Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K242996

    Validate with FDA (Live)

    Device Name
    EndyMed PRO MAX
    Date Cleared
    2025-06-10

    (257 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PRO MAX system is a Radiofrequency device intended for use in dermatologic procedures with separate indications for each handpiece.

    The TC Handpieces (Shaper Max, Mini Shaper Max, iFine Max) are intended for:
    • Relief of minor muscle aches and pain, relief of muscle spasm
    • Temporary improvement of local blood circulation

    When used up to 65W, the Mini Shaper Max and Small Max handpieces are also intended for the non-invasive treatment of mild to moderate facial wrinkles and rhytides.

    The FSR Max Handpiece is intended for dermatological procedures requiring ablation and resurfacing of the skin.

    The Intensif MAX Handpiece is intended for use in dermatologic procedures for electrocoagulation and hemostasis.

    The EndyMed Contour MAX Handpiece is intended for the treatment of the following medical conditions; using the Contour MAX applicator for delivery of non-thermal RF combined with massage:
    • Relief of minor muscle aches and pain, relief of muscle spasm
    • Temporary improvement of local blood circulation
    • Temporary reduction in the appearance of cellulite

    Device Description

    ENDYMED PRO™ MAX is a computerized system that generates pulses of radio frequency (RF) energy that are emitted onto patient skin. The device consists of a console and seven handpieces that emit either ablative or non-ablative radiofrequency.

    AI/ML Overview

    The provided document is a 510(k) clearance letter for the EndyMed PRO MAX device. It indicates that the device has been cleared by the FDA, but it primarily focuses on establishing "substantial equivalence" to predicate devices rather than directly presenting the results of a clinical study designed to prove the device meets specific acceptance criteria based on performance endpoints.

    The document describes performance testing but does not provide a table of acceptance criteria alongside detailed "reported device performance" in the context of a clinical study or specific performance metrics for the indications for use. Instead, it justifies substantial equivalence through comparison of specifications and generic performance testing.

    Here's an attempt to extract the requested information based on what is available in the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly state quantitative acceptance criteria for each indication (e.g., a specific percentage reduction in pain, or a defined improvement in wrinkles). Instead, it focuses on demonstrating that the device can achieve and maintain a therapeutic temperature similar to its predicate devices, which is presumed to lead to the intended effects. Therefore, the "Acceptance Criteria" here are inferred from the safety and performance testing described.

    Acceptance Criteria (Inferred from testing)Reported Device Performance (from "Tissue Heating Testing")
    Ability of handpiece to maintain therapeutic temperature of 39-45°C.Each handpiece (Shaper Max, iFine Max, Mini Shaper Max, and Contour Max) can maintain a therapeutic temperature of 39-45°C for at least 10 minutes of treatment.
    Bio-compatibility of patient-contacting materials.All handpieces have been tested and determined to be biocompatible.
    Compliance with electrical safety and EMC standards.Successfully completed IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2.
    Software verification and validation.Successfully completed software verification and validation testing in accordance with FDA guidance.
    Accuracy of RF energy output.Bench testing performed to verify the accuracy of the output of RF energy.
    Sterilization process effectiveness (for specific handpieces where applicable).Complied with ISO 11135 for sterilization of health-care products.
    Packaging integrity for sterilized devices.Complied with ISO 11607-1 for packaging for terminally sterilized medical devices.
    Risk management application.Complied with EN ISO 14971 for application of risk management to medical devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • For Tissue Heating Testing: "3 subjects with 3 anatomical sites each." (Total of 9 test sites across 3 subjects).
      • For other tests (Electrical Safety, EMC, Software, Biocompatibility, etc.): The document does not specify sample sizes in terms of human subjects or data sets. These are typically engineering or lab-based tests on the device itself or materials.
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective for the Tissue Heating Testing. Given the context of a 510(k) summary, these tests are typically part of a prospective verification and validation effort by the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish ground truth for the "test set" in the traditional sense of clinical endpoints for the indications. The tissue heating test measures temperature, which is an objective physical parameter.


    4. Adjudication Method for the Test Set

    The document does not mention any adjudication method like 2+1 or 3+1. This is expected as the described performance testing (tissue heating, electrical safety) involves objective measurements rather than subjective expert assessment.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done according to the provided document. The submission focuses on substantial equivalence based on technical specifications and bench/performance testing, not clinical comparative effectiveness with human readers.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device described is a physical medical device (Radiofrequency system with handpieces). While it contains software, the "performance testing" described (e.g., tissue heating) refers to the device's physical output characteristics, not an AI algorithm's standalone diagnostic or detection performance. Therefore, the concept of "standalone performance" for an AI algorithm is not applicable in the context presented.


    7. The Type of Ground Truth Used

    For the "Tissue Heating Testing," the "ground truth" is the measured temperature (39-45°C) achieved and maintained on the tissue. This is an objective, physiological measurement, not expert consensus, pathology, or outcomes data in a clinical trial sense for efficacy.


    8. The Sample Size for the Training Set

    The document does not mention a training set. This is because the device is a physical RF energy delivery system, and the "performance testing" described is for verification of its physical properties and safety, not for training a machine learning or AI algorithm.


    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161715

    Validate with FDA (Live)

    Device Name
    EndyGel
    Date Cleared
    2017-02-07

    (231 days)

    Product Code
    Regulation Number
    882.1275
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endy Gel is an Electroconductive gel media used with external electrode to reduce the impedance to alternating current) of the contact between the electrode surface and the skin.

    Device Description

    EndyMed's EndyGel™ is an Electroconductive gel media used with external electrode to enhance electrical conductivity by reducing the impedance (resistance to alternating current) of the radiofrequency signals path from the electrode surface between the electrode surface and the skin.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for EndyGel™, an electroconductive gel media. The purpose is to demonstrate substantial equivalence to a predicate device, Electro-Gel (K111717).

    Here's the analysis of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from Predicate Device)Reported Device Performance (EndyGel™)
    Device Class: Class IIClass II
    Classification Panel: NeurologyNeurology
    Product Code: GYBGYB
    Regulation Description: Electroconductive mediaElectroconductive media
    Regulation number: 21 CFR 882.127521 CFR 882.1275
    Indications for Use: Electro-Gel device is intended for use with external electrodes as the conductor between the scalp and the electrodes. It also reduces impedance (resistance to alternating current) between the electrode surface and the skin.EndyMed's EndyGel™ is an Electroconductive gel media used with external electrode to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin.
    Principles of Operation: Functions as a conductor between the skin and the electrodes. Intended to be used prior to Handpiece's electrodes placement to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin. The energy is then transferred via the gel to the patient's tissue.Same as predicate.
    Body contact: Intact SkinIntact Skin
    Biocompatibility: Complies with requirements of ISO 10993-1, including: Cytotoxicity (ISO 10993-5), Irritation and Sensitization (ISO 10993-10)Complies with requirements of ISO 10993-1, including: Cytotoxicity (ISO 10993-5), Irritation and Sensitization (ISO 10993-10)
    Sterilization: Provided non-sterileProvided non-sterile
    Shelf life: 3 years2 years
    Impedance (at 1 MHz): 549.71527.68
    Conductive material: Water (Aqua) with immersed Sodium chlorideWater (Aqua) with immersed Triethanolamine
    Composition: Water (solvent), Sodium Chloride (preservative), Aragum T-1998 (thickener), Potassium Bitartrate (Stabilizer), Glycerin (lubricant), Methylparaben (preservative), Propylparaben (preservative)Water (solvent), Carbomer (gel forming), Triethanolamine (Buffering), Methylisothiazolinone and Methylcloroisothiazolinone (preservative)
    pH: 6.997.15-7.33

    Study proving device meets acceptance criteria:

    The safety and efficacy of the EndyGel™ were established by a series of performance tests, including biocompatibility testing, electrical performance, and stability.

    • Biocompatibility testing:
      • Complies with ISO 10993-1, ISO 10993-5 (Cytotoxicity), and ISO 10993-10 (Irritation and Sensitization).
    • Electrical Performance testing:
      • The impedance of EndyGel™ (527.68 at 1 MHz) is within 4% of the predicate device (549.71).
    • Stability testing:
      • The EndyGel™ has a shelf life of 2 years (compared to 3 years for the predicate).
    • pH:
      • The pH of EndyGel™ (7.15-7.33) is within 2.2% of the predicate device (6.99).

    The document states that "The difference in total impedance and in pH between the two gels is negligible and do not impact the safety and effectiveness of the EndyGel." Also, "V&V and compliance testing, including: Biocompatibility testing, Electrical Performance testing and Stability testing, were successfully conducted and did not raise any new safety questions or identify any new risks."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size for the test set or the data provenance for the performance tests (biocompatibility, electrical performance, stability). It broadly refers to "performance tests" that were successfully conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is an electroconductive gel, and the "ground truth" for its performance is based on quantifiable physical and chemical properties and standard biological safety tests, not human expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human interpretation or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm-based device. Performance is based on intrinsic physical, chemical, and biological properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the acceptance criteria for this device are established by:

    • Industry standards (ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 13485, ISO 14971-1).
    • Quantifiable physical properties (impedance, pH, shelf life).
    • Chemical composition analysis.
    • Comparison to an existing legally marketed predicate device with a history of safe and effective use.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161199

    Validate with FDA (Live)

    Date Cleared
    2016-08-01

    (95 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage:

    • · Relief of minor muscle aches and pain, relief of muscle spasm
    • · Temporary improvement of local blood circulation
    • · Temporary reduction in the appearance of cellulite.
    Device Description

    EndyMed's EndyMed Contour Handpiece is a treatment handpiece to be attached to the FDA cleared EndyMed Imagine TC Skin Treatment System (Imagine K08346). The EndyMed Contour Handpiece combines suction module that creates vacuum energy and emits bipolar RF energy that flows between the electrodes to create thermal heating of the tissue.

    The Vacuum is used for the massaging of deep tissues by creating mild to deep suction. The vacuum massage contributes to the sub-dermal and adipose tissues shrinkage and improves the contact surface between electrodes and tissue.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the EndyMed Contour Handpiece. It details the device's intended use, technological characteristics, and a claim of substantial equivalence to predicate devices. However, this document does not contain the specific information requested regarding acceptance criteria for an AI/ML device study, and the detailed study description you've outlined.

    The document discusses performance tests for the EndyMed Contour Handpiece itself (e.g., power output, temperature, vacuum pressure), but this is validation of a physical medical device, not a study proving an AI/ML algorithm meets acceptance criteria.

    Here's why the requested information cannot be found in the provided text:

    • No AI/ML Component: The EndyMed Contour Handpiece is described as an electrosurgical cutting and coagulation device accessory that uses RF energy and vacuum mechanical manipulation. There is no mention of an AI/ML component or algorithm in its description or intended use.
    • Study Type: The "performance data" and "summary of clinical performance data" refer to lab performance tests, design validation, and software verification and validation for the physical device, and a claim that no clinical studies were necessary to show substantial equivalence with its predicate devices because it has the same intended use, clinical indication, and technology. This explicitly states new clinical studies were not performed for this submission.
    • Ground Truth: Since no clinical studies were performed, there is no discussion of establishing ground truth, expert consensus, pathology, or outcomes data for a test set.
    • Sample Sizes/Experts/Adjudication/MRMC: These are all concepts relevant to clinical or performance studies evaluating an AI/ML algorithm's effectiveness against a ground truth, which were not conducted for this submission.

    Therefore, I cannot populate the table or answer the specific questions about acceptance criteria and study proving device meets acceptance criteria, as the provided text does not describe such a study for an AI/ML-driven medical device.

    The document is a regulatory submission for a physical medical device (EndyMed Contour Handpiece), focusing on its technical specifications and substantial equivalence to existing devices, rather than the performance evaluation of a novel AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN150005

    Validate with FDA (Live)

    Date Cleared
    2015-12-18

    (336 days)

    Product Code
    Regulation Number
    878.4420
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndyMed Newa™ is an over-the-counter home use device intended for non-invasive treatment of mild to moderate facial wrinkles for adult women who have Fitzpatrick Skin Types I-IV.

    Device Description

    The Newa™ is an OTC, home use hand held device generating pulses of radiofrequency (RF) energy that are emitted into the skin. RF energy heats the tissue to improve the appearance of wrinkles and rhytides. It is a non-invasive, non-ablative device consisting of:

    • User interface
    • Programmable logic controller (PLC)
    • RF power module
    • Power supply
    • RF electrodes
    AI/ML Overview

    Acceptance Criteria and Device Performance Study for NEWATM

    1. Table of Acceptance Criteria and Reported Device Performance

    Special Control (Acceptance Criteria)Device Performance/Study Findings
    Non-clinical Performance Data: Demonstrate meets design specs and performance requirements (over-heating, power accuracy, radiofrequency, pulse cycle, waveform, pulse duration, and device characterization parameters).Over-heating safety: Device tested to ensure it does not overheat skin. Built-in thermistor monitors skin temperature, stops RF energy delivery when surface temperature exceeds 42°C. Acceptance criterion: temperature on gel/electrodes not to rise above 42±0.5℃. Test result: temperature rose to 42.1℃, then RF energy delivery stopped. Power accuracy: Validated at 10 Watts on a 360Ω load. Acceptance criterion: 10 Watt ±20% (8-12 Watt). Test result: measured total power was 10W. Newa™ parameter validation: Tested for radio frequency, pulse cycle, wave form, and pulse duration. Accepted values: 1 MHz RF, 750ms pulse cycle, modified sine wave (square wave), 300ms pulse duration for "1" mode, and 450ms for "2" mode. Test results: at the declared values.
    Label Comprehension and Self-Selection Performance Evaluation: Demonstrate intended OTC users can understand labeling and correctly choose device.Study: 247 subjects. 195 (78.9%) stated they could use the device; 52 (21.1%) stated they could not. All 52 who said no made correct self-selection decision. Of 195 who said yes, 180 (92.3%) made a correct self-selection decision. Acceptance Criterion: 85% target for self-selection success rate. Result: 92.3% achieved (for those who said they could use it). Validation test (Kappa test) showed strong agreement (k=0.81) between subjects' self-selection and expert opinion.
    Usability Performance Evaluation: Demonstrate layman can correctly use device based on directions.Study: 62 participants. Tasks: 1. Preparing self for treatment (3 tasks): 100% (62/62) completed correctly. 2. Preparing device for treatment (3 tasks): 100% (62/62) completed correctly. 3. Treatment performance (8 tasks): 100% (62/62) completed correctly. Result: All tasks performed correctly by 100% of subjects.
    Clinical Performance Evaluation: Demonstrate device performs as intended for aesthetic results.Study: 69 participants (62 completers). Effectiveness Criterion: At least 75% of patients show a success (≥1 score reduction on Fitzpatrick Wrinkle Severity Scale agreed by ≥2 of 3 dermatologists). Result (at 4-week post-treatment): * Reviewer #1: 91.93% (57/62) * Reviewer #2: 91.93% (57/62) * Reviewer #3: 90.32% (56/62) * ≥2 agree: 95.16% (59/62) Result (at 3-month follow-up): * Reviewer #1: 91.93% (57/62) * Reviewer #2: 96.77% (60/62) * Reviewer #3: 98.39% (61/62) * ≥2 agree: 100% (62/62) Conclusion: Study success criterion for treatment effectiveness was met.
    Biocompatibility: Patient-contacting components demonstrated to be biocompatible.Study: Electrodes and shell (b(4)) are same as Newa™ prescription device (K130793). Tested according to ISO 10993-1 and USP Class VI tests. Result: Met requirements for up to 30 days contact with human tissue.
    Instructions for Cleaning: Cleaning instructions validated.Study: Cleaning validation protocol and report provided. Result: Cleaning instructions are adequate. Device is non-sterile and reusable, for single user.
    Electromagnetic Compatibility (EMC) and Electrical Safety: Performance data provided.Study: Tested against IEC 60601-1:2005/EN 60601-1:2006, IEC 60601-1-2:2007, IEC 60601-1-11:2010. Result: Device passed all relevant testing portions, demonstrating electrical safety for home use.
    Software Verification, Validation & Hazard Analysis: Performed per guidelines.Study: Software is firmware on micro-controller, same as cleared prescription device (K130793). Considered moderate LOC. All elements for moderate LOC devices provided per FDA guidance. V&V testing conducted for system initialization, power button, power down, LEDs, vibration, temperature detection. Result: Satisfactory results. Software development procedures provide assurance of intended performance; all software risks adequately mitigated.
    Labeling: Include warnings, precautions, contraindications, and clinical summary.Content: User Manual, Quick Reference Guide, Box Labeling. Includes: When to use (indication), What is Newa, Who can benefit, Using Newa (regimen), Who cannot use (contraindications), What to do to avoid harm (warnings, precautions), Manufacturer details. <brImage (Figure 12): Skin types I-IV are shown as suitable, V-VI marked with an X as unsuitable. Clinical Summary: A summary of effectiveness and adverse events from the clinical study is specified to be included, along with other required statements like those regarding safety/effectiveness for non-indicated uses.

    Since this is a device for aesthetic use and not a diagnostic AI, many of the typical AI/ML study components (like training set details, MRMC studies, standalone performance with explicit metrics for algorithm output) are not directly applicable in the same way. The studies focus on safety, usability, and clinical effectiveness as a treatment device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Study (Effectiveness & Safety):
      • Enrollment: 69 participants (adult women, average age 54.4 ± 8.1 years, range 37-72; Fitzpatrick Skin Types I-IV).
      • Completers: 62 participants.
      • Provenance: Prospective, single-center study. Country of origin is not explicitly stated but implies the same location as the manufacturer or clinical trial site.
    • Label Comprehension & Self-Selection Study:
      • Sample Size: 247 subjects.
      • Provenance: Not explicitly stated but implies a general population capable of reading English, recruited for the study.
    • Usability Study:
      • Sample Size: 62 subjects (these were part of the clinical study participants).
      • Provenance: Part of the prospective clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Clinical Study (Effectiveness):

      • Number of Experts: Three independent board-certified dermatologists.
      • Qualifications: "Board-certified dermatologists." Specific years of experience are not mentioned.
    • Label Comprehension & Self-Selection Study:

      • Medical Expert for Validation: An interviewer (trained in assessing wrinkle severity) made an initial assessment. This was then provided to a "blinded outside expert" for final determination of correct self-selection.
      • Qualifications: The interviewer was "trained in assessing wrinkle severity." The "blinded outside expert" qualifications are not further detailed beyond being an "expert."

    4. Adjudication Method for the Test Set

    • Clinical Study (Effectiveness):

      • A patient was considered a "success" if at least 2 out of the 3 independent dermatologists agreed that the patient achieved at least one degree of wrinkle score reduction (per Fitzpatrick Wrinkle Severity Scale) when comparing baseline to 4-week follow-up. This is a 2+1 adjudication method.
    • Label Comprehension & Self-Selection Study:

      • The subject's decision and reasons, along with the interviewer's assessment of wrinkle severity, were provided to a "blinded outside expert" for a final determination of whether the subject had correctly self-selected. This implies a single expert adjudicating the correctness of the subject's self-selection, informed by other inputs.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a treatment device, not a diagnostic AI. The studies performed were to demonstrate the device's effect on wrinkles, not to enhance human reader performance in diagnosis. The "readers" (dermatologists) in the clinical study evaluated before-and-after photographs, but their role was to assess treatment outcomes, not to interpret an AI's output.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is a physical device intended for direct human use for treatment, not a standalone AI algorithm. The device itself performs actions (RF energy delivery) rather than providing diagnostic outputs that would be typically associated with a standalone AI algorithm's performance evaluation.

    7. The Type of Ground Truth Used

    • Clinical Study (Effectiveness): Expert Consensus. The ground truth for wrinkle reduction was established by the consensus of at least 2 out of 3 independent board-certified dermatologists, who assessed high-resolution photographs using the Fitzpatrick Wrinkle Severity Scale.
    • Safety Adjudication: Investigator observation and patient diaries for adverse events.
    • Self-Selection/Usability Studies: Expert opinion (interviewer + blinded outside expert) on the correctness of user actions and self-selection decisions.

    8. The Sample Size for the Training Set

    • Not Applicable (for the device's primary function). The NEWATM device is an electrosurgical device, not a machine learning algorithm in need of a "training set" in the conventional AI/ML sense for its core functionality. Its software controls basic operational and safety functions, which are validated through traditional software V&V processes and bench testing, not through a large-scale data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As noted above, a "training set" for an AI/ML algorithm is not described as being part of the device's development or validation for its primary function. Software validation relies on defined specifications and formal verification and validation protocols rather than a data-driven training approach.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130501

    Validate with FDA (Live)

    Device Name
    INTENSIF
    Date Cleared
    2014-03-13

    (379 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intensif Applicator is intended for use in Dermatologic and General Surgical procedures for electrocoagulation and hemostasis.

    Device Description

    EndyMed's Intensif Applicator is a treatment handpiece to be attached to the FDA cleared EndyMed Imagine TC Skin Treatment System (K08346) (EndyMed Pro). The Intensif Applicator has a custom grid array of micro electrodes that is used to deliver the energy into the tissue. The Intensif Applicator tip emits bipolar RF energy that flows between electrodes to create thermal heating of the tissue for hemostasis and coagulation. The Intensif Applicator consists of: . Disposable treatment element (tip) . RF electrodes on a disposable tip . Operation trigger.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Intensif Applicator, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / PerformanceDescription / Reported Performance
    SafetyCompliance with IEC 60601-1:2005 / EN 60601-1:2006 (Medical electrical equipment - General requirements for basic safety and essential performance, 3rd edition for basic safety and essential performance). Compliance with IEC/EN 60601-1-2 (Electromagnetic compatibility (EMC)). Lab performance tests, design validation, software verification and validation. Ex Vivo studies (laboratory and animal) showed similar thermal and biological effects to previously FDA cleared fractional skin hemostasis and coagulation devices.
    EfficacyLab performance tests, design validation, software verification and validation. Ex Vivo studies (laboratory and animal) showed similar thermal and biological effects to previously FDA cleared fractional skin hemostasis and coagulation devices. Device performs according to its specifications.
    Operational PerformanceIntensif Applicator's power control and accuracy. Intensif Applicator's safety features functionality. Intensif Applicator needle depth validation.
    Intended UseFor use in Dermatologic and General Surgical procedures for electrocoagulation and hemostasis. The results of preclinical testing indicate it offers a "minimally invasive, effective, safe device for skin hemostasis and coagulation."
    Substantial EquivalenceThe device has the same intended use and similar indications as predicate devices (Primaeva Medical System K080145, EndyMed FSR K101510, EndyMed Imagine TC Skin Treatment System K083461). Technology is similar. Power and frequency envelopes covered by predicate devices. Minor differences in human interface/accessories do not raise new safety/effectiveness issues, as verified by performance and ex vivo testing.

    Study Information

    This device did not involve a clinical study with human subjects. The supporting studies are entirely preclinical and ex vivo.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not specified. The document mentions "a few Ex Vivo laboratory studies and Ex Vivo animal studies tests" but does not quantify the number of samples or animals used.
    • Data Provenance: The studies were "Ex Vivo laboratory studies and Ex Vivo animal studies tests." The location where these studies were performed is not explicitly stated, but the company is based in Israel. This is a preclinical type of data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not applicable as the studies were preclinical (ex vivo laboratory and animal) and did not involve human interpretation or expert evaluation to establish ground truth in the context of clinical outcomes. The "ground truth" was based on measurable physical and biological effects in the lab.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no human expert adjudication of a test set as the studies were preclinical.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed. This device is not an AI-assisted diagnostic tool; it's an electrosurgical applicator.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • While the device itself operates without human intervention post-activation, this question is typically relevant for diagnostic algorithms or AI. For this medical device, the "standalone" performance refers to its ability to achieve its intended physical effect (electrocoagulation and hemostasis) as demonstrated in the ex vivo studies. It is not an "algorithm only" type of device.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance was established through measurable physical and biological effects in ex vivo laboratory settings and animal tissues. This included:
      • Evaluating thermal and biological effects.
      • Measuring power control and accuracy.
      • Testing safety features functionality.
      • Validating needle depth.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a medical device for electrocoagulation and hemostasis, not an AI/machine learning algorithm requiring a "training set" of data in the conventional sense. Its design and performance are based on engineering principles and physical testing.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there was no "training set" in the context of AI/machine learning for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120513

    Validate with FDA (Live)

    Device Name
    GLOW BY ENDYMED
    Date Cleared
    2012-05-08

    (77 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Glow by EndyMed is a noninvasive device intended for use in Dermatologic and General Surgical procedures. The TC applicator is indicated for mild to moderate facial wrinkles and rhytides. The FSR applicator is indicated for Dermatological procedures requiring ablation and resurfacing of the skin.

    Device Description

    The EndyMed Glow is a non-invasive unit consists of a user interface, programmable logic controller (PLC), an RF power module, internal electronics, and treatment handpieces. The user interface allows the selection of treatment parameters by pressing on the treatment buttons; A LCD screen displays the current treatment settings. The PLC is a specially configured computer that provides the operational and safety functions of the system. The system is connected to one or two out of 2 TC handpieces with different size electrode areas for different treatment sites, or to a FSR handpiece. The RF power module provides RF energy to the handpiece, producing a sinusoidal signal at 1MHz frequency.

    AI/ML Overview

    The provided text is a 510(k) Summary for the "Glow by EndyMed" device. This document is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. It does not contain information about a clinical study designed to prove the device meets specific acceptance criteria.

    The document primarily focuses on:

    • Device Description and Intended Use: Describing what the device is and what it's used for.
    • Predicate Devices: Identifying previously cleared devices to which "Glow by EndyMed" claims substantial equivalence.
    • Performance Standards: Stating compliance with general safety and EMC standards for medical electrical equipment.
    • Technological Characteristics Comparison with Predicates: Asserting that the device has the same technological characteristics as the predicate devices.

    Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and MRMC studies cannot be extracted from this document as such studies are not described.

    The document states:

    • "The Glow by EndyMed has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate devices."
    • "The TC applicator has exactly the same RF technology characteristics as the EndyMed Imagine TC Skin Treatment System (K0834610) and the FSR applicator has the same fractional RF technology as the EndyMed Fractional Skin Resurfacing (FSR) Applicator (K101510)."

    This indicates that the primary "proof" of meeting acceptance criteria for a 510(k) submission like this is typically through demonstrating substantial equivalence to already cleared devices, rather than through new clinical performance studies with defined acceptance criteria for a novel device. The implicit acceptance criteria are that the device performs as safely and effectively as the predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101510

    Validate with FDA (Live)

    Date Cleared
    2011-02-17

    (261 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fractional Skin Resurfacing (FSR) Applicator is intended for dermatological procedures requiring ablation and resurfacing of the skin

    Device Description

    EndyMed's Fractional Skin Resurfacing (FSR) Applicator is a treatment handpiece to be attached to the FDA cleared EndyMed Imagine TC Skin Treatment System (K08346). The FSR Applicator tip emits bipolar RF energy that flows between electrodes to create micro-ablation points on the skin, forming superficial ablation with a volumetric non ablative heating effect in the dermis.

    AI/ML Overview

    The provided text describes the EndyMed Fractional Skin Resurfacing (FSR) Applicator and its 510(k) submission. However, it does not contain the specific details required to fully address your request regarding acceptance criteria and a detailed study.

    The document states that a "performance testing" was conducted and its results "clearly indicate that the Fractional Skin Resurfacing (FSR) Applicator offers a non-invasive, effective, safe device for skin resurfacing." It also claims the device "is as safe, as effective, and performs at least as safely and effectively as the legally marketed device" (predicate devices).

    Here's a breakdown based on the information provided, highlighting what is missing:


    1. Table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Not specified in documentNon-invasive, effective, safe for skin resurfacing. Performs at least as safely and effectively as predicate devices.

    Missing Information: The document does not explicitly state specific quantitative or qualitative acceptance criteria (e.g., minimum percentage improvement in skin texture, reduction in wrinkles, safety endpoint thresholds). It only provides a general statement about the device's performance relative to safety and efficacy.


    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    Missing Information: The document states "a performance testing" was conducted but does not provide any details on:

    • The sample size of participants or data points used in the test set.
    • The country of origin for the data.
    • Whether the study was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Missing Information: The document mentions "performance testing" but does not describe how ground truth was established, nor does it refer to any experts involved in this process or their qualifications.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Missing Information: There is no information provided about any adjudication method used for establishing ground truth or evaluating the test results.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not Applicable (Based on available information): This device is an applicator for dermatological procedures (physical treatment), not an AI-assisted diagnostic device that would involve human readers interpreting results with or without AI. Therefore, an MRMC study related to AI assistance for human readers is not relevant in this context.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not Applicable (Based on available information): Similar to point 5, this is a physical treatment device, not an algorithm. Standalone performance (algorithm-only) is not applicable here. The device's performance inherently involves a human operator.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Missing Information: The document does not specify the type of ground truth used for evaluating the device's performance. Given it's a skin resurfacing device, potential ground truths could include clinical assessment scores (e.g., wrinkle scales, skin texture scores), photographic assessment, or potentially histological changes, but none are mentioned.


    8. The sample size for the training set:

    Missing Information: As the device is not an AI algorithm, there is no "training set" in the conventional sense. The "performance testing" described is likely a clinical trial or study on human subjects, not an algorithm training process.


    9. How the ground truth for the training set was established:

    Missing Information: As there is no AI algorithm training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1