Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K193159

    Validate with FDA (Live)

    Date Cleared
    2020-12-30

    (411 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EEG NeuroAmp II.5s is intended to acquire, display, store and archive the electroencephalographic (EEG) signals from a patient's brain obtained by placing two or more electrodes on the head. It must be used in combination with appropriate software and computer. The EEG NeuroAmp II.5s is also intended to be used to provide stimulation and recording of visual and auditory evoked response potentials with appropriate software, computer, and accessories.

    The EEG NeuroAmp II.5s can also be used for biofeedback and relaxation purposes of this training task, information for feedback may be derived from the electroencephalographic signals and from as many as eight channels of peripheral physiology measures such as are conventionally used in biofeedback.

    The EEG NeuroAmp II.5s is intended to be used only by trained professionals who can ensure sound handling practices. The device is intended for use in clinical environments (not surgical or emergency environments) as well as in medical and therapeutic practices. It can be used with patients of all ages but is not designed for fetal use.

    The device does not draw any diagnostic conclusion. Recorded data of electroencephalogram (EEG) signals or event related potentials (ERPs) need to be interpreted by a clinical expert. The results of such interpretation must be considered only in conjunction with other clinical findings.

    Device Description

    The EEG NeuroAmp II.5s is a user-friendly high-performance interface between client and clinician computer. It can be used in two different functionalities: In an EEG/ERP measurement and in an EEG biofeedback (neurofeedback) and/or peripheral biofeedback setting. The EEG NeuroAmp 11.5s device itself contains three function blocks:

      1. EEG amplifier, two channels, interface for 23- or 39-channel EEG recording front-end
      1. Impedance meter function
      1. peripheral channels for additional sensors, ERP synchronization and biofeedback

    Apart from the EEG-biofeedback functionality, together with the x23/x39 accessories, it is possible to utilize the equipment as an EEG/ERP measurement device. A 23- or 39-channel EEG can be measured with conventional EEG caps, which are not part of the device, according to the standard 10-20 system. The device does not draw any diagnostic conclusions and does not invite the clinician to do so. Recorded data of electroencephalogram (EEG) signals need to be interpreted by a clinical expert. The results of such interpretation must be considered in conjunction with other clinical findings to provide context. There is no therapeutic effect of the EEG NeuroAmp II.5s used in this setting.

    The NeuroAmp x23/x39 devices are a full 23-/39-channel digital EEG recording front-end with integrated impedance meter. They are an accessory to the EEG NeuroAmp 11.5s and work only in conjunction with it.

    AI/ML Overview

    The Corscience EEG NeuroAmp II.5s is an electroencephalograph device that functions for EEG/ERP measurement and biofeedback. The device received 510(k) clearance based on substantial equivalence to predicate devices, demonstrating its safety and effectiveness through performance testing against relevant IEC and ISO standards.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the EEG NeuroAmp II.5s are implicitly defined by its performance against the specified technical requirements, which are shown to be equivalent or superior to predicate devices. The "Discussion on safety and effectiveness" column in the comparative table effectively serves as the acceptance criteria and performance review.

    SpecificationAcceptance Criteria (from predicate comparison)Reported Device Performance (EEG NeuroAmp II.5s)
    EEG/ERP Measurement Functionality
    Indications for Use (short version)Acquire, display, store, and archive electrical activity of a patient's brain (Mitsar EEG and Natus Brain Monitor also allow stimulation and recording of evoked response potentials).Acquire, display, store and archive the electroencephalographic (EEG) signals from a patient's brain obtained by placing two or more electrodes on the head incl. stimulation and recording of visual and auditory evoked response potentials.
    Common name, Product code, Regulation numberGWQ, 21 CFR 882.1400 (Mitsar EEG); GWQ & OLV, 21 CFR 882.1400 (Natus Brain Monitor)EEG/ERP measurement & biofeedback device, GWQ, GWE, GWJ, OLV & HCC, 21 CFR 882.1400, 21 CFR 882.5050, 21 CFR 882.1890 & 21 CFR 882.1900
    Power supplyPrimarily USB port for power, with galvanic isolation (for EEG NeuroAmp, presumed for Mitsar EEG).Power supply via USB port (galvanic isolation according to IEC 60601-1)
    SoftwareEEG Studio (Mitsar EEG); Natus NeuroWorks™/SleepWorks (Natus Brain Monitor)ERPrec
    Windows 10 and 7 compatibleWindows 7, perhaps also Windows 10 (Mitsar EEG); Yes (Natus Brain Monitor)Yes
    EEG channels21 + 4 active/reference pairs, one auxiliary channel (Mitsar EEG); 40 (Natus Brain Monitor)up to 41 plus 8 peripheral channels
    Resolution EEG channels16 bit (Mitsar EEG); 24 bit (16 bit stored) (Natus Brain Monitor)24 bit, stored loss-free 24bit and full sampling rate
    Sampling frequency500 Hz per channel, multiplexed (Mitsar EEG); 256, 512, 1024, 2048, 4096 (Natus Brain Monitor)1Msample/second gross sampling rate, down sampled to 250 or 500 sps. synchronous sampling over all EEG- and peripheral channels
    Common-Mode Rejection Ratio (CMRR)at least 100 dB at 10 Hz (Mitsar EEG); >106db@60Hz (Natus Brain Monitor)>130 dB
    Notch filter-30 dB at 50(60) Hz (Mitsar EEG); No data (Natus Brain Monitor)<-60 dB at 50 or 60 Hz,optional setting in software. Recording always unfiltered raw signal.
    Input impedance EEG channels> 200 MΩ (Mitsar EEG); >1000 MOhm (Natus Brain Monitor)>1000 MOhm
    Bandwidth (3dB) and sample rate0.16 – 70 Hz (Mitsar EEG); 0.1 Hz – 100 Hz (Natus Brain Monitor)DC ... 100Hz/160Hz
    Impedance measurement EEG electrodes (electrode contact quality)Yes, 5kΩ – 40kΩ (Mitsar EEG); Yes, 2.5kΩ, 5kΩ, 10kΩ, 25kΩ (Natus Brain Monitor)Yes, 0 ... 140kΩ
    Input Noise< 1.5 µV peak to peak (Mitsar EEG); ≤ 2 uV pk-to-pk (0.1Hz to 100 Hz) (Natus Brain Monitor)< 0.01µV/√Hz; < 1.0 µV peak toPeak
    Overvoltage warningNo data (Mitsar EEG and Natus Brain Monitor)yes
    Input signal range10 - 5000 µV (Mitsar EEG); 20mV pk-to-pk, +/- 0.3VDC (Natus Brain Monitor)1100mVp-p
    Maximum Operational DC input voltage electrode offset±350 mV (Offset tolerance) (Mitsar EEG); ±300mV (Natus Brain Monitor)±550mV
    ERP measurement: synchronization mechanism and precision in msEvent button, No active synchronization, Accuracy/jitter: ±10ms/<20ms (Mitsar EEG); Integrated 8-bit trigger, Accuracy/jitter: ±10ms/<20ms (Natus Brain Monitor)Active audio and video synchronization with Sync device; Event button, 24 bit trigger.. Accuracy/jitter: ±1ms/<1ms
    Visual and auditory stimuli for ERP measurementStimulus presentation software incorporated, Audio and video sensors for initial calibration (Mitsar EEG); No data (Natus Brain Monitor)Stimulus presentation software incorporated in software ERPrec Audio and video sensors for continuous calibration and selftest of the system initial to each recording.
    Measurement of physiological dataMulti-purpose inputs for peripheral biosignal acquisition (Mitsar EEG); SpO2 Pulse Rate, Plethysmogram, PPG (Natus Brain Monitor)Combination sensor, applied to the finger, for heart rate and galvanic skin response (GSR). Skin temperature of the finger can be measured, not absolute values, only trends. No therapeutic effect of sensor.
    Video recordingPossible (Mitsar EEG and Natus Brain Monitor)Possible
    EEG-Biofeedback (Neurofeedback) Functionality
    Indications for UseBiofeedback and relaxation purposes (EEG NeuroAmp)Biofeedback and relaxation purposes
    Common name, Product code, Regulation numberBiofeedback device, HCC, 21 CFR 882.5050 (EEG NeuroAmp)EEG/ERP measurement & biofeedback device, GWQ, GWE, GWJ, OLV & HCC, 21 CFR 882.1400, 21 CFR 882.5050, 21 CFR 882.1890 & 21 CFR 882.1900
    Power supplyPower supply via USB port (galvanic isolation according to IEC 60601-1) (EEG NeuroAmp)Power supply via USB port (galvanic isolation according to IEC 60601-1)
    SoftwareCygnet, BEE Lab (EEG NeuroAmp)Cygnet, BEE Lab
    Windows 10 and 7 compatibleYes (EEG NeuroAmp)Yes
    x23/x39 EEG amplifier interfaceNo (EEG NeuroAmp)Yes
    No. peripheral channels inputs3 (EEG NeuroAmp)8
    No. peripheral channels outputs1 (EEG NeuroAmp)8
    EEG resolution13 bit (EEG NeuroAmp)32 bit
    Input noise< 1.5 µVp-p (EEG NeuroAmp)< 0.01 µV/ √Hz ; < 1.0 µVp-p
    Overvoltage warningYes (EEG NeuroAmp)yes
    EEG bandwidth0.056 ... 70Hz (EEG NeuroAmp)DC ... 100/160Hz
    Peripheral channel resolution13 bit (EEG NeuroAmp)24 bit
    Peripheral channel bandwidthDC ... 70Hz (EEG NeuroAmp)DC ... 100/160Hz

    2. Sample size used for the test set and data provenance

    The document does not describe a clinical study with a "test set" of patient data for the device to be evaluated. The performance testing conducted was:

    • Electrical Safety: Bench testing according to IEC 60601-1.
    • Electromagnetic Compatibility: Bench testing according to IEC 60601-1-2 and IEC 60601-2-26.
    • Performance Testing (Hardware): Bench testing in accordance with internal requirements and applicable clauses of IEC 62366-1, IEC 80601-2-26, and IEC 60601-2-40.

    The accessorizes were also tested to ISO 10993-1, IEC 60601-1 and IEC 60601-1-2.

    Therefore, there is no mention of sample size or data provenance in terms of patient data.

    3. Number of experts used to establish the ground truth for the test set and their qualifications

    This information is not applicable as there was no test set of patient data requiring expert ground truth establishment. The device capabilities were verified through technical bench testing against established engineering standards and specifications.

    4. Adjudication method for the test set

    This information is not applicable as there was no test set of patient data requiring adjudication.

    5. Multi-reader multi-case (MRMC) comparative effectiveness study

    No MRMC study was performed or described. The study focused on technical equivalence and performance against predicate devices and standards, not on human reader performance with or without AI assistance. The device does not draw any diagnostic conclusion and requires interpretation by a clinical expert.

    6. Standalone (i.e. algorithm only without human-in-the-loop performance) study

    No standalone performance study of an algorithm was conducted or described. The device is a hardware system for acquiring, displaying, storing, and archiving EEG signals, with capabilities for biofeedback and evoked response potentials. It's explicitly stated that the device "does not draw any diagnostic conclusion" and "Recorded data... need to be interpreted by a clinical expert."

    7. Type of ground truth used

    The "ground truth" for evaluating the EEG NeuroAmp II.5s was based on:

    • Adherence to international electrical safety and EMC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, ISO 10993-1 for accessories).
    • Compliance with device-specific performance standards for electroencephalographs and evoked response equipment (IEC 80601-2-26, IEC 60601-2-40).
    • Comparison of technical specifications against legally marketed predicate devices (Mitsar EEG K143233, Natus Brain Monitor K180290, EEG NeuroAmp K073557) to demonstrate substantial equivalence and often superior performance.

    8. Sample size for the training set

    This information is not applicable as the device is a medical measurement instrument and not an AI/ML algorithm that requires a training set of data.

    9. How the ground truth for the training set was established

    This information is not applicable as there was no training set of data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183369

    Validate with FDA (Live)

    Device Name
    COR12 ECG
    Date Cleared
    2018-12-28

    (24 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The COR12 ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on a screen or printed on paper. The COR12 ECG can be used for applications in patients age 4 years and older and a weight of 20 kg or higher. The COR12 ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital).

    Device Description

    The COR12 ECG is intended for the recording of 12-lead ECG data of a patient by trained staff, e.g., nurse, doctor's assistant or the doctor. This can be performed as a resting or a stress ECG. It transmits the ECG data in a specified format by Bluetooth transmission to a customer-defined host software running with its own display, which is not part of the COR12. There is no display of ECG waveforms on the COR12.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the COR12 ECG device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is claiming substantial equivalence to its predicate (Vyntus ECG – K150810) by stating that they are identical in all aspects. Therefore, the "acceptance criteria" and "reported device performance" are assumed to be the same as the predicate device, as per the submission's claim of identicality.

    Feature / Acceptance CriteriaReported Device Performance (COR12 ECG & Predicate)Comment
    Indications for Use
    Measures Surface ECGYesIdentical
    Records & Displays/Prints ECGYesIdentical
    Patient Age4 years and olderIdentical
    Patient Weight20 kg or higherIdentical
    Use ConditionsRoutine ECG, resting and stress conditionsIdentical
    UserTrained healthcare professionalsIdentical
    Environment of UseHealthcare facilities (doctor's office/hospital)Identical
    Technical Specifications
    Records ECG onlyYesIdentical
    Data outputRecorded and displayed on screen or printedIdentical
    Records underResting and stress conditionsIdentical
    Dimensions (W x H x D)8.0 x 9.3 x 2.1 cm (3.3 x 3.7 x 0.8 in)Identical
    Weight (without battery)200 g (0.4 lbs)Identical
    12-channel Surface ECGYesIdentical
    ElectrodesStandard ECG electrodesIdentical
    Electrode Connection4 mm snap, gildedIdentical
    ECG signal TechnologyImpedance measurementIdentical
    Accuracy of ECG signal1.94 μV/bit resolution (at 500 Hz sampling)Identical
    FilterBandpass 0.05 Hz – 150 Hz, No line filterIdentical
    Pacemaker detection4000 HzIdentical
    Operating Temperature10 – 37 °CIdentical
    Ambient Pressure700 to 1060 hPaIdentical
    Humidity5 – 95% RH (not condensing)Identical
    Power Supply1x AA Alkaline or rechargeable NiMH batteryIdentical
    Runtime (with AA battery)> 5hIdentical
    Runtime (with NiMH battery)> 8h (2850 mAh)Identical
    Current Consumption230 mA (Operation) / 185 mA (Idle)Identical
    Protection ClassificationDevice with internal power supply, Type CFIdentical
    Recovery Time (defibrillation)< 8 secondsIdentical
    EMC EmissionCISPR 11 Class BIdentical
    EMC ImmunityIEC 61000-4 parts 2, 3, 6, 8Identical
    Data TransmissionBluetooth 2.1+EDR Protocol: SPPIdentical

    2. Sample size used for the test set and the data provenance

    The submission explicitly states, "The subject and predicate devices are identical as the supplier of the predicate and thus the technical specifications are identical." This indicates that no separate test set was used for the COR12 ECG device in this submission. The device's performance is established by its claimed identicality to the predicate device. Therefore, no specific sample size or data provenance is reported for a unique test set for the COR12 ECG.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no independent test set was used for the COR12 ECG to demonstrate performance, there were no experts establishing ground truth for such a set. The claims are based on identicality to the predicate.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. No independent test set with adjudication was performed for the COR12 ECG in this submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electrocardiograph for recording and displaying ECG data and does not involve AI assistance for interpretation or comparative effectiveness studies with human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a hardware electrocardiograph for data acquisition, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the COR12 ECG, the "ground truth" for its performance is implicitly derived from the established performance and safety of its identical predicate device, the Vyntus ECG (K150810). No new ground truth was established for the COR12 ECG in this submission. The submission relies on the predicate's prior clearance which would have involved demonstrating accuracy against established standards for ECG measurement.

    8. The sample size for the training set

    Not applicable. This device is a measurement device (electrocardiograph) and does not involve machine learning or AI models that require a training set.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112077

    Validate with FDA (Live)

    Date Cleared
    2012-03-09

    (232 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HBldb product is to be used by qualified medical professionals for the post-hoc statistical evaluation of the human electroencephalogram (EEG), utilizing evoked response potentials (ERP).

    Device Description

    The Human Brain Index Software (HBldb) product is a software program for the post-hoc statistical analysis and comparison with a reference data (the Human Brain Index Reference Database (HBIRD)) of the human electroencephalogram (EEG) including spontaneous oscillations of the human brain potentials and event-related potentials (ERPs). The EEG is recorded on a separate device under the standardized HBIdb conditions and is transferred to the HBIdb in EDF+ format for

    • display .
    • . spectral analysis and
    • analysis of event-related potentials .
    • comparison of the gathered parameters against the Human Brain Index Reference Database . (HBIRD), and
    • compilation of a report. .
      Results of comparison of individual EEG/ERP parameters with the database are intended for use as an aid to diagnosis. No medication or treatment is applied based on this comparison alone. The results have to be considered only in conjunction with other clinical findings.
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the HBIdb software:

    Important Note: The provided document is a 510(k) summary for a medical device. This type of submission focuses on demonstrating substantial equivalence to existing, legally marketed predicate devices, rather than proving efficacy or meeting specific performance acceptance criteria through clinical trials. Therefore, the information typically found in a clinical study report regarding detailed acceptance criteria, effect sizes, and deep dives into ground truth establishment is largely absent. The "acceptance criteria" here implicitly refer to demonstrating equivalence in intended use, technology, and performance with predicate devices.

    Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) summary, formal "acceptance criteria" in the sense of specific numerical thresholds for performance metrics are not explicitly stated. The comparison is made against predicate devices, focusing on equivalency of features and intended use. The "device performance" is therefore implicitly that it functions similarly to the predicate devices.

    Feature / Criterion (Implicitly "Acceptance Criteria")HBIdb Software Performance (Reported)
    Intended UseUsed by qualified medical professionals for post-hoc statistical evaluation of human EEG, utilizing evoked response potentials (ERP). Intended for use on children and adults from age 7 to 80 years.
    EEG Data Comparison against Normative DatabaseYes (Functions as intended to compare EEG data against a normative database).
    ERP/EP Data Comparison against Normative DatabaseYes (Functions as intended to compare ERP/EP data against a normative database).
    Population7 to 80 years (Comparable to predicate devices: 6-90 years, Birth-82 years, Unknown, 6-90 years).
    Product CodeGWQ (Matches predicate devices).
    Classification882.1400 (Matches predicate devices).
    ICA based Artifact CorrectionYes (This is a distinguishing feature from predicates, indicating advanced capability, not a "performance" metric in this context).
    Decomposition of ERPs into Independent ComponentsYes (This is a distinguishing feature from predicates, indicating advanced capability, not a "performance" metric in this context).
    Software Development and Validation StandardsDeveloped and validated according to IEC 62304:2007. Software documentation and testing provided in accordance with FDA guidance for "Content of Premarket Submissions for Software Contained in Medical Devices." All software functions tested using specific EEG test files. Verification and validation performed on Windows XP 32 bit, Vista 32 bit, Windows 7 32 and 64 bit. Security dongle validated with a special utility.

    Study Information:

    1. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: The document does not specify a numerical sample size for a "test set" in terms of subject data. It states, "For verification and validation of the product, all software functions were tested by means of specific EEG test files." This implies a set of pre-defined, synthetic, or existing EEG data files were used to test the software's functionality, not necessarily a clinical dataset for performance evaluation.
      • Data Provenance: Not specified for these "specific EEG test files." Given the manufacturer is German, the test environment and data could be internal or from Germany, but this is not directly stated. The data used for the Human Brain Index Reference Database (HBIRD) is not described here, only that the device compares against it.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable/Not specified. The testing described focuses on software functionality (verification and validation against IEC and FDA guidance), not on diagnostic accuracy of the device's output against a clinical "ground truth" established by experts. The HBIdb product aids diagnosis but doesn't provide a definitive diagnosis on its own.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable/Not specified. Since the testing was for software functionality rather than clinical diagnostic performance adjudicated by experts, no such method is mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC comparative effectiveness study was not performed or described in this 510(k) summary. This type of study demonstrates clinical utility with human readers, which is beyond the scope of this type of premarket submission which is focused on substantial equivalence.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, in a sense. The testing performed was "standalone" in that it validated the software's functions (e.g., display, spectral analysis, ERP analysis, comparison against the HBIRD, report compilation) using "specific EEG test files." This focused on the algorithm's correct execution and output formation, not on diagnostic accuracy by itself. The device is an "aid to diagnosis" and meant to be used by "qualified medical professionals," indicating that human interpretation is always in the loop for diagnosis.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the software's functional testing, the "ground truth" would be the expected output or behavior of the software for the "specific EEG test files" based on its design specifications and algorithms. This is a technical ground truth, not a clinical diagnostic ground truth. The HBIdb compares data against a Human Brain Index Reference Database (HBIRD), which is a form of ground truth (normative data), but the method of establishing that database's ground truth is not detailed in this document.
    7. The sample size for the training set:

      • Not specified. The document does not mention a "training set" for the HBIdb software itself. This suggests the software is likely rule-based or statistical in nature, comparing inputs against a pre-established "reference database" (HBIRD), rather than a machine learning model that requires a distinct training phase with labeled data. The HBIRD itself would have been built from a large sample, but its size is not in this document.
    8. How the ground truth for the training set was established:

      • Not applicable, as a distinct "training set" for a machine learning model is not described. The HBIdb compares data against the Human Brain Index Reference Database (HBIRD). How the "ground truth" for this normative database (HBIRD) was established (e.g., participant selection, EEG acquisition protocols, data processing, statistical methods for defining "normative ranges") is beyond the scope of this 510(k) summary and not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K082077

    Validate with FDA (Live)

    Device Name
    BT3/6, BT12
    Date Cleared
    2008-09-02

    (41 days)

    Product Code
    Regulation Number
    870.2920
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BT3/6 (3/6-lead) and BT12 (12-lead), hereafter referred to as the "BT devices", are battery powered devices capable of acquiring and transmitting a standard electrocardiogram (EKG) to be applied by medically trained persons for the purpose of cardiac monitoring and diagnosis performed by medical professionals. The collected data is processed by the B1 device and then transmitted via a standard wireless link to a monitoring device, such as a P or hand-held device for display, review, printing, saving and post event processing by medical professionals. The collected data is not interpreted by the BT device as this is done by the monitoring device operated by medical professionals. Use of the BT devices is not restricted to adult population, but is also intended for infants weighing less than 10 kg (22 lbs.). Measurements taken by the BT devices are only significant if considered in connection with other clinical findings. No therapy or drugs can be administered based solely on ECG data derived from the BT devices. The BT devices are not intended for monitoring critical patients and are not intended for intracardial use.

    Device Description

    BT3/6. BT12 is a compact ECG device which is worn on the body and is capable of wireless data transmission. With the BT3/6 type, 2 ECG leads (II - III according to Einthoven) can be measured via a 4-pin cable. The receiving unit can calculate four additional leads (lead I according to Einthoven, aVL, aVR, aVF according to Goldberger).Type BT12 has a 10-pin cable, which, in addition to the leads of the BT3/6 variant, can also measure the 6 leads according to Wilson. With this cable, one can select among 3-, 6- or 12-channel ECG measurements. The cables of the BT are pluggable and therefore can be easily changed. The electrode clips on the patient cables are suitable for connection to button and tab electrodes. It has an integrated electrode contact detector and a heart rate meter, the results of which are displayed on an integrated LCD display. Thanks to the integrated radio data transmission technology, BT3/6, BT12 is able to transmit ECG data online over a range of up to 10 m to a monitor or recording unit for further evaluation. BT3/6, BT12 can be used in all medical applications where the long, stiff trunk cables are a problem due to their weight, their motion artifacts or their electromagnetic susceptibility or because they limit the patient in his freedom of motion. Especially significant are the applications in which it is necessary for the patient to move around freely, for example when the patient is undergoing a stress electrocardiogram. BT3/6, BT12 is compact and simple to operate. A clip allows the device to the patient's clothing. BT3/6, BT12 is powered by two standard AA cells. Patient contacting components are the electrodes. BT3/6, BT12 is intended for use with approved. biocompatible electrodes only.

    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information regarding the acceptance criteria and study for the BT3/6, BT12 device.

    Important Note: The provided document is a 510(k) Premarket Notification. 510(k) submissions typically demonstrate substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a full clinical comparative effectiveness study in the same way a PMA submission might. Therefore, some of the requested information (like effect size for MRMC studies or ground truth for training sets) may not be explicitly present because it's not a required component of a 510(k) for this type of device.


    Acceptance Criteria and Device Performance

    The device demonstrates compliance with recognized standards and shows performance characteristics that are comparable to or better than a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't define explicit "acceptance criteria" in a pass/fail sense for clinical performance. Instead, it compares the BT3/6, BT12 device's technical specifications and intended use against a legally marketed predicate device (Card Guard CG-7000DX-BT). The "acceptance" is based on demonstrating substantial equivalence, meaning the new device is as safe and effective as the predicate.

    Here's a table summarizing the comparison for key parameters:

    ParameterAcceptance Criteria (Predicate Device CG-7000DX-BT)Reported Device Performance (BT3/6, BT12)
    Intended UseSimilar - 12-lead ambulatory ECG for cardiac monitoring and diagnosis by medical professionals.Similar, with added intent for 3/6-lead options and infants < 10 kg. Both collect, process, and transmit ECG data; no interpretation by device.
    ECG interpretation softwareNo interpretationNo interpretation
    Input dynamic range+/- 5mV+/- 5 mV
    Frequency response bandwidthNot explicitly stated for predicate in comparison table, but predicate claims conformance to EC11 and EC38.0.05-150Hz / according to EC11 and IEC 60601-2-51
    Resolution10 bit24 bit A/D converter (15 bit transmitted); 2.58 µV/bit. (Better than predicate)
    Leads123/6 or 12
    CMRR (Common Mode Rejection Ratio)60 dB>94 dB (Better than predicate)
    Pacemaker detectionYesYes
    Input impedance> 10 MΩ20 MΩ (Better than predicate)
    DC offset correction± 75 mV± 190 mV (Better than predicate in stated performance claim)
    ECG storage capacity4-20s length, 40-10 recordings (event recorder)5 min/12 channel when transmission is interrupted (stores during interruption, not for general event recording)
    Temperature range (Operation)0...40°C0...50°C
    Temperature range (Storage)-20...65°C-20...70°C
    DisplayLCD (graphic, low resolution)LCD (displays on receiving unit only)
    A to D sampling rate500 samples/sec500 samples/sec
    Data transmissionWirelessWireless
    Degree of protection against penetration of waterIPX0IPX3 (Better than predicate)
    Defibrillation protectionNot defibrillation-proof (optional cable available)Device not defibrillation-proof, but cable is.
    Compliance to standardsPredicate claims conformance to ANSI/AAMI EC11:1991 and ANSI/AAMI EC38:1998Conforms to hardware requirements of EC11 and EC38. Full compliance not claimed due to lack of diagnostic software provided by manufacturer. Full compliance to IEC 60601-1 and IEC 60601-1-2.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a clinical "test set" in the context of diagnostic performance evaluation for a medical device that interprets ECGs. The BT3/6, BT12 device does not interpret ECG data; it merely acquires and transmits it.

    The testing mentioned relates primarily to hardware safety and performance standards. The submission states: "The BT3/6, BT12 has been tested according to IEC 60601-1 and IEC 60601-1-2 by accredited laboratories and has shown full compliance to these standards and the standards listed in chapter 9 of this 510(k)." This refers to engineering and electrical safety testing, not clinical studies with patient data to evaluate diagnostic accuracy.

    Therefore, there is no information on:

    • Sample size for a clinical test set.
    • Data provenance (country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As the device does not interpret ECG data, there was no need for experts to establish ground truth for a diagnostic test set. The device's function is to accurately acquire and transmit the raw ECG signal.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no clinical test set requiring adjudication in the context of diagnostic performance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a wireless ECG acquisition and transmission device, not an AI-assisted diagnostic tool. No MRMC study was mentioned or performed in the context of reader improvement with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The device's primary function is signal acquisition and transmission, not algorithmic interpretation by itself. Its performance is evaluated on its ability to accurately capture and transmit the raw ECG signal according to technical specifications, not on a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable for diagnostic ground truth. The "ground truth" for this device would be its adherence to technical specifications and standards for signal integrity, electrical safety, and wireless transmission, as outlined in IEC 60601-1, IEC 60601-1-2, and the hardware aspects of ANSI/AAMI EC11 and EC38. Compliance with these standards by accredited laboratories serves as the "proof" of its foundational performance.

    8. The sample size for the training set

    Not applicable. This device is a hardware product for signal acquisition and transmission, not a machine learning model that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073557

    Validate with FDA (Live)

    Date Cleared
    2008-02-28

    (71 days)

    Product Code
    Regulation Number
    882.5050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EEG NeuroAmp is intended for biofeedback and relaxation purposes. To perform its intended function it must be used in combination with a computer and appropriate software.

    For purposes of this training task, information for feedback may be derived from one or two channels of EEG and from as many as three channels of peripheral physiology measures such as are conventionally used in biofeedback.

    The EEG NeuroAmp is intended to be used in the office by trained professionals who can ensure sound handling practices.

    The EEG NeuroAmp is not intended for diagnostic purposes.

    Device Description

    The purpose of the EEG NeuroAmp is to act as user-friendly high-performance interface between client and clinician computer for EEG biofeedback) and/or peripheral biofeedback therapy. The EEG NeuroAmp contains four function blocks:

      1. EEG amplifier, two channels, aligned and high-resolution
      1. Built-in impedance meter for five electrodes helps achieve good electrode contact and warns if electrodes are worn out
      1. Peripheral sensor amplifiers, three channels, allow measurement of peripheral biofeedback signals such as GSR (Galvanic Skin Response) and temperature. The latter may be by means of a contact probe (thermistor) or non-contact infrared thermal sensor (infrared thermometer).
      1. An output for audio/visual/tactile feedback
    AI/ML Overview

    The Corscience GmbH & Co. KG EEG NeuroAmp is a biofeedback device intended for biofeedback and relaxation purposes.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to predicate devices and compliance with relevant standards. It does not explicitly define "acceptance criteria" in terms of specific performance metrics for the overall device's therapeutic effectiveness (e.g., a percentage reduction in a particular symptom). Instead, the acceptance is based on meeting safety and technical standards and demonstrating comparable performance to predicate devices for its intended use.

    However, some technical specifications can be viewed as performance characteristics that were likely expected to meet certain thresholds or be comparable to predicate devices.

    ParameterAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (EEG NeuroAmp)
    Intended UseBiofeedback and relaxation purposes. Must be used with computer and software. Data from 1-2 EEG channels and up to 3 peripheral physiology measures. For use by trained professionals in an office setting. Not for diagnostic purposes.The EEG NeuroAmp is intended for biofeedback and relaxation purposes. To perform its intended function it must be used in combination with a computer and appropriate software (Cygnet). For purposes of this training task, information for feedback may be derived from one or two channels of EEG and from as many as three channels of peripheral physiology measures such as are conventionally used in biofeedback. The EEG NeuroAmp is intended to be used in the office by trained professionals who can ensure sound handling practices. The EEG NeuroAmp is not intended for diagnostic purposes.
    Power SupplyAcceptable and safe power source.Power supply via USB port (galvanic isolation according to IEC 60601-1).
    SoftwareFunctional and appropriate for biofeedback.Cygnet
    No. of Input ChannelsComparable to predicate devices (BrainMaster had 2, ProComp had 8).5
    No. of Output ChannelsComparable to predicate devices (Predicates had 0).1
    ADC ResolutionComparable to predicate devices (BrainMaster 8 bit, ProComp 14 bit).13 bit
    Anti-aliasing FilterPresent for effective signal processing.5th order Butterworth
    DC Gain AccuracyComparable to predicate devices (BrainMaster ±5%, ProComp ±0.5%).±1%
    DC OffsetComparable to predicate devices (BrainMaster +4 LSB, ProComp +3 LSB).+1 LSB
    Bandwidth (3dB) & Sample RateAppropriate for EEG and peripheral biofeedback.EEG amplifier (2 channels): 0.08 Hz (1st order roll-off) ...70 (5th order roll-off). Sample rate for 50Hz line: 1000 sps. Sample rate for 60Hz line: 960 sps. Subsample rate = sample rate / 4 (250/240). Peripheral Amplifier (3 channels): DC ... 10Hz. Same sampling rate scheme.
    Full-scale Input Range, DCAppropriate for biofeedback signals.± 250 mV
    Overvoltage WarningDesirable for safety (ProComp/BrainMaster lacked this).Yes
    Impedance MeasurementAbility to check electrode contact quality.Both channels simultaneously; displayed by colored LED bars on device. Needs to be engaged explicitly, disabled during EEG neurofeedback.
    CMRR - EEG sensor onlyHigh common-mode rejection for signal quality (>100 dB for BrainMaster, >130 dB for ProComp).>130 dB
    Temperature Range (pIRx3 accessory)Suitable for forehead temperature measurement.29°C - 43°C or 84°F - 109°F
    Accuracy (pIRx3 accessory)Accurate enough for biofeedback (ProComp ±1.0°C).< 1°K
    Safety and EMC (Electrical Medical Safety)Compliance with IEC 60601-1 and IEC 60601-1-2.Full compliance demonstrated by accredited laboratories. Galvanically isolated from computer by medical grade isolating transformer.
    Software ValidationValidated for intended function.Tested as described in section 16 "Software" (details not provided in the excerpt).
    BiocompatibilityCompliance with ISO 10993-1 for patient-contacting parts.Housing material passed all tests for ISO 10993-1. Elastic belt: conforms to German "Oko-Tex-Standard," passed dermal irritation and sensitization tests, but did not pass cytotoxicity due to cotton material and is suitable for use on intact skin only (implies an acceptance of this limitation).

    Study Proving Acceptance Criteria are Met:

    The document describes a series of tests and comparisons to predicate devices to demonstrate substantial equivalence and safety rather than a single "study" proving therapeutic effectiveness against pre-defined clinical acceptance criteria.

    The primary "study" or demonstration of meeting acceptance criteria consists of:

    • Bench Testing and Compliance with Standards: The EEG NeuroAmp and its accessory were tested according to international standards for electrical medical devices and biocompatibility.
    • Comparison to Predicate Devices: A detailed comparison table (Section 5.12) is provided, highlighting the similarities in intended use and technical specifications, and justifying any differences.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document does not refer to a "test set" in the context of patient data or clinical outcomes. The testing primarily involved bench testing of the device's hardware and software against technical specifications and safety standards. There is no indication of patient data being used for device performance evaluation.
    • Data Provenance: Not applicable, as there is no patient data described as part of the performance evaluation. The data provenance described relates to manufacturing standards and regulatory compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Ground Truth Experts: Not applicable in the context of clinical ground truth. The "ground truth" for the device's acceptable performance was established by its compliance with recognized international standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-1) and comparison to the technical specifications of legally marketed predicate devices. The experts involved would be engineers and testing personnel at accredited laboratories. Their qualifications are implicitly that they are subject matter experts in the respective testing domains (e.g., electrical safety, EMC, biocompatibility).

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable. There was no clinical data requiring expert adjudication. Device performance was assessed against technical specifications and regulatory standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This device is a biofeedback device, not an AI-powered diagnostic or assistive tool for human readers. No MRMC comparative effectiveness study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Not applicable in the context of an algorithm's diagnostic performance. The device's "standalone" performance refers to its ability to accurately measure and process physiological signals according to its technical specifications, which was assessed through bench testing and compliance with standards. The device itself is an interface and amplifier, and its function is to provide accurate data to the clinician's software, which is then used in a human-in-the-loop biofeedback process.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: The ground truth used was technical specifications, international safety standards, and performance characteristics of legally marketed predicate devices. For example, a specified gain accuracy of ±1% or a CMRR of >130 dB would be the "ground truth" to which the device's measured performance was compared during testing.

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This device is a hardware biofeedback system, not an AI/ML algorithm that requires a training set of data.

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable. As there is no AI/ML algorithm involved, there is no training set or ground truth for such a set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1