Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K210445

    Validate with FDA (Live)

    Date Cleared
    2021-06-29

    (133 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nordiwell Medical Face Mask is intended to be worn to protect both patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate matters. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device(s) and provided non-sterile.

    Device Description

    The Nordiwell Medical Face Mask has a flat-folded design consisting of three layers that are comprised of an (polypropylene spunbond) inner and outer cover web and (polypropylene melt blown) filter web. The mask has earloops or tie strings to secure the mask over the users' mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. This is a single use, disposable device, provided non-sterile.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical face mask. It does not describe an AI/ML powered medical device, but rather a physical product. Therefore, many of the requested criteria regarding AI/ML model evaluation (e.g., training set, ground truth acquisition, MRMC studies, standalone performance with human-in-the-loop, number of experts for ground truth) are not applicable.

    However, I can extract the acceptance criteria and performance data for this physical medical device from the provided text.

    Here's the information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Test ItemAcceptance CriteriaReported Device PerformanceStudy Conclusion
    Fluid Resistance Performance (ASTM F1862-17)≥ 29 of 32 pass at 120 mmHg31 of 32 pass at 120 mmHgPass
    Particulate Filtration Efficiency (ASTM F2299-17)≥ 98%99.9%, 32 of 32 passPass
    Bacterial Filtration Efficiency (ASTM F2101-19)≥ 98%99.9%, 32 of 32 passPass
    Differential Pressure (Delta P) (EN 14683:2019, Annex C and ASTM F2100-19)< 6.0 mmH2O/cm²4.6 mmH2O/cm², 32 of 32 passPass
    Flammability (16 CFR 1610)Class 1Class 1, 32 of 32 passPass
    Biocompatibility (ISO 10993 series)Non-Cytotoxic, Non-Irritating, Non-SensitizingNon-Cytotoxic, Non-Irritating, Non-SensitizingPass

    2. Sample size used for the test set and the data provenance

    • Sample Size for performance tests: For each performance test (Fluid Resistance, Particulate Filtration, Bacterial Filtration, Differential Pressure, Flammability), the sample size was 32 units. The document also states "Bench test was conducted on 3 nonconsecutive lots of Medical Face Mask". This implies 32 units per lot, or a total of 32 units drawn across 3 nonconsecutive lots. The wording "32 of 32 pass" for each specific test strongly suggests that the sample size for each test was 32.
    • Data Provenance: The tests were conducted internally by the manufacturer (Changzhou Combat Protective Equipment Co., Ltd.) in China, as indicated by the contact information and location. The studies are essentially prospective in nature, as they involve testing newly manufactured devices to verify performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a physical product, not an AI/ML device that requires human expert annotation for ground truth. The "ground truth" here is the result of standardized physical and biological tests conducted against established ASTM and ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. This is a physical product. Test results are objective measurements from laboratory equipment and procedures, not subjective human interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI/ML medical device. No human reader studies were conducted as the device is a medical face mask.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an AI/ML medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this medical device's performance is based on established international and national standards (ASTM F1862-17, ASTM F2299-17, ASTM F2101-19, EN 14683:2019, 16 CFR 1610, ISO 10993 series). These standards define the test methodologies and acceptance criteria for medical face masks. The testing is performed in laboratory settings to objectively measure the device's physical and biological properties.

    8. The sample size for the training set

    • Not Applicable. This is a physical product, not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is a physical product, not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210181

    Validate with FDA (Live)

    Date Cleared
    2021-03-09

    (46 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When properly worn, the surgical face masks are intended to patient and healthcare workers from transfer of microorganisms, body fluids and particulate matter. This device is non-sterile and for single use only.

    Device Description

    The Nordiwell Surgical Face Masks are composed of 3-layers and are flat-pleated. The mask materials consist of an outer layer (polypropylene spunbond, blue), inner layer (polypropylene spunbond white), filter (polypropylene melt-blown, white) and ties (polypropylene spunbond white). The mask contains a malleable nosepiece which is centrally positioned between the opposed side edges along the top edge of the mask. The nosepiece is a stable anchor in the vicinity of the nose and readily bent to conform to contours of a person's face. Thus the nosepiece can provide a firm fit over the nose and then to secure the mask over the users' mouth and face. The mask is a single use, provided non-sterile. This device is not made with natural rubber latex and Shelf-life of the device is 2 years. The subject of this submission is a design change to the ear-loop in the Nordiwell Medical Face Mask(K202615). The ear-loop of the Nordiwell Medical Face Mask(K202615) is changed to tie in the subject device.

    AI/ML Overview

    This document describes the acceptance criteria and supporting study for the Nordiwell Surgical Face Mask (K210181), which is a surgical face mask.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is evaluated against the ASTM F2100 Level 3 standard for surgical face masks.

    Performance CharacteristicAcceptance Criteria (ASTM F2100 Level 3)Reported Device Performance
    Fluid Resistance160 mmHg32 of 32 pass at 160 mmHg
    Particulate Filtration Efficiency (PFE)≥ 98% at 0.1 micron> 99.9%
    Bacterial Filtration Efficiency (BFE)≥ 98%≥ 98%
    Differential Pressure (Delta P)< 6.0 mmH2O/cm²5.6 mmH2O/cm²
    FlammabilityClass 1Class 1
    Biocompatibility:
    - In Vitro CytotoxicityNon-cytotoxicNon-Cytotoxic
    - Skin IrritationNon-irritatingNon-Irritating
    - Skin SensitizationNon-sensitizingNon-Sensitizing

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for each specific test, but it indicates "32 of 32 pass" for fluid resistance, implying a sample size of 32 for that particular test. The data provenance is implied to be from the manufacturer, Changzhou Combat Protective Equipment Co., Ltd., based in China, as per the contact information provided in the 510(k) summary. The studies are non-clinical (bench testing) and therefore not retrospective or prospective in the human study sense.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable as the studies are non-clinical bench tests comparing the device's physical properties against established industry standards (ASTM F2100). No human experts were used to establish ground truth in the context of clinical interpretation or diagnosis.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies are non-clinical bench tests for physical properties according to standardized methods. No human adjudication was involved.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was conducted. The device is a surgical face mask, and performance is evaluated through bench testing against established standards, not through reader studies.

    6. Standalone Performance Study

    Yes, a standalone study was done. The document explicitly states: "Bench testing was conducted to demonstrate the performance of the subject device." The results of these bench tests are presented against the ASTM F2100 Level 3 criteria, indicating the algorithm's (device's) standalone performance for various physical and filtration properties.

    7. Type of Ground Truth Used

    The ground truth used for evaluation is established industry standards (ASTM F2100 Level 3) for surgical face mask performance and international standards for biocompatibility (ISO 10993 series). These are objective, quantifiable criteria.

    8. Sample Size for the Training Set

    This information is not applicable. The device is a physical product (surgical face mask), not an AI algorithm or a device requiring a training set in that context. The "training set" concept is irrelevant to a physical product's performance validation against standards.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202615

    Validate with FDA (Live)

    Date Cleared
    2020-12-31

    (113 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When properly worn, the surgical face masks are intended to patient and healthcare workers from transfer of microorganisms, body fluids and particulate matter. This device is non-sterile and for single use only.

    Device Description

    The Nordiwell Medical Face Masks are composed of 3-layers and are flat-pleated. The mask materials consist of an outer layer (polypropylene spunbond, blue), inner layer (polypropylene spunbond white), filter (polypropylene melt-blown, white) and ear-loops. The mask contains a malleable nosepiece to provide a firm fit over the nose and then to secure the mask over the users' mouth and face. The mask is a single use, provided non-sterile. This device is not made from Natural Rubber Latex. Shelf-life of the device is 2 years.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the Nordiwell Medical Face Mask (K202615) to demonstrate substantial equivalence to a predicate device. This is a medical device, specifically a surgical face mask, not an AI/ML powered device. Therefore, many of the requested categories in the prompt, such as those related to AI model performance, expert ground truth, adjudication methods, and training/test sets for AI, are not applicable.

    Here's the information that can be extracted from the provided text, focused on the device's physical and performance characteristics:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: Nordiwell Medical Face Mask (K202615)

    ItemAcceptance CriteriaReported Device PerformanceResult
    Fluid Resistance Performance (ASTM F1862-17)≥29 of 32 pass at 160 mmHg32 of 32 pass at 160 mmHgPass
    Particulate Filtration Efficiency (ASTM F2299-17)≥ 98%> 99.9%Pass
    Bacterial Filtration Efficiency (ASTM F2101-19)≥ 98%> 99.9%Pass
    Differential Pressure (Delta P) (EN 14683:2019, Annex C and ASTM F2100-19)<6.0 mmH2O/cm²5.6 mmH2O/cm²Pass
    Flammability (16 CFR 1610)Class 1Class 1Pass
    In Vitro Cytotoxicity Test (ISO 10993-5: 2009)Non-CytotoxicNon-CytotoxicPass
    Skin Irritation Test (ISO 10993-10:2010)Non-IrritatingNon-IrritatingPass
    Skin Sensitization Test (ISO 10993-10:2010)Non-SensitizingNon-SensitizingPass

    2. Sample size used for the test set and the data provenance

    The document indicates that for fluid resistance testing (ASTM F1862-17), a sample size of 32 units was used for testing against the acceptance criteria of "≥29 of 32 pass". For the shelf-life accelerated aging test, 3 non-consecutive lots were used (sample size per lot not specified for each performance test, but it can be inferred that appropriate sample sizes were used for each test on these lots).

    Data Provenance: The tests were conducted internally by the manufacturer, Changzhou Combat Protective Equipment Co., Ltd., in China, or by their designated testing facilities. The data is prospective, generated specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not Applicable. This device is a physical medical face mask. Ground truth is established through standardized laboratory bench testing rather than expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable. Adjudication methods are relevant for subjective assessments, typically in clinical or image-based studies. The performance of this device is measured using objective, standardized bench tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This is a physical medical device, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is derived from objective measurements against established industry standards and regulatory criteria through bench testing (e.g., ASTM and EN standards for filtration, pressure, flammability, and ISO standards for biocompatibility).

    8. The sample size for the training set

    Not Applicable. This product is a physical medical device, not a machine learning model; therefore, there is no "training set."

    9. How the ground truth for the training set was established

    Not Applicable. As there is no training set, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1