    K Number
    K222459

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-27

    (438 days)

    Product Code
    Regulation Number
    872.3260
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Treatment of Dental Hypersensitivity for use in adults over the age of 21.

    Device Description

    The Centrix FluoroSilver is a professional tooth desensitizer that is composed of Aqueous Silver Diamine Fluoride, 38.3% to 43.2% w/v. The solution is composed of water, ammonia, fluoride, silver and a colorant. The device is provided in a single 5 ml squeezable dropper bottle or a multi-pack which includes three 5 ml squeezable dropper bottles. Each variation is packaged in a preformed trav and includes applicators and the Instructions for Use (IFU). The primary mode of action is to reduce dentinal hypersensitivity by occluding open dentin tubules by the formation of a precipitate.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device (Centrix FluoroSilver Silver Diamine Fluoride 38%). This type of document is provided when a company seeks to demonstrate that their new device is substantially equivalent to a legally marketed predicate device. As such, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to demonstrating substantial equivalence through various performance and material characteristic comparisons, rather than clinical efficacy trials with predefined clinical acceptance criteria as would be typical for a PMA.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission, "acceptance criteria" are implicitly defined by the characteristics of the predicate device and relevant standards. The goal is to show the new device is "substantially equivalent."

    Characteristic / ParameterImplicit "Acceptance Criteria" (Predicate Device Performance/Standard)Reported Device Performance (Centrix FluoroSilver)Discussion / Substantial Equivalence
    Primary Mode of ActionReduce dentinal hypersensitivity by occluding open dentin tubules by formation of a precipitate.Reduce dentinal hypersensitivity by occluding open dentin tubules by formation of a precipitate.Identical
    Indications for UseTreatment of dentinal hypersensitivity for use in adults over the age of 21.Treatment of Dental Hypersensitivity for use in adults over the age of 21.Identical
    SDF Concentration (w/v)38.3% to 43.2% (of predicate, based on general description)38.3% to 43.2% w/v (stated in device description)Identical range. The comparison table states 38% for both.
    Silver Diamine Fluoride (SDF)360 - 400 g/L (as per predicate's typical specifications)360 - 400 g/LSpecifications match.
    Fluoride50 - 60 g/L50 - 60 g/LSpecifications match.
    pH8.5 - 9.58.5 - 9.5Specifications match.
    Density1.265 - 1.276 g/L1.265 - 1.276 g/LSpecifications match.
    Cytotoxicity (ISO 10993-5)Positive cytotoxicity (predicate device)"Unsuitable for analysis" due to high precipitate, and performed identically to predicate.Both devices showed positive cytotoxicity when tested, and their performance was identical ("unsuitable for analysis" due to precipitate formation). This is considered equivalent for the purpose of demonstrating safety profile similarity, given the context of a dentin desensitizer that causes precipitation.
    Skin Sensitization (ISO 10993-10)Not considered a contact skin sensitizer (presumably the predicate also meets this)NOT considered to be a contact skin sensitizer.Acceptable and considered safe. There is no explicit statement about the predicate's performance on this specific test, but the conclusion for the subject device is a positive safety finding.
    Irritation (ISO 10993-23)Non-Irritant to oral mucosa (presumably the predicate also meets this)Non-Irritant to oral mucosa.Acceptable and considered safe. Similar to skin sensitization, this is a positive safety finding.
    Acute Systemic Toxicity (ISO 10993-11)No signs of gross toxicity, adverse clinical effects, or abnormal behavior (predicate device)No signs of gross toxicity, adverse clinical effects, or abnormal behavior. No gross abnormalities noted after 3 day observation and necropsy. Performed identically to predicate.Identical performance with the predicate. Both met requirements.
    SEM AnalysisNo discernible difference in silver deposition (predicate expected to have robust deposition)Silver deposition on dentin and in tubules between Centrix SDF and Elevate SDF showed no discernible difference based on SEM analyses. Significant deposits of silver on cut surface of tooth and within dentinal tubules.Identical. Confirms similar physical mechanism of action (tubule occlusion).
    Fluoride Release0.024 +/- 0.006 ppmF/mm² (predicate device)0.023 +/- 0.003 ppmF/mm²Identical. No statistical difference confirmed equivalent fluoride release.
    Dentin Permeability Reduction97.5% +/- 2.5% (predicate device)97.5% +/- 4.3%Identical. Both devices reduced dentin permeability by an average of 97.5%. This is a key functional performance criterion directly related to the indication for use.
    Chemical Composition ProfileNearly Identical (predicate device)Strong match between subject and predicate device via MS analysis for chemical composition.Nearly identical. Confirmed by Mass Spectrometry. Differences in colorant (Fast Green FCF Dye vs. FD&C Blue 1) are noted but considered minor and not impacting substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a "test set" in the sense of a clinical trial for performance. Instead, it describes various laboratory and in vitro/in vivo (animal) studies for substantial equivalence.

    • Cytotoxicity (ISO 10993-5): No specific sample size mentioned, likely standardized by the ISO method.
    • Skin Sensitization (ISO 10993-10): No specific sample size mentioned, likely standardized by the ISO method (typically animal models).
    • Irritation (ISO 10993-23): No specific sample size mentioned, likely standardized by the ISO method (typically animal models for oral mucosa).
    • Acute Systemic Toxicity (ISO 10993-11): No specific sample size mentioned, likely standardized by the ISO method (typically animal models like rats or mice).
    • SEM Analysis: "Human tooth samples" used. Number not specified.
    • Fluoride Release Testing: "SDF treated tooth cross sections" used. Number not specified.
    • Dentin Permeability: "Human molar teeth samples" used. Number not specified, but the results include standard deviations, implying a sample size greater than one.

    Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. Given that these are bench and preclinical studies for a 510(k), they are generally prospective laboratory studies. "Human molar teeth samples" would typically be ethically sourced from dental extractions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This type of information is generally not applicable to a 510(k) submission for substantial equivalence based on bench/preclinical testing. There is no "ground truth" established by experts in the context of interpreting images or clinical outcomes data. The "ground truth" for the tests mentioned (e.g., cytotoxicity, fluoride release, dentin permeability) is determined by the objective measurements and protocols of the respective ISO standards or laboratory methodologies.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving human readers or expert consensus on qualitative data (e.g., image interpretation, clinical diagnosis). The studies presented are objective, quantitative laboratory tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study assesses the performance of human readers, typically in interpreting medical images, with and without AI assistance. The device in question (Centrix FluoroSilver Silver Diamine Fluoride 38%) is a chemical product for treating dental hypersensitivity, not an imaging or diagnostic AI device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a chemical substance, not a software algorithm or an AI device. The term "standalone performance" for an algorithm does not apply here. The performance of the substance itself was evaluated.

    7. The Type of Ground Truth Used

    The "ground truth" used for these substantial equivalence tests were objective, quantitative measurements and observations derived from standardized laboratory tests:

    • Cytotoxicity: Observation of cellular response in vitro according to ISO 10993-5.
    • Skin Sensitization: Dermal response in animal models according to ISO 10993-10.
    • Irritation: Mucosal response in animal models according to ISO 10993-23.
    • Acute Systemic Toxicity: Clinical observations and necropsy findings in animal models according to ISO 10993-11.
    • SEM Analysis: Direct microscopic observation of silver deposition on tooth samples.
    • Fluoride Release Testing: Quantitative measurement of fluoride concentration.
    • Dentin Permeability: Quantitative measurement of water flow through dentin.
    • Chemical Identity Testing: Mass Spectrometry data for chemical composition.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device where a "training set" would be used to develop an algorithm. The device is a chemical product.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200764

    Validate with FDA (Live)

    Device Name
    STATSTIX
    Date Cleared
    2020-10-15

    (205 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StatStix is applied to control temporary bleeding caused by minor cuts and abrasions that may occur on gingival tissue during dental procedures.

    Device Description

    StatStix is designed incorporating a hand-held polypropylene handle with a nylon polyamide applicator tip coated with a dried solution of 25% Aluminum Sulfate Hydrate Solution with Mint Flavor. The device is configured in two sizes to help control minor gingival bleeding intended for transient use by a dentist during dental restorative procedures.

    AI/ML Overview

    The provided FDA document describes the premarket notification (510(k)) for the StatStix device (K200764). It details the device's intended use and compares its characteristics to a predicate device for substantial equivalence. However, it explicitly states that clinical testing has not been performed on this product. Therefore, a study proving the device meets acceptance criteria based on clinical performance in humans is not present in this document.

    The document does, however, discuss non-clinical performance data and establishes acceptance criteria for those tests.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):

    Acceptance Criteria (Non-Clinical)Reported Device Performance (Non-Clinical)
    Biocompatibility:Meets ISO 10993-5 (cytotoxicity)
    - CytotoxicityMeets ISO 10993-10 (sensitization/irritation)
    - Sensitization/Irritation
    Physical Properties:Performed according to standard laboratory benchtop practices
    - FTIR
    - pH
    - Dosage Weight
    Shelf-Life:Allows for a two-year expiration dating
    - Consistent with ASTM F1980

    2. Sample Size Used for the Test Set and Data Provenance:

    • Clinical Test Set: Not applicable, as no clinical testing was performed for StatStix.
    • Non-Clinical Test Set: Not specified for biocompatibility, physical properties, or shelf-life testing. The document states "standard laboratory benchtop practices" for physical properties and "consistent with ASTM F1980" for shelf-life, implying standard sample sizes for these types of tests would have been used, but specific numbers are not detailed.
    • Data Provenance: Not explicitly stated, but assumed to be from Centrix Incorporated or their contracted laboratories given they are the submitter.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • Clinical Test Set: Not applicable.
    • Non-Clinical Test Set: Not specified. For non-clinical lab tests, "ground truth" is established by the methodologies and reference standards themselves, as opposed to expert human interpretation.

    4. Adjudication Method for the Test Set:

    • Clinical Test Set: Not applicable.
    • Non-Clinical Test Set: Not applicable. Adjudication methods are typically relevant for subjective assessments or when multiple human readers/experts are involved in ground truth establishment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states, "Clinical testing has not been performed on this product." Therefore, there is no information on how much human readers improve with AI vs. without AI assistance. The StatStix device is a physical product (applicator with astringent), not an AI-powered diagnostic or assistive tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • No, a standalone performance study in the context of an algorithm or AI was not done. StatStix is a physical device, not an algorithm.

    7. Type of Ground Truth Used:

    • Clinical Ground Truth: Not applicable, as no clinical testing was performed.
    • Non-Clinical Ground Truth: Established through standardized laboratory testing methodologies (e.g., ISO, ASTM, FTIR/pH/dosage weight analysis) where the outcomes are measurable and objective.

    8. Sample Size for the Training Set:

    • Not applicable. This device is a physical product, not an AI/machine learning algorithm requiring a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set for this type of device.