Found 25 results

510(k) Data Aggregation

    K Number
    K161482

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-27

    (27 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The water bottle cap irrigation system (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

    Device Description

    The EndoChoice Water Bottle Cap Irrigation System is a sterile, consumable, 24-hour multi-patient use device designed to fit commercially available sterile water bottles and irrigation pumps for providing sterile water during endoscopic procedures.

    The EndoChoice water bottle cap system is comprised of a polycarbonate standard water bottle cap, medical grade PVC tubing that extends from the endoscope connector into the water bottle, and a medical grade stainless steel cylinder that serves as both the water intake portion of the disposable water bottle cap and a weight to hold the tubing down while in the water bottle.

    The bottle tubing is added into the sterile water bottle and the irrigation system is fastened via the tubing system cap.

    Flexible pump tubing connected to the top of the water bottle cap runs to the tubing connector reducer. This flexible tubing is the part of the device that is placed into the irrigation pump. A tubing connector reducer joins the flexible tubing and rigid tubing. The rigid tubing runs to the tubing Luer connector. The tubing Luer connector has the first one-way check valve which prevents potential backflow of fluid from the endoscope into the tubing set. The separately packaged, single-use endoscope connectors (used to connect the tubing set to the auxiliary water channel) also contain one-way valves used to prevent potential backflow.

    As with the predicate, the separately packaged, single-use connectors are used to connect to different endoscope models, and all include the second one-way valve to prevent contamination due to potential backflow.

    Between patients, the endoscope connectors must be changed to ensure that there is no risk of cross contamination. EndoChoice water bottle irrigation tubing fits standard threaded lid disposable sterile water bottles of approximately 33 mm diameter.

    The water bottle cap system is supplied sterile and can be used up to 24 hours.

    AI/ML Overview

    The provided document is a 510(k) summary for the EndoChoice Water Bottle Cap Irrigation System. It discusses the device's equivalence to a predicate device based on non-clinical testing. However, the document does not contain information about acceptance criteria, device performance metrics, sample sizes for test sets (as this is not an AI/algorithm-based device and thus doesn't have a "test set" in that context), ground truth, or MRMC studies.

    Therefore, I cannot fulfill the request for a table of acceptance criteria and reported device performance, or details about the other points, as the necessary information is not present in the provided text.

    The document only states that:

    • Non-clinical testing (benchtop funcional performance testing, post-aging; and biocompatibility testing per ISO 10993-1) was performed.
    • "All test results passed, demonstrating that the device is safe and effective in comparison with the predicate device."

    This indicates that internal acceptance criteria for these tests were met, but the specific criteria and detailed results are not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161167

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-05-26

    (30 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from biopsy port throughout the endoscopic procedure, and provides access for irrigation.

    Device Description

    The Seal "Single-Use Biopsy valve is provided non-sterile, and is composed of a cylindrical base and a connected cap. The Single-Use Biopsy Valve provides sealable access via the working channel port of an endoscope for surgical instruments. The valve consists of an attached cap and valve body with a circular opening. The cap is pre-perforated, and ensures instrument access and removal is tight and leak-free.

    The valve is designed with a hollow body with a distal end that releasably attaches to the inlet port of the working channel. The valve body provides a flexible diaphragm seal that separates the body into a proximal chamber and distal chamber. The distal chamber secures onto the endoscope, while the proximal chamber secures the cap. Instruments are inserted through the cap and chambers, through the diaphragm seal, and into the endoscope. The diaphragm seal is configured such that a seal is formed due to deformation of the elastomer around an instrument when it is inserted through the opening, preventing fluids from passing through the biopsy valve.

    The cap serves two primary functions. One function of the cap is to provide an additional seal around surgical instruments to prevent leakage. The second function is to allow modification of pressure during insufflation by opening the cap, allowing gasses to flow from the endoscope, and out of the biopsy valve.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Seal Biopsy Valve) and does not contain information about acceptance criteria for a study or the study details you are asking for. It focuses on demonstrating substantial equivalence to a predicate device.

    Therefore, I cannot provide the requested information. The document details administrative information, the intended use of the device, its description, and a comparison to a predicate device to confirm substantial equivalence, rather than a performance study with acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160403

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-05-19

    (97 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartStart™ air/water valve is intended be used to control the air/water function on an endoscope during a GI endoscopic procedure.

    The SmartStart™ suction valve is intended to be used to control the suction function on an endoscope during a GI endoscopic procedure.

    Device Description

    The device is intended for single-use, and is supplied sterile, single-use air/water and suction valves help prevent potential safety risks and eliminate the need for manual cleaning and reprocessing. These valves easily incorporate into infection prevention policies as a single patient use item. The air/water valve is designed to be attached to the air/water port of the endoscope and the suction valve to the suction port of the endoscope. The activation of the suction valve allows the user to aspirate excess fluids or other debris obscuring the endoscopic image, while the activation of the air/water valve allows the user to control air and water flow to assist in cleansing the lens during procedures.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the SmartStart™ Air/Water and Suction Valves. It aims to demonstrate substantial equivalence to predicate devices, rather than establishing de novo performance criteria. Therefore, this document does not contain the information requested in the prompt regarding acceptance criteria and the comprehensive study details for a device meeting those criteria.

    Specifically, the document focuses on demonstrating that the SmartStart™ valves perform similarly to already approved devices. It does not provide:

    • A table of distinct acceptance criteria (e.g., sensitivity, specificity, accuracy targets) for the device's clinical performance.
    • Details of a study proving the device meets specific performance acceptance criteria beyond basic physical and functional tests (like air/water flow and leakage).
    • Information on sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth types for clinical efficacy.
    • Details regarding training sets for an AI/algorithm-based device, as this is a medical device and not an AI-driven software.

    The performance testing section (8.3) mentions "acceptance criteria for air flow, water flow and leakage," but it does not specify what those criteria are or provide detailed results from a study that quantifies the device's performance against these criteria. It simply states that the device "will perform in accordance with the acceptance criteria."

    This type of submission relies on comparative analysis with existing predicate devices, focusing on technical characteristics, indications for use, and a limited set of non-clinical performance and safety tests (sterilization, shelf-life, biocompatibility, and basic functional tests).

    Ask a Question

    Ask a specific question about this device

    K Number
    K153588

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-02-24

    (70 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoChoice water bottle cap system is intended to be used with an air source or carbon dioxide (CO2) with the purpose of supplying sterile water to the endoscope during endoscopic procedures. It is compatible with commercially available sterile water bottles.

    Device Description

    The EndoChoice water bottle cap system is a sterile disposable device designed to fit commercially available sterile water bottles for providing sterile water during endoscopic procedures. The water bottle cap system is designed with a clip that is placed on the tubing to stop water flow from the water bottle when the endoscope is not in use.

    The device attaches via a connector to the air/water port of an endoscope. The water bottle cap itself is attached to a standard sterile water bottle. Air supplied from the processor unit then pressurizes the water bottle, providing water flow when then air/water valve of the endoscope is depressed. Some models provide a feature that provides an option to use a CO2 air source. Instructions state that the device is to be used by trained endoscopy professionals.

    AI/ML Overview

    This FDA 510(k) premarket notification describes a medical device, the "EndoChoice Water Bottle Cap System," and makes a case for its substantial equivalence to a legally marketed predicate device. However, it does not contain the kind of detailed information about acceptance criteria and a study that proves the device meets them, as requested in your prompt.

    Here's why and what information is available:

    This document is a 510(k) submission, which is a premarket notification demonstrating that the device is at least as safe and effective as a legally marketed predicate device. It typically relies on comparing the new device to an existing one (the predicate) and showing that any differences do not raise new questions of safety or effectiveness. This is different from a PMA (Premarket Approval) which requires clinical trials to demonstrate safety and effectiveness for novel devices.

    Therefore, the document does not provide:

    • A table of acceptance criteria and reported device performance in terms of specific metrics like sensitivity, specificity, accuracy, etc.
    • Sample sizes used for a test set, data provenance, ground truth establishment for an AI/diagnostic device.
    • The number of experts or their qualifications for ground truth.
    • Adjudication methods.
    • A multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone algorithm performance.
    • Ground truth type for a diagnostic device.
    • Training set sample size or how its ground truth was established.

    What the document does provide, and how it relates to proving safety and effectiveness for this type of device (an accessory for endoscopes):

    1. Acceptance Criteria and Device Performance (Implicit):

    For this device, the "acceptance criteria" are implicitly tied to demonstrating that the modified device performs comparably to the predicate device and meets safety standards, particularly concerning its function in supplying sterile water and biocompatibility.

    Acceptance Criteria (Implicit from provided text)Reported Device Performance
    Functional Performance: Maintain intended function of supplying sterile water during endoscopic procedures."Benchtop functional performance testing, post-aging" was completed; "All test results passed."
    Biocompatibility: Meet safety standards for patient-contacting materials."Biocompatibility testing in conformance with ISO 10993-1" was completed; "All test results passed."
    Sterility: Maintain sterility until use.Sterilization method (Ethylene Oxide) is unchanged from the predicate.
    Compatibility: Remain compatible with existing endoscopes and water bottles.Compatibility with Olympus 140, 160, 180, and Fuse Endoscopes is unchanged. Compatible with Standard 33 mm water bottle caps (unchanged).
    Substantial Equivalence: Differences from predicate do not raise new issues of safety or effectiveness.Conclusion states: "The modified EndoChoice Water Bottle Cap System is equivalent to the legally marketed predicate device. It is the same or equivalent in terms of design, intended use, materials, and labeling."

    2. Sample Size for Test Set and Data Provenance:

    • Sample Size: Not specified quantitatively. The "benchtop functional performance testing" and "biocompatibility testing" would have involved a certain number of device units, but this specific number is not disclosed in the summary.
    • Data Provenance: The testing was "non-clinical testing" and "benchtop," meaning it was conducted in a laboratory/engineering environment, not on human patients. The country of origin for the testing facilities is not specified but would typically be where the manufacturer (EndoChoice, Inc. in Alpharetta, GA) conducts its R&D or uses contract labs. It is prospective testing performed on the modified device.

    3. Number of Experts and Qualifications for Ground Truth:

    • For this type of device, "ground truth" isn't established by medical experts in the way it would be for an AI diagnostic device. Instead, the "truth" is determined by engineering and material science standards (e.g., a pump correctly delivers water, a material does not cause a toxic reaction).
    • The experts involved would be engineers, material scientists, and quality control personnel who conducted and evaluated the benchtop and biocompatibility tests. Their specific number and qualifications are not mentioned in this summary but are implicit to the conduct of these types of tests under regulatory standards.

    4. Adjudication Method:

    • Not applicable in the context of this 510(k) submission for a non-diagnostic accessory. Test results would be objectively measured against predefined engineering specifications and ISO standards; there's no "adjudication" by multiple human reviewers for interpretation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data, and the device (often AI) is intended to assist or replace that interpretation. The EndoChoice Water Bottle Cap System is a physical accessory, not an interpretive diagnostic tool.

    6. Standalone (Algorithm Only) Performance:

    • No, a standalone performance study in the context of an algorithm or AI was not done. This device does not contain an algorithm or AI. It is a mechanical accessory.

    7. Type of Ground Truth Used:

    • For functional performance, the "ground truth" is adherence to engineering specifications for flow rate, pressure capability, leak prevention, and durability (post-aging). This is determined by objective physical measurements against established design requirements.
    • For biocompatibility, the "ground truth" is conformance to ISO 10993-1 standards, meaning the materials do not elicit an unacceptable biological response. This is determined by specified laboratory tests (e.g., cytotoxicity, sensitization, irritation).

    8. Sample Size for the Training Set:

    • Not applicable. There is no AI or algorithm that requires a "training set" for this device.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151475

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-02-08

    (251 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoChoice Select injection needle is intended for endoscopic injection into the gastrointestinal mucosa.

    Device Description

    The EndoChoice Select Injection Needle mainly consists of a cap, needle, boosting tube, and positioning ring, all made of 304 stainless steel. It also contains an outer tube and nut made of PTFE. The spring and spring tube are made of 302 stainless steel. The shell and injection handle are made of ABS. The bushing is made of TPE (SBS), the inner tube is made of nylon, and the color band indicator is made of CNBR.

    The products are intended for single use; an individual device is packed in a sealed pouch following ETO sterilization. The device is used for local injection via endoscope, with the average contact time of the products and the mucosa of the human digestive tract of less than 1 hour.

    AI/ML Overview

    The provided text describes a Special 510(k) Summary for the EndoChoice Select Injection Needle, which is a medical device. This type of submission is for modifications to a previously cleared device. The document primarily focuses on demonstrating "substantial equivalence" to a predicate device, rather than providing detailed acceptance criteria and a study to prove a device meets those criteria for a novel device.

    Therefore, many of the requested data points (like specific performance metrics, sample sizes for test and training sets, number and qualifications of experts, adjudication methods, MRMC studies, and detailed ground truth establishment) are generally not found in this type of submission. The focus is on showing the modified device performs as safely and effectively as the original, cleared device.

    However, I can extract the information that is present and indicate where the requested information is not typically presented in a 510(k) summary focused on substantial equivalence.

    1. Table of acceptance criteria and the reported device performance

    For the modified EndoChoice Select Injection Needle, the "acceptance criteria" are predominantly related to the equivalence of its characteristics and safety to the predicate device. The "reported device performance" is essentially that it passed these equivalence-based tests.

    Acceptance Criteria (Demonstrated Equivalence to Predicate K132065)Reported Device Performance (Result)
    Material Equivalence: The new patient-contacting material must be safe and effective.Biocompatibility testing for the new patient-contacting material passed.
    Sterilization Equivalence: Sterilization method and efficacy must be equivalent.Sterilization testing is unchanged from the predicate and passed.
    Accelerated Aging Equivalence: Shelf life and stability characteristics must be equivalent.Accelerated aging testing is unchanged from the predicate and passed.
    Functional Equivalence: The modified device must perform identically to the predicate.All test results passed, demonstrating safety and efficacy equivalent to the predicate device.
    Design, Intended Use, Labeling Equivalence: Must be the same or similar as the predicate.All characteristics (manufacturer, endoscope compatibility, sterility, sheath diameter, needle size, outer tubing, length, needle extension, indications for use, packaging) were found to be equivalent to the predicate.

    2. Sample sized used for the test set and the data provenance

    The document does not specify sample sizes for individual tests. The "test set" here refers to the samples of the modified device used for the mentioned non-clinical tests (biocompatibility, accelerated aging, sterilization). The data provenance is internal to EndoChoice, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. Ground truth, in the context of device performance, would typically refer to established standards or verifiable performance metrics against which the device is measured. For a K151475 submission, the "ground truth" for "substantial equivalence" is the performance and safety profile of the predicate device (K132065).

    4. Adjudication method for the test set

    Not applicable/Not specified in this document. Adjudication methods are more relevant for human-read evaluations, not typically for non-clinical engineering tests of medical devices like injection needles.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an injection needle, not an AI-powered diagnostic tool, so MRMC studies involving human readers and AI are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this substantial equivalence submission is the established safety and efficacy profile of the predicate device (EndoChoice Injection Needle K132065). The non-clinical tests conducted on the modified device (biocompatibility, sterilization, accelerated aging) were designed to demonstrate that the changes did not negatively impact these established characteristics of the predicate.

    8. The sample size for the training set

    Not applicable. This is a physical device, not an AI/machine learning model that would require a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a physical injection needle.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152580

    Validate with FDA (Live)

    Device Name
    rescuenet
    Manufacturer
    Date Cleared
    2015-11-03

    (54 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The retrieval net is intended for use in endoscopic retrieval of foreign objects, food bolus, tissue fragments and excised tissue such as polyps.

    Device Description

    The EndoChoice Retrieval Net is a disposable, single-use sterile device comprised of a flexible wire cable and snare loop with net which can be extended and retracted form the flexible outer sheath using a three ring handle. Objects are retrieved by advancing the finger rings of the handle to open the retracted net. The net is then endoscopically manipulated over the foreign object, food bolus, tissue fragment or excised tissue, in which the finger rings of the handle are retracted to close the net once the article has been retrieved.

    AI/ML Overview

    This document, a 510(k) premarket notification, describes the "EndoChoice Retrieval Net," a retrieval device intended for use in endoscopic procedures. The document aims to demonstrate the device's substantial equivalence to a legally marketed predicate device.

    Acceptance Criteria and Device Performance:

    The document focuses on demonstrating substantial equivalence to a predicate device, the Roth Net® retriever product line (K122462). The core acceptance criterion is that the new device performs at least as safely and effectively as the predicate, based on comparable characteristics.

    While a formal table of "acceptance criteria" with specific quantitative targets and reported performance isn't explicitly provided, the substantial equivalence table (Section 7.5) serves this purpose by comparing key characteristics of the new device to the predicate. The "reported device performance" is implicitly that the device meets or matches the predicate in the listed attributes.

    CharacteristicAcceptance Criteria (Predicate: Roth Net® retriever product line K122462)Reported Device Performance (EndoChoice Retrieval Net)
    Product CodeFDI, GCJFDI, GCJ
    Regulation No.21CFR § 876.430021CFR § 876.4300
    ClassificationEndoscopic electrosurgical unit and accessoriesEndoscopic electrosurgical unit and accessories
    Supplied SterileYesYes
    Sheath length160cm, 230cm (at least one matching)230cm
    Single useYesYes
    Sheath diameter2.5mm, 3.0mm (at least one matching)2.5mm
    Compatibility2.8mm, 3.2mm endoscope channel (at least one matching)2.8mm endoscope channel
    Net Size3 x 6 cm, 2 x 4.5 cm, 4 x 5.5 cm, 4 x 8 cm (at least one matching)3 x 6 cm
    ConfigurationOval, Octagonal, Hexagonal (at least one matching)Oval
    Indications for useIntended for use in endoscopic retrieval of excised polyps, tissue samples, foreign bodies and calculi during flexible and rigid endoscopy procedures.Intended for use in endoscopic retrieval of foreign objects, food boluses, tissue fragments and excised tissue such as polyps.
    PackagingSingle-use pouchSingle-use pouch

    Study Information:

    1. Sample size used for the test set and the data provenance:
      The document does not describe a clinical performance study with a "test set" in the traditional sense of evaluating diagnostic accuracy or a specific outcome. Instead, it relies on non-clinical testing (sterilization, shelf-life, biocompatibility, and general performance testing) to demonstrate that the device functions as intended and is safe. No specific sample sizes for these tests are provided in the summarized information, nor is data provenance (country of origin, retrospective/prospective) for a clinical test set.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable. This submission does not involve a "test set" requiring expert ground truth in the context of diagnostic or interpretive performance. The non-clinical tests rely on established scientific methods and standards.

    3. Adjudication method for the test set:
      Not applicable. There is no clinical "test set" or expert review process described for adjudication.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. The device is a physical retrieval tool, not an AI-assisted diagnostic or interpretive product. Therefore, an MRMC study related to AI assistance for human readers is not relevant.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable. This device is a physical medical instrument, not an algorithm, and intrinsically requires human operation.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For the non-clinical tests:

      • Sterilization and Shelf-Life: Ground truth is established by adherence to recognized international standards (ISO 11135:2014, AAMI ANSI ISO 11737-2:2009, AAMI ANSI ISO 11138-1:2006).
      • Biocompatibility: Ground truth is established by adherence to AAMI ANSI ISO 10993-1:2009.
      • Performance Testing: While details are not given, this would typically involve engineering tests against internal specifications or industry standards to confirm functional performance (e.g., net opening/closing, retrieval capability, material integrity).
    7. The sample size for the training set:
      Not applicable. As this is not an AI/machine learning device, there is no "training set."

    8. How the ground truth for the training set was established:
      Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142155

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-01-20

    (167 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoChoice water bottle cap system is intended to be used with an air source or carbon dioxide (CO2) with the purpose of supplying sterile water to the endoscope during endoscopic procedures. It is compatible with commercially available sterile water bottles.

    Device Description

    The EndoChoice water bottle cap system is a sterile disposable device designed to fit commercially available sterile water bottles for providing sterile water during endoscopic procedures. The water bottle cap system is designed with a clip that is placed on the tubing to stop water flow from the water bottle when the endoscope is not in use.

    The device attaches via a connector to the air/water port of an endoscope. The water bottle cap itself is attached to a standard sterile water bottle. Air supplied from the processor unit then pressurizes the water bottle, providing water flow when then air/water valve of the endoscope is depressed. Some models provide a feature that provides an option to use a CO2 air source. Instructions state that the device is to be used by trained endoscopy professionals.

    AI/ML Overview

    This document is a 510(k) premarket notification for the EndoChoice Water Bottle Cap System. It details the device, its intended use, and its substantial equivalence to a predicate device.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Benchtop Functional PerformanceAll test results passed
    Sterilization ValidationAll test results passed (conforms to validation testing)
    Biocompatibility in conformance with ISO 10993-1All test results passed

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "Benchtop functional performance testing" and "Laboratory validation testing of sterilization" without specifying exact sample sizes for each test. It also states "Biocompatibility testing in conformance with ISO 10993-1." The nature of these tests (benchtop, laboratory) suggests that they would likely use a dedicated number of device units or components for each specific test, rather than a broad "test set" like clinical trials.

    The document does not detail the provenance of data (e.g., country of origin, retrospective/prospective). Given these are non-clinical hardware tests, such details are typically not relevant or included.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This section is not applicable to this document. The document describes non-clinical performance and safety testing (functional, sterilization, biocompatibility). These tests do not involve human interpretation or the establishment of ground truth by clinical experts in the same way, for example, an imaging diagnostic device would. Performance is measured against engineering specifications and industry standards (e.g., ISO 10993-1) by qualified laboratory personnel.

    4. Adjudication Method for the Test Set:

    This section is not applicable. Adjudication methods like 2+1 or 3+1 are used to resolve discrepancies in expert interpretations (e.g., in clinical trials for diagnostic devices). Since the testing described is non-clinical and objective (functional, sterilization, biocompatibility), there is no need for such adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:

    This section is not applicable. The device (water bottle cap system for endoscopes) is a medical accessory, not an AI-assisted diagnostic or therapeutic device. It does not involve human readers interpreting data, nor does it have an AI component to assist human performance in diagnosis or treatment.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This section is not applicable. As mentioned above, the device is a medical accessory and does not involve an algorithm or AI.

    7. The Type of Ground Truth Used:

    For the non-clinical testing:

    • Benchtop functional performance testing: Ground truth would be established by the functional specifications and design requirements of the device (e.g., water flow rate, pressure resistance, connection integrity).
    • Laboratory validation testing of sterilization: Ground truth is established by validated sterilization methods and industry standards (e.g., microbial load reduction, sterility assurance level).
    • Biocompatibility testing in conformance with ISO 10993-1: Ground truth is established by the requirements of the ISO 10993-1 standard for biological evaluation of medical devices.

    8. The Sample Size for the Training Set:

    This section is not applicable. As a hardware medical accessory, this device does not utilize a "training set" in the context of machine learning or AI development.

    9. How the Ground Truth for the Training Set was Established:

    This section is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140472

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-17

    (234 days)

    Product Code
    Regulation Number
    878.3720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BONASTENT® Tracheal/Bronchial Stent System is indicated for the treatment of tracheobronchial strictures caused by malignant neoplasms.

    Device Description

    The BONASTENT® Tracheal/Bronchial (BTB) is a self-expanding tubular prosthesis designed to maintain patency of tracheobronchial strictures caused by malignant tumors. The stent is made of Nitinol wire and is weaved using a hook & cross wire construction; a silicon membrane covers the stent and is designed to prevent stent migration. Stent sizes range in diameter from 10mm to 20mm and in length from 50mm to 80mm.

    The BONASTENT® Tracheal/Bronchial is available on two types of delivery devices. The Y-Shape Handle Delivery Device is used for stents with an outer diameter of 10Fr and less. The I-Shape Handle Delivery Device which is used for stents with an outer diameter larger than 10F.

    AI/ML Overview

    The provided text describes the 510(k) Summary for the BONASTENT® Tracheal/Bronchial device. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with explicit acceptance criteria and device performance metrics in the typical sense of a clinical trial. However, it does outline performance characteristics compared to its predicate.

    Here's an analysis based on the information provided, framed to address your request for acceptance criteria and supporting studies where possible, while noting limitations due to the nature of a 510(k) summary for a stent:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this specific device (a tracheal/bronchial stent), the "acceptance criteria" are implicitly defined by demonstrating equivalence to a legally marketed predicate device (AERO Tracheobronchial Stent System). The performance data supplied are comparative benchmarks rather than direct acceptance criteria against a defined threshold.

    Performance MetricAcceptance Criteria (Implicit: Comparable to Predicate)Reported Device Performance (BONASTENT® Tracheal/Bronchial)Predicate Device Performance (AERO Tracheobronchial)
    Intended UseTreatment of tracheobronchial strictures caused by malignant neoplasms.Treatment of tracheobronchial strictures caused by malignant neoplasms.Treatment of tracheobronchial strictures caused by malignant neoplasms.
    Stent MaterialNitinolNitinolNitinol
    Covering Material(Compared for differences)SiliconePolyurethane membrane
    Stent Diameter (mm)10, 12, 14, 16, 18, 2010, 12, 14, 16, 18, 2010, 12, 14, 16, 18, 20
    Stent Length (mm)20-8020-8020-80
    Delivery Diameter(Compared for differences)8F - 12F16F and 22F
    Deployment Time(Reference, as predicate unknown)Non-aged stents average 16.97 sec; Aged stents 16.4 secUnknown
    Expansion Force (10x20 stent)Comparable to 0.390 lbs. (AERO)0.367 lbs.0.390 lbs.
    Expansion Force (20x80 stent)Comparable to 0.763 lbs. (AERO)0.748 lbs.0.763 lbs.
    Compression Force (10x20 stent)Comparable to 0.519 lbs. (AERO)0.517 lbs.0.519 lbs.
    Compression Force (20x80 stent)Comparable to 1.026 lbs. (AERO)1.026 lbs.1.026 lbs.
    Corrosion Resistance (in simulated gastric fluid)Comparable to predicate (which is unknown, but "Resistant" is the claim)Resistant to corrosion for a duration equivalent to 3 yearsUnknown
    BiocompatibilityDemonstratedPassed various tests (Cytotoxicity, Maximization sensitization, Acute systemic toxicity, Rabbit Pyrogen Study, Intracutaneous reactivity, Bacterial reverse mutation assay, Intramuscular implantation study)Passed (implied previous results for reference devices)
    Deployment TestingAccuracy and repeatability of the delivery system validated.All components, subassemblies, and/or full devices met required specifications.N/A (Internal test)
    Dimensional TestingReproducibility of stent length and diameter after deployment verified.All components, subassemblies, and/or full devices met required specifications.N/A (Internal test)
    Tensile Strength TestsPerformed for deployment system components with bonds/welds.All components, subassemblies, and/or full devices met required specifications.N/A (Internal test)

    2. Sample Size Used for the Test Set and Data Provenance

    The "test set" in this context refers to the samples used in bench testing.

    • Sample Size: Not explicitly stated for each test, but implied to be sufficient for the validation and comparative testing performed on the stent models (e.g., 10x20 and 20x80 stents) and deployment systems.
    • Data Provenance: The studies are described as "Bench Testing" and "in-vitro tests," meaning they were conducted in a laboratory environment. There is no information regarding country of origin for the data or whether it's retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable for this type of submission. This document pertains to bench testing and substantial equivalence, not a clinical study requiring expert ground truth for interpretation of medical data.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 refer to clinical data review, typically by medical experts for diagnoses or outcomes, which is not part of this bench testing summary.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (stent) submission, not an AI diagnostic or assistive device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device (stent) submission, not an algorithm.

    7. The Type of Ground Truth Used

    For the bench tests, the "ground truth" is defined by the physical properties measured against established engineering and material standards, or by direct comparison to the predicate device's measured performance. For instance, expansion force measurements are compared directly to the predicate's measurements. Biocompatibility relies on standardized tests.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not a machine learning or AI device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141598

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-09-30

    (109 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fuse® Colonoscopy System is intended for diagnostic visualization of the digestive tract. The system also provides access for therapeutic interventions using standard endoscopy tools. Fuse® Colonoscopes, in conjunction with the FuseBox™ processor, are indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve) for adult patients. The Fuse® Colonoscopy System consists of EndoChoice camera heads, endoscopes, video system, light source and other ancillary equipment.

    Device Description

    The Fuse Colonoscopy System is a GI platform indicated for diagnostic visualization and therapeutic intervention of the lower digestive tract in adult patients. The purpose of this submission is to expand the product sizes available for the legally marketed unmodified version Fuse 1C System (K132839) by adding two additional lengths: 133cm and 150cm to the current 168cm Fuse 1C colonoscope.

    This 510(k) submission also presents several design changes to the system. The indications for use, fundamental technology and operation principals of the legally marketed device were not changed. The system labeled for healthcare facilities/hospitals enables physicians to view a high- resolution wide field of view of up to 330°

    AI/ML Overview

    The provided text is a 510(k) summary for the "Fuse® Colonoscopy System," which is a medical device, not an AI algorithm. Therefore, the questions related to AI algorithm performance (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set) are not applicable.

    The document describes modifications to an existing device (Fuse™ 1C System, K132839) and aims to prove substantial equivalence to the predicate device, not to prove the device meets acceptance criteria for an AI algorithm.

    However, I can extract and present the information available regarding the device's performance testing and acceptance, which focuses on safety and effectiveness in comparison to its predicate.

    Here's a breakdown of the relevant information from the document:

    1. Table of Acceptance Criteria and Reported Device Performance (as pertains to device modification):

    The document does not explicitly state "acceptance criteria" in the traditional sense of numerical thresholds for an AI's performance metrics (e.g., AUC > 0.9). Instead, it focuses on demonstrating that the modified device's design outputs meet its design input requirements and that it performs "as well as the predicate" in terms of safety and effectiveness.

    CategoryAcceptance Criteria (Implicit for safety/effectiveness vs. Predicate)Reported Device Performance
    Colonoscope Working LengthIndustry-accepted standard for working lengths. Safety and effectiveness maintained despite adding shorter lengths (133cm, 150cm).Both designs (original and modified with additional lengths) utilize industry-accepted standard for working lengths.
    Angulation Knob BrakeIndustry-accepted standard for handle brake. Safety and effectiveness maintained despite adding R/L angulation.Both designs utilize industry-accepted standard for handle brake.
    Umbilical Cord Plug / SocketIndustry-accepted standard for usability. Safety and effectiveness maintained with integrated design.Both designs utilize industry-accepted standard for usability.
    Post-procedure Reprocessing MethodIndustry-accepted standard for reprocessing compliance. Safety and effectiveness maintained with automated reprocessing option.Both designs utilize industry-accepted standard for reprocessing compliance.
    Insertion Tube CoatingSimilar properties to predicate device's coating (TPU). Biocompatibility and performance maintained.Both materials are TPU, having similar properties.
    AdhesivesSimilar properties to predicate device's adhesives (Epoxy resin). Biocompatibility and performance maintained.Both materials are epoxy resins, having similar properties.
    SoftwareEnhanced features do not negatively impact safety or effectiveness, and may improve performance.Version 1.1.7 (modified) provides enhanced features over Version 1.0 (predicate). Specific performance metrics are not detailed but passed testing.
    General Performance (Safety & Effectiveness)The modified device must be as safe and effective as the predicate device for its intended use. Design outputs must meet design inputs."All test results passed, demonstrating that the device is safe and effective in comparison with the predicate device."

    2. Sample size used for the test set and the data provenance:

    • Not applicable for an AI algorithm.
    • For the device, the performance testing involved "benchtop functional, performance, and software testing," "Laboratory Safety / EMC compatibility re-testing," "Laboratory AER validation," and "Laboratory Biocompatibility testing." The sample sizes for these engineering and lab tests are not specified in this summary, but would typically involve a defined number of units or samples as per particular test standards.
    • Data Provenance: Testing was performed "within EndoChoice's laboratory and by accredited third parties." The origin of input "data" in the sense of patient data is not relevant here as it's a hardware/software device validation, not an AI model trained on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This relates to AI algorithm validation where expert consensus establishes ground truth for classification tasks. The Fuse® Colonoscopy System is a visualization and intervention tool, not an AI diagnostic system.

    4. Adjudication method for the test set:

    • Not applicable. This relates to AI algorithm validation using expert review.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No. This type of study is typically done for diagnostic tools (human readers interpreting images) where AI assistance is being evaluated. The Fuse® Colonoscopy System is an imaging system itself, not an AI assistant to image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This refers to AI algorithm performance. The device itself is "standalone" in the sense that its performance characteristics (e.g., optical quality, mechanical function, electrical safety) are evaluated intrinsically, but it requires a human user (physician) for its clinical application.

    7. The type of ground truth used:

    • For the engineering and laboratory tests performed, "ground truth" implies meeting the specified engineering requirements and international standards. For example, for biocompatibility, the ground truth is adherence to ISO 10993; for electrical safety, it's adherence to IEC 60601-1.
    • It's not "expert consensus, pathology, or outcomes data" in the context of validating an AI diagnostic algorithm.

    8. The sample size for the training set:

    • Not applicable. This refers to AI algorithms. The device does contain software (Version 1.1.7), but this appears to be for system control and image processing, not a machine learning model that requires a "training set" in the AI sense.

    9. How the ground truth for the training set was established:

    • Not applicable. This refers to AI algorithms.

    In summary, the provided document focuses on demonstrating substantial equivalence of a modified medical device to a legally marketed predicate device, rather than the validation of an AI algorithm against specific performance criteria. The evaluation is centered on engineering bench tests, laboratory validations, and compliance with recognized standards, ensuring the modifications do not negatively impact the device's safety and effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140760

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-16

    (82 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BONASTENT® Biliary Stent is indicated for the palliation of malignant strictures in the biliary tree.

    Device Description

    The Bonastent® Biliary stent is a self-expanding polygon mesh surface, tubular prosthesis designed to maintain patency of bile duct strictures caused by malignant tumors. The stent is made of Nitinol wire and is designed in such a way as to prevent migration and turnor ingrowth. The stent is provided preloaded on a delivery device and is available in 2 different diameters (8mm and 10mm) in various lengths.

    AI/ML Overview

    This document describes the 510(k) submission for the Bonastent® Biliary stent, K140760. This submission is for material changes to the delivery device of an already cleared product (K093003) and not for initial market clearance of a novel device or software. Therefore, the information regarding acceptance criteria, study design for performance, and ground truth establishment is significantly different from what would be expected for a software-as-a-medical-device (SaMD) or a new device effectiveness study.

    Based on the provided text, here’s a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this 510(k) submission, the "acceptance criteria" and "reported device performance" are framed around ensuring the safety and effectiveness of the device were maintained despite the material changes to the delivery system. The primary acceptance criterion was that the changes did not negatively impact biocompatibility or mechanical performance.

    Acceptance CriterionReported Device Performance
    Biocompatibility: No new safety issues due to material changes.Passed Cytotoxicity, Sensitization, and Irritation tests.
    Deployment: No negative impact on device deployment.Passed deployment testing.
    Tensile Strength: No negative impact on device tensile strength.Passed tensile strength testing.
    Overall Safety & Effectiveness: Device remains substantially equivalent to the predicate device.All tests support that no new safety and effectiveness issues are raised due to the changes.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for efficacy. Instead, the testing relates to bench testing for the physical device components.

    • Sample Size for Bench Tests: Not explicitly stated in terms of number of devices or components. The text mentions "bench tests were performed."
    • Data Provenance: N/A for clinical data. This refers to laboratory and bench testing of the device components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    N/A. As this submission pertains to material changes and not a clinical efficacy study, there was no "ground truth" derived from expert review of patient data. The "truth" here is established through validated laboratory and engineering tests against predefined specifications for biocompatibility and mechanical performance.

    4. Adjudication Method for the Test Set

    N/A. There was no clinical test set requiring adjudication from experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is typically conducted for diagnostic devices, especially those involving image interpretation by human readers, to assess the impact of AI on their performance. This 510(k) is for a physical medical device (biliary stent) and material changes to its delivery system, not an AI or diagnostic tool.

    6. If a Standalone Performance Study Was Done

    Yes, in a way. The "verification & validation activities" including biocompatibility and bench tests (deployment and tensile strength) can be considered standalone performance studies for the modified delivery device, aiming to demonstrate that it performs as intended and meets safety requirements. These studies are focused on the device's intrinsic mechanical and material properties.

    7. The Type of Ground Truth Used

    For the bench testing, the "ground truth" would be the engineering specifications and scientific standards for biocompatibility (e.g., ISO standards for cytotoxicity, sensitization, irritation) and mechanical performance (e.g., predefined force limits for deployment, tensile strength requirements). The results of these tests are compared against these established technical specifications.

    8. The Sample Size for the Training Set

    N/A. There is no training set mentioned, as this is not a machine learning or AI-based device.

    9. How the Ground Truth for the Training Set Was Established

    N/A. As there is no training set, this is not applicable.