Search Results
Found 8 results
510(k) Data Aggregation
(29 days)
The Omni Foot and Ankle Plating System is intended for use in internal fixation of arthrodeses, osteotomies, fractures and nonunions of the small bones of the foot & ankle including the fore-, mid-, and hind foot and ankle applications.
The Omni Foot and Ankle Plating System is a bone fixation system consisting of unalloyed Titanium plates and Titanium Alloy (Ti-6AL-4V) locking and non-locking plate screws, which meet ASTM F67 and ASTM F136, and a set of instruments used for implant site preparation and delivery. The plates are available in various configurations, essentially differing by the lengths and number of holes. The plate screws are provided in diameters of 2.8mm and 3.5mm in lengths from 8mm to 50mm. The System offers 3.5mm cannulated screws and 4.0mm solid screws in various lengths to be used as adjunctive fixation. The 3.5mm cannulated screw can also be used with a specialized locking screw ("Post") which contains a locking feature at the distal end for compression/stabilization.
The provided text is for a 510(k) premarket notification for a medical device called the "Omni Foot and Ankle Plating System." This document does not describe the acceptance criteria or a study that proves the device meets those criteria in the context of an AI/ML powered device. Instead, it details the regulatory process for establishing substantial equivalence to previously cleared devices.
Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, test set details, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set specifics for an AI/ML powered device from the provided text.
The document explicitly states: "No clinical testing was performed." and relies on "engineering analysis" for comparison to predicate devices, which is typical for non-AI/ML mechanical devices.
Ask a specific question about this device
(32 days)
The Axis Charcot Fixation System in diameters of 4.5 to 8.5mm is indicated for reconstruction procedures, non-unions and fusions of bones in the foot and ankle including the metatarsals, cuneiforms, cuboid, navicular, calcaneus and talus; specific examples include: medial and lateral column fusion resulting from neuropathic osteoarthropathy (Charcot).
The Axis Charcot Fixation System consists of 5.5, 6.5 and 7.5mm cannulated, titanium alloy fixation beams and accessories used for midfoot reconstruction. The modified device adds additional sizes, 4.5 and 8.5mm cannulated, titanium alloy fixation beams, previously cleared under Extremity Medical's 4.5 to 8.5 Screw System (K171018). The additional sizes offer the surgeon options for placement based on patient anatomy.
The provided text is a 510(k) summary for the Axis Charcot Fixation System, a medical device. It does not include information about AI/ML device performance, acceptance criteria, or studies involving human readers, as it describes a bone fixation system.
Therefore, most of the requested information cannot be extracted from this document as it pertains to AI/ML device evaluation.
However, based on the provided text, I can infer the following:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative manner typical for AI/ML performance. Instead, it refers to "engineering analyses evaluating mechanical strength" and "pullout strength". The performance determined was "substantially equivalent to the predicate devices for the intended use."
| Acceptance Criteria (Inferred from testing type) | Reported Device Performance |
|---|---|
| Mechanical strength of the smallest beam diameter (4.5mm) is substantially equivalent to predicate. | Demonstrated substantial equivalence to predicate devices. |
| Pullout strength of the shortest, smallest beam diameter is substantially equivalent to predicate. | Demonstrated substantial equivalence to predicate devices. |
| The largest beam diameter (8.5mm) is no worst case than the predicates. | No additional safety and effectiveness concerns presented. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable and not provided. The testing described is mechanical, not involving patient data or a "test set" in the context of an AI/ML device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable and not provided. The "ground truth" for this device would be established through engineering standards and physical measurements, not expert consensus on medical images or clinical outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and not provided. Mechanical testing does not involve adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No such study was performed or is relevant to this device. This device is a physical bone fixation system, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable. This device is a physical bone fixation system, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is based on engineering standards and physical measurements of mechanical properties (strength, pullout force).
8. The sample size for the training set
This information is not applicable and not provided. There is no AI/ML model being "trained" for this device.
9. How the ground truth for the training set was established
This information is not applicable and not provided. There is no AI/ML model being "trained" for this device.
In summary, the provided document describes a 510(k) submission for a physical medical device (bone fixation system) and does not contain information related to AI/Machine Learning device performance or ground truth establishment in the context of diagnostic or interventional AI.
Ask a specific question about this device
(28 days)
The Intraosseous Fixation System is intended for reduction of arthrodeses, osteotomies, intra- and extra-articular fractures and nonunions of the small bones and joints of the hand, wrist, foot and ankle, appropriate for the size of the device.
The Extremity Medical Intraosseous Fixation System is intended for reduction and internal fixation of arthrodeses, osteotomies, intra- and extra-articular fractures and nonunion of the small bones and joints of the hand, wrist, foot, and ankle. The system consists of solid and cannulated screws, with the option of adding Washers; a standard/flat Washer, a Washer Post (referred to as an X-Post) for engaging the head of the Screw, and an X-Clip for engaging the threads of the screw. The X-Post/X-Clip come in varying sizes and lengths. Similarly, the lag screws come in varying diameters and lengths, and are also available in short and long thread configurations. The modified device is a product line extension and improvement to the Screw and Washer System, cleared under K121349, intended to provide modularity to the end user. New components to the system consist of cannulated 3.5 mm screws and implant accessories, X-Posts and X-Clips, to offer surgeons options for placement. Additionally, the system will offer additional thread length options for the 4.5mm screws cleared under K121349 and 6.5mm screws substantially equivalent to those cleared under K121349.
This document, K201556, pertains to a 510(k) premarket notification for a medical device called the "Intraosseous Fixation System," manufactured by Extremity Medical, LLC. It is a Class II device (21 CFR 888.3040) classified as a smooth or threaded metallic bone fixation fastener.
The provided text does not contain any information about acceptance criteria or a study proving the device meets those criteria for software or an AI/ML algorithm. The document describes a traditional medical device (implants for bone fixation) and its substantial equivalence to predicate devices based on design, materials, and mechanical properties. There is no mention of an AI/ML component, image analysis, or any performance metrics typically associated with such systems.
Therefore, I cannot fulfill your request for the specific information regarding acceptance criteria, test set details, ground truth establishment, or human reader studies related to an AI/ML device, as none of that information is present in the provided FDA 510(k) clearance document.
Ask a specific question about this device
(268 days)
The KinematX Total Wrist Arthroplasty System is indicated for the replacement of wrist joints disabled by pain, deformity, and/or limited motion caused by:
- Non-inflammatory degenerative wrist disease of the radiocarpal joint including osteoarthritis, post-traumatic arthritis, and Kienbock's disease
- Revision where other devices or treatments have failed
- Scapholunate Advanced Collapse (SLAC)
- Rheumatoid Arthritis
The device is intended to be implanted with bone cement.
The KinematX Total Wrist Arthroplasty System is a semi-constrained implant system designed to replace the joints of the wrist to alleviate pain while restoring functionality and mobility of the wrist joints. The system consists of two main, modular components:
- Radial Implant Assembly: CoCr radial stem and an ultra-high molecular weight polyethylene (UHMWPE) bearing surface that interlocks onto a CoCr tray
- Carpal Implant Assembly: Baseplate with central peg, index metacarpal screw, hamate screw, and CoCr cap.
This FDA document, K191525, describes the KinematX Total Wrist Arthroplasty System. However, it does not include details about acceptance criteria, device performance tables, sample sizes for test or training sets, ground truth establishment, expert qualifications, or multi-reader multi-case studies. It explicitly states that "No clinical testing was performed."
Therefore, I cannot provide the requested information based on the provided text. The document focuses on establishing substantial equivalence to predicate devices through non-clinical testing (mechanical and cadaveric evaluations) and a comparison of indications for use, materials, and design.
Ask a specific question about this device
(62 days)
The Axis Plating System is indication and fixation of fractures or osteotomies, reconstruction procedures, nonunions and fusions of bones in the foot and ankle including the metatarsals, cuneiforms, cuboid, navicular, calcaneus and talus; specific examples include: medial and lateral column fusion resulting from neuropathy (Charcot).
The Axis Plating System is a bone fixation system consisting of Titanium Alloy (Ti-6AL-4V) plates, locking and non-locking plate screws, which meet ASTM F136, and a set of instruments used for implant site preparation and delivery. The plates are available in various configurations, essentially differing by geometry and number of holes. The plate screws are provided in diameters of 3.5mm and 4.5mm in lengths from 8mm to 50mm. The System offers 4.0mm cannulated screws and beams in various lengths to be used as adjunctive fixation. The 4.0mm cannulated screws and beams can also be used with a specialized locking screw ("Post") which contains a locking feature at the distal end for compression/stabilization.
This document is a 510(k) summary for the Axis Plating System, a medical device for bone fixation. It explicitly states: "No clinical testing was performed."
Therefore, based on the provided text, it's impossible to describe acceptance criteria or a study that proves the device meets those criteria from an AI/algorithm perspective, as no clinical study involving an AI or algorithmic component was conducted or described in this submission.
The document is about a hardware medical device (plates and screws), and its substantial equivalence is based on non-clinical testing (engineering analysis for plate and screw properties) and comparison to predicate devices, not on the performance of a diagnostic algorithm or AI.
If this was a misunderstanding and you intended to ask about the acceptance criteria and non-clinical testing specifically for the Axis Plating System (a physical device), then here's a summary of what's provided:
1. A table of acceptance criteria and the reported device performance:
The document doesn't provide a table of acceptance criteria with specific numerical targets and results for device performance. Instead, it lists the types of non-clinical engineering analyses performed. The "performance" is implicitly deemed acceptable if these tests demonstrate equivalence to predicate devices and satisfy relevant ASTM standards (ASTM F136 mentioned for Titanium Alloy).
| Acceptance Criteria (Implied) | Reported Device Performance (Tests Performed) |
|---|---|
| Equivalence in Material Properties | Titanium Alloy (Ti-6AL-4V) plates and screws meet ASTM F136 |
| Acceptable Mechanical Performance of Screws | Screw Engineering Analysis for: - Pullout - Torsion/bending - Torque to failure |
| Acceptable Mechanical Performance of Plates | Plate Engineering Analysis |
| Equivalence in Design, Principle of Operation | Comparison to predicate devices (K190365, K180808, K140792) |
Regarding the other points, as they relate to an AI/algorithm study, they are not applicable to this submission:
- 2. Sample sized used for the test set and the data provenance: Not applicable. No clinical test set or data described.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical test set requiring expert ground truth.
- 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done...: Not applicable. This is not a diagnostic AI device.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for this device's substantial equivalence is based on engineering principles, material standards, and comparison to existing legally marketed devices.
- 8. The sample size for the training set: Not applicable. No AI model trained.
- 9. How the ground truth for the training set was established: Not applicable.
In conclusion, the provided FDA document pertains to a physical medical device (bone fixation system) and explicitly states "No clinical testing was performed" for its substantial equivalence determination. Therefore, no information regarding acceptance criteria, test sets, ground truth establishment, or clinical studies involving AI or human readers can be extracted from this text.
Ask a specific question about this device
(102 days)
The Omni Foot Plating System is intended for use in internal fixation, reconstruction, or arthrodesis of the 1st Metatarsalphalangeal joint.
The Omni Foot Plating System is a bone fixation system consisting of unalloyed Titanium plates and Titanium Alloy (Ti-6AL-4V) locking and non-locking plate screws, which meet ASTM F67 and ASTM F136, and a set of instruments used for implant site preparation and delivery. The plates are available in various configurations, essentially differing by the lengths and number of holes. The plate screws are provided in diameters of 2.8mm and 3.5mm in lengths from 8mm to 50mm. The System offers 3.5mm cannulated screws in various lengths to be used as adjunctive fixation. The 3.5mm cannulated screw can also be used with a specialized locking screw ("Peg") which contains a locking feature at the distal end for compression/stabilization.
The provided document is a 510(k) premarket notification for the Omni Foot Plating System, which is a metallic bone fixation device. This type of device is classified as a Class II medical device. For such devices, clinical performance studies using patient data (like those involving AI-powered diagnostic tools) are typically not required, as substantial equivalence is often demonstrated through technological comparison, engineering analysis, and non-clinical bench testing.
Therefore, the requested information regarding acceptance criteria and studies proving the device meets those criteria, particularly aspects like sample size for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth for training data, is not applicable to this type of device submission.
The document explicitly states under "Clinical Testing": "No clinical testing was performed."
Instead, the acceptance criteria and proof of meeting them for the Omni Foot Plating System are based on:
- Technological Comparison: Demonstrating that the design, materials, and principle of operation are substantially equivalent to legally marketed predicate devices.
- Engineering Analysis: Detailed analysis of the device's design and function.
- Mechanical/Bench Testing: Physical tests conducted in a laboratory setting to verify performance specifications.
Here's a breakdown of the relevant information provided for the Omni Foot Plating System in relation to its acceptance criteria:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Demonstrated Equivalence through) | Reported Device Performance (Summary from 510(k)) |
|---|---|
| Design and Materials Similarity to Predicates | Omni Foot System, Bone Plates: Similar to predicate devices (K161864, K140397, K121349) regarding design and materials. Features a polyaxial locking feature identical to K161864. |
| Screw Design Similarity to Predicates | Omni Foot System, Bone Screws: Identical in design to Arix Foot System Screw (K161864). |
| Locking Peg Functionality Equivalence | Omni Foot System, Locking Peg: Acts like a locking screw and bone washer/nut for compression, similar to the crossing screw construct of IO FiX (K121349). Its 4.2mm diameter is equivalent to Paragon 28 ParaLock Plating System (K140397). |
| Mechanical Performance (Bench Testing) | Bench tests (pullout, torque, static bend testing) were conducted on the proposed 3.5mm screw and Peg device. The results indicate that the Omni Foot Plating System is equivalent to the predicate device in terms of these mechanical properties. |
| Material Compliance | Components meet ASTM F67 (unalloyed Titanium plates) and ASTM F136 (Titanium Alloy Ti-6AL-4V screws). |
| Intended Use Equivalence | The intended use for internal fixation, reconstruction, or arthrodesis of the 1st Metatarsalphalangeal joint is aligned with the presumed uses of the predicate devices for similar orthopedic applications. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not Applicable. No human subjects test set or clinical data was used for this submission. The "test set" primarily refers to the physical samples of the device components used for bench testing. The document does not specify the exact number of samples tested for each bench test (e.g., how many screws were tested for pullout strength).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. Ground truth, in the context of clinical studies with human data, is not relevant here. Device equivalence was established through engineering principles and mechanical testing, which would involve qualified engineers and technicians performing the tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. Adjudication methods apply to the interpretation of clinical data by multiple experts. Not relevant for this type of device submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. The device is a physical implant, not an AI-powered diagnostic tool. MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not Applicable. For this device, the "ground truth" for demonstrating safety and effectiveness is primarily established through adherence to engineering design specifications, material standards (ASTM), and the results of validated mechanical bench tests demonstrating properties (like pullout strength, torque, static bend) that are equivalent to or meet the requirements of predicate devices.
8. The sample size for the training set:
- Not Applicable. "Training set" refers to data used to train algorithms, which is not relevant for this physical device.
9. How the ground truth for the training set was established:
- Not Applicable. As per point 8, there is no training set in the context of this device submission.
In summary, the 510(k) submission for the Omni Foot Plating System relies on non-clinical data (engineering analysis and mechanical bench testing) to demonstrate substantial equivalence to predicate devices, rather than clinical studies with human data. Therefore, many of the questions related to clinical study design and data analysis are not applicable.
Ask a specific question about this device
Axis Charcot Fixation System:
The Axis Charcot Fixation System in diameters of 5.5, 6.5 and 7.5mm is indicated for reconstruction procedures, non-unions and fusions of bones in the foot and ankle including the metatarsals, cuneiforms, cuboid, navicular, calcaneus and talus; specific examples include: medial and lateral column fusion resulting from neuropathic osteoarthopathy (Charcot).
4.5 to 8.5 Screw System:
The 4.5 to 8.5mm Screw System is intended for fixation arthrodesis of the metatarsal-cuneiform, navicular-cuneiform, metatarsal-cuboid, talonavicular, and calcaneocuboid joints.
Axis Charcot Fixation System
The Axis Charcot Fixation System consists of 5.5, 6.5 and 7.5mm cannulated, titanium alloy fixation beams and accessories used for midfoot reconstruction.
4.5 to 8.5 Screw System
The 4.5 to 8.5 diameter screws consists of cannulated, titanium alloy fixation screws for use in bone reconstruction, osteotomy, arthrodesis and fracture repair and fixation in the foot.
Here's an analysis of the provided text regarding acceptance criteria and supporting studies:
This document is a 510(k) premarket notification for the "Axis Charcot Fixation System, 4.5 to 8.5mm Screw System." It's important to note that 510(k) submissions typically demonstrate substantial equivalence to a predicate device, rather than de novo approval requiring extensive clinical efficacy studies. Therefore, the "acceptance criteria" here are framed around demonstrating comparable performance to existing, legally marketed devices.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria Category | Specific Criteria/Tests Performed | Reported Device Performance (vs. Predicate) |
|---|---|---|
| Material Equivalence | Material of manufacture | Ti-6Al-4V (equivalent to predicate devices) |
| Screw Size Offering | Diameters (5.5, 6.5, 7.5mm for Axis Charcot; 4.5 to 8.5mm for Screw System) | Equivalent to predicate devices (WMT Salvation Beams and Bolts, Smith & Nephew 6.5 and 8.0mm Cannulated Screws; Extremity Medical Screw and Washer System, Paragon28 Monster Screw System) |
| Mechanical Performance | Pull-out strength | Bench testing performed and compared to predicate device |
| Static bending | Bench testing performed and compared to predicate device | |
| Dynamic bending | Bench testing performed and compared to predicate device | |
| Design Characteristics | Optional accessory washer/nut | Design differences in the clip of the washer/nut do not introduce new issues of safety or effectiveness compared to predicate washers/nuts. |
| Intended Use Equivalence | Indications for Use | "substantially equivalent" to predicate devices based on indications for use. |
| Principle of Operation | (Implicitly assessed) | "substantially equivalent" to predicate devices based on principles of operation. |
2. Sample size used for the test set and the data provenance
The document explicitly states: "No clinical testing was performed." Therefore, there is no "test set" in the sense of patient data.
- Sample Size for Mechanical Testing: Not explicitly stated in the summary, but typical for bench testing, multiple samples of each device configuration would be tested.
- Data Provenance: Not applicable as no clinical data was collected. The bench testing would have been conducted by Extremity Medical, LLC or a contracted lab.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. As no clinical testing was performed, there was no ground truth to establish from patient data using expert review. The "ground truth" for the mechanical testing would be the raw data collected from the testing equipment, and the comparison point would be the performance of the predicate device (likely from published data or internal testing).
4. Adjudication method for the test set
Not applicable. No expert adjudication of clinical data was performed.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a bone fixation system (screws), not an AI-powered diagnostic tool. MRMC studies are used for evaluating diagnostic performance in image interpretation.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is a medical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the mechanical testing:
- Ground Truth: The physical properties and performance measurements (e.g., pull-out force, bending stiffness, fatigue life) derived directly from the bench tests.
- Comparison Basis: The performance characteristics of the identified predicate devices, likely obtained through literature, publicly available predicate device information, or internal testing for comparison purposes.
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device. The design and manufacturing processes are informed by engineering principles, material science, and the performance characteristics of existing devices, rather than a "training set" in the context of data-driven models.
9. How the ground truth for the training set was established
Not applicable. See point 8.
Ask a specific question about this device
(122 days)
The HammerFiX device is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe and mallet toe.
The HammerFiX device is a bone fixation device consisting of a sterile, threaded PEEK implant and a set of instruments used for implant site preparation and delivery. The device is offered in small, medium and large implant sizes to allow for use in the proximal interphalangeal (PIP) joints of the lesser toes of the foot.
The provided text describes a 510(k) summary for the HammerFiX device, a bone fixation device. However, it explicitly states that no clinical testing was performed. Therefore, the information requested regarding acceptance criteria and a study proving the device meets them cannot be fully provided from the given document as it pertains to clinical performance. The 510(k) relies on non-clinical bench testing to establish substantial equivalence to predicate devices, rather than clinical efficacy or specific performance criteria from a clinical study.
Here's a breakdown of the available information based on the provided text, acknowledging the lack of clinical study data:
Acceptance Criteria and Study Details for HammerFiX Device (Based on Non-Clinical Testing)
1. Table of Acceptance Criteria and Reported Device Performance:
Since no clinical study was performed to evaluate performance against clinical acceptance criteria, the table below reflects the non-clinical testing performed and the general statement of equivalence based on those tests. Specific quantitative acceptance criteria or detailed numerical performance metrics from the bench testing are not provided in this summary.
| Acceptance Criteria Category (Non-Clinical) | Reported Device Performance |
|---|---|
| Mechanical Performance | |
| Pull-out Strength | Performed and compared to predicate devices. (Details not specified) |
| Torque Strength | Performed and compared to predicate devices. (Details not specified) |
| Static Bending | Performed and compared to predicate devices. (Details not specified) |
| Dynamic Bending | Performed and compared to predicate devices. (Details not specified) |
| Overall Equivalence | The HammerFiX device is substantially equivalent to its predicate device based on indications for use, principles of operation, design, and mechanical test data. |
2. Sample Size Used for the Test Set and Data Provenance:
- Test Set Sample Size: Not specified in the provided summary. This refers to the number of devices or constructs tested during non-clinical bench testing.
- Data Provenance: The testing was "bench testing," meaning it was conducted in a laboratory setting, likely at the manufacturer's facility or a contracted testing lab. No country of origin for data is stated, and it is inherently prospective for the specific bench tests conducted to support the 510(k).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
- Not Applicable. For non-clinical bench testing, "ground truth" as it relates to expert clinical opinion is not established. The "ground truth" for mechanical testing is typically defined by standardized test methods and engineering specifications. The tests compare the device's measured mechanical properties to those of predicate devices or established engineering standards.
4. Adjudication Method for the Test Set:
- Not Applicable. Adjudication methods (like 2+1, 3+1) are relevant for clinical studies, especially when multiple readers or experts assess subjective outcomes. For non-clinical mechanical testing, results are typically objective measurements, not subject to adjudication in this manner.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. The document explicitly states: "No clinical testing was performed." Therefore, no MRMC study was conducted.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. The HammerFiX device is a physical bone fixation implant, not a software algorithm or AI-based device. No standalone algorithm performance evaluation would be relevant.
7. The Type of Ground Truth Used:
- Non-Clinical Mechanical Standards and Predicate Device Performance: For the non-clinical testing, the "ground truth" was established by comparing the HammerFiX device's mechanical properties (pull-out, torque, static, and dynamic bending) against:
8. The Sample Size for the Training Set:
- Not Applicable. This question is relevant for AI/ML algorithm development. Since the HammerFiX device is a physical implant and no AI/ML component is mentioned, there is no "training set."
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable. See explanation above for training set.
Summary: The 510(k) clearance for the HammerFiX device was based on non-clinical bench testing demonstrating substantial equivalence to predicate devices, rather than a clinical study evaluating human performance or clinical outcomes. The document clearly states that "No clinical testing was performed."
Ask a specific question about this device
Page 1 of 1