD.

    The system has two modules:

    EarliPoint System consists of the following:

    • Eye-tracking module and a separate Operator Module that can control the Eye-tracking module remotely. The patient sits on a chair and the Eye-tracking module is adjusted by the operator such that the patient's eyes are within the specification of the eye tracking window
    • -Eye-tracking module captures the patient visual response to social information provided in the form of a series of age-appropriate videos
    • Operator's module is used to initiate and monitors the session remotely -
    • -WebPortal securely stores all patient information, analyzes the eye tracking data, and outputs the results. Users can retrieve the results directly from the web-portal.
    • -Artificial intelligence software analyzes the eye-tracking data and provides a diagnosis for ASD. In addition, it also outputs 3 indices (called EarliPoint Severity Indices) that proxy the ADOS-2 and Mullen validated ASD instruments
      • o Social Disability Index correlates and proxies ADOS-2
      • O Verbal Ability Index correlates and proxies the age equivalent Mullen Verbal Ability score
      • O Non-verbal Ability Index correlates and proxies the age equivalent non-verbal Mullen Ability score

    The eye-tracker used in the EarliPoint device has similar capability as the eye tracker used in the predicate device.

    AI/ML Overview

    The provided text describes a 510(k) submission for a modified EarliPoint System, which is an aid for diagnosing and assessing Autism Spectrum Disorder (ASD). The submission details performance testing and a pivotal clinical study.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria in this submission are derived from the clinical study's performance metrics against a reference standard (expert clinical diagnosis for ASD).

    Performance MetricAcceptance Criteria (from study results)Reported Device Performance (EarliPoint mITD)Reported Device Performance (EarliPoint CertainDx)
    SensitivityNot explicitly stated as a pre-defined acceptance criteria, but observed performance71% (95% CI: 64.6% - 76.9%)78.0% (95% CI: 70.5% - 84.3%)
    SpecificityNot explicitly stated as a pre-defined acceptance criteria, but observed performance80.7% (95% CI: 75.3% - 85.4%)85.4% (95% CI: 79.5% - 90.2%)
    SafetyNo serious adverse events related to device useNo reported serious adverse eventsNo reported serious adverse events
    Correlation with ADOS-2Positive correlation with EarliPoint Social Disability IndexCorrelates and proxies ADOS-2Correlates and proxies ADOS-2
    Correlation with Mullen Verbal AbilityPositive correlation with EarliPoint Verbal Ability IndexCorrelates and proxies age equivalent Mullen Verbal Ability scoreCorrelates and proxies age equivalent Mullen Verbal Ability score
    Correlation with Mullen Non-verbal AbilityPositive correlation with EarliPoint Non-verbal Ability IndexCorrelates and proxies age equivalent non-verbal Mullen Ability scoreCorrelates and proxies age equivalent non-verbal Mullen Ability score

    Note: The document does not explicitly state pre-defined numerical acceptance criteria for sensitivity and specificity. The reported performance is the outcome of the study, and its acceptance implies these values were deemed sufficient by the FDA for substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 475 evaluable patients for primary and secondary endpoint analysis. 25 patients had missing data for either the device or control diagnosis, bringing the total enrolled to 500.
    • Data Provenance:
      • Country of Origin: United States
      • Retrospective or Prospective: Prospective
      • Multi-center: Six sites

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not specified. The text mentions "expert clinicians" but does not quantify the number of individual experts or how many were involved per case.
    • Qualifications of Experts: Referred to as "expert clinicians" and their diagnosis is considered the "current best practice for diagnosis of ASD." No further specific qualifications like years of experience or board certifications are provided in the document.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not explicitly stated. The text mentions "expert clinician diagnosis" as the reference standard but does not detail how consensus was reached if multiple experts were involved or if there was a single expert per case.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in the traditional sense of comparing human readers with AI assistance versus human readers without AI assistance.
    • Effect Size: Therefore, no effect size for human reader improvement with AI assistance is reported. The study's design was a within-subject comparison where all patients were evaluated by both the EarliPoint system and expert clinicians, comparing the device's diagnostic output against the expert diagnosis.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: Yes, the study evaluated the "EarliPoint System diagnosis" relative to the expert clinical diagnosis. The device's output (ASD diagnosis and severity indices) is generated by "Artificial intelligence software [that] analyzes the eye-tracking data and provides a diagnosis for ASD." This indicates a standalone performance evaluation of the algorithm. The "EarliPoint CertainDx" analysis further refined this by focusing on cases where clinicians were certain of the diagnosis, still representing the device's standalone output being compared to this subset of expert diagnoses.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: "Expert clinician diagnosis (current best practice for diagnosis of ASD)." This is a form of expert consensus or clinical standard of care. It is further supported by correlation with validated ASD instruments (ADOS-2 and Mullen).

    8. The Sample Size for the Training Set

    • The document does not explicitly state the sample size for the training set. The clinical study described (the pivotal study) is an evaluation study (test set) for the device's performance, not the dataset used to train the AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not explicitly state how the ground truth for the training set was established. While it details the ground truth for the test set (expert clinician diagnosis), it does not provide information on the training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213882

    Validate with FDA (Live)

    Date Cleared
    2022-06-08

    (177 days)

    Product Code
    Regulation Number
    882.1491
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EarliPoint System is indicated for use in specialized developmental disabilities centers as a tool to aid clinicians in the diagnosis and assessment of Autism Spectrum Disorder (ASD) for patients ages 16 months through 30 months.

    Device Description

    The EarliPoint System is a medical device for diagnosis of Autism Spectrum Disorder (ASD) in children.

    EarliPoint System consists of the following:

    • EarliPoint WebPortal to enter the patient information and for access to the patient evaluation results,
    • EarliPoint Device with eye-tracking capability captures the patient visual response to social information provided in the form of a series of age-appropriate videos
    • Artificial intelligence software analyzes the eye-tracking data and provides a diagnosis for ASD.
    • Eye-tracking data also outputs 3 indices (called EarliPoint Severity Indices) that proxy the ADOS-2 and Mullen validated ASD instruments
      • Social Disability Index correlates and proxies ADOS-2
      • Verbal Ability Index correlates and proxies the age equivalent Mullen Verbal Ability score
      • Non-verbal Ability Index correlates and proxies the age equivalent non-verbal Mullen Ability score
    AI/ML Overview

    The EarliPoint System, a device for diagnosing Autism Spectrum Disorder (ASD), underwent a pivotal study to demonstrate its safety and effectiveness.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" but rather shows the device's performance (sensitivity and specificity) relative to a predicate device (CanvasDx) and a "gold standard" expert clinical diagnosis. The implication is that meeting or exceeding the predicate's performance, especially for certain populations, was a key aspect of acceptance.

    Population/MetricSensitivity Mean (n/N) [95% CI]Specificity Mean (n/N) [95% CI]
    CanvasDx (mITD), N=42551.6% (63/122) [42.8% - 60.5%]18.5% (56/303) [14.3% - 23.3%]
    EarliPoint (mITD), N=47571% (157/221) [64.6% - 76.9%]80.7% (205/254) [75.3% - 85.4%]
    EarliPoint CertainDx (N=335)78.0% (117/150) [70.5%, 84.3%]85.4% (158/185) [79.5% - 90.2%]
    EarliPoint UncertainDx (N=140)56.3% (p=1.00)68.1% (p=0.69)

    Note: The document states: "When compared to the predicate. CanvasDx, the sensitivity and specificity of the EarliPoint device are higher than the CanvasDx and hence the two devices are substantially equivalent." This implies that outperforming the predicate was a key acceptance criterion.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 475 evaluable patients for primary and secondary endpoint analysis (from an initial enrollment of 500).
    • Data Provenance: The study was "multi-center" with data collected from "six sites in the United States." The study was "prospective."

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not explicitly stated, but the ground truth was established by "expert clinicians" at "six sites." The document refers to "expert clinician diagnosis" and "clinicians' reference diagnoses."
    • Qualifications of Experts: Described as "expert clinician diagnosis (current best practice for diagnosis of ASD)." Specific qualifications (e.g., years of experience, specialization) are not detailed.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method for conflicting expert opinions. It refers to a "gold standard of the expert clinician diagnosis." It does, however, categorize the ground truth into "Certain Dx Population" (clinicians' certainty rating > 80%) and "Uncertain Dx Population" (clinicians' certainty rating ≤ 80%), indicating some level of consideration for diagnostic confidence.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was explicitly mentioned in the provided text, comparing human readers with AI assistance vs. without AI assistance. The study focuses on the device's performance compared to human expert diagnosis as the ground truth. The device is intended as an "aid clinicians," suggesting a human-in-the-loop scenario, but a specific MRMC study design is not detailed.

    6. Standalone (Algorithm Only) Performance

    • The reported sensitivity and specificity values represent the standalone performance of the EarliPoint System's algorithm in diagnosing ASD, as it provides a "diagnosis for ASD" from analyzing eye-tracking data. The study "evaluated for ASD by both the EarliPoint system and by expert clinician diagnosis."

    7. Type of Ground Truth Used

    • The primary ground truth used was "expert clinician diagnosis (current best practice for diagnosis of ASD)."
    • Additionally, the study correlated the three EarliPoint Severity Indices (Social Disability, Verbal Ability, Non-verbal Ability) against corresponding "expert clinical instruments of ADOS-2 and Mullen." This indicates a reliance on established clinical diagnostic tools as part of the expert ground truth.

    8. Sample Size for the Training Set

    • The document does not provide the sample size used for the training set. It only describes the design and results of the pivotal clinical study (test set).

    9. How the Ground Truth for the Training Set was Established

    • The document does not provide information on how the ground truth for the training set was established. It focuses solely on the test set's ground truth.
    Ask a Question