510(k) Data Aggregation

    K Number
    K201479

    Validate with FDA (Live)

    Device Name
    DemeMASK
    Date Cleared
    2020-07-24

    (50 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

    Device Description

    The proposed devices are 3-ply, flat-pleated surgical masks. The devices are manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filter. There surgical masks to be secured on users via earloops made of spandex and nylon. The devices are sold non-sterile and are intended to be single use and disposable.

    AI/ML Overview

    This document is a 510(k) summary for the DemeMASK Surgical Mask. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving the performance of an AI-powered device. Therefore, many of the requested categories related to AI device studies (like expert adjudication, MRMC studies, training set details) are not applicable to this document.

    However, I can extract the acceptance criteria and reported device performance for the physical properties of the surgical mask, along with details about the studies conducted to demonstrate these properties.

    1. Table of Acceptance Criteria and the Reported Device Performance (Bench Testing for DemeMASK Surgical Mask)

    TestAcceptance CriteriaReported Device PerformanceResult
    Resistance to penetration by synthetic blood (ASTM F1862)29 out of 32 passed in 160 mm Hg32 out of 32 passed in 160 mm HgPass
    Sub-micron particulate filtration efficiency at 0.1 micron (ASTM F2299)≥98%≥99%Pass
    Bacterial filtration efficiency (ASTM F2101-19)≥98%≥99%Pass
    Differential pressure (MIL-M-36954)< 6.0 mmH2O/cm²Average 3.6 mmH2O/cm²Pass
    Flame spread (16 CFR 1610)Class 1 Non-FlammableClass 1 Non-FlammablePass

    Biocompatibility Testing

    TestAcceptance Criteria (Implied by Predicate)Reported Device PerformanceResult
    Cytotoxicity (ISO 10993-5)Device is non-cytotoxic (based on predicate device comparison)Under the conditions of the study, the device is non-cytotoxic.Pass
    Skin Sensitization Test (ISO 10993-10)Device is non-sensitizing (based on predicate device comparison)Under the conditions of the study, the device is non-sensitizing.Pass
    Skin Irritation Test (ISO 10993-10)Device is non-irritating (based on predicate device comparison)Under the conditions of the study, the device is non-irritating.Pass

    2. Sample size used for the test set and the data provenance

    • Resistance to penetration by synthetic blood (ASTM F1862): 32 samples were tested.
    • Other performance tests (ASTM F2299, ASTM F2101-19, MIL-M-36954, 16 CFR 1610): The specific sample sizes are not explicitly stated in this summary for each test, but standard test methods typically involve a representative number of samples.
    • Biocompatibility tests (ISO 10993-5, ISO 10993-10): Sample sizes are not explicitly stated.
    • Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. These are typically bench tests performed on manufactured samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. These are bench tests for physical properties of a surgical mask, not a device requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. These are objective bench tests, not involving human adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to a surgical mask, not an AI-powered device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This document pertains to a surgical mask, not an algorithm.

    7. The type of ground truth used

    For the performance tests, the "ground truth" is established by the objective measurements obtained through standardized laboratory test methods (e.g., measuring filtration efficiency, pressure differential, and synthetic blood penetration). For biocompatibility, the ground truth is the biological response observed in standardized in vitro and in vivo (though not human clinical in this case) tests.

    8. The sample size for the training set

    Not applicable. This document pertains to a physical medical device (surgical mask) and not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a surgical mask.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191361

    Validate with FDA (Live)

    Date Cleared
    2019-11-07

    (170 days)

    Product Code
    Regulation Number
    878.4840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DemeDIOX Barbed polydioxanone suture is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.

    Device Description

    DemeDIOX Barbed Absorbable Surgical Suture is an absorbable, sterile surgical monofilament suture composed of polyester, poly (p-dioxanone). The DemeTECH Suture meets all requirements in the latest edition of the USP monograph for absorbable surgical sutures with the exception of diameter. DemeDIOX Barbed Suture consists of an absorbable monofilament strand thread with spiral unidirectional barbs and is available with or without needles in a suture size 5-0 to 2. The material is dyed with D&C Violet No.2. and contains no additives. The barbing allows for tissue approximation without the use of surgical knots.

    AI/ML Overview

    The provided text describes the submission of a 510(k) premarket notification for a medical device: DemeDIOX Barbed Absorbable Surgical Suture. This type of submission relies on demonstrating substantial equivalence to a previously legally marketed device (predicate device) rather than proving de novo safety and effectiveness through extensive clinical trials.

    Therefore, the study design described is focused on non-clinical testing to demonstrate conformance to established standards and comparison to predicate devices, rather than a clinical study measuring human-in-the-loop performance or outcomes in a clinical setting.

    Based on the provided information, here's an analysis of the acceptance criteria and the "study" (non-clinical testing) that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are largely derived from established standards for absorbable surgical sutures, primarily the United States Pharmacopeia (USP) and ISO 10993. The reported device performance demonstrates adherence to these standards and equivalence to predicate devices.

    Acceptance Criteria (Derived from USP 42, ISO 10993, and predicate equivalence)Reported Device Performance (as stated in the 510(k) Summary)
    Material CompositionComposed of polyester, poly (p-dioxanone), same as predicate devices.
    SterilitySterile. Achieved via Ethylene Oxide (EO) sterilization.
    BiocompatibilityConforms to ISO 10993-1. Material (polydioxanone) has
    "known biocompatibility" and "established history of use" in surgical sutures.
    In-vitro ResorptionTesting conducted to further demonstrate substantial equivalence to predicate devices. (No specific numerical results provided, but stated to support equivalence).
    USP 42 Requirements for "Absorbable Surgical Suture"Suture Material meets or exceeds requirements, with the exception of diameter (due to barbs). This includes:
        a. Suture DiameterMeets USP 42 requirements except for diameter. This deviation is explicitly acknowledged and attributed to the presence of barbs on the suture.
        b. Suture LengthMeets or exceeds the performance requirements defined in USP for "Suture Length Requirement" (95% of stated length).
        c. Tensile Strength (<881> USP)Meets or exceeds the performance requirements defined in USP 42 for "Tensile Strength."
        d. Needle Attachment (<871> USP)For sutures supplied with needles, meets or exceeds the performance requirements defined in USP 42 for "Needle Attachment."
    Extractable ColorAssured conformance to methods outlined in USP 42. (The device is dyed with D&C Violet No.2).
    Physical Properties & FunctionalityAssured conformance to suture diameter and length. (Reiterates USP conformance with the noted diameter exception due to barbs). The barbing allows for tissue approximation without knots.
    LabelingSterile, Single Use. Packaging and labeling conform to 21 CFR standards.
    Substantial EquivalenceDemonstrated to be substantially equivalent to the predicate devices (Ethicon PDS Barbed Suture K113004 and DemeTECH Polydioxanone Synthetic Monofilament K082097) in terms of intended use, technical characteristics, material, sterilization, and performance (with the acknowledged exception for diameter due to barbs).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes (e.g., number of sutures tested) for each non-clinical test (tensile strength, diameter, length, etc.). It generally states that "Non-clinical testing was conducted... to prove conformance to the requirements of USP for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1 and in-vitro resorption." and "Physical properties and functionality testing assured that the device conformed with suture diameter and suture length, extractable color and sterility to methods outlined in USP 42." These tests typically involve a statistically relevant number of samples to ensure robust results, as outlined in the respective standards (USP, ISO).
    • Data Provenance: The data is generated from non-clinical laboratory testing performed specifically for this 510(k) submission. No information on the country of origin of the data or whether it was retrospective or prospective is given, but non-clinical tests would inherently be prospective for the purpose of demonstrating conformance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of submission. Ground truth, in the context of diagnostic AI or image analysis, typically involves expert human annotation or clinical outcomes. For a medical device like a surgical suture, the "ground truth" and acceptance criteria are defined by established performance standards (USP, ISO) and comparison to predicate device characteristics as detailed in their respective 510(k) clearances. The performance is measured against these objective standards through physical and chemical testing, not through expert consensus on a test set.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (e.g., 2+1, 3+1) are used in studies where human experts are making subjective assessments, often in diagnostic accuracy studies. For testing a physical device like a suture, the "adjudication" is inherent in the standardized testing process where results are objectively measured against defined criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a MRMC comparative effectiveness study was not done.

    • This type of study is relevant for diagnostic devices, particularly those involving human interpretation (e.g., AI-assisted radiology).
    • The device in question, DemeDIOX Barbed Absorbable Surgical Suture, is a physical surgical suture, not a diagnostic or AI-powered device that assists human readers.
    • The submission explicitly states: "No clinical trials were conducted."

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone study (in the context of an algorithm or AI) was not done.

    • This is not an AI/algorithm-based device.
    • The performance assessment for this device is based on its physical properties, material characteristics, and biological compatibility as measured directly through laboratory tests and comparisons to predicate devices, not through an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" used for this device's acceptance is based on:

    • Established Industry Standards: Primarily the United States Pharmacopeia (USP 42) monographs for absorbable surgical sutures, which define physical and chemical performance requirements (tensile strength, diameter, length, etc.).
    • International Standards: ISO 10993-1 for biocompatibility.
    • Predicate Device Characteristics: Demonstrating that the new device is "substantially equivalent" to legally marketed predicate devices (Ethicon PDS Barbed Suture and DemeTECH Polydioxanone Synthetic Monofilament Suture) in terms of intended use, technology, and safety/performance profile.

    8. Sample Size for the Training Set

    This information is not applicable. This submission is for a physical medical device (surgical suture), not a machine learning model or an AI algorithm. Therefore, there is no "training set" in the sense of data used to train an algorithm. The development of the suture material and manufacturing process would involve internal R&D and quality control, but not a formally defined "training set" for an AI.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" for an AI algorithm in this context. The "ground truth" for the development and testing of the suture is established by the well-defined and often long-standing scientific and engineering principles of material science, biomaterials, and surgical product design, all guided by regulatory standards like USP and ISO.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190777

    Validate with FDA (Live)

    Date Cleared
    2019-07-31

    (127 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DemeFORCE Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries.

    Device Description

    DemeTECH DemeFORCE Suture is a nonabsorbable, sterile surgical monofilament suture composed of Ultra-high-molecular-weight polyethylene. The DemeTECH UHMWPE Suture meets or exceeds all requirements in the latest edition of the USP monograph for nonabsorbable surgical sutures and is provided sterile in various sizes and configurations. The material is available in different colors and may be provided with or without an attached needle(s).

    AI/ML Overview

    This document is a 510(k) summary for the DemeTECH DemeFORCE Nonabsorbable Surgical Suture. It describes non-clinical testing performed to establish substantial equivalence to predicate devices, rather than a study proving the device meets an acceptance criterion for an AI/ML algorithm or diagnostic tool. Therefore, much of the requested information regarding AI/ML studies is not applicable.

    Here's the information that can be extracted or inferred from the provided text, focused on the non-clinical tests performed:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Set by USP and FDA Guidance)Reported Device Performance (Summary)
    Conformance to USP for nonabsorbable surgical sutureMeets or exceeds all requirements in the latest edition of the USP monograph for nonabsorbable surgical sutures and USP 41.
    BiocompatibilityMeets requirements in accordance with ISO 10993-1.
    In-vitro and in-vivo resorptionDemonstrated substantial equivalence to predicate devices.
    Suture diameterConforms to methods outlined in USP 41.
    Suture length (95% of stated label length)Meets or exceeds the performance requirements defined in USP for "Suture Length Requirement".
    Knot pull tensile strengthMeets or exceeds the performance requirements defined in USP for "Tensile Strength" < 881 >.
    Needle attachment strengthMeets or exceeds the performance requirements defined in USP for "Needle Attachment" < 871 >.
    Extractable colorConforms to methods outlined in USP 41.
    SterilityConforms to methods outlined in USP 41.

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each non-clinical test. It states "Non-clinical testing was conducted on the device... to prove conformance to the requirements of USP." The provenance of the data is from in-house non-clinical testing performed by DemeTECH Corporation, as described in the "Non-Clinical Tests Performed" section. It is not retrospective or prospective patient data from a specific country, but rather laboratory testing of the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of submission. Ground truth in this context refers to established standards (USP monographs, ISO 10993) and the results of laboratory tests. There is no mention of human experts establishing ground truth for a "test set" in the context of imaging or diagnostic evaluations.

    4. Adjudication method for the test set

    Not applicable. This is not a study requiring adjudication of expert interpretations. The "test set" is the physical device being tested against predefined technical standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a surgical suture and does not involve AI, human readers, or diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This document is for a physical medical device (suture), not an algorithm.

    7. The type of ground truth used

    The ground truth used for the non-clinical tests were:

    • Established standards: United States Pharmacopeia (USP) monographs for nonabsorbable surgical sutures (specifically USP 41), and FDA's Special Control Guidance Document: Surgical Sutures.
    • International standards: ISO 10993-1 for biocompatibility.
    • Predicate device characteristics: For demonstrating substantial equivalence in areas like in-vitro and in-vivo resorption.

    8. The sample size for the training set

    Not applicable. This is a 510(k) submission for a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/ML algorithm in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181578

    Validate with FDA (Live)

    Date Cleared
    2018-10-11

    (118 days)

    Product Code
    Regulation Number
    878.5035
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DemeTECH PTFE Suture is indicated for use in all types of soft tissue approximation and/or ligation, including dental and general surgeries. The device is not indicated for use in ophthalmic surgery, microsurgery, and peripheral neural tissue.

    Device Description

    DemeTECH PTFE Suture is a nonabsorbable, sterile surgical monofilament suture composed of polytetrafluoroethylene. The DemeTECH PTFE Suture meets or exceeds all requirements in the latest edition of the USP monograph for nonabsorbable surgical sutures and is provided sterile in various sizes and configurations. The material is undyed and contains no additives and is white in appearance and may be provided with or without an attached needle(s).

    AI/ML Overview

    The provided document describes the FDA 510(k) clearance for the DemeTECH PTFE Nonabsorbable Surgical Suture. This is a medical device, not an AI/ML device, therefore, many of the requested criteria are not applicable. However, I can extract the relevant information regarding acceptance criteria and the studies performed.

    Here's the information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (USP 39)Reported Device Performance (Demetech PTFE Suture)
    Nonabsorbable Surgical Suture requirementsMeets or exceeds performance requirements
    Tensile Strength Requirements (<881>)Meets or exceeds performance requirements
    Needle Attachment Requirements (<871>)Meets or exceeds performance requirements
    Suture Length Requirement (95% of stated label length)Meets or exceeds performance requirements
    Biocompatibility (ISO 10993-1)Conforms to requirements
    SterilityConforms to requirements
    Suture DiameterConforms to requirements
    Extractable ColorConforms to requirements

    2. Sample size used for the test set and the data provenance:

    • The document does not specify the exact sample sizes for the non-clinical tests (e.g., number of sutures tested for tensile strength, needle attachment).
    • The data provenance is not explicitly stated in terms of country of origin, but the testing was conducted to meet USP (United States Pharmacopeia) and ISO (International Organization for Standardization) standards, which are internationally recognized. The studies were non-clinical (laboratory/bench top) and in-vitro/in-vivo.
    • The study was not a "test set" in the context of an AI/ML model; rather, it was a series of physical and biological tests on the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable as the device is a surgical suture, not an AI/ML system that requires expert ground truth for image interpretation or diagnosis. The "ground truth" for this device is its physical and biological performance as measured by established standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable as there was no "test set" in the context of human interpretation or AI/ML model output requiring adjudication. The tests involved direct measurement against predefined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. No MRMC study was conducted as this is a physical medical device (suture), not an AI/ML diagnostic or assistive tool. No human "readers" are involved in its performance assessment in this context.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This is not applicable. This is a physical nonabsorbable surgical suture, not a software algorithm or AI/ML device.

    7. The type of ground truth used:

    • The "ground truth" for the device's performance was established by internationally recognized standards:
      • United States Pharmacopeia (USP 39) monograph for nonabsorbable surgical sutures: This defines the physical and performance characteristics such as diameter, length, knot pull tensile strength, and needle attachment strength.
      • ISO 10993-1 for Biocompatibility: This defines the standards for evaluating the biological safety of medical devices.

    8. The sample size for the training set:

    • This is not applicable as the device is a physical medical device and does not involve AI/ML models that require training sets.

    9. How the ground truth for the training set was established:

    • This is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181582

    Validate with FDA (Live)

    Date Cleared
    2018-08-27

    (73 days)

    Product Code
    Regulation Number
    878.4840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DemeDIOX Absorbable Surgical Suture is indicated for use in general soft tissue approximation including pediatric cardiovascular tisue where growth is expected to occur and ophthalmic surgery, but not for use in adult cardiovascular, microsurgery and neural tissue. These sutures are useful where absorbable suture with extended wound support (up to six weeks) is desirable.

    Device Description

    DemeDIOX Absorbable Surgical Suture is an absorbable, sterile surgical monofilament suture composed of polyester, poly (p-dioxanone). The DemeTECH Suture meets all requirements in the latest edition of the USP monograph for absorbable surgical sutures and the modification from the existence 510k K082097 is that thread is winded around the needle, a foam needle park is applied to the needle to securely hold the needle in place with the thread. One to four needles will be placed in a suture size 6-0, 5cm length, no thread attached to the needle. The material is dyed violet or un-dyed and contains no additives.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the DemeDIOX Absorbable Surgical Suture. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study for an AI/algorithm-based medical device. Therefore, much of the requested information regarding AI study details (like sample size for test/training sets, experts for ground truth, MRMC studies, standalone performance, etc.) is not applicable or cannot be extracted from this document, as it pertains to a physical surgical suture.

    However, I can extract the acceptance criteria and the "device performance" in terms of compliance with standards and equivalence to the predicate device.


    1. Table of Acceptance Criteria and Reported Device Performance

    For DemeDIOX Absorbable Surgical Suture, the "acceptance criteria" are compliance with established regulatory and pharmacopeial standards, and "device performance" is demonstrated by meeting or exceeding these requirements and showing substantial equivalence to a predicate device.

    Acceptance Criteria (Standard/Requirement)Reported Device Performance (Compliance/Equivalence)
    USP Monograph for Absorbable Surgical Sutures (latest edition/USP 40/41)DemeTECH Suture meets all requirements in the latest edition of the USP monograph for absorbable surgical sutures.
    Suture Diameter and Suture LengthPhysical properties and functionality testing assured that the device conformed with suture diameter and suture length (methods outlined in USP 41). Suture Material meets or exceeds the performance requirements defined in USP for "Suture Length Requirement" (95% of stated label length).
    Extractable ColorPhysical properties and functionality testing assured that the device conformed with extractable color (methods outlined in USP 41).
    SterilityPhysical properties and functionality testing assured that the device conformed with sterility (methods outlined in USP 41). Labeling specifies "Sterile, Single Use". Sterilization method is Ethylene Oxide (EO).
    Biocompatibility (ISO 10993-1)Biocompatibility testing was conducted in accordance with ISO 10993-1. DemeTECH's material used was selected based on known biocompatibility (per ISO 10993) and established history of use in the surgical suture industry. Biocompatibility data demonstrated substantial equivalence.
    In-vitro and In-vivo ResorptionIn-vitro and in-vivo resorption testing was conducted to further demonstrate substantial equivalence to the predicate devices.
    Predicate Device Equivalence (K082097)DemeDIOX Absorbable Surgical Suture is composed of the same material (Polydioxanone) and has the same design (sterile, flexible, monofilament absorbable thread) as the predicate device. It shares the same product code (NEW), common name (Polydioxanone Absorbable Suture), suture characteristic (absorbable Polydioxanone surgical suture), labeling (Sterile, Single Use), intended use (soft tissue approximation), technical characteristics (monofilament, synthetic absorbable suture prepared from polyester, poly-(p-dioxanone)), material (prepared from polyester, poly-(p-dioxanone)), and sterilization method (Ethylene Oxide (EO)). Most USP performance requirements for "Absorbable Surgical Suture" in USP 40 are met or exceeded, with exceptions for Tensile Strength and Needle Attachment due to specific product configuration (see below).
    Tensile Strength Requirements (USP 40 for "Tensile Strength" <881>)Exception: Length of the thread is too short to perform the knot pull (for the specific configuration tested: suture size 6-0, 5cm length, no thread attached to the needle).
    Needle Attachment Requirements (USP 40 for "Needle Attachment" <871>)Exception: Thread is not attached to the needle (for the specific configuration tested: suture size 6-0, 5cm length, no thread attached to the needle).
    Suture PackagingSuture Material is packaged in the same or equivalent manner with sterile single or double package having labeling conforming to 21 CFR and USP 40.

    Regarding the study that proves the device meets the acceptance criteria:

    The "study" in this context refers to a series of non-clinical tests and demonstrations of equivalence to a legally marketed predicate device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated as a number of "samples" in the traditional sense of a clinical trial or AI dataset. The testing was done on the DemeDIOX Absorbable Surgical Suture itself, likely using batches or representatives of the manufactured suture according to standard test methods for medical devices (e.g., USP monographs).
    • Data Provenance: Not specified. Being a US FDA submission, the tests were presumably conducted in facilities adhering to relevant good laboratory practices (GLP), but the country of origin of the specific data is not mentioned.
    • Retrospective or Prospective: Not applicable in the conventional sense for non-clinical device testing. These were pre-market, non-clinical performance and equivalence tests conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a physical surgical suture, not an AI/algorithm-based diagnostic device where expert ground truth establishment for a test set would be relevant. Compliance was assessed against established pharmacopeial and ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/algorithm study involving human interpretation or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device (surgical suture), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is established by its compliance with:

    • The United States Pharmacopeia (USP) monographs for absorbable surgical sutures (e.g., USP 40, USP 41).
    • ISO 10993-1 for biocompatibility.
    • General device requirements and FDA Guidance for Surgical Sutures 510(k).
      These are objective, quantitative standards for material properties, sterility, and biological safety.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical medical device.