510(k) Data Aggregation

    K Number
    K251511

    Validate with FDA (Live)

    Device Name
    PhenoMATRIX
    Manufacturer
    Date Cleared
    2026-01-22

    (251 days)

    Product Code
    Regulation Number
    866.2190
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PhenoMATRIX is a WASPLab optional module intended for the automatic sorting of images of blood-based agar, chocolate agar, MacConkey agar, and CHROMagar Orientation culture media plates according to classification parameters based on Image Analysis Software results and clinical and demographic data.

    Image Analysis Software performs semi-quantitative and/or qualitative analysis of culture media plates by detecting microbial growth, estimating colony counts and differentiating isolates based on phenotypic colony characteristics.

    The system determines how the images should be sorted based on image analysis results in addition to patient data according to expert rules defined by the laboratory.

    All images shall be evaluated by trained personnel for final assessment and result definition.

    Device Description

    The PhenoMATRIX is an in vitro diagnostic software for automated classification of images of solid culture media plates streaked with microbiological samples derived from the human body.

    The PhenoMATRIX comprises software modules intended for image analysis and automatic classification of high-resolution digital images captured by WASPLab device for semi-quantitative and qualitative assessment of microbial growth.

    WLPostProcessing, its plugin and the Imaging Product SET perform microbial growth detection, colony counts estimations and isolates differentiation basing on phenotypic colony characteristics. The image analysis result is combined with LIS data (such as demographic data, clinical data and / or sample data) according to customizable logic rules defined by the laboratory, for image classification. The classification is used for image sorting into dedicated digital folders associated to suggested results.

    The PhenoMATRIX requires the WASPLab in order to operate. WASPLab is an in vitro diagnostic device for handling, incubation, digital imaging and sorting of agar culture plates.

    After PhenoMATRIX processing, the physical plates are left inside the WASPLab and images are available for digital inspection by the trained microbiologist through the WASPLab User Interface. The trained microbiologist shall assess the plate images in each digital classification folder. To each digital image a suggested result is assigned according to rules previously defined by the laboratory. The trained microbiologist reviews the plate images, folder by folder, and confirms (or changes) the assigned folder and result. After that, the plates follow the workflow that has been defined by the laboratory according to the assigned result.

    The main functionalities of PhenoMATRIX are:

    • Assignment of classification parameters to media plate images based on criteria defined by the user (e.g., LIS information, Image Analysis).
    • Result assignment according to the classification folder.
    • Visualization of media plate images in the WASPLab User Interface for digital inspection.

    All plate images shall be reviewed by trained microbiologist before any result definition.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K223245

    Validate with FDA (Live)

    Device Name
    Colibrí
    Manufacturer
    Date Cleared
    2023-03-20

    (151 days)

    Product Code
    Regulation Number
    866.3378
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Colibri™ is an automated in vitro diagnostic specimen preparation system for use with WASPLab® to prepare MALDI-TOF targets for the bioMérieux VITEK® MS or Bruker MALDI Biotyper® CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK® 2 Antimicrobial Susceptibility Testing (AST) system for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.

    The Colibri™ is an automated pre-analytical processor that nicks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/lonization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.

    The Collori™ software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

    Bacterial suspensions for AST and purity plates are identified by barcode label.

    The Colibri™ is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections.

    The Colibri™ has not been validated for use in the identification or processing of yeast species, or mycobacteria.

    Device Description

    The Colibrí is an instrument which automates the picking of selected colonies from plated media and prepares MALDI target slides for the bioMérieux VITEK MS or the Bruker MALDI Biotyper CA System that are used in clinical laboratories for identification and differentiation of organisms grown on plated media by Matrix-Assisted Laser Desorption/Tonization Time-of Flight Mass Spectrometry (MALDI-TOF MS). The Colibrí automates the preparation of microbial suspensions at known concentration for bioMérieux VITEK 2 System that is used in clinical laboratories for AST analyses. Moreover, the Colibrí is used for Purity Plates preparation for purity assessments.

    The Colibrí includes the following components:

    • . Colibrí instrument and software with on-board pipetting system and nephelometer
    • Colibrí Primary Tubes ●
    • . Colibrí Spreader
    • Colibrí Daily Verification kit. ●

    Colibri is designed to be used in conjunction with the WASPLab device for culture plate incubation and image analysis. After appropriate plate incubation, the operator selects the colonies from a digital image of culture media plate streaked with microbiological human specimen, available through WebApp software, the WASPLab User Interface.

    The operator assigns the automatic ID or AST tasks to the isolated colonies to be processed. Then, the operator loads the plates in the Colibri where colonies are automatically picked, spotted on the target slide and overlayed with the matrix or suspended into the dedicated solution for the preparation of the microbial suspension for AST purposes (Secondary Tube).

    When used in conjunction with the bioMérieux VITEK MS, the Colibrí can prepare the 48-spot target slides by performing the direct spotting of colonies. The calibrator used for quality control is manually applied by the operator at the end of the automated colony spotting. When used in conjunction with the Bruker MALDI Biotyper CA System, the Colibri can prepare either reusable 48-spot or disposable 96-spot targets by performing the Direct Transfer Sample Procedure. The BTS used for quality control is manually applied by the operator at the end of the automated colony spotting.

    When used in conjunction with the bioMérieux VITEK 2, the Colibrí can prepare the microbial suspension at the proper concentration by direct colony suspension method. The onboard nephelometer allows the preparation of Secondary Tubes (AST suspensions) at the correct concentration and the Colibrí Spreader is used for Purity Plates preparation.

    The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

    The traceability of prepared Secondary Tube and Purity Plates is maintained by dedicated labels applications.

    Colibrí requires four different calibrations, one on the nephelometer, three on the cameras. None of these calibration activities require user intervention if not in terms of periodical cleaning of the mechanical component as described in the dedicated section of the User Manual. The Set-up calibration of nephelometer and camera units are performed during the device initial setup. Autocalibration is performed at the end of the initial set-up and periodically during the preventive maintenance to check that all the mechanical references can be found inside the positioning tolerances, that the I/Os are responsive. Run-time calibration is performed during the normal usage to automatically check the proper functioning of the Colibrí.

    Colibrí requires a daily nephelometer verification to check the proper reading of suspensions at different turbidity values.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Colibrí

    The Colibrí is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for microbial identification and microbial suspensions for Antimicrobial Susceptibility Testing (AST).

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance CriteriaReported Device Performance
    Colony Picking Accuracy100% correct picking of designated colonies.100% of designated colonies were correctly picked without any event of picking a wrong colony.
    VITEK MS Identification Agreement (Gram-Negative)High agreement with expected strain identity.Gram-Negative Species: All species (Citrobacter koseri, Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa) showed 100% agreement with good confidence for VITEK MS identification. Across 92 picked colonies, there was a 100% agreement on species/group identification.
    VITEK MS Identification Agreement (Gram-Positive)High agreement with expected strain identity.Gram-Positive Species: Overall, 98.0% agreement with good confidence across 200 picked colonies. Individual species agreement ranged from 95.0% (Enterococcus faecium) to 100% (Staphylococcus aureus, Staphylococcus saprophyticus).
    AST Essential Agreement (EA)>98% agreement between Colibrí prepared samples and manually prepared samples.For all species and antimicrobial agents combined, 100% (1315/1315) of on-scale MIC results were in Essential Agreement (EA).
    AST Category Agreement (CA)>98% agreement between Colibrí prepared samples and manually prepared samples.For all species and antimicrobial agents combined, 99.4% (2703/2720) of results were in Category Agreement (CA).
    AST Very Major Category Error0%0%
    AST Major Category Error0%0%
    AST Minor Category ErrorAcceptable level (calculated based on the differences in CA from 100%).17 minor errors occurred out of 2720 tested, resulting in a minor error rate of 0.6%. This is considered acceptable given the high overall CA.

    2. Sample Size Used for the Test Set and Data Provenance

    The study utilized a test set consisting of various bacterial strains to evaluate the Colibrí's performance in preparing samples for MALDI-TOF identification and AST.

    • Sample Size for Identification:
      • Gram-Negative Species: 5 species (4 Enterobacterales, 1 Non-fermenter) with 92 colonies picked and analyzed.
      • Gram-Positive Species: 5 species (2 Enterococcus, 2 Staphylococcus, 1 Streptococcus) with 200 colonies picked and analyzed.
      • Total for Identification: 292 colonies.
    • Sample Size for AST:
      • Total AST Tested: 2720 tests (sum of "Total tested" column from AST summary tables).
      • This included: 1400 Enterobacterales, 200 Non-fermenters, 520 Staphylococci, 380 Enterococci, and 220 Streptococci.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. However, given that it is a submission to the U.S. Food & Drug Administration (FDA) by an Italian company (Copan WASP Srl, Brescia, Italy), the study was likely conducted to meet international regulatory standards, potentially with data from one or more clinical laboratories. The nature of the "Full workflow validation" study described suggests a prospective experimental design to assess the device's performance under controlled conditions.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not explicitly state the number of experts used or their specific qualifications for establishing ground truth. However, the study focuses on the accuracy of identification results against "expected strain identity" and comparison of MICs and SIR categories against "MICs obtained by bioMérieux VITEK 2 using manual sample preparation" and "FDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria."

    This implies that:

    • For identification, the "expected strain identity" would have been established using a gold standard method, likely by skilled microbiologists or reference laboratories with established credentials in microbial identification.
    • For AST, the "manual sample preparation" by bioMérieux VITEK 2 would serve as the comparator, and the interpretation of results would follow "FDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria," likely applied by qualified laboratory personnel.

    4. Adjudication Method

    The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for the test set.

    • For identification, performance was calculated as "percentage of spotted colonies matching the expected identity," suggesting a direct comparison to a single reference (ground truth).
    • For AST, the "MICs obtained by bioMérieux VITEK 2 using Colibrí were compared to the MICs obtained by bioMérieux VITEK 2 using manual sample preparation," indicating a direct method-to-method comparison. Discrepant SIR results were categorized, implying a systematic evaluation rather than a consensus-based adjudication in the traditional sense.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size of how much human readers improve with AI vs. without AI assistance. The Colibrí device is a pre-analytical automated system for sample preparation, not an AI-assisted diagnostic interpretation tool for human readers. Its function is to automate the preparation steps for subsequent analysis by other IVD systems (MALDI-TOF MS and VITEK 2 AST).

    6. Standalone Performance Study

    Yes, a standalone performance study was done for the algorithm/system. The "Full workflow validation" study directly assesses the Colibrí device's ability to accurately pick colonies and prepare samples for downstream analysis without human intervention in the picking and preparation steps themselves. The results presented for "VITEK MS Identification Result" and "AST summary of results" directly reflect the performance of the Colibrí-prepared samples when analyzed by the respective analytical instruments.

    7. Type of Ground Truth Used

    • For microbial identification: The ground truth was based on "expected strain identity," which implies confirmed identification results likely obtained through established, highly accurate microbiological methods serving as a reference.
    • For AST: The ground truth for comparative purposes was "MICs obtained by bioMérieux VITEK 2 using manual sample preparation," interpreted according to "FDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria."

    8. Sample Size for the Training Set

    The document explicitly states that the "Colibrí uses equally designed and developed hardware and software modules as the Colibrí System. Therefore, the performance of the predicate device represents the performance of the new device." It also mentions that "The results of the analytical studies were submitted to support the 510(k) Premarket Notifications K193138 and K220546." This indicates that the current submission (K223245) relies on the performance data of the predicate device (Colibrí System, K220546). The document does not provide details on the training set sample size for the Colibrí System's development.

    9. How the Ground Truth for the Training Set Was Established

    As the current submission relies on the predicate device's performance, the ground truth for any training set related to the predicate Colibrí System would have been established during its development and prior FDA submission (K220546). The current document does not provide details of how the ground truth for the predicate device's training set was established. However, based on the performance parameters listed (e.g., accuracy of colony picking, reproducibility of identification, accuracy of nephelometer), it likely involved:

    • Microbial cultures: Using well-characterized microbial strains with confirmed identities.
    • Manual reference methods: Comparing automated colony picking against visual confirmation by trained personnel.
    • Reference AST methods: Comparing automated suspension preparation and subsequent AST results against established manual AST methods and interpretive criteria to assess agreement in MICs and categorical susceptibility (Susceptible, Intermediate, Resistant).