Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K020925

    Validate with FDA (Live)

    Date Cleared
    2003-04-29

    (404 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Carbodissecting Endoscope is designed for use in intraluminal tissue separation during a femoro-popliteal endarterectomy procedure intended to achieve vascular reconstruction. The device is an adjunctive aid in the separation of vessel layers.

    Device Description

    The Carbodissecting Endoscope, sterile and single-use, consists of two components: The spatula is a surgical tool used with carbon dioxide (CO2) and saline for atherectomy using carbodissection. The scope is a flexible fiberscope intended for use with the Carbodissecting Endoscope Spatula to provide visual feedback.

    AI/ML Overview

    The provided 510(k) summary for the "Carbodissecting Endoscope" does not contain information related to acceptance criteria or a study proving the device meets performance criteria in the context of an AI/ML medical device.

    This document details the regulatory pathway for a surgical device (an intraluminal artery stripper) and focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and intended use. The "Performance Testing" section primarily refers to compliance with an electrical safety standard (IEC 601-2-18) and mentions "additional bench testing and biocompatibility testing," but provides no specifics on acceptance criteria or detailed study results for functional performance.

    Therefore, for almost all the requested information, the answer will be "Not Applicable" or "Not Provided in the document" because the device is a physical surgical tool, not an AI/ML algorithm requiring performance metrics like sensitivity, specificity, or human-in-the-loop studies.

    Here's a breakdown based on the provided text, indicating why most items cannot be answered:


    1. Table of Acceptance Criteria and Reported Device Performance

    | Criterion | Acceptance Criteria | Reported Device Performance |
    | :------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
    | Electrical Safety | Compliance with IEC 601-2-18 | Device meets the requirements of IEC 601-2-18 (Safety of Medical Electrical Equipment: Particular Requirements for Endoscopes). |
    | Biocompatibility | Not explicitly stated, but implied as part of safety testing | Biocompatibility testing has been completed. |
    | Bench Testing (Functional) | Not explicitly stated | Additional bench testing has been completed. (No specific functional criteria or results provided in this document). |

    2. Sample size used for the test set and the data provenance

    • Not Applicable. This is a physical surgical device, not an AI/ML algorithm that uses a test set of data. The document does not describe any clinical studies involving patient data for performance evaluation in a hypothesis-testing manner, but rather mentions "bench testing" and "biocompatibility testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. No ground truth or expert review process as described for AI/ML models is mentioned in relation to proving the device's performance for this surgical tool.

    4. Adjudication method for the test set

    • Not Applicable. No test set or adjudication process as typically defined for AI/ML models is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a surgical instrument, not an AI-assisted diagnostic or interpretive device. No MRMC study or AI assistance is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a manual surgical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not Applicable. Ground truth in the context of AI/ML performance evaluation is not relevant to this device's regulatory submission, which focuses on substantial equivalence based on prior predicate devices and safety testing.

    8. The sample size for the training set

    • Not Applicable. This is a physical device, not an AI/ML algorithm. There is no concept of a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. Consistent with item 8.

    Summary regarding AI/ML related questions:

    The provided 510(k) summary is for a traditional surgical device and predates the common regulatory framework for AI/ML medical devices. Therefore, the questions related to AI/ML specific performance criteria, data sets, ground truth, and human-in-the-loop studies are not applicable to the information contained within this document. The "Performance Testing" section primarily refers to compliance with an electrical safety standard, bench testing (without specifics), and biocompatibility testing, which are standard for physical medical devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013680

    Validate with FDA (Live)

    Date Cleared
    2002-03-04

    (117 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To gently dissect planes of soft tissue using controlled bursts of CO2 gas. The device may be used in both open and endoscopic procedures in which gentle, blunt dissection of soft tissue planes is desired.

    Device Description

    The Carbodissecting Endoscope consists of two components: The Spatula is a surgical tool used in combination with CO2 and saline, and is designed to be used with the supplied Scope to provide accurate visual feedback during a carbodissection procedure. The tip allows the surgeon to slide between the tissue planes. The front of the tip is slightly curved and has multiple CO2 outlets. The multiple CO2 outlets are used to fan the gas out along the entire front of the tip to help peel apart the two layers, to aid in the separation of the tissue The Scope is a flexible fiberscope intended for use with the Spatula.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving that a device meets such criteria. The document is an FDA 510(k) premarket notification summary for a "Carbodissecting Endoscope," indicating it has been deemed substantially equivalent to a predicate device.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria based on the provided text. The requested information regarding sample size, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, type of ground truth, and training set details are not present.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1