Search Filters

Search Results

Found 31 results

510(k) Data Aggregation

    K Number
    K241065

    Validate with FDA (Live)

    Device Name
    ChecQ (AC100)
    Manufacturer
    Date Cleared
    2025-03-21

    (337 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ChecQ is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.

    Device Description

    The ChecQ is a dental implant stability analyzer comprising a main unit and a charging cradle that uses Resonance Frequency Analysis (RFA) to measure implant stability non-invasively. It works with the ChecQPEG, which attaches to the dental implant, and measures stability by analyzing resonant frequencies generated by magnetic field stimulation from the probe tip. The stability result is displayed on the main unit's screen as an Implant Stability Quotient (ISQ), ranging from 1 to 99. The RFA method evaluates by applying transverse force to the implant using magnetism, measuring movement, and analyzing the resulting resonance frequency. The resonance frequency depends on the bone-implant gap and is calculated using a formula, with values expressed as ISQ scores (1-99) for clinical interpretation. The ChecQ mechanism converts voltage pulses generated by the DAC of the MCU into magnetic pulses, which induce vibrations in the ChecQPEG magnets. The device captures the free vibrations as magnetic pulses, processes them through an FFT to extract the natural frequency, and calculates the ISQ value based on this frequency.

    AI/ML Overview

    The provided text relates to the FDA 510(k) clearance for the ChecQ (AC100) device, a dental implant stability analyzer. However, the document outlines the equivalence to predicate devices based on non-clinical testing rather than providing detailed acceptance criteria and a study demonstrating the device meets those criteria through a multi-reader multi-case (MRMC) study or a standalone algorithm performance study.

    The document emphasizes "technological comparison" and "non-clinical testing" for substantial equivalence. It does not describe a clinical study in the format typically used to assess the performance of an AI/ML medical device, which would involve a test set, ground truth experts, adjudication, and performance metrics like sensitivity, specificity, or AUC.

    Therefore, many of the requested details cannot be extracted from the provided text.

    Here is what can be inferred or explicitly stated based on the given information:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria in terms of clinical performance metrics (e.g., sensitivity, specificity, accuracy) for the device. Instead, it focuses on demonstrating equivalence to a predicate device through non-clinical testing.

    The reported device performance is described qualitatively:

    • "The results confirmed that ChecQ exhibited equivalent performance, establishing its functional equivalence with the predicate device."
    • "Additional testing was conducted to evaluate the potential effects of the barrier sleeve on ISQ measurements, confirming that the presence of the barrier sleeve did not impact measurement accuracy."

    2. Sample size used for the test set and the data provenance

    The document mentions "multiple measurements using standardized test jigs" for comparative testing against the predicate device (Osstell Beacon). This indicates an in-vitro, non-clinical test using artificial setups rather than a test set derived from patient data. Therefore, there's no information on:

    • Sample size: Not applicable in the context of patient data; the "test set" consists of measurements on test jigs.
    • Data provenance: Not applicable as it's not human patient data. The testing appears to be conducted by the manufacturer, presumably in South Korea given the applicant's origin.
    • Retrospective or Prospective: Not applicable as it's not a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for the non-clinical comparison was established by the measurements from the predicate device (Osstell Beacon) and potentially engineering specifications for dimensional tests. There were no human experts involved in establishing ground truth for a clinical test set.

    4. Adjudication method for the test set

    Not applicable, as there was no clinical test set requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document explicitly states that the "ChecQ dental implant stability analyzer underwent comparative testing against the predicate device, Osstell Beacon, to assess its accuracy. The self-test methodology involved multiple measurements using standardized test jigs..." This is a non-clinical comparison of the device's output to a known standard (the predicate device) or engineering specifications, not an MRMC study with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in a sense, a "standalone" performance comparison was done, but it was comparing the device's measurement (ISQ score) against the measurement of another device (Osstell Beacon) on artificial test jigs, not against a clinical ground truth for diagnostic accuracy. The device (ChecQ) measures "Implant Stability Quotient (ISQ), ranging from 1 to 99," directly.

    7. The type of ground truth used

    The "ground truth" for the non-clinical comparison was the output of the predicate device (Osstell Beacon) under the same test conditions, and engineering specifications for mechanical properties and dimensions. It was not expert consensus, pathology, or outcomes data from patients.

    8. The sample size for the training set

    Not applicable. This device is an electro-mechanical measurement device, not an AI/ML algorithm that requires a "training set" in the context of machine learning. The device determines ISQ values based on resonance frequency analysis and established formulas, not through learned patterns from a training dataset.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233921

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-06

    (268 days)

    Product Code
    Regulation Number
    872.6250
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Luvis Chair is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.

    Device Description

    The Luvis Chair (LC700C) includes a chair used by dentists to provide seating for patients during dental procedures. It is designed to position the patient during treatment. The chair consists of a seat, backrest, and armrests that can be adjusted to accommodate the patient's position. Additionally, it may feature a headrest to offer additional support. Dental unit chairs come equipped with various tools and attachments, including dental lights, suction devices, and air- water syringes. The chair's controls are usually located within reach of the dentist, allowing for quick adjustments during the procedure. The Luvis Chair is intended to supply power to and serve as a base for dental devices and accessories.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a dental chair (Luvis Chair LC700C). However, it does not contain acceptance criteria or a study proving the device meets acceptance criteria in the manner requested.

    Instead, it asserts substantial equivalence to a predicate device based on similar intended use, principles of operation, and technological characteristics. It lists various standards the device complies with for safety, performance, and biocompatibility.

    Therefore, I cannot populate the table and answer the questions as requested because the specific information about acceptance criteria and detailed study results (like sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies) are not present in this 510(k) summary.

    The document indicates that "the conducted tests, as well as all verification and validation activities, demonstrate that the design specifications and technological characteristics of the Luvis Chair (LC700C) meet applicable requirements and standards." It also states, "The subject device had been subjected to the applicable safety and performance testing before release to ensure the device meets all its specifications." However, it does not provide the specific acceptance criteria (e.g., "device must perform with X% accuracy for Y function") or the detailed results of these tests for comparison or to be presented in a table format.

    Summary of what is present:

    • Device Name: Luvis Chair (LC700C)
    • Intended Use: To supply power to and serve as a base for dental devices and accessories. It includes a dental chair for use in a dental clinic environment by trained dental professionals.
    • Predicate Device: Osstem Implant Co., Ltd. K3 (K183347)
    • Reference Device: MeqaGen Implant Co., Ltd. N2 (K211556)
    • Testing Conducted (Non-Clinical Test Summary):
      • Thermal, electrical, mechanical safety & Electromagnetic Compatibility (compliance with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 80601-2-60)
      • Software Validations (basic documentation level, moderate level of concern, designed and developed according to a software development process, verified, and validated)
      • Biocompatibility (in vitro cytotoxicity test, oral mucosa irritation test, skin sensitization test; in accordance with ISO 10993-1 and ISO 7405)
      • Performance Test (in accordance with IEC 80601-2-60, ISO 7494-1, ISO 7494-2)
      • Cleaning and Sterilization Validation (in accordance with FDA guidance for reprocessing, ISO 17665-1, ISO TS 17665-2, ISO 16954, ISO 19458)
    • Conclusion: The device meets applicable requirements and standards, and the differences from predicate devices do not raise new questions of safety and effectiveness.

    Missing Information (and why it's missing from this document):

    The request for a detailed table of acceptance criteria vs. reported performance, sample sizes for test/training sets, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types are typical for AI/ML-enabled devices or those requiring specific performance metrics (e.g., diagnostic accuracy). This document is for a dental chair, which is a Class I medical device. For such devices, "acceptance criteria" are usually defined by compliance with recognized electrical safety, mechanical safety, biocompatibility, and performance standards (like the ISO and IEC standards listed). The "study" proving it meets these is often a series of engineering tests and validations against those standards, rather than clinical trials or AI performance evaluations with detailed ground truth analysis as would be done for a diagnostic or image analysis device.

    Therefore, the requested information (points 1-9) in the format of specific performance metrics is not applicable or not provided in this regulatory submission for a dental chair.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241063

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-07-25

    (98 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K240511

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-20

    (88 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dentis s-Clean SQ-SL Fixture is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

    Device Description

    Dentis s-Clean SQ-SL Fixture is two type as Mini and Regular according to the connection. Dentis s-Clean SQ-SL Fixture is a thread type implant made of Pure titanium according to ASTM F67 which will be placed in the alveolar bone to replace the function of the missing tooth. This device has connection between the upper prosthesis and the internal Hex.

    The surface of fixture is treated with SLA (Sandblasted with Large-grit and Acid-etching).

    The dimensions of fixtures are as following:

    No.Device NameDimension Ranges
    1s-Clean SQ-SL FixtureØ4.35, 4.8 and 5.2 (D) X 15.5 and 17.5mm
    2s-Clean SQ-SL Fixture MiniØ3.7 and 4.1 (D) X 15.5 and 17.5

    Tolerance of dimension shall be within ± 1% range.

    s-Clean SQ-SL Fixture is compatible with devices in K192688 and K210134 as below:

    510(K)Abutment NameDiameter(Ø)AngulationLength(mm)
    K192688s-Clean Cover ScrewØ3.6mm05.9mm
    K192688s-Clen (TiN Half Coating)Sole Abutment S-LineØ4.5, 5.5, 6.5 and7.5011.6, 12.6 and 13.6
    K210134s-Celan (TiN Half Coating)Angled AbutmentØ4.51512.6

    s-Clean SO-SL Fixture Mini is compatible with devices in K210080 as below:

    510(K)Abutment NameDiameter(Ø)AngulationLength(mm)
    K210080s-Clean Cover Screw MiniØ3.2mm05.0mm
    K2100080s-Clean Healing Abutment S-Line MiniØ4.3, 4.8 and 5.807.61, 8.61, 9.61, 10.61, 11.61,12.61 and 14.61
    K210080s-Celan (TiN Half Coating)Sole Abutment MiniØ4.5 and 5.5011.01, 12.01, 12.51, 13.01,13.521, 14.01, 14.51, 15.01,15.51, 16.01 and 17.01
    K210080s-Celan (TiN Half Coating)Couple Abutment MiniØ4.0, 4.5 and 5.508.35, 8.6, 9.35, 9.6, 9.85, 10.1,10.35, 10.6, 11.1, 11.35, 11.6,12.1, 12.85, 13.1, 13.35, 13.6,14.35 and 14.6
    K210080s-Celan (TiN Half Coating)Angled Abutment MiniØ4.0 and 4.51512.09, 12.34, 12.51 and 12.76

    Denis s-Clean SQ-SL Fixture is provided sterilized.

    Dentis s-Clean SQ-SL Fixture is packaged with Cover screw that was cleared in FDA as K192688 and K210080.

    AI/ML Overview

    This document is a 510(k) Summary for the Dentis s-Clean SQ-SL Fixture, a dental implant. It does not contain information about the acceptance criteria or a study proving the device meets those criteria from an AI/algorithm performance perspective. The document focuses on establishing substantial equivalence to predicate devices through non-clinical testing.

    Therefore, I cannot provide the requested information regarding AI/algorithm acceptance criteria, test set details, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details because this information is not present in the provided text.

    The "Non-Clinical Test Data" section describes engineering and biocompatibility tests conducted on the device, not an AI algorithm.

    Non-Clinical Test Data Summary (from the document):

    • Tests performed on the subject device: Fatigue Testing (ISO 14801:2016).
    • Tests performed on predicate devices and leveraged for the subject device:
      • Sterilization Validation Test on Fixtures (ISO 11137-1,2,3 referenced in K192688).
      • Shelf-Life Test on Fixtures (ASTM F1980 referenced in K153639).
      • Biocompatibility testing on fixtures (ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K153639).
      • Bacterial Endotoxin Test Report on Fixtures (ANSI/AAMI ST72:2011, USP <161>, and USP <85> referenced in K192688).
    • Acceptance Criteria for these non-clinical tests: "met the criteria of the standards" (e.g., ISO 14801:2016 for fatigue).
    • Study Design/Purpose: To demonstrate substantial equivalence to predicate devices.
    • MR Environment Condition: A non-clinical worst-case MRI review was performed using scientific rationale and published literature to evaluate magnetic safety, addressing magnetically induced displacement force and torque.

    This document describes a traditional medical device (dental implant) clearance process, not one involving Artificial Intelligence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233544

    Validate with FDA (Live)

    Device Name
    MESHEET
    Manufacturer
    Date Cleared
    2024-02-22

    (111 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MESHEET is indicated for the alignment of teeth during orthodontic treatment of malocclusion.

    Device Description

    MESEEHT is a thermoplastic aligner designed according to the patient to help maintain teeth to the final desired position. This MESHEET is to be removed for eating and for cleaning.

    MESHEET is sequence of transparent aligner manufacture from digital scan/plan.

    Each aligner delivers a unique combination of minor forces to create the planned tooth movement. MEHSEET has a main layer that material is poly ethylene and mesh layer that material is thermoplastic polyurethane. Among the main layer, mesh layer is added.

    MESEEHT is sequence of transparent aligners created from a digital orthodontic treatment plan and thermoplastic aligners designed according to the patient to help maintain teeth to the final desired position. This MESHEET is to be removed for eating and for cleaning.
    MESHEET is sequence of transparent aligner manufacture from digital scan/plan. The digital orthodontic treatment plan is created by a dental health professional. Each aligner delivers a unique combination of minor forces to create the planned tooth movement. MESHEET has cross shape holes MEHSEET has main layer that material is poly ethylene and mesh layer that material is thermoplastic polyurethane. Among the main layer. mesh layer is added.

    AI/ML Overview

    The provided text describes a medical device called MESHEET, an orthodontic plastic aligner, and its substantial equivalence to a predicate device. However, it does not contain information about the acceptance criteria or a specific study proving the device meets acceptance criteria in the context of an AI/software performance evaluation.

    The "Non-Clinical Testing" section mentions some tests were conducted, but these are general medical device validations and not a performance study for a diagnostic or prognostic AI algorithm. The tests mentioned are:

    • Biocompatibility tests (cytotoxicity, irritation, Acute systemic toxicity, Bacterial reverse mutation, In vitro chromosome aberration, Intramuscular tests) in accordance with ISO 10993-1. These are for material safety.
    • OTS software validation test according to FDA guidance. This is for the software used in the manufacturing process, not for an AI that interprets medical data.
    • Manufacturing Process Validation Testing to validate the fit, form, and function of the final product and ensure it achieves its intended use. This is a manufacturing quality control, not an AI performance study.

    Therefore, I cannot provide the requested information about acceptance criteria, device performance, sample sizes, ground truth establishment, or AI-specific study details because this information is not present in the provided document. The document primarily focuses on establishing substantial equivalence to a predicate device based on intended use, materials, technical characteristics, and standard medical device safety and manufacturing tests, not on the performance of an AI or software that makes diagnostic/prognostic decisions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230126

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-08-04

    (199 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dentis s-Clean Regular Abutment is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

    Device Description

    Dentis s-Clean Regular Abutment is intended for use as an aid in prosthetic restoration. It consists of Abutments, components, and Abutment screws.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies. It's important to note that this document is a 510(k) summary for a dental implant abutment, and as such, it focuses on demonstrating "substantial equivalence" to existing cleared devices rather than establishing novel safety and effectiveness criteria from scratch. Therefore, the "acceptance criteria" here are largely implied by the equivalence to predicate devices and adherence to relevant standards.

    General Observation: The document primarily focuses on demonstrating substantial equivalence (SE) based on comparing the subject device's (Dentis s-Clean Regular Abutment) technical characteristics, materials, and intended use to various predicate and reference devices. The "performance" being evaluated is largely through compliance with international standards and leveraging prior testing on similar devices. There isn't a direct "device performance" in terms of clinical outcomes with AI assistance or human reader improvement, as this is a physical medical device (dental abutment).


    1. A table of acceptance criteria and the reported device performance

    Given the nature of the device (dental abutment) and the submission type (510(k) for substantial equivalence), the "acceptance criteria" are predominantly about meeting material specifications, standardized mechanical testing, and demonstrating equivalence in design and intended use to legally marketed predicate devices. The "reported device performance" is largely the successful completion of these tests or the justification for leveraging prior testing.

    Acceptance Criterion (Implied by Standards & Equivalence)Reported Device Performance
    Mechanical Performance: - Resistance to fatigue under worst-case scenario. (ISO 14801:2016)- "Fatigue Testing under the worst-case scenario according to ISO 14801:2016" was performed on the subject device and "results of the above tests have met the criteria of the standards."
    Sterilization Efficacy (for SAVE Wide Cap): - Validation of sterilization process. (ISO 11137-1,2,3)- "Sterilization Validation Test for SAVE Wide Cap according to ISO 11137-1.2,3" was "performed for predicate devices and leveraged for the subject device." Implied to have met criteria.
    Shelf Life (for SAVE Wide Cap): - Confirmation of product stability over time. (ASTM F1980)- "Shelf Life Test for SAVE Wide Cap according to ASTM F1980 referenced in K171027" was "performed for predicate devices and leveraged for the subject device." Implied to have met criteria.
    End User Sterilization Efficacy (for Abutments): - Validation of end-user sterilization methods for various materials. (ANSI/AAMI ST79, ISO 17665-1,-2, ISO 11737-1,-2, ISO 11138-1)- "End User Sterilization Validation Test Report for Abutments made with Ti-6Al-4V ELI according to [standards] referenced in K111364" was "performed for predicate devices and leveraged for the subject device." - "End User Sterilization Validation Test Report for Abutments made with PEEK, POM and CP Titanium Grade 4 according to [standards]" was performed. - All implied to have met criteria.
    Biocompatibility: - Safety concerning biological interaction with human tissue. (ISO 10993-1:2009)- "Biocompatibility testing for subject Abutments according to ISO 10993-1:2009 referenced in K171027, K171694, and K222913" was performed on the "predicate device and leveraged for the subject device because both products are manufactured with same materials and manufacturing process." - "Demonstrates that the subject device is biocompatible and substantial equivalence with the predicate."
    Design and Material Equivalence: - Comparison of dimensions, materials, and general design to predicate devices.- Detailed comparison tables are provided for each component of the Dentis s-Clean Regular Abutment against specific predicate/reference devices, showing "same indications for use," similar or same dimensions (with differences justified as not impacting performance), and identical materials where applicable. - Example for s-Clean Couple Abutment: "Subject Device and Primary Predicate(K171027) have same indications for use, similar sizes...material, and sterilization method. Both devices are substantial equivalent."
    Intended Use Equivalence: - The subject device's intended use matches that of predicate devices.- "This system has the same intended use and fundamental scientific technology as its predicate devices." The Indications for Use statement is provided and implicitly compared to predicate devices.
    Manufacturing Process / Facility Equivalence: - For leveraged testing, assurance that processes are similar/identical.- For leveraged end-user sterilization tests: "because the product category, material, manufacturing process, facility, and packaging of the both products are exactly same."

    2. Sample size used for the test set and the data provenance

    The document specifies "Fatigue Testing under the worst-case scenario according to ISO 14801:2016" was performed on the subject device. However, it does not specify the sample size used for this specific test item.

    For the other tests (Sterilization Validation, Shelf Life, End User Sterilization Validation, Biocompatibility), the data provenance is that they were "performed for predicate devices and leveraged for the subject device" or "conducted on the predicate device and leveraged."

    • Sample Size: Not explicitly stated for any of the tests.
    • Data Provenance: Retrospective (leveraged from previously cleared predicate/reference devices) and some prospective (Fatigue Testing on the subject device). The country of origin of the data is not specified, but the manufacturer is Dentis Co., Ltd. (Korea) and the US correspondent is Withus Group Inc. (USA). The tests were presumably conducted in facilities capable of meeting the cited international standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this type of device submission. This is a physical dental implant component, not an AI/software device that requires expert adjudication for ground truth establishment. The "ground truth" here is compliance with engineering standards and material specifications, verified through laboratory testing.


    4. Adjudication method for the test set

    This question is not applicable. There is no human adjudication process described for the engineering tests performed on a physical dental abutment. The outcome of the tests (e.g., fatigue strength, material composition) is objectively measured against a standard.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This device is a physical dental abutment, not an imaging or diagnostic AI device that would involve human readers or AI assistance.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This device is a physical dental abutment, not an algorithm or software.


    7. The type of ground truth used

    The "ground truth" for the performance of this dental abutment is based on:

    • Engineering Standards and Specifications: Adherence to international standards like ISO 14801 (fatigue), ISO 11137 (sterilization), ASTM F1980 (shelf life), ISO 10993-1 (biocompatibility), and material specifications (e.g., ASTM F136 for Ti-6Al-4V ELI).
    • Material Composition Analysis: Ensuring materials match specified standards (e.g., Ti-6Al-4V ELI, Pure Titanium Gr4, PEEK, POM).
    • Dimensional Accuracy: Ensuring dimensions are within acceptable tolerances.
    • Demonstrated Performance of Predicate Devices: Relying on the established safety and effectiveness of legally marketed predicate devices through showing substantial equivalence in design, materials, and intended use.

    8. The sample size for the training set

    This question is not applicable. This is a physical device, not an AI/machine learning model, so there is no "training set."


    9. How the ground truth for the training set was established

    This question is not applicable. Since there is no training set, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230307

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-07-11

    (158 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dentis i-Clean System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

    Device Description

    Dentis i-Clean system is intended for use as an aid in prosthetic restoration. It consists of Abutments and Abutment screws.

    AI/ML Overview

    The Dentis i-Clean System is a dental implant system consisting of abutments and abutment screws. The device is not an AI/ML device, and therefore the standard acceptance criteria for such devices, along with the study design elements listed, do not apply.

    Here's an overview of the non-clinical testing performed to demonstrate substantial equivalence, as described in the provided document:

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML device, there are no performance metrics like sensitivity, specificity, or AUC with acceptance criteria described in the document. The substantial equivalence is based on meeting engineering and biological safety standards for medical devices.

    The non-clinical tests performed and their respective standards are:

    Acceptance Criteria (Test)Reference StandardReported Device Performance (Compliance)
    Fatigue TestingISO 14801:2016Performed under worst-case scenario.
    End-user Sterilization Validation (PEEK, CCM)ISO 17665-1, -2, ISO 11737-1, FDA guidance "Reprocessing Medical Devices..." (March 17, 2015)Validated the recommended sterilization for all subject devices delivered non-sterile to be end-user sterilized.
    End-user Sterilization Validation (Ti-6Al-4V ELI)ISO 17665-1, -2, ISO 11737-1 (leveraged from K111364)Leveraged from predicate device, demonstrating substantial equivalence.
    Biocompatibility (Ti-6Al-4V ELI)ISO 10993-1:2009 (leveraged from K171027 and K150344)Leveraged from predicate device, demonstrating substantial equivalence.
    Biocompatibility (PEEK, CCM)ISO 10993-1:2009 (leveraged from K171027)Leveraged from predicate device, demonstrating substantial equivalence.
    MRI Safety ReviewFDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment"Non-clinical worst-case MRI review performed using scientific rationale and published literature.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify a "test set" in the context of AI/ML. The tests mentioned are non-clinical (mechanical, sterilization, biocompatibility, MRI review). The sample sizes for these engineering and biological tests are not detailed in this summary but are typically defined by the respective ISO standards. The data provenance is presumed to be from laboratory testing performed by, or on behalf of, Dentis Co., Ltd. (Korea, as per submitter address) or the referenced predicate device manufacturers. These are not human data, so "retrospective or prospective" does not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as this is not an AI/ML device involving human experts for ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not an AI/ML device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable in the context of AI/ML. For medical devices, "ground truth" is typically defined by adherence to established engineering standards (e.g., ISO 14801 for fatigue) and biological safety standards (e.g., ISO 10993 for biocompatibility, ISO 17665 for sterilization).

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230246

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-25

    (115 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dentis s-Clean SQ-SL Narrow Implant System is intended for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The Dentis s-Clean SQ-SL Narrow Implant System is limited to the replacement of maxillary lateral incisors and mandibular incisors. It is intended for delayed loading.

    The Dentis s-Clean SQ-SL Narrow Implant System may be used as an artificial root structure for single tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors.

    The implants may be restored immediately

    1. with a temporary prosthesis that is not in functional occlusion,
    2. when splinted together as an artificial root structure for multiple tooth replacement of mandibular incisors, or
    3. for denture stabilization using multiple implants in the anterior mandible and maxilla.

    The implants may be placed in immediate function when good primary stability has been achieved and with appropriate occlusal loading.

    Device Description

    The Dentis s-Clean SQ-SL Narrow Implant System is used to replace missing teeth in various situations ranging from a single tooth loss to the complete loss of incisors teeth. This system is restricted to substitute the maxillary lateral incisors and mandibular incisors. It is one and two stage endosseous screw type implant with internal connection, intended for single use. Dentis s-Clean SQ-SL Narrow Implant System is a suitable treatment option when the possibility of placing a standard implant is limited due to physical conditions, where the horizontal space is limited by adjacent teeth and roots, or in situations with a narrow alveola ridge.

    AI/ML Overview

    The provided text is a 510(k) summary for a dental implant system. It focuses on demonstrating "substantial equivalence" to predicate devices, rather than presenting a study to prove a device meets specific performance acceptance criteria for a novel function. As such, the information required for a study proving device performance against acceptance criteria (sections 1-9 of your request) is largely absent from this type of regulatory document.

    However, I can extract information related to the acceptance criteria for ensuring safety and performance based on the type of testing performed, which is primarily non-clinical (bench) testing.

    Here's an attempt to answer your questions based on the provided text, highlighting where information is absent or not applicable to this type of regulatory submission:


    Device: Dentis s-Clean SQ-SL Narrow Implant System

    Indications for Use (from K230246): The Dentis s-Clean SQ-SL Narrow Implant System is intended for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The Dentis s-Clean SQ-SL Narrow Implant System is limited to the replacement of maxillary lateral incisors and mandibular incisors. It is intended for delayed loading.


    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria with specific numerical performance metrics. Instead, it lists the types of tests performed and states that the device was found "substantially equivalent." For specific performance criteria, one would typically refer to the referenced ISO standards (e.g., ISO 14801:2016 for fatigue). The "reported device performance" is implicitly that it met the requirements of these standards.

    Acceptance Criteria (General Category)Reported Device Performance (Implicitly Met)
    Mechanical/Fatigue StrengthComplies with ISO 14801:2016 (Fatigue Testing under worst-case scenario)
    BiocompatibilityComplies with ISO 10993-1:2009 (leveraged from predicate devices)
    Sterilization EfficacyComplies with ANSI/AAMI ST79, ISO 17665-1,-2, ISO 11737-1,-2, and ISO 11138-1 (End User Sterilization Validation, leveraged from predicate devices)
    Material Composition EquivalenceDemonstrated substantial equivalence in chemical composition between subject and predicate (specifically for TiN coating via SEM, EDS, adhesion, thickness)
    MR Environment CompatibilityAssessed via non-clinical worst-case MRI review using scientific rationale and published literature per FDA guidance (magnetically induced displacement force and torque assessed)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any of the tests. For fatigue testing per ISO 14801, a minimum number of samples is typically tested, but the exact number isn't provided here.
    • Data Provenance: Not specified, but the submission is from Dentis Co., Ltd. in Korea. The studies are described as "non-clinical test data," implying bench testing, not patient data. Therefore, "retrospective or prospective" is not applicable in the clinical trial sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a 510(k) submission for a dental implant system, demonstrating substantial equivalence through non-clinical (bench) testing, not through expert-reviewed clinical imaging data or diagnostic performance studies. "Ground truth" in this context refers to the defined metrics and standards of the bench tests (e.g., fatigue cycles to failure, material composition analysis).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept applies to clinical studies involving human readers and interpretations, not to non-clinical bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a dental implant system, not an AI-powered diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical dental implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's non-clinical testing is defined by international and national standards for material properties, mechanical performance, biocompatibility, and sterilization (e.g., ISO 14801:2016, ISO 10993-1:2009, ANSI/AAMI ST79). For the MR environment assessment, it relies on "scientific rationale and published literature."

    8. The sample size for the training set

    Not applicable. This is not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. This is not a machine learning model.


    Summary of the "Study" (Non-Clinical Testing):

    The "study" described in this 510(k) summary is a set of non-clinical (bench) tests designed to demonstrate that the Dentis s-Clean SQ-SL Narrow Implant System is as safe and effective as previously cleared predicate devices.

    • Purpose: To demonstrate substantial equivalence to predicate devices (K161244, K182194, K210080, K171027, K171694) in terms of materials, design, mechanical performance, biocompatibility, and sterilization, rather than proving performance against novel clinical acceptance criteria.
    • Key Tests Performed:
      • Fatigue Testing: Performed on the subject device according to ISO 14801:2016 under worst-case scenarios.
      • Biocompatibility Testing: Leveraged from predicate devices (K171694) and based on ISO 10993-1:2009. The rationale is that the subject device uses the same materials and manufacturing process.
      • End User Sterilization Validation: Leveraged from predicate devices (K171694 and K171027) based on ANSI/AAMI ST79, ISO 17665-1,-2, ISO 11737-1,-2, and ISO 11138-1. The rationale is similar product category, material, manufacturing process, facility, and packaging.
      • Surface Modification (TiN Coating): Data provided to compare chemical composition, SEM, EDS, adhesion, and thickness between subject and predicate (K171694).
      • MR Environment Compatibility: Non-clinical worst-case MRI review performed based on scientific rationale and published literature, referencing FDA guidance.
    • Conclusion: The device "constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use," based on these non-clinical tests and comparisons to predicate devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230523

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-24

    (86 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    s-Clean Link Abutment Narrow is intended for use with dental implant as a support for single or multiple tooth prostheses in mandibular central and lateral incisors and maxillary lateral incisors.

    s-Clean Link Abutment Narrow is intended for use with the s-Clean OneQ-SL Narrow Fixture according to connector part in the chart. All digitally designed zirconia for use with s-Clean Link Abutment Narrow are intended to be manufactured at a Dentis validated milling center.

    Device Description

    The s-Clean Link Abutment Narrow is intended to provide support for customized prosthetic restorations such as crowns and bridges. The s-Clean Link Abutment is composed of a two-piece abutment that is a link abutment at the bottom and a coping (CAD/CAM patient specific superstructure) as the top-half. The submission also includes an abutment fixation screw. The s-Clean Link Abutment are pre-manufactured (stock) abutments, made from a titanium alloy conforming to ASTM F136.

    The top half of the s-Clean link abutment composed of zirconia is referenced from our previously cleared K222913 submission.

    s-Clean Abutment Screw is made of Ti-6Al-4V ELI (ASTM F136) and cleared in K161244.

    The diameters of s-Clean Link Abutment Narrow are 4.0 and 4.5mm.

    s-Clean Link Abutment Narrow is provided non-sterile and must be sterilized after the cementation of the patientspecific superstructure on the Link Abutment Narrow.

    AI/ML Overview

    Accessing the provided text, there is no information about an AI/ML device, its acceptance criteria, or a study proving that it meets those criteria. The document describes a traditional medical device, specifically a dental implant abutment (s-Clean Link Abutment Narrow), and its 510(k) clearance process.

    Therefore, I cannot provide the requested information. The text does not contain details regarding:

    1. A table of acceptance criteria and reported device performance for an AI/ML device.
    2. Sample sizes or data provenance for an AI/ML test set.
    3. The number or qualifications of experts for establishing ground truth for an AI/ML device.
    4. Adjudication methods for an AI/ML test set.
    5. A multi-reader multi-case (MRMC) comparative effectiveness study for AI assistance.
    6. Stand-alone performance of an AI algorithm.
    7. The type of ground truth used for an AI/ML device.
    8. The sample size for an AI/ML training set.
    9. How ground truth for an AI/ML training set was established.

    The document focuses on non-clinical testing for equivalence to a predicate device, material testing, sterilization validation, biocompatibility, and MRI compatibility for a physical dental implant component.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230203

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-03

    (98 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dentis I-FIX Abutment is intended for use as an aid in prosthetic rehabilitation.

    Device Description

    Dentis I-FIX Abutment consists of I-FIX Angled Type Fixture Healing Abutment, I-FIX Angled Type Fixture Healing Abutment Screw, and I-FIX Abutment Screw.

    I-FIX Angled Type Fixture Healing Abutment and I-FIX Angled Type Fixture Healing Abutment Screw are compatible with following Implant Systems:
    Proprietary Name: I-FIX System (Angled Type Fixture)
    Compatible Implants (K number): K083586
    Implant diameter size: Ø 2.0, 2.5, 3.0 mm
    Implant length: 10~16 mm

    I-FIX Abutment Screw is only used with I-FIX Cemented Abutment and I-FIX Free Abutment cleared in K083586.

    The dimensions of abutments are as following:

    1. I-FIX Angled Type Fixture Healing Abutment: (D) 4.03 X (L) 3.2, 4, 5 and 7 mm, Angulation 0°
    2. I-FIX Angled Type Fixture Healing Abutment Screw: (D) 2.15 X (L) 4.0, 4.8, 6.1 and 8.0 mm, Angulation 0°
    3. I-FIX Abutment Screw: (D) 2.15 X (L) 4.7 mm, Angulation 0°

    Tolerance of dimension for Abutments shall be within ± 1% range.

    The Abutments have below featured:

    • I-FIX Angled Type Fixture Healing Abutment: Uses - The healing Abutment is used for protecting inner hole of fixture and adjusting the appropriate height during the healing period. Surface - Non. Materials - Titanium Grade 4.
    • I-FIX Angled Type Fixture Healing Abutment Screw: Uses - The healing Abutment is used for protecting inner hole of fixture and adjusting the appropriate height during the healing period. Surface - Non. Materials - Ti-6Al-4V ELI.
    • I-FIX Abutment Screw: Uses - I-FIX Abutment screw is used with the previously cleared devices, K083586 such as I-FIX Cemented and Free Abutment for connecting with fixture. Surface - Non. Materials - Ti-6Al-4V ELI.

    I-Fix Angled Type Fixture Healing Abutment and I-FIX Angled Type Fixture Healing Abutment Screw are packaged as a set and provided sterilized. I-FIX Abutment Screw is provided non-sterilized.

    Materials:

    • I-FIX Angled Type Fixture Healing Abutment is fabricated from pure titanium (Conforming to ASTM Standard F67)
    • I-FIX Angled Type Fixture Healing Abutment Screw and I-FIX Abutment Screw are fabricated from Ti-6Al-4V ELI (Conforming to ASTM Standard F136)
    AI/ML Overview

    This FDA 510(k) submission for the Dentis I-FIX Abutment indicates that the device is substantially equivalent to predicate devices, meaning it does not require a new study to prove its effectiveness or safety. Instead, the submission relies on existing data from predicate devices and non-clinical testing. Therefore, there is no specific de novo study described for this device with acceptance criteria and reported device performance in the traditional sense of a clinical trial.

    However, I can extract the acceptance criteria that were met through non-clinical testing by referring to the standards and guidance mentioned.

    Here's a breakdown of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table with quantitative acceptance criteria and specific numerical reported device performance for the Dentis I-FIX Abutment as one would expect from a de novo clinical study. Instead, it states that various non-clinical tests were performed and that "The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device."

    Here are the types of criteria that were met, based on the standards cited:

    Acceptance Criteria TypeReported Device Performance
    Material ConformanceMeets ASTM Standard F67 (for CP Titanium Gr4) and ASTM Standard F136 (for Ti-6Al-4V ELI)
    Sterilization ValidationMeets ISO 11137-1, 2, 3 (for gamma sterilization) and ANSI/AAMI ST79, ISO 17665-1, -2, ISO 11737-1, -2, ISO 11138-1 (for end-user sterilization)
    Shelf LifeMeets ASTM F1980
    BiocompatibilityMeets ISO 10993-1:2009
    Tolerance of DimensionWithin ± 1% range (as stated on page 4)
    MRI SafetyScientific rationale and published literature (e.g., Woods, Terry O., et al., 2019) used to assess magnetically induced displacement force and torque, meeting parameters per FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify sample sizes for the non-clinical tests performed. These tests are typically conducted on a limited number of device samples in a laboratory setting. The data provenance is also unspecified beyond the general reference to "Non-Clinical Test Data" and the identification of Dentis Co., Ltd. as the manufacturer in Korea.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the submission relies on non-clinical engineering and materials testing, not a study requiring expert clinical assessment for ground truth. The "ground truth" here is objective measurement against established engineering and biocompatibility standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable as there was no clinical study with a test set requiring adjudication of findings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The Dentis I-FIX Abutment is a physical dental device, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. The device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" used for non-clinical testing was based on:

    • Established international and national standards (e.g., ISO, ASTM, ANSI/AAMI).
    • Material specifications (e.g., ASTM F67, ASTM F136).
    • Engineering specifications (e.g., dimensional tolerances).

    8. The sample size for the training set

    This section is not applicable. There was no "training set" in the context of machine learning or AI development for this physical device. The device design and manufacturing processes are based on established engineering principles and materials science.

    9. How the ground truth for the training set was established

    This section is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4