    Intended Use

    INDICATIONS FOR USE AS A RESECTION ELECTRODE:

    The Dornier Bi Polar Electrode is designed and intended for use in endoscopic urological surgical procedures involving the resection, ablation, or removal of soft tissue and where hemostasis is required. The specific urological indications include use in the prostate, bladder and bladder neck. The procedure for which the devices can be used are: Transurethral resection in saline (TURis), Transurethral prostatectomy, transurethral resection of the prostate (TURP), for benign prostatic hyperplasia, Transurethral incision of the prostate (TUIP) or bladder neck, Transurethral resection of bladder tumors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.

    The following electrodes are intended to be used as resection electrodes:

    • BIP12CLL: BIPOLAR 12° Large Cutting Loop 24 Fr Electrode-Sterile, Single-Use
    • BIP12CLM: BIPOLAR 12° Medium Cutting Loop 24 Fr Electrode-Sterile, Single-Use
    • BIP30CLL: BIPOLAR 30° Large Cutting Loop 24 Fr Electrode-Sterile, Single-Use
    • BIP30CLM: BIPOLAR 30° Medium Cutting Loop 24 Fr Electrode-Sterile, Single-Use
    • BIPMBLA: BIPOLAR Medium Bladder Loop 24 Fr Electrode-Sterile, Single-Use
    • BIPNEEL: BIPOLAR Needle 24 Fr Electrode-Sterile, Single-Use

    INDICATIONS FOR USE AS A VAPORIZATION ELECTRODE:

    The Dornier Bi Polar Electrode is designed and intended for use in urological surgical procedures involving vaporization, ablation, coagulation, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue indication include: Use in the prostate, bladder, and bladder neck. The specific treatment indications include benign prostate hyperplasia (BPH), bladder cancer, tumors, lesions and neoplasms. The specific urological indications include Transurethal Electro vaporization (TUVP), (TVP), (TUEVP) also known as Transurethal Vapor Resection of the prostate (TUVRP), Transurethal Vaporization in Saline (TUVis). These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.

    The following electrode is intended to be used for plasma vaporization:

    • BIPDDSC: BIPOLAR Dornier Disc™ 24 Fr Electrode-Sterile, Single-Use
    Device Description

    The Dornier Bipolar electrodes are sterile packaged, bipolar electrodes designed to deliver high frequency electrosurgical energy at the treatment area of the patient. Dornier bipolar electrodes are intended for single-use only.

    The electrodes are designed to be connected through a compatible resectoscope to the electrosurgical unit for the delivery of electrosurgical energy at the treatment area of the patient. It is suitable for surgical cutting, coagulation and removal of soft tissues by controlled heating effect.

    The Dornier Bipolar Electrode shall be used in conjunction with compatible resectoscopes such as Olympus Bipolar 12° and Olympus Bipolar 30° to the compatible bipolar electrosurgical generator such as Gyrus PlasmaKinetic™ SuperPulse Generator and Olympus ESG 400.

    Electrical energy is supplied by the generator. A cable is connected from the generator to the working element of the Resectoscope. The Positive lead attaches to the bottom of the working element and the negative to the top of the working element. Electrical current flows from the generator into the Connector, down the Inner Tube and into the tip of the probe, creating a plasma field. The current returns via the saline, through Forks and Scope, into the negative lead and back to the generator. The strength of the plasma field can be adjusted by increasing or lowering the power setting on the generator. A foot switch is used to turn the electrode on and off.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243820

    Validate with FDA (Live)

    Date Cleared
    2025-01-31

    (50 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dornier Hoover Negative Pressure Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. It is designed to establish a conduit for the treatment of urinary stones or other urinary diseases during the endoscopic procedures.

    Device Description

    The Dornier Hoover Negative Pressure Access Sheath is designed to establish a conduit for the treatment of urinary stones or other urinary diseases during the endoscopic procedures. It is for the passage of endoscopes and other urological devices for Ureteroscopy and used for the treatment of stones and for the removal of foreign bodies in the urinary tract.

    The packaged product includes a Hydrophilic-coated Dilator with a locking mechanism and a Hydrophilic- coated guide sheath with joint.

    The guide sheath is fitted with a Y connector on the Y connector is bifurcated straight tube and oblique tube. One segment of the Y connector is straight and is contiguous with the sheath. The other is constructed in an oblique angle with a longitudinal pressure control vent. The oblique tube is to be connected to a negative pressure aspirator with clear tube or alternatively, connected to a specimen collector then onto a negative pressure aspirator, then to collect stones or foreign bodies during the endoscopic procedure.

    A dilator is included for the insertion of the dilator is radiopaque and is fitted with a hand buckle on the proximal end. The dilator can be locked to the proximal end of the straight tube using a locking mechanism.

    The Dornier Hoover Neqative Pressure Access Sheath is reinforced with a stainless-steel coil, constructed of a medical grade thermoplastic elastomer. All colorants used are compliant with FDA standards.

    The device is offered of inner diameter in French size 10Fr, 12Fr and length 40cm, 50cm. The device is supplied sterile and intended for single use only.

    The Dornier Hoover Neqative Pressure Access Sheath has six different models. The structural composition and materials of the different models are identical. The only difference is in dimension including inner diameter, and length.

    AI/ML Overview

    I am sorry, but the provided text does not contain specific acceptance criteria, detailed study information, or performance metrics for the Hoover Negative Pressure Access Sheath.

    The document is a 510(k) premarket notification letter and summary for a medical device. It states that non-clinical performance bench testing was conducted to demonstrate substantial equivalence to a predicate device and that the device meets "pre-determined testing and acceptance criteria." However, it does not specify what those criteria are or provide results in a way that would allow for the construction of the requested table and detailed study description.

    The text mentions general categories of testing:

    • General performance testing: Connection Strengths, Coefficients of Friction, Dimensional Analysis, Fluid Leakage.
    • Biocompatibility testing: Cytotoxicity, Irritation, Sensitization.
    • Sterilization validation: Conforming to ENISO 11135:2014.

    It also indicates that testing data and results demonstrating compliance are "included in this submission," but these specifics are not present in the provided text. Therefore, I cannot furnish the detailed information you've requested regarding acceptance criteria, reported performance, sample sizes, ground truth establishment, or any MRMC study details.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233380

    Validate with FDA (Live)

    Date Cleared
    2024-06-26

    (268 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trident Mobile Fluoroscopy System is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

    Device Description

    The Trident Mobile Fluoroscopy System is a mobile Image Intensified Fluoroscopic X-ray unit with a flat panel image receptor system. The Trident Mobile Fluoroscopy System consists of the following components: an X- ray generator and tube housing and a flat panel detector. An X-ray cabinet contains system elements such as the X- ray generator, power electronics for the imaging chain. The system offers an optional monitor for viewing the captured images. The X-Ray Generator is located in the base of the C-Arm. High voltage is carried to the X-Ray tube across a set of cables and the X-Ray tube emits the X-Rays that are directed toward the control of the operator. The X-Rays pass through the patient and are captured by the flat panel image detector (FPD). The flat panels used in the Trident Mobile Fluoroscopy System are Varex models 3030DX or 2121DXV. This Varex series have been used in similar cleared devices (K192541, K220871). The Varex flat panel system used in the Trident uses Cesium lodide as the image scintillator which is identical to that used in the reference device (K220871). These flat panel models differ only in the dimensions of the image receptor. FPD images are processed and can be displayed on the optional image monitor located on the Workstation. The Physician or system operator can view the images as they are displayed on the attached monitor or they may also choose to store the images for later review via the connection to the hospital DICOM system. The C-Arm is designed to move into the required positions to allow for proper x-ray locations in relation to the patient for the procedure that is being performed. These procedures can be performed with or without a patient table. In order to accomplish this, the system is designed with the ability to rotate/swivel to obtain different appropriate viewing angles. The systems movement is manually adjusted with the exception of the z-axis. This is motorized with the controls on the C-arm. The release of x-ray is controlled using a footswitch and/or a hand controller.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Dornier MedTech America Trident Mobile Fluoroscopy System, based on the provided FDA 510(k) summary:

    This device is an X-Ray system, and the provided document is a 510(k) summary for substantial equivalence. For such devices, "acceptance criteria" and "device performance" in the context of clinical studies (like sensitivity, specificity, etc.) are generally not directly applicable in the same way as they would be for an AI-powered diagnostic device. Instead, substantial equivalence is proven through non-clinical performance testing against recognized standards and comparative image quality assessments.

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The FDA 510(k) summary for this type of device (an X-ray system) primarily relies on demonstrating compliance with recognized performance standards and comparative image quality rather than traditional clinical performance metrics like sensitivity/specificity. Therefore, "acceptance criteria" here refers to successful completion of non-clinical tests and "reported device performance" refers to the results of those tests.

    Acceptance Criteria (Compliance Goal)Reported Device Performance (Achievement)
    Electrical Safety and EMC
    IEC 60601-1:2005, AMD1:2012 (Basic safety and essential performance)Confirmed applicable requirements met.
    IEC 60601-1-2:2014 (EMC)Confirmed applicable requirements met.
    EN 60601-1-6:2010+A1:2015, IEC 62366-2:2007, A1:2014 (Usability)Confirmed applicable requirements met.
    IEC 60601-1-3:2008, A1:2013 (Radiation protection in diagnostic X-ray)Confirmed applicable requirements met.
    IEC 60601-2-54:2009, A1 2015, A2 2019 (X-ray for radiography/radioscopy)Confirmed applicable requirements met.
    IEC 60601-2-43:2010 (X-ray for interventional procedures)Confirmed applicable requirements met.
    Software Performance
    IEC 62304-2006/A1:2015 (Software life cycle)Software documented for Enhanced Documentation Level (Moderate Level of Concern) and meets requirements.
    Image Quality (Clinical Acceptability)
    Images found clinically acceptable based on comparison to predicateComparative images using a pelvic phantom and a Primus phantom were reviewed by a certified radiologist and found to be clinically acceptable in comparison to the predicate GE device.
    Substantial EquivalenceAchieved. Performance data (non-clinical tests and image comparison) demonstrated that the Trident Mobile Fluoroscopy System is as safe and effective as the OEC 9900 Elite and Dornier Nautilus, sharing similar indications for use, intended use, technological characteristics, and principles of operation, without raising new or different questions of safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document mentions "Comparative images using the Trident Mobile Fluoroscopy System and the GE predicate device were taken to reflect usage conditions of the device. The images taken used a pelvic phantom as well as a Primus phantom." This indicates that the "test set" consisted of images generated from at least two phantoms (a pelvic phantom and a Primus phantom), but the exact number of images or imaging sequences is not specified beyond "images."
    • Data Provenance: The images were acquired from phantoms, not human patients. The country of origin of the phantoms or the location of the imaging is not specified, but it would have been conducted as part of the manufacturer's non-clinical testing. This is a prospectively generated dataset for testing the device's image quality.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: "Upon review by a certified radiologic they were found to be clinically acceptable..." This indicates that one certified radiologist reviewed the images.
    • Qualifications of Experts: The expert was described as "a certified radiologist." Specific years of experience are not mentioned.

    4. Adjudication Method for the Test Set

    • Adjudication Method: The document states that a single certified radiologist reviewed the images and found them to be clinically acceptable. This implies no formal adjudication process (like 2+1 or 3+1 consensus) as only one expert was involved in the assessment for clinical acceptability.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. The study described involves a single radiologist reviewing phantom images to assess clinical acceptability compared to a predicate device. This is not an MRMC study comparing human reader performance with and without AI assistance.

    6. If a standalone (algorithm only without human-in-the-loop performance) was done

    • Yes, in essence, standalone performance was assessed. The device itself generates images. The "image comparative testing" described focuses on the quality of the images produced by the device (Trident Mobile Fluoroscopy System) compared to a predicate device, as assessed by a radiologist. This is an assessment of the device's output and its intrinsic image quality, separate from its operation by a human user in a clinical workflow. The "algorithm" here refers to the imaging system's capabilities in producing a diagnostic image.

    7. The Type of Ground Truth Used

    • Expert Consensus/Opinion on Image Quality: The ground truth for the image quality assessment was based on the "clinical acceptability" determined by a certified radiologist, comparing the images from the subject device to those from the predicate GE device. Since these were phantom images, "pathology" or "outcomes data" are not relevant here.

    8. The Sample Size for the Training Set

    • Not Applicable: This device is an X-ray imaging system, not an AI/ML-powered diagnostic algorithm that requires a "training set" in the conventional sense. The "training set" concept is typically relevant for machine learning models that learn from data. The reference here is to standard engineering design and testing, where components and systems are tested against specifications, not trained on data.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable: As there is no "training set" for an AI/ML model for this traditional imaging device, the establishment of its ground truth is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221903

    Validate with FDA (Live)

    Device Name
    Delta III Pro
    Date Cleared
    2023-02-22

    (237 days)

    Product Code
    Regulation Number
    876.5990
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Delta III Pro Lithotripter is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

    Device Description

    The Delta III Pro Lithotripter is a modular urological work station designed for extracorporeal shock wave lithotripsy ("ESWL") and for diagnostic and therapeutic procedures usual in urology.
    The Delta III is composed of the following modules:

    • . Basic Unit with integrated X-ray C-arm and Therapy Arm with camera for Shockwave Treatment;
    • Patient Table: .
    • Control Desk/Image Storage (UIMS). .
      The basic unit contains the power supplies, control unit, power electronics for motor drives, components for shockwave generation, and an integrated Therapy C-arm and an X-Ray C-Arm. The housing can be positioned with its back close to the room wall and has wide side doors for easy service.
      The therapy and X-ray C-arm house the shock wave source ("EMSE") and the complete X- ray unit. The X-ray unit consists of the X-ray generator, the X-ray tube, a flat panel image receptor system, and a high resolution imaging chain. This provides the imaging to perform the procedures. The C-arms allow for a wide range of movement to facilitate performing urological procedures. The shock wave circuit supplies the shock wave energy needed for the treatment of kidney stones.
      The Delta III Pro's urological patient table provides longitudinal, lateral and vertical travel range to allow easy positioning of the stone in the shock wave focus for lithotripsy and urological procedures.
      The image processing system (UIMS) with DICOM 3 capability supports PACS connection and offers complete X-rav control and image handling.
    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study that proves the device meets those criteria in the context of typical AI/ML device performance studies (e.g., sensitivity, specificity, accuracy against a ground truth).

    Instead, the document is a 510(k) submission for a Delta III Pro Lithotripter, which is a medical device for fragmenting urinary tract stones. The submission focuses on demonstrating substantial equivalence to a predicate device (Dornier Delta III Lithotripter, K201074) based on minor modifications.

    The "Performance Data" section describes compliance with electrical safety, electromagnetic compatibility, radiation protection, safety of lithotripsy equipment, safety and performance of x-ray equipment, and usability engineering standards (IEC standards), as well as software verification and validation. These are engineering and regulatory compliance criteria, not performance metrics like sensitivity or specificity for a diagnostic AI.

    Here's a breakdown of why the requested information for acceptance criteria and related study details cannot be found in the provided text:

    • This is not an AI/ML device in the context of diagnostic performance: The device is a lithotripter, which physically breaks down stones. Its performance is measured by its safety, efficacy in stone fragmentation, and compliance with various engineering standards, not by diagnostic accuracy.
    • The "Performance Data" section refers to regulatory compliance and non-clinical testing: "Non-clinical functional testing successfully passed and met all design requirements" and compliance with IEC standards are mentioned. This refers to engineering verification and validation, not a clinical study comparing its diagnostic performance against a ground truth with human experts.
    • No mention of human readers, AI assistance, or diagnostic metrics: The text does not discuss human interpretation of images/data, AI algorithms for analysis, or metrics like accuracy, sensitivity, specificity, AUC, etc.

    Therefore, many of your specific questions are not applicable to the information provided in this 510(k) summary. I cannot create the table or answer the questions the way they are designed for an AI/ML diagnostic device.

    However, I can extract the information that is present regarding "performance" in the context of this device and its submission:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for diagnostic performance, but it lists standards with which the device complies. The "reported device performance" is primarily stated as successful compliance with these standards and established design requirements.

    Acceptance Criteria (Implied)Reported Device Performance
    Electrical safety (IEC 60601-1:2006/AC:2010+A1:2013)In compliance with the standard
    Electromagnetic compatibility (IEC 60601-1-2: 2015)In compliance with the standard
    Radiation protection (IEC 60601-1-3: 2008)In compliance with the standard
    Safety of equipment for extracorporeally induced lithotripsy (IEC 60601-2-36:2014)In compliance with the standard
    Basic safety and essential performance of x-ray equipment (IEC 60601-2-54:2009 + Cor.:2010 + Cor.:2011)In compliance with the standard
    Application of usability engineering (IEC 62366-1:2015 + COR1:2015)In compliance with the standard
    Software functionality (identical to predicate device)Identical functionality to predicate device, documentation for Moderate Level of Concern included (FDA Guidance)
    Non-clinical functional requirementsSuccessfully passed and met all design requirements
    Safety and effectiveness (overall)Device is as safe and effective as the cited predicate device

    2. Sample size used for the test set and the data provenance
    This information is not provided because it's a non-clinical, engineering verification for a physical device, not an AI/ML diagnostic study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. No "ground truth" in the diagnostic sense is established for this type of device.

    4. Adjudication method for the test set
    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not applicable in the diagnostic sense. The "ground truth" for this device would be established engineering specifications, safety standards, and performance benchmarks for stone fragmentation, which are not detailed in this summary.

    8. The sample size for the training set
    Not applicable. This is not an AI/ML device that requires a training set in that context.

    9. How the ground truth for the training set was established
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220871

    Validate with FDA (Live)

    Device Name
    Nautilus
    Date Cleared
    2022-04-18

    (24 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dornier Nautilus is an image intensified, fluoroscopic x-ray system that is intended for use in a wide field of applications, including all general examinations in urology and gynecology, as well as endoscopic and contrast examinations, imaging with radiography and/or fluoroscopy on patients in either the horizontal or vertical position.

    Device Description

    The Dornier Nautilus is an Image Intensified Fluoroscopic X-ray System with a flat panel image receptor system. The Nautilus consists of the following components: an X- ray generator and tube housing, flat panel detector, monitors and procedure table. An X-ray cabinet contains system elements such as the X-ray generator, power electronics for the imaging chain.

    The Dornier Nautilus is a radiographic and fluoroscopy examination table with the X-ray tube housing mounted under the table on a fixed arm. A flat panel detector is mounted above the patient table. The flat panel used is a Varex model 4343DXV. These Varex 4343 series have been used in similar cleared devices (K192541). The Varex flat panel system uses Cesium lodide as the image scintillator which is identical to that used in the predicate device. While the X-ray tube and detector are fixed in their positions relative to each other when the system is in use, the table top and X-ray/detector unit can be moved in a variety of planes to position the patient in the desired imaging position. The captured images are processed and can be stored in the users DICOM system.

    AI/ML Overview

    The provided text describes the Nautilus, an image intensified fluoroscopic x-ray system. However, it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of device performance metrics like sensitivity, specificity, or accuracy.

    The "Performance Data" section primarily addresses adherence to electrical safety and electromagnetic compatibility (EMC) standards, and software verification and validation. These are crucial for the safety and basic functionality of the device, but they are not the typical performance metrics associated with demonstrating the clinical efficacy or diagnostic accuracy of an imaging system compared to a ground truth or a human reader.

    Here's a breakdown of what is and isn't present, based on your request:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria Mentioned: The text implicitly states acceptance criteria by listing the standards that the Nautilus was tested against (e.g., IEC 60601-1, IEC 60601-1-2). The "Performance Data" states that "Performance testing confirmed that the Nautilus met the requirements of the following standards."
    • Reported Device Performance: The document only reports conformance to these safety and software standards, not specific performance metrics in terms of image quality, diagnostic accuracy, or clinical outcomes. There are no numerical results (e.g., contrast-to-noise ratio, spatial resolution, diagnostic sensitivity/specificity) provided.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No clinical or diagnostic performance test set is described. The testing mentioned is for electrical safety, EMC, and software, which typically involves engineering tests, not patient data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No ground truth establishment for diagnostic performance is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No diagnostic performance test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states: "Clinical testing is not necessary for the subject system Nautilus, based on the same basic technology as the predicate device and based on existing minor differences." This indicates that an MRMC study or any clinical effectiveness study was not performed or deemed necessary for this 510(k) submission. There is no AI assistance mentioned, so no effect size for human readers with AI vs. without AI assistance can be provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This device is an imaging system (fluoroscopic x-ray system) and not an AI algorithm. Its performance is evaluated fundamentally in conjunction with a human operator / clinician interpreting the images. No standalone algorithm performance is applicable or discussed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. No diagnostic ground truth is mentioned. The "ground truth" for the non-clinical tests would be the requirements defined by the referenced international standards for electrical safety, EMC, and software quality.

    8. The sample size for the training set

    • Not Applicable. As no AI algorithm or diagnostic performance study is described, there's no training set for such a purpose.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.

    Summary of Device Acceptance and Study as Described in the Document:

    The Nautilus device's acceptance is based on its substantial equivalence to a predicate device (Dornier Genesis K151485) and its demonstrated conformance to established international standards for:

    • Electrical Safety: IEC 60601-1, EN 60601-1-6, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54
    • Electromagnetic Compatibility (EMC): IEC 60601-1-2
    • Usability: EN 60601-1-6, IEC 62366-2
    • Software Verification and Validation: Adherence to FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Moderate Level of Concern).

    The study that "proves the device meets the acceptance criteria" is non-clinical testing (bench testing) against these referenced standards. No clinical studies, human reader studies, or diagnostic performance studies with explicit acceptance criteria (e.g., sensitivity/specificity thresholds) and corresponding test results are provided in this submission document. The rationale provided for not conducting clinical testing is that the device uses "the same basic technology as the predicate device and based on existing minor differences."

    Ask a Question

    Ask a specific question about this device

    K Number
    K210394

    Validate with FDA (Live)

    Date Cleared
    2021-05-19

    (98 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dornier Bipolar Cable is indicated to be used with a compatible electrosurgical generator and an HF electrode for endoscopic treatment in urological applications: Cutting, ablation, resection, vaporization and coagulation with HF current.

    Device Description

    The Dornier Bipolar Cable is a sterile, single use, disposable device that allows connection of an electrosurgical generator to an HF electrosurgical resection and vaporization electrode. The outer body of the Dornier Bipolar Cables are constructed of a medical grade thermoplastic elastomer. The cable consists of a double wire zip cord with one plug end which has four (4) banana and one (1) straight pin, style of connector that is then attached to the generator. The other end is separated and has two (2) overmolded connectors, one male pin on a wire and one female pin on a wire which are compatible with the HF Electrode.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Dornier Bipolar Cable. This is a submission for a medical device that does not involve Artificial Intelligence (AI). The submission focuses on demonstrating substantial equivalence to a predicate device through conventional engineering and safety testing.

    Therefore, many of the requested categories related to AI performance, such as sample sizes for test and training sets, expert consensus, adjudication methods, and MRMC studies, are not applicable to this document.

    Here's the information that can be extracted or deduced from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryReported Device Performance
    SterilityFound to be acceptable
    PackagingFound to be acceptable
    BiocompatibilityFound to be acceptable
    Electrical Safety Testing (IEC 60601-1-2:2014)Limited to Radiated Emissions (EN 55011), Radiated and Conducted Immunity (EN 61000-4-3 and EN 61000-4-6). Found to be acceptable.
    Electrical Safety Testing (ANSI/AAMI/IEC 60601-2-2: 2009)ACTIVE ACCESSORY HF leakage, ACTIVE ACCESSORY HF dielectric strength, ACTIVE ACCESSORY mains frequency dielectric strength. Found to be acceptable.
    Dimensional TestsFound to be acceptable
    Connector grip testFound to be acceptable
    Continuity testFound to be acceptable
    Cable insulation HF and Mains Dielectric Strength testFound to be acceptable
    Cable tensile strength testsFound to be acceptable
    Activation recognition and operationFound to be acceptable
    Insulation ResistanceFound to be acceptable
    Endurance TestFound to be acceptable
    Connector pull forceFound to be acceptable
    Surface temperature measurement testFound to be acceptable

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as the device is a physical electrosurgical cable and its evaluation did not involve a "test set" in the context of an AI algorithm or data analysis study. The testing performed was engineering and safety validation. The provenance of test data for physical components is not usually detailed in this manner for 510(k) submissions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. "Ground truth" in this context typically refers to the confirmed diagnostic or clinical reality for evaluating an AI model. For a physical medical device, the "ground truth" is established through adherence to engineering standards and validated test methods by qualified engineers and technicians, not through expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as there was no test set requiring expert adjudication in the context of an AI system.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the device is a physical electrosurgical cable and does not involve AI assistance for human readers or direct comparative effectiveness in that manner.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as there is no algorithm or AI component in this device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is based on meeting established engineering standards and validated test methods for its physical and electrical properties. This is typically determined by physical measurements, electrical tests, and adherence to regulatory standards (e.g., IEC 60601 series).

    8. The sample size for the training set

    This information is not applicable as there is no AI component or training set for this device.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no AI component or training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201815

    Validate with FDA (Live)

    Date Cleared
    2020-11-20

    (142 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dornier MAGELLAN Ureteral Access Sheath is indicated for use in endoscopic urology procedures where ureteral dilation and ureteral access is desired for injection of fluids and insertion and removal of endoscopes and related instruments. The target population is for adults only (at least 22 years old).

    Device Description

    The Dornier MAGELLAN Ureteral Access Sheath is a two component ureteral dilatation system that provides an open conduit to the upper urinary tract to facilitate Ureteroscopy, which contains a single lumen for injection of fluids as well as passage of endoscopes and related instruments. The packaged product includes a Hydrophilic-coated Dilator with a locking mechanism and a Hydrophilic-coated sheath with hub. The Access Sheath is a sterile, single use, disposable device that allows access to the ureter to facilitate scope and urological tool passage. The Dornier MAGELLAN Ureteral Access Sheaths are constructed of a medical grade thermoplastic elastomer (Pebax®). This material has been USP Class VI tested. All colorants used are compliant with FDA standards.

    AI/ML Overview

    This FDA 510(k) summary describes the Dornier MAGELLAN Ureteral Access Sheath, a medical device, and its substantial equivalence to a predicate device. As such, it does not detail acceptance criteria and a study that proves the device meets specific performance metrics in the way that an AI/ML device submission would. Instead, the submission focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device through a comparison of characteristics and performance tests.

    Here's a breakdown of the requested information based on the provided text, and where certain details are not applicable (N/A) because this is a traditional medical device submission, not an AI/ML B.S.E submission:


    Acceptance Criteria and Device Performance

    The submission states that the Dornier MAGELLAN Ureteral Access Sheath underwent various performance tests, and "All testing was found to be acceptable and substantially equivalent to those of the predicate device." While specific numerical acceptance criteria are not provided in this summary, the implicit acceptance criterion for each test is that the device's performance must be comparable to, or meet the established benchmarks for, the predicate device.

    Table of Acceptance Criteria and Reported Device Performance

    Test PerformedAcceptance Criteria (Implicit)Reported Device Performance
    SterilityPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    PackagingPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    BiocompatibilityPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    RadiopacityPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Sheath IDPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Dilator ODPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Sheath Distal Tip IDPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Dilator Taper lengthPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Kink ResistancePerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Assembly Flexibility-3 point bend testPerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Sheath/Hub TensilePerformance comparable to predicate device / established standards"All testing was found to be acceptable"
    Dilator /Hub TensilePerformance comparable to predicate device / established standards"All testing was found to be acceptable"

    Study Details

    1. Sample size used for the test set and the data provenance:
      This information is N/A for this type of submission. The performance data section lists the types of tests conducted (e.g., kink resistance, tensile strength), which are typically bench tests or material tests, not clinical studies with patient data. Therefore, there isn't a "test set" in the context of patient data, nor is there data provenance in terms of country of origin or retrospective/prospective in the way an AI/ML device would report it.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This information is N/A. The "ground truth" for these types of mechanical and material tests is defined by engineering specifications and established test methodologies, not by expert medical interpretation.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      This information is N/A. Adjudication methods are relevant for clinical studies where multiple experts evaluate cases that might have ambiguous findings, which is not applicable to the performance tests listed.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      This information is N/A. This device is a physical ureteral access sheath, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and the concept of "human readers improving with AI" are not relevant.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      This information is N/A. This is a physical medical device, not an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For the performance tests described (e.g., sterility, biocompatibility, kink resistance, tensile strength), the "ground truth" is typically based on engineering specifications, material standards, and validated test methods. It's not based on expert medical consensus, pathology, or outcomes data. For example, the ground truth for "kink resistance" would be defined by a specific force or angle at which the sheath begins to kink, as measured by a standardized physical test.

    7. The sample size for the training set:
      This information is N/A. This device does not involve a training set as it is not an AI/ML model.

    8. How the ground truth for the training set was established:
      This information is N/A. This device does not involve a training set or ground truth establishment in the context of AI/ML.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201074

    Validate with FDA (Live)

    Date Cleared
    2020-09-03

    (134 days)

    Product Code
    Regulation Number
    876.5990
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Delta III Lithotripter is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

    Device Description

    The Delta III Lithotripter is a modular urological work station designed for extracorporeal shock wave lithotripsy ("ESWL") and for diagnostic and therapeutic procedures usual in urology. The Delta III is composed of the following modules: Basic Unit with integrated X-ray C-arm and Therapy Arm with camera for Shockwave Treatment; Patient Table; Control Desk/Image Storage (UIMS).

    AI/ML Overview

    This document is a 510(k) summary for the Delta III Lithotripter, detailing minor changes to an existing device. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with acceptance criteria and a human-in-the-loop performance evaluation. The information provided primarily consists of engineering and software validation.

    Here's an analysis based on the provided text, addressing your questions to the extent possible:

    1. Table of acceptance criteria and the reported device performance

    The document does not present a formal table of acceptance criteria for clinical performance in the way one might expect for a new diagnostic or AI device. Instead, the "acceptance criteria" are implied by compliance with standards and functional validation tests, with the reported performance being "does not impact the performance" and "equal to or better than the predicate."

    Acceptance Criteria (Implied)Reported Device Performance
    Standards Compliance:
    IEC 60601-1 (Electrical safety)Compliant
    IEC 60601-1-2 (EMC)Compliant
    IEC 60601-1-3 (Radiation protection)Compliant
    IEC 60601-2-36 (Lithotripsy safety)Compliant
    IEC 60601-2-28 (X-ray tube assemblies)Compliant
    IEC 60601-2-54 (X-ray equipment safety)Compliant
    IEC 62366-1 (Usability engineering)Compliant
    IEC/TR 62366-2 (Usability guidance)Compliant
    Bench Testing (Functional Validation):
    New ultrasound device provides visual and accurate images for stone visualizationPerformance assured; provides visual and accurate images
    New camera image quality is equal to or better than the predicateImage quality equal to or better than the predicate
    UIMS software with AGFA package does not impact equipment functioning and processes X-ray imagesDoes not impact functioning; provides ability to further process X-ray images
    Electrical safety, EMC, functional usability are fully addressedDemonstrated compliance with standards and equivalent performance to predicate
    No unanticipated new risks identifiedNo unanticipated new risks identified

    2. Sample size used for the test set and the data provenance

    The document describes bench testing and standards compliance, not a clinical test set with patient data. Therefore, there is no mention of a "sample size for the test set" in the context of patient data, nor its provenance (country of origin, retrospective/prospective). The tests mentioned are engineering and software validation tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information is not applicable to the submission described. The described tests are about technical performance, safety, and functionality, not about expert interpretation of medical images or patient outcomes. Ground truth in this context would refer to engineering specifications or validated functional requirements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the submission describes engineering and software validation, not a clinical study requiring adjudication of diagnoses or findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or is mentioned. The submission is for minor changes to an existing lithotripter, focusing on hardware component replacements and software updates, not AI integration for improved human reading.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device (Delta III Lithotripter) is a medical device for shock wave lithotripsy. While it includes image processing software (UIMS with AGFA MUSICA), the evaluation described is for the functional performance of these components within the overall system, not a standalone algorithm being evaluated for diagnostic accuracy without human involvement. The AGFA MUSICA software enhances X-Ray image processing, but it's part of the lithotripter system for visual guidance during procedures, not a diagnostic AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the described tests, the "ground truth" would be established by:

    • Engineering specifications and regulatory standards: For electrical safety, EMC, radiation protection, and usability.
    • Predicate device performance: For the ultrasound image quality and camera image quality, the ground truth for comparison is the performance of the components being replaced.
    • Functional requirements: For the UIMS software validation, the ground truth is that the software should perform its intended function (e.g., image processing, PACS connection) without negatively impacting other equipment functions.

    There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for these specific validation tests.

    8. The sample size for the training set

    Not applicable. This submission does not involve a machine learning or AI model that requires a training set in the conventional sense. The UIMS software update involves incorporating the AGFA MUSICA package, which is likely a pre-existing, validated image processing software, not a custom-trained model for this specific application.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190612

    Validate with FDA (Live)

    Date Cleared
    2019-12-06

    (270 days)

    Product Code
    Regulation Number
    876.5470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dornier EQUINOX Balloon Dilatation Catheter is indicated for dilation of the urinary tract.

    Device Description

    The Dornier EQUINOX Balloon Dilatation Catheter is a multiple lumen catheter with a dilatation balloon mounted at the distal tip. Dilatation balloon catheters are used to exert radial force to dilate narrow ureteral segments.

    The Dornier EQUINOX Balloon Dilatation Catheters are sterile, single-use devices. The EQUINOX Balloon Dilatation Catheters are available in 5.8 French (Fr) diameter, with a catheter length of 80 centimeters (cm). There will be three (3) balloon diameters, (4mm, 5mm and 6mm) and two (2) balloon lengths (4cm and 10cm)

    The Dornier EQUINOX Balloon Dilatation Catheters are constructed of a medical grade nylon material. This material has been USP Class VI tested. The base material is compounded with BaSO4 to render the Dornier EQUINOX Balloon Dilatation Catheter radiopaque under x-ray fluoroscopy. The catheter also has radio-opaque marker bands on the proximal and distal ends of the balloon for visualization during use. All colorants used are compliant with FDA standards.

    AI/ML Overview

    This FDA 510(k) summary describes a medical device, the Dornier EQUINOX Balloon Dilatation Catheter, and its substantial equivalence to a predicate device, rather than providing details of a study on AI or diagnostic performance. Therefore, most of the requested information about acceptance criteria, study details, expert involvement, and ground truth is not applicable.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists performance tests but does not provide specific quantitative acceptance criteria or detailed numerical results. Instead, it states that "All testing was found to be acceptable and substantially equivalent to those of the predicate device."

    Test CategoryAcceptance Criteria (Not explicitly stated with numerical values in the document)Reported Device Performance
    Sterility(Implied: Device must be sterile)Found to be acceptable
    Packaging(Implied: Packaging integrity maintained)Found to be acceptable
    Biocompatibility(Implied: Material is biocompatible)Found to be acceptable (USP Class VI tested material)
    Balloon Burst Durability Test(Implied: Balloon withstands pressure, does not burst prematurely)Found to be acceptable
    Balloon Deflation Time(Implied: Balloon deflates within an acceptable time)Found to be acceptable
    Balloon/Scope Compatibility(Implied: Compatible with endoscopic scopes)Found to be acceptable
    Radiopacity(Implied: Visible under x-ray fluoroscopy)Found to be acceptable (compounded with BaSO4, marker bands)
    Effective Working Length(Implied: Meets specified length)Found to be acceptable
    Catheter Tip Length(Implied: Meets specified length)Found to be acceptable
    Balloon Diameter @ RBP (Rated Burst Pressure)(Implied: Achieves specified diameter at rated pressure)Found to be acceptable
    Balloon Length @ OP (Operating Pressure)(Implied: Achieves specified length at operating pressure)Found to be acceptable
    Catheter Shaft OD (Outer Diameter)(Implied: Meets specified diameter)Found to be acceptable

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided in a 510(k) summary. The document describes bench testing, not clinical studies with patients or data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. This document is about a mechanical device's performance characteristics, not a diagnostic or AI device that requires expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods are typically used in clinical trials or studies where human expert consensus is needed to determine ground truth for diagnostic accuracy, which is not the subject of this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This document is for a medical device (balloon catheter), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This document is for a medical device (balloon catheter), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Not applicable. The "ground truth" for this device's performance would be engineering specifications and established test methods (e.g., measuring dimensions, pressure resistance, sterility, biocompatibility standards).

    8. The sample size for the training set:

    Not applicable. This document does not describe an AI/machine learning device that uses a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This document does not describe an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190312

    Validate with FDA (Live)

    Date Cleared
    2019-11-15

    (276 days)

    Product Code
    Regulation Number
    876.4620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dornier CASCADE Ureteral Stent is indicated to relieve obstruction in a variety of benign, malignant and post-traumatic conditions in the ureter such as presence of stones and/or stone fragments, or other ureteral obstructions such as those associated with ureteral stricture, carcinoma of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL). The stent may be placed using endoscopic surgical techniques or percutaneously using standard radiographic technique.

    Device Description

    The Dornier CASCADE Ureteral Stents are sterile, single-use devices. The stents are available in 4.9 to 8.0 French (Fr) diameter, with lengths ranging from 14.0 to 32.0 centimeters (cm). The stents are constructed of a thermoplastic polyurethane elastomer. The base polyurethane material is compounded with BaSO4 to render the stent radiopaque under x-ray fluoroscopy.

    Drainage holes extend along the body of the stent as well as on the pigtails to provide drainage. The stents are secured in the urinary tract with pigtail loops on the proximal and distal ends. A monofilament tether for repositioning and removal of the device is located on the proximal pigtail (bladder end) of the stent. Along the stent are graduation marks to provide visualization during stent advancement and placement. The CASCADE Ureteral Stent includes a pigtail straightener, a small polymer tube to aid in placement over a wire guide. The pigtail straightener is removed prior to use. The CASCADE Ureteral Stents are labeled for a maximum 30day indwell time.

    The stents are packaged into a pre-formed tray and sealed within a Tyvek pouch. The individually sealed Tyvek pouches are labeled with identification labeling and then packaged into a corrugated shipping box. The packaged stents are sterilized by a contract sterilizer using ethylene oxide.

    AI/ML Overview

    The provided text is a 510(k) summary for the Dornier CASCADE Ureteral Stent. It describes the device, its indications for use, and the studies conducted to demonstrate its substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria or detailed results of a study designed to prove the device meets those criteria in a format applicable to evaluating AI/ML models.

    The document states: "All testing was found to be acceptable and bench performance testing comparable to the predicate device." This is a general statement of compliance, not a detailed report of acceptance criteria and performance data for a particular device function.

    Therefore, many of the requested details about acceptance criteria, specific device performance, sample sizes for test and training sets, expert qualifications, adjudication methods, or MRMC studies are not present in this regulatory document.

    However, I can extract the information that is available from the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of specific numerical acceptance criteria or reported device performance values. It generally states that "All testing was found to be acceptable and bench performance testing comparable to the predicate device." The types of tests performed are listed, implying that the device met the requirements of these tests.

    Summary of available information for Section 1:

    Acceptance Criteria (General)Reported Device Performance (General)
    Sterilization validation requirementsFound to be acceptable
    Shelf life, including packaging validation and functional performance testing requirementsFound to be acceptable
    Biocompatibility requirements (cytotoxicity, skin sensitization, skin irritation, acute systemic toxicity, material-mediated pyrogenicity, genotoxicity, implantation studies, chemical characterization)Found to be acceptable for stent and pusher
    Bench Performance Testing requirements (visual inspection, dimensional analysis, pigtail retention, suture retention, coefficient of friction, tensile strength, elongation, radiopacity, and flow rate)Found to be acceptable and comparable to the predicate device

    2. Sample size used for the test set and the data provenance

    Not applicable/Not provided in this document. This document pertains to a physical medical device (ureteral stent), not a software or AI/ML device that operates on a "test set" of data in the common sense of the term for AI/ML. The "testing" refers to physical and biological evaluations of the stent itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable/Not provided in this document. Ground truth as typically defined for AI/ML models (e.g., expert consensus on medical images) is not relevant for the types of physical and biological tests described for this ureteral stent.

    4. Adjudication method for the test set

    Not applicable/Not provided in this document. Adjudication methods are typically for evaluating discrepancies in human or AI interpretations of data, which is not what this document describes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document does not mention any MRMC study or AI assistance, as it concerns a physical medical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This document does not describe an algorithm or AI performance.

    7. The type of ground truth used

    Not applicable/Not provided in this document in the context of expert consensus or pathology for data interpretation. The "ground truth" for the physical device would be the validated standards and specifications against which the stent's physical properties and biological interactions were measured (e.g., ISO standards, ASTM standards for material properties, specified biocompatibility limits).

    8. The sample size for the training set

    Not applicable/Not provided. This concept is for AI/ML models, not for physical device testing.

    9. How the ground truth for the training set was established

    Not applicable/Not provided. This concept is for AI/ML models, not for physical device testing.


    In summary, the provided document is a 510(k) summary for a physical medical device, not an AI/ML device. Therefore, most of the requested information regarding AI/ML model evaluation, such as sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, and MRMC studies, is not present or applicable. The document confirms that the Dornier CASCADE Ureteral Stent underwent standard sterilization, shelf-life, biocompatibility, and bench performance testing, and was found to be acceptable and comparable to its predicate device.