510(k) Data Aggregation

    K Number
    K252894

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-01-06

    (117 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aprevo® cervical ACDF interbody system includes interbody fusion devices indicated at one or more levels of the cervical spine (C2-T1) in patients with cervical disc disease, instability, trauma including fractures, deformity defined as kyphosis, lordosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. Cervical disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The devices are to be filled with autograft bone and/or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone. The aprevo® cervical ACDF interbody system must be used with supplemental fixation (e.g., cervical plate or cervical posterior fixation).

    The aprevo® cervical ACDF-X interbody system includes interbody fusion devices indicated at one or more levels of the cervical spine (C2-T1) in patients with cervical disc disease, instability, trauma including fractures, deformity defined as kyphosis, lordosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. Cervical disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The devices are to be filled with autograft bone and/or allogenic bone graft comprised of cancellous, cortical, and/or corticocancellous bone. When used with the screws that accompany the device, the aprevo® cervical ACDF-X interbody system is intended for use as a standalone system. Deformity procedures to correct coronal angulation or use of a hyperlordotic device (>20° lordosis) must include supplemental fixation (e.g., cervical plate or cervical posterior fixation).

    Device Description

    The aprevo® cervical interbody system, which is comprised of the aprevo® cervical ACDF interbody system and the aprevo® cervical ACDF-X interbody system configurations, is designed to stabilize the cervical spinal column and facilitate fusion. The personalized aprevo® devices incorporate patient specific features and include an aperture intended for the packing of bone graft. The individualized surgical correction plan and device configurations are developed using patient radiological images.

    The aprevo® cervical interbody system interbody devices are additively manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 and provided sterile, and the screws are manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136 and provided sterile. The associated instruments, which facilitate the placement, adjustment, and removal, if necessary, of the interbody devices, are manufactured from stainless steel per ASTM A564 and provided sterile packaged for single patient use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252611

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-22

    (126 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aprevo® cervical plating system is intended for anterior cervical fixation (C2-T1) for the following indications:

    • Degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies),
    • Spondylolisthesis,
    • Trauma (i.e., fracture or dislocation),
    • Spinal stenosis,
    • Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
    • Tumor,
    • Pseudoarthrosis, and
    • Failed previous fusion.
    Device Description

    The aprevo® cervical plating system, which is comprised of the aprevo® cervical segmental plating system and the aprevo® cervical multilevel plating system configurations, is intended for anterior fixation of the cervical spine. The system consists of a variety of segmental and multilevel plates that are additively manufactured from titanium alloy (Ti-4Al-6V ELI) per ASTM F3001 as well as a range of fixation screws manufactured from titanium alloy (Ti-4Al-6V ELI) per ASTM F136. The associated instruments, which facilitate the placement, adjustment, and removal, if necessary, of the implants, are manufactured from stainless steel per ASTM A564. The plates, screws, and instruments are provided sterile packaged for single patient use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aprevo® anterior and lateral lumbar interbody fusion devices are intended for interbody fusion in skeletally mature patients and are to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches may include anterior lumbar interbody fusion or lateral lumbar interbody fusion.

    The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® anterior and lateral lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches.

    The aprevo® transforaminal lumbar interbody fusion device is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® transforaminal lumbar interbody fusion device is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches.

    The aprevo® transforaminal lumbar interbody fusion device is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® transforaminal lumbar interbody fusion devices are to be filled with auto graft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches.

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is intended for interbody fusion in skeletally mature patients. The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device is intended to be used with autograft and/or allogenic bone graft comprised of cancellous bone and/or corticocancellous bone and is to be used with supplemental fixation cleared for use in the lumbar spine. The device may be implanted via an open or minimally invasive approach.

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis, kyphosis, or sagittal), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is intended for standalone use at one or two levels of the spine when used with the screws that accompany the implant and with implants less than or equal to 20° of lordosis. At more than two levels or with implants greater than 20° of lordosis, the aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is intended to be used with the screws that accompany the implant and with supplemental fixation. When used at more than one level in patients with degenerative scoliosis and/or sagittal deformity, the aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) must be used with the screws that accompany the implant and with supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine. These patients should be skeletally mature and have had at least six months of non-operative treatment. The device is to be filled with autograft bone and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. The device may be implanted via an open or minimally invasive approach.

    The aprevo® TLIF-C Articulating System is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® TLIF-C Articulating System is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches.

    The aprevo® TLIF-C Articulating System is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® TLIF-C Articulating System devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches.

    The aprevo® Cervical ACDF lnterbody System are interbody fusion devices indicated at one or more levels of the cervical spine (C2-T1) in patients with the following degenerative cervical conditions: cervical disc disease, instability, trauma including fractures, deformity defined as kyphosis, lordosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. Cervical disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or cortico-cancellous bone. The aprevo® Cervical ACDF lnterbody System must be used with supplemental fixation systems. For hyperlordotic corrections (≥20° lordosis), the system must be used with at least an anterior cervical plate as supplemental fixation.

    The aprevo® Cervical ACDF-X lnterbody System are interbody fusion devices indicated at one or more levels of the cervical spine (C2-T1) in patients with the following degenerative cervical conditions: cervical disc disease, instability, trauma including fractures, deformity defined as kyphosis, lordosis, or scoliosis, cervical spondylotic myelopathy, spinal stenosis, and failed previous fusion. Cervical disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or cortico-cancellous bone. When used as a standalone system, the aprevo® Cervical ACDF-X lnterbody implant with integrated screw fixation is intended for use at multiple contiguous levels, or up to two levels when used in trauma, deformity or failed previous fusions. Deformity procedures to correct coronal angulation or any use of hyperlordotic correction (≥20°) must include supplemental fixation such as posterior cervical screw fixation or anterior plating.

    Device Description

    The aprevo® Lumbar Intervertebral Body Fusion Devices include ALIF, LLIF, TLIF, ALIF-X and TLIF-CA interbodies. The aprevo® lumbar interbody fusion devices are designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® device incorporates patient specific features to allow the surgeon to tailor the procedure to the individual needs of the patient and includes an aperture intended for the packing of bone graft. The individualized surgical correction plan and device configurations are developed using patient CT scans. The aprevo® devices are manufactured from titanium alloy (Ti-6Al-4V) per ASTM F3001, while the screws that accompany ALIF-X are machined from material per ASTM F136. The devices are accompanied by an inserter instrument which facilitates the placement of the interbodies. Both the interbody devices and instruments are provided as single use, sterile-packed product to the end user.

    The aprevo® Cervical Intervertebral Body Fusion Devices include ACDF, ACDF-X and ACDF-X no cams. The aprevo® Cervical ACDF System, which includes the aprevo® Cervical ACDF Interbody, aprevo® Cervical ACDF-X Interbody, and the aprevo® Cervical ACDF-X (NO CAM) Interbody are designed to stabilize the cervical spinal column and facilitate fusion. The personalized aprevo® devices incorporate patient-specific features to allow the surgeon to tailor the deformity correction to the individual needs of the patient and include an aperture for the packing of bone graft. The aprevo® Cervical ACDF System interbodies are additively manufactured from Ti-6Al-4V ELI titanium alloy per F3001, while the screws are machined from Ti alloy per ASTM F136. The devices are accompanied by an inserter instrument which facilitates the placement of the interbodies. Both the interbody devices and the instruments are provided to the end-user as single use, sterile-packed products.

    The purpose of this 510(k) is to extend the expiration of imaging used to create 3D surgical plans. There have been no changes made to the software used in the surgical planning process.

    AI/ML Overview

    The provided FDA 510(k) clearance letter (K250827) for Carlsmed, Inc.'s "aprevo®" interbody fusion devices primarily concerns the renewal of previously cleared devices and the extension of the imaging expiration dating used for 3D surgical planning. Therefore, the document does not contain details of a clinical study assessing the performance of an AI-based device meeting acceptance criteria in the way one might expect for a novel AI/ML medical device submission.

    Instead, the performance testing summary indicates that "Clinical data was used to evaluate imaging expiration. Expiration dating was validated using statistical analysis (DICE score) with an acceptance criterion of >0.80. Additionally, the performance was evaluated across key cohorts." This suggests that the "device" being evaluated for performance in this context is likely related to the software's ability to accurately segment or generate 3D models from imaging data over time, rather than the interbody fusion implants themselves. The clearance is for the physical implants, which are considered substantially equivalent to previously cleared predicates. The "performance testing" here seems to be an update or validation related to the software component of the surgical planning.

    Given this context, I will extract information relevant to the software's performance, as that is the only "acceptance criteria" and "study" mentioned, even if it's not a typical AI model performance study against clinical outcomes.

    Here's an interpretation based on the limited information provided, assuming the "device performance" refers to the imaging processing capabilities of the aprevo® system's software, specifically regarding the "expiration" of imaging data for 3D surgical plans:


    Acceptance Criteria and Device Performance Study for Imaging Expiration

    The context of this FDA 510(k) submission (K250827) indicates that the performance study was specifically undertaken to extend the expiration of imaging used to create 3D surgical plans. This suggests the "device performance" being assessed relates to the consistency and accuracy of the software's ability to process imaging data for surgical planning over time, implying that "aging" of the input imaging data does not degrade the software's output beyond an acceptable threshold.

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance CriteriaReported Device Performance
    DICE Score (for imaging expiration/segmentation accuracy)> 0.80Not explicitly stated, but implied to meet the criterion as the 510(k) was cleared. The text says "Expiration dating was validated using statistical analysis (DICE score) with an acceptance criterion of >0.80."

    Note: The document only specifies the acceptance criterion. The specific numerical performance result (e.g., the exact DICE score achieved) is not provided in this public clearance letter. However, the clearance implies that the criterion was met.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not specified in the provided text.
    • Data Provenance: "Clinical data was used." No details on the country of origin or whether the data was retrospective or prospective are provided.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not specified. (e.g., 2+1, 3+1, none).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done?: No, an MRMC study comparing human readers with and without AI assistance is not mentioned or implied by the description of this performance evaluation, which focuses on "imaging expiration" and DICE score.

    6. Standalone (Algorithm Only) Performance

    • Standalone Performance Done?: Yes, the use of a "DICE score" for "imaging expiration" strongly suggests an algorithm-only performance evaluation, likely comparing the software's segmentation or 3D reconstruction output against a form of ground truth derived from the clinical data. The phrase "There have been no changes made to the software used in the surgical planning process" further indicates an assessment of the software's intrinsic capabilities.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The use of "DICE score" typically implies that the ground truth involves expert-annotated segmentations or regions of interest within the medical images. The DICE score measures the spatial overlap between the algorithm's output and the ground truth. While not explicitly stated, it's highly probable that this ground truth was derived from expert consensus or manual annotation of anatomical structures on the clinical imaging data.

    8. Sample Size for the Training Set

    • Training Set Sample Size: Not mentioned. This submission is for an extension/validation of existing software, not the initial clearance of a newly developed AI model, so details on original training are not typically included here.

    9. How Ground Truth for the Training Set Was Established

    • Ground Truth Establishment for Training Set: Not mentioned. As above, this document focuses on a validation study for imaging expiration, not the initial development or training of the AI algorithm.

    Summary of Limitations Based on Provided Text:

    The provided 510(k) summary (specifically the "Performance Testing Summary") is extremely concise regarding details of the study. It states that "Clinical data was used to evaluate imaging expiration" and that "Expiration dating was validated using statistical analysis (DICE score) with an acceptance criterion of >0.80. Additionally, the performance was evaluated across key cohorts."

    Therefore, many typical details requested for robust AI/ML performance studies, such as specific sample sizes, expert qualifications, and adjudication methods, are not present in this public clearance letter. This is common for 510(k) summaries, which are often condensed versions of the full submission. The key takeaway is that the "performance testing" here addresses a specific aspect ("imaging expiration" for 3D surgical planning) of the software rather than a direct clinical efficacy study of the implant or a broad AI diagnostic/prognostic model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243802

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-03-17

    (96 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    aprevo® anterior and lateral lumbar interbody fusion device:

    The aprevo® anterior and lateral lumbar interbody fusion devices are interbody fusion in skeletally mature patients and are to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches may include anterior lumbar interbody fusion or lateral lumbar interbody fusion.

    The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) months of non-operative treatment. aprevo® anterior and lateral lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implanted via a variety of open or minimally invasive approaches.

    aprevo® anterior lumbar interbody fusion device with interfixation:

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is intended for interbody fusion in skeletally mature patients and is to be used with supplementation cleared for use in the lumbar spine. The aprevo® anterior lumbar interbody fusion device with interfixation is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous bone and/or cortico-cancellous bone. These implants may be implanted via a variety of open or minimally invasive approaches. ALIF-X implants are intended for standalone use at one or two levels of the spine when used with the screws that accompany the implant and with implants less than or equal to 20° of lordosis. At more than two levels or with implants greater than 20° of lordosis, ALIF-X is intended to be used with the screws that accompany each implant and with supplemental fixation. When used at more than one level in patients with degenerative scoliosis and/or sagittal deformity, the aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    Device Description

    aprevo® anterior and lateral lumbar interbody fusion device:

    The aprevo® anterior and lateral lumbar interbody fusion devices are designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® devices incorporate patient specific features to allow the surgeon to tailor the deformity correction to the individual needs of the patient and include an aperture intended for the packing of bone graft. The individualized surgical correction plan and device configurations are developed using patient radiological images.

    The aprevo® anterior and lateral lumbar interbody fusion devices are additively manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 and are provided sterile. The associated instruments, which facilitate the placement, adjustment, and removal, if necessary, of the interbody devices, are manufactured from stainless steel and provided sterile packaged for single use.

    aprevo® anterior lumbar interbody fusion device with interfixation:

    The aprevo® anterior lumbar interbody fusion device with interfixation is designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® device incorporates patient specific features to allow the surgeon to tailor the procedure to the individual needs of the patient and include an aperture intended for the packing of bone graft. The individualized surgical correction plan and device configurations are developed using patient radiological images.

    The aprevo® anterior lumbar interbody fusion device with interfixation is additively manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 and is provided sterile. The device includes screws that are manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136 and are offered in a range of lengths. The associated instruments, which facilitate the placement, adjustment, and removal, if necessary, of the device, are manufactured from stainless steel and provided sterile packaged for single use.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "aprevo® anterior and lateral lumbar interbody fusion device" and "aprevo® anterior lumbar interbody fusion device with interfixation." This document explicitly states that "Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data."

    Therefore, no clinical study was conducted to prove the device meets acceptance criteria. The FDA's substantial equivalence determination was based on non-clinical testing.

    Here's a breakdown of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Since no clinical study was performed, there are no specific clinical acceptance criteria or reported device performance in this context. The non-clinical testing primarily demonstrates that the subject device has similar mechanical properties and performance as the predicate devices.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Not explicitly stated as clinical criteria, but implied by non-clinical testing)Reported Device Performance
    Mechanical PerformanceDevice must exhibit similar mechanical properties (e.g., strength, durability) to predicate devices under various loads."The subject configuration's manufacturing processes are identical to those of the predicate devices, and mechanical testing confirmed that the subject devices have the same mechanical properties as the predicate devices."
    Specific tests were performed for:Passed / Met expectations (implied by conclusion of substantial equivalence)
    - Static and dynamic compression
    - Static and dynamic compression shear
    - Tensile testing (per ASTM F3001)
    Material PropertiesDevice material (titanium alloy Ti-6Al-4V ELI) must conform to ASTM F3001 and F136 standards."additively manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F3001" and "screws that are manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136"; Microstructure and chemical composition assessed.
    Design/FunctionalityDevice should perform its intended function of stabilizing the lumbar spinal column and facilitating fusion, consistent with predicate devices."The subject devices have identical intended use, indications for use, mechanical properties, raw materials, sterilization, and packaging as the predicate devices."

    2. Sample size used for the test set and the data provenance

    Not applicable, as no clinical test set was used for proving acceptance criteria in a comparative clinical study. The non-clinical testing would involve physical samples of the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical test set requiring expert ground truth was utilized.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set was utilized.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is an intervertebral body fusion device, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical implant, not a software algorithm.

    7. The type of ground truth used

    For the purpose of non-clinical testing, the "ground truth" would be the established engineering and material standards (e.g., ASTM F3001, ASTM F136) and the performance characteristics of the legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable, as this is a physical medical device and not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242599

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-20

    (112 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aprevo® Digital Planning software is intended to be used by trained, medically knowledgeable design personnel to generate surgical plans for skeletally mature patients. The device processes a 3D spine model in conjunction with inputs from healthcare professionals to produce 3D models of spinal corrections and anatomical measurements.

    Device Description

    The aprevo® Digital Planning, is a rules-based, semi-automated surgical planning software that, based on surgeon input, measures the 3D spine model and results in a spinal correction plan. The outputs consist of a surgical plan and 3D spinal correction assets, which are then reviewed by the surgeon. The device is operated by trained, medically knowledgeable Carlsmed design personnel.

    AI/ML Overview

    The provided document is a 510(k) summary for the aprevo® Digital Planning software. While it describes the software and its intended use, it does not contain specific details on acceptance criteria or a comprehensive study report with quantitative performance metrics that would typically be found in a detailed validation study.

    The relevant section for performance criteria and testing is under "Non-clinical Testing" (page 6). It states: "The surgical plans generated by aprevo® Digital Planning were shown to be substantially equivalent, with 90% confidence and reliability, to plans manually created by medically knowledgeable operators." This is a summary statement of a study result, but it lacks the detailed breakdown requested.

    Therefore, many of the requested fields cannot be filled directly from the provided text. I will fill in what can be inferred or explicitly stated and highlight where information is missing.


    Acceptance Criteria and Study for aprevo® Digital Planning Software

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Quantitative / Qualitative)Reported Device Performance
    Surgical plans generated by aprevo® Digital Planning are substantially equivalent to plans manually created by medically knowledgeable operatorsSubstantially equivalent with 90% confidence and reliability to manually created plans.
    Software performs according to specifications.Verified.
    Software performs according to user requirements.Verified.

    Missing Information:

    • Specific quantitative metrics for "substantial equivalence" (e.g., specific measurement tolerances for anatomical landmarks, angular measurements, or correctional values). The document states "90% confidence and reliability," which is a statistical measure of the comparison, not the quantitative metric itself.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in the provided text.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study involved comparing software-generated plans to "plans manually created by medically knowledgeable operators," implying retrospective data processed by operators.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not explicitly stated.
    • Qualifications of Experts: Described as "medically knowledgeable operators" and "Internal Carlsmed personnel only" (page 6). Specific qualifications like "radiologist with 10 years of experience" are not provided.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not explicitly stated. The comparison is between the automated software plans and "manually created plans," but how discrepancies, if any, were resolved for the ground truth is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done? Not explicitly mentioned as an MRMC study in the standard sense (human readers assisting with/without AI). The study described is a comparison between software-generated plans and manually created plans. It's not a study on how human readers improve with AI assistance, but rather a performance validation comparing the AI's output to human-generated output.
    • Effect Size of Human Reader Improvement: Not applicable given the type of study described.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Standalone Study Done? Yes, in a sense. The "Non-clinical Testing" section validates the performance of the software-generated surgical plans against manually created plans. While the software takes "inputs from healthcare professionals" and trained personnel operate it, the evaluation seems to focus on the output of the algorithms ("surgical plans generated by aprevo® Digital Planning") compared to a human-derived ground truth. The device is described as "rules-based, semi-automated."

    7. Type of Ground Truth Used

    • Ground Truth Type: "Plans manually created by medically knowledgeable operators" (page 6). This is a form of expert consensus or expert-derived ground truth.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not explicitly stated in the provided text. The document describes the software as "rules-based, semi-automated" and uses algorithms for landmark determination and surgical plan generation. This implies a rule-based system or potentially a machine learning model, but training set details are not provided.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment for Training Set: Not explicitly stated. Given the description of the device as "rules-based, semi-automated," it suggests that the "rules" and "algorithms" were likely developed based on medical knowledge and potentially expert-verified data, but the specific process for establishing ground truth for any potential training data is not detailed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243635

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-13

    (18 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® anterior lumbar interbody fusion device with interfixation is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous bone and/or cortico-cancellous bone. These implants may be implanted via a variety of open or minimally invasive approaches. ALIF-X implants are intended for standalone use at one or two levels of the spine when used with the screws that accompany the implant and with implants less than or equal to 20 degrees of lordosis. At more than two levels or with implants greater than 20 degrees of lordosis, ALIF-X is intended to be used with the screws that accompany each implant and with supplemental fixation. When used at more than one level in patients with degenerative scoliosis and/or sagittal deformity, the aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

    Device Description

    The aprevo® anterior lumbar interbody fusion device with interfixation is designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® device incorporates patient specific features to allow the surgeon to tailor the procedure to the individual needs of the patient and includes an aperture intended for the packing of bone graft. The individualized surgical correction plan and device configurations are developed using patient CT scans. The aprevo® devices are manufactured from titanium alloy (Ti-6Al-4V) per ASTM F3001, while the screws are machined from material per ASTM F136. The devices are accompanied by an inserter instrument which facilitates the placement of the interbodies. Both the interbody devices and instrument are provided as single use, sterile-packed product to the end user. The purpose of this 510(k) is to modify the indications for use of the subject device to include standalone indications.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study proving that a device meets such criteria. The document is an FDA 510(k) clearance letter and a 510(k) Summary for a medical device called "aprevo® anterior lumbar interbody fusion device with interfixation."

    This type of document primarily focuses on establishing substantial equivalence to previously cleared devices, rather than detailing specific performance acceptance criteria and studies demonstrating adherence to those criteria for software or AI/ML components. The text describes the device, its intended use, comparison to predicate devices, and regulatory classification. It does not mention any AI/ML components, software, or studies evaluating algorithm performance, human reader improvement with AI assistance, or related aspects.

    Therefore, I cannot fulfill your request for information about acceptance criteria and a study that proves the device meets them based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241477

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-09

    (138 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® anterior lumbar interbody fusion device with interfixation is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an OD! >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous bone and/or cortico-cancellous bone. These implanted via a variety of open or minimally invasive approaches. The aprevo® anterior lumbar interbody fusion device with interfixation is intended to be used with the screws that accompany each implant and with supplemental fixation.

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with mvelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches. The aprevo® anterior lumbar interbody fusion device with interfixation is intended to be used with the screws that accompany each implant and with supplemental fixation.

    Device Description

    The aprevo® anterior lumbar interbody fusion device with fixation (ALIF-X is) designed to stabilize the lumbar spinal column and promote fusion. The personalized aprevo® interbodies incorporate patient specific features to allow the surgeon to tailor the deformity correction to the individual needs of the patient and include an aperture intended for the packing of bone graft. The aprevo® ALIF-X device is accompanied by three screws designed to be used to provide fixation.

    The aprevo® ALIF-X interbodies are additively manufactured from Ti-6A1-4V ELI titanium alloy per ASTM F3001, while the screws are machined from material per ASTM F136. The devices are accompanied by an inserter instrument which facilitates the placement of the interbodies. Both the interbody devices and instrument are provided as single use, sterile-packed product to the end user.

    AI/ML Overview

    The provided text is a 510(k) Summary for the aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X). It outlines the device, its intended use, and a summary of technological similarities and differences to a predicate device.

    However, the document does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of AI/ML performance, such as diagnostic accuracy, sensitivity, or specificity. The "Performance Testing Summary" section refers to mechanical testing standards (ASTM F2077 and ASTM F3001) for physical properties of the implant, not clinical or AI/ML performance.

    Therefore, I cannot provide the requested information from the given text. The questions about test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details are not addressed in this filing, as they are not relevant to the type of device being described (a physical spinal implant, not an AI/ML-driven diagnostic or therapeutic tool).

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® anterior lumbar interbody fusion device with interfixation is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous bone and/or cortico-cancellous bone. These implanted via a variety of open or minimally invasive approaches. The aprevo® anterior lumbar interbody fusion device with interfixation is intended to be used with the screws that accompany each implant and with supplemental fixation.

    The aprevo® anterior lumbar interbody fusion device with interfixation (ALIF-X) is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment, aprevo® lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches. The aprevo® anterior lumbar interbody fusion device with interfixation is intended to be used with the screws that accompany each implant and with supplemental fixation.

    aprevo anterior and lateral lumbar interbody fusion devices (ALIF and LLIF):

    The aprevo® anterior and lateral lumbar interbody fusion devices are intended for interbody fusion in skeletally mature patients and are to be used with supplementation cleared for use in the lumbar spine. The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI>40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches may include anterior lumbar interbody fusion or lateral lumbar interbody fusion.

    The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non- operative treatment. aprevo anterior and lateral lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implanted via a variety of open or minimally invasive approaches.

    aprevo® transforaminal lumbar interbody fusion devices (TLIF):

    The aprevo® transforaminal lumbar interbody fusion device is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® transforaminal lumbar interbody fusion device is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches.

    The aprevo® transforaminal lumbar interbody fusion device is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® transforaminal lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implanted via a variety of open or minimally invasive approaches.

    Device Description

    The aprevo® Intervertebral Body Fusion Devices include ALIF-X, ALIF, LLIF, TLIF-O, and TLIF-C interbodies. The aprevo® interbody fusion devices are designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® devices incorporate patient specific features to allow the surgeon to tailor the deformity correction to the individual needs of the patient.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "aprevo® anterior and lateral lumbar interbody fusion device". It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices, particularly regarding changes in a manufacturing process (addition of a new heat treatment).

    Based on the document, this is not a study of a device that uses Artificial Intelligence (AI) or machine learning (ML), nor does it involve a multi-reader multi-case (MRMC) comparative effectiveness study, human readers, or ground truth establishment in the context of diagnostic or prognostic AI/ML applications. The device in question is an implantable interbody fusion device, and the "performance" described relates to mechanical properties and material characteristics, not diagnostic accuracy or expert interpretation.

    Therefore, many of the requested elements for describing acceptance criteria and study that proves a device meets them (especially those related to AI/ML, human readers, sample sizes for test/training sets, ground truth establishment for interpretative tasks, and effect sizes of AI assistance) are not applicable or extractable from this document.

    The document primarily focuses on non-clinical testing/performance data to demonstrate substantial equivalence following a manufacturing change.

    Here's a breakdown of the relevant and non-relevant information based on your request:


    Acceptance Criteria and Study for aprevo® Interbody Fusion Device

    Study Type: Non-clinical (mechanical and material performance) to demonstrate substantial equivalence for a manufacturing process change. This is not a clinical study or a study involving AI/ML performance.

    Applicable Sections from Request:

    • 1. A table of acceptance criteria and the reported device performance: This can be partially derived from the Discussion of Non-Clinical Testing/Performance Data section, as it outlines the tests performed to show equivalence.
    • 2. Sample size used for the test set and the data provenance: For mechanical testing, this refers to the number of devices tested. Data provenance is not explicitly stated but is implied to be from internal testing by the manufacturer.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This is not applicable in the context of an implantable device's performance, but if interpreted as device-only performance (without human input during its function), then yes, the mechanical tests are "standalone" in that sense.
    • 7. The type of ground truth used: For mechanical properties, the "ground truth" would be established engineering standards (ASTM F2077-18, ASTM F3001-14) that define acceptable performance.

    Non-Applicable Sections from Request (as per the provided document's content):

    • 3. Number of experts used to establish the ground truth... and qualifications: Not applicable for mechanical tests of an implantable device.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done... effect size: Not applicable. This is not an AI/ML diagnostic or image interpretation device.
    • 8. The sample size for the training set: Not applicable. This is not an AI/ML device.
    • 9. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.

    Detailed Response:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details that the "subject device underwent mechanical performance testing to validate that the performance of the worst-case subject devices is substantially equivalent to the previously cleared predicate devices." The acceptance criteria implicitly involve meeting the performance requirements of the cited ASTM standards, and the reported performance is that substantial equivalence was demonstrated.

    Acceptance Criteria CategorySpecific Test Standard / GoalReported Device Performance (Summary)
    Mechanical PerformanceASTM F2077-18: Methods for intervertebral body fusion devicesDemonstrated substantial equivalence to predicate devices.
    - Static/Dynamic CompressionThe results "did not identify any new or increased risks".
    - Static/Dynamic Compression Shear
    Material PerformanceASTM F3001-14: Standard specification for additively manufactured titanium-6 aluminum-4 vanadium alloysDemonstrated substantial equivalence to predicate devices.
    - Tensile TestingThe results "did not identify any new or increased risks".
    - Microstructure Assessment per ASTM F3001
    - Chemical Composition Assessment per ASTM F3001
    Overall ComparisonSubstantial Equivalence to Predicate Devices (K222009, K222082, K231140) in Function and PerformanceConcluded to be substantially equivalent in design, function, and performance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the number of devices (samples) used for each mechanical test. Standard practice for such tests under ASTM typically involves a minimum number of samples (e.g., n=5 or n=6 per condition) to achieve statistical significance. However, the exact sample sizes are not provided in this summary.
    • Data Provenance: The data provenance is implied to be from internal testing conducted by Carlsmed, Inc., as part of their premarket notification submission to the FDA. The country of origin for the data is not specified, but the company is based in Carlsbad, California, USA. The testing is prospective in the sense that it was conducted specifically to demonstrate equivalence for this submission, although the data itself is on device performance.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not Applicable. This is a mechanical/material performance study of an implantable medical device, not a study involving human interpretation of medical images or data. "Ground truth" for these tests refers to the objective physical measurements against established engineering standards.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No human adjudication process is involved as it's not an interpretive or diagnostic study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size:

    • No. This type of study is typically performed for diagnostic or AI-assisted diagnostic devices. This device is an implantable interbody fusion device, and its performance is assessed via mechanical and material properties, not human reader accuracy with or without AI assistance.

    6. If a Standalone Performance Study was done:

    • Yes (in principle for mechanical device testing). The "Discussion of Non-Clinical Testing/Performance Data" describes tests performed on the device itself (e.g., static/dynamic compression, tensile testing) to evaluate its mechanical and material properties against established standards. These tests are performed on the device samples in a laboratory setting, independent of human clinical application for their evaluation of device performance in terms of strength, integrity, etc.

    7. The Type of Ground Truth Used:

    • The "ground truth" for this study is established by validated engineering standards and specifications (ASTM F2077-18 for mechanical properties and ASTM F3001-14 for material properties). The performance of the devices is measured against the requirements outlined in these standards and compared to the performance of predicate devices.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is not an AI/ML device, so there is no training set in the context of machine learning.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. As there is no training set for an AI/ML model, this question does not apply.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232282

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-22

    (235 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    aprevo® anterior and lateral lumbar interbody fusion devices (ALIF and LLIF):

    The aprevo® anterior and lateral lumbar interbody fusion devices are intended for interbody fusion in skeletally mature patients and are to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The devices are intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches may include anterior lumbar interbody fusion or lateral lumbar interbody fusion.

    The aprevo® anterior and lateral lumbar interbody fusion devices are indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® anterior and lateral lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implanted via a variety of open or minimally invasive approaches.

    aprevo® transforaminal lumbar interbody fusion devices (TLIF):

    The aprevo® transforaminal lumbar interbody fusion device is intended for interbody fusion in skeletally mature patients and is to be used with supplemental fixation instrumentation cleared for use in the lumbar spine. The aprevo® transforaminal lumbar interbody fusion device is indicated for use as an adjunct to fusion at one or more levels of the lumbar spine in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. These patients should have had six months of non-operative treatment. The device is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft. These implants may be implanted via a variety of open or minimally invasive approaches.

    The aprevo® transforaminal lumbar interbody fusion device is indicated for use at one or more levels of the lumbosacral spine as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. aprevo® transforaminal lumbar interbody fusion devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems). These implants may be implanted via a variety of open or minimally invasive approaches.

    Device Description

    The aprevo® Interbody Fusion Devices include ALIF, LLIF, and TLIF interbodies. The aprevo® Interbody Fusion Devices are designed to stabilize the lumbar spinal column and facilitate fusion. The personalized aprevo® devices incorporate patient specific features to allow the surgeon to tailor the deformity correction to the individual needs of the patient.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (aprevo® anterior and lateral lumbar interbody fusion device, aprevo® transforaminal lumbar interbody fusion device). It does not contain information about an AI/ML component, nor does it discuss acceptance criteria and studies related to AI/ML performance.

    Instead, the document focuses on:

    • Substantial Equivalence: The primary goal of a 510(k) submission is to demonstrate that a new device is substantially equivalent to a legally marketed predicate device.
    • Device Description and Indications for Use: It describes the device and its intended uses.
    • Technological Similarities and Differences: It explicitly states that the subject devices are identical to the predicate devices in terms of intended use, indications for use, dimensions, mechanical properties, raw materials, sterilization, and packaging. The only noted difference is the addition of a new Additive Manufacturing supplier.
    • Non-Clinical Testing/Performance Data: The study described is mechanical performance testing (e.g., ASTM F2077, ASTM F3001) to validate that the new supplier's manufacturing processes result in devices that perform equivalently to the predicate devices. This is crucial for demonstrating that the change in supplier does not introduce new risks or alter performance.

    Therefore, I cannot extract the requested information regarding acceptance criteria and performance studies for an AI/ML device from this document. The questions you've asked are specific to AI/ML device validation (e.g., sample size for test/training sets, expert adjudication, MRMC studies, standalone performance, ground truth). This document is for a physical medical implant where "performance" is assessed through biomechanical testing, not AI/ML metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231955

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-11-03

    (123 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    aprevo® Digital Segmentation software is intended to be used by trained, medically knowledgeable design personnel to perform digital image segmentation of the spine, primarily lumbar anatomy. The device inputs DICOM images and outputs a 3-D model of the spine.

    Device Description

    The device is a software medical device that will use DICOM images as input and provide 3D model of the spine structure. Pre-processing will be performed on the uploaded DICOM files to filter soft tissue and identifying spine. Upon removal of soft tissue and identification of spine structure, the software will utilize an AI-based algorithm to segment the structure and render a 3D model as an output.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the "aprevo® Digital Segmentation" software.

    Here's a breakdown of the requested information:


    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    IOU (Intersection Over Union) score > 80%Exceeded 80%
    Vertebral body labeling accuracy > 90%Exceeded 90% overall
    Vertebral body labeling sensitivity > 80%Exceeded 80%
    Vertebral body labeling specificity > 80%Exceeded 80%

    Study Details:

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: Not explicitly stated in the provided text, but it mentions that "Independent training and validation datasets were selected to ensure model performance would reflect real clinical performance" and "Validation datasets represented diversity in populations and equipment."
      • Data Provenance: Not explicitly stated, however, the phrase "diversity in populations and equipment" suggests data from various sources but does not specify countries of origin. The study was a "Non-Clinical Testing" which implies retrospective data.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • Not specified. The document states that the ground truth for the algorithm was used for model performance evaluation, but does not detail how this ground truth was established, or the number/qualifications of experts involved.
    3. Adjudication Method for the Test Set:

      • Not specified.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • If done: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or indicated. The document states "CLINICAL TESTING: Not applicable." The study solely focuses on the standalone performance of the software.
      • Effect size of human readers improvement: Not applicable, as no MRMC study was conducted.
    5. Standalone Performance (Algorithm only without human-in-the-loop):

      • If done: Yes, a standalone performance evaluation was done. The "NON-CLINICAL TESTING" section describes the evaluation of the "software performance" using IOU and accuracy metrics for segmentation and labeling, without human intervention in the reported performance metrics.
    6. Type of Ground Truth Used:

      • The type of ground truth used is not explicitly stated as expert consensus, pathology, or outcomes data. However, for "segmentation" and "vertebral body labeling," the ground truth would typically be established by expert annotation or a similar gold standard, refined through a consensus process, but this is not detailed.
    7. Sample Size for the Training Set:

      • Not explicitly stated. It only mentions that "Independent training and validation datasets were selected to ensure model performance would reflect real clinical performance."
    8. How the Ground Truth for the Training Set Was Established:

      • Not explicitly stated. The document mentions that "Independent training and validation datasets were selected," but does not elaborate on the method used to establish the ground truth for the training data (e.g., expert annotations, manual segmentation).
    Ask a Question

    Ask a specific question about this device