Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K161048

    Validate with FDA (Live)

    Date Cleared
    2016-07-26

    (103 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Conventus Orthopaedics Ulna Fixation System is indicated for treatment of distal or proximal ulna fractures when internal fixation is desired, and fracture fragments are not too numerous and/or too small to be stabilized with the use of the device.

    Device Description

    The Conventus Orthopaedics Ulna Fixation System is an intramedullary device intended for distal and proximal ulna fractures. The Ulna Fixation System is a self-expanding implant which is deployed into the medullary canal and provides a scaffold to which bone fragments are attached using fragment screws. The implant is made from titanium alloy (Ti-6Al-4V ELI) and Nitinol.

    AI/ML Overview

    This document is a 510(k) summary for the Conventus Orthopaedics Ulna Fixation System. It does not describe a study involving an AI/CADe/CADx device and human readers. Instead, it describes a medical device for fracture fixation and preclinical testing done to prove substantial equivalence to predicate devices. Therefore, I cannot extract the information requested as it is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151379

    Validate with FDA (Live)

    Date Cleared
    2015-08-13

    (83 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Conventus PRSTM System is indicated for treatment of proximal radial fractures when internal fixation is desired, and fracture fragments are not too numerous and/or too small to be stabilized with the use of the device.

    Device Description

    The Conventus PRSTM is an intramedullary device intended for proximal radius fractures. The PRSTM System is a self-expanding implant which is deployed into the medullary canal and provides a scaffold to which bone fragments are attached using fragment screws. The implant is made from titanium alloy (Ti-6Al-4V ELI) and Nitinol.

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA to Conventus Orthopaedics, Inc. It concerns the Conventus PRSTM System, an intramedullary device for proximal radial fractures. The letter confirms substantial equivalence to predicate devices based on pre-clinical testing mentioned in the 510(k) Summary.

    Here's an analysis of the provided information, focusing on acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    The provided document does not explicitly state specific acceptance criteria (e.g., pass/fail thresholds) for the pre-clinical tests. It lists the types of tests performed and concludes that "The results demonstrate that the PRSTM System is substantially equivalent to the legally marketed predicate devices." This implies that the device performance met the criteria necessary to demonstrate equivalency to the predicate devices, whatever those inherent criteria were for each test.

    Acceptance Criteria Type (Implicit)Reported Device Performance (Summary)
    Static and cyclic axial/bend testingMet performance equivalent to predicate devices.
    Static and cyclic torsional testingMet performance equivalent to predicate devices.
    Screw pullout testingMet performance equivalent to predicate devices.
    Corrosion testingMet performance equivalent to predicate devices.
    Wear testingMet performance equivalent to predicate devices.
    Nickel ion release testingMet performance equivalent to predicate devices.
    Surface analysis testingMet performance equivalent to predicate devices.
    Nitinol phase compositionMet performance equivalent to predicate devices.
    Nitinol transition temperature (Af)Met performance equivalent to predicate devices.
    Biocompatibility testingMet performance equivalent to predicate devices.
    Animal TestingMet performance equivalent to predicate devices.
    MRI TestingMet performance equivalent to predicate devices.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not provide specific sample sizes for any of the listed pre-clinical tests. It also does not specify the data provenance in terms of country of origin or whether the studies were retrospective or prospective. Given these are pre-clinical (laboratory and animal) tests, the concept of retrospective/prospective human data provenance does not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable to the type of pre-clinical studies described. These studies involve material science, mechanical engineering, and animal models, not human diagnostic interpretation requiring expert consensus on ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable as the studies are physical/mechanical/biological tests, not diagnostic interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not conducted. This is a pre-clinical evaluation of a surgical implant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This refers to AI algorithm performance and is not applicable to the pre-clinical evaluation of a medical device implant.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For these pre-clinical studies, "ground truth" refers to the established scientific and engineering principles, material properties, and biological responses measured by validated test methods. For example:

    • Mechanical tests: Ground truth is derived from engineering standards, material specifications, and the established performance characteristics of the predicate devices.
    • Corrosion, wear, nickel ion release, surface analysis, Nitinol properties: Ground truth is based on established material science standards and acceptable ranges for medical implant materials.
    • Biocompatibility: Ground truth is established by ISO standards and recognized biological response criteria (e.g., cytotoxicity, sensitization, irritation).
    • Animal Testing: Ground truth involves histological analysis, imaging, and functional assessment to determine biological response and healing in comparison to expected outcomes or predicate device performance.

    8. The sample size for the training set

    This refers to AI model training and is not applicable to the pre-clinical evaluation of a medical device implant.

    9. How the ground truth for the training set was established

    This refers to AI model training and is not applicable to the pre-clinical evaluation of a medical device implant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141737

    Validate with FDA (Live)

    Date Cleared
    2015-04-08

    (285 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Conventus Orthopaedics PHS™ System is indicated for the fixation of proximal humerus fractures except when there are too many fracture fragments to repair the articular surface.

    Device Description

    The Conventus PHS™ is an intramedullary device intended to treat proximal humerus fractures. The PHS™ System is a self-expanding implant which is deployed into the medullary canal and provides a scaffold to which bone fragments are attached using fragment screws. The implant is made from titanium alloy (Ti-6Al-4V ELI) and Nitinol.

    AI/ML Overview

    The provided text describes a medical device called the Conventus PHS™ System, an intramedullary device for fixing proximal humerus fractures. However, it does not include information about acceptance criteria or a study that proves the device meets those criteria in the context of diagnostic or AI-assisted performance.

    The document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device, primarily through mechanical and biocompatibility testing. It outlines the device's physical properties and mechanical performance, not its diagnostic accuracy or the performance of any AI component.

    Therefore, I cannot provide details for most of your requested information as it is not present in the provided text.

    Here is what I can extract:

    1. A table of acceptance criteria and the reported device performance
    The document does not explicitly state acceptance criteria or performance metrics related to diagnostic accuracy or AI performance. The pre-clinical tests performed are mechanical and material characterization tests (e.g., static and cyclic axial/bend testing, screw pullout testing, corrosion testing, wear testing, biocompatibility testing, animal testing). The conclusion is simply that "The results demonstrate that the PHS™ System is substantially equivalent to the legally marketed predicate devices," without providing specific numerical acceptance criteria or performance values for these tests.

    2. Sample size used for the test set and the data provenance
    Not applicable. There is no "test set" in the context of diagnostic or AI performance evaluation described. The "testing" refers to mechanical and material characterization of the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. No ground truth establishment for a diagnostic test set is described.

    4. Adjudication method for the test set
    Not applicable. No test set adjudication method is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    No. An MRMC study or any study involving human readers or AI assistance is not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    No. There is no mention of an algorithm or AI performance.

    7. The type of ground truth used
    Not applicable. No ground truth for diagnostic or AI performance is established. The "ground truth" equivalent for the pre-clinical tests would be established engineering standards and material properties, but these are not for diagnostic performance.

    8. The sample size for the training set
    Not applicable. No training set for an AI or diagnostic algorithm is mentioned.

    9. How the ground truth for the training set was established
    Not applicable. No training set for an AI or diagnostic algorithm is mentioned.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1