510(k) Data Aggregation

    K Number
    K251206

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-24

    (159 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital X-Ray DentiMax Pro Imaging System is intended to be used with standard digital X-ray systems to collect dental x-rays photons and convert them into dental images that may be stored, viewed, and manipulated by dentists for the diagnosis of disease of the teeth, jaw, and oral structures.

    Device Description

    The Digital X-Ray DentiMax Pro Imaging System consists of intraoral digital sensors and the optional DentiMax Imaging Software application. This system provides image acquisition, storage, and display functions for dental radiography. The digital intraoral x-ray sensors are based on complementary metal-oxide semiconductor (CMOS) technology. A scintillating material converts the incident x-rays into visible light, the light is coupled optically to a CMOS technology light detection imager and then is converted to digital data. A high-speed direct USB interface enables a simple connection to a computer without need for an additional control box. The design of the sensor assembly supports the automatic detection of incident x-rays to generate digital images for intraoral applications, once armed via a software command. The DentiMax Imaging Software application functions as a viewing software for image acquisition, enhancement, and display; however, it is not required for use of the sensors. The sensors are also compatible with other FDA-cleared dental imaging software commonly available in the marketplace. The optional DentiMax positioner kit helps ensure the system is aligned correctly to capture accurate dental images. The major function of the Digital X-Ray DentiMax Pro Imaging System is to convert X-rays into high-resolution digital images suitable for dental diagnostic use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K220556

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-04-13

    (44 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OpenSensorX Series is used in conjunction with dental Radiography in medical units. The product is used for dental X-ray examination and the diagnosis of structural diseases. The product is expected to be used in hospitals and clinics, operated and used by trained professionals under the guidance of doctors. This device is not intended for mammography and conventional photography applications. This device is suitable for providing dental radiography imaging for both adult and pediatric.

    Device Description

    OpenSensorX series are digital intra-oral sensors. It features a 20um pixel pitch CMOS sensor with directly deposited CsI:Tl scintillator which ensures optimal resolution. An easy to use hi-speed direct USB interface enables a simple connection to a PC without need for an additional control box. The optional iRay intra-oral software application makes it easy to acquire, enhance, analyze, view and share images from the sensor. The major function of the OpenSensorX series is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. This detector is the key component of intra-oral DR system, enables to complete the digitalization of the medical X-ray imaging with the intra-oral DR system software. The OpenSensorX series has two device models, OpenSensor0001X and OpenSensor0002X.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria and a study that proves the device meets those criteria in a structured format. The document is primarily a 510(k) summary for the OpenSensorX Series, focusing on establishing substantial equivalence to a predicate device. It describes general non-clinical studies and states that clinical data is not needed.

    However, I can extract the non-clinical considerations from Section 9, which lists performance parameters tested against standard requirements to establish substantial equivalence. These can be interpreted as implicit acceptance criteria for maintaining equivalence with the predicate device.

    Here's the information extracted and organized to best fit your request:

    Acceptance Criteria and Device Performance

    The "Nonclinical Considerations" section (Section 9.3) outlines various performance aspects that were tested to demonstrate substantial equivalence to the predicate device (DentiMax Digital X-ray Imaging System, K092547). While the specific acceptance values are not always explicitly stated (e.g., "meet standard requirements" or "assured the same as the predicate device"), the reported performance indicates that the OpenSensorX series meets or is substantially equivalent to the predicate for these criteria.

    Acceptance Criteria (Implied from Non-clinical Studies)Reported Device Performance (OpenSensorX Series)
    Electrical Safety & EMCMet standard requirements (IEC/ES 60601-1, IEC60601-2-65, IEC 60601-1-2)
    Biocompatibility (Patient Contact Materials)Evaluated with ISO10993-1; results assured safety same as predicate.
    Sensor Position Frame SafetyEvaluated and assured same as predicate device.
    Dose to Output Signal Transfer FunctionDemonstrated substantial equivalence to predicate device.
    Signal to Noise Ratio (SNR)Demonstrated substantial equivalence to predicate device.
    UniformityDemonstrated substantial equivalence to predicate device.
    DefectDemonstrated substantial equivalence to predicate device.
    Minimum Triggering Dose RateDemonstrated substantial equivalence to predicate device.
    Modulation Transfer Function (MTF)0.1 at 12.5lp/mm (Predicate: 0.1 at 7lp/mm) - Note: This indicates an improvement in spatial frequency response.
    Spatial ResolutionDemonstrated substantial equivalence to predicate device.
    Low Contrast ResolutionDemonstrated substantial equivalence to predicate device.
    Image Acquisition TimeDemonstrated substantial equivalence to predicate device.
    Software Hazard ClassificationHazards classified, requirements and design specifications defined and passed all test cases, complying with intended design.

    Study Details

    The provided text describes non-clinical studies performed to establish substantial equivalence, rather than a clinical study or a comparative effectiveness study with human readers.

    1. Sample size used for the test set and the data provenance: Not explicitly stated for each non-clinical test. The tests are described generally (e.g., "Electrical, mechanical, environmental safety and performance testing"). The product is an intraoral X-ray sensor; thus, the "test set" would refer to the physical devices themselves and the images generated by them under various conditions, rather than a dataset of patient images in the typical sense of AI/clinical studies. Data provenance is not specified beyond being "non-clinical."

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The studies mentioned are primarily engineering and performance tests (e.g., electrical safety, EMC, sensor performance metrics). Ground truth, in this context, would be established by validated test equipment and standard measurement protocols.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. This refers to consensus methods for expert interpretation, which is not relevant for the described non-clinical performance and safety tests.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No. The document explicitly states: "Clinical data is not needed to characterize performance and establish substantial equivalence." This indicates no clinical study, and therefore no MRMC study, was performed. The device is an image acquisition sensor, not an AI-assisted diagnostic tool.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, in effect. The performance metrics (MTF, SNR, uniformity, etc.) are inherent characteristics of the device itself, measured objectively without human interpretation or intervention in the measurement process. The software component also underwent verification and validation tests independently.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the non-clinical performance tests, the "ground truth" is based on established engineering standards, physical measurements, and comparison to the performance specifications of the predicate device. For biocompatibility, it's based on ISO 10993-1. For software, it's based on predefined requirements and design specifications.

    7. The sample size for the training set: Not applicable. This device is an X-ray sensor, not an AI/ML algorithm that requires a training set in the conventional sense. The software mentioned (iRayDR) handles image acquisition, enhancement, analysis, and viewing, implying more traditional software engineering and validation rather than machine learning.

    8. How the ground truth for the training set was established: Not applicable, as there is no mention of a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092547

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-10-30

    (72 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DentiMax Digital X-ray Imaging System is intended to be used with standard digital X-ray systems to collect dental x-rays photos and convert them into electronic impulses that may be stored, and manipulated by dentists for the diagnosis of diseases of the teeth, jaw, and oral structures.

    Device Description

    Digital Dental intraoral Xray Sensor. This device is intended to be used for dental radiographic examination for the diagnosis of diseases of the teeth, jaw, and oral structures.

    AI/ML Overview

    The provided text is a 510(k) summary for the DentiMax Digital X-ray Imaging System. It describes the device, its intended use, and claims substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria, a specific study proving the device meets those criteria, or details regarding ground truth establishment, expert adjudication, or reader studies.

    The 510(k) summary focuses on demonstrating that the DentiMax device has "similar technological characteristics" to a legally marketed predicate device (Accent™ Sensor Digital X-ray System, K050693) and is "equivalent." This is a regulatory pathway that often relies on comparison to existing devices rather than extensive de novo clinical studies with predefined acceptance criteria.

    Therefore, I cannot populate the table or answer the specific questions directly from the provided text. The requested information (acceptance criteria, study details, sample sizes, expert qualifications, etc.) would typically be found in a more detailed clinical or performance study report, which is not part of this 510(k) summary.

    Here's a breakdown of why each requested item cannot be answered with the given text:

    1. Table of acceptance criteria and reported device performance: This information is not present. The 510(k) summary makes a general claim of equivalence but does not list specific performance metrics or acceptance thresholds.
    2. Sample size used for the test set and the data provenance: No test set or study specific to the DentiMax device's performance against acceptance criteria is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no such test set or ground truth establishment process is detailed.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: The document does not mention an MRMC study. It focuses on device characteristics comparison.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device described is an X-ray imaging system, which inherently involves human interpretation. No mention of an algorithm-only performance study is made.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as no specific performance study requiring ground truth is described.
    8. The sample size for the training set: Not applicable, as machine learning training sets are not mentioned.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided FDA 510(k) documentation does not contain the detailed performance study information you are asking for. It focuses on demonstrating substantial equivalence to a predicate device based on technological characteristics and intended use.