Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K230893

    Validate with FDA (Live)

    Date Cleared
    2023-10-23

    (206 days)

    Product Code
    Regulation Number
    876.4480
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swiss LithoClast® Trilogy is indicated for fragmentation and removal of urinary tract calculi in the kidney, ureter and bladder.

    Device Description

    The Swiss LithoClast® Trilogy offers three modes of lithotripsy: 1) ultrasound alone; 2) shock wave alone; or 3) combined ultrasound and shock wave.. The LithoClast Trilogy system consists of the console used to set the treatment parameters and generate the treatment energy, a dual-energy reusable handpiece, and a variety of probe sizes to enable use of the system with a wide range of commercially available endoscopes. Delivery of energy is controlled using a two-step foot pedal. The purpose of this 510(k) is to increased the number of validated reprocessing cycles for the reusable handpiece and add the option of a wireless foot pedal.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device, the Swiss LithoClast® Trilogy, which is an electrohydraulic lithotriptor used for fragmenting and removing urinary tract calculi. The submission is to address two changes: the addition of an optional wireless foot pedal and an increase in the validated number of reprocessing cycles for the reusable handpiece from 60 to 100.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly present a single table labeled "Acceptance Criteria and Reported Device Performance" for the overall device's indication. Instead, it describes performance testing related to the changes being submitted in this 510(k). The implicit acceptance criteria are that the device with the changes performs equivalently to the predicate device and meets relevant safety and performance standards.

    CharacteristicAcceptance CriteriaReported Device Performance
    Wireless Foot Pedal - Software ValidationThe software in the wireless foot pedal and dongle must be validated commensurate with a moderate level of concern and compliant with IEC 62304 Edition 1.1 2015-06. All unit, integration, and system level tests must meet acceptance criteria.All tests met the acceptance criteria and were passed, demonstrating compliance with IEC 62304.
    Wireless Foot Pedal - Electrical Safety & EMCThe Swiss LithoClast Trilogy, when operated by the wireless foot pedal, must comply with IEC 60601-1:2005 + A1:2012 and IEC 60601-1-2:2014. The Wireless Foot Pedal Kit must also comply with EN 300 328 V2.2.2 (2019-07) and EN 62311:2008.Demonstrated compliance with all specified electrical safety and EMC standards.
    Wireless Foot Pedal - Device Performance TestingThe wireless foot pedal and dongle must meet the design input requirements for functional performance, battery performance, mechanical integrity, dongle connection/disconnection integrity, chemical resistance, expected use lifetime, and RF quality of service and coexistence.All test results demonstrated that the wireless foot pedal and dongle meet the design input requirements.
    Handpiece Lifetime Testing (Reprocessing Cycles)Handpieces must meet performance specifications and be effective in clearing stones in a simulated use set-up within 20% of the time needed using new handpieces (at time 0) after 133 cleaning, disinfection, and steam sterilization cycles. This must demonstrate validation for 100 reprocessing cycles with 95% reliability and 95% confidence.Performance testing confirmed that handpieces met performance specifications and were effective in clearing stones within 20% of the time needed by new handpieces after 133 reprocessing cycles. The conclusion was validation for 100 reprocessing cycles with 95% reliability and 95% confidence.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Wireless Foot Pedal Testing (Software, Electrical Safety, Device Performance): The document does not specify a distinct "test set" in terms of patient data or clinical samples for the wireless foot pedal. The testing described is primarily engineering and bench testing on the device hardware and software itself. Therefore, sample sizes would relate to the number of production units tested, but this detail is not provided.
    • Handpiece Lifetime Testing:
      • Sample Size: Twenty (20) handpiece units were tested.
      • Data Provenance: This was bench testing conducted by the manufacturer, not patient data. The country of origin of this data is not explicitly stated but would be from EMS Electro Medical Systems SA, Nyon, Switzerland, or a contract testing facility. This is prospective testing performed for the purpose of this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable to the studies described. The "ground truth" for these tests relates to engineering specifications and performance benchmarks (e.g., successful software operation, compliance with electrical safety standards, stone clearance within a specified time). These are not studies involving human interpretation of data for which expert consensus would establish ground truth in a clinical context.

    4. Adjudication Method for the Test Set:

    Not applicable. There was no clinical test set requiring expert adjudication. The tests were objective engineering and performance verification.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The changes addressed in this 510(k) are related to an optional wireless foot pedal and increased reprocessing cycles for a handpiece, neither of which involves human readers interpreting cases or AI assistance in a diagnostic or interventional workflow that would necessitate an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The term "standalone" performance typically refers to the performance of an AI algorithm independent of human interaction. This submission does not involve an AI algorithm with such a function. The software validation mentioned for the wireless foot pedal assesses the functional correctness and safety of the embedded software, not its standalone performance in interpreting medical data.

    7. The Type of Ground Truth Used:

    • Wireless Foot Pedal Testing: The ground truth was based on pre-defined engineering design input requirements and international standards for electrical safety, EMC, and wireless communication (e.g., functional performance specifications, battery life targets, compliance with IEC 60601 series, EN 300 328, EN 62311).
    • Handpiece Lifetime Testing: The ground truth was based on performance specifications for stone clearance (i.e., effectiveness in clearing stones in a simulated use set-up within 20% of new handpiece performance) and statistical reliability/confidence targets (95% reliability, 95% confidence for 100 cycles).

    8. The Sample Size for the Training Set:

    Not applicable. There is no AI model or algorithm being trained with a dataset mentioned in this 510(k) submission. The software validation is for embedded control software, not for an AI/ML component that learns from data.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there was no training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220538

    Validate with FDA (Live)

    Date Cleared
    2022-07-11

    (136 days)

    Product Code
    Regulation Number
    890.5660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DolorClast® Radial is indicated for:

    • Relief of minor muscle aches and pains
    • Temporary increase in local blood circulation
    • Activation of connective tissue
    Device Description

    The DolorClast® Radial is a portable radial pressure pulse device with an integrated air compressor. The DolorClast® Radial system is composed of the DolorClast® Radial console, handpiece, interchangeable applicator heads, and optional cart. Swiss DolorClast® Gel, previously approved under P050004, is also provided for coupling the treatment applicator to the patient's skin.
    The DolorClast® Radial console generates and controls the pressure pulse frequency and intensity. Air pressure from the console drives a projectile within the handpiece against the applicator head to generate the pressure pulses at the set frequency and intensity. Device activation is done by the practitioner via a trigger button on the handpiece. The applicator at the distal end of the handpiece transmits the radial pressure pulses to the targeted tissues via the coupling gel applied to the treated skin.

    AI/ML Overview

    This document is an FDA 510(k) summary for a medical device (DolorClast® Radial), which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific acceptance criteria in the manner of a clinical trial for diagnostic performance. Therefore, many of the requested elements (e.g., sample sizes for training/test sets, ground truth establishment for AI, MRMC studies, expert qualifications) are not applicable or not present in this type of regulatory submission.

    However, based on the provided text, I can extract information related to the device's functional performance testing against internal protocols.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes several functional performance tests and generally states that "All tests were passed," indicating the device met internal EMS protocols. Specific numerical acceptance criteria are generally not provided in this summary, but the reported performance is that the device met those criteria.

    Acceptance Criterion (Internal EMS Protocol)Reported Device Performance
    System Functional Performance Testing
    Compressor activation only when handpiece trigger pressedAll tests were passed (confirming activation met pre-defined tolerance).
    Pressure output at each pressure setting met pre-defined pressure setting toleranceAll tests were passed (confirming pressure output met pre-defined tolerance).
    Correct operation of console at all treatment settings (including Smart Protocols and Special Modes)All tests were passed (confirming correct operation).
    Correct device response to error conditionsAll tests were passed (verifying correct device response).
    Correct operation of handpiece (pressure adjustment, starting/stopping treatment, recording piston use)All tests were passed (confirming correct operation).
    Acoustic output characterization (Displacement of applicator heads)Displacement of applicator heads: 0.18 - 0.28 mm
    Acoustic output characterization (Maximum penetration depth)Maximum penetration depth: 40 mm
    Acoustic output characterization (Energy flow density)0.29 mJ/mm² at 4 bar; 0.14 mJ/mm² at 2.4 bar
    Acoustic output characterization (Positive peak pressure amplitudes)17 MPa at 4 bar; 11.24 MPa at 2.4 bar
    Acoustic output characterization (Negative peak pressure amplitudes)10 MPa at 4 bar; 7.2 MPa at 2.4 bar
    Acoustic output characterization (Derived focal acoustic pulse energy)5.9 mJ at 4 bar*; 2.2 mJ at 2.4 bar*
    Acoustic output characterization (Derived pulse-intensity integral (max))0.29 mJ/mm² at 4 bar; 0.14 mJ/mm² at 2.4 bar
    Acoustic output characterization (Rise time)3.2 µs
    Acoustic output characterization (Compressional pulse duration)2.6 µs (1st peak); 62.7 µs (1st phase)
    Lifetime Testing
    Console functions and meets specs for 7 years or 50,000,000 pulsesAll tests were passed (confirming specified useful lifetime).
    Handpiece functions and meets specs for 2 years or 5,000,000 pulsesAll tests were passed (confirming specified useful lifetime).
    Applicators function and meet specs for 2 years or 5,000,000 pulsesAll tests were passed (confirming specified useful lifetime).
    Biocompatibility (Applicator tips and gel)Previously approved under P050004; no changes to material/manufacturing.
    Transportation Testing (ISTA 2A and ISTA 3A)All tests were passed (devices and accessories functional after conditioning).
    Software Validation (IEC 62304 Ed 1.1)Software validated in software testing and system level testing.
    Electrical Safety and EMC (IEC 60601-1:2005 + A1:2012, IEC 60601-1-2:2014)Demonstrated to comply with stated standards.

    Note: The document states "Averaged across all applicator sizes" for derived focal acoustic pulse energy.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated for any of the performance tests. The testing described appears to be engineering verification and validation, likely involving a limited number of test units or prototypes.
    • Data Provenance: The origin of the data is stated as "internal EMS protocols" and "internal EMS testing." It is prospective testing conducted specifically for this regulatory submission on the device itself. No mention of geographical data origin (e.g., country) is made as it's not a clinical data study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not a study assessing diagnostic performance against a "ground truth" established by experts. The "ground truth" for these tests refers to engineering specifications and regulatory standards.

    4. Adjudication method for the test set

    Not applicable. This is not a study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a therapeutic massager, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical therapeutic device, not an algorithm. Its performance is inherent in its physical operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" effectively used for this regulatory submission is adherence to engineering specifications, internal EMS protocols, and recognized industry standards (e.g., IEC 60601-1, IEC 60601-1-2, ISTA 2A/3A). For biocompatibility, the "ground truth" was established by prior approval (P050004) of identical materials.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191124

    Validate with FDA (Live)

    Date Cleared
    2019-05-29

    (30 days)

    Product Code
    Regulation Number
    876.4480
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swiss LithoClast® Trilogy is indicated for fragmentation and removal of urinary tract calculi in the kidney, ureter and bladder.

    Device Description

    The Swiss LithoClast® Trilogy offers three modes of lithotripsy: 1) ultrasound alone; 2) shock wave alone; or 3) combined ultrasound and shock wave.. The LithoClast Trilogy system consists of the console used to set the treatment parameters and generate the treatment energy, a dual-energy reusable handpiece, and a variety of probe sizes to enable use of the system with a wide range of commercially available endoscopes. Delivery of energy is controlled using a two-step foot pedal.

    Two models of the LithoClast Trilogy are available: one with a peristaltic pump and one with a pinch valve. Both versions can be used to suction stone fragments into the optional Stone Catcher. An external vacuum system is required to enable suction with the pinch valve model of the console.

    AI/ML Overview

    This is an FDA 510(k) summary for a medical device (Swiss LithoClast® Trilogy), not an AI/algorithm-driven device. Therefore, much of the requested information (such as AI performance metrics, sample sizes for AI training/test sets, expert adjudication methods, MRMC studies, or standalone algorithm performance) is not applicable or available in this document.

    However, I can extract the relevant acceptance criteria and study information related to the device accessory.

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    No adverse impact on the handpiece sterile barrier system (pouch or wrap) during the steam sterilization process due to the use of the new Steri Holder accessory."Performance testing was conducted to verify that use of the Steri Holder during steam sterilization has no adverse impact on the handpiece sterile barrier system (pouch or wrap) during the steam sterilization process." (Implicitly, the device met this criterion by passing the testing).

    Study Details (as applicable to this non-AI device modification):

    1. Sample size used for the test set and the data provenance: Not specified in the document. The testing involved evaluating the sterile barrier system of the handpiece when sterilized with the Steri Holder. The document does not provide details on the number of handpieces or sterilization cycles tested, nor the provenance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For this type of device modification, ground truth would typically be established through standardized sterilization validation protocols and integrity testing, not "experts" in the context of diagnostic interpretation.
    3. Adjudication method for the test set: Not applicable in the context of expert consensus. Performance testing results are typically evaluated against pre-defined engineering and sterility validation standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document pertains to a medical device (lithotripter accessory), not an AI diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.
    6. The type of ground truth used: For this specific modification, the "ground truth" would be established by the validated integrity of the sterile barrier system after sterilization, assessed through standard sterility and packaging integrity tests (e.g., visual inspection, dye penetration, microbial ingress tests). The document states the testing was to verify "no adverse impact on the handpiece sterile barrier system."
    7. The sample size for the training set: Not applicable. There is no training set for an AI algorithm.
    8. How the ground truth for the training set was established: Not applicable. There is no training set for an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182490

    Validate with FDA (Live)

    Date Cleared
    2018-10-09

    (28 days)

    Product Code
    Regulation Number
    876.4480
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swiss LithoClast® Trilogy is indicated for fragmentation and removal of urinary tract calculi in the kidney, ureter and bladder.

    Device Description

    The Swiss LithoClast® Trilogy has three possible modes of operation: 1) pneumatic lithotripsy alone; 2) ultrasound lithotripsy alone; and 3) and combined pneumatic and ultrasound lithotripsy.

    The LithoClast Trilogy system consists of the console used to set the treatment parameters and generate the treatment energy, a reusable handpiece, and a variety of probe sizes to enable use of the system with a wide range of commercially available endoscopes. Delivery of energy is controlled using a two-step foot pedal.

    Two models of the LithoClast Trilogy are available: one with a peristaltic pump and one with a pinch valve. Both versions can be used to suction stone fragments into the optional Stone Catcher. An external vacuum system is required to enable suction with the pinch valve model of the console.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the "Swiss LithoClast® Trilogy". This document primarily focuses on demonstrating substantial equivalence to a predicate device for the purpose of adding new sterilization methods for a reusable handpiece. It does not contain information about a study proving the device meets acceptance criteria related to its clinical performance (e.g., fragmentation and removal of urinary tract calculi) or a multi-reader multi-case (MRMC) comparative effectiveness study.

    Therefore, many of the requested sections regarding clinical performance, expert ground truth, sample sizes for test and training sets, and human reader improvement with AI assistance cannot be extracted from this document.

    However, I can extract information related to the sterilization validation, which is the core of this particular 510(k) submission.

    Here's a summary of what can be extracted:

    1. Table of acceptance criteria and the reported device performance (for sterilization):

    Acceptance Criterion (for sterilization)Reported Device Performance (for sterilization)
    Achievement of Sterility Assurance Level (SAL) 10⁻⁶Demonstrated SAL 10⁻⁶ for each sterilizer

    2. Sample sized used for the test set and the data provenance: Not applicable. This document describes a sterilization validation, not clinical performance testing on a test set of patient data. The "test set" in this context refers to the items (e.g., handpieces) subjected to the sterilization process. The provenance is internal validation testing by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as this is a sterilization validation, not a study requiring expert clinical interpretation.

    4. Adjudication method for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, an MRMC study was not done, as this document concerns a medical device for stone fragmentation, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No, this is not an AI algorithm.

    7. The type of ground truth used:

    • For sterilization: The ground truth for sterilization validation is the scientific standard of achieving a Sterility Assurance Level (SAL) of 10⁻⁶, based on established microbiological testing methods.

    8. The sample size for the training set: Not applicable, as this is not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established: Not applicable.


    Additional information related to the sterilization validation from the document:

    • Sterilization Standard: The validation was conducted in accordance with ISO 14937:2009 "Sterilization of health care products. General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices."
    • Sterilization Parameters: The recommended sterilization parameters for each H₂O₂ sterilizer were used.
    • Sterilizers Validated:
      • STERIS V-PRO 1
      • STERIS V-PRO 60
      • STERIS V-PRO maX
      • STERRAD 100S Short
      • STERRAD NX Standard
    • Device Component Subject to Validation: The reusable Trilogy handpiece.

    In summary, this document provides evidence for the sterilization efficacy of the reusable handpiece, but not for the clinical performance of the device in fragmenting and removing urinary tract calculi.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181997

    Validate with FDA (Live)

    Date Cleared
    2018-08-30

    (35 days)

    Product Code
    Regulation Number
    876.4480
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swiss LithoClast Trilogy is indicated for fragmentation and removal of urinary tract calculi in the kidney, ureter and bladder.

    Device Description

    The Swiss LithoClast® Triloqy is an intracorporal lithotripter with three possible modes of operation: 1) pneumatic lithotripsy alone; 2) ultrasound lithotripsy alone; and 3) and combined pneumatic and ultrasound lithotripsy.

    The LithoClast Trilogy system consists of the console used to set the treatment parameters and generate the treatment energy, a reusable handpiece, and a variety of probe sizes to enable use of the system with a wide range of commercially available endoscopes. Delivery of energy is controlled using a twostep foot pedal.

    Two models of the LithoClast Trilogy are available: one with a peristaltic pump and one with a pinch valve. Both versions can be used to suction stone fragments into the optional Stone Catcher. An external vacuum system is required to enable suction with the pinch valve model of the console.

    AI/ML Overview

    I am sorry, but based on the provided text, there is no information about acceptance criteria, reported device performance, or any study details that would allow me to populate the requested table and answer the study-related questions.

    The document is an FDA 510(k) clearance letter for the "Swiss LithoClast Trilogy". It describes a special 510(k) submission for an engineering modification to improve the durability of a reusable handpiece. It states that "The revised handpiece met the acceptance criteria of the validation testing," but it does not provide what those acceptance criteria were, what the reported performance was, or any details about the validation study itself.

    Therefore, I cannot provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181364

    Validate with FDA (Live)

    Date Cleared
    2018-06-20

    (28 days)

    Product Code
    Regulation Number
    876.4480
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swiss LithoClast Trilogy is indicated for fragmentation and removal of urinary tract calculi in the kidney, ureter and bladder.

    Device Description

    The Swiss LithoClast® Trilogy has three possible modes of operation: 1) pneumatic lithotripsy alone; 2) ultrasound lithotripsy alone; and 3) and combined pneumatic and ultrasound lithotripsy.

    The LithoClast Trilogy system consists of the console used to set the treatment parameters and generate the treatment energy, a reusable handpiece, and a variety of probe sizes to enable use of the system with a wide range of commercially available endoscopes. Delivery of energy is controlled using a twostep foot pedal.

    Two models of the LithoClast Trilogy are available: one with a peristaltic pump and one with a pinch valve. Both versions can be used to suction stone fragments into the optional Stone Catcher. An external vacuum system is required to enable suction with the pinch valve model of the console.

    AI/ML Overview

    This document describes a Special 510(k) submission for the Swiss LithoClast® Trilogy device. The primary purpose of this submission is to add hydrogen peroxide (H2O2) sterilization as an option for reprocessing the reusable handpiece, in addition to the previously validated steam sterilization. Therefore, the acceptance criteria and study described below specifically pertain to the validation of this new sterilization method for the handpiece.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    For the Handpiece: Successful validation of H2O2 sterilization method to achieve a sterility assurance level (SAL) of 10-6 in accordance with ISO 14937:2009 using the STERRAD 100NX sterilizer on the Standard cycle, ensuring the device remains safe and effective.**Sterilization validation was conducted using the STERRAD 100NX sterilizer on the Standard cycle. The study demonstrated the achievement of sterility assurance level (SAL) 10-6 for the Swiss LithoClast® Trilogy reusable handpiece. This indicates the new sterilization method effectively renders the handpiece sterile to the required standard without compromising its functionality or safety. The submission concludes that "the Swiss Lithoclast® Trilogy reusable handpiece, when sterilized using the STERRAD 100NX H2O2 sterilizer, is substantially equivalent to the handpiece when sterilized using steam."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific "sample size" in terms of a number of handpieces tested for the H2O2 sterilization validation. However, sterilization validation studies typically involve:
    * Multiple samples (e.g., biological indicators, inoculated devices) within each test run to demonstrate the killing of a specified number of microorganisms.
    * Multiple replicate cycles (e.g., three consecutive successful cycles) to prove reproducibility.

    The data provenance is related to the performance of the Swiss LithoClast® Trilogy reusable handpiece when subjected to H2O2 sterilization. Given that this is a Special 510(k) for a manufacturing change (adding a sterilization method), the data would be prospective and generated specifically for this validation study. The location of the test facilities or the country of origin of the data is not specified beyond the applicant being EMS Electro Medical Systems SA in Nyon, Switzerland.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of study does not involve "experts" establishing a ground truth in the same way clinical diagnostic studies do. The "ground truth" for sterilization validation is fundamentally defined by the successful inactivation of biological indicators (BIs) or inoculated devices, as measured by standard microbiological assays (e.g., lack of growth). The expertise lies in:
    * Microbiologists and sterilization engineers who design, execute, and interpret the validation studies according to ISO 14937 and other relevant standards.
    * Regulatory affairs personnel who ensure compliance with medical device regulations.

    The document does not specify the number or qualifications of these individuals, as this is standard practice in such validation studies.

    4. Adjudication Method for the Test Set

    Not applicable for a sterilization validation study. The outcome (sterile or not sterile, SAL achieved or not achieved) is determined by objective laboratory testing (microbiological culture results) rather than expert adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No. This is not a diagnostic device or a device whose performance is assessed by human readers. It is a device sterilization validation.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    N/A. This is not an AI/algorithm-based device. "Standalone performance" in this context would refer to the device's functional integrity and sterility post-sterilization, which is inherently part of the validation.

    7. The Type of Ground Truth Used

    The ground truth used for this sterilization validation is microbiological sterility testing. This involves:
    * Inoculating representative devices or biological indicators (BIs) with a known, resistant microorganism (e.g., Geobacillus stearothermophilus for H2O2 sterilization).
    * Processing these inoculated items through the sterilization cycle.
    * Aseptic retrieval and culture of the BIs or device samples to determine if any viable microorganisms remain.
    * The "ground truth" is established by the lack of microbial growth (kill) conforming to the specified SAL (10-6).

    8. The Sample Size for the Training Set

    Not applicable. This is a sterilization validation study, not a machine learning study that would use training sets.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of validation. The ground truth (sterility) is established through standardized microbiological testing methods as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123038

    Validate with FDA (Live)

    Date Cleared
    2013-01-16

    (110 days)

    Product Code
    Regulation Number
    876.4480
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SWISS LITHOBREAKER is intended to be used for the fragmentation of urinary tract calculi in the kidney, ureter, and bladder.

    Device Description

    The EMS SWISS LITHOBREAKER is a handheld, standalone, battery operated intracorporeal lithotripter which uses a mechanical impact drive to fragment urinary tract calculi in the kidney, ureter, and bladder. The device generates around 3 impacts in continuous mode per second which are transmitted through the probes on the stones or a lower frequency by "single-pulse" mode, depending on the activation time of the button. The probes are introduced into the working channels of suitable endoscopes. The energy is sustained and stable throughout the life of the battery and the frequency drops only at the end of the battery life. An optional suction handpiece allows extraction of fragmented stones.

    The proposed device consists of the following:

    • Handpiece including powerpack .
    • LithoClast probes, non-sterile reusable and sterile single use .
    • LithoVac lv3 suction system with suction tubes .

    The SWISS LITHOBREAKER is also supplied with accessories for cleaning (endcap), periodic maintenance (O-rings), probe accessories (interface and probe guide), AAA batteries and sterile drapes.

    AI/ML Overview

    The provided 510(k) summary for the E.M.S. Electro Medical Systems SA SWISS LITHOBREAKER outlines the device's technical specifications and a summary of non-clinical performance testing. However, it does not include a table of specific acceptance criteria and detailed reported device performance for each criterion, nor does it describe a study that explicitly "proves the device meets the acceptance criteria" in a typical clinical study format.

    Based on the provided document, here's an analysis of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document describes performance specifications that the device fulfills, but does not provide a quantitative table of acceptance criteria with corresponding performance metrics. It generally states that the device "fulfills the prospectively defined performance specifications" and that testing "demonstrated that the SWISS LITHOBREAKER fulfills the prospectively defined performance specifications."

    The performance aspects tested and considered in comparison to predicate devices were:

    • Probe Tip Displacement
    • Probe Tip Velocity
    • Stone Breakage/Clearance
    • Stone Displacement (Retropulsion)
    • Energy Output

    Without specific numbers or target ranges, a formal table cannot be created from the provided text. The document asserts that the results of this testing confirm that the SWISS LITHOBREAKER is substantially equivalent to the predicates based on these parameters.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for the non-clinical performance testing. It also does not specify the data provenance in terms of country of origin or whether the testing was retrospective or prospective, although the statement "prospectively defined performance specifications" implies the criteria were set before testing. The testing appears to be laboratory-based rather than involving human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable as the described testing is non-clinical (laboratory-based performance testing of mechanical characteristics) and does not involve establishing ground truth from expert assessments of medical images or patient outcomes.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable as the described testing is non-clinical and does not involve expert review or adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done as the device is a medical instrument (lithotripter) and not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone medical instrument. The performance testing described (Probe Tip Displacement, Probe Tip Velocity, Stone Breakage/Clearance, Stone Displacement (Retropulsion), Energy Output) inherently evaluates the device's standalone mechanical performance. There is no "algorithm" in the context of AI being evaluated here; it's a mechanical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance testing, the "ground truth" would be established by physical measurements and standardized testing methodologies using calibrated equipment and reference materials (e.g., artificial stones for stone breakage/clearance tests). It would not involve expert consensus, pathology, or outcomes data, as these are clinical measures.

    8. The sample size for the training set

    This information is not applicable as the SWISS LITHOBREAKER is a mechanical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as the SWISS LITHOBREAKER is a mechanical medical device, not an AI model that requires a training set.


    In summary:

    The 510(k) summary primarily relies on non-clinical performance testing to demonstrate substantial equivalence to predicate devices, rather than clinical studies with human subjects or AI-specific evaluation metrics. The "acceptability criteria" are implied through the fulfillment of "prospectively defined performance specifications" and comparison to known predicate devices on specific mechanical performance characteristics (probe tip displacement, velocity, stone breakage, retropulsion, energy output), without providing granular data or specific numerical targets. The basis for clearance is demonstrating that the new device performs comparably to already cleared devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110173

    Validate with FDA (Live)

    Date Cleared
    2011-04-07

    (77 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AIR-FLOW MASTER PIEZON combines the functions of an ultrasonic scaler and airpolishing unit within a single chassis. The AIR-FLOW MASTER PIEZON is intended for use in the following dental and periodontal applications:

    • Removing supra and subgingival calculus deposits and stains from teeth .
    • Periodontal pocket lavage with simultaneous ultrasonic tip movement .
    • Scaling and root planing .
    • Releasing crowns, bridges, inlays, and posts as well as condensing gutta percha .
    • Plugging for amalgam condensation .
    • Amalgam burnishing .
    • Preparing, cleaning and irrigating root canals .
    • Cavity preparation .
    • Cementing inlays and onlays .
    • · Retrograde preparation of root canals

    The AIR-FLOW MASTER PIEZON is intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of teeth.

    The AIR-FLOW MASTER PIEZON can be used for the following cleaning procedures:

    • plaque removal for placement of sealants .
    • surface preparation prior to bonding/cementation of inlays, onlays, crowns and veneers .
    • surface preparation prior to placing composite restorations .
    • effective plaque and stain removal for orthodontic patients .
    • cleaning prior to bonding ortho brackets .
    • cleaning implant fixture prior to loading .
    • · stain removal for shade determination
    • plaque removal prior to fluoride treatment .
    • plaque and stain removal prior to whitening procedure .

    The AIR-FLOW MASTER PIEZON is also intended for use as an air-polisher in patients suffering from periodontal disease. The AIR-FLOW MASTER PIEZON is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.

    Device Description

    The EMS AIR-FLOW MASTER PIEZON is a dental device that combines the functions of an ultrasonic scaler and air-polishing unit within a single chassis. The proposed device consists of the base control unit, hoses and connectors for two handpieces, and a foot control containing a four-position switch. There are two mounts on the top of the base unit, one for an irrigation liquid container and the other for one of the two air-polishing chambers supplied with the device.

    The AIR-FLOW MASTER PIEZON is supplied with the Piezon Handpiece LED for performing ultrasonic scaling functions. The proposed device is compatible with EMS instruments legally marketed for ultrasonic scaling procedures.

    The AIR-FLOW MASTER PIEZON is also supplied with the AIR-FLOW Handpiece and PERIO-FLOW Handpiece for performing air-polishing procedures. The proposed device is compatible with the AIR-FLOW CLASSIC (sodium bicarbonate), AIR-FLOW SOFT (glycine), and AIR-FLOW PERIO (glycine) prophylaxis powders.

    The AIR-FLOW MASTER PIEZON is supplied with accessories for attaching and removing instruments and nozzles from the handpieces and containers for storage and sterilization of the reusable components and accessories.

    AI/ML Overview

    Here's an analysis of the provided text regarding the EMS AIR-FLOW MASTER PIEZON, focusing on acceptance criteria and supporting studies, formatted as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Specifications)Reported Device Performance
    Compliance with ANSI/AAMI ST79:2006 (Steam Sterilization & Sterility Assurance)Fulfilled
    Compliance with ISO 17664 (2004) (Sterilization of Medical Devices Information to be Provided by the Manufacturer)Fulfilled
    Compliance with ANSI/AAMI/ISO 17665-1:2006 (Sterilization, Moist Heat)Fulfilled
    Compliance with IEC 60601-1:1988 +A1: 1991 + A2: 1995 (Medical Electrical Equipment Safety)Fulfilled
    Compliance with IEC 60601-1-2 (Medical Electrical Equipment EMC Requirements and Tests)Fulfilled
    Compliance with ISTA 2A (2008) (Pre-shipment Testing for Packaged Products)Fulfilled
    Substantial Equivalence to Predicate Devices (AIR-FLOW S2, PIEZON MASTER 700, AIR-FLOW MASTER, AIR-FLOW MASTER STANDARD) based on: - Intended Use - Operational Characteristics - Functional Technological CharacteristicsDemonstrated through direct comparison and testing of integrated functionalities. The device fulfills prospectively defined design and performance specifications.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical performance testing against established standards. These involve engineering and laboratory tests, not tests on a "test set" in the sense of patient data. Therefore, notions of sample size and data provenance (country of origin, retrospective/prospective) are not applicable in the context of this submission. The testing was performed on the device itself and its components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the study involved non-clinical performance testing against pre-defined engineering and medical device standards, not clinical diagnostic accuracy or efficacy requiring expert ground truth in the traditional sense. The "ground truth" here is the adherence to the specifications outlined in the standards.

    4. Adjudication Method for the Test Set

    This is not applicable as the evaluation involved non-clinical performance testing against established standards. There was no "test set" requiring adjudication by clinicians for diagnostic accuracy or outcome.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC study was conducted. This device is an ultrasonic scaler and air-polishing unit, not an AI software or diagnostic tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No standalone algorithm performance study was done. This device is a physical dental instrument, not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the compliance with recognized international and national consensus standards for medical device safety, electrical safety, electromagnetic compatibility, and sterilization/reprocessing. These standards define objective, measurable criteria for performance.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical dental instrument, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093723

    Validate with FDA (Live)

    Date Cleared
    2010-09-08

    (279 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The miniMaster LED is an ultrasonic scaler that is intended for the following:

    • Removing supra and subgingival calculus deposits and stains from the teeth
    • Periodontal pocket lavage with simultaneous ultrasonic tip movement
    • Scaling and root planning
    • Releasing crowns, bridges, inlays and posts as well as condensing gutta percha
    • Plugging for amalgam condensation
    • Amalgam burnishing
    • Preparing, cleaning, and irrigating root canals
    • Preparing approximal cavities
    • Cementing inlays and onlays
    • Retrograde preparation of root canals
    Device Description

    The miniMaster LED is an ultrasonic scaler consisting of a main chassis containing an external electric power supply, controls and displays, ultrasonic generator, and a bottle-fed irrigation system. A 2-step footswitch is connected to the main chassis by a footswitch cord. A handpiece is connected to the main chassis by a handpiece cord, with irrigant flow control located on the cord itself. Instruments designed for specific dental procedures are attached to the distal end of the handpiece.

    The miniMaster LED is a modification of the miniMaster Ultrasonic Scaler that was the subject of K050710. The overall design of the miniMaster LED is identical to the design of the miniMaster. The introduction of Piezon handpiece LED, materials changes to the peristaltic pump tubing, and design modifications implemented for the miniMaster LED were made to improve the functional performance and ease of use of the ultrasonic scaler. These design changes are listed in Table 1.

    AI/ML Overview

    The provided text describes the 510(k) summary for the miniMaster LED Ultrasonic Scaler, which is a medical device. The information below is extracted and organized to answer your specific questions.


    Acceptance Criteria and Device Performance Study for miniMaster LED Ultrasonic Scaler

    The miniMaster LED Ultrasonic Scaler is a modified version of the miniMaster Ultrasonic Scaler (K050710), with enhancements primarily focused on illumination and material changes for improved functional performance and ease of use. The acceptance criteria for this device were met through non-clinical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardReported Device Performance
    Electrical SafetyUL 60601-1, "Medical Electrical Equipment—Part 1: General Requirements for Safety" (2nd edition including amendments 1&2)The device met the general requirements for safety as defined by UL 60601-1.
    Electromagnetic Compatibility (EMC)IEC 60601-1-2, "Medical electrical equipment. Part 1 General requirements for safety - Part 2 Collateral standard: Electromagnetic compatibility: Requirements and Tests" (1993)The device met the electromagnetic compatibility requirements as defined by IEC 60601-1-2.
    Package IntegrityISTA 2A, "Packaged-Products 150 lb (68 kg) or Less"The packaging met the standards for integrity during transport as defined by ISTA 2A.
    Sterilization ValidationANSI/AAMI ST79:2006, "Comprehensive guide to steam sterilization and sterility assurance in health care facilities" (for reprocessing instructions)Reprocessing instructions for reusable components (handpiece, instruments, accessories) were validated to ensure sterility assurance.
    ANSI/AAMI/ISO 17665-1:2006 "Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices" (for steam sterilization process)The steam sterilization process was validated to achieve a Sterility Assurance Level (SAL) of $10^{-6}$ using a half-cycle method.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in the context of clinical data, as no clinical testing was submitted. For the non-clinical tests, the sample size is not explicitly mentioned, but it would typically involve a statistically relevant number of devices or components to demonstrate compliance with the referenced standards. The data provenance is derived from non-clinical laboratory and engineering testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. No clinical testing was performed, and therefore, no expert-established ground truth for a clinical test set was required or reported. The ground truth for the non-clinical tests were the measured values and compliance with the specified engineering and safety standards.

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical testing was performed, there was no need for an adjudication method by medical experts. Compliance with non-clinical engineering and safety standards is typically determined through direct measurement and comparison to predefined thresholds.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. The miniMaster LED is an ultrasonic scaler, a physical dental device, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm-only) performance study was not relevant or performed. This device is a physical, electromechanical tool used by a human operator, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for this device was engineering standards compliance and objective measurements during non-clinical safety, performance, and sterilization validation tests.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML product and does not involve a "training set" in the machine learning sense. The "training" in the context of a medical device refers to the design, development, and iterative testing process, not a data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no "training set" in the AI/ML context, there is no ground truth established for it. The performance of the device is assessed against established engineering and safety standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093000

    Validate with FDA (Live)

    Date Cleared
    2010-07-30

    (305 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2