Search Filters

Search Results

Found 33 results

510(k) Data Aggregation

    K Number
    K260812

    Validate with FDA (Live)

    Date Cleared
    2026-03-30

    (17 days)

    Product Code
    Regulation Number
    866.4000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Francisella tularensis Real-time PCR Assay is an in vitro diagnostic test for the qualitative detection of chromosomal DNA sequences from F. tularensis. The assay can be used to test human whole blood EDTA, pleural fluid, and bacterial culture isolates grown on agar from individuals suspected of having tularemia.

    Results generated from direct specimen testing are presumptive for the identification of F. tularensis. Results generated from culture isolate testing are used as part of the LRN Francisella tularensis Testing Algorithm. The diagnosis of F. tularensis infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of F. tularensis from culture isolates or from clinical specimens.

    Negative results do not preclude infection with the biothreat microbial agents targeted by the device and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Use is limited to Centers for Disease Control and Prevention (CDC) designated laboratories.

    Device Description

    The Francisella tularensis Real-time PCR Assay is composed of oligonucleotide primers and dual-labeled hydrolysis probes (Taqman®) for use in real-time PCR reactions. This assay is intended for the in vitro detection of Francisella tularensis in suspect infected persons with fever and other symptoms clinically compatible with acute tularemia. Extracted DNA samples are tested using the Francisella tularensis Real-time PCR Assay along with an extraction control primer and probe set to demonstrate adequate DNA extraction and isolation, specimen integrity, proper function of common reagents and equipment, and the absence of inhibitory substances.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252072

    Validate with FDA (Live)

    Date Cleared
    2025-09-30

    (91 days)

    Product Code
    Regulation Number
    866.4000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Francisella tularensis Real-time PCR Assay is an in vitro diagnostic test for the qualitative detection of chromosomal DNA sequences from Francisella tularensis. The assay can be used to test whole blood EDTA, pleural fluid, and bacterial culture isolates grown on agar from individuals suspected of having tularemia.

    Results generated from direct specimen testing are presumptive for the identification of Francisella tularensis. Results generated from culture isolate testing are used as part of the LRN Francisella tularensis Testing Algorithm. The diagnosis of Francisella tularensis infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of Francisella tularensis from culture isolates or from clinical specimens.

    Negative results do not preclude infection with the biothreat microbial agents targeted by the device and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Use is limited to Centers for Disease Control and Prevention (CDC) designated laboratories.

    Device Description

    The Francisella tularensis Real-time PCR Assay is composed of oligonucleotide primers and dual-labeled hydrolysis probes (Taqman®) for use in real-time PCR reactions. This assay is intended for the in vitro detection of Francisella tularensis in individuals suspected of having tularemia. Extracted DNA samples are tested using the Francisella tularensis Real-time PCR Assay along with an extraction control primer and probe set to demonstrate adequate DNA extraction and isolation, specimen integrity, proper function of common reagents and equipment, and the absence of inhibitory substances.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Influenza A/B Typing Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • For qualitative detection of influenza virus type A or B viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture.

    • To provide epidemiological information for surveillance of circulating influenza viruses.

    Performance characteristics for influenza were established during a season when seasonal influenza viruses A(H1N1) and A(H3N2) were the predominant influenza A viruses in circulation and during a season when the A(H1N1)pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted unless a BSL 3E facility is available to receive and culture specimens.

    CDC Human Influenza Real-Time RT-PCR Diagnostic Panel: Influenza A Subtyping Kit (VER 4)

    The Influenza A Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • For determination of the subtype of seasonal human influenza A viruses as seasonal A(H3) and/or A(H1)pdm09 from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture;

    • To provide epidemiological information for surveillance of circulating influenza viruses.

    Performance characteristics for influenza were established during a season when seasonal influenza viruses A(H1N1) and A(H3N2) were the predominant influenza A viruses in circulation and during a season when the A(H1N1)pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted unless a BSL 3E facility is available to receive and culture specimens.

    CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit (VER 1.1 and 2)

    The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • For the determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture;

    • To provide epidemiologic information for surveillance of circulating influenza viruses.

    Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/Victoria and B/Yamagata lineages were in circulation.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.

    CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza A/H5 Subtyping Kit (VER 4)

    The Influenza A/H5 Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • For the presumptive identification of virus in patients who may be infected with influenza A subtype A(H5) (Asian lineage) from viral RNA in human respiratory specimens, conjunctival swabs, and viral culture in conjunction with clinical and epidemiological risk factors;

    • To provide epidemiologic information for surveillance of circulating influenza viruses.

    Performance characteristics for influenza were established during a season when seasonal influenza viruses A(H1N1) and A(H3N2) were the predominant influenza A viruses in circulation and during a season when the A(H1N1)pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.

    Testing with the influenza H5a and H5b primer and probe sets should not be performed unless the patient meets the most current U.S.Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspect A(H5) specimens. The definitive identification of influenza A(H5) (Asian lineage) either directly from patient specimens or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted unless a BSL 3E facility is available to receive and culture specimens.

    Device Description

    The CDC Human Influenza Real-Time RT-PCR Diagnostic Panel is used in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with the CDC device. The panel is configured in four separate kits. Each kit consists of oligonucleotide primers, fluorescently labeled hydrolysis probes, and controls which are used in rRT-PCR assays for the in vitro qualitative detection and characterization of influenza virus RNA in respiratory specimens from patients presenting with influenza-like illness (ILI) or from viral culture.

    Oligonucleotide primers and probes for detection of influenza A, influenza B, and influenza A of swine origin were selected from highly conserved regions of the matrix (M), non-structural (NS), and nucleoprotein (NP) genes, respectively. Oligonucleotide primers and probes for characterization and differentiation of influenza A(H3) and A(H1)pdm09 viruses, avian influenza A(H5) viruses, and genetic lineages of influenza B were selected from highly conserved regions of their hemagglutinin (HA) genes. Oligonucleotide primers and probes to detect the human RNase P gene (RP) in control samples and clinical specimens is also included in the panel.

    AI/ML Overview

    The provided document, K243274: CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, is an FDA 510(k) clearance letter. This document primarily describes the device, its intended use, and confirms its substantial equivalence to a previously cleared predicate device (K243931). Critically, it states that no analytical or clinical testing was performed for this specific modification, as the submission was to add a Predetermined Change Control Plan (PCCP).

    Therefore, the document does not contain the detailed study results, acceptance criteria, sample sizes, ground truth establishment, or expert involvement for the original validation of the device's performance. It explicitly states that "The performance characteristics of the CDC Flu rRT-PCR Dx Panel... were previously established and remain the same as the predicate device (K243931)."

    As such, I cannot provide the specific information requested in the prompt based on the provided text alone. The prompt asks for details of "the device" meeting acceptance criteria, and this document pertains to a modification that did not involve re-evaluating performance.

    However, I can interpret what would typically be sought for such an analysis in the context of an in vitro diagnostic (IVD) PCR panel like the one described. I will outline what the acceptance criteria and the study that proves the device meets them would likely entail for an IVD, but with the explicit understanding that the provided document does not contain these details.


    Based on the provided document, the device's performance characteristics were "previously established" for the predicate device (K243931) and are simply carried over here as "remaining the same." The current submission (K243274) is for adding a Predetermined Change Control Plan (PCCP) and states that "No analytical testing was performed for this modification" and "No clinical testing was performed for this modification."

    Therefore, the detailed information regarding acceptance criteria, reported performance, sample sizes, expert involvement, and ground truth establishment for the original validation of this device is NOT present in the provided text. The tables and descriptions below represent what would typically be expected for an FDA cleared RT-PCR diagnostic panel, assuming the original studies were conducted to industry standards, but are not extracted directly from the given document.


    Acceptance Criteria and Device Performance (Hypothetical for an RT-PCR IVD)

    For a real-time RT-PCR diagnostic panel like the CDC Human Influenza Virus Panel, acceptance criteria and performance would typically focus on analytical sensitivity (Limit of Detection - LoD), analytical specificity (cross-reactivity, inclusivity), and clinical performance (sensitivity, specificity).

    1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical)

    Performance MetricTarget Analyte(s)Acceptance Criteria (Hypothetical)Reported Device Performance (Hypothetical)
    Analytical Sensitivity (LoD)Influenza ADetect 95% of replicates at a specified viral RNA concentration (e.g., < 500 copies/mL)Achieved 100% detection for Influenza A (H1N1, H3N2, pdm09) at [X] copies/mL; 95% at [Y] copies/mL.
    Influenza BDetect 95% of replicates at a specified viral RNA concentration (e.g., < 500 copies/mL)Achieved 100% detection for Influenza B (Victoria, Yamagata) at [A] copies/mL; 95% at [B] copies/mL.
    H5 SubtypeDetect 95% of replicates for relevant H5 strains at a specified concentration (e.g., < 1000 copies/mL)Achieved 100% detection for specified H5 strains at [C] copies/mL; 95% at [D] copies/mL.
    Analytical Specificity (Inclusivity)All targetsDetect 100% of tested strains/isolates representing known genetic diversity within each target (e.g., various clades/lineages).100% detection across all 50 tested Influenza A strains (H1N1, H3N2, pdm09) and 30 Influenza B strains (Victoria, Yamagata); 5 specific H5 strains.
    Analytical Specificity (Exclusivity/Cross-Reactivity)All targetsNo cross-reactivity with common respiratory pathogens or human nucleic acid (e.g., < 5% false positives/inhibitory effects).No cross-reactivity observed with 30 non-influenza respiratory pathogens (e.g., RSV, HRV, Adenovirus) or human nucleic acid for any target.
    Clinical SensitivityAll targets> 90% Positive Percent Agreement (PPA) compared to a gold standard (e.g., viral culture, sequencing).Overall PPA: 95.8% (Influenza A), 94.2% (Influenza B), 97.1% (H5 presumptive). Ranges for specific kits within these bounds.
    Clinical SpecificityAll targets> 95% Negative Percent Agreement (NPA) compared to a gold standard.Overall NPA: 98.5% (Influenza A), 99.1% (Influenza B), 99.5% (H5 presumptive). Ranges for specific kits within these bounds.
    Reproducibility/PrecisionAll targetsCoefficient of Variation (CV) < 10% for Cq values at low and high positive concentrations across different sites, lots, operators.Inter-site CV < 5%, Inter-lot CV < 3%, Inter-operator CV < 4% for all targets at specified concentrations.

    Note: The actual quantitative values for LoD, PPA, and NPA would be specific to each kit and its validated performance.

    Study Details (Hypothetical for an RT-PCR IVD)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size (Hypothetical):
      • Analytical Validation: Typically involves a statistically significant number of replicates (e.g., 20-60 replicates per concentration level) for LoD, and a panel of 50-100 isolates/strains for inclusivity/exclusivity testing.
      • Clinical Validation (Prospective/Retrospective): For an IVD like this, hundreds to thousands of clinical specimens would be evaluated. For example:
        • Influenza A/B Typing Kit: ~500-1000 positive specimens (e.g., 200-400 A, 100-200 B, remaining co-infections/untouched negatives) and ~500-1000 negative specimens.
        • Influenza A Subtyping Kit: ~300-500 Influenza A positive specimens, including a balanced representation of H1 and H3.
        • Influenza B Lineage Genotyping Kit: ~100-200 Influenza B positive specimens, balanced for Victoria and Yamagata lineages.
        • Influenza A/H5 Subtyping Kit: Could involve a smaller number of true H5 positives (often spiked or well-characterized rare clinical samples) and a larger set of other influenza A types and negatives.
    • Data Provenance (Hypothetical):
      • Country of Origin: Likely multi-center studies within the United States given CDC's role and FDA clearance. Could also include international sites for broader strain diversity.
      • Retrospective or Prospective: Clinical validation typically involves both.
        • Retrospective: Utilizing banked, well-characterized clinical specimens to ensure adequate representation of rare or specific targets/lineages.
        • Prospective: Collecting new clinical specimens as they arrive, processed independently to mimic real-world use conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts (Hypothetical)

    • Number of Experts: For an RT-PCR assay, ground truth is usually established by highly reliable reference methods rather than human interpretation of images. However, if any interpretation or adjudication was needed (e.g., for discordant clinical results), it would involve a small panel.
      • For PCR-based diagnostics, ground truth is typically established by:
        • Confirmatory PCR methods (e.g., an FDA-cleared reference assay, sequencing of the target region, or a highly sensitive lab-developed test).
        • Viral culture (particularly for original isolation and characterization of strains).
        • Next-generation sequencing (NGS) for comprehensive genetic characterization.
    • Qualifications of Experts (if applicable for discordant results/clinical adjudication):
      • Senior Clinical Microbiologists, PhD-level or MD, with extensive experience (e.g., 10-20+ years) in diagnostic virology and molecular diagnostics in CLIA-certified laboratories.
      • Researchers with expertise in influenza genomics and epidemiology (e.g., from CDC's influenza division).

    4. Adjudication Method (Hypothetical)

    • For discordant results between the investigational device and the initial reference method in clinical studies, an adjudication method would be employed.
      • Method: Often involves a third, more definitive method. For an RT-PCR assay, this would frequently be bi-directional Sanger sequencing of the viral target region detected, or a second, orthogonal PCR method run by an expert laboratory.
      • Process: Results would be reviewed by a qualified laboratory director or team, and the definitive result (ground truth) would be assigned based on the adjudicating test.
      • "None" is typically not an acceptable approach for discordant results in regulated clinical trials.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance (Hypothetical)

    • No, an MRMC study would generally not be applicable nor performed for this type of device. This device is a diagnostic kit (RT-PCR) that generates quantitative cycle threshold (Cq) values and qualitative (positive/negative/subtype) results. It does not involve "human readers" interpreting images or complex data in the same way an AI-assisted diagnostic imaging device would.
    • The "human-in-the-loop" aspect here is the laboratory technician performing the test and interpreting the Cq values based on predefined cut-offs. The device itself is not "AI-assisted."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done (Hypothetical)

    • Yes, a "standalone" or "algorithm only" performance is inherently measured for this device. An RT-PCR assay's performance (sensitivity, specificity, LoD) is determined by its analytical component (primers, probes, reagents, thermal cycling) and the software's ability to interpret raw fluorescence data into Cq values and apply predefined cut-offs for positive/negative calls.
    • This is the primary way the analytical performance is established and is the foundation for clinical performance. The "algorithm" here is the defined Cq cut-offs and interpretive logic built into the system.

    7. The Type of Ground Truth Used (Hypothetical)

    • Expert Consensus / Reference Methods:
      • Confirmatory Molecular Methodologies: Often a consensus derived from an FDA-cleared reference RT-PCR assay, or an independently validated, highly sensitive laboratory-developed PCR test.
      • Nucleotide Sequencing: Gold standard for definitive identification and subtyping (e.g., Sanger sequencing or Next-Generation Sequencing of key viral genes مثل Hemagglutinin (HA) or Neuraminidase (NA) for influenza).
      • Viral Culture: Historically a gold standard for detecting viable virus, though less sensitive than PCR for direct clinical specimens. Often important for initial characterization of novel strains.
      • Clinical Outcomes Data: While not the primary ground truth for the presence of the pathogen, clinical outcomes (e.g., patient recovery, severity of illness, epidemiological links) can provide supportive context for the clinical utility of the test.
    • For an influenza diagnostic, the molecular sequencing data and confirmed viral culture (for positive samples) along with a robust negative control (absence confirmed by multiple methods) would be the primary components of ground truth.

    8. The Sample Size for the Training Set (Hypothetical)

    • Not applicable in the typical sense of machine learning/AI. This is a traditional RT-PCR assay kit. There isn't a "training set" for an AI model.
    • The "training" analogue would be the extensive research and development process to design primers and probes that are specific and sensitive to the target viruses, and to optimize the reaction conditions. This involves screening numerous viral strains and non-target organisms, but it's not a "training set" in the machine learning context.

    9. How the Ground Truth for the Training Set Was Established (Hypothetical)

    • Not applicable due to the nature of the device. As mentioned, it's a traditional RT-PCR kit, not an AI/ML device that requires a training set and associated ground truth for model development. The design and optimization of such a kit rely on fundamental molecular biology principles, known genetic sequences of influenza viruses, and iterative laboratory testing.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CDC Human Influenza Real-Time RT-PCR Diagnostic Panel: Influenza A/B Typing Kit (VER 2)

    The Influenza A/B Typing Kit contains reagents and controls of the CDC Human Influenza Virus RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and evidemiological information:

    · For qualitative detection of influenza virus type A or B viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture.

    · To provide epidemiological information for surveillance of circulating influenza viruses.

    CDC Human Influenza Real-Time RT-PCR Diagnostic Panel: Influenza A Subtyping Kit (VER 4)

    The Influenza A Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    · For determination of the subtype of seasonal human influenza A viruses as seasonal A(H3) and/or A(H1)pdm09 from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture:

    · To provide epidemiological information for surveillance of circulating influenza viruses.

    CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza B Lineage Genotyping Kit (VER 1.1 and 2)

    The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic realtime PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • For the determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture;

    · To provide epidemiologic information for surveillance of circulating influenza viruses.

    CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza A/H5 Subtyping Kit (VER 4)

    The Influenza A/H5 Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    · For the presumptive identification of virus in patients who may be infected with influenza A subtype A(H5) (Asian lineage) from viral RNA in human respiratory specimens, conjunctival swabs, and viral culture in conjunction with clinical and epidemiological risk factors;

    · To provide epidemiologic information for surveillance of circulating influenza viruses.

    Device Description

    The CDC Human Influenza Real-Time RT-PCR Diagnostic Panel consists of four real-time RT-PCR (rRT-PCR) assays used on IVD-labeled real-time PCR instruments that has been FDA-cleared for use with this device. The panel is configured in four separate kits: Influenza A/B Typing kit, Influenza A/H5 Subtyping kit, and Influenza B Genotyping kit. Each kit consists of oligonucleotide primers, fluorescently labeled hydrolysis probes, and controls which are used in rRT-PCR assays for the in vitro qualitative detection of influenza virus RNA in respiratory and conjunctival specimens from patients presenting with influenza-like illness (IL). Oligonucleotide primers and probes for detection of influenza A, influenza B, and 2009 influenza A (swine origin) were selected from highly conserved regions of the matrix (M), non-structural (NS), and nucleoprotein (NP) genes, respectively. Oligonucleotide primers and probes for characterization and differentiation of influenza A(H3) and A(H1)pdm09 viruses, genetic lineages of influenza A(H5) viruses were selected from highly conserved regions of their HA genes. Oligonucleotide primers and probes to detect the human RNase P gene (RP) in control samples and clinical specimens are also included in the panel.

    AI/ML Overview

    The provided text describes the acceptance criteria and study proving the device meets those criteria for the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel: Influenza A/H5 Subtyping Kit (VER 4), specifically concerning the addition of conjunctival swabs as a specimen type.

    Here’s a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance:

    The document does not explicitly present a table of acceptance criteria in terms of specific quantitative thresholds (e.g., minimum sensitivity/specificity percentages) for clinical performance. Instead, it describes an evaluation based on demonstrating successful detection capabilities and similarity to a predicate device. The performance is reported descriptively based on the clinical study results.

    Reported Device Performance (for Influenza A/H5 Subtyping Kit (VER 4) on conjunctival swabs):

    Performance MeasureValue/DescriptionContext/Details
    Limit of Detection (LoD)10^-5 EID50/mlConfirmed for conjunctival swab specimens.
    Interfering SubstancesNo interference observedTested with two common over-the-counter eyedrops (containing olopatadine hydrochloride and naphazoline hydrochloride respectively). Tested at 2x LoD in contrived specimens (conjunctival matrix and simulated respiratory matrix) and by adding directly to RT-PCR reaction.
    Clinical Performance (on Confirmed Cases with Conjunctivitis Symptoms)26/33 (78.8%) positive, 3/33 (9.1%) inconclusiveFor A/H5 on Conjunctival Swab (CS) specimens.
    Clinical Performance (on Confirmed Cases with Conjunctivities but URT Positive Only)4/33 (15.2%) positiveFor A/H5 on Upper Respiratory Tract (URT) specimens only.
    Clinical Performance (on Confirmed Cases without Conjunctivitis Symptoms)1/2 (50%) positive on CS, 1/2 (50%) positive on URT onlyOne case had CS sample, other had URT only.
    Clinical Performance (on Probable Cases with Conjunctivitis Symptoms)2/2 (100%) negative on CS
    Subject Matched Paired URT and CS Specimens (Confirmed Cases)10/32 (31.3%) positive on at least one URT and CS specimens
    15/32 (46.9%) positive in CS specimens only
    4/32 (12.5%) positive in at least one URT specimen only
    3/32 (9.4%) CS inconclusive, URT negative
    Subject Matched Paired URT and CS Specimens (Probable Cases)All URT and CS specimens tested negativeFrom 1 probable human case.

    2. Sample size used for the test set and the data provenance:

    • Clinical Test Set Sample Size:
      • 46 total cases (44 confirmed, 2 probable) initially considered.
      • 35 confirmed cases were assessed for performance due to exclusion of 9 cases where not all necessary assay reagents were used.
      • 2 probable cases were assessed.
      • The document also specifies the number of clinical specimens available for testing: e.g., 32 confirmed cases had subject-matched paired URT and CS specimens, 3 confirmed cases had only CS specimens, 1 probable case had subject-matched paired URT and CS specimens, and 1 probable case had only a CS specimen.
    • Data Provenance:
      • Country of Origin: US (from public health investigations in the US).
      • Retrospective or Prospective: The data was collected from specimens from March to November 2024 as part of public health investigations into outbreaks. This indicates a retrospective collection of existing specimens from human cases.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The ground truth for confirmed/probable cases of A/H5N1 was established based on "the criteria described in the current Council of State and Territory Epidemiologists (CSTE) document 24-ID-09." The document does not specify the number of individual experts or their qualifications directly. This refers to established public health guidelines rather than individual expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • The ground truth was established by adherence to existing CSTE criteria, not through a human reader adjudication process of the device's output. The device results were compared against these established "confirmed or probable case status."

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This document describes the validation of a laboratory diagnostic kit (RT-PCR) with specific specimen types and does not involve human readers interpreting images or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, this is effectively a standalone performance study. The device itself is an RT-PCR diagnostic panel. The "performance" described is that of the kit (reagents and controls run on a PCR instrument) in detecting viral RNA, without human interpretation of raw data beyond reading the instrument's output based on established cut-offs (e.g., Ct values). It's an "algorithm only" in the sense that the test results in a definitive positive, negative, or inconclusive outcome based on chemical reactions and signal detection.

    7. The type of ground truth used:

    • The ground truth for the clinical performance evaluation was based on "confirmed or probable case status for each case on the criteria described in the current Council of State and Territory Epidemiologists (CSTE) document 24-ID-09." This can be categorized as outcomes data/epidemiological criteria as determined by public health authorities.

    8. The sample size for the training set:

    • The document describes the validation of a diagnostic panel (PCR-based), not an AI algorithm. Therefore, there is no specific "training set" in the machine learning sense. The device's components (primers, probes) and their performance characteristics are established through analytical studies (e.g., LoD, interfering substances) and clinical validation with patient samples. The development of such a kit is based on extensive prior knowledge of influenza virus genetics and PCR technology, but not a "training set" in the context of supervised learning for AI.

    9. How the ground truth for the training set was established:

    • As there is no "training set" in the AI sense, this question is not applicable to the device described. The "ground truth" for the analytical development of the RT-PCR panel would be based on well-characterized viral samples (e.g., BPL inactivated virus with known concentrations) and established laboratory techniques for nucleic acid detection.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241110

    Validate with FDA (Live)

    Date Cleared
    2024-05-21

    (29 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Influenza A Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    · For determination of the subtype of seasonal human influenza A viruses as seasonal A(H3) and/or A(H1)pdm09 from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture;

    · To provide epidemiologic information for surveillance of circulating influenza viruses.

    Device Description

    The Influenza A Subtyping Kit contains components of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel that is used in rRT-PCR assays on an FDA-cleared in vitro diagnostic real-time PCR instrument. The primer and probe sets contained in the Influenza A Subtyping Kit are designed for the detection and characterization of influenza type A viruses that infect humans.

    The Influenza A Subtyping Kit consists of oligonucleotide primers and dual-labeled hydrolysis (TagMan®) probes and controls. which may be used in rRT-PCR assays for the in vitro qualitative detection and characterization of the human influenza virus RNA in respiratory specimens from patients presenting with influenza-like illness (IL). The oligonucleotide primers and probes for detection of Influenza A and 2009 Influenza A (swine origin) were selected from highly conserved regions of the matrix (M), and the nucleoprotein (NP), respectively. Oligonucleotide primers and probes for characterization and differentiation of seasonal influenza A(H3) and A(H1)pdm09 viruses were selected from highly conserved regions of their respective HA genes. Detection of viral RNA not only aids in the diagnosis of illness caused by seasonal. newly emerging, and novel influenza viruses in patients with ILI, but also provides epidemiological and surveillance information on influenza and aids in the presumptive laboratory identification of specific novel influenza A viruses.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied through the performance metrics evaluated and compared to the predicate device and NGS as a gold standard. The core idea is that the modified assay should perform equivalently or better than the previous version, especially for strains with the 3' mutation.

    Acceptance Criterion (Implied)Reported Device Performance (H3_v2 ZEN & BHQ)
    Analytical Sensitivity (Limit of Detection - LoD) equivalent to or better than predicate.A/Darwin/09/2021 (with 3' mutation): LoD was 10^(1.66) EID50/mL for both H3_v2 (ZEN) and H3_v2 (BHQ) with Invitrogen Platinum III SuperScript™. For Quanta qScript™, LoD was 10^(1.18) EID50/mL for H3_v2 (ZEN) and 10^(1.66) EID50/mL for H3_v2 (BHQ).A/HongKong/4801/2014 (without 3' mutation): LoD was 10^(2.12) EID50/mL for both H3_v2 (ZEN) and H3_v2 (BHQ) with Invitrogen Platinum III SuperScript™. For Quanta qScript™, LoD was 10^(2.12) EID50/mL for both H3_v2 (ZEN) and H3_v2 (BHQ).Overall confirmed LoD of the H3 v2 assay for both ZEN and BHQ quenchers was equivalent to the current H3 IVD assay (10^(-12) or 1.23x10^(2-04) EID50/mL).
    Inclusivity for diverse A(H3) strains equivalent to or better than predicate.All influenza A(H3) strains tested, representing temporal, geographic, and genetic diversity, were detected by the modified H3 v2 assay (both ZEN and BHQ) at low and high titers.Inclusivity of influenza A(H3) strains was not impacted.
    Analytical Specificity (Cross-Reactivity with other influenza subtypes) – No cross-reactivity with non-target influenza.No cross-reactivity was seen with the H3 v2 assay with either ZEN or BHQ quenchers when tested against various other influenza A subtypes (H1N1pdm09, H1N2v, H1N1v, H3N8, H5N8, H7N9, H9N2) and influenza B and C viruses.
    Analytical Specificity (Cross-Reactivity with non-influenza respiratory pathogens) – No cross-reactivity.None of the tested non-influenza human respiratory viruses, bacteria, or yeast were detected with either the H3 v2 ZEN or BHQ assays.
    Positive Percent Agreement (PPA) with NGS for A(H3) strains equivalent to or better than predicate, especially with 3' mutation.Invitrogen Superscript™ III: H3 v2 (ZEN) 100% (93.4-100), H3 v2 (BHQ) 100% (93.4-100) vs. H3 IVD 100% (93.4-100).Quanta qScript™: H3 v2 (ZEN) 96.67% (89-99), H3 v2 (BHQ) 96.67% (89-99) vs. H3 IVD 95% (86-98).For both PPA and NPA, the modified H3 v2 assay performed equivalent or better than the current H3 IVD assay.
    Negative Percent Agreement (NPA) for negative specimens equivalent to or better than predicate.Invitrogen Superscript™ III: H3 v2 (ZEN) 100% (93.4-100), H3 v2 (BHQ) 100% (93.4-100) vs. H3 IVD 100% (93.4-100).Quanta qScript™: H3 v2 (ZEN) 100% (93.4-100), H3 v2 (BHQ) 100% (93.4-100) vs. H3 IVD 100% (93.4-100).For both PPA and NPA, the modified H3 v2 assay performed equivalent or better than the current H3 IVD assay.
    No significant shift in Ct values compared to predicate in common positive specimens.No significant shift in Ct values was seen with the modified H3 v2 assay when comparing average Ct values for positive specimens generating a positive result with both sets of primers and probes.
    Improved sensitivity for strains with the 3’ mutation.Illustrative example: One positive specimen with a double mutation showed a shift in Ct value from an average of 31.1 for H3 IVD to 22.94 (H3 v2 ZEN) and 22.77 (H3 v2 BHQ), indicating improved detection.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Evaluation):
      • Positive Specimen Panel: 60 influenza A(H3) specimens (30 with 3' mutation, 30 without 3' mutation).
      • Negative Specimen Panel: 60 negative specimens (from symptomatic patients known to be positive for influenza H1N1).
      • Total Clinical Test Set: 120 specimens (60 positive, 60 negative).
    • Data Provenance: Retrospective study. Clinical specimens were collected from patients during previous influenza seasons in the United States.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The ground truth for the clinical test set was established using Next Generation Sequencing (NGS) and direct clinical specimen classification, not through expert consensus of visual or diagnostic interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth for the clinical test set was established via Next Generation Sequencing (NGS), which directly determines the genetic identity of the virus, rather than a consensus among human reviewers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an in vitro diagnostic real-time RT-PCR diagnostic panel, not an AI-assisted diagnostic tool that would involve human readers interpreting AI output.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the performance evaluation in this submission is of the standalone device (algorithm only). The RT-PCR assay itself provides the result, without human interpretation of the assay's primary output (e.g., a visual scan of a reaction). The results are considered definitive from the machine output.

    7. The Type of Ground Truth Used

    • Analytical Performance: The ground truth for analytical sensitivity (LoD) and inclusivity studies was based on known, quantified viral stocks (EID50/mL or ID50/mL) of specific influenza strains.
    • Clinical Performance: The ground truth for positive and negative clinical specimens was established using Next Generation Sequencing (NGS) as the comparator assay. This method directly confirms the identity and specific genetic characteristics (e.g., presence of 3' mutation) of the influenza virus in the samples.

    8. The Sample Size for the Training Set

    The document does not explicitly state a "training set" in the context of device development. This is a modification to an existing RT-PCR assay, and the "development" or "training" of such a device primarily involves bioinformatic analysis, primer/probe design adjustments, and analytical testing with known isolates. The "in-silico analysis" described (Process 1: assessment of primers against global H3N2 sequence information from GISAID EpiFlu database; Process 2: BLAST against NCBI nr/nt database) serves a similar function to a training set for algorithm-based devices by informing the optimal primer and probe sequences. No numerical sample size is provided for these bioinformatic databases.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, for an RT-PCR assay, the "training set" concept is different from an AI/ML context. The ground truth for informing the primer/probe design (which is effectively the "training" data for the assay's specificity and sensitivity for detecting target sequences) was established through:

    • GISAID EpiFlu database: This database contains comprehensive, publicly shared influenza sequence information, used to assess potential primer and probe sets against known H3N2 sequences and calculate nucleotide mismatches. The sequences in this database are derived from laboratories globally undergoing influenza surveillance.
    • NCBI BLAST+ against the nr/nt database: This is a vast database of non-redundant nucleotide sequences, against which primer sequences were compared to confirm inclusivity for H3 Influenza virus HA segments and exclusivity against non-target sequences.

    This information is based on established genetic sequences, which serve as the fundamental "ground truth" for designing molecular diagnostic assays.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222558

    Validate with FDA (Live)

    Date Cleared
    2022-08-30

    (6 days)

    Product Code
    Regulation Number
    866.3315
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Non-variola Orthopoxvirus Real-ime PCR Primer and Probe Set is intended for the in vitro qualitative presumptive detection of non-variola Orthopoxvirus DNA extracted from human pustular rash specimens and viral cell culture lysates submitted to a Centers for Disease Control and Prevention designated laboratory. The assay detects nonvariola Orthopoxvirus DNA, including vaccinia, cowpox, monkeypox and ectromelia viruses at varying concentrations. This assay does not differentiate vaccinia virus or monkeypox virus from other orthopoxviruses detected by this assay and does not detect variola virus. Refer to the CDC algorithm, Acute, Generalized Vesicular or Pustular Rash Illness Testing Protocol in the United States for recommended testing and evaluation algorithms for patients presenting with acute, generalized pustular or vesicular rash illness.

    Results of this assay are for the identification of non-variola Orthopoxvirus DNA. These results must be used in conjunction with other diagnostic assays and clinical observations to diagnose Orthopoxvirus infection. The assay should only be used to test specimens with low/moderate risk of smallpox exists, viral culture should not be attempted. Negative results obtained with this device do not prection virus infection and should not be used as the sole basis for treatment or other patient management decisions.

    Use is limited to Centers for Disease Control and Prevention designated laboratories.

    Device Description

    Unchanged from original submission (K221834).

    AI/ML Overview

    The provided text is a 510(k) Summary for the "Non-variola Orthopoxvirus Real-time PCR Primer and Probe Set". This document primarily focuses on demonstrating substantial equivalence to a predicate device and outlines the intended use and some general performance characteristic categories.

    However, the document does not include detailed information regarding:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes used for test sets or their data provenance.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness study results.
    • Standalone (algorithm only) performance.
    • Specific types of ground truth used (beyond implying laboratory testing).
    • Sample size for training sets.
    • How ground truth for training sets was established.

    Instead, for performance characteristics, the document states: "Inquiries regarding performance characteristics for the Non-variola Orthopoxvirus Real-time PCR Primer and Probe Set should be directed to the Centers for Disease Control and Prevention." It mentions that the Limit of Detection (LOD) was determined through an analytical sensitivity study and broadly references analytical sensitivity and specificity, as well as clinical performance, but without providing the detailed results or methodologies for these studies within this summary.

    The document highlights the device's intended use for in vitro qualitative presumptive detection of non-variola Orthopoxvirus DNA and the need to use results in conjunction with other diagnostic assays and clinical observations. It also specifies use limitations to "Centers for Disease Control and Prevention designated laboratories."

    Therefore, based solely on the provided text, I cannot provide the requested detailed information regarding acceptance criteria and study particulars. The document explicitly directs inquiries about these performance characteristics to the CDC.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221834

    Validate with FDA (Live)

    Date Cleared
    2022-06-24

    (1 days)

    Product Code
    Regulation Number
    866.3315
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Non-variola Orthopoxvirus Real-ime PCR Primer and Probe Set is intended for the in vitro qualitative presumptive detection of non-variola Orthopoxvirus DNA extracted from human pustular rash specimens and viral cell culture lysates submitted to a Laboratory Response Network (LRN) reference laboratory. The assay detects non-variola Orthopoxvirus DNA, including vaccinia, cowpox, monkeypox and ectromelia viruses at varying concentrations. This assay does not differentiate vaccinia virus or monkeypox virus from other orthopoxviruses detected by this assay and does not detect variola virus. Refer to the CDC algorithm, Acute, Generalized Vesicular or Pustular Rash IIIness Testing Protocol in the United States for recommended testing and evaluation algorithms for patients presenting with acute, generalized pustular or vesicular rash illness.

    Results of this assay are for the identification of non-variola Orthopoxvirus DNA. These results must be used in conjunction with other diagnostic assays and clinical observations to diagnose Orthopoxvirus infection. The assay should only be used to test specimens with low/moderate risk of smallpox exists, viral culture should not be attempted. Negative results obtained with this device do not prection virus infection and should not be used as the sole basis for treatment or other patient management decisions.

    Use is limited to Centers for Disease Control and Prevention designated laboratories.

    Device Description

    Unchanged from original submission (K221658).

    AI/ML Overview

    This document, a 510(k) summary for the "Non-variola Orthopoxvirus Real-time PCR Primer and Probe Set" (K221834) from the Centers for Disease Control and Prevention (CDC), does not provide details on acceptance criteria or the specific study proving the device meets those criteria directly within the provided text.

    Instead, it refers to a predicate device (K221658) and indicates that the current submission (K221834) is similar, with the main change being labeling. It explicitly states:

    • "Unchanged from original submission (K221658)" for Device Description, Principle of Operation, Sample Types, and Instrumentation/Software.
    • For "Analytical Limit of Detection (LoD)", "Analytical Sensitivity and Specificity", and "Clinical Performance," it repeatedly states: "Inquiries regarding performance characteristics for the Non-variola Orthopoxvirus Real-time PCR Primer and Probe Set should be directed to the Centers for Disease Control and Prevention."

    Therefore, based solely on the provided text, the specific information requested cannot be fully extracted. However, I can provide a template of what such an answer would look like if the information were available, and address what can be inferred or is explicitly absent.


    Unable to fully answer based on provided text. The document refers to a predicate device and directs inquiries for performance characteristics to the CDC, rather than detailing them.

    Here's an overview of the requested information based on what is and isn't present in the document:

    1. Table of Acceptance Criteria and Reported Device Performance
    This information is not provided in the document.

    2. Sample Size Used for the Test Set and Data Provenance
    This information is not provided in the document. The document refers to performance characteristics from a previous submission (K221658) and states inquiries should be directed to the CDC.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
    This information is not provided in the document. As this is an in vitro diagnostic (PCR test), ground truth would likely be established through a combination of culture, sequencing, or other highly sensitive and specific reference methods, rather than expert interpretation of images/clinical findings.

    4. Adjudication Method
    This information is not applicable/provided in the context of a PCR assay. Adjudication methods like 2+1 or 3+1 are typically used for subjective evaluations (e.g., image interpretation by radiologists or pathologists).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
    This information is not applicable/provided in the context of this device. An MRMC study is relevant for devices involving human interpretation (e.g., AI for radiology). This device is a molecular diagnostic assay.

    6. Standalone Performance Study
    Yes, this device is inherently a standalone algorithm/assay without human-in-the-loop performance measurement beyond the technician performing the PCR and interpreting the quantitative PCR cycles. The document implies such a performance study must have been conducted for the predicate device (K221658) when it mentions "Analytical Limit of Detection (LoD)," "Analytical Sensitivity and Specificity," and "Clinical Performance" studies, even though the results are not detailed here.

    7. Type of Ground Truth Used
    For a PCR assay, the ground truth for performance studies (LoD, analytical sensitivity, clinical performance) would typically be established by:

    • Reference Methods: Culture, sequencing, or a validated gold-standard molecular test for the detection of non-variola Orthopoxviruses.
    • Well-characterized samples: Use of known positive and negative controls, spiked samples, and clinical samples with confirmed infection status.
      This specific detail is not provided in the document, but these are the standard ground truth types for such a device.

    8. Sample Size for the Training Set
    This information is not provided in the document. For PCR primer and probe design, a "training set" might refer to the genomic sequences used to design the primers and probes. For assay validation, there isn't a "training set" in the machine learning sense; rather, there are analytical and clinical validation samples.

    9. How the Ground Truth for the Training Set was Established
    This information is not provided in the document, and the concept of "ground truth for a training set" is less directly applicable in the same way as it is for machine learning models. For a PCR assay, primer and probe sequences are designed based on conserved regions of target genomes. Validation samples would then have their true status established by highly reliable reference methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221658

    Validate with FDA (Live)

    Date Cleared
    2022-06-10

    (2 days)

    Product Code
    Regulation Number
    866.3315
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Non-variola Orthopoxvirus Real-time PCR Primer and Probe Set is intended for the in vitro qualitative presumptive detection of non-variola Orthopoxvirus DNA extracted from human pustular rash specimens and viral cell culture lysates submitted to a Laboratory Response Network (LRN) reference laboratory. The assay detects non-variola Orthopoxvirus DNA, including Vaccinia, Cowpox, Monkeypox and Ectromelia viruses at varying concentrations. This assay does not differentiate Vaccinia virus or Monkeypox virus from other Orthopoxviruses detected by this assay and does not detect Variola virus. Refer to the CDC algorithm, Acute, Generalized Vesicular or Pustular Rash Illness Testing Protocol in the United States for recommended testing and evaluation algorithms for patients presenting with acute, generalized pustular or vesicular rash illness.

    Results of this assay are for the identification of non-variola Orthopoxvirus DNA. These results must be used in conjunction with other diagnostic assays and clinical observations to diagnose Orthopoxvirus infection. The assay should only be used to test specimens with low/moderate risk of smallpox exists, viral culture should not be attempted. Negative results obtained with this device do not prection and should not be used as the sole basis for treatment or other patient management decisions.

    Use is limited to Laboratory Response Network (LRN) designated laboratories.

    Device Description

    The Non-variola Orthopoxvirus Real-time PCR Primer and Probe Set uses a fluorogenic probe, consisting of an oliqonucleotide with a reporter dye (FAM) attached to the 5′ end and a quencher dye (BHQ1) attached at or near the 3′ end. The probe anneals to a specific target sequence located between the forward and reverse primers. During the extension phase of the PCR cycle, the 5' nuclease activity of the Taq polymerase degrades the probe causing the reporter dye to separate from the quencher dye, thereby generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved from their respective probes and the fluorescence intensity is monitored during the PCR in real-time. The Taq polymerase used in this assay is inactive at room temperature and is activated by incubation at 95°C, thus minimizing the production of nonspecific amplification products.

    Each extracted DNA sample is tested using the Non-variola Orthopoxvirus Real-time PCR Primer and Probe set alonq with an internal control primer and probe set(s) to demonstrate adequate DNA extraction, proper function of common reagents and equipment, and the absence of inhibitory substances.

    AI/ML Overview

    This document is a 510(k) summary for the Non-variola Orthopoxvirus Real-time PCR Primer and Probe Set, submitted by the Centers for Disease Control and Prevention (CDC). It is a re-submission (K221658) of a previously cleared device (K181205). The document largely states that many aspects of the device are unchanged from the previous submission and directs further inquiries about performance characteristics to the CDC.

    Based on the provided text, many of the requested details are not explicitly stated in this FDA summary document. The document refers to an "Original Submission (K181205)" for comparison and states that for most performance characteristics, inquiries should be directed to the CDC. However, based on the information available:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a tabulated format or provide detailed "reported device performance" in this summary. It mentions an "Analytical Limit of Detection (LoD)" and "Analytical Sensitivity and Specificity" studies were performed. It also indicates "Clinical Performance" studies were conducted, but the results for these are not presented here. Given that this is a re-submission with an "Unchanged" intended use and principle of operation, the acceptance criteria and performance would likely be similar to or supported by the original submission (K181205).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not explicitly provided in the text. It states that inquiries regarding performance characteristics should be directed to the CDC.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not explicitly provided in the text.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not explicitly provided in the text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This device is a PCR primer and probe set for detecting viral DNA, not an AI-assisted diagnostic device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is a molecular diagnostic assay. Its performance is inherent to the chemical reactions and detection system, and it operates "stand-alone" in the sense that the test result is directly outputted by the instrumentation without human real-time modification of the analytical process. However, human interpretation of the results and clinical context is required as stated in the Indications for Use: "Results of this assay are for the identification of non-variola Orthopoxvirus DNA. These results must be used in conjunction with other diagnostic assays and clinical observations to diagnose Orthopoxvirus infection."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document refers to "identification of non-variola Orthopoxvirus DNA" as the output. The ground truth for such molecular tests typically involves:

    • Validated reference strains/samples: For analytical sensitivity and specificity, the ground truth would be precisely known viral DNA concentrations or the known presence/absence of target and non-target organisms.
    • Clinical diagnosis/outcome: For clinical performance, the ground truth would be established through a combination of other diagnostic assays, clinical observations, and potentially follow-up outcomes, as implied by the statement "These results must be used in conjunction with other diagnostic assays and clinical observations to diagnose Orthopoxvirus infection."

    Specific details are not provided in this summary.

    8. The sample size for the training set

    This information is not provided in the text. As a PCR diagnostic kit, there isn't typically a "training set" in the machine learning sense. Instead, development involves assay design, optimization, and then validation studies.

    9. How the ground truth for the training set was established

    Not applicable as there is no "training set" in the machine learning context for this type of device. Assay optimization and validation (which could be considered analogous to "training" and "testing" in a broad sense for a traditional diagnostic) would rely on well-characterized samples with established viral content.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CDC Human Influenza Real-Time RT-PCR Diagnostic Panel: Influenza A/B Typing Kit (VER 2)
    The Influenza A/B Typing Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • · For qualitative detection of influenza virus type A or B viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture.
    • · To provide epidemiological information for surveillance of circulating influenza viruses.

    CDC Human Influenza Real-Time RT-PCR Diagnostic Panel: Influenza A Subtyping Kit (VER 3)
    The Influenza A Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • · For determination of the subtype of seasonal human influenza A viruses as seasonal A(H3) and/or A(H1)pdm09 from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture;
    • · To provide epidemiological information for surveillance of circulating influenza viruses.

    CDC Human Influenza Real-Time RT-PCR Diagnostic Panel: Influenza A/H5 Subtyping Kit (VER 4)
    The Influenza A/H5 Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR instrument that has been FDA-cleared for use with this kit in conjunction with clinical and epidemiological information:

    • · For the presumptive identification of virus in patients who may be infected with influenza A subtype A(H5) (Asian lineage) from viral RNA in human respiratory specimens and viral culture in conjunction with clinical and epidemiological risk factors;
    • · To provide epidemiological information for surveillance of circulating influenza viruses.
    Device Description

    The CDC Human Influenza Real-Time RT-PCR Diagnostic Panel is used in real-time RT-PCR (rRT-PCR) assays on an in vitro diagnostic real-time PCR system. The panel is configured in four separate kits. Each kit consists of oligonucleotide primers, fluorescently labeled hydrolysis probes, and controls which are used in rRT-PCR assays for the in vitro qualitative detection and characterization of influenza virus RNA in respiratory specimens from patients presenting with influenza-like illness (ILI). Oligonucleotide primers and probes for detection of influenza A, influenza B, and 2009 influenza A (swine origin) were selected from highly conserved regions of the matrix (M), non-structural (NS), and nucleoprotein (NP) genes, respectively. Oligonucleotide primers and probes for characterization and differentiation of influenza A(H3) and A(H1)pdm(9 viruses and genetic lineages of influenza B were selected from highly conserved regions of their HA genes. Oligonucleotide primers and probes to detect the human RNase P gene (RP) in control samples and clinical specimens is also included in the panel.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily related to analytical sensitivity (Limit of Detection - LOD) and inclusivity for various influenza A strains, and analytical specificity (cross-reactivity and exclusivity). The clinical performance evaluation also serves as a form of acceptance criteria for positive and negative agreement with previously characterized samples.

    Acceptance Criteria CategorySpecific Criteria/MetricTarget Performance (Implied from Study Design)Reported Device Performance (Summary)
    Analytical Sensitivity (LOD Equivalency)100% positivity (3/3 replicates) at either the same endpoint LOD concentration or within a 5-fold dilution of the predicate device for benchmark strains.Demonstrated.Met for all tested benchmark and current strains across both enzyme systems (Invitrogen Superscript and Quanta qScript).
    Analytical Sensitivity (LOD Confirmation)≥95% of 20 individually extracted replicates testing positive at the confirmed LOD.Demonstrated.Met for all modified InfA, pdmInfA, and pdmH1 assays with various influenza strains and both enzyme systems.
    Analytical Sensitivity (Inclusivity)All tested influenza A strains (10 for pdmInfA/pdmH1, 24 for InfA) at low titer (near LOD) should result in 3/3 positive replicates.100% positive agreement for all inclusivity strains.Achieved 3/3 positive replicates for all 10 pdmInfA/pdmH1 inclusivity strains and all 24 InfA inclusivity strains.
    Analytical Specificity (Cross-Reactivity)No cross-reactivity with non-targeted influenza viruses at high titers, with the exception of specific known cross-reactivity where noted.Limited to no cross-reactivity.Modified InfA assay showed expected positive results for B/Victoria and B/Yamagata lineages. Modified pdmInfA assay showed cross-reactivity with one non-targeted influenza virus at very high titer (specifying A/Iowa/1/2006, A/Texas/14/2008, A/Ohio/09/2015 [A(H1N1)v], A/Minnesota/19/2011, A/Ohio/35/2017 [A(H1N2)v], A/Ohio/13/2017 [A(H3N2)v], A/gyrfalcon/Washington/41088-6/2014 [A(H5N8)]).
    Analytical Specificity (Exclusivity)No cross-reactivity with 35 common non-influenza respiratory pathogens (bacteria, yeast, other viruses) at high titers.No amplification for non-influenza pathogens.No cross-reactivity observed with any of the 35 tested non-influenza respiratory pathogens.
    Clinical Performance (Positive Agreement)High positive agreement with the predicate device on residual clinical specimens.100% agreement expected.Modified InfA Assay: 100% positive agreement across all specimen types and both enzyme systems (e.g., NPS, NS: 51/51). Modified pdmInfA and pdmH1 Assays: 100% positive agreement across all specimen types and both enzyme systems (e.g., NPS, NS: 28/28).
    Clinical Performance (Negative Agreement)High negative agreement with the predicate device on residual clinical specimens.100% agreement expected.Modified InfA Assay: 100% negative agreement (54/54 NPS) for both enzyme systems. Modified pdmInfA and pdmH1 Assays: 100% negative agreement (54/54 NPS) for both enzyme systems.

    Study Details:

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • LOD Equivalency & Confirmation:

      • Sample Size: Varies by virus strain and specific assay. For LOD equivalency, triplicate samples of serial dilutions were tested. For LOD confirmation, 20 individually extracted samples were tested for each target.
      • Data Provenance: Virus strains are identified by name and origin (e.g., A/Michigan/45/2015, A/Illinois/20/2018, A/Hong Kong/4801/2014, A/Abu Dhabi/240/2018, A/duck/Vietnam/NCVD-1544/2012, A/duck/Vietnam/NCVD-17A231/2016). Specific country of origin for all strains is not explicitly stated but implied from nomenclature.
      • Retrospective/Prospective: Experimental, controlled laboratory studies using characterized virus stocks.
    • Inclusivity:

      • Sample Size: 10 influenza A(H1N1)pdm09 viruses and 24 influenza A viruses of various types/subtypes. Each virus sample was tested in triplicate.
      • Data Provenance: Virus strains represented temporal, geographic, and genetic diversity (e.g., A/Florida/81/2018, A/Alaska/35/2018, A/Switzerland/8060/2017).
      • Retrospective/Prospective: Experimental, controlled laboratory studies using characterized virus stocks.
    • Analytical Specificity (Cross-Reactivity & Exclusivity):

      • Sample Size: Cross-reactivity: Various influenza viruses (e.g., A/Perth/16/2009, B/Maryland/15/2016) tested in triplicate. Exclusivity: 35 organisms (16 viruses, 18 bacteria, 1 yeast) tested.
      • Data Provenance: Organisms are identified by strain name.
      • Retrospective/Prospective: Experimental, controlled laboratory studies using characterized stocks of various organisms.
    • Clinical Performance Evaluation:

      • Sample Size:
        • Modified InfA Assay: 62 positive (35 A(H1N1)pdm09, 27 A(H3N2)) and 50 negative residual human respiratory clinical specimens.
        • Modified pdmInfA and pdmH1 Assays: 35 positive (for A(H1N1)pdm09) and 50 negative residual human respiratory clinical specimens.
      • Data Provenance: Residual human respiratory clinical specimens collected from patients during previous influenza seasons in the United States (2011-12 and 2013-14).
      • Retrospective/Prospective: Retrospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts used to establish ground truth. For the analytical studies (LOD, inclusivity, specificity), the ground truth is based on the known identity and titer of the cultured virus or bacterial/yeast strains. For the clinical performance evaluation, the ground truth for "positive" or "negative" status was established by prior testing (comparator) with the FDA-cleared CDC Human Influenza Real-Time RT-PCR Diagnostic Panel (K190302), which implicitly would have been validated using established laboratory methods or expert consensus in its own clearance process.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    No explicit adjudication method (like 2+1 or 3+1 consensus) is described. The analytical studies often cite "number of positive replicates out of three total replicates tested," implying a direct comparison to the expected outcome from the known sample. For clinical studies, the comparator is the already FDA-cleared predicate device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an in vitro diagnostic real-time RT-PCR diagnostic panel, not an AI-assisted diagnostic tool that humans interpret. There are no "human readers" in the context of interpreting the device's output, nor is there a multi-reader multi-case study described.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are standalone because the device itself is a diagnostic assay (a collection of reagents and controls) that produces a result (detection/characterization of viral RNA) through a real-time RT-PCR instrument. While laboratory personnel operate the instrument and interpret the output, the core performance studies evaluate the assay's ability to detect the target without human intervention influencing the assay's chemical and enzymatic reactions. The term "algorithm" is not directly applicable in the same way as with AI software, but the assay's 'logic' for detection is intrinsic to its design.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For Analytical Studies (LOD, Inclusivity, Specificity): The ground truth is based on known, characterized viral and microbial strains with established titers (e.g., TCID50/mL or EID50/mL) or concentrations (CFU/mL, ng/µL). This is a highly controlled laboratory ground truth.
    • For Clinical Performance Evaluation: The ground truth was established by prior testing with the FDA-cleared CDC Human Influenza Real-Time RT-PCR Diagnostic Panel (K190302), acting as the comparator or reference method for the collected residual human respiratory clinical specimens.

    8. The sample size for the training set

    The document does not explicitly describe a separate "training set" in the context of machine learning or AI. This device is a RT-PCR based assay, and its development would typically involve empirical optimization and validation against a variety of known strains and clinical samples rather than a formal training set for an AI model.

    9. How the ground truth for the training set was established

    As there is no explicit "training set" described for an AI model, this question is not directly applicable. For the development and optimization of the RT-PCR assays, ground truth for evaluating probe and primer design would have been established through a combination of sequencing data to identify conserved regions and target specificity, and testing against known, characterized viral isolates/strains to ensure desired reactivity.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192871

    Validate with FDA (Live)

    Date Cleared
    2019-11-07

    (30 days)

    Product Code
    Regulation Number
    866.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B. anthracis Real-Time PCR Assay is an in vitro diagnostic test for the qualitative detection of plasmid and chromosomal DNA sequences from B. anthracis. The assay can be used to test human respiratory samples, whole blood, serum, plasma, swabs from lesions, cerebrospinal fluid, and bacterial culture isolates from individuals suspected of having anthrax.

    Results generated from direct specimentesting are presumptive for the identification of B. anthracis. Results generated from culture isolate testing should be used in conventional methods for identification of Bacillus anthracis isolates as part of the LRN Bacillus anthracis Testing Algorithm. The diagnosis of anthrax infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidences, in addition to the identification of B. anthracis from cultures or detection directly in clinical specimens.

    Use is limited to Laboratory Response Network (LRN) designated laboratories.

    The B. anthracis Real-time PCR Assay is also intended for environmental specimen testing for biothreat detection and response. FDA has not evaluated claims related to the use of this assay on environmental specimens.

    Device Description

    The B. anthracis Real-time PCR Assay uses a fluorogenic probe, consisting of an oligonucleotide with a reporter dve (FAM) attached to the 5' end and a quencher dye (BHQ1) attached at or near the 3′ end. The probe anneals to a specific target sequence located between the forward and reverse primers. During the extension phase of the PCR cycle, the 5′ nuclease activity of Taq polymerase degrades the probe causing the reporter dye to separate from the quencher dye and a fluorescent signal is generated. With each cycle, additional reporter dye molecules are cleaved from their respective probes and the fluorescence intensity is monitored during the PCR. The Tag polymerase used in this assay is inactive at room temperature. It must be activated by incubation at 95°C, which also minimizes the production of nonspecific amplification products.

    Each extracted DNA sample is tested with three B. anthracis primer and probe sets run as individual reactions. The primer and probe sets target genes encoding virulence factors as well as conserved regions of DNA from the B. anthracis chromosome. All primer and probe sets must be positive for the overall result of the B. anthracis Real-time PCR Assay to be interpreted as positive. Any result that is positive for some, but not all target regions, is still considered equivocal and follow-up laboratory investigation should be performed per the LRN Bacillus anthracis testing algorithm.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the B. anthracis Real-time PCR Assay. It indicates substantial equivalence to a predicate device and provides information on its intended use, device description, and a comparison with the predicate. However, it explicitly directs inquiries regarding performance characteristics (including analytical limit of detection, analytical sensitivity and specificity, and clinical performance) to the Centers for Disease Control and Prevention. Therefore, the acceptance criteria and the study that proves the device meets those criteria are not detailed within this document.

    Here is a summary of the information that is available in the document, and what is explicitly stated as not available:


    1. Table of Acceptance Criteria and Reported Device Performance

    • Not available in the provided document. The document directs inquiries regarding performance characteristics to the Centers for Disease Control and Prevention.

    2. Sample size used for the test set and the data provenance

    • Not available in the provided document. The document states that inquiries regarding analytical LoD, analytical sensitivity and specificity, and clinical performance should be directed to the Centers for Disease Control and Prevention. This would include information about sample sizes and data provenance for validation studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable / Not available in the provided document. For a PCR assay, the "ground truth" is typically established by definitive laboratory methods (e.g., bacterial culture, sequencing, or a reference method) rather than expert opinion on images or clinical assessments. The details of how ground truth was established for the analytical and clinical performance studies are not present.

    4. Adjudication method for the test set

    • Not applicable / Not available in the provided document. See point 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a diagnostic PCR assay, not an AI-assisted imaging or diagnostic tool that involves human readers in the traditional sense of an MRMC study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, implicitly. The device is a Real-time PCR Assay, which is an in vitro diagnostic test. Its performance inherently refers to its standalone analytical and clinical performance in detecting target DNA sequences. While human operators are involved in running the assay and interpreting results, the "algorithm" (the PCR chemistry and detection) itself functions stand-alone. However, specific details of "standalone performance" metrics (like sensitivity, specificity, accuracy against a reference standard) are not provided and are referred to the CDC.

    7. The type of ground truth used

    • For a PCR assay detecting DNA sequences from B. anthracis, the ground truth for performance studies would typically be established by a "gold standard" laboratory method, such as:
      • Bacterial culture and identification: Confirmation of B. anthracis growth and subsequent identification using conventional microbiological methods (e.g., phenotypic differences, susceptibility testing).
      • Reference molecular methods: Other validated PCR assays or DNA sequencing for definitive identification.
    • Specific details of how ground truth was established for the validation studies are not available in the provided document.

    8. The sample size for the training set

    • Not applicable. For a PCR assay, there isn't a "training set" in the machine learning sense. The assay is developed based on known genetic targets of B. anthracis. Development involves optimizing primer/probe design and assay conditions, but not "training" an algorithm on a dataset.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 4