510(k) Data Aggregation

    K Number
    K200654

    Validate with FDA (Live)

    Date Cleared
    2020-07-22

    (132 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rubicon screening device is a small battery-powered device designed to assess and record nasal and oral airflow in adult patient during sleep in the home setting. The device is intended as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography (PSG) based on the patient's test data.

    Device Description

    The Rubicon Screening Device is comprised of 3 components and software. For the physical device. Rubicon is comprised of an Airflow sensor, cable connecting air flow sensor and a single inline module (SIM) as described below:

    Airflow sensor: The airflow sensor is placed under the user's nose where it converts temperature and pressure changes to a voltage signal. Polarized Polyvinylidene Fluoride (PVDF) is a temperature-sensing material. PVDF generates a small electrical charge in response to temperature change. The sensor is permanently connected to the SIM via two lead wires. The temperature difference between patient exhaled air and ambient air, as well as airflow pressure on the sensor generates a small voltage signal, which can be recorded as airflow by recording equipment. The vibration of breath sounds (snoring) produces a small voltage signal, which can be recorded as snoring by recording equipment. As integrated into the Rubicon, the airflow sensor provides two channels of output data from the same PVDF sensor. One channel focuses on sinusoidal airflow waveforms for identification of airflow events, and the second channel provides snore data.

    Cable: The lead wires are a typical cable pair. The lead wires were reviewed as part of the IEC 60601-1 safety evaluation. The lead wires transmit signals from the airflow sensor to the SIM and are permanently attached to the airflow sensor and the SIM.

    SIM (Housing and PCB): The SIM's primary function is to collect raw electrical signals from the airflow sensor. These signals include airflow and snore analog data. These analog signals are filtered and acquired by the SIM's A/D converter, then transmitted in real time to a nearby Smartphone, which must be running the Rubicon companion app. The SIM PCB is powered by a small coin cell battery.

    Software: The basic purpose of the software is to: Collect data from the sensor device by transmitting the data to a collection hub, which is a Smartphone + a mobile software application (app). Transmit data to a remote data center, which is accomplished via the cellular network. Store, process, and display data to a health care professional or end user via the Rubicon Analysis Software PC application and its report. The system consists of three major components: Sim Software (firmware), Rubicon App and Rubicon Desktop Software

    AI/ML Overview

    The Rubicon Screening Device is intended to assess and record nasal and oral airflow in adult patients during sleep in a home setting. It functions as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography (PSG) based on the patient's test data.

    The provided document, K200654, mainly focuses on establishing substantial equivalence to a predicate device (IMS-SleepCheck, K022294) and a reference device (Dymedix – Airflow Sensor, K040069). It describes the device's design, components, and indications for use, and outlines non-clinical testing performed (biocompatibility, electrical safety, electromagnetic compatibility, and comparative performance testing). However, it does not present a detailed study demonstrating specific acceptance criteria with reported performance metrics, nor does it provide the detailed information requested in points 2-9.

    The document states, under "Comparative testing," that "Data and scoring compared to PSG collected and analyzed data and reviewed by independent sleep specialists." This implies a comparative study was conducted, likely to establish the performance of Rubicon in identifying airflow events relevant to sleep disorders. However, the results, sample sizes, and other specifics of this study are not provided in the submitted text.

    Therefore, based on the provided text, a full answer to all your questions cannot be completed. Below is a summary of what can be extracted and what information is missing.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and specific performance metrics for the Rubicon Screening Device are not explicitly stated in the provided document. The document mentions "comparative performance testing" and that "Data and scoring compared to PSG collected and analyzed data and reviewed by independent sleep specialists," but it does not present the results of this comparison or any pre-defined acceptance thresholds.

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size used for the test set in any comparative performance study. It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document states that data was "reviewed by independent sleep specialists" to establish ground truth through comparison with PSG. However, it does not specify the number of experts, nor their detailed qualifications (e.g., years of experience, board certifications).

    4. Adjudication method for the test set

    The adjudication method used for establishing ground truth from the "independent sleep specialists" is not described in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    The document mentions "comparative performance testing" where "Data and scoring [were] compared to PSG collected and analyzed data and reviewed by independent sleep specialists." This implies a comparison against PSG, which is typically considered the gold standard for sleep studies. However, it does not specify if this was an MRMC study designed to assess reader performance with and without AI assistance, nor does it provide an effect size for human reader improvement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document describes the device's components, including software for data collection, transmission, storage, processing, and display to a healthcare professional. The statement "We are not diagnosing but reporting detected airflow events for further consideration by the clinician" suggests that the device's output (airflow events) is intended to be used by a clinician, implying a human-in-the-loop scenario. It does not explicitly state whether a standalone algorithm performance study was done without human interpretation.

    7. The type of ground truth used

    The ground truth for the comparative testing was established by comparing the device's collected and analyzed data and scoring to Polysomnography (PSG), reviewed by independent sleep specialists. PSG is generally considered the gold standard for diagnosing sleep disorders.

    8. The sample size for the training set

    The document does not provide any information about a training set or its sample size. This submission focuses on substantial equivalence and non-clinical testing rather than detailed algorithm development and validation.

    9. How the ground truth for the training set was established

    As no training set is mentioned, there is no information on how its ground truth would have been established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192564

    Validate with FDA (Live)

    Date Cleared
    2020-01-21

    (125 days)

    Product Code
    Regulation Number
    882.1320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Gold Cup EEG Electrodes are intended for non-invasive use with recording and monitoring equipment, (active and reference), of Electroencephalograph (EEG), electromyography (EMG), and Evoked Potentials (EP).

    Device Description

    The EEG electrodes are provided to the healthcare provider in either a set of 10 which are typically used for sleep studies or a set of 15 which are used for EEG monitoring. Each set of electrodes are the same length, however, sets can be ordered in lengths varying from 1.0 to 3.0 meters. Each electrode within a set has a different color to aid in placement.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study information for the Dymedix Diagnostics Disposable Gold Cup EEG Electrodes (K192564), based on the provided document:

    This document is a 510(k) clearance letter, which means the device is being cleared based on demonstrating substantial equivalence to a predicate device, rather than requiring extensive clinical trials to establish its efficacy and safety from scratch. Therefore, the "study" described focuses on bench testing and biocompatibility to show that the new device performs comparably to established devices.


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria / RequirementReported Device Performance / Evidence
    Indications for UseIdentical to predicate device: Non-invasive use with recording and monitoring equipment for EEG, EMG, and EP.The device's Indications for Use are identical to the predicate devices. The electrodes can be used with Electroencephalograph (EEG), Electromyography (EMG), and Evoked Potentials (EP) monitoring equipment.
    Technology & ConstructionSimilar to predicate device, with comparable signal collection and transmission.The technology and construction are similar to the predicate devices, utilizing an electrode of similar shape and size to collect and transmit signals through connection wires to the monitoring device. The electrode materials are different from Daehan (silver/silver chloride vs. gold plated), but the Dymedix electrodes are identical to Technomed Europe Gold Cup Electrodes in material (gold plated brass) indicating equal or better signal sensitivity.
    Environment of UseSimilar to predicate device: Healthcare settings where monitoring studies may be performed.The environments of use are similar to the predicate devices: healthcare settings where monitoring studies may be performed.
    Patient PopulationSimilar to predicate device: Adults and children.The patient population is similar to the predicate devices: adults and children.
    BiocompatibilityMeet applicable ISO 10993 standards for external communicating device with limited duration surface contact.Biocompatibility testing performed to applicable ISO 10993 standards. Patient contacting materials are reported as non-cytotoxic, non-sensitizers, non-irritants, and non-toxic.
    Electrical PerformanceMeet the ANSI/AAMI EC12:2000/(R)2015 standard.Electrical performance testing was performed to the ANSI/AAMI EC12:2000/(R)2015 standard, and the electrodes are reported to meet the standard.
    Mechanical IntegrityAdequate performance throughout device use, exceeding requirements.Performance bench testing verified the electrodes will perform adequately through the duration of use. Mechanical bond testing and surface adhesive testing demonstrate the device integrity exceeds requirements for the intended use.

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a "test set" in terms of patient data or clinical images. The data provenance is from bench testing for electrical, mechanical, and biocompatibility properties. No specific "sample sizes" (e.g., number of electrodes tested for each bench test) are provided in this summary, which is typical for 510(k) summary documents. The testing is prospective in the sense that the manufacturer conducted these tests specifically for this submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the context of this 510(k) summary. "Ground truth" in this type of submission refers to established engineering standards (like ISO 10993 and ANSI/AAMI EC12) and the technical specifications of the predicate devices. Experts involved would be those qualified to conduct and interpret the specified standard tests (e.g., test engineers, toxicologists for biocompatibility), but their specific number and qualifications are not detailed here.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically for clinical trials or image annotation, which are not described here.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done. This device is a passive EEG electrode and does not involve AI or human reader interpretation for its function.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a hardware component (electrode), not an algorithm or software.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for demonstrating substantial equivalence relies on:
        • Industry standards: ISO 10993 for biocompatibility, ANSI/AAMI EC12:2000/(R)2015 for electrical performance.
        • Predicate device specifications: The technical characteristics, indications for use, and performance of the legally marketed predicate devices (K180232 Daehan Disposable CUP Electrodes and K072016 Technomed Europe Cup Electrodes).
    7. The sample size for the training set:

      • Not applicable. There is no "training set" as this is a hardware device, not an AI/ML algorithm requiring data for training.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set.