    Device Name
    VITALHEAT
    Date Cleared
    2004-06-22

    (75 days)

    Product Code
    Regulation Number
    870.5900
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VitalHeat™ is designed to Non-Invasively treat hypothermic patients by warming their body core. This is achieved by local application of negative pressure and thermal load (heat) to a distal appendage.

    Device Description

    The Dynatherm Medical, Inc. VitalHeat™ - Warming Mitt - Control Unit The VitalHeat™ is a compact, thermal warming device for use in health care facilities to help patients recover from the discomfort and consequences of lowered core temperature. The device utilizes a technology, which combines sub-atmospheric pressure (SAP) and a heating element on one heat tooknology, which cornent design is to be utilized on a hand). The combination of sub-atmospheric pressure and a heating element allow for the maximum transfer of heat through the heat exchange prosoulature. The compact design allows for minimum coverage of the patient (hand), which should not impede standard patient care and/o full body access.

    AI/ML Overview

    The VitalHeat™ device is a thermal warming device intended to non-invasively treat hypothermic patients by rewarming their body core through the application of negative pressure and a thermal load to the hand.

    Here's an analysis of its acceptance criteria and the study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text focuses on the comparative table with predicate devices rather than explicit acceptance criteria with specific performance values. However, the comparative table implicitly sets the performance expectations for the VitalHeat™ by aligning with or improving upon predicate devices.

    Acceptance Criterion (Implicit)VitalHeat™ Reported Performance
    Intended UsePatient Temperature Control and Maintain
    Intended Environment of UseHealthcare Facilities
    ContraindicationsPatients < 18 years, Peripheral Vascular Disease
    Type of TechnologySub Atmospheric Pressure/Water Paddle, Disposable Mitt
    Pressure DeviceYes - Negative
    Sub-Atmospheric Pressure (mmHg)40 ± 5 mmHg
    Electrical (AC)Yes
    Temperature Range≤ 45 ° C
    Application SiteHand
    Control TypeMicro - Logic
    Size - Controller16 x 6 x 6 In.
    Weight15.0 Lbs.
    MobilityHand-Held, IV Pole MTG, Table Top
    Water Tank200 ml
    Flow Rate> 1000 ml/Min.
    High Temperature AlarmYes
    Water Level AlarmYes - Water Flow
    Sub-Atmospheric Pressure IndicatorYes
    Alarms/IndicatorsLED and Audible
    TimerYes
    SealYes

    2. Sample Size Used for the Test Set and Data Provenance

    The provided summary does not contain information about a test set sample size or data provenance (e.g., country of origin, retrospective or prospective data) for a clinical study. The submission relies on a comparison to predicate devices, electrical safety, EMC, and environmental testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    No information is provided regarding experts used to establish a ground truth for a clinical test set. The submission is a 510(k) premarket notification, which often relies on comparison to predicate devices and bench testing rather than extensive clinical trials with expert-adjudicated ground truth.

    4. Adjudication Method for the Test Set

    No information is provided regarding an adjudication method, as no clinical test set with a need for ground truth adjudication is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No information is provided to suggest that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed. The document focuses on the technical specifications and substantial equivalence to predicate devices, not on human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The VitalHeat™ is a physical medical device, not an algorithm that performs diagnostic or analytical tasks requiring standalone performance evaluation. Therefore, no standalone algorithm-only performance study was conducted or is applicable.

    7. Type of Ground Truth Used

    Based on the provided information, the primary "ground truth" used for substantial equivalence is the performance and specifications of the legally marketed predicate devices (Aquarius Medical Corporation Thermo-STAT - K970367 and AcroTherm - K003368).

    For the device's technical specifications, the ground truth would be established by bench testing and engineering measurements to confirm performance against design specifications (e.g., temperature range, pressure levels, flow rate, alarm functionality).

    8. Sample Size for the Training Set

    The device is a physical therapeutic device, not an AI/ML algorithm that requires a training set. Therefore, the concept of a "training set sample size" is not applicable to this submission.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for an AI/ML algorithm involved, this question is not applicable. The device's functionality and safety are established through adherence to engineering specifications and comparison to predicate devices.