Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163570

    Validate with FDA (Live)

    Date Cleared
    2017-08-15

    (239 days)

    Product Code
    Regulation Number
    862.3250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Test System (COCM) is for the qualitative determination of benzoylecgonine (cocaine metabolite) in human urine at a cutoff value of 300 ng/mL. The assay is designed for professional use with a clinical chemistry analyzer. For in vitro diagnostic use only.

    This assay provides a rapid screening procedure for determining the presence of benzoylecgonine in urine. The assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. Clinical considerations and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

    Device Description

    Carolina Liquid Chemistries Cocaine Metabolite Test System (COCM) is a ready-to-use, liquid reagent, homogenous enzyme immunoassay. The assay uses a specific antibody that can detect benzoylecgonine (cocaine metabolite) in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs. The assay is based on competition between benzoylecgonine and glucose-6-phosphate dehydrogenase (G6PDH) enzyme, and free drug from the urine sample for a fixed amount of the specific antibody. In the absence of free drug from the urine sample the specific antibody binds to the drug labeled with G6PDH enzyme causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically and 340nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

    AI/ML Overview

    This document describes the validation of the Carolina Liquid Chemistries Cocaine and Cocaine Metabolite Test System (COCM), an enzyme immunoassay for the qualitative detection of benzoylecgonine (cocaine metabolite) in human urine at a cutoff value of 300 ng/mL. The study aims to demonstrate substantial equivalence to a legally marketed predicate device (Lin-Zhi International, Inc. Cocaine Metabolite Enzyme Immunoassay).

    Based on the provided information, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document implicitly uses the performance of the predicate device and LCMS/GCMS as the "acceptance criteria" for demonstrating substantial equivalence. The key performance characteristics evaluated for the candidate device are Accuracy, Interfering Substances/Cross-Reactivity, and Precision.

    Performance CharacteristicAcceptance Criteria (Implicit, based on Predicate/Reference Method)Reported Device Performance (Carolina Liquid Chemistries COCM)
    AccuracyHigh concordance with LCMS for positive and negative samples at various concentrations relative to the 300 ng/mL cutoff, similar to predicate device's performance against GCMS.Candidate device compared to LCMS (n=100 specimens) - Drug Free (0.00 ng/mL): 30 Negative (100% negative) - < 50% of cutoff (75-149 ng/mL): 11 Negative (100% negative) - > 50 to < 100% of cutoff (150-299 ng/mL): 9 Negative (100% negative) - > 100 to < 150% of cutoff (301-449 ng/mL): 24 Positive (100% positive) - > 150% of cutoff (>= 450 ng/mL): 26 Positive (100% positive) - Overall: All 100 samples matched LCMS results (100% agreement).
    Interfering Substances/Cross-Reactivity/SpecificityDemonstrate a specific antibody that can detect benzoylecgonine with minimal cross-reactivity to various common prescription and abused drugs. pH (3-11) and specific gravity (1.000-1.030) should not interfere.- Specific Gravity & pH: "Study results show that neither pH (range tested: 3-11), nor specific gravity conditions (range tested 1.000-1.030) will interfere with this assay."- Specificity (Cross-Reactivity): - Structurally Related Cocaine Compounds showed cross-reactivity (e.g., Cocaine (30 µg/mL), Norcocaine (60 µg/mL), Ecgonine, Methyl Ester (350 µg/mL) were "Positive"). Benzoylecgonine (0.3 µg/mL) also positive, as expected. - Numerous other common drugs/substances (e.g., Acetaminophen, Amphetamine, Codeine, Morphine, Nicotine, etc.) tested up to 1000-2000 µg/mL showed "Negative" cross-reactivity.
    PrecisionConsistent and reproducible results across runs and within runs, especially around the cutoff concentration.- Within Run (Qualitative Analysis): 20 replicates of spiked human urine at 9 concentrations. At the 300 ng/mL cutoff, 15 Negative / 5 Positive results were obtained (implies some borderline results). All other concentrations (0-225 ng/mL and 375-600 ng/mL) showed 100% agreement with expected negative/positive results.- Run-to-Run (Qualitative Analysis): 90 replicates of spiked human urine at 9 concentrations. At the 300 ng/mL cutoff, 84 Positive / 6 Negative results were obtained (implies some borderline results). All other concentrations (0-225 ng/mL and 375-600 ng/mL) showed 100% agreement with expected negative/positive results. - Within/Run-to-Run (ΔmA/min): Provided mean, SD, %CV for different concentrations, indicating good precision (e.g., %CV < 2% for all reported concentrations).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Accuracy Test Set: n = 100 samples. The origin of the data (country, retrospective/prospective) is not specified in the provided text. It indicates "real human urine" was used for the precision study, which might apply to accuracy as well.
    • Precision Test Set:
      • Within Run: 20 replicates for each of the 9 spiked concentrations.
      • Run-to-Run: 90 replicates for each of the 9 spiked concentrations.
      • The samples were "real human urine" spiked with benzoylecgonine. The provenance (country, retrospective/prospective) is not specified.
    • Interfering Substances/Cross-Reactivity: Not a "test set" of patient samples, but rather analytical samples spiked with various substances into a "drug-free urine calibrator matrix." The number of samples for each substance is not specified, but multiple concentrations were tested.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This study does not involve human experts interpreting results. The ground truth for the test set is established using analytical reference methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as the ground truth is established by objective analytical methods (LCMS/GCMS). There is no "adjudication" of human interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a study validating an in vitro diagnostic (IVD) device for drug detection, not an AI-assisted diagnostic imaging study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, this study is inherently a standalone performance evaluation of the device. The "algorithm" here is the enzyme immunoassay itself, which provides a quantitative or qualitative result without human-in-the-loop interpretation beyond reading the instrument output. The comparison is made directly against a confirmatory analytical method (LCMS).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for the accuracy study is:

    • Highly sensitive and specific analytical methods: Liquid Chromatography/Mass Spectrometry (LCMS) for the candidate device, and Gas Chromatography/Mass Spectrometry (GCMS) for the predicate device. These are considered gold standards for confirming the presence and concentration of drugs/metabolites in biological samples.

    8. The sample size for the training set

    This document describes a validation study for a fully developed IVD device, not an AI model that requires a "training set." The device is a "ready-to-use, liquid reagent, homogenous enzyme immunoassay" and its "chemical make-up for both Reagent 1 and Reagent 2...is identical to what is in use with the predicate device." Therefore, there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for an AI model. The chemical composition and reaction principles of the immunoassay are based on established scientific methods and the predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1