510(k) Data Aggregation

    K Number
    K091871

    Validate with FDA (Live)

    Device Name
    DK50 DS
    Manufacturer
    Date Cleared
    2009-10-23

    (122 days)

    Product Code
    Regulation Number
    868.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for supplying compressed air for medical ventilators.

    Device Description

    The Ekom model DK50 DS is a portable air compressor designed to supply air to medical equipment, i.e., ventilators.

    The Model DK50 DS contain an oil-free piston compressor driven by a low maintenance single phase electric motor. Compressed air is cooled where condensed water is separated into a vessel. Incoming air passes through 2 filters and undergoes double filtration as it passes through the system. Constant output pressure is maintained by a pressure regulator. There is a built-in air tank which allows peak air consumption of 200 Lpm.

    The device may be used as a standby source of air. In this configuration, the respiratory equipment is supplied with compressed air from the facilities central air distribution. Air pressure in this central distribution is sensed by the Ekom pressure sensor. If the pressure is sufficient, the compressor stays in the STANDBY mode. If the pressure falls the compressor automatically starts and becomes the main air supply.

    When the compressor is used as the main source, the control unit determines its operation according to the current need for air. If air consumption is zero, the device switches to STANDBY.

    The compressor is equipped with indicators for output pressure, operation hours, power status, drying efficiency and battery condition as well as acoustic and optic alarms to warn of high operating temperature, low output pressure and loss power.

    AI/ML Overview

    The provided document describes a medical device (portable air compressor DK50 DS) and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving that the device meets those criteria in the context of an AI/ML or diagnostic device.

    The document focuses on:

    • Device Description: A portable air compressor designed to supply air to medical ventilators.

    • Indications for Use: Supplying compressed air for medical ventilators in hospital or sub-acute institution environments.

    • Substantial Equivalence: A comparison of the DK50 DS with two predicate devices (K060781 - Ekom Model DK50 D and K041406 - Newport Model C250) based on specifications like output flow, power, alarms, and physical dimensions.

    This is a 510(k) premarket notification for a traditional medical device (a portable air compressor), not an AI/ML-driven diagnostic or image analysis tool. Therefore, the concepts of acceptance criteria for diagnostic performance (like sensitivity, specificity), ground truth, test/training sets, expert adjudication, or MRMC studies are not applicable to the information provided.

    The "acceptance criteria" discussed in the document are effectively the technical specifications and performance characteristics compared against the predicate devices to establish substantial equivalence.

    Therefore, I cannot provide the requested information because it is not present in the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060781

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-10-03

    (195 days)

    Product Code
    Regulation Number
    868.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DK50 D and DK50 DM Medical Compressors are indicated for supplying compressed air for medical ventilators.

    Device Description

    The DK50 D and DK50 DM Medical Compressors are electric and produces air from the normal environment to supply compressed air for medical ventilators.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for two medical compressors, DK50 D and DK50 DM. It details their intended use, describes them as substantially equivalent to a predicate device, and lists the standards they were tested against. However, the document does NOT contain information about acceptance criteria or a study proving device performance against such criteria in the way typically expected for an AI/ML device.

    This document is for a traditional medical device (compressors), not an AI/ML device, and therefore the concepts of "acceptance criteria" and "device performance" are addressed differently. The "study" mentioned is a general statement about adherence to industry standards, not a specific clinical or performance evaluation with statistical metrics.

    Given the information provided, here's how the questions can be addressed:

    1. A table of acceptance criteria and the reported device performance
      • Acceptance Criteria: Not explicitly stated as quantifiable metrics in the document. The general acceptance criterion is that the device "meets its specifications and is safe and effective for its intended use" and is "substantially equivalent" to the predicate device.
      • Reported Device Performance: The document states that "Comprehensive testing has been conducted on The DK50 D and DK50 DM Medical Compressors in accordance with various industry recognized standards, including: IEC 60601-1-2:2001, EN 55011:1998 &A1:1999." The outcome of this testing is that the device "meets its specifications." Specific performance metrics (e.g., flow rate, pressure, noise levels, durability) are not detailed in this summary.
    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with IEC 60601-1-2:2001, EN 55011:1998 & A1:1999Testing conducted; device meets specifications
    Safety and Effectiveness for intended use (supplying compressed air for medical ventilators)Deemed safe and effective; substantially equivalent to predicate
    1. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not applicable / Not provided. The testing mentioned refers to engineering and quality assurance testing against standards, not a clinical study with a "test set" of patients or data in the context of AI/ML.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable / Not provided. Ground truth is not established in this context. The "truth" is whether the device adheres to engineering specifications and safety standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable / Not provided. Adjudication methods like 2+1 or 3+1 are relevant for interpreting complex data, often in clinical imaging or diagnostic studies, where human consensus is needed for ground truth. This is not
        pertinent to the testing of an air compressor.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable / Not done. MRMC studies are for evaluating human performance, often with or without AI assistance, usually in diagnostic tasks. This device is an air compressor, not a diagnostic tool, and has no human-in-the-loop AI component.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable / Not done. This refers to AI algorithm performance. The device is a mechanical compressor.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not applicable / Not provided. In the context of this device, "ground truth" would be the engineering specifications and performance parameters as defined by internal design documents and relevant industry standards. The summary does not provide specific details on how conformity to these was verified beyond stating "comprehensive testing."
    7. The sample size for the training set

      • Not applicable / Not provided. This concept is for AI/ML models.
    8. How the ground truth for the training set was established

      • Not applicable / Not provided. This concept is for AI/ML models.