Search Filters

Search Results

Found 15 results

510(k) Data Aggregation

    K Number
    K220533

    Validate with FDA (Live)

    Date Cleared
    2022-11-16

    (265 days)

    Product Code
    Regulation Number
    868.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endoscopy Oxygen Mask is a single patient, disposable device intended for delivering supplemental oxygen and monitoring expired gases from the patient, with ports to allow the clinician to insert scopes, probes, or tubes. It is for non-intubated, spontaneously breathing patients greater than 30 kg.

    Device Description

    The Endoscopy Oxygen Mask is a multi-port mask that serves several functions: A standard oxygen mask for when a patient requires supplemental oxygen; Sampling of exhaled gases for monitoring, typically end-tidal CO2; Additional ports (membranes) to allow for most types of scopes, probes, and tubes to be inserted while still delivering supplemental O2 and sampling exhaled gases.

    AI/ML Overview

    The provided document describes the Endoscopy Oxygen Mask (K220533) and its substantial equivalence to a predicate device, the Panoramic Oxygen Mask (POM) (K172365). The acceptance criteria for the new device are primarily demonstrated through comparative non-clinical testing against the predicate device, showing similar performance in key aspects.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state "acceptance criteria" in a numerical or pass/fail format for each performance metric. Instead, it relies on demonstrating similarity or equivalence to the legally marketed predicate device (K172365) through non-clinical comparative testing. The performance is reported in terms of this similarity.

    Acceptance Criteria (Inferred from Comparison to Predicate)Reported Device Performance (Endoscopy Oxygen Mask)
    Indications for Use: Single patient, disposable device intended for delivering supplemental oxygen and monitoring expired gases from the patient, with ports to allow the clinician to insert scopes, probes, or tubes for non-intubated, spontaneously breathing patients > 30 kg.Similar to predicate.
    Patient Population: Non-intubated spontaneously breathing patients (Adults to Children).Similar to predicate. (Note: Subject device tested for Adults, whereas predicate was for Adults to Children).
    Environment of Use: Locations where procedures are performed where the patient requires supplemental oxygen, monitoring exhaled gases, and scope access (Hospital, sub-acute, clinic, physician offices, pre-hospital).Similar to predicate.
    Duration of Use: Single patient, disposable, < 24 hours.Similar to predicate.
    Mode of Operation: O2 delivery through standard oxygen supply tubing and simultaneous exhaled gas monitoring via a gas sampling line connected from mask to capnography or oxygen only delivery. Slit access ports for introduction of a scope.Similar to predicate.
    Gas Sampling Connection: Luer slip fit.Similar to predicate.
    Profile: Over the nose / mouth.Similar to predicate.
    Face Strap: Yes.Similar to predicate.
    Entrainment Vents: One-way valves to prevent rebreathing.Similar to predicate.
    %CO2 accuracy and Respiration rate: Tested at different simulated patient settings for breath rate, Tidal Volume at different CO2 concentrations with waveforms with similar performance to predicate.Performed similarly to the predicate, with the proposed device performing better, but no claim made other than equivalence.
    Biocompatibility: External Communicating (indirect), Tissue contact; Surface Communicating (direct), Skin contact; Limited duration of use (< 24 hours) as per ISO 10993-5, ISO 10993-10, ISO 18562-2.Similar to predicate. Cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material mediated pyrogenicity, and particulate matter testing performed.
    Shelf-life: Equivalent to predicate's shelf-life.1 year accelerated age testing (Predicate had 3 years real-time). Implies similarity in shelf-life characteristics.
    FiO2 results: Similar across the range of tests.Similar across the range of tests. The differences did not have a pre-defined clinical criteria and are reported as relative performance for clinician guidance.
    Internal Volume: No adverse impact on comparative performance.Subject device has a larger internal volume (314 ml for adult vs. predicate's 198 ml for adult), but this difference does not impact comparative performance and provides more access.

    2. Sample size used for the test set and the data provenance

    The document does not specify a "sample size" in terms of number of patients or clinical samples because it describes non-clinical comparative performance testing using simulated patient settings. Therefore, there is no data provenance in terms of country of origin or retrospective/prospective study design as this was not a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The study was non-clinical and did not involve human experts establishing ground truth for a test set. Performance metrics were based on simulated parameters and physical measurements.

    4. Adjudication method for the test set

    This information is not applicable as there was no test set requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This device is an Oxygen Mask with Gas Sampling, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a medical device, not an algorithm. The performance testing was for the device itself in simulated environments.

    7. The type of ground truth used

    For %CO2 accuracy and Respiration rate, the "ground truth" was established by simulated patient settings at different breath rates, tidal volumes, and known CO2 concentrations.

    For Biocompatibility, the "ground truth" was established by standards such as ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 18562-2.

    8. The sample size for the training set

    This information is not applicable. There was no "training set" as this is not a machine learning or AI-based device.

    9. How the ground truth for the training set was established

    This information is not applicable as there was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062104

    Validate with FDA (Live)

    Device Name
    TALKING TRACH
    Date Cleared
    2007-04-20

    (270 days)

    Product Code
    Regulation Number
    868.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EMS Talking Trach tracheostomy tube, cuffed, is intended to provide tracheal access for airway management of tracheostomized patient. The device allows tracheostomy patients with a functional larynx and an unobstructed upper airway to vocalize / phonate. The Inner Speech cannula allows vocalization with the tracheostomy tube's cuff to remain inflated for ventilator dependent patients. For adult patients > 30 kg.

    Device Description

    The proposed EMS Talking Trach is a tracheostomy tube which allows patients who are ventilator dependent or spontaneously breathing to speak or phonate. It can be used with patients on ventilators while the cuff is "up" or inflated. It is offered in sizes from 4.0 to 10.0 mm with disposable standard and speaking inner cannula.

    AI/ML Overview

    The provided text describes the EMS Talking Trach, a tracheostomy tube designed to allow patients to speak or phonate, even while on a ventilator with the cuff inflated. The submission focuses on demonstrating substantial equivalence to predicate devices rather than presenting a study with detailed acceptance criteria and performance data for a novel device.

    Therefore, the requested information about acceptance criteria and a study proving the device meets them cannot be fully extracted in the typical sense for a diagnostic or AI device. However, based on the provided document, here's an interpretation of the performance claims and the study design where applicable:

    1. A table of acceptance criteria and the reported device performance

    The submission claims "substantial equivalence" as its primary acceptance criterion. The performance attributes mentioned are expected to be "substantially equivalent to predicate" for the EMS Talking Trach.

    Acceptance CriterionReported Device Performance (Claimed)
    Ventilator Performance:
    Peak Pressure (ventilator)Substantially equivalent to predicate when testing at various flow rates.
    Peak Pressure (intra-pulmonary)Substantially equivalent to predicate when testing at various flow rates.
    Expiratory Flow (average)Substantially equivalent to predicate.
    Expiratory Flow (Peak)Substantially equivalent to predicate.
    General EquivalenceThere are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices (Passy Muir – K962714, Portex – Trach Talk – K972385, and Shiley – Trach tubes – K962173).
    Standards ComplianceComply with ISO 5366.1, 5366-3.

    2. Sample size used for the test set and the data provenance

    The document does not describe a clinical study with a "test set" in the context of device performance evaluation based on patient data for diagnostic accuracy. The performance claims are based on engineering tests for ventilator compatibility rather than human subject data. Therefore, details like data provenance or test set sample sizes for human subjects are not provided. The testing appears to be primarily bench testing (in vitro simulation) to assess ventilator performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The submission focuses on physical and functional equivalence to predicate devices, not on a diagnostic task requiring expert-established ground truth. The "ground truth" for ventilator performance would be established by the physical measurements from testing equipment.

    4. Adjudication method for the test set

    Not applicable, as no human-reviewed test set data for diagnostic performance is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical instrument (tracheostomy tube), not an AI-powered diagnostic tool, and therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the ventilator performance testing, the "ground truth" would be the objectively measured values of peak pressure and expiratory flow from the testing equipment, compared against the values obtained from the predicate devices under similar conditions. This is a form of engineering/bench test data ground truth.

    8. The sample size for the training set

    Not applicable. There is no training set in the context of machine learning or AI algorithms for this device submission.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023683

    Validate with FDA (Live)

    Date Cleared
    2003-12-01

    (395 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient interface accessory for use with CPAP and bi-level systems used in the treatment of adult OSA and / or ventilatory support. A minimum pressure of > 3.0 cm H₂O at the mask is required. Single patient, multi-use

    Device Description

    The EMS Multi-strap Full Face mask covers both the nose and mouth and includes a nonrebreathing / anti-asphyxia valve, which is activated under flow / pressure from a CPAP or bilevel ventilator. It is open to ambient air when the ventilator is not ON allowing the patient to breath ambient air. It has a quick release mask harness system. It is single patient, multi-use.

    AI/ML Overview

    The information provided describes the Engineered Medical Systems, Inc. Multi-strap Full Face Mask and its 510(k) summary for clearance. However, the document does NOT contain a study that proves the device meets specific acceptance criteria with reported device performance metrics in the way a clinical or performance study for an AI/ML medical device would.

    Instead, this 510(k) summary focuses on demonstrating substantial equivalence to predicate devices for a physical medical device (a CPAP mask). The "acceptance criteria" here are characteristics and design features that are compared to legally marketed predicate devices, rather than performance metrics from a formal study with a test set, ground truth, or expert review as typically seen in AI/ML device submissions.

    Therefore, many of the requested fields cannot be directly answered from the provided text, as they pertain to performance studies not conducted or reported in this type of submission.

    Here's the breakdown based on the provided text, addressing the points where information is available or inferable within the context of a substantial equivalence claim for a physical device:


    1. Table of Acceptance Criteria and the Reported Device Performance

    For this type of device, "acceptance criteria" are typically defined by functional specifications and comparison to predicate devices, demonstrating that the new device performs as intended and is as safe and effective. The "reported device performance" is a demonstration that these specifications are met and that the device is substantially equivalent to predicates.

    Attribute / Acceptance Criteria (as per functional specifications and comparison to predicates)Reported Device Performance / Status (as per 510(k) summary)
    Indications for use: A patient interface accessory for use with CPAP and bi-level systems used in the treatment of adult OSA and / or ventilatory support. A minimum pressure of > 3.0 cm H₂O at the mask is required.Meets this indication, as it is the stated intended use. Substantially equivalent to predicates for this use.
    Single patient, multi-useYes (device is designed for this)
    PrescriptionYes (device is intended for prescription use)
    Intended population: Any patientMeets this intention.
    Intended Environment of Use: Hospital, Sub-acute Institutions, HomeMeets this intention.
    Mask covers nose and mouthYes (design feature)
    Quick release mask harnessYes (design feature)
    Non-rebreathing / anti-asphyxia valveYes (design feature)
    Must be used with exhalation valve in circuitYes (design feature/requirement)
    Open to ambient when ventilator offYes (design feature)
    Valve opens at > 3 cm H₂OYes (design feature, aligns with minimum pressure)
    Can be cleanedYes (design feature, implies reusability for single patient)
    Materials: Mask cone and Elbow - PC; Mask cushion - PVC; Flap valve - SiliconeYes (materials specified and used)
    Performance Standards: None under Section 514No specific 514 standards apply or are cited as met.
    Overall Safety and EffectivenessDemonstrated safe and effective, substantially equivalent to predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The submission focuses on substantial equivalence based on design, materials, and intended use comparison to predicate devices, not on a formal clinical or performance test study with a "test set" in the context of an AI/ML device.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. There was no "test set" requiring expert ground truth in the context of this 510(k) for a physical medical device.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No "test set" requiring adjudication was reported.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/provided. This is a submission for a physical medical device (CPAP mask), not an AI/ML diagnostic tool, and therefore no MRMC study would be relevant or expected.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/provided. This is a physical medical device; there is no "algorithm" or "standalone performance" in the context described.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not applicable/provided. For this physical medical device, the "ground truth" for demonstrating equivalence largely relies on engineering specifications, material properties, and comparison of functional design features to existing legally marketed devices.


    8. The sample size for the training set

    This information is not applicable/provided. This document does not describe a training set as would be relevant for an AI/ML device.


    9. How the ground truth for the training set was established

    This information is not applicable/provided. This document does not describe a training set or its ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013728

    Validate with FDA (Live)

    Device Name
    TRACH VOX
    Date Cleared
    2002-07-30

    (263 days)

    Product Code
    Regulation Number
    868.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For patients breathing spontaneously via a tracheostomy tube, the TrachVox provides heat and humidity for entrained air and incorporates a cough relief and suction port valve. The unit may be closed by the patient's finger, which directs the air through the vocal cords to facilitate speech.

    For patients breathing spontaneously via an endotracheal tube (ET tube) the TrachVox provides heat and moisture for the inhaled air.

    Device Description

    The TrachVox connects to a 15 mm tracheostomy or endotracheal tube connector of a spontaneously breathing patient. It incorporates four (4) key features:

    • HME media for humidifying inhaled air
    • Suction port
    • Method for closing off exhalation to outside and divert the exhaled breathe around the tracheostomy tube and through the vocal cords for speech
    • Connector for delivery of supplemental oxygen
    AI/ML Overview

    The provided text describes a medical device called TrachVox, a Tracheal HME with speech valve and suction port. The submission is a 510(k) premarket notification and focuses on establishing substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a dedicated study.

    Therefore, many of the requested sections (acceptance criteria table, sample sizes, expert qualifications, adjudication, MRMC studies, standalone performance, ground truth types for test/training sets, and training set details) are not directly applicable or available in this type of submission.

    However, I can extract information related to the device's technical specifications and how its performance is compared to predicate devices, which implicitly serves as the "study" for establishing substantial equivalence.

    Here's a breakdown of the available information structured to answer your request where possible, and indicating when information is not present:


    Acceptance Criteria and Study Details for TrachVox

    1. Table of Acceptance Criteria and Reported Device Performance

    This 510(k) submission establishes substantial equivalence by comparing the TrachVox device's attributes and performance characteristics to those of legally marketed predicate devices. The "acceptance criteria" here are implicitly defined by the attributes and performance ranges of these predicate devices, which the TrachVox aims to match or perform comparably to.

    AttributePredicate Device Performance (Implicit Acceptance Criteria)Reported TrachVox Device Performance
    For spontaneously breathing tracheostomy patientsYesYes
    Incorporated speech valveYesYes
    Provides heat and humidityYesYes
    Can provide supplemental oxygenYesYes
    Intended to connect to tracheostomy tube of spontaneously breathing patientYesYes
    Can connect to an ET tube of spontaneously breathing patientYesYes
    Intended for single patient up to 24 hoursYesYes
    PrescriptionYesYes
    Intended populationAny patientAny patient
    Intended Environment of UseHome, Hospital, sub-acute, ICUHome, Hospital, sub-acute, ICU
    Placement in various locations in circuitYesYes
    Supplemental oxygen connectionYesYes
    Standard 15/22 mm connectorsYesYes
    Dead Space (ml)(Not explicitly stated for predicates, implied to be comparable)9.5 ml
    Resistance to flow(Not explicitly stated for predicates, implied to be comparable)0.23 cm H₂O @ 30 Lpm
    0.9 cm H₂O @ 60 Lpm
    Weight(Not explicitly stated for predicates, implied to be comparable)2.9 gm
    Normally closed suction slit valveYesYes
    Speech valve normally open for inhalationYesYes
    Allows exhalation to atmosphere freelyYesYes
    Method for diverting exhaled air through vocal cord for facilitating speech (Speech valve)Mechanically similar mechanisms on predicatesCover suction valve with fingers, depress housing, compressing foam and closes off exhalation to atmosphere vents
    Method for revert to normal exhalation after speakingMechanically similar mechanisms on predicatesRemove finger, foam spring-action raises housing, opening exhalation to atmosphere vents
    Moisture output (mg H₂O/l) ISO 9360(Not explicitly stated for predicates, but implies compliance with ISO 9360 standards)20.1 mg H₂O /l @ 500 ml TV
    Moisture output (mg H₂O/l) ISO 9360 with supplemental oxygen(Not explicitly stated for predicates, but implies compliance with ISO 9360 standards)14.9 mg H₂O /l @ 10 Lpm
    Housing components - polypropyleneYesYes
    HME mediaFoamFoam
    ValveSiliconeSilicone
    None under Section 514YesYes
    ISO 5356-1 Conical 15/22YesYes
    ISO 9360 - HME moisture outputYesYes

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the "acceptance criteria" (defined by substantial equivalence to predicate devices) is primarily a comparative analysis of specifications and functional features. For quantitative metrics like Dead Space, Resistance to Flow, Weight, and Moisture Output, it is implied that bench testing was conducted on the TrachVox device to obtain these performance values, and these values were deemed acceptable when compared to relevant predicate devices or industry standards (like ISO 9360).

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not applicable. This submission relies on engineering specifications and bench testing of the single device model, not a patient-based "test set" in the context of clinical trials or AI/diagnostic studies.
    • Data Provenance: The data primarily comes from engineering design specifications and bench testing of the TrachVox device itself, and comparison to publicly available specifications/regulatory submissions of predicate devices. The country of origin for the data generation would be the manufacturer, Engineered Medical Systems, in Indianapolis, IN, USA. The data is prospective in the sense that the new device's performance was measured.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is not a study involving human interpretation of data where "ground truth" is established by experts.

    4. Adjudication method for the test set:

    • Not applicable. No expert adjudication was done for this type of submission.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done:

    • No, an MRMC comparative effectiveness study was not done.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a mechanical medical device, not an AI algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the TrachVox's performance metrics (dead space, resistance, moisture output) would be derived from direct physical measurements and laboratory bench testing according to established engineering and ISO standards (e.g., ISO 9360 for HME moisture output, ISO 5356-1 for connectors). For functional features, the "ground truth" is its design and operational mechanism. For the comparison to predicate devices, the "ground truth" is the regulatory and published specifications of those predicate devices.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning study.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/machine learning study.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013122

    Validate with FDA (Live)

    Date Cleared
    2002-04-19

    (212 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use with ventilators, anesthesia machines, and open flow systems where filtration of inspired and / or expired gases is desired and to add maintain and retain moisture for the exhaled breathe of the patient. Model with tidal volumes ranging from 20-70 cc for neonate, 70-250 cc for pediatric and >150 cc for adults. Use up to 24 hours.

    Device Description

    The EMS Filter and combined Filter / HME are available in multiple sizes and shapes, rectangular and round, and incorporate standard 15 / 22 mm connectors with a gas sampling luer port. The depth filter uses electrostatic media for filtration and a foam media for the HME media.

    AI/ML Overview

    Here's an analysis of the provided text regarding the EMS Electra Filter and Filter/HME device, focusing on acceptance criteria and the study proving it meets those criteria:

    The provided document is a 510(k) Premarket Notification, which is a submission to the FDA to demonstrate that the device to be marketed is substantially equivalent to a legally marketed predicate device. This type of submission relies on comparing the new device's characteristics and performance to those of existing, legally marketed devices, rather than comprehensive clinical trials for novel devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly define "acceptance criteria" as a separate, formal list against which the device was tested as one might see in a clinical trial. Instead, it presents general technical characteristics of the proposed device and demonstrates their equivalence to predicate devices. The "acceptance criteria" are implicitly defined by the performance of the predicate devices and existing standards for such devices.

    AttributeAcceptance Criteria (Implied by Predicate/Standard)Reported Device Performance (EMS Electra)
    Dead Space (ml)- (Should be comparable to predicate devices)10 to 65 ml
    Resistance to flow (Adult)≤3 cm H2O @ 60 lpm≤3 cm H2O @ 60 lpm
    Resistance to flow (Pediatric)≤1.0 cm H2O @ 20 lpm1.0 cm H2O @ 20 lpm
    Resistance to flow (Neonate)≤0.5 cm H2O @ 5 lpm0.5 cm H2O @ 5 lpm
    Bacterial Filtration Efficiency (BFE)99.999+% (Based on Nelson Lab. results for predicate)99.999+%
    Viral Filtration Efficiency (VFE)99.99+% (Based on Nelson Lab. results for predicate)99.99+%
    Weight (gm)- (Should be comparable to predicate devices)10 to 35 gm
    Humidification Output (Adult, 1000cc TV)32 mg H2O/L32 mg H2O/L
    Humidification Output (Pediatric, 250cc TV)32 mg H2O/L32 mg H2O/L
    Humidification Output (Neonate, 50cc TV)30 mg H2O/L30 mg H2O/L
    Tidal Volume Range (Neonate)20-70 cc20-70 cc
    Tidal Volume Range (Pediatric)70-250 cc70-250 cc
    Tidal Volume Range (Adult)>150 cc>150 cc
    Standards ComplianceISO 5356-1, ISO 594-2, ISO 9360Yes (Compliance asserted)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document details performance data derived from laboratory testing, specifically mentioning "Nelson Lab." for Bacterial and Viral filtration efficiency. This indicates the data is likely from professional testing laboratories rather than human clinical trials.

    • Sample Size: Not explicitly stated for each test. For filtration efficiency, usually multiple samples of the filter media are tested. For resistance and humidification, often a set number of devices are tested according to a standard protocol.
    • Data Provenance: The document states "Nelson Lab." for BFE and VFE, implying a specialized testing facility. The country of origin is not specified, but Nelson Labs is a well-known US-based testing laboratory. The data is prospective in the sense that the new device was fabricated and then tested to determine its performance characteristics. It is not retrospective in the sense of analyzing past patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not applicable as the evaluations are based on objective physical and biological performance tests, not subjective expert assessment of images or patient outcomes. The "ground truth" for these tests are the established scientific methodologies and standards by which filter performance, resistance, and humidification are measured.

    4. Adjudication Method for the Test Set:

    This is not applicable as the evaluation relies on direct measurement of physical properties and performance characteristics, not on subjective assessments that would require adjudication (e.g., medical image interpretation).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This is not applicable. The device is a filter/heat and moisture exchanger (HME), not an imaging or diagnostic AI system. Therefore, an MRMC study related to human reader improvement with AI assistance is irrelevant to this device.

    6. Standalone (Algorithm Only) Performance:

    This is not applicable. The device is a physical medical device (a filter), not an algorithm or AI system.

    7. Type of Ground Truth Used:

    The ground truth used is objective laboratory measurement data and established performance standards. For example:

    • For resistance, the ground truth is the measured pressure drop across the filter at specified flow rates.
    • For filtration efficiency, the ground truth is the measured percentage of bacteria/viruses removed according to standardized laboratory testing protocols (e.g., using bacterial aerosols like Staphylococcus aureus).
    • For humidification output, the ground truth is the gravimetric measurement of water vapor produced by the HME under specific temperature, humidity, and tidal volume conditions as per ISO 9360.

    8. Sample Size for the Training Set:

    This is not applicable. Since this device is a physical product and not an AI/machine learning model, there is no "training set." Its design and performance capabilities would be based on engineering principles, materials science, and testing iterations, not data training.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable for the same reason as point 8. The "ground truth" for the device's design would stem from engineering specifications, industry standards (like ISO for connectors and HME moisture output), and the performance characteristics observed in predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013123

    Validate with FDA (Live)

    Date Cleared
    2002-03-14

    (176 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use with pulmonary function testing. To filter air between the patient's exhaled air and the testing equipment. Single patient use.

    Device Description

    The EMS PFT Filter is acompact, electrostatic filter with various end-fitting adaptable to various pulmonary function testing circuits. It has 75 ml deadspace and resistance of 07. cm H2O at 720 Ipm per ATS spirometry or 0.5 cm I-120 @ 60 lpm. There are various connectors to allow connection to various PFT equipment. Single patient use.

    AI/ML Overview

    The provided text describes a Premarket Notification 510(k) for the EMS PFT Filter. This type of submission is for demonstrating substantial equivalence to a predicate device, not typically for reporting on a study that establishes performance against predefined acceptance criteria for a novel device.

    However, the document does list performance attributes for the EMS PFT Filter and compares them to those of a predicate device (PDS KoKo K934475), which can be interpreted as demonstrating that the EMS device meets performance levels comparable to a legally marketed device.

    Here's an attempt to structure the information according to your request, with the understanding that this is a 510(k) submission, and the "study" is more of a comparison against a predicate, rather than a de novo clinical trial with specific acceptance criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given this is a 510(k) for substantial equivalence, the "acceptance criteria" are derived from the performance of the predicate device. The EMS PFT Filter is considered acceptable if its performance is comparable to or better than the predicate's.

    AttributeAcceptance Criteria (Predicate Device K934475)Reported Device Performance (EMS PFT Filter Model #5813)
    Intended useFor use with pulmonary function testing. To filter air between the patient's exhaled air and the testing equipment.Same
    Intended for single patientYesYes
    PrescriptionYesYes
    Intended populationAny patientSame
    Intended Environment of UseHospital, sub-acuteSame
    Can be used with several different PFT machinesYesYes
    Design Features - Compact housingYesYes
    Design Features - Various end-fittingsYesYes
    Dead Space (ml)60 ml75 ml
    Resistance to flow at 720 lpm (per ATS standard for spirometry)<1.5 cm H2O0.7 cm H2O
    Resistance to flow at 60 lpm<1.5 cm H2O0.5 cm H2O
    Bacterial filtration99.99+%99.9999%
    Viral filtration99.99+%99.999+%
    Housing materialPolystyrenePolystyrenc
    Filter mediaElectrostatic polypropyleneElectrostatic polypropylene

    Note on Differences: The dead space of the EMS PFT Filter (75 ml) is higher than that of the predicate (60 ml). However, the resistance to flow (0.7 cm H2O at 720 lpm and 0.5 cm H2O at 60 lpm) is better (lower) than the predicate's criteria of <1.5 cm H2O. The bacterial and viral filtration efficiencies are also better (higher) for the EMS filter. The document states, "There are no significant differences between the intended device and the predicates." This implies that the differences observed (like dead space, which is only slightly higher, and improved filtration/resistance) are not considered clinically significant enough to preclude substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a traditional "test set" in the context of a clinical trial or algorithm validation. The performance data presented (e.g., dead space, resistance, filtration efficiency) are typically derived from laboratory testing of the device itself.

    • Sample size: Not specified. It would usually involve testing a certain number of manufactured units to ensure consistency, but this is not detailed in the summary.
    • Data provenance: The data would originate from bench testing performed by the manufacturer, Engineered Medical Systems, Inc. The country of origin is implied to be the USA (Indianapolis, IN). The data is prospective in the sense that it's generated for the device being submitted, but not from a prospective patient study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. For a medical device like a PFT filter, "ground truth" for its physical performance (like filtration efficacy or resistance) is established through standardized engineering and microbiological testing methods, not by expert consensus on clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. As described in point 3, this type of device performance is measured objectively through specified tests, not through expert adjudication of images or clinical outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is relevant for diagnostic imaging or AI algorithms where human readers interpret cases. This device is a passive filter; its performance is measured mechanistically, not through human interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device is a physical filter, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance metrics (e.g., bacterial/viral filtration, resistance, dead space) is established by standardized laboratory testing methods against established industry or regulatory standards (e.g., ATS spirometry standards mentioned for resistance). These are objective physical measurements.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013089

    Validate with FDA (Live)

    Date Cleared
    2002-03-14

    (178 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is desired and to add maintain and retain moisture for the exhaled breathe of the patient

    Device Description

    The EMS HEPA Filter and combined HEPA filter / HME are available in multiple sizes and shapes, recogular and round, and incorporate standard 15 / 22 mm connectors with a gas sampling luer port. recoalignar and round, and morporativer for filtration and a foam media for the HME media. Each filter is individually tested for rating performance according to BS 3928 Sodium Flame for Air filters. The "HEPA" performance was in accordance to DOE 3202-97 and ASTM D2986 – DOP.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the EMS HEPA Filter and HEPA Filter / HME, which are bacterial/viral filters and heat and moisture exchangers. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a full study report with detailed performance metrics against those criteria.

    However, the "General Technical Characteristics" table [1] lists several performance attributes of the proposed devices, which can be interpreted as acceptance criteria based on established standards and the characteristics of predicate devices. The "Reported Device Performance" column reflects the device's measured or designed specifications.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" and "reported device performance" in separate columns, nor does it present a formal study showing the device meets these criteria with numerical results from a specific test. Instead, it lists technical characteristics, which by their nature in a 510(k) summary, imply that the device performance meets or is equivalent to these values for substantial equivalence.

    Acceptance Criteria (Implied from "General Technical Characteristics")Reported Device Performance (Implied from "EMS - Proposed Devices" column)
    Indications for use - To filter inspired and / or expired gases.Same (as predicate)
    Intended for single patient, up to 24 hoursYes
    PrescriptionYes
    Intended populationAny patient
    Intended Environment of UseHome, Hospital, sub-acute, Emergency services
    Placement in various locations in circuitYes
    Gas sampling portYes
    Standard 15/22 mm connectorsYes
    Dead Space (ml)45 to 80 ml
    Resistance to flow< 3.4 cm H2O @ 60 lpm
    Bacterial filtration - BFE - Nelson Lab.99.9999%
    Viral filtration - VFE - Nelson Lab.99.9999%
    Weight (gm)27 to 35 gm
    Humidification output (mg H2O/l)27 mg H₂O /L at TV of 500 cc, where applicable
    Housing materialPolystyrene
    Filter mediaPaper fiber
    Performance Standards (Compliance)
    ISO 5356-1 Conical 15/22Yes
    ISO 594-2 Luer FittingsYes
    ISO 9360 - HME moisture outputYes
    DOE 3202-97 and ASTM D2986 - DOP99.97% (performance, implies compliance)

    2. Sample size used for the test set and the data provenance

    The document states that "Each filter is individually tested for rating performance according to BS 3928 Sodium Flame for Air filters." and "The "HEPA" performance was in accordance to DOE 3202-97 and ASTM D2986 – DOP." [0]. For BFE and VFE, it mentions "Nelson Lab." [1].

    • Sample Size: The phrase "Each filter is individually tested" suggests that all manufactured filters undergo performance testing, but it doesn't specify the sample size of filters used for the validation testing that led to the reported performance figures in the table. It refers to ongoing quality control rather than a specific study sample size.
    • Data Provenance: The testing for BFE and VFE was performed by "Nelson Lab." [1]. Other tests refer to specific standards (BS 3928, DOE 3202-97, ASTM D2986), implying these tests were conducted internally or by a contracted lab to meet those standards. The country of origin of the data is not explicitly stated, but the company is based in Indianapolis, IN, 46241, USA [0], and Nelson Labs is a US-based laboratory, suggesting US provenance. The data appears to be prospective in the sense of being generated specifically for the device's design verification and validation, as opposed to retrospective analysis of existing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a physical medical device (filter/HME) and its performance metrics (e.g., filtration efficiency, resistance to flow, dead space) are determined by objective physical and biological testing methods against established standards, not by expert interpretation or consensus on images or clinical cases.

    4. Adjudication method for the test set

    Not applicable, as the tests are objective measurements against standards, not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical filter/HME, not an AI-based diagnostic or assistive technology for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical filter/HME, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the performance claims are the established international and national standards for medical device filters and HMEs. These include:

    • BS 3928 Sodium Flame for Air filters [0]
    • DOE 3202-97 and ASTM D2986 – DOP for HEPA performance [0, 1]
    • ISO 5356-1 Conical 15/22 [1]
    • ISO 594-2 Luer Fittings [1]
    • ISO 9360 - HME moisture output [1]
    • Bacterial and Viral Filtration Efficiency testing methods (likely based on ASTM F2101 for BFE and ASTM F2100 for VFE, implicitly performed by Nelson Labs as this is their specialty expertise and standard industry practice for these claims). [1]

    These standards define the test methods and acceptable performance ranges.

    8. The sample size for the training set

    Not applicable. This device is a physical medical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K003497

    Validate with FDA (Live)

    Date Cleared
    2000-11-30

    (17 days)

    Product Code
    Regulation Number
    868.2600
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K000215

    Validate with FDA (Live)

    Device Name
    SWIVEL ELBOW
    Date Cleared
    2000-05-26

    (123 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K000314

    Validate with FDA (Live)

    Date Cleared
    2000-04-12

    (71 days)

    Product Code
    Regulation Number
    868.5690
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2