Found 19 results

510(k) Data Aggregation

    K Number
    K251025

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-29

    (271 days)

    Product Code
    Regulation Number
    868.1920
    Age Range
    0 - 999
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mon-a-Therm™ Esophageal Stethoscope with Temperature Sensor 400TM is indicated for use in the routine monitoring of temperature, as well as heart and respiratory sounds, in an anesthetized patient. The device is intended for insertion into the esophagus.

    Device Description

    The subject device of this premarket 510(k) notification is referred to as the Mon‑a‑Therm™ Esophageal Stethoscope with Temperature Sensor 400TM.

    The legacy Esophageal Stethoscope (K811862) was originally cleared on July 10, 1981, under the name "Hi-Lo Temp® Esophageal Stethoscope with a Thermistor Sensor." It has since been renamed to "Mon-a-Therm™ Esophageal Stethoscope with Temperature Sensor 400TM."

    The Mon-a-Therm™ Esophageal Stethoscope with Temperature Sensor 400TM is currently the marketed device that is undergoing a material modification to eliminate DEHP and Phthalates.

    The Mon‑a‑Therm™ Esophageal Stethoscope with Temperature Sensor 400TM monitors temperature and heart and breath sounds.

    Features and benefits:

    • Soft, thin cuff
    • Male Luer fitting conveniently attaches to standard acoustical earpieces
    • Long lead wire keeps connector away from the surgical field
    • Compatible with most multifunction patient monitors

    The Mon‑a‑Therm™ Esophageal Stethoscope with Temperature Sensor 400TM is packaged individually as a sterile, single-use device and is available in the following sizes: 12 Fr 50/case CFN 90041, 18 Fr 50/case CFN 90042, 24 Fr 50/case CFN 90043, and 9 Fr 50/case CFN 90049. The device and its packaging are not made with natural rubber latex or phthalates. The type of probe and device size are designated on the unit package.

    A 400 series thermistor is placed in a PVC clear tube and sealed by a blue plastisol cuff. This end of the tube is inserted into the esophagus and directly contacts mucosa. In the opposite end of the tube, there is a luer connector and a thermistor connector. The luer connector should be attached to a standard earpiece and the thermistor connector should be inserted to an appropriate interface cable.

    Each probe is electrically connected to a compatible interface cable which is specified in Table 1. The interface cable connects the probe to a patient monitor, which is compatible with the 400 series thermistor, so that the temperature measurement value is displayed on the screen of the monitor. All patient monitors that meet the specifications for the 400 series thermistor, temperature accuracy, and compatible interface cables are compatible. Refer to Figure 3: Illustration of Patient Monitor Compatibility.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These devices are intended for use in providing tracheal access for airway management.
    These devices are also intended for use with percutaneous dilatational tracheotomy (PDT) procedures.

    Device Description

    The subject devices are sterile, single use dual cannula tracheostomy tube that has 90° curve and biocompatible radiopaque outer cannula with a distal or proximal extension (XLT) designed for patients with challenging anatomy.

    All tracheostomy tubes feature flexible disposable inner cannula with full circumference at proximal end and a flexible laser-etched flange with integrated standard 15mm connector. The cuffed devices are provided with low-pressure TaperGuard™ cuff, featuring a thin compliant wall that, when inflated, adapts and conforms to the irregular borders of the tracheal wall. The cuffed devices have an inflation system consisting of an inflation line, pilot balloon and self-sealing inflation valve, allowing for inflation and deflation of the cuff.

    The subject devices are manufactured from medical grade polyvinyl chloride (PVC) with non-Di(2-ethylhexyl) phthalate (DEHP) plasticizer and are sterilized by ethylene oxide (EO) sterilization method.

    The cuffed and cuffless configurations of subject devices with distal and proximal extensions are available in four (4) sizes from 5.0mm to 8.0mm.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250725

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-03

    (268 days)

    Product Code
    Regulation Number
    878.4964
    Age Range
    18 - 150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hugo™ Robotically Assisted Surgery (RAS) System is intended to assist in the accurate control of instruments and accessories including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/graspers, needle holders, electrosurgical tools and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrosurgery, and suturing during minimally invasive urologic surgical procedures. The system is indicated for adult use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative surgical procedures set forth in the Hugo™ RAS System Indications Document.

    Device Description

    The Hugo™ RAS System is a modular assembly of devices intended to assist a surgeon with the accurate control of instruments and accessories for performing tele-operative surgical procedures. It is a system of software-operated, electro-mechanical, cable-connected capital equipment with compatible instruments and accessories.

    The major components of the Hugo™ RAS System are capital equipment: (1) System Tower, (2) Surgeon Console, and (3) Arm Cart Assemblies (ACAs). Each Arm Cart Assembly (ACA) consists of a movable cart with casters supporting setup arm connected to a modular and extendable robotic arm containing an instrument drive unit (IDU). Instruments and accessories attach to the IDUs of the robotic arms to perform robotically assisted surgery. The Surgeon Console is the surgeon's interface for controlling the Hugo RAS System. The Surgeon Console is operated by the user to control the robotic arms and attached instruments. The System Tower is a mobile cart that houses system computers, high-definition operating room team interactive (ORTI) display, power-management system and cable hooks for cable storage.

    Hugo™ RAS system instruments include wristed instruments enabling surgeons to precisely grab, dissect, manipulate and suture tissue. There are nine different wristed instruments currently part of the Hugo™ RAS System grouped into basic instruments and electrosurgical instruments, including Large Needle Driver, Extra Large Needle Driver, Double Fenestrated Grasper, Toothed Grasper, Cadiere Forceps, Secure Cadiere Forceps, Monopolar Curved Shears, Bipolar Maryland Forceps and Bipolar Fenestrated Grasper. Hugo™ RAS system accessories include 3D glasses, surgical drapes, Endoscope Adapters, Sterile Interface Modules, Instrument Key, and Monopolar Tip Covers. These accessories enable the Hugo™ RAS system and instruments to achieve their intended use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K223158

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-23

    (108 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LigaSure XP Maryland Jaw Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, thick tissue (tissue bundles), and lymphatics is desired. The LigaSure XP Maryland Jaw Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as colorectal, bariatric, urologic, vascular, thoracic, and gynecologic. These may include, but are not limited to, such procedures as Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, and so forth.

    The LigaSure XP Maryland Jaw Sealer/Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure XP Maryland Jaw Sealer/Divider for these procedures.

    Device Description

    The proposed LigaSure™ XP Maryland Jaw Sealer/Divider is a sterile, single-use bipolar vessel sealer. It is labeled as prescription use only. The device connects to the Valleylab™ FT10 Electrosurgical Platform generator for tissue ligation. Energy is applied to tissue interposed between the nano-coated Maryland-style jaws creating a ligation. The jaws contain an independent cutting blade for division of tissue. A distinctive characteristic is the new continuous rotation capability of the jaws and shaft. The shaft is a common 5 mm diameter available in three lengths (23 cm, 37 cm, 44 cm) for various general surgical procedures in both open and minimally invasive (laparoscopic) approaches. The instrument body is a pistol grip design which can be used by right or left-handed users to access the controls. The new device is offered with two handle body designs, One-Step Sealing (LXMJ23S, LXMJ37S, LXMJ44S) or Latching Handle (LXMJ23L, LXMJ37L, LXMJ44).

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria and study proving a digital health device's performance. Instead, it describes a substantial equivalence determination for a surgical instrument, the LigaSure™ XP Maryland Jaw Sealer/Divider.

    Therefore, many of the requested categories in your prompt, such as "number of experts used to establish the ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "training set size," and "ground truth for the training set," are not applicable to this medical device submission, which focuses on hardware and electrosurgical performance rather than an AI-powered diagnostic or assistive tool.

    However, I can extract the relevant performance testing and acceptance criteria information that is present in the document for the LigaSure™ XP Maryland Jaw Sealer/Divider.

    Here's the information structured to the best of my ability given the provided document:


    Acceptance Criteria and Study for LigaSure™ XP Maryland Jaw Sealer/Divider

    This submission details the substantial equivalence of an electrosurgical cutting and coagulation device, the LigaSure™ XP Maryland Jaw Sealer/Divider. The performance testing described focuses on various engineering, biological, and functional aspects of the surgical instrument, not on the performance of a digital health or AI device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Predetermined Device Requirements)Reported Device Performance
    SterilizationCompliance with ISO 11135, ISO 11737-1, ISO 11737-2, ISO 10993-7Met acceptance criteria for sterilization by ethylene oxide (EO).
    Shelf LifeCompliance with ISO 11607-1, ISO 11607-2, ASTM D4169, ASTM F1980Packaging and product integrity acceptable for five-year shelf life.
    BiocompatibilityCompliance with ISO 10993-1 for: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity.Met requirements for all specified endpoints.
    SoftwareN/A (Device does not contain software)N/A
    Electromagnetic Compatibility (EMC) & Electrical SafetyCompliance with IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-2-18Met applicable clauses of standards.
    Bench PerformancePredetermined device requirements for: mechanical/functional performance, visual inspection, device reliability, ex vivo vessel sealing/burst, and ex vivo lymphatic sealing/burst testing.Met all predetermined device requirements.
    Animal PerformancePredetermined device requirements for: in vivo acute hemostasis, acute thermal spread, and chronic hemostasis tissue testing.Met all predetermined device requirements.
    Clinical PerformanceSufficiency of clinical literature to support safety and effectiveness for stated indications.Clinical literature studies were evaluated to further support safety and effectiveness.
    Technological CharacteristicsComparison to predicate devices for aspects like indications for use, sterility, packaging, instrument design, lever latching mechanism, energy activation, shaft/jaws rotation, jaw type/shape, electrical characteristics, compatible energy platform, operating amperage/voltage, direct tissue contacting materials, patient contacting materials, jaw coating, and shaft material.Differences (e.g., continuous 360° rotation, removal of heat shrink on shaft) were deemed not to affect safety and effectiveness and not to raise different questions of safety and effectiveness as demonstrated through performance testing.

    2. Sample size used for the test set and the data provenance
    The document does not specify exact sample sizes for each test set (e.g., number of vessels, lymphatic segments, or animals).

    • Provenance: The tests described (bench, animal) are laboratory-based performance studies, likely conducted by the manufacturer Covidien, llc. The document does not mention data provenance in terms of country of origin or whether human data was retrospective or prospective, as clinical trials for efficacy/safety in humans were not the primary evaluation method shown (literature review was used for clinical support).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not applicable and not provided. The "ground truth" for this device's performance would be objective measurements in engineering, biological models, and animal studies (e.g., burst pressure, hemostasis, thermal spread), not expert consensus on interpretations.

    4. Adjudication method for the test set
    This is not applicable and not provided. Performance outcomes are likely based on direct measurement against predefined physical/biological thresholds.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This is not applicable. This device is a surgical instrument, not an AI or diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This is not applicable. This device is a surgical instrument, not an algorithm.

    7. The type of ground truth used
    The "ground truth" for this device's performance is based on:

    • Engineering standards (e.g., ISO, ASTM, IEC for sterilization, packaging, electrical safety).
    • Biocompatibility standards (ISO 10993-1).
    • Predefined physical and biological thresholds/measurements from ex vivo (vessel/lymphatic sealing/burst) and in vivo (hemostasis, thermal spread) studies.
    • "Clinical literature studies were evaluated to further support the safety and effectiveness use of the proposed LigaSure™ XP Maryland Jaw Sealer/Divider indications for use." This implies that existing clinical evidence for similar devices and principles contributes to the overall justification, but no new clinical study data from the sponsor is explicitly referenced.

    8. The sample size for the training set
    This is not applicable. This is a hardware device; there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established
    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213602

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-14

    (60 days)

    Product Code
    Regulation Number
    876.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ritus™ tunneling tool kit is used to both create a tunnel tract through subcutaneous tissue and seal the end of the catheter. This allows the external portion of the Covidien™ peritoneal dialysis catheter to be temporarily buried in advance of dialysis for patients that are candidates for delayed peritoneal dialysis treatment.

    Device Description

    The Ritus™ tunneling tool kit comes with a separate tunneling device and plug. This kit is designed to create a subcutaneous tunnel to temporarily bury the external portion of the peritoneal dialysis catheter within the abdominal, subcutaneous tissue in advance of needing dialysis.

    The tunneling device is made of stainless steel with a rigid plastic handle on the proximal end. There is an angled tip with a removable plastic cap. Functioning as a dilator, the tunneling device separates the tissue to create a tunnel tract, then is used to pull the catheter through the tunnel to desired exit site.

    The titanium catheter plug is designed to seal the end of the catheter for the length of time that the catheter is buried within the subcutaneous tissue.

    AI/ML Overview

    This is not an AI/ML device, therefore, the requested information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211561

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-05

    (199 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INVOS™ Patient Monitor, model PM7100, is a noninvasive cerebral/somatic oximetry system intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor. It is intended for use on individuals > 2.5 kg at risk for reduced-flow or no-flow ischemic states. For patients ≤ 2.5kg, the INVOS™ Patient Monitor, model PM7100 is only intended for adjunct trend monitoring of regional hemoglobin oxygen saturation of blood tissue beneath the sensor.

    It is also intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood tissue beneath the sensor in any individual.

    The INVOS™ Pediatric rSO2 sensor is indicated for single patient use when cerebral/somatic monitoring of site-specific regional oxygen saturation (rSO2) is required in patients weighing greater than 4 kg and less than 40 kilograms. This sensor is only intended to be used with INVOS™ PM7100 system with INVOS™ Near Infrared Spectroscopy (NIRS) technology. For additional information regarding setup and use of the INVOS™ PM7100 System including indications for use, contraindications, warnings and cautions, consult the Monitoring System Operator's Manual.

    The INVOS™ OxyAlert NIRSensor disposable sensor Model IS is indicated for single patient use when cerebrallsomatic monitoring of site-specific regional oxygen saturation (rSO2) is required in patients weighing > 2.5 kilograms and < 40 kilograms or for trend monitoring of rSO2 of blood tissue beneath the sensor in any individual < 40kg. This sensor is only intended to be used with INVOS™ Near Infrared Spectroscopy (NIRS) technology including monitoring systems and devices integrated with INVOS™ NIRS technology. For additional information regarding setup and use of the INVOS™ System including indications for use, contraindications, warnings and cautions, consult the Monitoring System Operator's Manual.

    Device Description

    The INVOS™ PM7100 Patient Monitor is a cerebral/somatic tissue oximeter intended for use as an adjunct trend monitor of regional hemoglobin and oxygen saturation monitoring. The monitor utilizes a near infrared diffuse reflectance spectroscopy system employing near infrared light at four wavelengths for the adult and pediatric system configurations. One pair of wavelengths is used to estimate the percentage of hemoqlobin saturated with oxygen in tissue beneath the sensor; another pair of wavelengths is used for the sensor on/off detection algorithm. The infant system configuration currently only employs the two wavelengths needed to estimate regional oxygenated hemoglobin.

    The subject device is non-sterile and consists of a multi-channel touch screen display, preamplifier, cables, and three single use sensor types for use in the adult (PMSENS71-A; cleared in K182868), pediatric (PMSENS71-P; new to subject device), and infant (IS; new to subject device) populations.

    The subject device utilizes up to four detachable sensors to collect signals, and up to two preamplifiers receive signals from the sensors, digitize the signals, process the data and then periodically estimate the rSQ2 at each sensor site. The preamplifiers then transmit the measured and calculated parameter data to the monitor where the information is displayed. The oximeter is powered primarily by AC power at 100 VAC to 240 VAC ±10% and is equipped with an internal rechargeable lithium-ion battery for intra hospital transport and back-up purposes. The INVOS™ PM7100 Patient Monitor is intended for use in hospitals, and is not intended for home use or out-of-hospital transport.

    The PM7100 Monitor configures the PMPAMP71 preamplifier modules for monitoring and allows the user to configure sensor placements on a patient's body as well as establish baseline rSQ2 values. Device features include a user configurable rSO2 baseline, alarms, signal strength indicator and area under the curve thresholds. The monitor measures and displays an rSO2 trend line in a graph for the estimated regional oxygen saturation value unique to the specific area under each sensor. the baseline rSQ2 value, the current estimated rSQztrend accuracy value and percent change from patient rSO2 baseline. The PM7100 Monitor also displays alarm information and indicates connected sensor type. The Monitor is equipped with technical (system status) and physiological (patient status) alarms. Alarm conditions are detected via the sensor, the physiological and technical information are then processed in the preamplifier/processor, which then communicates this information to the monitor then provides a visual and audio alarm notification. The device permits the user to silence alarms, mark events, and manage case history data.

    The PMPAMP71 preamplifier interfaces with the PM7100 monitor via a cable for communications and power, and with one or two sensors via reusable sensor cables to receive optical signals. The optical signal flows from the sensor into the PMPAMP71 which in turn generates saturation (rSO2) for tissue under the sensor that is communicated to the PM7100 monitor.

    The INVOS™ Adult SpO2Sensor, PMSENS71-A, is a non-sterile, non-invasive, disposable sensor intended for application on cerebral and somatic sites. The PMSENS71-A was designed for use with the INVOS™ PM7100 and 5100C cerebral/somatic monitoring systems for monitoring of site-specific regional oxygen saturation (rSO2) in adult patients weighing >40 kilograms. There have been no significant changes to this sensor since clearance under K182868.

    The INVOS™ Pediatric rSO2 Sensor, PMSENS71-P, is a new non-sterile, non-invasive, disposable sensor intended for application on cerebral and somatic sites in pediatric patients greater than 4 kg and less than 40 kilograms. The PMSENS71-P was designed to support the existing two wavelength rSO2algorithm along with a new two wavelength sensor on/off detection algorithm. Accordingly, the PMSENS71-P sensor is designed to emit and collect sensor data with a total of four wavelengths.

    The INVOS™ Infant Regional Saturation Sensor, model IS, is a non-sterile, non-invasive, disposable sensor intended for application on cerebral and somatic sites in the infant and neonate patient population. The IS sensor configuration currently only employs the two wavelength m. Accordingly, the IS sensor is designed to emit and collect sensor data with a total of two wavelengths. The INVOS™ Reusable Infant Sensor Adapter Cable (PMAC71RIC) is a new non-sterile, non-invasive reusable cable intended for adapting the infant sensor connection to the PM7100 preamplifier.

    The INVOS™ Docking Station connects to the INVOS™ PM7100 Patient Monitor and is designed with a mounting surface on the back of the docking station which can connect to the Patient Monitor Stand or any other compatible mounting service. The docking station is equipped with inputs for AC power, USB, serial port and VGA port. The docking station also facilitates recharging of the back-up battery. There have been no significant changes to the docking station since clearance in K182868.

    The INVOS™ Patient Monitor Stand is an accessory that connects to the INVOS™ Docking Station. The monitor stand has not changed since clearance in K 182868.

    AI/ML Overview

    The document provided describes the acceptance criteria and the study that proves the device meets the acceptance criteria for the INVOS™ PM7100 Patient Monitor, INVOS™ Pediatric rSO2 Sensor, and INVOS™ Infant Regional Saturation Sensor.

    Here's the breakdown of the information requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    Device: INVOS™ PM7100 System (with Infant/Neonatal IS sensor)

    Acceptance CriteriaReported Device Performance
    rSO2 trend precision ≤ 2.9% standard deviation of the rSO2 trend error2.09% rSO2 for cerebral
    1.96% rSO2 for somatic

    Device: INVOS™ PM7100 System (with Pediatric rSO2 sensor, PMSENS71-P)

    Acceptance CriteriaReported Device Performance
    Mean bias of rSO2 measurements (compared to predicate SPFB sensor with INVOS™ 5100C) does not exceed ±4% rSO2Mean bias = ±0.466% rSO2
    Successfully detecting sensor removal within 1 minute100%
    Successfully detecting sensor application within 15 seconds100%
    Successfully detecting the sensor is off the subject96.43%
    Successfully detecting the sensor is on the subject100%

    2. Sample Size Used for the Test Set and Data Provenance

    Infant/Neonatal Sensor Study (Hypoxia Study):

    • Sample Size: 24 subjects
    • Data Provenance: Not explicitly stated regarding country of origin, but described as a non-invasive hypoxia study conducted on healthy, non-smoking adults and adolescent volunteers. Given the FDA submission context, it's highly likely to be conducted in the US. This was a prospective clinical study.

    Pediatric Sensor Feasibility Study (Room Air Study):

    • Sample Size: 28 subjects
    • Data Provenance: Not explicitly stated regarding country of origin. This was a prospective clinical feasibility study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly mention the use of experts to establish a "ground truth" in the traditional sense of image annotation or disease diagnosis for the clinical studies.

    • For the Infant/Neonatal sensor study, the "ground truth" or reference for the rSO2 trend precision appears to be the cleared INVOS™ 5100C system in conjunction with the Infant/Neonatal Sensor (predicate system).
    • For the Pediatric sensor feasibility study, the "ground truth" or reference for rSO2 performance was the predicate INVOS™ 5100C system in conjunction with the Pediatric sensor, SPFB. Sensor on/off detection accuracy was evaluated against observed physical states.

    Therefore, this aspect of ground truth establishment by human experts, as typically seen in AI/ML performance studies, is not applicable in the context of this device's testing.


    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are typically used for establishing ground truth in studies involving human interpretation (e.g., radiology reads). As the studies described here involve direct physiological measurements and comparisons to a predicate device, or assessment of a sensor's physical state detection, an adjudication method in this sense is not applicable. The "ground truth" was established by the measurements from the predicate device or the direct observation of sensor states.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improving with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. The device is an oximeter, and the studies focused on its measurement accuracy and trend precision compared to a predicate device, and the functionality of new features like sensor on/off detection. It is not an AI/ML-based diagnostic device that assists human readers.


    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, in essence, the "Performance Data Summary" describes standalone performance for the specific algorithms and hardware of the device.

    • Non-clinical Performance Testing: This section details bench-top optical phantom testing to verify static rSO2 calculation, confirming the subject device's rSO2 algorithm performs within acceptable limits compared to the predicate system. This is a standalone performance test.
    • Clinical Studies: The clinical studies evaluated the direct output of the device (rSO2 measurements, trend precision, sensor on/off detection) against a reference (predicate device or physical observation), without direct human-in-the-loop interpretation impacting the device's output. While a human uses the device, the performance metrics are about the device's intrinsic measurement capabilities.

    7. The Type of Ground Truth Used

    • Infant/Neonatal Sensor Study: Comparative "ground truth" was established based on measurements from the cleared INVOS™ 5100C system in conjunction with the Infant/Neonatal Sensor (predicate system) during induced hypoxia. This is a form of comparative reference device data.
    • Pediatric Sensor Feasibility Study (rSO2 performance): Comparative "ground truth" was established based on measurements from the cleared INVOS™ 5100C system in conjunction with the Pediatric sensor, SPFB (predicate system) at room air. This is also a form of comparative reference device data.
    • Pediatric Sensor Feasibility Study (sensor on/off detection): The ground truth was based on direct observation of the sensor's physical state (on subject, off subject, face up/down on sheet, hanging).

    8. The Sample Size for the Training Set

    The document does not describe the use of a "training set" in the context of machine learning. The studies described are verification and validation studies (test sets) for the device's performance, not for training a model. The algorithms are based on Near Infrared Spectroscopy (NIRS) technology, not on machine learning that requires a specific training phase with a dedicated dataset.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no machine learning training set is described or utilized in the regulatory submission for this device. The device's operation is based on established NIRS principles and algorithms, rather than data-driven machine learning models requiring large labeled training sets.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212301

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-11-15

    (115 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sonicision™ 7 curved jaw cordless ultrasonic dissection device is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The device can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, and other open and endoscopic procedures. The device can be used to coagulate isolated vessels up to and including 7 mm in diameter, using the minimum mode. The device can be used to coagulate isolated vessels up to and including 5 mm in diameter, using the maximum mode.

    The Sonicision™ 7 13 cm device is also indicated for use in otorhinolaryngologic (ENT) procedures.

    The Sonicision™ reusable battery pack is a non-sterile, rechargeable, lithium-ion battery that provides power to compatible surgical devices. For full device indications for use, reference the indications for use statements of the compatible devices.

    Device Description

    A functional Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device is assembled by the user within the operating room. The functional device is assembled from the following three components: (1) a sterile, single-use dissector, (2) a user cleaned and sterilized reusable generator, and (3) a user cleaned and disinfected reusable battery. There are four different lengths of sterile single-use dissectors to allow surgeons a range of options to meet the needs of each specific surgical case.

    The assembled functional device is a hand-held, battery-powered device used to dissect through tissue and to coagulate vessels up to and including 7 mm in diameter. The clinical intended use is achieved by the surgeon when pressure is applied to tissue placed between the clamping jaw and the exposed portion of the probe while activating ultrasonic energy by pressing a button on the handpiece.

    AI/ML Overview

    The provided FDA 510(k) summary for the Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device does not contain information about acceptance criteria for AI/ML performance, nor does it describe a study specifically proving the device meets AI/ML acceptance criteria. This document describes a medical device, not an AI/ML algorithm.

    However, I can extract the performance data and comparisons that were provided for this medical device to demonstrate its substantial equivalence to predicate devices. This will address the general request for acceptance criteria and the study that proves the device meets them, but it will be in the context of a medical device, not an AI/ML algorithm.


    Acceptance Criteria and Device Performance for Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device

    The submission demonstrates substantial equivalence to predicate devices through various performance tests. The "acceptance criteria" here are implied by achieving comparable performance to the predicate devices or meeting recognized standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicImplied Acceptance Criterion (Comparable to Predicates / Standard Met)Reported Device Performance (Sonicision™ 7)
    BiocompatibilityNo new materials introduced compared to the primary predicate (K180149), thus demonstrating similar biocompatibility.Evidence presented in previously submitted primary predicate 510(k) [K180149]. The proposed device did not introduce any new materials, therefore, by inference, it met this criteria.
    SoftwareCompliance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005) and IEC 62304:2015.Software tested in accordance with the specified guidance. The software development process met the requirements of IEC 62304:2015. (Note: Software was modified to allow 7 mm coagulation using minimum mode, compared to the primary predicate's 5 mm coagulation. This implies the software changes met the safety and effectiveness criteria for the expanded capability.)
    Electromagnetic Compatibility (EMC)Meeting applicable requirements of IEC 60601-1-2, 4th edition.The assembled device met the applicable requirements of IEC 60601-1-2, 4th edition.
    Electrical SafetyMeeting applicable requirements of IEC 60601-1:2005/A1:2012, Edition 3.1.The device met all applicable requirements of IEC 60601-1:2005/A1:2012, Edition 3.1.
    Ex-vivo Burst Strength (Vessel Coagulation)Comparable burst strength to the same type of blood vessels coagulated by predicate devices. This criterion is crucial for demonstrating effective hemostasis.Ex-vivo burst testing showed that blood vessels coagulated by the Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device had comparable burst strength to the same type of blood vessels coagulated by the predicates. (Specifically, the device is able to coagulate vessels up to and including 7 mm, an improvement over the primary predicate's 5 mm).
    Ex-vivo Tissue Testing (Thermal Effects)Maximum temperatures and cool down times of the active blade and shaft comparable to predicate devices after multiple activations on mesentery, implying acceptable thermal injury control.Ex-vivo tissue testing showed that the maximum temperatures and cool down times of the Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device's active blade and shaft were comparable to the maximum temperatures and cool down times of the predicate's active blade and shaft after multiple activations on mesentery.
    Acute In-vivo Testing (Hemostasis & Lateral Thermal Spread)Comparable rates of hemostasis and comparable lateral thermal spread to predicate devices, indicating equivalent immediate surgical performance.Acute in-vivo testing showed that the Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device and the predicates achieved comparable rates of hemostasis and comparable lateral thermal spread.
    Chronic In-vivo Testing (Sustained Hemostasis)Maintenance of hemostasis for at least 21 days for vessels up to and including 7 mm in diameter. This confirms long-term efficacy of vessel sealing.Chronic in-vivo testing showed that vessels (up to and including 7 mm in diameter) coagulated by the Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device maintain hemostasis for at least 21 days.
    Human FactorsMeeting user needs and demonstrated ability for intended users to use the device safely and effectively in intended use environments.Human factors activities demonstrated that the Sonicision™ 7 Curved Jaw Cordless Ultrasonic Dissection Device meets user needs and is able to be used for the intended users, intended use, and use environments.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the numerical sample sizes for each of the performance tests (e.g., how many vessels were tested in ex-vivo burst testing, how many animals in acute/chronic in-vivo studies). It only mentions "several bench and animal tests."

    The data provenance (country of origin, retrospective/prospective) is not specified in this summary. These types of tests are typically prospective studies conducted by the device manufacturer or contracted labs.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This device is a surgical instrument, not an AI/ML diagnostic tool requiring expert interpretation of images for ground truth. Therefore, the concept of "experts establishing ground truth" in the diagnostic AI sense does not apply here. The "ground truth" or reference standards are established through direct measurements (e.g., burst pressure), histological analysis, and observed clinical outcomes in animal models. The experts involved would be qualified biomedical engineers, veterinarians, and possibly surgeons overseeing the animal studies, but their specific number and qualifications are not detailed.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is a medical device performance study, not an AI/ML algorithm evaluation with human readers.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device performance study, not an AI/ML algorithm evaluation with human readers.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device performance study, not an AI/ML algorithm evaluation. While the device has "control software," its performance is evaluated in conjunction with the mechanical and ultrasonic components, as a complete system operated by a human surgeon.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is established through:

    • Direct physical measurements: Ex-vivo burst pressure, temperature measurements.
    • Biological observations: Visual assessment of hemostasis, measurement of lateral thermal spread.
    • Pathology/Histology: Although not explicitly stated, chronic in-vivo studies often involve histological examination of tissue to confirm vessel sealing and assess tissue damage.
    • Outcomes data: For the chronic in-vivo study, "maintaining hemostasis for at least 21 days" refers to an outcome measurement.

    8. The sample size for the training set

    Not applicable. This is a medical device and not an AI/ML algorithm. There is no "training set" in the context of machine learning. The device development involved engineering design, prototyping, and iterative testing, but not machine learning training.


    9. How the ground truth for the training set was established

    Not applicable. As there is no AI/ML training set, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203303

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-17

    (38 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

    The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures.

    The system's optional 3D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted trajectory on a computer monitor. The size and shape of the predicted zone relative to the position of the antenna are displayed on the enhanced ultrasound image.

    Device Description

    The Emprint™ SX Ablation Platform with Thermosphere™ Technology is microwave-based system intended to deliver energy through an antenna inserted into soft tissue for the purpose of coagulating (ablation) a defined tissue volume. The Emprint™ SX Ablation Platform utilizes a 2450 MHz 100W generator to deliver power to a single microwave ablation antenna. The platform is software-controlled and a touchscreen Graphical User Interface is used to select the desired ablation time (up to 10 minutes) and power (5 to 100W) settings. Using an optional temperature probe, the ablation platform can be set to monitor the temperature of a desired target and to automatically turn the generator off, when the target reaches a pre-set temperature. The ablation platform uses circulating, room-temperature, normal saline to cool the non-radiating portion of the antenna shaft and to provide a consistent ablation zone. The normal saline is pumped from an IV bag through the antenna shaft and back to the IV bag in a closed system.

    Using an optional/selectable navigation feature, the Emprint™ SX Ablation Platform assists physicians in the accurate placement of the trackable Emprint™ SX Navigation Antenna into a target anatomical structure by overlaying the image of the antenna and its trajectory onto an ultrasound image in real time.

    Sensors are attached to platform-compatible, open and laparoscopic ultrasound probes and to the platform's trackable antenna. These sensors detect fluctuations in an electromagnetic field emitted from the system's Table Top Field Generator (TTFG). The TTFG is placed underneath the patient and transmits positional information that is processed by the system's navigation hardware and interpreted by its software. The platform's software allows the surgeon to see a representation of the ablation zone on an auxiliary display and the system's operator to control the ablation system via the touchscreen interface.

    AI/ML Overview

    This document is a 510(k) summary for a software update (Emprint™ SX 1.1.1 Software Update) to an existing device, the Emprint™ SX Ablation Platform with Thermosphere™ Technology. It focuses on demonstrating substantial equivalence to the predicate device, which is the previous version of the same system. Therefore, the document primarily addresses the software change and its impact rather than the original device's full acceptance criteria and clinical efficacy.

    Since this submission is a software update to an already cleared device and not an initial clearance for a new device type, there is no detailed acceptance criteria table, clinical study data, or ground truth establishment presented in sections 1-7 of your request. The submission argues that the software update does not affect the device's fundamental performance, indications for use, or underlying principles of operation.

    Here's a breakdown based on the provided text, addressing as many of your points as possible:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Acceptance Criteria: Not explicitly stated in terms of quantitative clinical performance metrics for the device as a whole in this software update submission. The acceptance criteria for this specific submission relate to the successful implementation and verification of the software change without introducing new safety or effectiveness concerns.
    • Reported Device Performance: The document states that "Comprehensive SW Verification testing was performed by Medtronic in support of the proposed change." The conclusion is that "The results of comprehensive verification testing raised no new questions of safety or effectiveness." This implies the software update met internal verification criteria to ensure it did not negatively impact the previously established performance.
    Acceptance Criterion (for software update)Reported Device Performance
    Error handling allows clearing general system errorsUser can clear general system error without power-cycling the system.
    No adverse effect on ablation therapy deliveryMitigates risk of delay in procedure due to error.
    No effect on other software architecture componentsVerified that changes affect only Emprint API code.
    No effect on system's underlying technological characteristicsVerified.
    No effect on intended use, essential function, or performanceVerified.
    No new questions of safety or effectiveness raisedComprehensive SW Verification testing led to this conclusion.

    2. Sample size used for the test set and the data provenance:

    • Test Set: Not specified for this software update. "Comprehensive SW Verification testing" was performed, but details on the number of test cases or scenarios are not provided in this summary.
    • Data Provenance: The testing was performed by Medtronic, the manufacturer. The document doesn't specify if external data was used for this software update's verification. It's internal testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for this software update submission. The "ground truth" for a software verification typically involves predefined expected software behaviors and outputs, which are assessed by software engineers and quality assurance personnel. It doesn't involve clinical experts in the same way a diagnostic device's clinical performance study would.

    4. Adjudication method for the test set:

    • Not applicable/Not specified in the provided summary. Software verification typically follows defined test protocols and review processes by the development and quality assurance teams, rather than a clinical adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This document describes a software update to a microwave ablation platform, not an AI-assisted diagnostic or interpretative device that would typically involve MRMC studies. The device's navigation feature does assist in antenna placement, but the current submission is about a software bug fix related to error handling, not an enhancement to the navigation's clinical efficacy or human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, in essence. The "software verification testing" would assess the algorithm's (software's) performance in isolation, ensuring it functions as intended (e.g., clearing errors, processing messages correctly) without human intervention in the execution of the software's core logic. However, since this is a system that always involves human-in-the-loop operation (a surgeon using it for ablation), the "standalone" here refers to the software component's functional correctness.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For this software update, the ground truth would be the expected and correct behavior of the software based on its design specifications. This is established during the software development lifecycle through requirements definition, design documents, and coding standards, and is then verified through testing. It's not clinical "ground truth" like pathology for a diagnostic device.

    8. The sample size for the training set:

    • Not applicable. This submission is not for a machine learning or AI algorithm that requires a training set. It's for a traditional software update that fixes an error handling behavior.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of software update.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193056

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-20

    (385 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Puritan Bennett 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.66lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively to patients who require the following types of ventilator support

    • Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trachectorny tube) or non-invasively (via mask or nasal prongs)

    • Assist/Control, SIMV, or Spontaneous modes of ventilation

    Device Description

    The Puritan Bennett 980 Series Ventilator System is a dual-microprocessor-based, touch-screen controlled; critical care ventilator intended to provide continuous ventilation for neonate to adult patients who require either invasive ventilation or non-invasive ventilation. It can be used in hospitals and institutions and for intra -hospital transport applications with access to the appropriate services.

    The ventilator system offers features for patient comfort while delivering sensitive, precise breaths to critically ill patients. The product ventilates Neonatal, Pediatric, and Adult patients with predicted body weights from 0.3 kg, and with tidal volumes for mandatory volume-controlled breaths from 2 mL to 2500 mL.

    The following options are being made to the Subject Device:

    • Initial release -
      • Integrated Nebulizer option ●
      • High Flow Oxygen Therapy (HFO2T) option ●
      • NIV+ Software option ●
      • New Monitored Parameter Porive ●
    • Modification to previously cleared feature -
      • Updated IE Sync Algorithm
    AI/ML Overview

    This document is an FDA 510(k) Premarket Notification for the Puritan Bennett 980 Series Ventilator System. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria in the context of an AI/ML medical device.

    Therefore, none of the requested information regarding acceptance criteria, device performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, ground truth types, or training set details can be extracted from the provided text. The document states that "Clinical evidence was not necessary to show substantial equivalence," and the non-clinical evidence focuses on verification and validation against standards and software/system tests, not a performance study as would be expected for a new or modified algorithm with performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200427

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-07-10

    (140 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sonicision™ cordless ultrasonic dissection device is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The device can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures. The Sonicision™ cordless ultrasonic dissection device can be used to coagulate isolated vessels up to 5 mm in diameter.

    The Sonicision™ 13 cm device is also indicated for use in otorhinolaryngologic (ENT) procedures.

    Device Description

    The Sonicision™ Cordless Ultrasonic Dissector is a component of the Sonicision™ Cordless Ultrasonic Dissection Device, which is a hand-held surgical device consisting of three interdependent components that. when assembled, enable ultra-high-frequency mechanical motion (ultrasonic energy) to transect, dissect, and coagulate tissue.

    The dissector is a sterile, single-use component to which the Sonicision reusable generators and battery packs attach. This component contains features essential to the control and performance of the assembled device; such as the clamping jaw, active blade, speaker, two-stage energy button, rotation wheel, and jaw lever.

    Four configurations are available, differing principally by shaft lengths are 13 cm, 26 cm, 39 cm, and 48 cm; corresponding with catalog numbers SCD13, SCD26, SCD396, and SCD48, respectively. The only difference between SCD391 and SCD396 is the packaging configuration. There is no difference in the design of the device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Sonicision™ Cordless Ultrasonic Dissector, a surgical device. While it details performance testing, it does not explicitly state specific "acceptance criteria" numerical values that the device had to meet. Instead, the performance studies aim to demonstrate comparability or non-inferiority to the predicate device.

    Therefore, I cannot populate a table with explicit acceptance criteria and corresponding reported performance numerical values as they are not presented in that format in the document. The document describes the type of performance tests conducted and the finding of comparability.

    However, I can extract and present the information related to the studies in the requested format, interpreting "acceptance criteria" as the goal of the test (e.g., comparable burst strength) and "reported device performance" as the outcome in relation to the predicate.

    Here's the breakdown of the information based on the provided text:


    Acceptance Criteria and Device Performance

    As explicit numerical acceptance criteria are not provided, the "acceptance criteria" are inferred as demonstrating comparability or non-inferiority to the predicate device for various performance aspects.

    Acceptance Criteria (Inferred Goal)Reported Device Performance (Study Outcome)
    Ex-vivo burst strength of coagulated blood vessels comparable to predicate.Blood vessels coagulated by the Sonicision™ Cordless Ultrasonic Dissector had comparable burst strength to the same type of blood vessels coagulated by the predicate.
    Maximum temperatures and cool-down times of active blade and shaft comparable to predicate after multiple activations on mesentery.The device's maximum temperatures and cool down times of the active blade and shaft were comparable to the predicate's active blade and shaft after multiple activations on mesentery.
    Rates of hemostasis and lateral thermal spread comparable to predicate in acute in-vivo testing.The Sonicision™ Cordless Ultrasonic Dissector and the predicate achieved comparable rates of hemostasis and comparable lateral thermal spread in acute in-vivo testing.
    Vessel hemostasis retention (up to 5 mm) for at least 21 days for coagulated vessels.Vessels (up to 5 mm in diameter) coagulated by the Sonicision™ Cordless Ultrasonic Dissector maintain hemostasis for at least 21 days (demonstrated in chronic in-vivo testing).
    Meets user needs and FDA expectations (Human Factors Validation).Human factors validation in a porcine model and a human cadaver model demonstrated that the Sonicision™ Cordless Ultrasonic Dissector meets user needs and FDA expectations. (This implies successful completion of simulated use tasks without critical errors).

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Ex-vivo Burst Testing: Specific sample size not provided, but performed on "blood vessels."
      • Ex-vivo Tissue Testing: Specific sample size not provided, but performed on "mesentery."
      • Acute In-vivo Testing: Specific sample size not provided.
      • Chronic In-vivo Testing: Specific sample size not provided.
      • Human Factors Validation: Performed in a "porcine model and a human cadaver model." Specific number of participants/cadavers not provided.
      • Data Provenance: The studies are described as "bench and animal tests" and human cadaver models, implying controlled laboratory or clinical simulation environments. The country of origin is not specified but is presumably where Covidien (Boulder, Colorado) conducts its R&D or contracted research. The studies are prospective in nature, designed specifically to test the device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number or qualifications of experts involved in establishing ground truth for any of these performance tests. For animal and cadaver studies, typically veterinarians, pathologists, and surgeons/anatomists would be involved, but this information is not provided.
    3. Adjudication method for the test set:

      • The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the outcomes of these performance tests. The results seem to be presented as direct measurements or observations.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study involving human "readers" or AI assistance was not done. This device is a surgical instrument, not an imaging diagnostic device that would typically involve human "readers" or AI for image interpretation. The human factors validation is about user interaction with the device, not diagnostic interpretation.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This is not applicable as the device is a physical surgical instrument and does not involve an AI algorithm with standalone performance. Its performance is always integrated with human use.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Burst Strength: Likely measured mechanically (physics-based ground truth).
      • Temperatures/Cool Down Times: Measured using temperature sensors (physics-based ground truth).
      • Hemostasis/Lateral Thermal Spread: Likely assessed visually, histologically (pathology-based), or through direct measurement during animal studies (outcomes/observational ground truth).
      • Hemostasis Retention: Assessed through long-term observation in chronic animal studies (outcomes data).
      • Human Factors Validation: Established by objective task completion metrics (e.g., successful dissection, coagulation, no user errors) and subjective user feedback, compared against pre-defined user needs and FDA expectations for usability.
    7. The sample size for the training set:

      • This concept (training set) is not applicable as the device is a physical surgical instrument, not a machine learning/AI model. Therefore, there is no "training set" in the conventional sense. The development of the device would involve engineering, design, and iterative testing, but not data-driven "training" like an AI.
    8. How the ground truth for the training set was established:

      • As there is no "training set" for an AI model, this question is not applicable to this device submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2