Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K981226

    Validate with FDA (Live)

    Date Cleared
    1998-09-28

    (188 days)

    Product Code
    Regulation Number
    866.3400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMAGEN™ Parainfluenza virus Group (Types 1, 2 and 3) is a qualitative direct immunofluorescence test for the presumptive detection and confirmation of parainfluenza virus type 1, 2 and 3 antigens in respiratory specimens (nasopharyngeal aspirates) from paediatric populations and in cell culture preparations.

    The IMAGEN™ Parainfluenza virus Types 1. 2 and 3 is a qualitative direct immunofluorescence test for the presumptive detection and differentiation of parainfluenza virus type 1, 2 and 3 antigens respectively in respiratory specimens (nasopharyngeal aspirates) from paediatric populations and in cell culture preparations.

    A negative result obtained following direct staining of nasopharyngeal aspirates should be considered presumptive until confirmed by culture.

    Device Description

    IMAGEN™ Parainfluenza Virus Group (Types 1, 2 and 3) is for the detection and confirmation of the presence of Parainfluenza virus antigens in cell culture preparations and direct specimens (nasopharyngeal aspirates) from paediatric populations. A single reagent is provided which contains a mix of purified murine monoclonal antibodies specific to Parainfluenza virus types 1, 2 and 3, conjugated to fluorescein isothiocyanate (FITC).

    IMAGEN™ Parainfluenza Virus Types 1, 2 and 3 is for the detection and differentiation of Parainfluenza virus type 1, 2 and 3 antigens respectively in cell culture preparations and direct specimens (nasopharyngeal aspirates) from paediatric populations. Three individual reagents are provided which each contain a purified murine monoclonal antibody specific to either Parainfluenza virus type 1, 2 or 3, conjugated to fluorescein isothiocyanate (FITC).

    The technological characteristics of the IMAGEN ™ Parainfluenza Virus Group (Types 1, 2 and 3) and IMAGEN'''' Parainfluenza Virus Types 1, 2 and 3 differ from those of the Predicate Device in that they consist of directly FITC labelled type specific mouse monoclonal antibodies which are used in a one step direct immunofluorescence technique.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the DAKO IMAGEN™ Parainfluenza Virus diagnostic kits:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve a sensitivity of X% against the predicate device"). Instead, it presents the performance of the new device and compares it to a predicate device and viral isolation. Therefore, I will derive the implied acceptance criteria from the reported performance and the overall conclusion.

    Acceptance Criteria (Implied)IMAGEN™ Parainfluenza Virus Group (Types 1, 2 and 3) PerformanceIMAGEN™ Parainfluenza Virus Types 1, 2 and 3 Performance (Specific Types)
    Against Predicate Device:
    • High Correlation98.4% (125/127)Type 1: 100% correlation, sensitivity, specificity Type 2: 100% correlation, sensitivity, specificity Type 3: 98.4% (125/127) correlation, 75% (3/4) sensitivity, 99.1% (122/123) specificity
    • High Relative Sensitivity96.2% (25/26)Type 1: 100% sensitivity Type 2: 100% sensitivity Type 3: 75% (3/4) sensitivity
    • High Relative Specificity99.0% (100/101)Type 1: 100% specificity Type 2: 100% specificity Type 3: 99.1% (122/123) specificity
    Against Viral Isolation (Gold Standard):
    • High Correlation97.6% (164/168)Type 1: 100% correlation, sensitivity, specificity Type 2: 100% correlation, sensitivity, specificity Type 3: 97.6% (164/168) correlation, 91.6% (22/24) sensitivity, 98.6% (142/144) specificity
    • High Relative Sensitivity97.9% (47/48)Type 1: 100% sensitivity Type 2: 100% sensitivity Type 3: 91.6% (22/24) sensitivity
    • High Relative Specificity97.5% (117/120)Type 1: 100% specificity Type 2: 100% specificity Type 3: 98.6% (142/144) specificity
    Overall Conclusion (as stated in the document)The IMAGEN™ Tests will provide an accurate means of detecting Parainfluenza virus antigens in respiratory specimens.The IMAGEN™ Tests will provide an accurate means of detecting Parainfluenza virus antigens in respiratory specimens.

    Study Proving Device Meets Criteria:

    The "clinical performance characteristics" study described in the 510(k) summary is the study that proves the device meets the (implied) acceptance criteria. This study compared the IMAGEN™ Parainfluenza Virus kits against the Bartels Viral Respiratory Screening and Identification kit (Predicate Device) and viral isolation.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The IMAGEN™ Parainfluenza Virus Group (Types 1, 2 and 3) test and the IMAGEN™ Parainfluenza Virus Types 1, 2 and 3 test were both assessed on a total of 184 direct specimens.
    • Data Provenance: The study was conducted in 3 routine diagnostic laboratories:
      • 1 in the US
      • 2 in the UK
      • The data is retrospective/observational as it involves assessing specimens within routine diagnostic laboratories.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts or their qualifications for establishing the ground truth.

    • The ground truth for comparison was established by:
      • The Predicate Device (Bartels Viral Respiratory Screening and Identification kit)
      • Viral isolation (considered the gold standard for viral detection).

    For viral isolation, it's generally understood that highly trained laboratory personnel (e.g., virologists, clinical microbiologists, medical technologists with specialized training) perform and interpret the results, but their specific "expert" qualifications, years of experience, or number are not detailed in this document.


    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for conflicting results. The comparisons are presented directly against the predicate device and viral isolation. It implies that if there were discrepancies between the new device and the predicate device, viral isolation was used as the definitive arbiter.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not explicitly described. This document pertains to the performance of an in vitro diagnostic (IVD) device (a lab test), not an imaging AI or a device requiring human reader interpretation in the context of improving diagnostic accuracy. The "readers" here would be the lab technicians performing the tests, but the study focuses on the performance of the reagents themselves, not a human-AI interaction.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this was a standalone performance study of the diagnostic kit. The "algorithm" here refers to the immunofluorescence assay protocol itself, which is executed by laboratory personnel, but the performance metrics (sensitivity, specificity, correlation) are attributed to the kit's ability to detect the virus, independent of real-time human interpretation enhancement or human-in-the-loop decision making.


    7. The Type of Ground Truth Used

    Two types of ground truth were used:

    1. Predicate Device: The Bartels Viral Respiratory Screening and Identification kit.
    2. Viral Isolation: This is considered the definitive gold standard for confirming the presence of live virus and is often used as the ultimate ground truth in virology.

    8. The Sample Size for the Training Set

    The document does not mention a training set or any machine learning/AI development. This is a 510(k) submission for a diagnostic reagent kit, not an AI/ML medical device. Therefore, the concept of a "training set" as it relates to AI is not applicable here.


    9. How the Ground Truth for the Training Set Was Established

    As no training set was mentioned or implied (see point 8), this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973954

    Validate with FDA (Live)

    Date Cleared
    1997-12-22

    (74 days)

    Product Code
    Regulation Number
    866.3330
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMAGEN™ Respiratory Screen is a qualitative indirect immunofluorescence screening test for the presumptive detection of respiratory viruses; Respiratory Syncytial Virus (RSV). Influenza A and B. Parainfluenza types 1. 2 and 3 and Adenovirus in respiratory specimens (nasopharyngeal aspirates) and in cell cultures.

    Device Description

    Not Found

    AI/ML Overview

    Here's an analysis of the provided text regarding the IMAGEN™ Respiratory Screen, structured according to your requested information:

    IMAGEN™ Respiratory Screen Acceptance Criteria and Study Details

    The provided documents (K973954) consist of a summary of safety and effectiveness, the FDA's clearance letter, and the Indications for Use statement for the IMAGEN™ Respiratory Screen. It is a supplement to K962037.

    1. A table of acceptance criteria and the reported device performance

    The provided text does not explicitly state numerical acceptance criteria in a table format, nor does it provide detailed performance metrics (like sensitivity, specificity, accuracy) for the IMAGEN™ Respiratory Screen. Instead, it states that "Performance characteristics for the additional intended uses have been established by external clinical evaluation against the Bartels Viral Respiratory Screening and Identification Kit and standard viral isolation reference methods."

    To fill this table accurately, we would need to refer to "Exhibit E," which is mentioned as containing the detailed performance data. Without "Exhibit E," specific numerical acceptance criteria and reported device performance cannot be provided.

    Hypothetical Table (Illustrative, as actual data is missing from the provided text):

    Performance MetricAcceptance Criteria (Hypothetical)Reported Device Performance (Hypothetical)
    Sensitivity≥ 90% for all target virusesNot Reported (Refer to Exhibit E)
    Specificity≥ 95% for all target virusesNot Reported (Refer to Exhibit E)
    Overall Agreement≥ 92% with reference methodsNot Reported (Refer to Exhibit E)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated in the provided documents. The text mentions "clinical evaluation," but the number of specimens tested is not given.
    • Data Provenance: The study was an "external clinical evaluation." The specific country of origin for the data is not mentioned. Given the regulatory contact is from the UK, it's possible some or all of the clinical evaluation was conducted there or in other European countries, but this is not confirmed. It is a retrospective evaluation, as the data was collected to establish performance characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • The ground truth for the test set was established using "standard viral isolation reference methods" and the "Bartels Viral Respiratory Screening and Identification Kit." These methods are considered the gold standard for viral detection.
    • The text does not specify the number of individual experts or their qualifications involved in interpreting these reference methods for establishing ground truth. The implication is that the reference methods themselves (e.g., viral culture followed by identification) are the "experts" in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • The provided text does not describe any specific adjudication method among human readers for the test set. Since the evaluation was against "standard viral isolation reference methods" and a predicate device (Bartels Kit), the ground truth was inherently established by these objective methods rather than through expert consensus requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. This device is an in vitro diagnostic (IVD) immunofluorescence screening test, not an AI-powered diagnostic tool. The performance description focuses on the agreement of the device's output with reference methods, not on human-reader performance with or without AI assistance.
    • Therefore, an effect size for human readers with/without AI assistance is not applicable and not reported.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, a standalone performance evaluation was done. The IMAGEN™ Respiratory Screen itself is an "algorithm only" in the sense that it produces a result (presumptive detection of respiratory viruses) based on the immunofluorescence reaction. The "clinical evaluation against the Bartels Viral Respiratory Screening and Identification Kit and standard viral isolation reference methods" directly assesses the standalone performance of the IMAGEN™ system. The text indicates it's for use in laboratories where "qualified technicians are familiar with routine indirect immunofluorescence testing," suggesting that while human technicians perform the test, the performance being evaluated is that of the assay itself compared to the gold standard.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The ground truth used was "standard viral isolation reference methods" (e.g., cell culture with subsequent viral identification) and comparison against a legally marketed predicate device, the "Bartels Viral Respiratory Screening and Identification Kit." Viral isolation is considered a gold standard for detecting viable viruses. These are objective laboratory methods, not subjective expert consensus or pathology.

    8. The sample size for the training set

    • The provided documents do not mention a "training set" or its sample size. This is consistent with the nature of an immunofluorescence assay development, which typically undergoes analytical validation and then clinical validation against known specimens rather than learning from a "training set" like an AI model would.

    9. How the ground truth for the training set was established

    • As there is no mention of a "training set" in the context of an immunofluorescence assay, this question is not applicable. The development process for such a device would involve extensive internal validation using characterized specimens, but it's not typically referred to as a "training set" with ground truth established in the same way as machine learning models.
    Ask a Question

    Ask a specific question about this device

    K Number
    K962037

    Validate with FDA (Live)

    Date Cleared
    1996-12-26

    (216 days)

    Product Code
    Regulation Number
    866.3330
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMAGENTM Respiratory Screen is a qualitative indirect immunofluorescence test for the detection of Respiratory Syncytial Virus, Influenza A virus, Parainfluenza virus type 3 and Adenovirus directly in respiratory specimens and Respiratory Syncytial Virus, Influenza A and B virus. Parainfluenza virus types 1, 2 and 3 and Adenovirus in cell culture monolayers.

    Device Description

    The test consists of the following reagents: a Screening reagent, a Negative control reagent, a Fluorescein Isothiocyanate (FITC) Conjugate reagent, Mounting fluid, and Positive and Negative Control slides. It is a two-step direct immunofluorescence staining method.

    AI/ML Overview

    This document describes the IMAGENTM Respiratory Screen, a qualitative indirect immunofluorescence test for detecting various respiratory viruses. The information provided focuses on its performance characteristics and safety.

    Acceptance Criteria and Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity thresholds). Instead, it describes performance characteristics established through external clinical evaluations against a predicate device (Bartels Viral Respiratory Screening and Identification Kit) and standard viral isolation reference methods.

    The reported device performance is presented as a summary of claims that have been established (or are ongoing).

    Acceptance Criteria (Implied)Reported Device Performance
    Directly in respiratory specimens:
    Detection of Respiratory Syncytial virus (RSV)Claims established
    Detection of Influenza A virusClaims established
    Detection of Parainfluenza virus type 3Claims established
    Detection of AdenovirusClaims established
    Detection of Influenza B virusTrials ongoing to collect adequate data
    Detection of Parainfluenza virus type 1Trials ongoing to collect adequate data
    Detection of Parainfluenza virus type 2Trials ongoing to collect adequate data
    In cell culture isolates:
    Detection of Respiratory Syncytial virus (RSV)Data submitted to support claims
    Detection of Influenza A virusData submitted to support claims
    Detection of Influenza B virusData submitted to support claims
    Detection of Parainfluenza virus type 1Data submitted to support claims
    Detection of Parainfluenza virus type 2Data submitted to support claims
    Detection of Parainfluenza virus type 3Data submitted to support claims
    Detection of AdenovirusData submitted to support claims

    Missing critical information includes:

    • Specific sensitivity and specificity values (or other relevant metrics) that were considered acceptable.
    • The statistical methods used to determine if the performance met any implicit criteria.

    Study Information

    The document provides limited details on the specific studies conducted:

    • Sample size used for the test set and the data provenance: The document does not specify the sample size used for the external clinical evaluation (test set). It also does not mention the country of origin of the data or whether the studies were retrospective or prospective. It only states "external clinical evaluation."
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided.
    • Adjudication method for the test set: The document does not describe any adjudication method.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: An MRMC study is not mentioned. The evaluation was against a predicate device and standard viral isolation reference methods, not described as a comparative effectiveness study with human readers.
    • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The IMAGENTM Respiratory Screen is a diagnostic kit (reagents) used by qualified technicians. Its performance is inherently linked to human interpretation, so a purely standalone (algorithm-only) performance is not applicable in the traditional sense for this type of test. Its "standalone" performance would be the performance of the assay itself when processed and interpreted by a technician, which is what the clinical evaluation would assess.
    • The type of ground truth used: The ground truth for the test set was established using "the Bartels Viral Respiratory Screening and Identification Kit and standard viral isolation reference methods." This implies that viral isolation (a microbiological gold standard) was the primary ground truth, with the Bartels kit potentially used as a comparator or secondary reference.
    • The sample size for the training set: This information is not provided. As an immunofluorescence assay kit, it doesn't typically have a "training set" in the machine learning sense. The "training" of the product would involve its development, antibody selection, and optimization, not a data-driven training set like an AI algorithm.
    • How the ground truth for the training set was established: Not applicable in the context of an immunofluorescence test kit, as there's no "training set" in the AI sense. The development of the kit would rely on known positive and negative viral samples for reagent specificity and sensitivity optimization.
    Ask a Question

    Ask a specific question about this device

    K Number
    K940564

    Validate with FDA (Live)

    Date Cleared
    1996-01-11

    (707 days)

    Product Code
    Regulation Number
    866.3120
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1