The Ceribell Instant EEG Headset is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired.
The Ceribell Instant EEG Headset is a single-use, non-sterile, disposable EEG electrode device that includes up to 20 EEG electrodes that are placed on the subject's scalp. The headset is a prescription-use device intended to collect and provide EEG signals to an EEG recording or monitoring device.
The Ceribell Instant EEG Headset includes the following components:
- Elastic fabric headset
- Electrode assemblies, each consisting of the following:
- Passive silver/silver chloride (Ag/AgCl) electrode
- Reservoir filled with conductive electrolyte gel
- Mechanism for dispensing gel onto patient scalp
- Scalp-contacting prongs to prepare scalp for electrode contact
- Cable attached to the headset to allow connection to an EEG acquisition/recording device
N/A
Ask a specific question about this device
(167 days)
The Ceribell Delirium Monitor System is intended to analyze features associated with diffuse slowing electroencephalogram (EEG) patterns that may be indicative of delirium. The Ceribell Delirium Monitor software is intended to aid in the screening and monitoring of delirium with clinical assessments in adult patients in critical care settings within hospitals.
The Ceribell Delirium Monitor System analyzes discrete segments of EEG to notify clinicians when EEG patterns associated with delirium are detected while monitoring the patient. Changes in patient condition that are detected by the device should be verified before commencing any interventions.
Ceribell's Delirium Monitor System is a novel device that uses EEG to aid in the detection of delirium. The device is comprised of the Delirium Monitor software as well as Ceribell's previously cleared EEG acquisition system (K191301) and EEG headband (K232052) for recording and processing the EEG.
Once the EEG is recorded and processed the Delirium Monitor software performs the following steps:
- Extraction of EEG features relevant to delirium assessment in 30-minute segments
- Analysis of the extracted features of each segment with a machine-learning based delirium assessment model
- Providing delirium-positive or delirium-negative assessment output to the intended users after the first 30 minutes of recording, including audible and visible notifications on the Ceribell Pocket EEG Device in the event of delirium-positive findings
- Thereafter, providing delirium-positive or delirium-negative assessment output to the intended users once every 15 minutes, generating a notification when the assessment transitions from negative to positive.
Once generated, the output of the Delirium Monitor will be interpreted by the clinician alongside other relevant information to determine whether a diagnosis of delirium is appropriate.
This document describes the acceptance criteria and the study proving the Ceribell Delirium Monitor System meets these criteria, based on the provided FDA 510(k) Clearance Letter.
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance | Comments |
|---|---|---|
| Sensitivity (lower-bound of 95% CI) $\ge$ 70% | 81% [76-87] for individual assessments | Met criteria |
| Specificity (lower-bound of 95% CI) $\ge$ 70% | 81% [78-84] for individual assessments | Met criteria |
Study Details
-
Sample Size used for the Test Set and Data Provenance:
- Sample Size: 225 adult patients (22 years or older).
- Data Provenance: Retrospective analysis of EEG data collected from patients in a critical care environment. The country of origin for the data is not specified in the provided text.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
- Number of Experts: The text states, "Clinical assessment of the patient (according to the DSM-5 criteria for delirium diagnosis) was also conducted up to three times a day by a qualified clinician." It does not specify the exact number of clinicians involved or their precise qualifications beyond "qualified clinician," though it implies expertise in delirium assessment.
-
Adjudication Method for the Test Set:
- The provided text does not explicitly detail an adjudication method beyond "clinical assessment ... by a qualified clinician." It does not mention committee review, multiple independent clinicians followed by arbitration (e.g., 2+1, 3+1), or other formal adjudication processes for establishing ground truth from disagreements. It implies a single "qualified clinician" established the ground truth for each assessment.
-
Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:
- No such study was explicitly reported. The provided text focuses on the standalone performance of the algorithm against clinical ground truth. There is no mention of human readers improving with AI assistance, nor any reported effect size for such improvement.
-
Standalone Performance:
- Yes, a standalone (algorithm only without human-in-the-loop) performance study was done. The performance metrics (Sensitivity, Specificity, AUROC, PPV, NPV, Cohen's Kappa) reported in Table 2 and Table 3 directly represent the algorithm's performance against the established clinical ground truth without human intervention in the device's output interpretation.
-
Type of Ground Truth Used:
- Expert Consensus (Clinical Diagnosis): The ground truth was established by "clinical assessment of the patient (according to the DSM-5 criteria for delirium diagnosis) ... by a qualified clinician." This indicates expert clinical diagnosis based on established medical criteria as the reference standard.
-
Sample Size for the Training Set:
- The provided document does not specify the sample size for the training set. It mentions that the "recorded EEG data and corresponding clinical delirium diagnosis were analyzed retrospectively to determine the performance of the Delirium Monitor," and later refers to "the validation dataset." This implies a separate training phase but does not provide its details. The PCCP mentions "expansion of training data and optimization of the algorithm," further suggesting a training set was used but its size is not disclosed here.
-
How the Ground Truth for the Training Set Was Established:
- The document does not explicitly state how the ground truth for the training set was established. However, given the validation set's ground truth was established by "clinical assessment ... by a qualified clinician" according to DSM-5 criteria, it is highly probable that a similar methodology was used for the training data. The PCCP notes "Ceribell's data management and algorithm development practices," which would likely encompass ground truth establishment for training data.
Ask a specific question about this device
(143 days)
The Ceribell Infant Seizure Detection Software is intended to mark previously acquired sections of EEG recordings in newborns (defined as preterm or term neonates of 25-44 weeks postmenstrual age) and infants less than 1 year of age that may correspond to electrographic seizures in order to assist qualified clinical practitioners in the assessment of EEG traces. The Seizure Detection Software also provides notifications to the user when detected seizure prevalence is "Frequent", "Abundant", or "Continuous", per the definitions of the American Clinical Neurophysiology Society Guideline 14. Delays of up to several minutes can occur between the beginning of a seizure and when the Seizure Detection notifications will be shown to a user.
The Ceribell Infant Seizure Detection Software does not provide any diagnostic conclusion about the subject's condition and Seizure Detection notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert.
The Ceribell Infant Seizure Detection Software is a software-only device that is intended to mark previously acquired sections of EEG recordings that may correspond to electrographic seizures in order to assist qualified clinical practitioners in the assessment of EEG traces.
Ceribell Infant Seizure Detection Software: Acceptance Criteria and Supporting Study
1. Table of Acceptance Criteria and Reported Device Performance
| Activity Category | Metric | Acceptance Criteria | Device Performance (Overall) | 95% Confidence Interval | Meets Criteria? |
|---|---|---|---|---|---|
| Seizure Episodes with Seizure Burden ≥10% (Frequent) | PPA | Lower bound of 95% CI ≥ 70% | 91.36% | [85.71, 94.91] | Yes |
| FP/hr | Upper bound of 95% CI ≤ 0.446 FP/hr | 0.204 | [0.180, 0.230] | Yes | |
| Seizure Episodes with Seizure Burden ≥50% (Abundant) | PPA | Lower bound of 95% CI ≥ 70% | 91.23% | [82.67, 96.57] | Yes |
| FP/hr | Upper bound of 95% CI ≤ 0.446 FP/hr | 0.083 | [0.069, 0.100] | Yes | |
| Seizure Episodes with Seizure Burden ≥90% (Continuous) | PPA | Lower bound of 95% CI ≥ 70% | 91.18% | [75.00, 100.00] | Yes |
| FP/hr | Upper bound of 95% CI ≤ 0.446 FP/hr | 0.057 | [0.045, 0.072] | Yes |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 713 patients.
- 25-36 weeks PMA: 155 patients
- 37-44 weeks PMA: 321 patients
-
44 weeks PMA: 237 patients
- Data Provenance: The EEG recordings were obtained from patients less than 1 year of age who received continuous EEG monitoring within the hospital environment. The study was retrospective. The country of origin is not explicitly stated in the provided text.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: 3
- Qualifications of Experts: Expert pediatric neurologists who were fellowship-trained in epilepsy or clinical neurophysiology.
4. Adjudication Method for the Test Set
- Adjudication Method: A two-thirds majority agreement among the 3 expert pediatric neurologists was required to form a determination of seizures, establishing the reference standard for the test set.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. The study focused on the standalone performance of the algorithm against an expert-adjudicated ground truth.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, a standalone performance study was done. The performance metrics (PPA and FP/hr) were evaluated for the Ceribell Infant Seizure Detection Software algorithm without human intervention in the detection process. The reviewing neurologists for ground truth establishment were explicitly blinded to the software's output.
7. The Type of Ground Truth Used
- Type of Ground Truth: Expert consensus (adjudication by a panel of 3 expert pediatric neurologists).
8. The Sample Size for the Training Set
- The sample size for the training set is not provided in the document. The document states, "Importantly, none of the data in the validation dataset were used for training of the Seizure Detection algorithm; the validation dataset is completely independent."
9. How the Ground Truth for the Training Set Was Established
- The document does not explicitly state how the ground truth for the training set was established. It only mentions that the validation dataset was independent and not used for training.
Ask a specific question about this device
(182 days)
The Ceribell Instant EEG Headcap is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired. The headcap shall be placed on the head of infants, including newborns and preterm babies.
The Ceribell Instant EEG Headcap is a single-use, non-sterile, disposable EEG electrode device that includes a minimum of 9 EEG electrodes that are placed on the subject's scalp. The Headcap is intended to collect and provide EEG signals to an EEG recording or monitoring device.
The Ceribell Headcap includes the following components:
- An elastic fabric headcap
- An elastic fabric chin strap
- Nine (9) silver/silver chloride (Ag/AgCl) electrodes
- A cable attached to the headcap to allow connection to an EEG acquisition/recording device
N/A
Ask a specific question about this device
(310 days)
The Ceribell Seizure Detection Software is intended to mark previously acquired sections of EEG recordings in patients greater or equal to 1 year of age that may correspond to electrographic seizures in order to assist qualified clinical practitioners in the assessment of EEG traces. The Seizure Detection Software also provides notifications to the user when detected seizure prevalence is "Frequent", "Abundant", or "Continuous, per the definitions of the American Clinical Neurophysiology Society Guideline 14. Delays of up to several minutes can occur between the beginning of a seizure and when the Seizure Section notifications will be shown to a user.
The Ceribell Seizure Detection Software does not provide any diagnostic conclusion about the subject's condition and Seizure Detection notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a training expert.
The Ceribell Seizure Detection Software is a software-only device that is intended to mark previously acquired sections of EEG recordings that may correspond to electrographic seizures in order to assist qualified clinical practitioners in the assessment of EEG traces.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter:
1. Table of Acceptance Criteria & Reported Device Performance
| Metric / Category | Acceptance Criteria (95% CI) | Reported Device Performance (95% Confidence Interval) | Pass / Fail |
|---|---|---|---|
| Positive Percent Agreement (PPA) | Lower bound ≥ 70% PPA for each threshold | ||
| Seizure Burden ≥10% (Frequent) | |||
| Ages 1-11 | Lower bound ≥ 70% | 96.12% [88.35, 99.28] | Pass |
| Ages 12-17 | Lower bound ≥ 70% | 87.01% [73.16, 93.55] | Pass |
| Ages 18+ | Lower bound ≥ 70% | 95.71% [91.30, 98.43] | Pass |
| Overall | Lower bound ≥ 70% | 93.93% [90.03, 96.52] | Pass |
| Seizure Burden ≥50% (Abundant) | |||
| Ages 1-11 | Lower bound ≥ 70% | 96.67% [87.50, 100.00] | Pass |
| Ages 12-17 | Lower bound ≥ 70% | 95.45% [73.33, 100.00] | Pass |
| Ages 18+ | Lower bound ≥ 70% | 96.72% [88.37, 100.0] | Pass |
| Overall | Lower bound ≥ 70% | 96.50% [92.12, 98.77] | Pass |
| Seizure Burden ≥90% (Continuous) | |||
| Ages 1-11 | Lower bound ≥ 70% | 92.59% [76.00, 100] | Pass |
| Ages 12-17 | Lower bound ≥ 70% | 100.0% [100, 100] | Pass |
| Ages 18+ | Lower bound ≥ 70% | 93.55% [78.26, 100.0] | Pass |
| Overall | Lower bound ≥ 70% | 94.12% [85.45, 98.48] | Pass |
| False Positive rate per hour (FP/hr) | Upper bound ≤ 0.446 FP/hr for each threshold | ||
| Seizure Burden ≥10% (Frequent) | |||
| Ages 1-11 | Upper bound ≤ 0.446 | 0.2700 [0.2445, 0.2986] | Pass |
| Ages 12-17 | Upper bound ≤ 0.446 | 0.2141 [0.1920, 0.2394] | Pass |
| Ages 18+ | Upper bound ≤ 0.446 | 0.1343 [0.1250, 0.1445] | Pass |
| Overall | Upper bound ≤ 0.446 | 0.1763 [0.1670, 0.1859] | Pass |
| Seizure Burden ≥50% (Abundant) | |||
| Ages 1-11 | Upper bound ≤ 0.446 | 0.1561 [0.1369, 0.1772] | Pass |
| Ages 12-17 | Upper bound ≤ 0.446 | 0.0921 [0.0776, 0.1082] | Pass |
| Ages 18+ | Upper bound ≤ 0.446 | 0.0547 [0.0480, 0.0615] | Pass |
| Overall | Upper bound ≤ 0.446 | 0.08180 [0.0754, 0.0885] | Pass |
| Seizure Burden ≥90% (Continuous) | |||
| Ages 1-11 | Upper bound ≤ 0.446 | 0.0843 [0.0697, 0.1006] | Pass |
| Ages 12-17 | Upper bound ≤ 0.446 | 0.0399 [0.0301, 0.0511] | Pass |
| Ages 18+ | Upper bound ≤ 0.446 | 0.0249 [0.0204, 0.0299] | Pass |
| Overall | Upper bound ≤ 0.446 | 0.03951 [0.0351, 0.0443] | Pass |
2. Sample Size and Data Provenance for the Test Set
- Sample Size for Test Set:
- Total Number of Patients: 1701
- Ages 1-11: 450 patients
- Ages 12-17: 392 patients
- Ages 18+: 859 patients
- Total Number of Patients: 1701
- Data Provenance: The EEG recordings dataset used for performance validation was gathered from real-world clinical usage of the Ceribell Pocket EEG Device. The specific country of origin is not explicitly stated, but it's implied to be from acute care hospital settings where the predicate device (Ceribell Pocket EEG Device) is used. The data is retrospective as it was previously acquired. There were no patient inclusion or exclusion criteria applied, indicating a representative sample of the intended patient population.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: More than three expert neurologists (implied by "a two-thirds majority agreement was required"). The tables later specify "3 expert reviewers" for the seizure burden distribution, suggesting at least 3, and possibly more given the "two-thirds majority" rule.
- Qualifications of Experts: Fellowship trained in epilepsy or neurophysiology. No specific years of experience are mentioned.
4. Adjudication Method for the Test Set
- Adjudication Method: A two-thirds majority agreement among the expert neurologists was required to establish the ground truth for seizures. This implies a method of consensus. The experts were fully blinded to the outputs of the Seizure Detection Software.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No. The documentation describes a standalone algorithm performance study, not a comparative effectiveness study involving human readers with and without AI assistance.
- Effect Size: Not applicable, as no MRMC study was performed.
6. Standalone Algorithm Performance
- Was a standalone study done? Yes. The study directly evaluates the "performance of the Seizure Detection algorithm" by comparing its output (algorithm marks/notifications) against the expert-established ground truth. The algorithm's PPA and FP/hr metrics are presented, which are standard for standalone AI performance.
7. Type of Ground Truth Used
- Type of Ground Truth: Expert consensus (specifically, a two-thirds majority agreement among fellowship-trained neurologists reviewing EEG recordings). This is clinical expert ground truth based on visual review of EEG.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not explicitly stated. The document only mentions that "none of the data in the validation dataset were used for training of the Seizure Detection algorithm; the validation dataset is completely independent." This ensures the integrity of the test set but does not provide information about the training set size.
9. How Ground Truth for the Training Set Was Established
- Ground Truth Establishment for Training Set: Not explicitly stated. The document focuses exclusively on the validation dataset's ground truth methodology. However, given the nature of deep learning models, it's highly probable that the training data also underwent a rigorous ground truth labeling process, likely similar to (or potentially identical in methodology to) the validation set, though this is not detailed here.
Ask a specific question about this device
(29 days)
The Ceribell Instant EEG Headband is an electroencephalogram (EEG) electrode array intended for single patient use in the recording of EEGs. The Instant EEG Headband is intended for prescription use in the home, healthcare facility, or clinical research environment.
The Ceribell Instant EEG Headband is a 10 electrode EEG headband is non-sterile and disposable for single patient use and designed to be used exclusively with the Ceribell Pocket EEG Device (K191301) for EEG acquisition and recording.
The Ceribell Instant EEG Headband is comprised of the following components:
- An elastic fabric headband
- A cable attached to the headband to allow connection to an EEG acquisition/recording device
- 10 electrode assemblies, each consisting of the following:
- Passive Silver/silver-chloride electrode
- Reservoir filled with conductive electrolyte gel
- Mechanism for dispensing gel onto patient scalp
- Scalp-contacting prongs to prepare scalp for electrode contact
This FDA 510(k) summary for the Ceribell Instant EEG Headband (K232052) explicitly states "Performance Data: None required." and that the device is identical to its predicate (K210805) with only a modified Indications for Use Statement and associated labeling.
Therefore, the provided document does not contain the information necessary to answer the questions regarding acceptance criteria and a study that proves the device meets those criteria. It indicates that no performance data was needed for this specific submission because the device is essentially a re-labeled version of a previously cleared device.
Ask a specific question about this device
(182 days)
The Ceribell Status Epilepticus Monitor software is indicated for the diagnosis of Electrographic Status Epilepticus in patients greater than or equal to 18 years of age who are at risk for seizure. The Ceribell Status Epilepticus Monitor software analyzes EEG waveforms and identifies patterns that may be consistent with electrographic status epileptious as defined in the American Clinical Neurophysiology Society's Guideline 14.
The diagnostic output of the Ceribell Status Epilepticus Monitor is intended to be used as an aid for determining patient treatment in acute-care environments. The device's diagnosis of Electrographic Status Epilepticus provides one input to the clinician that is intended to be used in conjunction with other elements of clinical practice to determine the appropriate treatment course for the patient.
The Ceribell Status Epilepticus Monitor is intended for diagnosis of Electrographic Status Epilepticus only. The device does not substitute for the review of the underlying EEG by a qualified clinician with respect to any other types of pathological EEG patterns. The device is not intended for use in Epilepsy Monitoring Units.
The Ceribell Status Epilepticus Monitor is a software as medical device that analyzes EEG waveforms for the intended use of recognizing electrographic status epilepticus (ESE). The subject device software is intended for use only with the Ceribell Pocket EEG Device (K191301), which is also the predicate device. The predicate device contains a software module that performs detection of seizures in a similar manner as the subject device. The user workflow and instructions for starting an EEG recording on a patient are unchanged compared to the predicate device.
The user places the Ceribell Instant EEG Headband (K210805) on the patient, the headband contains 10 electrodes that are arranged in a bipolar montage and correspond to the following locations following the 10-20 electrode naming convention: Fp1, F7, T3, T5, O1, Fp2, F8, T3, T6, O2. The 10 electrodes form 8 channels (4 on the left hemisphere, 4 on the right hemisphere) that are analyzed by the subject device's ESE detection algorithm.
Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
The Ceribell Status Epilepticus Monitor underwent clinical validation to assess its performance in diagnosing Electrographic Status Epilepticus (ESE). The primary performance metrics were sensitivity and specificity.
1. Table of Acceptance Criteria and Reported Device Performance
| Measure | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Sensitivity | High sensitivity for ESE detection to minimize false negatives and ensure timely treatment. Specific quantitative acceptance criteria were not explicitly stated as numerical thresholds (e.g., "X% sensitivity") but the document emphasizes "100% sensitive" for this aspect. | 100% |
| Specificity | Sufficient specificity to avoid unnecessary treatment in ESE-negative cases. Specific quantitative acceptance criteria were not explicitly stated as numerical thresholds. | 94% |
2. Sample Size for the Test Set and Data Provenance
- Sample Size: 350 subjects
- Data Provenance: Retrospective study using previously-collected EEG data from 6 hospitals of varying size and geographic locations. The data represents "real-world data" from a fixed one-year time period. All included subjects were ≥ 18 years of age.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: A team of "qualified neurologists" (specific number not given, but plural).
- Qualifications of Experts: Described as "qualified neurologists." No further details on years of experience or specific subspecialties are provided in the document.
4. Adjudication Method for the Test Set
- Adjudication Method: "Majority opinion of the expert reviewers." This indicates a consensus-based approach. For example, if there were three reviewers, at least two would need to agree.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly stated as having been performed. The study focused on the standalone performance of the algorithm against expert-established ground truth.
- Effect Size of Human Readers with vs. Without AI Assistance: Not applicable, as an MRMC study with human-in-the-loop performance was not described.
6. Standalone (Algorithm Only) Performance
- Standalone Performance: Yes, the study describes the standalone (algorithm-only) performance of the Ceribell Status Epilepticus Monitor. The device algorithm was run on the collected dataset, and its output was compared against the expert-established ground truth.
7. Type of Ground Truth Used
- Ground Truth Type: Expert consensus. Specifically, a "team of qualified neurologists independently review[ed] and categorize[d] each EEG; the ground-truth reference standard is established by a majority opinion of the expert reviewers."
8. Sample Size for the Training Set
- The document does not explicitly state the sample size used for the training set. The clinical validation study described is a retrospective study of the algorithm's performance on a test set (350 subjects).
9. How the Ground Truth for the Training Set Was Established
- The document does not explicitly state how the ground truth for the training set was established, as details about the training phase itself are not provided in this summary. The focus is on the clinical validation study.
Ask a specific question about this device
(173 days)
The Ceribell Instant EEG Headcap is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired.
The Ceribell Instant EEG Headcap is a single-use, non-sterile, disposable EEG electrode device that includes a minimum of 9 EEG electrodes that are placed on the subject's scalp. The Headcap is intended to collect and provide EEG signals to an EEG recording or monitoring device.
The Ceribell Instant EEG Headcap is comprised of the following components:
- An elastic fabric headcap
- An elastic chin strap
- A minimum of 9 silver/silver chloride (Ag/AgCl) electrodes
- A cable attached to the headcap to allow connection to an EEG acquisition/recording device
The Ceribell Instant EEG Headcap is a cutaneous electrode device. The provided text outlines the acceptance criteria and performance data for this device, comparing it to a predicate device (Wuhan Greentek Disposable EEG Electrodes, K200162) to establish substantial equivalence.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Ceribell Instant EEG Headcap are based on the same requirements as its predicate device and defined by FDA-recognized consensus standards, primarily ANSI/AAMI EC12 Disposable ECG Electrodes and ISO 10993 series for biocompatibility.
| Attribute | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Electrical Performance | ||
| Average AC Impedance | ≤ 2 kΩ (individual pairs ≤ 3kΩ) (ANSI/AAMI EC12) | Met the same requirements per FDA guidance document. |
| DC Offset Voltage | ≤ 100 mV (ANSI/AAMI EC12) | Met the same requirements per FDA guidance document. |
| Combined Offset Instability and Internal Noise | ≤ 150 μV (ANSI/AAMI EC12) | Met the same requirements per FDA guidance document. |
| Bias Current Tolerance (DC Voltage Offset) | ≤ 100 mV (ANSI/AAMI EC12) | Met the same requirements per FDA guidance document. |
| Biocompatibility | ||
| Cytotoxicity | ISO 10993-5 | Met the same requirements per FDA guidance document "Use of International Standard ISO 10993-1". |
| Sensitization, Irritation or Intracutaneous Reactivity | ISO 10993-10 | Met the same requirements per FDA guidance document "Use of International Standard ISO 10993-1". |
| Electrical Connection Compliance | Conductive Connection Compliance per ES 60601-1, IEC 60601-1 clause 8.5.2.3, 21 CFR 898.12 | Met the same requirements per FDA guidance document. |
| Shelf Life | ANSI/AAMI EC12 and IEC 60601-2-2 | Testing completed (results not explicitly detailed, but implied compliance). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample size for the test sets (e.g., number of EEG headcaps or subjects) used for each performance test. It only mentions that performance data was "provided" and "completed testing" was done.
Data provenance (e.g., country of origin, retrospective or prospective) is also not specified in the provided text.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not applicable to this type of device and study. The Ceribell Instant EEG Headcap is a physical medical device (an electrode headcap), not an AI algorithm or diagnostic tool that requires expert interpretation for ground truth establishment. The performance criteria are objective measurements of electrical properties and biocompatibility.
4. Adjudication Method for the Test Set
This is not applicable as the tests performed are objective physical and electrical measurements, not interpretive clinical assessments requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) use AI assistance to improve their diagnostic accuracy. The Ceribell Instant EEG Headcap is a hardware device for signal acquisition.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical electrode headcap, not an algorithm. Its performance is measured through objective electrical and biocompatibility testing, not algorithmic accuracy.
7. The Type of Ground Truth Used
The "ground truth" for the tests performed on the Ceribell Instant EEG Headcap is based on established industry standards and objective measurable physical and electrical properties. For example:
- Electrical Performance: The ground truth is defined by the limits set in ANSI/AAMI EC12.
- Biocompatibility: The ground truth is defined by the pass/fail criteria of ISO 10993 tests (e.g., absence of cytotoxicity, sensitization, irritation).
8. The Sample Size for the Training Set
This is not applicable. The Ceribell Instant EEG Headcap is a physical medical device, not an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the same reason as above.
Ask a specific question about this device
(166 days)
The Ceribell Instant EEG Headband is an electroencephalogram (EEG) electrode array intended for single patient use in the recording of EEGs in patients of 2 years and older. The Instant EEG Headband is intended for prescription use in the home, healthcare facility, or clinical research environment.
The Ceribell Instant EEG Headband is a 10 electrode EEG headband is non-sterile and disposable for single patient use and designed to be used exclusively with the Ceribell Pocket EEG Device (K191301) for EEG acquisition and recording.
The Ceribell Instant EEG Headband is comprised of the following components:
- An elastic fabric headband
- A cable attached to the headband to allow connection to an EEG acquisition/recording device
- 10 electrode assemblies, each consisting of the following:
- Passive Silver/silver-chloride electrode
- Reservoir filled with conductive electrolyte gel
- Mechanism for dispensing gel onto patient scalp
- Scalp-contacting prongs to prepare scalp for electrode contact
The provided document is a 510(k) premarket notification decision letter from the FDA to Ceribell, Inc. regarding their Instant EEG Headband. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study report of the device's diagnostic performance against established acceptance criteria.
The letter explicitly states:
"We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..."
The "Performance Data" section within the 510(k) Summary lists the types of testing performed:
- Biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10
- Bench testing and simulated use testing to verify system performance including testing and evaluation to ANSI EC12-2000 and IEC 60601-1 requirements.
- Shelf life testing
These tests are related to the safety and functionality of the electrode array itself (biocompatibility, electrical safety, durability, and signal acquisition quality) and not to the diagnostic performance of an AI algorithm or human-in-the-loop performance related to interpreting EEG signals. The device described is a cutaneous electrode array, not an AI-powered diagnostic tool.
Therefore, the requested information regarding acceptance criteria for AI algorithm performance, sample size for test sets, expert adjudication, MRMC studies, or standalone algorithm performance is not present in this document because the device being cleared is an EEG headband (hardware), not a software algorithm for interpreting EEG.
To answer your request, if this were an AI/software as a medical device (SaMD) submission, the following sections would typically be present, but they are absent here due to the nature of the device:
1. Table of Acceptance Criteria and Reported Device Performance: This would typically define metrics (e.g., sensitivity, specificity, AUC) and their target values, along with the actual results from the study. (Not applicable/Not present)
2. Sample size used for the test set and data provenance:
* Sample Size: Not applicable.
* Data Provenance: Not applicable.
3. Number of experts used to establish ground truth and qualifications:
* Number of Experts/Qualifications: Not applicable, as there is no diagnostic AI component requiring ground truth for interpretation.
4. Adjudication method for the test set:
* Adjudication Method: Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
* MRMC Study: No, this is a hardware device.
6. If a standalone (algorithm only) performance was done:
* Standalone Performance: No, this is a hardware device.
7. The type of ground truth used:
* Ground Truth: Not applicable.
8. The sample size for the training set:
* Training Set Sample Size: Not applicable.
9. How the ground truth for the training set was established:
* Training Set Ground Truth Establishment: Not applicable.