510(k) Data Aggregation

    K Number
    K212758

    Validate with FDA (Live)

    Device Name
    Autoplaque
    Date Cleared
    2023-05-19

    (626 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Autoplaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.

    Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.

    The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present.

    Autoplaque must be installed on a suitable commercial computer platform. It is the user's responsibility to ensure the monitor quality and ambient light conditions are consistent with the clinical applications.

    Typical users of Autoplaque are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.

    Device Description

    Autoplaque 3.0. a stand-alone software, performs post-processing of coronary Computed Tomography Angiography (CTA) images and measurements of components of images using computerized algorithms.

    Autoplaque 3.0 aids the physician with measurement of coronary artery stenosis and provides measurements for coronary plaque and coronary artery remodeling. Autoplaque 3.0 does not replace standard clinical practice or clinician decision making.

    Autoplaque 3.0 allows for standardized characterization of plaque and stenosis from DICOM image data (loaded from the local computer hard drive) and includes the following features:

    • Review of heart and coronary vessels in Multiplanar Reformatting (MPR), curved MPR, and straightened MPR views:
    • Measurement of vessel diameter and area;
    • Characterization and measurement of plaque parameters; and
    • Measurement of lumen diameter, area, and luminal stenosis.

    Autoplaque 3.0 includes automated vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.

    Autoplaque 3.0 can run on Windows or Mac computer platforms. The minimum hardware specifications are specified in the user manual.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the studies conducted for Autoplaque 3.0, based on the provided text:

    Acceptance Criteria and Device Performance for Autoplaque 3.0

    The acceptance criteria for Autoplaque 3.0 focused on establishing concordance and agreement for various plaque and stenosis measurements compared to expert reader measurements and the predicate device. The primary statistical measures used were the intraclass correlation coefficient (ICC) and correlation coefficient. While specific numerical thresholds for "excellent agreement" or "excellent correlation" are not explicitly stated as acceptance criteria, the studies report achieving this level of agreement.

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Implied)Reported Device Performance
    For Plaque Volume Measurements (Total Plaque, Calcified Plaque, Non-Calcified Plaque) and Diameter Stenosis:
    Agreement with Expert Reader (US Population)Excellent agreement (ICC and correlation coefficient)Excellent agreement shown (ICC and correlation coefficient)
    Agreement with Expert Reader (Non-US Population)Excellent agreement (ICC and correlation coefficient)Excellent agreement shown (ICC and correlation coefficient)
    Agreement with Predicate DeviceExcellent correlation and intraclass correlation agreementExcellent correlation and intraclass correlation agreement
    Analysis Time Per Lesion (Compared to Experts)Significantly faster than experts (for plaque analysis)5.7 seconds (Autoplaque 3.0) vs. 25-30 minutes (experts)
    Analysis Time Per Lesion (Compared to Predicate)Significantly faster than predicate device (for plaque analysis)<2 seconds (Autoplaque 3.0) vs. 30 seconds (predicate)

    2. Sample Sizes and Data Provenance

    Test Set - Comparison with Expert Reader (US Population):

    • Sample Size: 201 patients, with a total of 781 lesions analyzed.
    • Data Provenance: Retrospective, clinically indicated coronary computed tomography angiography (CCTA) images from multiple U.S. sites. Ethnicity information was available for 80 patients: White alone (not Hispanic or Latino) 18%, Black 19%, Hispanic or Latino 61%, and American Indian and Alaska native alone 1%.

    Test Set - Comparison with Expert Reader (Non-US Population):

    • Sample Size: 175 patients, with a total of 1081 lesions analyzed.
    • Data Provenance: Retrospective, clinically indicated coronary computed tomography angiography (CCTA) images from multiple non-U.S. external sites.

    Test Set - Comparative Study with Predicate Device:

    • Sample Size: 27 patients, with a total of 30 lesions analyzed.
    • Data Provenance: Not explicitly stated as retrospective or prospective, but derived from clinically indicated CCTA images obtained from a commercial Siemens CT scanner.

    Training Set:

    • Sample Size: Not specified in the provided text.
    • Data Provenance: Not specified in the provided text.

    3. Number and Qualifications of Experts for Ground Truth

    For US Patient Population Study Validation:

    • Number of Experts: One
    • Qualifications: Board-certified imaging cardiologist.

    For Non-US Patient Population Study Validation:

    • Number of Experts: Two cardiologists and one additional highly experienced radiologist.
    • Qualifications: Two cardiologists, one highly experienced radiologist.

    4. Adjudication Method for the Test Set

    For US Patient Population Study Validation:

    • Adjudication Method: Not explicitly stated, as only one expert was mentioned for ground-truthing. This suggests a direct comparison method.

    For Non-US Patient Population Study Validation:

    • Adjudication Method: Discrepancies between the two cardiologists were resolved by an additional highly experienced radiologist. This implies a 2+1 adjudication model.

    For Comparative Study with Predicate Device:

    • Adjudication Method: Not applicable, as this study compared the two devices rather than devices to human ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was specifically described where human readers improved with AI vs. without AI assistance. The clinical validation studies focused on comparison of the device's output to expert measurements, rather than assessing an AI-assisted workflow for human readers. However, the significantly reduced analysis time for the device compared to experts (5.7 seconds vs. 25-30 minutes) implies a substantial efficiency improvement when using the AI tool.

    6. Standalone Performance Study

    Yes, a standalone performance study was conducted. Both the "Comparison with Expert Reader Measurements in a Multicenter US Patient Population" and "Comparison with Expert Reader Measurements in a Non-US Patient Population" evaluated the Autoplaque 3.0 device's performance independently against expert ground truth. The device generates its output (plaque volumes, diameter stenosis) without human intervention in the measurement process, although the output can be reviewed and edited by clinicians.

    7. Type of Ground Truth Used

    For comparison with expert readers (both US and Non-US populations):

    • Ground Truth Type: Expert Consensus/Measurements. The ground truth was established by human expert readers (board-certified imaging cardiologist, cardiologists, and highly experienced radiologist) directly performing the measurements on the CCTA images.

    8. Sample Size for the Training Set

    The sample size for the training set is not explicitly mentioned in the provided text. The document focuses on the validation studies performed after the device was developed.

    9. How Ground Truth for Training Set was Established

    The method for establishing ground truth for the training set is not explicitly mentioned in the provided text. However, given that the device uses "Deep learning-based vessel, plaque and lumen segmentation," it is highly probable that the training data's ground truth would have been established through expert manual annotations or segmentations on CCTA images, similar to how the validation ground truth was established. The text only mentions that the deep learning-based segmentation was "validated with ground truth by clinician expert readers."

    Ask a Question

    Ask a specific question about this device

    K Number
    K172609

    Validate with FDA (Live)

    Date Cleared
    2017-11-03

    (64 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cedars-Sinai Cardiac Suite of applications is intended to enable an automated display, review, and quantification of Nuclear Medicine Cardiology medical images and datasets. Cedars-Sinai Cardiac Suite may be used in multiple settings including the hospital, clinic, doctors office, or remotely. The results provided should be reviewed by qualified healthcare professionals (e.g., radiologists, cardiologists, or general nuclear medicine physicians) trained in the use of medical imaging devices.

    Device Description

    The Cedars-Sinai Cardiac Suite is a stand-alone software solution for Cardiac SPECT and PET imaging processing and review. The Indication/Intended Use and the fundamental scientific technology of the modified device has not been changed and the same as those described in the unmodified predicate device. CSMC Cardiac Suite will be marketed as a comprehensive application suite that includes QPS+QGS (Quantitative Perfusion SPECT + Ouantitative Gated SPECT) and CSImport applications. This allows automatic processing and review of quantitative and qualitative information generated by nuclear medicine studies. Purchasable Options consist of Quantitative Blood Pool SPECT (QBS), ARG/QARG (for reporting purposes), Fusion (SPECT/CT/CTA and/or PET/CT/CTA), Motion Correction (MOCO), Automated Reconstruction (AutoRECON), CSView (Cedars-Sinai Viewer) and QPET. QPET also includes viability quantification and two additional databases (rubidium and ammonia) for processing PET studies.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria or a dedicated study proving the device meets them. Instead, it describes a 510(k) premarket notification for the "CSMC Cardiac Suite 2017" software, which asserts substantial equivalence to a previously cleared device (CSMC Cardiac Suite v2015- Nuclear Medicine Software, K141652).

    While "software validation, which confirms all features of the device were compliant with the software requirements" is mentioned, it is a general statement and does not provide specifics that would allow for a comprehensive answer to your questions regarding acceptance criteria, study design, or performance metrics.

    Therefore, many of your requested items cannot be extracted from the given text.

    Here's an attempt to answer what can be inferred or directly stated from the provided document, with many sections marked as "Not provided in the text."


    Acceptance Criteria and Device Performance

    The document does not explicitly state specific quantitative acceptance criteria (e.g., minimum sensitivity, specificity, or agreement metrics) or report detailed device performance against such criteria. The "Performance Data" section merely states: "Support of the substantial equivalence of the CSMC Cardiac Suite was provided as a result of software validation, which confirms all features of the device were compliant with the software requirements."

    Acceptance Criteria (Not Explicitly Stated)Reported Device Performance (Not Explicitly Stated)
    Not provided in the text.The device underwent "software validation, which confirms all features of the device were compliant with the software requirements." No specific performance metrics (e.g., accuracy, precision, recall) are reported.

    Study Details

    1. Sample size used for the test set and the data provenance: Not provided in the text.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided in the text.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not provided in the text.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not provided in the text. The document refers to the device as a "stand-alone software solution" and states "The results provided should be reviewed by qualified healthcare professionals." It does not describe a study involving human readers or their improvement with AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document states, "The Cedars-Sinai Cardiac Suite is a stand-alone software solution for Cardiac SPECT and PET imaging processing and review." However, it does not explicitly describe a standalone performance study with specific metrics that would compare its outputs to a ground truth, independent of human review. The "software validation" mentioned is very general.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not provided in the text.
    7. The sample size for the training set: Not provided in the text.
    8. How the ground truth for the training set was established: Not provided in the text.

    Summary of what is known:

    • The submission is for the "CSMC Cardiac Suite 2017," a software for automated display, review, and quantification of Nuclear Medicine Cardiology medical images and datasets.
    • It is a "stand-alone software solution."
    • The basis for substantial equivalence is primarily on its similarity in intended use, principle of operation, and technological characteristics to a previously cleared device (K141652).
    • "Software validation" was performed to confirm compliance with software requirements. This is a general statement about software quality assurance, not a specific clinical performance study.
    • The results from the device "should be reviewed by qualified healthcare professionals (e.g., radiologists, cardiologists, or general nuclear medicine physicians)." This indicates it's an aid, not a definitive diagnosis.

    The document focuses on establishing substantial equivalence to a predicate device rather than detailing a specific clinical performance study with defined acceptance criteria and results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141652

    Validate with FDA (Live)

    Date Cleared
    2014-07-22

    (32 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cedars-Sinai Cardiac Suite of applications is intended to enable an automated display, review, and quantification of Nuclear Medicine Cardiology medical images and datasets. Cedars-Sinai Cardiac Suite may be used in multiple settings including the hospital, clinic, doctors office, or remotely. The results provided should be reviewed by qualified healthcare professionals (e.g., radiologists, cardiologists, or general nuclear medicine physicians) trained in the use of medical imaging devices.

    Device Description

    The Cedars-Sinai Cardiac Suite is a stand-alone software solution for Cardiac SPECT and PET imaging processing and review. Cedars-Sinai Cardiac Suite (non-viewer) minimum system requirements include a computer with at least IGB RAM, 50MB hard disk space for software installation. a display resolution of at least 1024x768 with 16-bit color, a network adapter, a mouse (or other pointer device; trackpad, trackball, etc ... ) and one of the following operating systems: Windows XP, Windows Vista, Windows 7, Windows 8, Windows Server 2003. Windows Server 2008. Windows Server 2012. Mac OS X (10.6. 10.7 & 10.8). The viewer component of the Cedars-Sinai Cardiac Suite minimum system requirements is iPad 2, iPad (3rd generation), iPad (4th generation) and iPad Mini. The Cedars-Sinai Cardiac Suite operates on camera independent reconstructed SPECT and/or PET image files. CSMC Cardiac Suite will be marketed as a comprehensive application suite that includes QPS+QGS (Quantitative Perfusion SPECT + Quantitative Gated SPECT) and CSImport applications. This allows automatic processing and review of quantitative and qualitative information generated by nuclear medicine studies. Purchasable Options consist of Quanitative Blood Pool SPECT (QBS), QARG (for reporting purposes), Fusion (SPECT/CT/CTA and/or PET/CT/CTA), Motion Correction (MOCO), Automated Reconstruction (AutoRECON), Remote Viewer (Viewer) and OPET. OPET also includes viability quantification and two additional databases (rubidium and ammonia) for processing PET studies.

    AI/ML Overview

    The provided 510(k) summary does not contain the detailed information required to answer all parts of your request regarding acceptance criteria and a specific study proving the device meets them.

    Here's what can be extracted and what is missing:


    Acceptance Criteria and Reported Device Performance

    The provided document describes the device as substantially equivalent to a predicate device, and mentions "performance" in a general sense, stating:

    "Support of the substantial equivalence of the CSMC Cardiac Suite was provided as a result of software validation, which confirms all features of the device were compliant with the software requirements."

    This statement confirms software validation was performed to ensure compliance with software requirements, but it does not specify clinical performance acceptance criteria (e.g., sensitivity, specificity, accuracy targets) or provide a table of reported device performance against such criteria. The document focuses on demonstrating substantial equivalence based on intended use, technological characteristics, and design.

    Therefore, a table of acceptance criteria and reported device performance cannot be generated from the provided text.


    Missing Information from the Study Description:

    The 510(k) summary primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical performance study with the requested metrics. As such, the following information cannot be found in the provided text:

    • Sample size used for the test set and the data provenance: Not mentioned.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    • Adjudication method for the test set: Not mentioned.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned. The document primarily discusses the software as a computational/display tool for existing imagery, not a system that assists human interpretation in a comparative study.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document states it is "stand-alone software solution for Cardiac SPECT and PET imaging processing and review" and "allows automatic processing and review of quantitative and qualitative information." While this indicates standalone processing, it doesn't describe an 'algorithm only' standalone performance study against a clinical endpoint or ground truth. The "results provided should be reviewed by qualified healthcare professionals," implying human-in-the-loop is part of its intended clinical use.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not mentioned.
    • The sample size for the training set: Not mentioned. Since it's a software for processing and quantifying existing images derived from SPECT/PET, the concept of a "training set" for an AI model (as often understood in current AI contexts) is not explicitly discussed. The validation described is "software validation, which confirms all features of the device were compliant with the software requirements."
    • How the ground truth for the training set was established: Not mentioned.

    In conclusion, the provided 510(k) summary focuses on demonstrating that the Cedars-Sinai Cardiac Suite Nuclear Medicine Software is substantially equivalent to a previously cleared device (CSMC Cardiac Suite- Nuclear Medicine Software K112474) based on intended use, technological characteristics, and design, along with general software validation. It does not include the detailed clinical performance study design, acceptance criteria, or results that your request asks for.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123433

    Validate with FDA (Live)

    Date Cleared
    2013-02-26

    (111 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K112474

    Validate with FDA (Live)

    Date Cleared
    2011-09-23

    (28 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K100714

    Validate with FDA (Live)

    Date Cleared
    2011-04-01

    (385 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    0 - 60
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CSMC Cardiac Suite is intended to enable an automated display, review, and quantification of Nuclear Medicine Cardiology medical images and datasets. CSMC Cardiac Suite may be used in multiple settings including the hospital, clinic, doctors office, or remotely. The results provided should be reviewed by qualified healthcare professionals (e.g., radiologists, cardiologists, or general nuclear medicine physicians) trained in the use of medical imaging devices.

    Device Description

    The Cedars-Sinai Cardiac Suite is a stand-alone software solution for Cardiac SPECT and PET imaging processing and review. Cedars-Sinai Cardiac Suite minimum system requirements include a computer with at least 1GB RAM, 50MB hard disk space for software installation, a display resolution at least 1024x768 with 16-bit color, a network adapter, a mouse (or other pointer device; trackpad, trackball, etc.) and one of the following operating systems: Windows XP Professional. Windows Vista Professional. Windows 7 Professional. The Cedars-Sinai Cardiac Suite operates on camera independent reconstructed SPECT and/or PET image files. CSMC Cardiac Suite will be marketed as a comprehensive application suite that includes QGS (Quantitative Gated SPECT), QPS (Quantitative Perfusion SPECT) and CSImport applications. This allows automatic processing and review of quantitative and qualitative information generated by nuclear medicine studies. Purchasable Options consist of Quantitative Blood Pool SPECT (QBS), QARG (for reporting purposes), Fusion (SPECT/CT/CTA and/or PET/CT/CTA), Motion Correction (MOCO) and QPET. QPET also includes viability quantification and two additional databases (rubidium and ammonia) for processing PET studies.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Cedars-Sinai Cardiac Suite (K100714)

    The Cedars-Sinai Cardiac Suite is a stand-alone software solution for Cardiac SPECT and PET imaging processing and review, including applications such as QGS, QPS, QBS, QARG, CSImport, MoCo, and QPET. The device's substantial equivalence to predicate devices (AutoQUANT® Plus and Cedars-Sinai BPGS and MoCo Software Programs) was demonstrated through performance testing.

    1. Acceptance Criteria and Reported Device Performance

    The core of the performance testing revolved around demonstrating that the algorithms in the CSMC Cardiac Suite operate identically to those in the predicate devices.

    Acceptance CriteriaReported Device Performance
    Functional Equivalence: Each software application functions as per its specifications.QPS Verification: 1211 tests passed, 0 failed. QGS Verification: 1211 tests passed, 0 failed. QBS Verification: 632 tests passed, 0 failed. QARG Verification: 1161 tests passed, 0 failed. CSImport Verification: 111 tests passed, 2 failed (These two failures are likely acceptable as the overall submission was cleared). MOCO Verification: 23 tests passed, 0 failed. Installation Verification: 82 tests passed, 0 failed. Regression Tests: 912 tests passed, 0 failed (across multiple daily runs).
    Algorithmic Equivalence: All algorithms (with the exception of two new ones: "Quality Score" and "RV algorithm") produce the same results as previous versions of the software (predicate devices).A set of 16 studies had their algorithm results compared to those of the predicate device. These comparisons were included in regression tests, which consistently passed (912 tests passed, 0 failed). This implies identical results for the algorithms common to both the device and its predicates.
    New Algorithms Functionality: The two new algorithms ("Quality Score" and "RV algorithm") are integrated and perform as expected as part of the updated QPS / QGS.The document states that the updated version of QPS and QGS includes an improved contouring algorithm based on a quality score and RV (right ventricle) extraction and analysis. The comprehensive nature of the functional and regression tests (which passed) suggests these new features are functional and integrated correctly.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 16 studies were used for direct comparison of algorithm results between the device and its predicate. Beyond this, thousands of tests were conducted for functional and regression testing, but the specific number of unique patient cases for these tests is not explicitly stated.
    • Data Provenance: The document does not specify the country of origin of the data. It also does not explicitly state whether the data was retrospective or prospective, but given the nature of comparing to predicate device results, it's highly likely to be retrospective data that was previously processed by the predicate devices.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not indicate the involvement of experts for establishing ground truth for the test set in the context of clinical interpretation or diagnosis. The focus was on the performance of the algorithms themselves compared to predicate device results, not against a human-derived clinical ground truth for a diagnostic outcome.

    4. Adjudication Method

    Not applicable. The study focused on algorithmic equivalence to predicate devices, not on a diagnostic outcome requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or performed. The study aimed to demonstrate that the device's algorithms operate identically to previously cleared predicate devices.

    6. Standalone Performance Study

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was effectively done. The performance testing section directly addresses the functionality and output of the software's algorithms, comparing them to predicate device outputs. The statement, "With the exception of two new algorithms ("Quality Score" and "RV algorithm") all algorithms operate identically (i.e. produce the same results) as with previous versions of the software (referenced in predicate devices)", indicates a standalone evaluation of the algorithmic output.

    7. Type of Ground Truth Used

    The ground truth used for the algorithmic equivalence testing was the results generated by the predicate devices (AutoQUANT® Plus and Cedars-Sinai BPGS/MoCo Software Programs). For the functional verification tests (QPS, QGS, etc.), the ground truth was the expected output or behavior defined by the software's specifications.

    8. Sample Size for the Training Set

    The document does not provide any information regarding a "training set" or its sample size. This suggests that the device was not developed using traditional machine learning supervised training paradigms that require distinct training sets, or at least, such details were not considered relevant for this 510(k) submission focused on substantial equivalence to existing devices. The new algorithms ("Quality Score" and "RV algorithm") are mentioned, but their development methodology (including if they involved new training) is not detailed.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set information is provided or implied in the document.