510(k) Data Aggregation

    K Number
    K250778

    Validate with FDA (Live)

    Device Name
    Sil-Flow
    Date Cleared
    2025-05-23

    (70 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sil-Flow is indicated for
    [A] Temporary inlay and onlay treatments of the cavity
    [B] Sealing of openings for implant screws
    [C] Relining material for temporary crowns and bridges
    [D] Block-out material for retentive areas in the dental arch, e.g. before taking impressions
    [E] Covering of the gingival margin
    [F] Fixing of resin matrix during filling placement
    [G] Temporary filling of cavities

    Device Description

    Sil-Flow is a flowable light-cured temporary filling material used for filling and sealing.

    AI/ML Overview

    The provided FDA Premarket Notification (510(k)) for the Sil-Flow device is for a Tooth Shade Resin Material and NOT an AI/Software as a Medical Device (SaMD) product. Therefore, the information typically found in such a submission regarding AI device performance, such as:

    • Acceptance Criteria for AI performance metrics (e.g., sensitivity, specificity, AUC)
    • Study design for AI performance evaluation (e.g., standalone, human-in-the-loop, MRMC)
    • Sample size and data provenance for AI test sets
    • Ground truth establishment (number/qualifications of experts, adjudication methods)
    • Training set details for AI models
    • Effect size of AI assistance for human readers

    are not present in this document.

    This 510(k) submission focuses on demonstrating substantial equivalence to a predicate device (Clip Flow) based on physical and chemical properties, biocompatibility, and intended use for a dental material.

    Here's a breakdown of the requested information based on the provided document, highlighting what is present and what is absent from the perspective of an AI/SaMD device:


    Acceptance Criteria and Study Proving Device Meets Criteria (for Dental Material)

    The acceptance criteria and supporting study for Sil-Flow relate to its physical, chemical, and biological properties as a tooth shade resin material, demonstrating its safety and performance for its indicated uses.

    1. A table of acceptance criteria and the reported device performance

    Test ParameterAcceptance Criteria (from ISO 4049 Protocol)Reported Device PerformanceComments
    Physical/Chemical Properties
    Visual Appearance(Implied to meet standard)Met criteria"All test results met the criteria in the standard."
    Depth of Cure (mm)Not explicitly stated (but referenced ISO 4049)3.1 mmPredicate device (Clip Flow) reported 11mm, suggesting a difference. However, "Both the subject devices and the predicate device have physical parameters which meet the requirements of the relevant ISO standards," implying 3.1mm is acceptable per the standard for Sil-Flow.
    Shor Hardness DNot explicitly stated52 Shor DPredicate (Clip Flow) reported 90 Shore A. Note the difference in Shore scale (D vs. A), making direct comparison difficult without a conversion or further context, but still stated to meet relevant ISO standard.
    Sensitivity to Ambient Light(Implied to meet standard)Met criteria"All test results met the criteria in the standard."
    Water Sorption (µg/mm³)<40 µg/mm³ (from predicate)<40 µg/mm³Explicitly stated and met.
    Water Solubility (µg/mm³)<7.5 µg/mm³ (from predicate)<7.5 µg/mm³Explicitly stated and met.
    Color Stability(Implied to meet standard)Met criteria"All test results met the criteria in the standard."
    Shade Stability(Implied to meet standard)Met criteria"All test results met the criteria in the standard."
    Biocompatibility Testing
    CytotoxicityPass (according to ISO 10993-05)PassedExplicitly stated.
    IrritationPass (according to ISO 10993-10)PassedExplicitly stated.
    SensitizationPass (according to ISO 10993-10)PassedExplicitly stated.
    Acute Systemic ToxicityPass (according to ISO 10993-11)PassedExplicitly stated.
    Stability (Shelf Life)
    Shelf LifeDemonstrated at 3 years3 years"All shelf life determinations use the same testing protocols as the characterization testing which are based on ISO 4049."

    2. Sample sizes used for the test set and the data provenance

    • Sample Size for Test Set: Not specified in the document. The testing summary refers to "protocols based on ISO 4049," which would define the number of samples for each test (e.g., number of specimens for depth of cure, water sorption, etc.), but the exact numbers are not provided.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the testing is in vitro (laboratory testing of the material itself) and biocompatibility testing on biological models, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. For a dental material, "ground truth" is established through standardized laboratory measurements and validated biological tests, not human expert interpretation of clinical images or data in a diagnostic context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. See point 3. This concept applies to human-reviewed clinical data, not material property testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI/SaMD device. No MRMC study was performed as it's typically for assessing the impact of AI on diagnostic accuracy.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an AI/SaMD device. "Standalone performance" refers to an AI algorithm's performance without human interaction.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Laboratory Standard Measurements and Biocompatibility Test Results: The "ground truth" for this device is derived from the results of standardized physical property tests (e.g., specific gravity measurements for water sorption, calibrated hardness testers for Shor hardness, optical measurements for depth of cure) and the results of validated biocompatibility assays following ISO 10993 standards.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/SaMD device. There is no "training set" in the context of an AI model for this dental material. The material itself is manufactured and tested.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200188

    Validate with FDA (Live)

    Date Cleared
    2020-04-01

    (68 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DSI Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patients esthetics and chewing function. DSI implants are intended for single or multiple unit restorations on splinted or non-splinted applications. Premium Spiral implants are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period.

    Device Description

    DSI Dental Implant System is an internal hex implant system with one model of implant at this time, Premium Spiral is a slightly tapered spiral internal hex implant with micro-rings at the implant neck which comes in 3.5, 3.75, 4.2, 5.0, and 6.0 diameter. The implants come in lengths of 8, 10, 11.5, 13 and 16. The 6.0 diameter implant does not come in lengths of 13 or 16mm. All implants and abutments are made of ASTM F136 Ti 6Al 4V ELI. The implants have a grit blasted and acid etched surface. Loc-in abutments are anodized. All types of abutments can be used with all implants. Straight abutments, standard narrow abutments, anatomic angled abutments, straight shoulder abutments, and straight wide shoulder abutments are all screw-retained permanent abutments for single or multiple restorations. The system also includes abutment fixation screws. Straight abutments come in 7, 9, or 11mm lengths and shouldered straight abutments come 1, 2, 3, or 4mm in gingival height. Wide straight shoulder abutments come in gingival heights of 1, 2, or 3mm. Standard narrow abutments come in lengths of 7, 9, 11 mm. Straight wide abutments come in lengths of 9 or 11 mm. 15° anatomic angulated abutments come in gingival heights of 1, 2, or 3 mm. 25° anatomic angled abutments come in gingival heights of 1, 2, or 3mm. Straight multi-unit abutments, Loc-in abutments, overdenture abutments, and ball attachments are permanent threaded abutments which are for multi-restorations only. Loc-in abutments come in collar heights of 0.5, 1, 2, 3, 4, 5, and 6 mm. Ball attachments come in gingival heights of 0.5, 1, 2, 3, 4, 5, or 6mm. Overdenture abutments come in gingival heights of 0.5, 1.5 mm. Overdenture abutments are used with removable dentures and partial dentures. Straight multi-units come in gingival heights of 1, 2, 3 or 4 mm. Multi-unit abutments are used for fixed restorations. Straight multi-unit mounted dentures can only be removed by the dentures mounted on flat abutments, overdentures or ball attachments can be removed by the patient. The UCLA is 4.5mm in diameter and is for making gold restorations which are taller than 4mm. Healing caps (3.8, 4.5, 5.5, and 6.0 mm diameter) are temporary use abutments used during the healing phase which come in different gingival heights in order to account for tissue thickness differences. The 3.8mm diameter healing caps come in gingival heights of 3, 4, 5, 6 and 7 mm. The 4.5 and 5.5 mm diameter healing caps come in gingival heights of 2, 3, 4, 5, 6 and 7mm. The 6.3 mm diameter healing cap comes in gingival heights of 2, 3, 4, and 5 mm.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the DSI Dental Implant System, asserting its substantial equivalence to legally marketed predicate devices. The document details the device, its indications for use, and a summary of testing conducted to support this claim.

    It does not describe an AI/ML powered medical device, nor does it present acceptance criteria and a study proving the device meets those criteria in the way typically expected for such products (e.g., performance metrics like sensitivity, specificity, AUC). Instead, the "acceptance criteria" for this dental implant system are framed in terms of demonstrating substantial equivalence to a predicate device, primarily through materials, design, and fatigue performance.

    Therefore, many of the requested points regarding AI/ML device performance (like sample size for test set, ground truth experts, MRMC studies, standalone performance, training set details) are not applicable or cannot be extracted from this document as it pertains to a traditional medical device (dental implants).

    However, I can still extract information related to the device's performance testing and comparison to predicates, interpreting "acceptance criteria" in the context of demonstrating substantial equivalence for a non-AI device.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (for Substantial Equivalence)Reported Device Performance (DSI Dental Implant System)
    Mechanical Strength (Fatigue Performance)Run out limit about the same or higher than other implant systems (predicates).
    BiocompatibilitySurface treatment does not adversely change cytotoxicity. Materials are common in dental implants and meet ASTM F136.
    Sterilization EfficacyImplants sterilized according to ISO 11137-1, ISO 11137-2, and ISO 11137-3. Abutment steam sterilization done according to ISO 17665-1 and ISO 17665-2.
    Package Integrity & Shelf LifePackage integrity testing and accelerated aging conducted.
    Endotoxin ContentEndotoxin testing according to USP 161 conducted.
    Indications for Use (Similarity)Substantially equivalent to SpiralTech Dental Implant System.
    Material (Similarity)Ti6Al4V, same as predicate devices.
    Design/Connection (Similarity)Internal Hex connection, similar to predicate devices. Slight differences in geometry and size are considered minor.

    2. Sample size used for the test set and the data provenance
    The document does not specify exact sample sizes for each test (e.g., how many implants were fatigue tested) or the provenance (country of origin, retrospective/prospective) of any clinical data, as this is primarily a non-clinical, performance-based submission. The testing appears to be lab-based.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. Ground truth in the context of AI/ML evaluation by experts is not relevant here. The evaluation is against engineering standards and comparison to predicate devices, which is typically conducted by engineers and regulatory experts.

    4. Adjudication method for the test set
    Not applicable for this type of device and submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    The "ground truth" or reference for evaluating this device's performance is based on recognized international standards (ISO, ASTM, USP) for medical devices, specifically dental implants, and comparison to legally marketed predicate devices. For example, mechanical strength is compared against the run-out limits established by ISO 14801.

    8. The sample size for the training set
    Not applicable. This is not an AI/ML device, and therefore does not have a "training set" in that context.

    9. How the ground truth for the training set was established
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1