Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230318

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-07

    (60 days)

    Product Code
    Regulation Number
    870.3800
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Memo 3D device is indicated for use in patients suffering from congenital or acquired mitral insufficiencies or stenoinsufficiencies with dilatation and deformation of the mitral annulus. Type I insufficiencies, with no manifest lesions in the subvalvular apparatus, can be treated with the implant of the annuloplasty ring on its own. For type II insufficiencies, characterised by valve prolapse sustained by elongation/ breakage of the chordae tendineae and papillary muscles, and type III insufficiencies, characterised by partially immobilised leaflets due to the fusion/hypertrophy of the chordae tendineae, the device implantation must be accompanied by corrective valvuloplasty.

    The use of the MEMO 3D ReChord device is indicated for use in patients suffering from congenital or acquired mittal insufficiencies or steno-insufficiencies with dilatation and deformation of the mitral annulus. Type I insufficiencies, with no manifest lesions in the subvalvular apparatus, can be treated with the implant of the annuloplasty ring on its own. For type II insufficiencies, characterised by valve prolapse sustained by elongation' breakage of the chordae and papillary muscles, and type III insufficiencies, characterised by partially immobilised leaflets due to the fusion/hypertrophy of the chordae tendineae, the device implantation must be accompanied by corrective valvuloplasty.

    The use of the Memo 4D device is indicated for use in patients suffering from congenital or acquired mitral insufficiencies or steno-insufficient with dilatation of the mitral annulus. Type I insufficiencies, with no manifest lesions in the subvalvular apparatus, can be treated with the implant of the annuloplasty ring on its own. For type II insufficiencies, characterised by valve prolapse sustained by elongation' breakage of the chordae tendineae and payillary muscles, and type III insufficiencies, characterised by partially immobilised leaflets due to the fusion/hypertrophy of the chordae tendineae, the device implantation must be accompanied by corrective valvuloplasty.

    Device Description

    Like their predicate devices, the subject MEMO 3D, MEMO 3D ReChord and MEMO 4D Semirigid Annuloplasty Rings are supplied as sterile, nonpyrogenic ring pre-mounted on disposable holders.

    Both the predicate and subject devices are manufactured by embedding a superelastic metallic alloy inner core with medical grade silicone.

    The resulting silicone sheath around the inner core is then encased within a tubular knitted fabric coated with a thin layer of turbostratic carbon (Carbofilm™). The fabric is then sewn along its length with a Carbofilm™ coated polyester thread.

    Both the predicate and subject devices MEMO 3D ReChord and MEMO 4D Semirigid Annuloplasty Rings feature a fully removable system (i.e., the ReChord System) in the posterior curve of the annuloplasty ring, composed by a series of loops made by a single piece of yellow surgical thread retained in place by a single piece of blue surgical thread.

    The ReChord System is designed to provide a temporary reference element to facilitate the sizing of the artificial chord length at the annular plane level when performing replacement of mitral chordae tendineae in concomitance with the implant of the annuloplasty ring.

    The MEMO 3D and MEMO 3D ReChord Annuloplasty Rings are manufactured in 8 different sizes, from 24 to 38 mm, with 2 mm increments.

    The MEMO 4D Annuloplasty Ring is manufactured in 10 different sizes, from 24 to 42 mm, with 2 mm increments.

    A complete set of accessories is available separately to properly size the annulus and implant the MEMO 3D, MEMO 3D ReChord and MEMO 4D annuloplasty ring.

    AI/ML Overview

    The provided document is a 510(k) summary for the Memo 3D Semirigid Annuloplasty Ring, Memo 3D ReChord Semirigid Annuloplasty Ring, and Memo 4D Semirigid Annuloplasty Ring. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a standalone study.

    Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving device performance in the way a clinical study report would.

    Specifically:

    1. A table of acceptance criteria and the reported device performance: Not provided. The submission focuses on showing that changes in raw materials and manufacturing (alternative suppliers for PET fabric and suture thread, and a different sewing path) do not alter the fundamental characteristics or performance compared to the predicate devices.
    2. Sample size used for the test set and the data provenance: Not applicable in the context of a 510(k) submission primarily relying on non-clinical testing for material and manufacturing changes.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    4. Adjudication method: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an annuloplasty ring, not an AI-powered diagnostic tool.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used: Not applicable.
    8. The sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    Summary of Non-Clinical Performance Data Provided (Section VII):

    The submission relies on the following non-clinical performance data to demonstrate substantial equivalence:

    • Biocompatibility testing
    • Mechanical characterization of the PET fabric from the alternative supplier.
    • Mechanical characterization of the PET suture thread from the alternative supplier and the alternative sewing path.
    • Mechanical testing of the final device manufactured with the new suppliers' materials and sewing path.
    • Quality control testing to evaluate the Carbofilm™ coating on the PET fabric and suture thread from the alternative suppliers.
    • LAL test to evaluate endotoxin contamination on the new materials and final devices.

    Clinical Performance Data (Section VIII):

    The document explicitly states: "No clinical testing was conducted in support of the subject devices, as the indications for use are equivalent to those of their predicate devices (K071327, K142221, K180411). The non-clinical testing referred to in this submission supports the substantial equivalence of these devices."

    Therefore, the approval is based on the substantial equivalence of the modified devices to already cleared predicate devices, supported by non-clinical testing to ensure that the changes in raw materials and manufacturing methods do not adversely affect the safety and effectiveness of the device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1