Found 2 results

510(k) Data Aggregation

    K Number
    K130476

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-01-08

    (317 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dental Morelli Orthodontic Dental Implant is a threaded titanium dental implant screw intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. The device is used temporarily and is removed after orthodontic treatment.

    Device Description

    The Orthodontic Miniscrew for Absolute Anchorage is an implantable medical device used as a temporary skeletal anchorage point for orthodontic movements. The mini screw is intended to receive low intensity static clinical loads. Maximum dynamic forces may occur at low frequencies for short periods; which is inadequate to compromise the performance of the product through fatigue. The device is available in different lengths (6, 8, 10, mm), diameter 1.5 mm, and supplied in sealed sterile polyethylene bags.

    AI/ML Overview

    The provided document is a 510(k) summary for the "Orthodontic Miniscrew for Absolute Anchorage" and focuses on demonstrating substantial equivalence to a predicate device, the "Aarhus Anchorage System K041527." It does not describe a study involving a comparison of device performance against detailed acceptance criteria in the manner typically seen for novel device efficacy. Instead, it relies on demonstrating equivalence through material properties, dimensions, intended use, and general performance characteristics.

    Therefore, many of the requested sections (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable or cannot be extracted from this document, as the submission primarily focuses on equivalence based on established standards and similarity to a predicate.

    Here's an analysis based on the information available in the provided text:


    Acceptance Criteria and Device Performance

    The acceptance criteria are implicitly defined by compliance with established material and sterilization standards, and by demonstrating performance comparable to the predicate device.

    Acceptance Criterion (Implicit)Reported Device Performance
    Material Composition (conforms to surgical implant standards)Titanium Alloy ASTM F 136 (same as predicate). Adheres to chemical, mechanical, and metallurgical requirements for wrought annealed titanium-6aluminum-4vanadium ELI (extra low interstitial) alloy (R56401).
    BiocompatibilityTitanium Alloy in medical grade according to ASTM-F 136 is accepted for endosseous implant (same as predicate).
    SterilityGamma Radiation (same as predicate). Conforms to ISO 11137 (requirements for developing, validating, and routinely controlling sterilization process). Conforms to ASTM F1980-07 (addresses loss of sterile barrier system integrity).
    Mechanical Safety/PerformanceTensile strength of material according to ASTM - F 136 is accepted for endosseous implants (same as predicate). Performance testing conducted to determine torque levels, forces, and insertion lengths. Measured performance is "comparable to the Aarhus Anchorage System K041527."
    Dimensions (comparable to predicate)Diameter ranges from 1.5 mm, length ranges from 6-10 mm. (Predicate: Diameter ranges from 1.5-2 mm, length ranges from 9-11 mm). The submission states "similar dimensions."
    Performance Characteristics (comparable to predicate)Self-trapping/Self-drilling (same as predicate).
    Intended Use / Indications for UseThreaded titanium dental implant screw intended to serve as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. Used temporarily. (Identical to predicate).
    Substantial Equivalence Conclusion"in all respects based on Intended Use, Indications of Use, Construction Material, Performance and Bench Tests that the Orthodontic Miniscrew for Absolute Anchorage and the Aarhus Anchorage System K041527 are substantially equivalent." Specifically, "Potential adverse effects are identical." and "made from the same material ... and has similar dimensions and characteristics."

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes "Performance Testing" which included "testing was done to determine the torque levels, forces and insertion lengths while inserting the Orthodontic Miniscrew for Absolute Anchorage." However, no specific sample sizes (N) for these tests are mentioned.

    The data provenance is implied to be from bench testing conducted by Dental Morelli Ltda. The country of origin for the device manufacturer is Brazil. The data itself would be from laboratory experiments, not patient data, in this context of a 510(k) submission for mechanical equivalence.


    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not Applicable. This submission is for a medical device (orthodontic miniscrew) and focuses on demonstrating substantial equivalence through material and mechanical properties, as well as adherence to standards. It does not involve a diagnostic algorithm or a scenario where expert-established ground truth is required in the context of a "test set" as one might find in an AI/ML medical device submission. The "ground truth" here is adherence to engineering standards and established performance of the predicate.


    4. Adjudication Method for the Test Set

    Not Applicable. See point 3. No clinical or diagnostic "test set" requiring expert adjudication is described.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No. This document does not mention any MRMC comparative effectiveness study. This type of study is typically used for diagnostic devices to assess human reader performance with and without AI assistance. This device is a physical implant.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is a physical medical device, not an algorithm or AI system.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the performance testing mentioned (torque levels, forces, insertion lengths), the ground truth is based on instrumentation measurements comparing the new device against the performance characteristics of the predicate device. For material and sterilization claims, the ground truth is demonstrated by conformance to established international standards (ASTM F136, ASTM F1980-07, ISO 11137). There is no "expert consensus" or "pathology" in the typical sense for this submission regarding performance.


    8. The sample size for the training set

    Not Applicable. This is not an AI/ML device, so there is no concept of a "training set."


    9. How the ground truth for the training set was established

    Not Applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131197

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-08-27

    (123 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Edgewise and Roth Ceramic Brackets are intended for use in orthodontic treatments. The brackets are affixed to teeth so that pressure can be exerted on the teeth.

    Device Description

    Edgewise and Roth Ceramic Brackets are intended to be bonded to teeth, upon which an orthodontic wire is used to move the teeth to new positions. The Morelli bracket consists of a translucent alumina body. The bracket is uncoated. The Morelli bracket is not self-bonding and requires a primer and adhesive to bond the bracket to the teeth.

    AI/ML Overview

    The provided text is a 510(k) summary for Edgewise Ceramic Brackets and Roth Ceramic Brackets by Dental Morelli Ltda. It describes the device, its intended use, and compares it to a predicate device (Clarity Advanced Ceramic Brackets by 3M Unitek Corporation, K102803).

    This document does not describe a study proving the device meets specific acceptance criteria in the context of an AI/ML medical device. Instead, it demonstrates substantial equivalence to an already legally marketed device to obtain FDA clearance for a physical medical device (orthodontic brackets).

    Therefore, I cannot provide an answer that includes:

    1. A table of acceptance criteria and reported device performance (as these are not defined for an AI/ML device in this document).
    2. Sample size for a test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, or training set details (as these are relevant for AI/ML studies, not this type of medical device clearance).
    3. Type of ground truth (as it would be relevant for an AI/ML study, not for demonstrating substantial equivalence of a physical bracket).

    The document focuses on comparing the new device's characteristics (materials, indications for use, anatomical site, location of use, etc.) to a predicate device. The "acceptance criteria" here is effectively the demonstration of sufficient similarity and safety/effectiveness compared to the legally marketed predicate device.

    However, I can extract the information related to the device comparison, which serves as the basis for establishing "acceptance" for clearance in this context:


    Acceptance Criteria and Device Performance (based on Substantial Equivalence to Predicate Device)

    The "acceptance criteria" for the Edgewise and Roth Ceramic Brackets are met by demonstrating their substantial equivalence to the predicate device, Clarity Advanced Ceramic Brackets (K102803). This means the new devices exhibit similar performance and safety characteristics.

    Table 1: Device Comparison ("Acceptance Criteria" through Substantial Equivalence)

    CharacteristicEdgewise Ceramic Brackets by Dental MorelliRoth Ceramic Brackets by Dental MorelliPredicate: Clarity Advance Ceramic Brackets by 3M Unitek Corporation (K102803)
    Indications for useIntended for use in orthodontic treatments. Affixed to teeth to exert pressure.Intended for use in orthodontic treatments. Affixed to teeth to exert pressure.Intended for use in orthodontic treatments. Affixed to teeth to exert pressure.
    Target PopulationPatients in need of teeth alignment correctionPatients in need of teeth alignment correctionPatients in need of teeth alignment correction
    Anatomical SiteTeethTeethTeeth
    Location of useUse only by professional orthodontistsUse only by professional orthodontistsUse only by professional orthodontists
    MaterialsAluminum oxideAluminum oxideAluminum oxide
    BiocompatibilityAluminum oxide is medical grade and accepted for ceramic brackets.Aluminum oxide is medical grade and accepted for ceramic brackets.Aluminum oxide is medical grade and accepted for ceramic brackets.
    Compatibility with the environment and other devicesAluminum oxide is medical grade and accepted for ceramic brackets.Aluminum oxide is medical grade and accepted for ceramic brackets.Aluminum oxide is medical grade and accepted for ceramic brackets.
    SterilityNon-sterileNon-sterileNon-sterile
    Maxillary In-out (mm)0.940.6-1.20.53-.089
    Maxillary Torque0-7 to +8-7 to +17
    Maxillary Angulation00 to +120 to +8
    Mandibular In-out (mm)0.940.6 - 1.20.51-1.14
    Mandibular Torque0-22 to 0-17 to 0
    Slot0.022"0.022"0.022"

    Study/Evaluation Proving Acceptance:

    The non-clinical data and biocompatibility evaluation presented in the 510(k) submission served as the "study" to prove the device meets acceptance criteria by demonstrating substantial equivalence.

    1. Sample size used for the test set and the data provenance: Not applicable. This is not an AI/ML study involving a test dataset for algorithm performance. The "test set" is the physical device being compared. Data provenance is not described in this context beyond the device's characteristics and materials.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used: For a physical device like an orthodontic bracket, "ground truth" relates to material properties, dimensions, and performance characteristics consistent with established standards or the predicate device. The core "ground truth" for clearance is the performance and safety profile of the legally marketed predicate device. The new device's characteristics are compared against this established profile.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    In summary: The "acceptance criteria" for these orthodontic ceramic brackets were met by demonstrating that they are substantially equivalent to a predicate device (Clarity Advanced Ceramic Brackets) in terms of intended use, indications for use, materials, design, and performance characteristics. The evaluation was a non-clinical comparison and biocompatibility assessment, not a clinical trial or AI/ML performance study.