Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K232732

    Validate with FDA (Live)

    Date Cleared
    2024-01-17

    (132 days)

    Product Code
    Regulation Number
    862.3870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid Marijuana (THC) Test Strip 20 and Rapid Marijuana (THC) Test Dipcard 20 is a rapid, screening test for the qualitative detection of Marijuana and their metabolites in human urine at the cut-off concentration of 20 ng/mL.

    Rapid Marijuana (THC) Test Strip 50 and Rapid Marijuana (THC) Test Dipcard 50 is a rapid, screening test for the qualitative detection of Marijuana and their metabolites in human urine at the cut-off concentration of 50 ng/mL.

    The tests contain two formats:1) Test Strip and 2) Test Dipcard. The tests are intended for in vitro diagnostics use. They are intended for over-the-counter use.

    The tests provide only a preliminary result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

    Device Description

    Rapid Marijuana (THC) Test Strip 20 and Rapid Marijuana (THC) Test Dipcard 20 are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Marijuana and their metabolites at or above the cut-off concentration of 20 ng/mL. The tests can be performed without the use of an instrument.

    Rapid Marijuana (THC) Test Strip 50 and Rapid Marijuana (THC) Test Dipcard 50 are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Marijuana and their metabolites at or above the cut-off concentration of 50 ng/mL. The tests can be performed without the use of an instrument.

    Test Strip and Test Dipcard use identical test strips made with same chemical formulation and manufacturing procedures.

    AI/ML Overview

    This document describes the performance of the Rapid Marijuana (THC) Test Strip and Dipcard devices, available in 20 ng/mL and 50 ng/mL cutoff concentrations, for the qualitative detection of Marijuana and its metabolites in human urine.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of specific performance metrics (e.g., minimum sensitivity, specificity, or agreement percentages for accuracy studies). However, the performance data presented implies a standard of acceptable qualitative detection around the cutoff concentration. The precision and accuracy studies evaluate the device's ability to correctly identify positive and negative samples at various concentrations relative to the cutoff. The home-use study assesses user comprehension and ease of use.

    Since no explicit numerical acceptance criteria are provided, I will present key performance indicators from the precision and accuracy studies. For the purpose of this table, I will infer that a high percentage of correct classifications at and around the cutoff, along with 100% correct classification for samples significantly above or below the cutoff, would be considered acceptable.

    Inferred Performance Acceptance Criteria & Reported Device Performance for Cut-off 20 ng/mL:

    Performance MetricImplied Acceptance Criteria (Inferred from data presentation)Rapid Marijuana (THC) Test Strip 20 Reported Performance (Positive/Negative counts at specified concentrations across 3 lots)Rapid Marijuana (THC) Test Dipcard 20 Reported Performance (Positive/Negative counts at specified concentrations across 3 lots)
    Precision StudyFor -75% Cutoff (5 ng/mL): 100% Negative calls or very closeLot 1: 0 positive, 60 negativeLot 2: 0 positive, 60 negativeLot 3: 0 positive, 60 negativeLot 1: 0 positive, 60 negativeLot 2: 0 positive, 60 negativeLot 3: 0 positive, 60 negative
    For -50% Cutoff (10 ng/mL): 100% Negative calls or very closeLot 1: 0 positive, 60 negativeLot 2: 0 positive, 60 negativeLot 3: 0 positive, 60 negativeLot 1: 0 positive, 60 negativeLot 2: 0 positive, 60 negativeLot 3: 0 positive, 60 negative
    For -25% Cutoff (15 ng/mL): Predominantly Negative calls, some positive possibleLot 1: 6 positive, 54 negativeLot 2: 4 positive, 56 negativeLot 3: 4 positive, 56 negativeLot 1: 4 positive, 56 negativeLot 2: 6 positive, 54 negativeLot 3: 4 positive, 56 negative
    For Cutoff (20 ng/mL): Mix of Positive and Negative calls (around 50% for each)Lot 1: 34 positive, 26 negativeLot 2: 36 positive, 24 negativeLot 3: 32 positive, 28 negativeLot 1: 36 positive, 24 negativeLot 2: 38 positive, 22 negativeLot 3: 34 positive, 26 negative
    For +25% Cutoff (25 ng/mL): Predominantly Positive calls, some negative possibleLot 1: 54 positive, 6 negativeLot 2: 56 positive, 4 negativeLot 3: 58 positive, 2 negativeLot 1: 56 positive, 4 negativeLot 2: 56 positive, 4 negativeLot 3: 58 positive, 2 negative
    For +50% Cutoff (30 ng/mL): 100% Positive calls or very closeLot 1: 60 positive, 0 negativeLot 2: 60 positive, 0 negativeLot 3: 60 positive, 0 negativeLot 1: 60 positive, 0 negativeLot 2: 60 positive, 0 negativeLot 3: 60 positive, 0 negative
    For +75% Cutoff (35 ng/mL): 100% Positive calls or very closeLot 1: 60 positive, 0 negativeLot 2: 60 positive, 0 negativeLot 3: 60 positive, 0 negativeLot 1: 60 positive, 0 negativeLot 2: 60 positive, 0 negativeLot 3: 60 positive, 0 negative
    For +100% Cutoff (40 ng/mL): 100% Positive calls or very closeLot 1: 60 positive, 0 negativeLot 2: 60 positive, 0 negativeLot 3: 60 positive, 0 negativeLot 1: 60 positive, 0 negativeLot 2: 60 positive, 0 negativeLot 3: 60 positive, 0 negative
    Accuracy StudyOverall Agreement: High agreement with LC/MS, especially for samples far from cutoffSite 1,2,3 (Agreement with LC/MS categories):- Neg. (drug free): 27 Negative- Neg. (<-50% cutoff): 5 Negative- Near cutoff neg (-50% to cutoff): 7 Negative, 1 Positive (discordant)- Near cutoff pos (cutoff to +50% cutoff): 1 Negative (discordant), 8 Positive- Pos. (>+50% cutoff): 31 PositiveSite 1,2,3 (Agreement with LC/MS categories):- Neg. (drug free): 27 Negative- Neg. (<-50% cutoff): 5 Negative- Near cutoff neg (-50% to cutoff): 7 Negative, 1 Positive (discordant)- Near cutoff pos (cutoff to +50% cutoff): 1 Negative (discordant), 8 Positive- Pos. (>+50% cutoff): 31 Positive
    Home Use Consumer StudyAgreement at 0 ng/mL: 100% Negative100% Agreement (30 Negative)100% Agreement (30 Negative)
    Agreement at -50% Cutoff (10 ng/mL): 100% Negative100% Agreement (30 Negative)100% Agreement (30 Negative)
    Agreement at -25% Cutoff (15 ng/mL): High percentage of Negative (e.g.,>90%)93% Agreement (2 Positive, 28 Negative)90% Agreement (3 Positive, 27 Negative)
    Agreement at +25% Cutoff (25 ng/mL): High percentage of Positive (e.g.,>90%)90% Agreement (27 Positive, 3 Negative)93% Agreement (28 Positive, 2 Negative)
    Agreement at +50% Cutoff (30 ng/mL): 100% Positive100% Agreement (30 Positive)100% Agreement (30 Positive)
    Agreement at +100% Cutoff (40 ng/mL): 100% Positive100% Agreement (30 Positive)100% Agreement (30 Positive)

    Inferred Performance Acceptance Criteria & Reported Device Performance for Cut-off 50 ng/mL:

    Performance MetricImplied Acceptance Criteria (Inferred from data presentation)Rapid Marijuana (THC) Test Strip 50 Reported Performance (Positive/Negative counts at specified concentrations across 3 lots)Rapid Marijuana (THC) Test Dipcard 50 Reported Performance (Positive/Negative counts at specified concentrations across 3 lots)
    Precision StudyFor -75% Cutoff (12.5 ng/mL): 100% Negative calls or very closeLots 1,2,3: 0 positive, 60 negativeLots 1,2,3: 0 positive, 60 negative
    For -50% Cutoff (25 ng/mL): 100% Negative calls or very closeLots 1,2,3: 0 positive, 60 negativeLots 1,2,3: 0 positive, 60 negative
    For -25% Cutoff (37.5 ng/mL): Predominantly Negative calls, some positive possibleLot 1: 4 positive, 56 negativeLot 2: 6 positive, 54 negativeLot 3: 2 positive, 58 negativeLot 1: 6 positive, 54 negativeLot 2: 2 positive, 58 negativeLot 3: 6 positive, 54 negative
    For Cutoff (50 ng/mL): Mix of Positive and Negative calls (around 50% for each)Lot 1: 36 positive, 24 negativeLot 2: 38 positive, 22 negativeLot 3: 34 positive, 26 negativeLot 1: 38 positive, 22 negativeLot 2: 34 positive, 26 negativeLot 3: 36 positive, 24 negative
    For +25% Cutoff (62.5 ng/mL): Predominantly Positive calls, some negative possibleLot 1: 56 positive, 4 negativeLot 2: 56 positive, 4 negativeLot 3: 58 positive, 2 negativeLot 1: 58 positive, 2 negativeLot 2: 56 positive, 4 negativeLot 3: 54 positive, 6 negative
    For +50% Cutoff (75 ng/mL): 100% Positive calls or very closeLots 1,2,3: 60 positive, 0 negativeLots 1,2,3: 60 positive, 0 negative
    For +75% Cutoff (87.5 ng/mL): 100% Positive calls or very closeLots 1,2,3: 60 positive, 0 negativeLots 1,2,3: 60 positive, 0 negative
    For +100% Cutoff (100 ng/mL): 100% Positive calls or very closeLots 1,2,3: 60 positive, 0 negativeLots 1,2,3: 60 positive, 0 negative
    Accuracy StudyOverall Agreement: High agreement with LC/MS, especially for samples far from cutoffSite 1,2,3 (Agreement with LC/MS categories):- Neg. (drug free): 21 Negative- Neg. (<-50% cutoff): 10 Negative- Near cutoff neg (-50% to cutoff): 8 Negative, 1 Positive (discordant)- Near cutoff pos (cutoff to +50% cutoff): 1 Negative (discordant), 9 Positive- Pos. (>+50% cutoff): 30 PositiveSite 1,2,3 (Agreement with LC/MS categories):- Neg. (drug free): 21 Negative- Neg. (<-50% cutoff): 10 Negative- Near cutoff neg (-50% to cutoff): 8 Negative, 1 Positive (discordant)- Near cutoff pos (cutoff to +50% cutoff): 1 Negative (discordant), 9 Positive- Pos. (>+50% cutoff): 30 Positive
    Home Use Consumer StudyAgreement at 0 ng/mL: 100% Negative100% Agreement (30 Negative)100% Agreement (30 Negative)
    Agreement at -50% Cutoff (25 ng/mL): 100% Negative100% Agreement (30 Negative)100% Agreement (30 Negative)
    Agreement at -25% Cutoff (37.5 ng/mL): High percentage of Negative (e.g.,>90%)97% Agreement (1 Positive, 29 Negative)93% Agreement (2 Positive, 28 Negative)
    Agreement at +25% Cutoff (62.5 ng/mL): High percentage of Positive (e.g.,>90%)93% Agreement (28 Positive, 2 Negative)93% Agreement (28 Positive, 2 Negative)
    Agreement at +50% Cutoff (75 ng/mL): 100% Positive100% Agreement (30 Positive)100% Agreement (30 Positive)
    Agreement at +100% Cutoff (100 ng/mL): 100% Positive100% Agreement (30 Positive)100% Agreement (30 Positive)

    2. Sample Size Used for the Test Set and the Data Provenance

    • Precision Study:
      • Sample Size: For each device type (Strip 20, Dipcard 20, Strip 50, Dipcard 50), 3 lots were tested. Each lot involved 60 determinations for each of 9 concentration levels, across 6 operators at 3 Point-of-Care sites over 10 non-consecutive days. This totals to 1620 observations per device type in the precision study.
      • Data Provenance: The document does not explicitly state the country of origin for the urine samples. The study involved spiking drug-free urine specimens. The study itself was conducted at 3 Point-of-Care sites, but their locations are not specified. It is a prospective study as samples were prepared and tested specifically for this evaluation.
    • Accuracy Study:
      • Sample Size: 80 clinical urine specimens were used for each device type (Strip 20, Dipcard 20, Strip 50, Dipcard 50) and each operator site. Since there were 3 operator sites, this amounts to 80 * 3 = 240 tests (device results vs. LC/MS) for each specificity (20 ng/mL and 50 ng/mL).
      • Data Provenance: The samples were described as "clinical urine specimens," implying they were collected from human subjects. The country of origin is not specified, and it's unclear if they were retrospective or prospectively collected for the study.
    • Home Use Consumer Study:
      • Sample Size: 180 lay users participated in the study. Each user performed 1 test on a provided specimen. The samples were prepared for the study. For each device type and concentration category, 30 determinations were made.
      • Data Provenance: The document does not specify the country of origin of the lay users or the urine samples used. It is a prospective study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Precision Study: The ground truth for the spiked samples was established by the precise concentration of "-)-11-nor-9-Carboxy-Δ⁹-THC" added to drug-free urine, confirmed with LC/MS. No human experts were involved in establishing this ground truth, as it was based on laboratory preparation and analytical confirmation.
    • Accuracy Study: The ground truth for the clinical urine specimens was established by LC/MS (Gas Chromatography/Mass spectrometry or Liquid chromatography/Mass spectrometry). LC/MS is described as the "preferred confirmatory method." LC/MS is an analytical chemistry technique, not a human expert. Therefore, no human experts directly established the ground truth in this study; it was based on objective analytical measurement.
    • Home Use Consumer Study: Similar to the precision study, the ground truth for samples was established by spiking drugs into urine with concentrations confirmed by LC/MS. No human experts were involved in establishing this ground truth.

    4. Adjudication Method for the Test Set

    • Precision Study: The results were read by operators comparing the test line intensity to the control line. There is no mention of an adjudication process for discrepancies in reading within the precision study; each determination (positive/negative) was recorded independently.
    • Accuracy Study: The device results were compared directly to the LC/MS results. Discordant results are noted in tables (e.g., device positive when LC/MS was negative below cutoff, or device negative when LC/MS was positive above cutoff), but no formal adjudication method (like 2+1 reading) is described beyond this direct comparison.
    • Home Use Consumer Study: The "Layer user Results" likely represent the interpretation of the test by the lay users themselves. No specific adjudication method is mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study done. The device is a rapid, screening immunoassay test (similar to a pregnancy test) for qualitative detection of THC metabolites in urine, designed for Over-The-Counter use. It is interpreted visually by the user or an operator. There is no AI component in this device, nor is there a comparison of human readers with vs. without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a lateral flow immunochromatographic assay requiring human visual interpretation. It is not an algorithm-only device. The results are read visually.

    7. The Type of Ground Truth Used

    • Cross-reactivity and Interference Studies: Ground truth was established by precisely preparing samples with known concentrations of various compounds and expected drug-free or drug-spiked urine.
    • Precision Study: Ground truth for the concentration levels of (-)-11-nor-9-Carboxy-Δ⁹-THC was confirmed using LC/MS.
    • Accuracy Study: Ground truth for clinical urine specimens was established by LC/MS (Gas Chromatography/Mass spectrometry or Liquid chromatography/Mass spectrometry).
    • Home Use Consumer Study: Ground truth for the spiked samples was confirmed using LC/MS.

    In summary, the primary ground truth method for quantitative confirmation of drug concentrations in samples was LC/MS.

    8. The Sample Size for the Training Set

    This information is not applicable to this device. As a rapid, screening immunoassay test, it does not involve machine learning or AI models that require a "training set." Its performance is based on the chemical reactions and visual detection inherent in the lateral flow assay.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for this device type.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153050

    Validate with FDA (Live)

    Date Cleared
    2016-04-26

    (189 days)

    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test Dipcard is a rapid drug screening test designed to qualitatively detect the presence of drugs and drug metabolites in human urine at the following cut-off concentrations:

    TestCalibratorCut-off level
    Marijuana (THC)Delta-9-THC-COOH50 ng/mL
    Cocaine (COC)Benzoylecgonine300 ng/mL
    Amphetamine (AMP)D-Amphetamine1000 ng/mL
    Methamphetamine (MET)D-Methamphetamine1000 ng/mL
    Morphine 2000 (MOP)Morphine2000 ng/mL
    Barbiturates (BAR)Secobarbital300 ng/mL
    Benzodiazepines (BZO)Oxazepam300 ng/mL
    Methylenedioxymethamphetamine (MDMA)3,4-Methylenedioxymethamphetamine500 ng/mL
    Methadone (MTD)Methadone300 ng/mL
    Oxycodone (OXY)Oxycodone100 ng/mL
    Phencyclidine (PCP)Phencyclidine25 ng/mL
    Buprenorphine(BUP)Buprenorphine10ng/mL
    2-ethylidene-1, 5-dimethyl-3,2-ethylidene-1, 5-dimethyl-3,300 ng/mL
    3-diphenylpyrrolidine(EDDP)3-diphenylpyrrolidine
    Morphine(MOP300)Morphine300 ng/mL
    Propoxyphene(PPX)Propoxyphene300 ng/mL
    Tri-cyclic Antidepressants (TCA)Nortriptyline1000 ng/mL

    The tests contain two formats:1) Test Cup and 2) Test Dipcard. The tests may be configured as single drug tests in any combination of the drug analytes listed in the table above up to a maximum of 15 analytes. Only one cutoff concentration will be included per analyte per device. The tests are intended for in vitro diagnostics use. They are intended for prescription use including point of care sites and over-the-counter use.

    The tests will vield preliminary positive results when prescription drugs Barbiturates. Benzodiazenine, Methadone, Propoxyphene or Tricyclic Antidepressants are ingested, even at or above there are no unformly recognized drug levels for Barbiturates, Buprenorphine, Benzodiazepine, Propoxyphene and Tricyclic Antidepressants in urine.

    This assay provides only a preliminary analytical test result. Gas Chromatography Mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result. particularly when preliminary positive results are indicated.

    Device Description

    Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test Dipcard are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Buprenorphine, Benzodiazepines, Cocaines, 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine, Methylenedioxymethamphetamine, Methamphetamine, Morphine300, Morphine2000, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Tri-cyclic Antidepressants, Marijuana and their metabolites ( specifically THC ) at or above the cut-off levels as indicated. The tests can be performed without the use of an instrument. Test Cup and Test Dipcard use identical test strips made with same chemical formulation and manufacturing procedures.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study findings for the "Rapid Single/Multi-drug Test Cup, Rapid Single/Multi-drug Test Dipcard" device, based solely on the provided text:

    This document is a 510(k) Premarket Notification summary to the FDA, indicating that the device is seeking clearance based on substantial equivalence to a legally marketed predicate device, not necessarily showcasing comprehensive clinical trials for novel devices. Therefore, the "study" described focuses on demonstrating equivalence rather than proving novel clinical utility with large-scale, multi-reader, or outcomes-based studies.


    Acceptance Criteria and Reported Device Performance

    The core acceptance criteria are based on the ability to qualitatively detect specific drugs and their metabolites in human urine at predefined cut-off concentrations. The performance data demonstrates agreement with these cut-off concentrations, indicating the device meets the intended detection capability.

    Table of Acceptance Criteria and Reported Device Performance (Specificity and Sensitivity implicitly)

    For each drug, the acceptance criteria are to correctly identify negative samples and positive samples relative to the specified cut-off. The performance is reported in terms of the number of positive/negative results at various concentrations around these cut-offs during precision and accuracy studies.

    Precision Study (Example for AMP, BAR, BUP, BZO, COC, EDDP, MDMA, MET, MOP300, MOP2000, MTD, OXY, PCP, PPX, TCA, THC)

    The precision study aimed for accurate detection at concentrations:

    • 0 ng/mL: All negative
    • -75% cutoff: All negative or predominantly negative
    • -50% cutoff: All negative or predominantly negative
    • -25% cutoff: Mixture of positive and negative, demonstrating sensitivity around the cut-off
    • Cutoff: Mixture of positive and negative, demonstrating sensitivity around the cut-off
    • +25% cutoff: Mixture of positive and negative, demonstrating sensitivity around the cut-off
    • +50% cutoff, +75% cutoff, +100% cutoff: All positive or predominantly positive

    Observed Performance (Summarized from Precision Tables)

    For all drugs tested (AMP, BAR, BUP, BZO, COC, EDDP, MDMA, MET, MOP300, MOP2000, MTD, OXY, PCP, PPX, TCA, THC) across three lots for both the Test Cup and Test Dipcard:

    • At 0 ng/mL, -75% cutoff, and -50% cutoff: All (or overwhelmingly almost all) reported results were negative. This indicates good specificity below the cutoff.
    • At +50% cutoff, +75% cutoff, and +100% cutoff: All (or overwhelmingly almost all) reported results were positive. This indicates good sensitivity above the cutoff.
    • At -25% cutoff, cutoff, and +25% cutoff: A mixed distribution of positive and negative results was consistently observed, demonstrating the device's ability to differentiate samples around the cut-off concentration, as expected for a qualitative assay. For instance, at the exact cutoff, typically around 30-40 positive and 20-30 negative results were seen out of 60 determinations per lot, for most analytes.

    Accuracy Study (Clinical Specimens - Example for AMP, BAR, BUP, BZO, COC, EDDP, MDMA, MET, MOP300, MOP2000, MTD, OXY, PCP, PPX, TCA, THC)

    The accuracy study categorizes samples into five groups based on GC/MS or HPLC reference:

    • Drug free
    • Less than half the cutoff
    • Near cutoff negative (between 50% below cutoff and cutoff)
    • Near cutoff positive (between cutoff and 50% above cutoff)
    • High positive (greater than 50% above cutoff)

    The acceptance criteria here are that results from the Rapid Multi-drug Test Cup/Dipcard should align with the GC/MS or HPLC classifications.

    Observed Performance (Summarized from Accuracy Tables)

    For both the Test Cup and Test Dipcard:

    • Drug-free, Less than half the cutoff, Near Cutoff Negative samples: Overwhelmingly resulted in negative reports from the device. A very small number of "false positives" (e.g., 1 or 2 positive results in the "Near Cutoff Negative" category for some drugs) were noted, and some "false negatives" for Buprenorphine, Morphine, Methadone, Oxycodone, TCA, and THC in the "Near Cutoff Negative" or "Near Cutoff Positive" categories when compared to the reference methods. The "Analysis of Discordant Results" section details these specific discrepancies.
    • Near Cutoff Positive and High Positive samples: Overwhelmingly resulted in positive reports from the device. Similar to above, a very small number of discordant results (e.g., some "Near Cutoff Positive" samples were reported negative by the device) were observed, which are explicitly detailed in the "Analysis of Discordant Results" tables.

    Overall Conclusion on Acceptance Criteria: The studies presented generally demonstrate that the device performs as expected for a qualitative immunoassay, showing good agreement with GC/MS/HPLC results for samples significantly above or below the cut-off. The mixed results near the cut-off are typical for such tests.


    Study Details

    1. Sample size used for the test set and the data provenance:

      • Precision Study: For each drug, there were 9 concentrations tested. For each concentration, 60 determinations were made (3 lots x 3 runs/day x 10 days = 90 determinations per concentration, but the table indicates 60 determinations per lot for each concentration, which suggests (2 operators * 3 aliquots tested per lot per day = 6 determinations per lot per day) * 10 days leads to 60 determinations per lot). With 16 drugs, this means roughly 16 x 9 x 60 = 8,640 determinations for the precision study (excluding the implicit number of lots).
      • Accuracy Study (Clinical Samples): 80 clinical urine specimens per drug were used for both the Multi-drug Test Cup and Multi-drug Test Dipcard. With 16 drugs, this amounts to 16 x 80 = 1,280 samples per device format (Cup and Dipcard), so a total of 2,560 unique clinical specimens.
        • Data Provenance: The document does not specify the country of origin but states "Clinical urine specimens." It also states these were "retrospective" as they were analyzed by GC/MS/HPLC prior to testing with the device, and then retrospectively analyzed against the device results. The absence of details on where these clinical samples were collected typically means their origin is unknown or not considered relevant for this type of submission (often laboratory-sourced for drug testing evaluations).
      • Lay User Study: 1720 lay users participated. Each participant performed 1 test. Urine samples were prepared at 0, +/- 50% cutoff, +/- 25% cutoff, and +100% cutoff. This study does not explicitly state the number of unique urine samples used but rather the number of user interactions with a test sample. For each drug and each concentration, there were 10 or 30 or 360 determinations across two lots (one for Cup, one for Dipcard, it seems implied).
        • Data Provenance: Not specified, but involved "lay users" performing tests, implying a prospective collection for this specific study.
      • Interference Study: For each drug, 4 concentrations (0, -50% cutoff, +50% cutoff, +100% cutoff), 3 lots, and 15 determinations per lot per concentration. This amounts to 16 x 4 x 3 x 15 = 2,880 determinations.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Ground Truth Method: For the accuracy studies, the ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS) or High-Performance Liquid Chromatography (HPLC). These are analytical chemistry "gold-standard" methods, not expert human interpretation.
      • Number/Qualifications of Experts: Not applicable in the context of GC/MS or HPLC as ground truth. The results are based on analytical measurements, not human interpretation of images or observations.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • No formal adjudication method is described for establishing ground truth, as the ground truth relies on quantitative analytical methods (GC/MS, HPLC) rather than subjective human assessment.
      • For the precision study, results were based on multiple determinations (e.g., 60 per lot per concentration), and for the lay user study, the "agreement" was calculated based on the collective results of the lay users against the known spiked concentrations. Discordant results in the accuracy study are individually listed, implying a comparison against the GC/MS/HPLC result without further adjudication panels described for the device's outcome.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was performed or described. This device is a rapid diagnostic test (Test Cup/Dipcard) for drug screening, not an AI-assisted diagnostic imaging device. Therefore, the concept of "human readers improving with AI vs without AI assistance" does not apply to this specific device. The closest equivalent is the "Lay User Study," which assesses how well lay users can interpret the test independently using the provided instructions. Their agreement rates for correctly classifying samples are provided.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, in essence, standalone performance was assessed in the "Precision" and "Accuracy" studies if "standalone" is interpreted as the device's intrinsic ability to detect drugs in controlled and clinical samples when read by trained operators. The device itself (the dipcard or cup) produces a visual qualitative result (lines appearing or not appearing), which is then read. The "precision study" and "accuracy study" data (Table 8.5, 8.6) represent this "algorithm only" (device only) performance when operated and interpreted by presumably trained personnel. The "Lay User Study" (Table 8.7) assesses reading by an untrained human-in-the-loop, demonstrating usability rather than optimal device performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth for the accuracy studies was established by analytical chemistry methods: Gas Chromatography/Mass Spectrometry (GC/MS) or High-Performance Liquid Chromatography (HPLC). These methods provide quantitative measurements of drug concentrations, which are considered the definitive "ground truth" for drug screening device validation. For spiked samples in precision, lay user, and interference studies, the ground truth was the known spiked concentration of the drug.
    7. The sample size for the training set:

      • This document describes a 510(k) submission, which is for a traditional in-vitro diagnostic device (a rapid test cup/dipcard with reagent strips). There is no "training set" in the context of machine learning or AI models, as this device does not utilize such technologies. The studies described are for validation and performance assessment of the physical chemical assay.
    8. How the ground truth for the training set was established:

      • As there is no "training set" in the AI/ML sense, this question is not applicable to the device described. The studies involve testing the device against samples with known drug concentrations (either spiked or determined by reference analytical methods) to evaluate its performance.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142800

    Validate with FDA (Live)

    Date Cleared
    2014-12-11

    (73 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid Single/Multi-drug test Cup and Rapid Single/Multi-drug test Dipcard are lateral flow chromatographic immunoassays designed to qualitatively detect the presence of drug metabolites in human urine at the following cut-off concentrations:

    TestCalibratorCut-off level
    Marijuana (THC)Delta-9-THC-COOH50ng/mL
    Cocaine (COC)Benzolecgonine300ng/mL
    Amphetamine (AMP)D-Amphetamine1000ng/mL
    Methamphetamine (MET)D-Methamphetamine1000ng/mL
    Morphine 2000 (MOP)Morphine2000ng/mL
    Barbiturates (BAR)Secobarbital300ng/mL
    Benzodiazepines (BZO)Oxazepam300ng/mL
    Methylenedioxymethamphetamine (MDMA)3,4-Methylenedioxymethamphetamine500ng/mL
    Methadone (MTD)Methadone300ng/mL
    Oxycodone (OXY)Oxycodone100ng/mL
    Phencyclidine (PCP)Phencyclidine25ng/mL

    The tests contain two formats: 1) Test Cup and 2) Test Dipcard. The tests may be configured as single drug tests or multiple drug tests in any combination of the drug analytes listed in the table above. These tests are intended for in vitro diagnostics use. They are intended for prescription use including point of care sites and over-the-counter use. The tests will yield preliminary positive results when prescription drugs Barbiturates, Benzodiazepine, and Methadone are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturates and Benzodiazepine in urine.

    The assays provide only a preliminary analytical test result. Gas Chromatography/Mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

    Device Description

    Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test Dipcard are competitive binding, lateral flow immunochromatographic assays for the qualitative detection of Amphetamine, Cocaine, Marijuana, Methamphetamine, Barbiturates, Benzodiazepines, Methylenedioxymethamphetamine, Methadone, Oxycodone, Phencyclidine and their metabolites ( specifically THC ) at or above the cut-off levels as indicated. The tests are performed without the use of an instrument.

    The test cup and test dipcard formats use identical test strips made with the same chemical formulation and manufacturing procedures.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Rapid Single/Multi-drug Test Cup and Dipcard, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for precision or accuracy studies in a quantitative manner (e.g., "must achieve X% agreement"). However, given the nature of the data presented, the implied acceptance criterion for precision is that the device should reliably identify samples at or above the cutoff as positive, and samples significantly below the cutoff as negative, with a reasonable degree of accuracy for concentrations near the cutoff. For accuracy, the implied acceptance criterion is that the device should demonstrate high agreement with GC/MS analysis. For lay user readability, the implied criterion is high user comprehension.

    Let's synthesize the data into a table based on these implied criteria.

    Note on "Reported Device Performance": The document provides raw numbers of positive/negative results at different concentration levels for precision and individual count breakdowns for accuracy against GC/MS. To create a concise table, I will report the general trends and highlight key metrics. Specific percentages are primarily for the lay-user study.

    Acceptance Criteria CategorySpecific Metric (Implied)Reported Device Performance (Summary)
    Precision (for each drug)Reliable detection: - 0% false positives for negative samples - 0% false negatives for high positive samples - Consistent results at +/- 25% cutoff levelsAcross all drugs and lots (single/multi-drug, cup/dipcard formats): - Negative samples (0 ng/mL, -75% & -50% cutoff): Universally reported as 0 positive, 60 negative (100% accurate negative). - High positive samples (+50%, +75%, +100% cutoff): Universally reported as 60 positive, 0 negative (100% accurate positive). - Near Cutoff (-25% & +25% cutoff): Showed variability (e.g., AMP -25% cutoff ranged 4-10 positives out of 60, +25% cutoff ranged 50-58 positives out of 60). This variability near cutoff is expected for qualitative tests.
    Accuracy (vs. GC/MS)High agreement between device result and GC/MS for all concentration categories. Minimal discordant results.For clinical specimens: - Drug free, less than half cutoff, and high positive: Generally 100% agreement with GC/MS. - Near Cutoff Negative and Near Cutoff Positive: Showed some mixed results (expected for qualitative tests near cutoff). - Discordant Results: A few specific cases reported where the device result (e.g., positive) did not match GC/MS (e.g., device positive for MET at 867ng/mL vs. 1000ng/mL cutoff, device negative for THC at 61ng/mL vs. 50ng/mL cutoff). These are mostly near the cutoff.
    InterferenceNo interference from common substances found in urine at 100µg/mL.None of the tested compounds (listing many common drugs and substances) were shown to interfere with the device's performance when added to drug-free or drug-positive urine.
    pH EffectDevice performance unaffected by urinary pH range of 3 to 9.The results demonstrated that varying ranges of pH (3 to 9 in 1 pH unit increments) do not interfere with the test's performance when spiked at 50% below and 50% above cutoff levels.
    Specific Gravity EffectDevice performance unaffected by urinary specific gravity range (1.002-1.040).The results demonstrated that varying ranges of urinary specific gravity (1.002, 1.010, 1.020, 1.030, 1.040) do not affect the test result when spiked at 50% below and 50% above cutoff.
    Lay User StudyHigh agreement with expected results and high user comprehension for home use.Agreement: For lay users, agreement percentages were high, generally 100% for clear negatives and clear positives, and ranging from 83% to 97% for samples near the +/- 25% cutoff for various drugs and test formats. Readability/Ease of Use: 100% of users felt the test was easy to use and the instructions were easy to understand. The labeling was assessed at a 6th-grade reading level.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Study:
      • Sample Size: 60 determinations per lot per concentration (Total: 9 concentrations * 60 determinations/lot * 3 lots = 1620 observations for each drug type). This was for both single drug and multi-drug formats, and both cup and dipcard.
      • Data Provenance: Not explicitly stated, but based on the overall document, it is likely internal laboratory testing or conducted at the specified Point-of-Care sites, implying prospective. The "nonparticipant" blinding suggests a controlled, prospective design.
    • Accuracy Study (Clinical Urine Specimens):
      • Sample Size: 80 clinical urine specimens for each drug type. These were divided into five categories (drug free, less than half cutoff, near cutoff negative, near cutoff positive, high positive).
      • Data Provenance: "Clinical urine specimens" suggests real-world samples. The study was conducted by nurses at two Point-of-Care sites, indicating prospective, real-world data collection using the device.
    • Home Use Consumer Study:
      • Sample Size: 360 lay users. Urine samples were prepared at various concentrations (0, +/- 50% cutoff, +/- 25% cutoff, +100% cutoff). Each participant performed 1 test on a provided specimen.
      • Data Provenance: Prospective, controlled study with lay users using prepared samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Precision Study: Ground truth was established by spiking known concentrations of drugs into drug-free urine. The concentrations were "confirmed with GC/MS." No human experts were used for ground truth directly for precision.
    • Accuracy Study (Clinical Urine Specimens): The ground truth for the clinical urine specimens was established by Gas Chromatography/Mass Spectrometry (GC/MS) analysis. This is an analytical chemistry technique, not human expert interpretation.
    • Home Use Consumer Study: Ground truth was established by spiking known concentrations of drugs into drug-free urine. The concentrations were "confirmed with GC/MS." No human experts were used for ground truth directly for the lay user study's analytical performance.

    4. Adjudication Method for the Test Set

    • Precision Study: The samples were "blindly labeled by a nonparticipant" and "randomized prior to testing." Results were qualitative (Positive/Negative) based on the device's output. No explicit multi-reader adjudication method (e.g., 2+1) is mentioned; it appears to be a direct comparison of the device's qualitative result to the known spiked concentration.
    • Accuracy Study (Clinical Urine Specimens): "All samples were blindly labeled by a nonparticipant" and "randomized prior to testing." The device results were compared directly to the GC/MS analysis. No multi-reader adjudication is mentioned for the device's interpretation or the GC/MS results.
    • Home Use Consumer Study: Each participant performed one test and filled out a questionnaire. The device's qualitative result (positive/negative) from the lay user was compared against the known spiked concentration. No explicit multi-reader adjudication is mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This device is a qualitative in-vitro diagnostic (drug test cup/dipcard) that provides a direct positive or negative result, read by an individual (either a healthcare professional or a lay user). It is not an imaging AI device that assists human readers in interpreting complex medical images. Therefore, an MRMC comparative effectiveness study involving AI-assisted human readers is not applicable and was not performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, in essence. The device itself is a standalone qualitative test. Its performance (producing a positive or negative line) is inherent to the device's chemical assay. While a human physically observes the lines, the "algorithm" is the chemical reaction and visual indicator built into the device. The precision and accuracy studies directly reflect this standalone chemical "algorithm's" performance against known concentrations or GC/MS. The lay user study evaluated the human reading the standalone device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The primary ground truth used for evaluating the performance of the device was known spiked concentrations of drugs (confirmed by GC/MS) for precision and lay user studies, and Gas Chromatography/Mass Spectrometry (GC/MS) analysis for clinical accuracy studies. GC/MS is a highly accurate analytical method considered a gold standard for confirming the presence and concentration of drugs in urine.

    8. The sample size for the training set

    The document describes performance studies, but it does not mention a "training set" in the context of machine learning or AI models. This device is a lateral flow immunoassay, which does not typically involve training data in the AI sense. The design and validation of this type of device rely on analytical and clinical performance studies, not machine learning model training.

    9. How the ground truth for the training set was established

    As there is no mention of a "training set" for an AI or machine learning model, this question is not applicable to the information provided in the document. The ground truth for the performance evaluation (test sets) was established through spiking known drug concentrations into urine and confirmation via GC/MS, as detailed in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140215

    Validate with FDA (Live)

    Date Cleared
    2014-06-16

    (139 days)

    Product Code
    Regulation Number
    862.3150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are lateral flow chromatographic immunoassay designed to qualitatively detect the presence of drugs and drug metabolites in human urine at the following cut-off concentrations:

    TestCalibratorCut-off level
    Barbiturates (BAR)Secobarbital300 ng/mL
    Benzodiazepines (BZO)Oxazepam300 ng/mL
    Methylenedioxymethamphetamine (MDMA)3,4-Methylenedioxymethamphetamine500 ng/mL
    Methadone (MTD)Methadone300 ng/mL
    Oxycodone (OXY)Oxycodone100 ng/mL
    Phencyclidine (PCP)Phencyclidine25 ng/mL

    The tests contain two formats:1) Test Cup, 2) Test Dipcard, The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for in vitro diagnostics use. They are intended for prescription use in clinical laboratories only and not for point-of-care use.

    This assay provides only a preliminary analytical test result. Gas Chromatography/Mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

    Device Description

    One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Barbiturates, Benzodiazepines, Methylenedioxymethamine, Methadone, Oxycodone, Phencyclidine and their metabolites at or above the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    The provided document describes the Co-Innovation Biotech Co.,Ltd. One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard for qualitative detection of drugs of abuse in human urine. Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for qualitative immunoassay tests like these are typically assessed by comparing the device's results (positive/negative) against a gold standard (GC/MS) especially around the cutoff concentration. While explicit numerical acceptance criteria (e.g., % agreement, sensitivity, specificity targets) are not explicitly stated as 'acceptance criteria' in the document, the performance data implicitly serves as the demonstration that the device performs acceptably. The study evaluates the device's ability to correctly identify drug-free, less than half cutoff, near cutoff negative, near cutoff positive, and high positive samples.

    The tables below synthesize the reported device performance for both the Test Cup and Test Dipcard formats for single and multi-drug tests, against the GC/MS analysis. The "Total" column in the original tables sums up the sample distribution across concentration categories, not the total number of samples processed for each drug test condition. The key performance indicator here is the consistency of results, particularly for samples around the cutoff concentration.

    Device Performance for One Step Single/Multi-drug Test Cup & Dipcard (Single Drug Test Example - All Drugs follow similar pattern)

    Drug TestCo-Innovation ResultGC/MS Analysis (Drug-free)GC/MS Analysis (<0.5x Cutoff)GC/MS Analysis (Near Cutoff Negative)GC/MS Analysis (Near Cutoff Positive)GC/MS Analysis (>1.5x Cutoff)Total Samples
    BAR+00063480
    -33070080
    BZO+00173380
    -31080080
    MDMA+00053480
    -32351080
    MTD+00153580
    -32250080
    OXY+00063480
    -35050080
    PCP+00153580
    -35040080

    Analysis of Discordant Results (Example for Test Cup and Dipcard, Single and Multi-drug)

    Drug TestCutoff (ng/mL)Device ResultGC/MS Drug Concentration (ng/mL)Drug in Urine
    BZO300Positive188Oxazepam
    MDMA500Negative7153,4-Methylenedioxymethamphetamine
    MTD300Positive209Methadone
    PCP25Positive23Phencyclidine
    (Note: These discordant results are consistent across all four tables presented for Single Drug Test Cup, Single Drug Test Dipcard, Multi-drug Test Cup, and Multi-drug Test Dipcard)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 80 clinical urine specimens were used for each drug tested, for each device format (Single Drug Test Cup, Single Drug Test Dipcard, Multi-drug Test Cup, Multi-drug Test Dipcard). Therefore, for each drug, a total of 320 samples were analyzed (80 * 4). The samples were categorized into five groups based on concentration relative to the cutoff (drug-free, less than half the cutoff negative, near cutoff negative, near cutoff positive, and high positive).
    • Data Provenance: The data is from "clinical urine specimens," implying human origin. However, the country of origin is not specified, and it is a retrospective analysis as the specimens were analyzed by GC/MS for comparison with the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS) analysis, which is considered the "preferred confirmatory method" for drug of abuse testing.
    • No human "experts" are mentioned as having established the ground truth through consensus or individual reading of the device results for the purpose of the primary accuracy study. The device results were compared directly against the GC/MS findings.

    4. Adjudication Method for the Test Set

    • None in the traditional sense of multiple human readers adjudicating results. The device's qualitative results (positive/negative) were directly compared to the quantitative GC/MS results for each sample. Discordant results were individually listed with the GC/MS concentration.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a MRMC comparative effectiveness study was not performed. The study's focus was on the standalone performance of the devices against a gold standard (GC/MS). Human readers' improvement with or without AI assistance is not relevant here as it's a diagnostic test that provides a clear positive/negative result, not an AI-assisted diagnostic imaging interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, this was a standalone performance study. The "One Step Single/Multi-drug Test Cup" and "One Step Single/Multi-drug Test Dipcard" are qualitative, lateral flow immunochromatographic assays designed to be read directly without an instrument or human interpretation loop beyond reading the presence or absence of test lines. The performance data presented directly reflects the device's output (positive/negative) compared to GC/MS.

    7. The Type of Ground Truth Used

    • The ground truth used was quantitative analytical data from Gas Chromatography/Mass Spectrometry (GC/MS), which is the preferred confirmatory method for drug of abuse testing. This is a highly objective and precise method for determining drug concentration.

    8. The Sample Size for the Training Set

    • This document describes a performance evaluation of a medical device, not an AI algorithm that requires a training set. Therefore, no training set for an algorithm is mentioned or applicable in this context. The devices are immunoassay tests, not machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    • As mentioned above, there is no training set for an algorithm as the device is an immunoassay test.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131110

    Validate with FDA (Live)

    Date Cleared
    2014-01-15

    (271 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    One Step Single/Multi-drug Test Cup Single/Multi-drug Test Dipeard are lateral flow chromatographic immunoassrys designed to qualitatively detect the presence of drugs and drug metabolites in human urine at or above the following cut-off concentrations:

    Marijuana (THC) Calibrato r · Delta-9-THC-COOH Cut-off level 50 ng/ml.
    Cocaine (COC) Benzoylecgonine 300 ng/ml.
    Amphetamine (AMP) D-Amphetamine 1000 ng/ml.
    Methamphetamine (MET) D-Methamphetamine 1000 ng/mL
    Morphine 2000 (MOP) Morphine 2000 ne/ml.

    The tests contain two formats: 1) Test Cup, 2) Test Dipeard. The test configuration comes with single drug sereening test or any combinations of multiple drug screening tests. The test is intended for in vitro diagnostics use. They are intended for preseription use in clinical laboratories only and not for point-of-care use.

    These esss provide only a prefininary analytical test result and are the first step in a two-step process for detecting drugs of abuse in urine. The second step is continuing the results in a certified laboratory. For a quantitative result or to confirm preliminary positive positive positive results obtained by the One Step Multi-drug Test Cup Insert or One Step Single/Multi-drug Test Dipeard Insert, a more specific alternaire method such as Gas Chromatography/Mass Spectomerry (GCMS) must be used. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

    Device Description

    One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Cocaines, Marijuana, Methamphetamine,Morphine and their metabolites at or above the cut-off levels as indicated. The tests can be performed without the use of an instrument.

    AI/ML Overview

    The provided document describes the performance of the Co-Innovation Biotech Co., Ltd. One Step Single/Multi-drug Test Cup and Dipcard for the qualitative detection of drugs and drug metabolites in human urine.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" with numerical thresholds for performance metrics (like sensitivity, specificity, or agreement percentages). Instead, it presents the results of clinical studies (accuracy studies) designed to demonstrate the device's performance against a gold standard (GC/MS). The implied acceptance is that the device demonstrates a high level of agreement with GC/MS, especially near and above the cutoff concentrations.

    For the purpose of this response, I infer the performance target is to achieve high agreement with GC/MS. The reported device performance is shown in the tables below, demonstrating the number of positive and negative results from the device compared to GC/MS at different concentration levels.

    Performance Summary (based on provided data for both formats: Cup and Dipcard)

    Drug TestCut-off Level (ng/mL)Device FormatAgreement with Drug Free (Negative Result)Agreement with < Half Cutoff (Negative Result)Agreement Near Cutoff NegativeAgreement Near Cutoff PositiveAgreement High Positive (Positive Result)Number of Discordant Results (Device Pos/GCMS Neg or Device Neg/GCMS Pos)
    AMP1000Single Test Cup100% (117/117)Not explicitly given in a clear format (combined into one cell)11/12 (Negatives)10/11 (Positives)31/31 (Positives)2 (1 Positive at 867 ng/mL, 1 Negative at 1175 ng/mL)
    AMP1000Single Test Dipcard100% (117/117)100% (7/7)11/12 (Negatives)10/11 (Positives)29/29 (Positives)2 (1 Positive at 867 ng/mL, 1 Negative at 1175 ng/mL)
    AMP1000Multi-drug Test Cup100% (25/25)100% (3/3)11/12 (Negatives)10/11 (Positives)29/29 (Positives)2 (1 Positive at 867 ng/mL, 1 Negative at 1175 ng/mL)
    AMP1000Multi-drug Test Dipcard100% (25/25)100% (3/3)11/12 (Negatives)10/11 (Positives)29/29 (Positives)2 (1 Positive at 867 ng/mL, 1 Negative at 1175 ng/mL)
    COC300Single Test Cup100% (133/133)100% (0/0)12/13 (Negatives)0/0 (Positives)31/31 (Positives)1 (1 Positive at 172 ng/mL)
    COC300Single Test Dipcard100% (133/133)100% (0/0)12/13 (Negatives)9/9 (Positives)31/31 (Positives)1 (1 Positive at 172 ng/mL)
    COC300Multi-drug Test Cup100% (27/27)100% (0/0)12/13 (Negatives)9/9 (Positives)31/31 (Positives)1 (1 Positive at 172 ng/mL)
    COC300Multi-drug Test Dipcard100% (27/27)100% (0/0)12/13 (Negatives)9/9 (Positives)31/31 (Positives)1 (1 Positive at 172 ng/mL)
    THC50Single Test Cup100% (107/107)100% (4/4)9/10 (Negatives)12/13 (Positives)31/31 (Positives)2 (1 Positive at 40 ng/mL, 1 Negative at 59 ng/mL)
    THC50Single Test Dipcard100% (107/107)100% (4/4)9/10 (Negatives)12/13 (Positives)31/31 (Positives)2 (1 Positive at 40 ng/mL, 1 Negative at 59 ng/mL)
    THC50Multi-drug Test Cup100% (26/26)100% (4/4)9/10 (Negatives)12/13 (Positives)27/27 (Positives)2 (1 Positive at 40 ng/mL, 1 Negative at 59 ng/mL)
    THC50Multi-drug Test Dipcard100% (26/26)100% (4/4)9/10 (Negatives)12/13 (Positives)27/27 (Positives)2 (1 Positive at 40 ng/mL, 1 Negative at 59 ng/mL)
    MET1000Single Test Cup100% (85/85)100% (4/4)13/14 (Negatives)17/18 (Positives)44/44 (Positives)2 (1 Positive at 904 ng/mL, 1 Negative at 1248 ng/mL)
    MET1000Single Test Dipcard100% (85/85)100% (4/4)13/14 (Negatives)17/18 (Positives)44/44 (Positives)2 (1 Positive at 904 ng/mL, 1 Negative at 1248 ng/mL)
    MET1000Multi-drug Test Cup100% (24/24)100% (2/2)13/14 (Negatives)12/13 (Positives)27/27 (Positives)2 (1 Positive at 904 ng/mL, 1 Negative at 1248 ng/mL)
    MET1000Multi-drug Test Dipcard100% (24/24)100% (2/2)13/14 (Negatives)12/13 (Positives)27/27 (Positives)2 (1 Positive at 904 ng/mL, 1 Negative at 1248 ng/mL)
    MOP2000Single Test Cup100% (67/67)100% (5/5)15/17 (Negatives)19/19 (Positives)57/57 (Positives)2 (2 Positives at 1608 ng/mL and 1875 ng/mL)
    MOP2000Single Test Dipcard100% (67/67)100% (5/5)15/17 (Negatives)19/19 (Positives)57/57 (Positives)2 (2 Positives at 1608 ng/mL and 1875 ng/mL)
    MOP2000Multi-drug Test Cup100% (21/21)100% (2/2)15/17 (Negatives)11/11 (Positives)29/29 (Positives)2 (2 Positives at 1608 ng/mL and 1875 ng/mL)
    MOP2000Multi-drug Test Dipcard100% (21/21)100% (2/2)15/17 (Negatives)11/11 (Positives)29/29 (Positives)2 (2 Positives at 1608 ng/mL and 1875 ng/mL)

    2. Sample Size Used for the Test Set and Data Provenance

    • Single Drug Test:
      • Test Set Size: 165-186 clinical urine specimens for each drug (AMP, COC, THC, MET, MOP) were used for each device format (Cup and Dipcard).
      • Data Provenance: The specimens were described as "clinical urine specimens," implying they were collected from human subjects in a real-world clinical setting. No specific country of origin is mentioned, but the submitter's address is in Guangzhou, P.R. China, suggesting the data may originate there. The study appears to be retrospective, as existing clinical specimens were analyzed.
    • Multi-Drug Test:
      • Test Set Size: 80 clinical urine specimens for each drug were used for each device format (Cup and Dipcard).
      • Data Provenance: Similar to the single drug test, these were "clinical urine specimens." Provenance details are identical to the single drug test.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS). This is a laboratory-based analytical method, not a human expert interpretation. Therefore, the concept of "number of experts" or their "qualifications" in the context of establishing ground truth does not apply as it would for image-based diagnostics. The accuracy of GC/MS itself is well-established as a confirmatory method for drug testing.

    4. Adjudication Method for the Test Set

    No human adjudication method (like 2+1, 3+1 consensus) was used. The study compared the device results directly against the quantitative results obtained from GC/MS. Discordant results (where the device and GC/MS disagreed) are individually listed with the specific drug concentration from GC/MS. This suggests that GC/MS was considered the definitive gold standard.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This device is a biochemical assay (lateral flow immunoassay) for qualitative detection, not an AI-assisted diagnostic tool where human readers interpret medical images or data. The device provides a direct positive/negative result, not an interpretation that humans would then review.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, this was effectively a standalone study. The device (One Step Single/Multi-drug Test Cup/Dipcard) itself produces a result (qualitative positive or negative) without human interpretation. The study evaluates the performance of the device only against the GC/MS ground truth. Human involvement is limited to performing the test procedure and reading the physical test line, which is a direct visual cue, not an interpretive task that would significantly affect the "algorithm's" performance.

    7. The Type of Ground Truth Used

    The ground truth used was GC/MS Analysis (Gas Chromatography/Mass Spectrometry). This is an objective, analytical method that provides quantitative drug concentrations in the urine specimens. It is considered the gold standard for drug confirmation.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" for the device. As an immunochromatographic assay, these devices are typically developed through bench-top experimentation and optimization of reagents and manufacturing processes, rather than machine learning training that would require a distinct "training set." The performance data provided is for the rigorous testing of the finalized device.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no explicit training set mentioned in the context of machine learning, this question is not directly applicable. The development and optimization of the immunoassay itself would rely on established biochemical principles and analytical methods to determine antigen-antibody reactions and cutoff sensitivities, likely using reference standards and known concentrations of drugs and metabolites during the R&D phase.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132085

    Validate with FDA (Live)

    Date Cleared
    2013-11-12

    (130 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Co-Innovation One Step HCG Test Strip is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. It is intended for over-the-counter (OTC) use.

    The Co-Innovation One Step HCG Test Cassette is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. It is intended for over-the-counter (OTC) use.

    The Co-Innovation One Step HCG Test Midstream is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. It is intended for over-the-counter (OTC) use.

    Device Description

    Co-Innovation One Step Human Chorionic Gonadotropin (HCG) Test is a rapid sandwich immunoassay device designed for the qualitative determination of human chorionic gonadotropin (hCG) concentration in human urine samples, as an aid in the early detection of pregnancy. The test devices are in three different formats: Strip, Cassette and Midstream . Three test formats use identical strips and each test strip in the device consists of:

    1. A conjugate pad contains colloidal gold conjugated with mouse monoclonal anti-B-HCG antibody specific to the beta subunit of hCG.

    2. A nitrocellulose membrane which is striped with the mouse monoclonal anti-a-HCG antibody in the test line (T line) and goat anti-mouse IgG polyclonal antibody in the control line (C line).

    The Cassette format has the same performance specifications as the Test Strip format. The difference is that the urine sample is dispensed by dropper onto the sample well on the cassette.

    The Midstream format has the same performance specifications as the Test Strip format. The difference is that the device is placed into the urine stream or dipped into the urine collection cup for 5 seconds.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary, detailing the acceptance criteria and the studies conducted to demonstrate the device's performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the successful outcomes presented in the study results, particularly for precision and sensitivity, as no explicit numerical acceptance targets are stated (e.g., "sensitivity must be >99%"). However, the performance data clearly demonstrates that the device meets the implicit requirement for accurate qualitative detection of HCG at the clinical cut-off of 25mIU/mL.

    Acceptance Criterion (Implicit)Reported Device Performance
    Precision: Consistent positive results at/above 25mIU/mL and negative results below 25mIU/mL.Precision (Strip, Cassette, Midstream - all methods): - 0mIU/mL, 12.5mIU/mL, 18.75mIU/mL: 100% negative (30/30 for each LOT and concentration). - 25mIU/mL, 50mIU/mL, 100mIU/mL: 100% positive (30/30 for each LOT and concentration).
    Sensitivity: Detect HCG at the clinical cut-off of 25mIU/mL.Sensitivity: 25mIU/mL (derived from precision study, showing 100% detection at this concentration).
    Specificity: No cross-reactivity with common related hormones.Specificity: No cross-reaction observed with LH at 500mIU/mL, FSH at 1000mIU/mL, and TSH at 1000uIU/ml in urine specimens containing 0 mIU/ml and 25mIU/ml HCG.
    Interference: No interference from common exogenous compounds.Interference: No interferences observed from Acetaminophen, Aspirin, Ascorbic acid, Atropine, Caffeine (all at 20mg/dL), Glucose (2000mg/dL), Hemoglobin (500mg/dL), Tetracycline, Ampicillin (all at 20mg/dL), Albumin (2000mg/dL), Bilirubin (2mg/dL) for both negative and positive HCG samples.
    HCG ß-core fragment interference: No interference from high levels of HCG ß-core fragment.HCG ß-core fragment interference: No interference was observed when spiked at 125,000, 250,000, 500,000, and 1,000,000 pmol/mL in urine samples containing 0 and 25 mlU/ml HCG.
    pH interference: Stable performance across a physiological pH range.pH interference: Varying pH ranges (3 to 9 in 1 pH unit increments) did not interfere with test performance for 0mlU/ml and 25mlU/ml HCG samples.
    Specific gravity interference: Stable performance across a physiological range of specific gravity.Specific gravity interference: No interference was observed when specific gravity was between 1.01-1.04 for purified water and 25mIU/mL HCG spiked specimens.
    Urinary system diseases interference: No interference from markers of urinary conditions.Urinary system diseases interference: No interference observed from strongly positive white blood cells, urine occult blood, uric acid, and urine ketone in urine samples containing 0mlU/ml and 25mlU/ml HCG.
    Hook effect: Maintain positive results at high HCG concentrations.Hook effect: Generated positive results for HCG concentrations from 62,500mIU/ml to 1,000,000mIU/ml. (Note: "T line get to light as the concentration above 125000mlU/ml" indicates the test still works, but potentially with reduced band intensity, which is acceptable for qualitative detection and does not lead to false negatives at very high concentrations).
    Method Comparison (Professional User): High agreement with predicate device results.Professional Method Comparison (Strip, Cassette, Midstream - all methods): 100% agreement with the predicate device for all formats. (e.g., for Strip, Professional A: 38 Positive, 42 Negative; Professional B: 38 Positive, 42 Negative, all matching the predicate).
    Method Comparison (Lay User): High agreement with professional test results.Lay User Method Comparison (Strip, Cassette, Midstream - all methods): 100% agreement with professional laboratory results using the candidate device for all formats. (e.g., for Strip, Lay Users: 38 Positive, 42 Negative, all matching the professional results).
    OTC User Performance: Untrained users can obtain correct results reliably.OTC User Study: - Strip: 29/30 correct positive (31.25mIU/ml), 30/30 correct negative (18.75mIU/ml). - Cassette: 30/30 correct positive, 30/30 correct negative. - Midstream (dip): 30/30 correct positive, 30/30 correct negative. - Midstream (simulated midstream): 30/31 correct positive, 29/29 correct negative. Conclusion: Untrained operators can get correct results.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Study: 30 clinical samples (normal, non-pregnant female urine) were spiked with HCG at 6 different concentrations (0mIU/ml, 12.5mIU/ml, 18.75mIU/ml, 25mIU/ml, 50mIU/ml, 100mIU/ml). Each concentration had 30 samples tested per LOT, across 3 lots, yielding 90 tests per concentration per format. The study was conducted over 10 days, 3 runs/day.
      • Provenance: Clinical samples from "normal, non-pregnant females." The document does not explicitly state the country of origin, but the submitter is from Guangzhou, China, and the OTC user study sites are in Guangzhou, Shanghai, and Wuhan, suggesting a Chinese origin. The study appears to be a prospective experimental study using spiked samples.
    • Specificity Study: 60 urine specimens (containing 0 mIU/ml and 25mIU/ml HCG) were spiked with LH, FSH, and TSH. Tested across 3 lots.
      • Provenance: Not explicitly stated, likely similar to precision study. Prospective experimental study.
    • Interfering Substances Study: Normal, non-pregnant female urine specimens (0 and 25 mIU/ml HCG) were spiked with 11 different compounds. 3 batches of each format were tested for each interferent.
      • Provenance: Not explicitly stated, likely similar to precision study. Prospective experimental study.
    • HCG ß-core fragment Interference Study: Normal, non-pregnant female urine specimens (0 and 25 mlU/ml HCG) were spiked with 4 different concentrations of HCG ß-core fragment. 3 batches of each format were tested.
      • Provenance: Not explicitly stated, likely similar to precision study. Prospective experimental study.
    • pH Interference Study: Aliquoted negative urine pool adjusted to pH 3-9, spiked with 0mlU/ml and 25mlU/ml HCG. 3 batches were tested repeatedly.
      • Provenance: Not explicitly stated, likely similar to precision study. Prospective experimental study.
    • Specific Gravity Interference Study: Purified water and 25mlUml HCG spiked specimens formulated with specific gravity 1.01-1.04. 3 lots tested.
      • Provenance: Not explicitly stated, likely similar to precision study. Prospective experimental study.
    • Urinary System Diseases Interference Study: Urine specimens with strongly positive markers (white blood cells, occult blood, uric acid, urine ketone) prepared with 0mlU/ml and 25mlU/ml HCG. 3 lots tested repeatedly.
      • Provenance: Not explicitly stated, likely clinical samples. Prospective experimental study.
    • HOOK Effect Study: HCG free specimens spiked with 5 different very high HCG concentrations (62,500mIU/m1 to 1,000,000mIU/ml). Three lots tested.
      • Provenance: Not explicitly stated, likely similar to precision study. Prospective experimental study.
    • Professional Method Comparison: 353 urine samples collected from women at a hospital laboratory. "Approximately half of the women were pregnant in the early stage of less than 5 weeks." Ages 18-45 years.
      • Provenance: Clinical samples from a hospital laboratory. Not explicitly stated, but likely China given other study locations. Retrospective (samples collected, then tested) or prospective (samples collected specifically for the study) clinical study.
    • Lay User Method Comparison: The same 353 urine samples as the Professional Method Comparison.
      • Provenance: Clinical samples, likely China. Retrospective or prospective clinical study.
      • Specific breakdown for lay users: 80 using strip, 113 using cassette, 160 using midstream (actual + dip methods combined).
    • OTC User Performance: 240 female subjects (untrained operators, ages 18-45) tested masked spiked urine samples (120 containers of 18.75mlU/ml and 120 of 31.25mlU/ml). Each subject performed 1 test on one format/method.
      • Provenance: Samples were spiked. Subjects from Guangzhou, Shanghai, and Wuhan, China. Prospective experimental study (simulated home use).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Precision, Specificity, Interference, pH, Specific Gravity, Urinary System Diseases, HOOK Effect Studies: The ground truth (HCG concentration, presence/absence of interferents) was established by spiking known concentrations of HCG or interfering substances, traceable to WHO standards or determined by immunoassay (ELISA). This does not involve human experts establishing ground truth for individual test samples, but rather analytical methods.
    • Professional Method Comparison: The "ground truth" for this study was the result obtained by the predicate device when tested by laboratory professionals. The exact number of professionals is not specified, but the study was conducted at "two laboratories (namely Professional A and Professional B)", implying at least two professionals. Their qualifications are stated as "laboratory professionals."
    • Lay User Method Comparison: The "ground truth" for this study was the result obtained by laboratory professionals using the candidate device. The number of laboratory professionals is not specified, but it was conducted at "a laboratory." Their qualifications are "laboratory professionals."
    • OTC User Performance: The "ground truth" for this study was established by masked spiked urine samples (known HCG concentrations) and independently verified by professional laboratory personnel at the manufacturer site. The number of professionals is not specified, and their exact qualifications are not detailed beyond "professional laboratory personnel."

    4. Adjudication Method for the Test Set

    • For the analytical performance studies (Precision, Specificity, Interference, etc.), adjudication generally involves comparing the device's output to the known, spiked concentration or condition. There's no human adjudication process described beyond analytical confirmation of spike levels.
    • For the Professional Method Comparison and Lay User Method Comparison, the comparison is against the predicate device or professional results of the candidate device. The agreement is tabulated directly, implying no specific "adjudication" if disagreement occurred, as 100% agreement was reported.
    • For the OTC User Performance study, the lay users' results were compared against the "masked spiked sample professional users" results (i.e., known spiked concentration verified by professionals). Again, no specific adjudication process for discrepancies is described, as the focus is on the agreement rate.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study done, as this device is a qualitative diagnostic test (HCG pregnancy test) not an imaging device typically analyzed by human readers with or without AI assistance. The studies involve human users (professional or lay) interpreting the test results visually, but not in a context typically requiring MRMC analysis.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This device is a visual qualitative immunochromatographic assay. There is no "algorithm" in the sense of a software-based AI. The performance is inherently "standalone" in that it produces a visual result (a line) and the interpretation is human-in-the-loop (reading the line). The "performance" studies are effectively standalone because they evaluate the test's ability to produce the correct visual output given an HCG concentration. The OTC user study then evaluates the human (lay user) interpretation of this visual output.

    7. The Type of Ground Truth Used

    • Spiked Samples (Precision, Specificity, Interference, Hook Effect, pH, Specific Gravity, Urinary System Diseases, OTC User Study): The ground truth was established by laboratory control, adding known concentrations of HCG (traceable to WHO 3rd IS) or interfering substances to urine samples. This is a form of analytical or engineered ground truth.
    • Professional Method Comparison: The ground truth was the result from the predicate device.
    • Lay User Method Comparison: The ground truth was the result from professional users testing the candidate device.

    8. The Sample Size for the Training Set

    There is no explicit training set described in this 510(k) summary. This type of device (lateral flow immunoassay) typically does not involve a machine learning model that requires a distinct "training set." The development process for such a device relies on chemical and biological engineering principles, optimization of reagents, and initial analytical testing, rather than an iterative machine learning training process with labeled data.

    9. How the Ground Truth for the Training Set Was Established

    As no explicit training set for a machine learning algorithm is mentioned, pertinent to the nature of this device, the question of how ground truth was established for a training set is not applicable. The ground truth for the performance evaluations (test sets) was established as described in section 7.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1