510(k) Data Aggregation

    K Number
    K103600

    Validate with FDA (Live)

    Date Cleared
    2011-04-08

    (121 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Portable X-ray System (Model: BIOX) is intended to be used by trained dentists and dental technicians as extra-oral X-ray source for producing diagnostic X-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.

    Device Description

    The Portable X-ray System (Model: BIOX) is portable dental X-ray system that operates on 24VDC supplied by a rechargeable Li-Polymer battery pack. The X-ray controls and power source are assembled into a single hand-held enclosure. The package includes battery charger. The Portable X-ray System generates and controls X-ray in order to diagnose teeth and jaw. It is composed of X-ray generator, controller and beam limiting device. Operating principle is that X-ray generated by high voltage electricity into X-ray tube, which penetrates teeth and jaw, and makes X-ray images on a receptor (chemical film or digital sensor). The Portable X-ray System (Model: BIOX) is a diagnostic x-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic x-ray images using intra-oral image receptors. Its use is intended for both adult and pediatric subjects. This device includes a high frequency inverter that changes direct current to alternating current, X-ray tube, electrical protective devices, and other elements. The portable X-ray system (Model: BIOX) provides sharp and clear images and protects patients and dentists from leakage radiation by using a lead protective seal.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a medical device, the Portable X-Ray System (Model: BIOX). This document focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific acceptance criteria in terms of diagnostic performance (e.g., sensitivity, specificity for detecting a particular condition).

    Instead, the acceptance criteria and proof provided relate to the device's technical specifications, safety, and performance as an X-ray source, ensuring it is as safe and effective as the predicate device.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to the predicate device and compliance with various IEC standards. The "reported device performance" refers to the BIOX model's specifications.

    Acceptance Criteria (Implied by Predicate & Standards)Reported Device Performance (BIOX Model)
    Intended Use EquivalenceIntended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic X-ray images using intraoral image receptors. For both adult and pediatric subjects.
    Energy SourceRechargeable 24V DC Lithium polymer battery pack Or DC 24V Power Supply
    Source to Skin Distance210 Cm
    Cone Diameter6 Cm
    Exposure Time Range0.01-1.6 seconds, 0.01 increments
    Time Accuracy±(10% + 1ms)
    mA (Tube Current)3mA fixed
    kVp (Tube Voltage)60kV fixed
    Wave FormConstant Potential (DC)
    Safety, EMC, and Performance StandardsIEC 60601-1, IEC 60601-1-2, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32
    User InterfaceExposure time: up, down select buttons of parts of teeth, adult and child, film and sensor with display
    Exposure SwitchControl panel and remote controller
    Tube Head MountingYes

    2. Sample size used for the test set and the data provenance

    The document does not describe a clinical study in the form of a diagnostic image interpretation trial with a "test set" of patient data. The "tests" mentioned are:

    • Bench testing
    • Clinical evaluation (not detailed, but likely referring to a general assessment of functionality and safety, not a diagnostic accuracy study).

    Therefore, there is no sample size for a test set of patient cases or information about data provenance in the context of diagnostic performance. The testing was focused on engineering and performance parameters of the X-ray device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no diagnostic accuracy study with a "test set" and "ground truth" for patient conditions was described, no expert consensus was established for such a purpose.

    4. Adjudication method for the test set

    Not applicable. There was no diagnostic accuracy study described that would require an adjudication method for case interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The device is an X-ray source, not an AI-powered image analysis tool. Its purpose is to produce images, not to interpret them or assist human readers with AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is an X-ray source, not an algorithm, and does not operate in a "standalone" or "human-in-the-loop" performance context related to AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not describe the establishment of ground truth for diagnostic purposes. The "ground truth" in this context is the physical and electrical performance parameters of the X-ray machine itself, which are verified through various engineering tests and compliance with recognized standards.

    8. The sample size for the training set

    Not applicable. Since this is an X-ray imaging device and not an AI algorithm, there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI algorithm, there is no ground truth established for this purpose.

    Summary of the Study Proving Acceptance Criteria:

    The study proving the device meets the acceptance criteria is described as a combination of:

    • Bench Testing: This likely involved engineering tests to verify the device's electrical, mechanical, and radiation output specifications (e.g., kVp, mA, exposure time accuracy, radiation leakage).
    • Clinical Evaluation: While not detailed, this would typically involve assessing the device's functionality, ease of use, and ability to produce clinically usable images in a practical setting. This is distinct from a diagnostic accuracy study.
    • Compliance with Standards: The device was tested and found to comply with a series of international standards:
      • UL/IEC 60601-1 (General requirements for basic safety and essential performance of medical electrical equipment)
      • IEC 60601-1-3 (Collateral Standard: General requirements for radiation protection in diagnostic X-ray equipment)
      • IEC 60601-2-7 (Particular requirements for the safety of high voltage generators of diagnostic X-ray generators) - likely a typo in the original document, often 2-7 is related to X-ray tubes
      • IEC 60601-2-28 (Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis)
      • IEC 60601-2-32 (Particular requirements for the safety of associated equipment of X-ray equipment)
      • EN/IEC 60601-1-2 (Electromagnetic compatibility - Requirements and tests)

    The conclusion states that "The result of bench testing and clinical evaluation indicates that the subject device is as safe and effective as the predicate devices." This implies that the BIOX device met the safety and performance benchmarks set by the predicate device and the relevant international standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082167

    Validate with FDA (Live)

    Date Cleared
    2008-09-26

    (57 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Portable X-ray System (Model: DIOX-602, PROX) is intended to be used by trained dentists and dental technicians as extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.

    Device Description

    The Portable X-ray System (Model: DIOX-602, PROX) is portable dental X-ray system that operates on 24VDC supplied by a rechargeable Li-Polymer battery pack. The X-ray controls and power source are assembled into a single hand-held enclosure. The package includes battery charger.

    The Portable X-ray System generates and controls X-ray in order to diagnose of tooth and jaw. It is composed of X-ray generator, controller and beam limiting device. Operating principle is that X-ray generated by high voltage electricity into X-ray tube, which penetrates tooth and jaw, and makes X-ray images on receptor (chemical film or digital sensor).

    The Portable X-ray System (Model: DIOX-602, PROX) is a diagnostic x-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral xray source for producing diagnostic x-ray images using intra-oral image receptors.

    Its use is intended for both adult and pediatric subjects. This device includes a high frequency inverter that changes direct current to alternating current, X-ray tube, electrical protective devices, and other elements. The portable X-ray system (Model: DIOX-602, PROX) provides with sharp and clear images and keeps patients and dentists away from radiation using small dose of radiation.

    AI/ML Overview

    The provided text is a 510(k) summary for the DIGIMED Portable X-Ray System. It primarily focuses on demonstrating substantial equivalence to a predicate device and outlines technical specifications rather than specific performance acceptance criteria or a detailed study proving the device meets them.

    Therefore, many of the requested details, especially those related to AI algorithm performance, sample sizes for training/test sets, expert adjudication, and ground truth establishment, are not present in this document, as the device is an X-ray system, not an AI-powered diagnostic tool.

    Here's a breakdown of what can be extracted and what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state numerical acceptance criteria or performance metrics for diagnostic accuracy in the way an AI algorithm study would. Instead, it relies on demonstrating substantial equivalence to an existing predicate device based on technical specifications and safety/EMC testing.

    FeaturePredicate Device (PORT-X II) PerformanceSubject Device (DIOX-602, PROX) PerformanceAcceptance Criteria (Implied)
    Energy Source22.2V DC Li-Polymer battery24V DC Li-Polymer batteryComparable rechargeable battery power
    Exposure Time0.01-2.0 seconds0.01-1.6 secondsWithin clinically acceptable range for dental X-rays
    Time Accuracy$\pm$ (10%+1ms)$\pm$ (10%+1ms)Same accuracy as predicate device
    mA2mA fixed2mA fixedSame fixed current as predicate
    kVp60kV fixed60kV fixedSame fixed voltage as predicate
    Wave FormConstant Potential (DC)Constant Potential (DC)Same wave form as predicate
    Safety & EMCIEC 60601-1, -2-7, -2-28, -2-32IEC 60601-1, -2-7, -2-28, -2-32Compliance with international safety and EMC standards
    Source to Skin Dist20cm20cmSame distance as predicate (important for radiation dose)
    Cone Diameter7cmDIOX-602: 6.5cm, PROX: 6cmComparable beam limiting
    User InterfaceExposure time: up, down. Selection buttons of parts of teeth, adult/child, film/sensor with display.Exposure time: up, down. Two buttons for modes and selection of parts of teeth, adult/child, film/sensor with display.Comparable user-friendly interface for dental practitioners
    Exposure SwitchControl panel and remote controllerControl panel and remote controllerSame flexible control options
    Intended UseExtraoral x-ray source for diagnostic x-ray images using intraoral image receptors for adult/pediatric subjects.SameSame intended use as predicate

    Study Proving Device Meets Acceptance Criteria:

    The document states: "The result of bench testing and clinical evaluation indicates that the subject device is as safe and effective as the predicate devices."

    And "EMS testing was conducted by EMC Compliance Co., Ltd. for DIOX-602 and SGS Testing Korea Co., Ltd. for PROX in accordance with Standard EN/IEC 60601-1-2. All test results were satisfactory."

    This indicates that the "study" was primarily bench testing for technical specifications and EMC (Electromagnetic Compatibility) testing to ensure compliance with relevant IEC standards (e.g., IEC 60601 series). A clinical evaluation was also performed, but no details are provided about its design, methodology, or results beyond the general statement of safety and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified.
    • Data Provenance: Not specified. The "clinical evaluation" is mentioned, but no details on where or how the data was collected. Given the manufacturers are Korean, it's possible some data was from Korea, but this is speculative.
    • Retrospective/Prospective: Not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • This information is not provided. As this is an X-ray hardware submission, not an AI diagnostic algorithm, the concept of "ground truth" derived from expert consensus on images is less central. The "clinical evaluation" would likely focus on image quality and safety, possibly involving dentists or technicians evaluating the images produced.

    4. Adjudication Method:

    • Not specified.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. This is not an AI-assisted diagnostic device, so an MRMC study comparing human readers with and without AI assistance is not applicable and was not performed.

    6. Standalone (Algorithm Only) Performance:

    • No. This is an X-ray hardware system, not an algorithm. Standalone algorithm performance is not relevant.

    7. Type of Ground Truth Used:

    • For the "clinical evaluation," the ground truth implicitly would be the diagnostic quality of the X-ray images produced by the device, evaluated by trained professionals, and compared to the images produced by the predicate device and the clinical needs of dentistry. This is not explicitly pathology or outcomes data but rather an assessment of image utility for diagnosis.
    • For the technical and safety testing, the "ground truth" is adherence to the specified technical parameters (e.g., kVp, mA, exposure time accuracy) and compliance with international standards (IEC 60601 series) as verified by tests performed in certified labs.

    8. Sample Size for the Training Set:

    • Not applicable/Not specified. This device does not involve a "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable/Not specified. As there is no AI component, there is no training set and therefore no ground truth established for it.