510(k) Data Aggregation

    K Number
    K181772

    Validate with FDA (Live)

    Date Cleared
    2018-08-01

    (29 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovoStitch Pro Mensical Repair System is intended for approximation of soft tissue in meniscal repair procedures.

    Device Description

    The NovoStitch Pro Meniscal Repair System, size 0 is a new configuration of the predicate device. While the predicate device passes size 2-0 suture, this new configuration passes size 0 braided, nonabsorbable, surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with suture are attached.

    AI/ML Overview

    Here's an analysis of the provided text regarding the NovoStitch Pro Meniscal Repair System, CTX-A004 (size 0), focusing on acceptance criteria and the supporting study:

    The provided document is a 510(k) summary for the NovoStitch Pro Meniscal Repair System, CTX-A004 (size 0). This document is primarily concerned with demonstrating substantial equivalence to a predicate device (NovoStitch Pro Meniscal Repair System, CTX-A003 (size 2-0)) rather than detailing a comprehensive study with specific acceptance criteria in the manner one might find for a novel AI/software device. The device in question is a surgical instrument that deploys sutures.

    Therefore, the information you've requested regarding AI/software performance metrics (like sensitivity, specificity, MRMC studies, training set details, ground truth establishment for training, etc.) is not applicable to this submission, as it describes a mechanical surgical device and not a software-based diagnostic or therapeutic aid.

    However, I can extract the relevant information concerning the device's acceptance criteria and performance as described in the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly present a table of numerical acceptance criteria with corresponding performance values in the way software performance data might be presented (e.g., "sensitivity > X%" and "achieved Y%"). Instead, it describes general categories of testing and the successful outcomes. The acceptance criteria are implied by the nature of the tests conducted and the statements that the device "meet all pre-established acceptance criteria."

    Acceptance Criteria Category (Implied)Reported Device Performance
    Device StrengthTesting performed in a simulated use environment; device met all pre-established acceptance criteria.
    Device ReliabilityTesting performed in a simulated use environment; device "can reliably deliver up to six stitches. This includes the preloaded handle assembly and up to five additional cartridges." Device met all pre-established acceptance criteria.
    Device InsertionTesting performed in a simulated use environment; device performed as expected. Device met all pre-established acceptance criteria.
    Suture DeploymentTesting performed in a simulated use environment; device performed as expected. Device met all pre-established acceptance criteria.
    Device RemovalTesting performed in a simulated use environment; device performed as expected. Device met all pre-established acceptance criteria.
    Ability to Reload with Suture CartridgesTesting performed in a simulated use environment; device demonstrated the ability to reload with new suture cartridges to perform additional device insertions, deployments, and removals. Device met all pre-established acceptance criteria.
    Device Joint StrengthVerification testing included device joint strength evaluation; device met all pre-established acceptance criteria.
    Device Performance EvaluationsVerification testing included device performance evaluations; device met all pre-established acceptance criteria.
    SterilizationValidated or adopted from previous testing.
    PackagingValidated or adopted from previous testing.
    Shelf LifeValidated or adopted from previous testing.
    Biocompatibility (Suture)"met all biocompatibility requirements for permanent implants in contact with tissue and/or bone" (ISO 10993-1 compliant).
    Biocompatibility (Delivery Handle - patient contact portions)"met all requirements for limited duration contact with tissue and/or bone" (ISO 10993-1 compliant).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each test category, nor does it detail data provenance in terms of country of origin or retrospective/prospective studies. It states "Product performance testing for the NovoStitch Pro Meniscal Repair System, size 0 was performed in a simulated use environment." This implies these were bench tests focused on the device's mechanical and functional capabilities.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable as the device is a surgical instrument, not a diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device would be its ability to physically perform its intended function (e.g., deploy a suture correctly, maintain strength).

    4. Adjudication Method for the Test Set

    This is not applicable for the reasons stated above. Adjudication methods are typically for subjective assessments or disagreements in diagnostic interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and Effect Size with AI Assistance

    This is not applicable. The device is a surgical tool, not an AI-assisted diagnostic or therapeutic system.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a mechanical surgical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for this device's performance evaluation is based on engineering and functional specifications in a simulated use environment (bench testing) and ISO standard requirements for biocompatibility. For instance, whether the suture was reliably deployed, if the device maintained its strength, and whether material components met biocompatibility standards.

    8. The Sample Size for the Training Set

    This is not applicable. The device is not a machine learning model, and thus does not have a "training set."

    9. How the Ground Truth for the Training Set was Established

    This is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180531

    Validate with FDA (Live)

    Date Cleared
    2018-03-29

    (29 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovoStitch Pro Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

    Device Description

    The NovoStitch Pro Meniscal Repair System is a next generation of the predicate device, the NovoStitch Plus Meniscal Repair System. The NovoStitch Pro Meniscal Repair System passes size 2-0 braided, nonabsorbable, polyethylene surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with polyethylene suture are attached.

    AI/ML Overview

    This document describes the NovoStitch Pro Meniscal Repair System, a device for meniscal repair procedures. The provided text is a 510(k) summary for the device, which focuses on demonstrating substantial equivalence to a predicate device. As such, it does not detail specific acceptance criteria for performance metrics or present a comprehensive study proving the device meets those criteria in the way one might find for a novel device with new performance claims (e.g., diagnostic accuracy for an AI system). Instead, it discusses the types of tests performed to ensure the device's basic functionality, safety, and equivalence to the predicate.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria in a table format with corresponding performance metrics like sensitivity, specificity, or AUC, as these types of metrics are not typically applicable to a mechanical surgical device like a suture delivery system in the same way they are for diagnostic AI.

    However, the text describes types of tests performed and some general statements of successful verification:

    Acceptance Criteria (Implied from Testing)Reported Device Performance
    Device strengthVerified
    Device reliabilityVerified
    Device insertionValidated
    Suture deploymentValidated
    Device removalValidated
    Ability to reload with new cartridgesValidated
    Ability to reliably deliver up to six stitchesDemonstrated
    Device joint strengthEvaluated
    Device performance evaluationsEvaluated
    Biocompatibility (suture implant)Met all requirements for permanent implants in contact with tissue and/or bone
    Biocompatibility (patient contact portions of delivery handle)Met all requirements for limited duration contact with tissue and/or bone
    Human factors and usability processes for six essential tasksConducted and conformed to FDA guidance

    2. Sample Size for Test Set and Data Provenance

    • Test Set (Human Factors/Usability): The study used 15 orthopedic surgeons.
    • Data Provenance: Not explicitly stated regarding country of origin or retrospective/prospective for the engineering/bench testing. For human factors, it was a usability study conducted in conformance with FDA guidance, implying a prospective design.

    3. Number of Experts and Qualifications for Ground Truth

    • For the human factors and usability testing, 15 orthopedic surgeons served as participants. Their qualifications are stated simply as "orthopedic surgeons." The study involved them being "trained on the use of the NovoStitch Pro per the Instructions for Use, followed by a quantitative assessment." In this context, their performance of tasks is the data being evaluated, rather than them establishing a ground truth for an external dataset.
    • For bench testing and biocompatibility, the ground truth is established by standards (e.g., ISO 10993-1) and internal engineering specifications, not by clinical experts in the same way as, for example, reading medical images.

    4. Adjudication Method for Test Set

    • Not Applicable in the traditional sense for diagnostic performance. For the human factors study, the method was a "quantitative assessment of the human factors and usability processes for six essential tasks." This likely involved metrics such as task completion rates, errors, and time to completion, rather than an expert adjudication process to resolve disagreements on a subjective classification.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is primarily relevant for diagnostic imaging devices or AI systems where multiple human readers interpret cases, and their performance with and without AI assistance is compared. The NovoStitch Pro is a surgical device, not a diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone (algorithm only) performance study was not done. This is also not applicable as the NovoStitch Pro is a mechanical surgical device, not an AI algorithm.

    7. Type of Ground Truth Used

    • Bench Testing: Ground truth is based on engineering specifications, physical measurements, and industry standards (e.g., for strength, reliability, proper articulation).
    • Biocompatibility: Ground truth is based on international standards, specifically ISO 10993-1, and specific testing protocols to determine material compatibility with biological systems.
    • Human Factors/Usability: Ground truth is based on established human factors principles and the successful or unsuccessful completion of predefined essential tasks by the users (orthopedic surgeons) according to the device's Instructions for Use.

    8. Sample Size for Training Set

    • Not Applicable. The NovoStitch Pro is a mechanical device, not an AI system, so there is no "training set" in the context of machine learning.

    9. How Ground Truth for Training Set Was Established

    • Not Applicable. As there is no training set for an AI algorithm. The device's design and manufacturing rely on engineering principles, material science, and established medical device development standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143356

    Validate with FDA (Live)

    Date Cleared
    2015-07-21

    (239 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovoStitch Plus Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.

    Device Description

    The NovoStitch Plus Meniscal Repair System passes size 2-0 braided, non-absorbable, polyethylene surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with polyethylene suture are attached.

    AI/ML Overview

    The FDA 510(k) summary for the NovoStitch Plus Meniscal Repair System provides information on its acceptance criteria and the studies conducted. However, it does not contain all the requested information, especially regarding detailed statistical analysis, sample sizes for specific tests, and expert qualifications for ground truth establishment that would typically be found in a separate study report.

    Here's what can be extracted and inferred from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document mentions "pre-established acceptance criteria" but does not explicitly list them with specific values. Instead, it describes general claims of meeting these criteria through various tests.

    Acceptance Criteria CategoryReported Device Performance (as stated in the document)
    Mechanical Performance
    Device Strength & ReliabilityValidated
    Device InsertionValidated
    Suture DeploymentValidated
    Device RemovalValidated
    Reloading CapabilityValidated; device can reliably deliver up to six stitches (1 provided + 5 additional)
    Device Joint StrengthVerified
    Delivery Device PerformanceVerified
    Suture Tensile StrengthVerified (per USP requirements for non-absorbable surgical suture)
    Suture DiameterVerified (per USP requirements for non-absorbable surgical suture)
    Usability
    Handle Deployment ForceDemonstrated equivalent usability to predicate device
    Biocompatibility
    BiocompatibilityDemonstrated compliance with ISO 10993-1 requirements for permanent implants in contact with tissue and/or bone
    Sterilization & Packaging
    SterilizationValidated
    PackagingValidated
    Shelf LifeValidated

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the various tests.

    • Test Set Sample Size: Not specified. The document mentions "product performance testing ... in a simulated use environment" and "bench testing."
    • Data Provenance: The document implies testing was conducted by Ceterix Orthopaedics Inc. ("Product performance testing for NovoStitch Plus Meniscal Repair System was performed..."), but does not specify the country of origin of the data or if it was retrospective or prospective. Given it's premarket notification, it would be prospective testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The tests seem to be primarily engineering and bench testing, not reliant on expert human interpretation for "ground truth" in the way an AI diagnostic device would be.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided as the "ground truth" for this device appears to be based on objective physical measurements and engineering specifications, rather than subjective expert assessment requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI or diagnostic imaging devices where human readers interpret results, and the document describes a surgical repair system.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable to the NovoStitch Plus Meniscal Repair System, as it is a physical surgical device, not an algorithm or AI system. Its performance inherently involves a human operator (surgeon).

    7. The Type of Ground Truth Used

    The ground truth used for this device's performance evaluation appears to be:

    • Engineering Specifications/Standards: Such as USP requirements for suture tensile strength and diameter, and ISO 10993-1 for biocompatibility.
    • Pre-established Acceptance Criteria: Defined by the manufacturer for mechanical functions (insertion, deployment, removal, strength, reliability).
    • Simulated Use Environment Outcomes: Verification that the device performs its intended functions (e.g., reliably delivers stitches).

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable in the context of this device. A "training set" is relevant for machine learning algorithms, not for physical medical devices being evaluated through bench and simulated use testing.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided and is not applicable for the same reasons as #8.