510(k) Data Aggregation

    K Number
    K213467

    Validate with FDA (Live)

    Date Cleared
    2022-06-14

    (229 days)

    Product Code
    Regulation Number
    886.4670
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVA NEXUS™ Ophthalmic Surgical System is indicated for both anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e., vitreoretinal) ophthalmic subretinal microinjection of gases or aqueous fluids.

    In addition, the optional laser is indicated for the following:

    ConditionTreatment
    Diabetic Retinopathy
    • Proliferative Diabetic Retinopathy• Panretinal Photocoagulation
    • Clinically Significant Macular Edema• Focal or Grid Laser
    Retinal Tear and DetachmentsLaser Retinopathy
    Lattice DegenerationRetinal Photocoagulation
    Sub-retinal (choroidal) NeovascularizationFocal Laser
    Retinal Vascular Occlusion
    • Neovascularization secondary to Brand or Central retinal vein occlusion• Scatter Laser Photocoagulation
    • Chronic macular edema secondary to Branch or Central retinal vein occlusion• Focal or Grid Laser
    Glaucoma
    • Primary Open-angle• Trabeculoplasty
    • Closed Angle• Iridotomy or Iridoplasty
    Device Description

    The EVA NEXUS™ Ophthalmic Surgical System (EVA NEXUS) is a combined anterior and posterior procedure ophthalmic system that is modular in design and is identical in most respects to the recently cleared predicate EVA Ophthalmic Surgical System (K190875). EVA NEXUS (see Figures 1, and 2 for external views) is designed for use in anterior and posterior procedures that require infusion, vitreous cutting, aspiration, illumination, irrigation, lens emulsification and fragmentation, cautery, diathermy as well as photocoagulation.

    AI/ML Overview

    This document describes the EVA NEXUS™ Ophthalmic Surgical System, a combined anterior and posterior procedure ophthalmic system. The submission (K213467) seeks to demonstrate substantial equivalence to its predicate device, the EVA Ophthalmic Surgical System (K190875), and other components cleared under different K numbers.

    The device includes several modifications and new components compared to its predicate, such as a redesigned infusion pole, a second infusion/irrigation port, "Smart IOP" functionality, microinjection capability, a digitally controlled phaco board, increased cutter speed, and a video overlay device.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" in the typical sense of quantitative thresholds for efficacy or safety endpoints. Instead, it relies on demonstrating substantial equivalence to predicate devices through a combination of bench testing, compliance with international standards, and verification/validation testing. The performance aspects are primarily evaluated by comparison to the predicate device and relevant standards.

    However, based on the "Comparison of Technological Characteristics with the Predicate Devices" (Table 5 and subsequent detailed comparisons for individual components like vitrectomy probes, tubing, and laser probes) and "Performance (Bench) Testing" sections, we can infer some performance requirements and reported outcomes:

    Feature/ParameterAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (EVA NEXUS)
    GeneralSimilar functional overview to predicate, with added features (e.g., Micro Injection, Digital Overlay).Provides Phaco-emulsification, Vitrectomy, Diathermy, Irrigation/Aspiration, Illumination, Fluid/Air exchange, Viscous fluid control (injection, extraction, Micro Injection), Laser, Visualization (Digital overlay). (Different)
    DimensionsSimilar to predicate (72x167x60 cm)70x185x60 cm (Similar)
    WeightSimilar to predicate (142 kg)128 kg (Similar)
    User InterfacesSame principle of operation, console screen, multifunction footswitch, remote.Same principle of operation, 19-inch glass color display touchscreen, multifunction footswitch, remote. (Same for core interfaces, Digital Overlay is Different/New)
    Phaco-emulsificationSame principle of operation as predicate.Same principle of operation. (Same)
    Phaco Module (Design)To drive Ultrasound power Phaco handpiece.Digitally controlled driver board. (Similar - change from analog to digital)
    Power Output50 ± 20% [W] (predicate: 50 [W])50 ± 20% [W] (Same)
    Phaco Frequency40kHz, Auto-tuning.40kHz, Auto-tuning. (Same)
    Phaco Stroke Length80 µm +/-20 µm (3002.M), 100 µm +/-20 µm (3002.P).80 µm +/-20 µm (3002.M), 100 µm +/-20 µm (3002.P). (Same)
    Max U/S Velocity of Tip OutputVelocity of tip max. 13.2 m/s.Velocity of tip max. 13.2 m/s. (Same)
    VitrectomySame principle of operation.Same principle of operation. (Same)
    Vitrectomy Pulse Rate20-8000 ppm, tolerance +/- 20% (predicate).Single and 20-10000 ppm, tolerance +/- 20%. (Similar - increased max rate)
    Vitrectomy Cutting RateUp to 8000 cpm (standard), up to 16000 cpm (dual cut) (predicate).Up to 10000 cpm (standard), up to 20000 cpm (dual cut). (Similar - increased max rate)
    Vitrectome Durability (New)Withstand overpressure of 4.5 bar, function throughout lifetime (5 min at 5000 CPM, 5 min at max cut rate), ≥ 150,000 cutting cycles.Successfully met in Disposable Pneumatic Vitrectomes Product Verification Report (RPRT 30612000). (Met)
    DiathermySame principle of operation.Same principle of operation. (Same)
    Diathermy TypeBipolar Coagulation.Bipolar Coagulation. (Same)
    Diathermy Shape of HF-voltageUn-modulated square voltage (predicate).Un-modulated sinusoidal voltage. (Different)
    Diathermy Frequency1 MHz (±10%).1 MHz (±10%). (Same)
    Diathermy Max. Output Power10 W ±10% (at 150 Ohm) (predicate).10 W ±20% (at 100 Ohm). (Similar - different impedance, wider tolerance)
    Diathermy Voltage (peak-peak)200 Vp-p max. (predicate).180Vp-p max. (Similar - lower max voltage)
    Fluidics ModuleTo maintain anterior/posterior chamber stability.Same principle of operation. (Same)
    Irrigation (Infusion) PrincipleGravity or Air pressurized Infusion bottle (predicate).Pressurized infusion by plunger pump, pressure control by squeezing membranes of cartridge. (Different - from gravity to pumped)
    Irrigation (Infusion) ModesGravity/AIC/VGPC (predicate).Fixed/AIC/SMART IOP. (Different - new Smart IOP mode)
    Infusion/Irrigation Pressure0-150 mmHg.0-150 mmHg. (Same)
    Aspiration ModeFlow mode or vacuum mode.Flow mode or vacuum mode. (Same)
    Aspiration - Flow ModeNominal flow 0-90 [ml/min], Vacuum limit 0 to -680 [mmHg].Nominal flow 0-90 [ml/min], Vacuum limit 0 to -680 [mmHg]. (Same)
    Aspiration - Vacuum ModeVacuum 0 to -680 [mmHg], Rise time 300 [ms].Vacuum 0 to -680 [mmHg], Rise time 300 [ms]. (Same)
    Aspiration - BackflushMicro Backflush (Pressure 0-50 mmHg, Time 0-250 ms), Auto Backflush (Pressure 0-50 mmHg, Time 0-250 ms), Manual Backflush (Pressure as Infusion/Irrigation).Micro Backflush (Pressure 0-50 mmHg, Time 0-250 ms), Auto Backflush (no user selectable parameters), Manual Backflush (Pressure as Infusion/Irrigation), Proportional Backflush (Pressure 0-50mmHg). (Similar - Auto Backflush parameters changed, Proportional Backflush added)
    Cartridge and Tubing SetsSame design principle as predicate.Same design principle (collection bag, tubing for irrigation/aspiration, administration set). (Same - with modifications to tubing design for new infusion mechanisms)
    Endo-IlluminationSame principle of operation, LED source, 3 port connections, 40 lm ±20% output, 425-680 nm spectrum, NA=0.6.Same principle of operation, LED source, 3 port connections, 40 lm ±20% output, 425-680 nm spectrum, NA=0.6. (Same)
    Fluid/Air ExchangeSame principle of operation and design.Same principle of operation and design. (Same)
    Pressure Range0 - 150 mmHg.0 - 150 mmHg. (Same)
    Pressure Accuracy<3 mmHg.<3 mmHg. (Same)
    Flow Rate1200 [ml/min].1200 [ml/min]. (Same)
    VFC Module
    Viscous Fluid ExtractionSame principle of operation, 0 to -660 mmHg pressure.Same principle of operation, 0 to -660 mmHg pressure. (Same)
    Viscous Fluid InjectionSame principle of operation, 0 - 6 bar pressure.Same principle of operation, 0 - 6 bar pressure. (Same)
    Micro Injection (New Feature)Not applicable for predicate. Required to inject small amounts of fluid (min 50 µl) through high gauge delivery device.Inject small amount of aqueous fluid through high gauge delivery device, 0 - 2 bar pressure, footswitch controlled. (New Capability)
    LaserSame principle of operation.Same principle of operation. (Same)
    Treatment Laser TypeDiode pumped frequency doubled Nd:YAG Laser, 532nm wavelength.Diode pumped frequency doubled Nd:YAG Laser, 532nm wavelength. (Same)
    Laser ClassClass IV.Class IV. (Same)
    Laser ModesSingle shot, pulsed or continuous.Single shot, pulsed or continuous mode. (Same)
    Laser WavelengthGreen - 532 nm ±5 nm.Green - 532 nm ±5 nm. (Same)
    Laser Power0.05-1.2 W ±20%.0.05-1.2 W ±20%. (Same)
    Laser Pulse Duration50-5000 ms.50-5000 ms. (Same)
    Laser Interpulse Time50-5000 ms.50-5000 ms. (Same)
    Aiming Beam WavelengthRed - 630 - 645 nm.Red - 630 - 645 nm. (Same)
    Aiming Beam PowerLess than 1 mW.Less than 1 mW. (Same)
    SoftwareWindows embedded OS, GUI for parameter setting and control.Windows embedded OS, EVA NEXUS GUI for parameter setting, interfaces for operator to control different modules, and service purpose. (Similar - upgraded software with new features)
    BiocompatibilityCompliance with ISO 10993-1, 10993-5, 10993-10, 10993-12 for patient contact components.Confirmed acceptable by cytotoxicity, Kligman maximization, and intracutaneous irritation testing. (Met)
    Electrical Safety & EMCCompliance with IEC 60601-1, IEC 60601-2-2, IEC 80601-2-58, EN 60601-1-2, 47 CFR Part 15 Subpart B.System complies with these standards. (Met)
    Software V&VCompliance with FDA Guidance for Software in Medical Devices ("Major" level of concern).Verification and validation testing conducted, documentation provided as recommended. (Met)
    Performance (Bench) TestingCompliance with ISO 15004-2 (illuminator), IEC 60601-2-22 (laser), IEC 80601-2-58 (lens removal/vitrectomy).Testing to ensure compliance reported as performed. (Met)
    EVA/EVA NEXUS Performance Comparison (Irrigation, Aspiration, Micro Injection)Functions "similar" to predicate, in relation to IEC 80601-2-58:2016.Report RPRT 30609300 concludes functions are "very similar", substantiating substantial equivalence. (Met)
    Pneumatic CompatibilityCompatible with compressed nitrogen (N2).Reports (RPRT 30554100) conclude no changes impact N2 compatibility, substantially equivalent to predicate. (Met)
    Cartridge VerificationDesigned for EVA NEXUS is substantially equivalent to EVA cartridge.Report RPRT 30518300 shows substantial equivalence. (Met)
    1364.DD BiocompatibilityISO 10993-1 (cytotoxicity) for new components, with prior 510(k) for syringe.Cytotoxicity testing (RPRT 30527100) and adoption of prior 510(k) (K063280) for syringe satisfied requirement. (Met)
    1364.DD SterilizationISO 11135 validation, adequate for product.Sterilization validation (chapter 14.1.3, adoption 2020031 STER) concludes product is less worst case and adopted. (Met)
    1364.DD Shelf LifeAdequate packaging system for storage throughout specified shelf life.Shelf life testing (chapter 14, adoption 2015052 SLAD) concludes 1364.DD is less worst case, studies adopted. (Met)
    1364.DD Performance TestsAll tests performed successfully (RPRT 30492800, section 7.14).All tests performed successfully. (Met)
    1364.DD Package & Performance Post ShippingPackage integrity and functionality post shipping.Testing (chapter 14, adoption 2015052 PIAD) concludes 1364.DD is not worst case, studies adopted. (Met)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes bench testing and compliance with international standards, rather than clinical studies with human participants. Therefore, concepts like "test set sample size" and "data provenance" (country of origin, retrospective/prospective) are not directly applicable in the way they would be for a clinical trial or AI model validation with patient data.

    • Sample Size for Bench Tests: Specific sample sizes for individual bench tests (e.g., number of vitrectomy probes tested for durability, units for electrical safety) are generally not explicitly stated in this high-level summary but would be detailed in the individual RPRT reports (e.g., RPRT 30612000 for vitrectomes). For example, vitrectome durability tests refer to "≥ 150,000 cutting cycles," indicating a performance threshold rather than a sample size of devices.
    • Data Provenance: All testing appears to be laboratory/bench data, performed by the manufacturer, D.O.R.C. Dutch Ophthalmic Research Center (International) B.V., located in the Netherlands.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a surgical system, not a diagnostic AI device that requires expert-annotated ground truth from medical images or similar data. The "ground truth" for its performance is established through adherence to engineering specifications, international medical device standards (e.g., IEC, ISO), and performance equivalence to a legally marketed predicate device.

    Therefore, the concept of "experts establishing ground truth for a test set" with specific qualifications (like radiologists) is not applicable here. The "experts" involved would be the engineers, quality control personnel, and regulatory affairs specialists who designed the tests, interpreted the results, and compiled the submission, ensuring compliance with established norms and standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this is not a diagnostic device requiring human adjudication of results from a test set. The evaluation is based on objective measurements against engineering specifications and international standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human readers or MRMC studies. The device is a surgical system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The EVA NEXUS is a surgical system operated by a human surgeon ("human-in-the-loop" by definition for a surgical tool). It does not function as a standalone algorithm for diagnosis or decision-making.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the EVA NEXUS Ophthalmic Surgical System is:

    • Compliance with established international medical device standards (e.g., IEC 60601 series, ISO 10993 series, ISO 15004-2).
    • Engineering specifications and performance metrics derived from device design and functionality.
    • Substantial equivalence to legally marketed predicate devices, implying that the predicate's established safety and effectiveness forms the basis for the new device's acceptable performance.
    • Successful completion of verification and validation testing against these standards and specifications.

    8. The sample size for the training set

    Not applicable. The EVA NEXUS is a hardware-based surgical system with embedded software for control; it is not an AI/ML model that undergoes "training" on a "training set" of data in the conventional sense. The software verification and validation activities are for ensuring its functional requirements and safety, not for machine learning model training.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" or "ground truth" for such a set in the context of this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190875

    Validate with FDA (Live)

    Date Cleared
    2019-09-25

    (174 days)

    Product Code
    Regulation Number
    886.4670
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVA Ophthalmic surgical system is indicated for both anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e. vitreoretinal) ophthalmic surgery.

    In addition, the optional laser is indicated for the following:

    Condition:Treatment:
    Diabetic Retinopathy
    * Proliferative Diabetic RetinopathyPanretinal Photocoagulation
    * Clinically Significant Macular EdemaFocal or Grid Laser
    Retinal Tear and DetachmentsLaser Retinopathy
    Lattics DegenerationRetinal Photocoagulation
    Sub-retinal (choroidal) NeovascularizationFocal laser
    Retinal Vascular Occlusion
    * Neovascularization secondary to Branch or Central retinal vein occlusionScatter Laser Photocoagulation
    * Chronic macular edema secondary to Branch or Central retinal vein occlusionFocal or Grid Laser
    Glaucoma
    * Primary Open-angleTrabeculoplasty
    * Closed AngleIridotomy or Iridoplasty
    Device Description

    The EVA Ophthalmic Surgical System (EVA) is a combined anterior and posterior procedure ophthalmic system that was cleared by FDA in March, 2015 (K142877). The EVA is designed for use in anterior and posterior procedures that require infusion, vitreous cutting, aspiration, illumination, irrigation, lens emulsification and fragmentation, cautery, diathermy as well as photocoagulation.

    The purpose of this 510(k) is to obtain clearance for minor improvements to the cleared device. These changes include:

    • Footswitch: The footswitch of the EVA has been modified to improve ergonomics and simultaneously integrate the laser pedal functionality. An optional separate laser footswitch will remain available if surgeons prefer.
    • Illumination: Due to quality improvements in the LED output, the light output has been increased to improve illumination with small gauge fibers. However, the system controls the absolute output of the module to 40 lumens as previously cleared. As an example, the illumination output for a 27 gauge fiber has been improved from 5 lumens as cleared to 7 lumens in the proposed system. However, the maximum output, and maximum exposure to patient, remains at 40 lumens for any fiber as previously cleared. Thus there are no risks introduced with this change.
    • The Posterior Module (Air Functionality): To support use of EVA for fluid/air exchanges (F-AX), independent of the compressed air / gas supply available, the hardware design of the Posterior (VFIE, Air, Proportional Scissors) module has been modified to provide air through two independent circuits, instead of one as previously cleared. As a result, the air for F-AX is provided by a separate circuit that draws air from the environment (with appropriate filtration), whilst the compressed gas drive for other EVA functions is provided from the pneumatic input (supplied by the surgical setting - typically compressed air or Nitrogen) as per the current design. With this improved design, the compressed gas input of the system will only be used for internal system operation while air needed for surgical use will be derived from filtered, ambient air. Thus, the consumption of compressed gas used to operate the EVA is reduced.
    • Software: To support these changes the EVA software was upgraded as well as anomaly/bug fixes. A complete detail of these minor improvements can be found in the Software portion of this 510(k) (Section 16 and Annex 5).
    • Sterilization of Reusable Accessories: Minor changes to the conditions recommended in the labeling have been made and are supported by validation and performance testing included in this 510(k).
    • Shelf life of Disposable Accessories: The shelf-life of peel pouch packed disposable accessories has been extended to 5-years and is supported by validation and performance testing included in this 510(k).
    • Accessory Changes: Some packs (combinations of accessories) were discontinued and new configurations added. No new accessories that were not previously cleared were added.
    AI/ML Overview

    The provided text is a 510(k) summary for the D.O.R.C. EVA Ophthalmic Surgical System. This document focuses on demonstrating substantial equivalence to a predicate device, primarily through modifications to ergonomic features, illumination, air functionality, software updates, and accessory sterilization/shelf-life.

    Based on the available information, the document does not describe a study involving an AI/machine learning component that generates performance metrics (e.g., sensitivity, specificity, AUC) against acceptance criteria. Instead, it details various performance tests to ensure compliance with standards and functionality.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study proving the device meets them in the context of an AI/ML-based performance evaluation. The document does not contain information about:

    1. A table of acceptance criteria and reported device performance (in terms of AI metrics).
    2. Sample sizes used for a test set (for AI evaluation).
    3. Number of experts used to establish ground truth or their qualifications.
    4. Adjudication methods for a test set.
    5. MRMC studies or effect sizes for human readers with/without AI assistance.
    6. Standalone (algorithm-only) performance evaluations.
    7. Type of ground truth used for AI evaluation.
    8. Sample size for a training set (for AI).
    9. How ground truth for a training set was established (for AI).

    The performance data section in the 510(k) summary focuses on biocompatibility, electrical safety/EMC, software verification/validation (for general software functionality, not AI performance), and bench performance testing related to physical and functional aspects of the surgical system (e.g., illumination, pneumatic system, compliance with IEC/ISO standards).

    In summary, the provided document does not contain the information requested about AI performance testing or a related study with acceptance criteria for such a system.