510(k) Data Aggregation

    K Number
    K172986

    Validate with FDA (Live)

    Device Name
    Zmachine Synergy
    Date Cleared
    2017-12-19

    (83 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zmachine Synergy is an EEG and respiratory signal recorder. The device is intended for use by adult patients in the home or clinical environment, under the direction of a qualified healthcare practitioner, to aid in the diagnosis of sleep disorders.

    Device Description

    The Zmachine Synergy system is a portable, battery operated, medical device housed within an ABS enclosure. The system combines the single-channel EEG recording capability of the Zmachine DT-100 (K101830) with the respiratory signal recording capability of the Resmed ApneaLink Air (K143272). The Zmachine Synergy is thereby capable of recording EEG, respiratory airflow, respiratory effort, blood oxygen saturation, pulse rate, and body position during sleep.

    AI/ML Overview

    The provided text focuses on establishing substantial equivalence for the Zmachine Synergy device by comparing it to predicate devices (Zmachine DT-100 and ApneaLink Air) through technical and bench testing. It does not contain information about clinical study data, ground truth establishment by experts, or MRMC studies that would typically be associated with AI/algorithm performance claims.

    Therefore, many of the requested categories cannot be fully addressed based on the provided document.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes "bench comparison testing" where the Zmachine Synergy was compared against predicate devices. The "acceptance criteria" are implied to be "substantially equivalent" or "high agreement" between the Zmachine Synergy and the predicate devices for each measured parameter.

    TestAcceptance Criteria (Implied)Reported Device Performance
    EEGSubstantially equivalent EEG amplifier characteristics.High agreement with design limits and each other for amplifier gain, highpass/lowpass filter cutoff frequency, DC offset, and noise floor. Found substantially equivalent.
    Respiratory AirflowStrong linear relationship / substantially equivalent signals.Pearson's correlation coefficient revealed a strong linear relationship. Found substantially equivalent.
    Respiratory EffortStrong linear relationship / substantially equivalent signals.Pearson's correlation coefficient revealed a strong linear relationship. Found substantially equivalent.
    Pulse OximetryHigh agreement and low mean squared error.Heart rate and oxygen saturation readings were in high agreement with calibrator output levels and showed low mean squared error when comparing the two systems. Found substantially equivalent.
    Body PositionVery high agreement with angular reference.Angular readings and angular reference positions were in very high agreement throughout 360 degrees of rotation. Found substantially equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    The study described is bench testing, not a clinical study involving patients or a test set of data in the typical sense for algorithm performance. The "test set" consisted of:

    • A multi-channel EEG analog playback system for EEG.
    • A variable pressure air pump for Respiratory Airflow.
    • Controlled belt stretching and relaxing against a linear scale for Respiratory Effort.
    • A patient simulator for Pulse Oximetry.
    • Rotation against an angular reference for Body Position.

    Therefore, traditional "sample size" is not applicable, as these were controlled bench tests. Data provenance is specific to the synthetic signals generated by the test equipment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. The "ground truth" for these bench tests was established by the precise outputs of the test equipment (e.g., specific frequencies, known pressure values, specific angles) and the design specifications of the devices themselves. No human expert adjudication was involved in establishing this "ground truth."

    4. Adjudication Method for the Test Set

    Not applicable. There was no human adjudication as the tests were performed against known physical inputs from laboratory equipment.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. The provided text does not describe an MRMC study. The study focuses on comparing the fundamental signal acquisition characteristics of the Zmachine Synergy with its predicate devices at a technical, hardware level, not on assessing how human readers perform with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Partially applicable, but for hardware performance, not an algorithm's diagnostic output. The "standalone" performance described is the ability of the Zmachine Synergy's hardware components to accurately acquire and record various physiological signals, as compared to predicate devices or known inputs, without human interpretation of derived diagnostic information. The device is an "EEG and respiratory signal recorder" intended to "aid in the diagnosis of sleep disorders," implying subsequent human interpretation of the recorded signals. The described tests confirm the accuracy of the recording capabilities, which is a form of standalone performance for the signal acquisition hardware. Specific algorithm performance for detecting sleep disorders based on these signals is not detailed.

    7. The Type of Ground Truth Used

    The ground truth used was synthetic signals from laboratory equipment and design specifications/known outputs.

    • For EEG: Broad spectrum shaped white noise and zero-level output signal.
    • For Respiratory Airflow: Stepped variable air pressure signal.
    • For Respiratory Effort: Controlled belt stretching and relaxing against a linear scale.
    • For Pulse Oximetry: Specified calibrator output levels from a patient simulator.
    • For Body Position: An angular reference the device was rotated against.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a pre-market notification for a medical device (Zmachine Synergy), focusing on demonstrating substantial equivalence through technical bench testing of hardware components. It does not involve a training set for an AI or machine learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set mentioned or implied for an AI/ML algorithm within this document, the method of establishing its ground truth is irrelevant here.