Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K090307

    Validate with FDA (Live)

    Date Cleared
    2009-06-18

    (132 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DDI Ultrasonic Nebulizer System is an ultrasonic (vibrating mesh) nebulizer system designed to aerosolize medications for inhalation by the patient. The device may be used with pediatric and adult patients in the home, hospital, and sub-acute care settings.

    Device Description

    The DDI Ultrasonic Nebulizer System is a mesh screen ultrasonic nebulizer that operates in an identical fashion as other mesh screen nebulizers. The device creates aerosols of liquid medication by ejection of droplets from a mesh vibrated at ultrasonic frequencies and provide fine particles to the patient's lungs. The DDI Ultrasonic Nebulizer System is powered by a cable connecting to a 115 VAC power source with the AC adapter or two AA alkaline batteries.

    AI/ML Overview

    Here's an analysis based on the provided text, outlining the acceptance criteria and the study details for the DDI Ultrasonic Nebulizer System:

    The provided document (K090307) is a 510(k) summary for a Medical Device. For 510(k) submissions, the primary "study" is often a comparison of technical characteristics and performance to a legally marketed predicate device, rather than a traditional clinical trial with acceptance criteria for new device performance metrics. The goal of a 510(k) is to demonstrate substantial equivalence, not necessarily to prove absolute efficacy or safety with new, independent studies.

    Therefore, the "acceptance criteria" for this device are implicitly the performance specifications and safety standards met by its predicate devices, and the "study" is the comparison against those predicates to establish substantial equivalence.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as quantitative targets in the document. Instead, the DDI Ultrasonic Nebulizer System is deemed acceptable because its performance parameters are "statistically identical" to the predicate devices. The table below presents the key performance characteristics reported for the new device and its predicates, implying that matching or being substantially similar to these predicate values constitutes acceptance.

    ParameterAcceptance Criteria (Implied by Predicate A)Reported Device Performance (DDI Ultrasonic Nebulizer System)
    Intended UseAerosolize medications for inhalationAerosolize medications for inhalation
    Target PopulationPediatric and adultPediatric and adult
    Environment of UseHome, hospital, sub-acute careHome, hospital, sub-acute care
    Anatomical SiteMouthMouth
    Mode of OperationUltrasonicUltrasonic
    Energy TypeElectricityElectricity
    Energy Source115 VAC or 2 AA Alkaline batteries115 VAC or 2 AA Alkaline batteries
    Battery Life4 Hours4 Hours
    Aerosolization ElementMetal MeshMetal Mesh
    Nebulizing MethodVibrating meshVibrating mesh
    Nebulization Rate0.2 ml/min (Predicate A) / 0.25 ml/min (Predicate B)0.2 ml/min
    Particle Size (MMAD)Approx 5 micronsApprox 5 microns
    Patient ConnectorMouthpiece, Optional MaskMouthpiece, Optional Mask
    Patient InterfaceHand-heldHand-held
    UseSingle patientSingle patient
    Dimensions (in)4.3" x 2.2" x 1.7" (Predicate A)4.3" x 2.2" x 1.7"
    Weight98 grams (Excluding batteries) (Predicate A)98 grams (Excluding batteries)
    PortableYesYes
    Reservoir (mL)8 ml maximum (Predicate A)8 ml maximum
    Ultrasonic NebulizerYesYes
    Nebulizer components cleanableYesYes
    Materials of ConstructionPlastic and metalPlastic and metal
    Electrical SafetyMeets IEC 60601:1988 + A1:1991 + A2:1996, EN 60601-1:1990 + A1:1993 + A2:1995 + A13:1996, EN 60 601-1-2:2001Meets the same standards

    Study Details Proving Acceptance

    The "study" for this 510(k) submission is primarily a nonclinical comparison of technical characteristics and performance against two legally marketed predicate devices:

    1. Omron Micro-Air Vibrating Mesh Nebulizer (K062263)
    2. Health & Life Co. Ultrasonic Nebulizer System Model HL 100 (K081738)

    The conclusion states that "All of the performance parameters of the devices are statistically identical, and do not raise any new safety or efficacy." This forms the basis of the substantial equivalence determination.

    1. Sample size used for the test set and data provenance:

      • The document does not describe a clinical "test set" in the traditional sense of patient data. The evaluation appears to be based on bench testing of the device's technical specifications and electrical safety.
      • Data provenance is not explicitly stated as country of origin, but it is implied to be from the manufacturer's internal testing as part of their regulatory submission. The data is retrospective in the sense that it's comparing a new device against established predicate device specifications, and the testing of the new device likely occurred prior to the submission date.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is not applicable as there was no "test set" requiring expert ground truth in the context of diagnostic interpretation. Ground truth for device performance parameters (e.g., nebulization rate, particle size) would have been established through standardized engineering and laboratory testing protocols.
    3. Adjudication method for the test set:

      • Not applicable, as there was no "test set" requiring human adjudication in the context of interpretation.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

      • No, a MRMC comparative effectiveness study was not done. This submission is for a medical device (nebulizer) that delivers medication, not a diagnostic imaging or screening tool that would involve human readers interpreting cases.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This is not applicable, as the device is a physical medical device (nebulizer), not an algorithm or AI system. Its performance is inherent to its design and mechanical/electrical operation.
    6. The type of ground truth used:

      • The "ground truth" for the device's performance characteristics (e.g., nebulization rate, particle size, battery life, dimensions) would have been established through laboratory measurements and engineering specifications. For electrical safety, the ground truth is defined by compliance with the referenced international standards (IEC 60601, EN 60601-1, EN 60 601-1-2).
    7. The sample size for the training set:

      • Not applicable. This device is not an AI or machine learning system, so there is no concept of a "training set" in this context.
    8. How the ground truth for the training set was established:

      • Not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060467

    Validate with FDA (Live)

    Date Cleared
    2006-07-13

    (140 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Prodigy Blood Glucose Test System is intended for use in the quantitative measurement of glucose in capillary whole blood from the fingertip and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and are not intended for use on neonates.

    Device Description

    The Prodigy glucose test system consists of a meter and test strips. The system utilizes an electrochemical method-based meter and dry reagent biosensor (test strips) for blood glucose testing. The size of the current is proportional to the amount of glucose present in the sample, providing a quantitative measurement of glucose in fresh whole blood and control solutions. The system also provides voice broadcast function when the meter is used to test.

    AI/ML Overview

    The provided document (K060467) pertains to the 510(k) summary for the Prodigy Blood Glucose System. After careful review, the document does not contain the detailed acceptance criteria or the specific study results that would allow for a complete description of the device's performance against such criteria.

    The 510(k) submission primarily focuses on establishing substantial equivalence to a predicate device (TaiDoc Technology Corporation, K042005) rather than providing a detailed performance study against pre-defined acceptance criteria.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, nor the detailed information regarding sample sizes, ground truth establishment, or multi-reader multi-case studies, as this information is not present in the provided text.

    Here's what can be extracted and inferred from the text regarding safety and effectiveness:

    1. A table of acceptance criteria and the reported device performance:
    This information is not provided in the document. The document states: "The results of clinical and non-clinical testing indicate that the new device is as safe and effective as the predicate." However, it does not detail what specific acceptance criteria were used or the quantitative performance results against them.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    This information is not provided in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    This information is not provided in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not applicable/provided. The Prodigy Blood Glucose System is an in-vitro diagnostic device for measuring glucose, not an AI-assisted diagnostic tool requiring multi-reader studies with human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    The Prodigy Blood Glucose System is an electrochemical method-based meter. Its performance is inherent to the device itself (strip and meter interaction). The concept of "standalone algorithm only" is not directly applicable in the same way it would be for image analysis AI. The "standalone" performance would be its accuracy in measuring glucose values, which is generally evaluated in clinical trials using reference methods. However, the details of such a trial are not provided in this summary.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    For blood glucose measurement systems, the "ground truth" is typically established by comparing the device's readings against a laboratory reference method (e.g., a YSI analyzer) with high accuracy and precision. While the document mentions "clinical and non-clinical testing data," it does not explicitly state the type of ground truth method used.

    8. The sample size for the training set:
    This information is not provided in the document. (Note: For a traditional electrochemical blood glucose meter, the concept of a "training set" as understood in machine learning is not directly applicable. Calibration data and internal validation data would be collected during development and manufacturing, but specific "training set" size for an algorithm is not relevant here).

    9. How the ground truth for the training set was established:
    As above, this information is not provided and the concept of "training set ground truth" is not directly applicable in the context of this device type.

    Summary of available device performance information (from the "Similarities" table):

    ItemNew Device: ProdigyPredicate K042005
    EnzymeGlucose oxidaseGlucose oxidase
    Test range20-600 mg/dL20-600 mg/dL
    Sample volume0.7 uL1.8-2.5 uL
    Test time7 sec.10 sec.

    This table highlights technical specifications and equivalence, but not detailed accuracy performance against clinical acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1