510(k) Data Aggregation

    K Number
    DEN200055

    Validate with FDA (Live)

    Device Name
    GI Genius
    Date Cleared
    2021-04-09

    (213 days)

    Product Code
    Regulation Number
    876.1520
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.

    Device Description

    The GI GENIUS™ is an artificial intelligence/machine learning (AI/ML) device system comprised of software, hardware, and accessories that is intended for polyp detection during standard white-light colonoscopy. The device system generates a video on the main endoscopy display that contains the original live video together with superimposed markers (in the form of green boxes) that appear when a lesion is detected.

    The GI Genius takes the Serial Digital Interface (SDI) output stream from the video endoscope processor as an input and then generates an SDI output stream to the existing monitor/display system containing the original video stream with additional markers superimposed on it. In essence, the system is inserted into the video stream just prior to it being displayed to the user/operator.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the GI Genius device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    I. Nonclinical/Bench Studies
    Video DelayTime $\leq$ 5.75 milliseconds(b)(4) microseconds ((b)(4)%)
    Annotation DelayTime $\leq$ 120 milliseconds(b)(4) or (b)(4) milliseconds ((b)(4) frames)
    Video Quality IntegrityNo degradation in image quality (identical pixels in 3 color channels, excluding marker overlays)Test data show no pixel-level discrepancies except for pixels overlaid with markers
    II. Standalone Performance
    Activation TimeDevice detects lesion faster than endoscopist reaction timeGI Genius detected polyp 1270 ms (95% CI: 857 ms, 1684 ms) before average endoscopist
    Object-Level Sensitivity (persistence > 0ms)No specific threshold given; presented as trade-off with FP rate. The device detects polyps before endoscopist detection time.81.96% (95% CI: 77.35%; 85.97%)
    Object-Level False Positives (persistence > 0ms)No specific threshold given; presented as trade-off with sensitivity.156.31 (95% CI: 135.61; 177.00) FP Objects/Patient
    Frame-Level False Positive Rate (FPR / FRAME)Expected: 4.85% or lower (based on estimation from video database)Logistic Regression Mixed Model: 1.44% (95% CI: 1.27%; 1.63%) Non-parametric Cluster Bootstrap: 2.02% (95% CI: 1.72%; 2.35%)
    Frame-Level True Positive Rate (TPR / FRAME)No specific threshold given; presented.Logistic Regression Mixed Model: 47.46% (95% CI: 42.51%; 52.45%) Non-parametric Cluster Bootstrap: 49.57% (95% CI: 45.24%; 54.06%)
    Standalone Performance (Overall)Sufficient to fulfill indications for use; adequate benchmark for improved lesion detection.Met pre-defined performance criteria and found adequate for benchmarking.
    III. Clinical Performance (AID Study)
    Primary Endpoint: Adenoma Detection Rate* (ADR*)Non-inferiority (10% margin) to standard colonoscopy; then superiority if non-inferiority met.Superiority Demonstrated: GI Genius ADR* 55.1% (95% CI: 44.0% to 65.8%) vs. Standard 42.0% (95% CI: 31.3% to 53.4%). Difference 13.1% (95% CI: 0.09; 23.3), p < 0.05.
    Secondary Endpoint: Adenomas per Colonoscopy* (APC*)Superiority to standard colonoscopy.Superiority Demonstrated: GI Genius APC* 0.81 (95% CI: 0.57 to 1.15) vs. Standard 0.57 (95% CI: 0.39 to 0.82). Ratio 1.43 (95% CI: 1.03; 1.98), p = 0.03.
    Secondary Endpoint: Positive Percent Agreement (PPA)Non-inferiority (15% margin) to standard colonoscopy.Non-inferiority Demonstrated: GI Genius PPA 62.1% (95% CI: 43.4% to 77.8%) vs. Standard 65.2% (95% CI: 46.0% to 80.4%). Difference -3.1% (95% CI: -14.3; 4.8), p < 0.05.
    IV. Usability
    Semi-quantitative QuestionsThreshold score for semi-quantitative questions > 3.Average scores ranged from 3.87 to 4.73.
    Qualitative QuestionsNo usability errors; no changes to colonoscopy workflow.No usability errors reported; no changes to workflow.
    V. EMC & Electrical SafetyPasses acceptance criteria of ANSI/AAMI/IEC 60601-1-2:2014 and IEC 60601-1:2005 + A1:2012 (Ed. 3.1).Test results pass acceptance criteria.
    VI. Software/CybersecurityIdentified as moderate level of concern; submitted all required documentation (hazard analysis, V&V, threat model, etc.).Documentation included. Acceptable verification and validation activities at unit, integration, and system level. Cybersecurity documentation complete.

    Study Details

    2. Sample Sizes and Data Provenance

    • Test Set (Standalone Performance):
      • Sample Size: 150 colonoscopy videos.
        • 105 videos included 338 excised polyps with histology confirmation.
        • 45 videos did not include polyps or lesions.
      • Data Provenance: Originally from a study titled, "The Safety and Efficacy of Methylene Blue MMX® Modified Release Tablets Administered to Subjects Undergoing Screening or Surveillance Colonoscopy" [NCT01694966]. This study was a multi-arm study conducted without specific country information, but the clinical study (AID study) was performed in Italy. Given the standalone performance data was derived from the MMX trial, it is retrospective. The 150 videos for the Holdout Test Set were specifically without methylene blue.
    • Clinical Study (AID Study):
      • Sample Size (mITT Population): 263 patients.
        • 136 patients randomized to GI Genius+colonoscopy.
        • 127 patients randomized to standard colonoscopy.
      • Data Provenance: Randomized, prospective, multicenter, controlled clinical investigation performed in Italy (Humanitas Research Hospital [Milan], Nuovo Regina Margherita Hospital [Rome], and Valduce Hospital [Como]).

    3. Number of Experts and Qualifications for Test Set Ground Truth

    • Standalone Performance Test Set:
      • Experts: A panel of five expert endoscopists reviewed video clips to establish the critical time frame for lesion detection and to define the endoscopists' first detection time for object-level performance.
      • Qualifications: Not explicitly stated beyond "expert endoscopists."
    • Clinical Study Test Set (AID Study):
      • Experts: Six endoscopists were involved in conducting the colonoscopies and establishing ground truth through biopsy and histological confirmation.
      • Qualifications: Endoscopists with moderate endoscopy expertise, defined as an Adenoma Detection Rate (ADR) between 25% to 40%.

    4. Adjudication Method for Test Set

    • Standalone Performance Test Set:
      • The reference standard for true positives, true negatives, false positives, and false negatives was established by having endoscopists review video clips around histologically confirmed polyps and placing an annotation box around the polyps visible in each frame. The device's markers were then assessed for overlap with these annotations using an Intersection over Union (IoU) criterion. This implies expert consensus/annotation was the primary method, followed by a quantitative comparison. No explicit adjudication for disagreements among experts is described, rather, the experts created the ground truth.
    • Clinical Study Test Set:
      • Ground truth for polyps in the clinical study was established by histological confirmation of excised lesions. The study design does not explicitly describe a multi-reader, multi-case (MRMC) adjudication process for detection, but rather the gold standard for lesion presence/type relied on pathology.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Yes, a form of MRMC comparative effectiveness study was done, in the clinical study (AID Study).
      • The study design involved six endoscopists, performing colonoscopies both with AI assistance (GI Genius) and without (standard colonoscopy). This is a direct comparison of human performance with and without AI assistance.
      • It was a randomized, prospective, multicenter, controlled clinical investigation.
      • Effect Size of Human Reader Improvement:
        • For the primary endpoint, Adenoma Detection Rate (ADR*): The GI Genius + colonoscopy group had an adjusted ADR* of 55.1%, while the standard colonoscopy group had an ADR* of 42.0%. This represents an absolute improvement of 13.1 percentage points with AI assistance (p < 0.05 for superiority).
        • For the secondary endpoint, Adenomas per Colonoscopy (APC*): The GI Genius + colonoscopy group had an adjusted APC* of 0.81, while the standard group had an APC* of 0.57. The estimated counts ratio was 1.43 (meaning 43% more adenomas/carcinomas detected per colonoscopy with AI assistance, p = 0.03).

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone performance study was done. This is explicitly detailed under "PERFORMANCE TESTING - BENCH - STANDALONE PERFORMANCE."
      • It demonstrated the algorithms' object-level and frame-level detection performance (sensitivity, false positive rate, ROC performance) without human-in-the-loop interaction for the detection phase.
      • The activation time of the device detecting a lesion was measured against the average endoscopist's detection time, showing the AI's ability to detect before a human.

    7. Type of Ground Truth Used

    • Standalone Performance Test Set:
      • Expert Consensus/Annotation: For creating bounding boxes around polyps in video frames, reflecting expert identification of lesions.
      • Histology Confirmation: For the presence of polyps, their type (adenoma/non-adenoma), and size. This provided the definitive ground truth that experts then used for annotation.
    • Clinical Study Test Set:
      • Histology Confirmation: All primary and secondary endpoints (ADR*, APC*, PPA) relied on histologically confirmed polyps, adenomas, or carcinomas as the ground truth. This is a gold standard for lesion characterization.

    8. Sample Size for Training Set

    • Training Set (for initial algorithm training):
      • Number of Videos: A subset of 1205 videos from the original MMX study.
        • Polyps + MMX: 349 videos
        • Polyps + No MMX: 219 videos
      • Number of Subjects: 568 subjects (from Table 1, "Training" column).

    9. How Ground Truth for Training Set Was Established

    • The training of the GI Genius AI software was done on a subset of videos in which polyps were identified.
    • The original study from which the videos were sourced ("The Safety and Efficacy of Methylene Blue MMX® Modified Release Tablets Administered to Subjects Undergoing Screening or Surveillance Colonoscopy" [NCT01694966]) inherently involved the process of identifying and characterizing polyps, likely through a combination of endoscopic observation and subsequent histological confirmation of excised lesions. The text states polyps were "identified, either in the presence or absence of methylene blue," implying the clinical process of polyp detection and confirmation.
    • The specific method of "labeling" or "ground-truthing" for the AI training process (e.g., whether it involved manual annotation of frames by experts for every polyp, similar to the standalone test set ground truth) is not explicitly detailed for the training set itself, but it is reasonable to infer it followed similar principles of expert-based identification and/or histological confirmation as seen in the test sets.