Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K210324

    Validate with FDA (Live)

    Date Cleared
    2021-03-04

    (28 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V. Mueller Cosgrove™ Flex Clamps are indicated to occlude a blood vessel temporarily. Used in pulmonary and gastrointestinal procedures and can be used to clamp over indwelling catheters. Also used in minimally invasive and standard open cardiovascular procedures for temporary occlusion of a blood vessel.

    Device Description

    Each V. Mueller Cosgrove™ Flex Clamp device consists of a ring handle, flexible shaft and jaw. The design of the flexible shaft allows the end user to bend the shaft following clamping of the blood vessel providing the user a clear field of visualization to the surgical site.

    The proposed device and the predicate device have a similar design, the same materials and the same manufacturing and fabrication methods/processes. The main difference between the subject device and the predicate device is a change to the wire configuration of the cable. In addition, the cleaning and sterilization instructions in the instructions for use were updated in accordance with current guidance. These changes do not raise different questions of safety or effectiveness.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "V. Mueller Cosgrove Flex Clamps." This device is a vascular clamp used to temporarily occlude blood vessels.

    Based on the provided text, there is no information about a study proving the device meets acceptance criteria related to an AI or algorithm-driven system. The document, K210324, is for a physical medical device (vascular clamps) and its substantial equivalence to previously marketed predicate devices. The "Performance Data" section explicitly states: "N/A - No clinical tests were conducted for this submission." and "The results of the non-clinical tests demonstrate the V. Mueller Cosgrove™ Flex Clamps meet all performance requirements and are substantially equivalent to the predicate devices."

    Therefore, I cannot provide the requested information regarding acceptance criteria and studies for an AI/algorithm-driven device, as this document is not about such a device. The "acceptance criteria" and "performance data" discussed in the document relate to the physical properties, reusability, sterilization, and biocompatibility of the vascular clamp itself, as demonstrated by non-clinical tests.

    If you intended to ask about a different document or a different type of device, please provide that information.

    However, I can extract and present the existing acceptance criteria and performance data for the V. Mueller Cosgrove Flex Clamps as presented in the document, though it won't match the format for an AI/algorithm study:


    Acceptance Criteria and Device Performance for V. Mueller Cosgrove Flex Clamps (K210324)

    1. Table of Acceptance Criteria and the Reported Device Performance (Non-Clinical)

    Acceptance Criteria (Characteristic)Standard / Test / FDA GuidanceReported Device Performance (Summary of Results)
    Instrument strengthStrength Tests, Life Cycle TestPASS
    Device must be reusableCleaning and Sterilization TestsPASS
    Device is able to be cleaned and sterilizedAAMI TIR12, AAMI TIR30, ANSI AAMI ST79, ISO 11138, ISO 17664, ISO 17665PASS
    Device materials are biocompatibleISO 10993PASS

    Note: These criteria pertain to the physical characteristics and functionality of the vascular clamp, not to an AI or algorithm.

    Regarding the other points you requested for an AI/algorithm study, they are not applicable to this document as it concerns a physical medical device:

    1. Sample size used for the test set and the data provenance: Not applicable. The performed tests are non-clinical (e.g., strength tests, material biocompatibility tests), not involving data sets like those for AI.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI validation (e.g., expert consensus on medical images) is not relevant here. Device testing would be performed by qualified engineers/technicians according to established protocols.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is not an AI.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this physical device, "ground truth" would be the engineering specifications and performance standards (e.g., tensile strength, sterilization efficacy), which are determined by established scientific methods and industry standards. It's not a "truth" derived from human interpretation of data in the AI sense.
    7. The sample size for the training set: Not applicable. This device is not an AI.
    8. How the ground truth for the training set was established: Not applicable. This device is not an AI.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163615

    Validate with FDA (Live)

    Date Cleared
    2017-05-04

    (133 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Snowden-Pencer MicroLap 3mm Laparoscopic Instruments are indicated to be used in laparoscopic and other minimally invasive procedures for cutting, clamping, suturing and dissecting and to allow high frequency monopolar cutting and coagulation.

    Device Description

    Snowden-Pencer MicroLap 3mm Laparoscopic Instruments consists of a handle and insert assembly. The insert assembly contains a shaft and end-effector (jaw pattern). There are several device models that encompass various lengths, diameters and jaw patterns based on the surgeon's needs.

    AI/ML Overview

    The provided document is a 510(k) summary for the Snowden-Pencer MicroLap 3mm Laparoscopic Instruments. This type of submission is for medical devices that demonstrate substantial equivalence to a legally marketed predicate device, and it typically relies on non-clinical performance data rather than extensive clinical studies as seen with AI/ML devices. Therefore, much of the information requested in your prompt (e.g., sample sizes for training/test sets, expert adjudication methods, MRMC studies, standalone algorithm performance) is not applicable or not found in this document.

    However, I can extract the relevant information regarding the device's acceptance criteria and the studies that prove it meets those criteria based on the provided text.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria / CharacteristicPerformance Results (Acceptance Criteria implicitly met if "PASS")
    Instrument strengthPASS (Strength Tests)
    Device must be reusablePASS (Cleaning and Sterilization Tests)
    Device must maintain pneumoperitoneumPASS (Pneumoperitoneum Test)
    Device to meet required electrosurgical safety standardsPASS (IEC 60601-1, IEC 60601-2, IEC 60601-2-2, IEC 60601-2-18)
    Device is able to be cleaned and sterilizedPASS (AAMI TIR12, AAMI TIR30, ANSI AAMI ST79, ANSI AAMI ST81, ISO 11138, ISO 17664, ISO 17665)
    Device materials are biocompatiblePASS (ISO 10993)

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: This document does not specify the sample sizes (i.e., number of instruments tested) for each of the non-clinical performance tests. It states "Results Summary," which typically means the overall outcome for the tested batch/type.
    • Data Provenance: Not specified, but these are typically lab-based, internal company tests conducted to industry standards. They are non-clinical and not derived from patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The "ground truth" for non-clinical performance tests like instrument strength or sterilization efficacy is established by adherence to recognized industry standards (e.g., ISO, AAMI, IEC standards), not by expert consensus in the medical sense for diagnostic tasks. The "experts" would be the engineers and technicians performing and verifying the tests in accordance with these standards.

    4. Adjudication method for the test set:

    • Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in diagnostic studies involving human interpretation or clinical outcomes. For performance tests of a physical device, the "adjudication" is the objective measurement against a defined standard.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a physical surgical instrument, not an AI/ML diagnostic tool. Therefore, an MRMC study is not relevant or performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a physical surgical instrument, not an algorithm. Standalone performance for an algorithm is not applicable.

    7. The type of ground truth used:

    • The "ground truth" for these tests is defined by the specific parameters and pass/fail criteria outlined in the referenced national and international performance standards (e.g., IEC 60601 series for electrosurgical safety, ISO 10993 for biocompatibility, AAMI/ANSI standards for cleaning and sterilization). For example, a "strength test" would have a defined force it must withstand without breaking or deforming.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" in the context of a physical medical device like a laparoscopic instrument. (This term is relevant for AI/ML algorithms).

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162432

    Validate with FDA (Live)

    Date Cleared
    2017-01-18

    (140 days)

    Product Code
    Regulation Number
    870.2900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Multi-Link Cable and Lead Wire System is intended to transmit ECG signals from patient monitors for monitoring purposes. The Multi-Link Cable and Lead Wire System is limited to indications for use of the connected monitoring equipment. The Multi-Link trunk cables) are reusable, nonsterile and can be reprocessed. The Multi-Link lead wires are available reusable (single patient use). The Multi-Link Cable and Lead Wire System is compatible with GE Healthcare, Philips, Mindray, Spacelabs and Nihon Kohden electrocardiograph monitors.

    Device Description

    The Multi-Link X2 ECG Cable and Lead Wire Systems are a combination of reusable ECG trunk cables and single-patient use lead wires (SPUL) used to transmit signals from patient electrodes to various electrocardiograph monitors for monitoring purposes. This type of device is common to both the industry and to most medical establishments. The Multi-Link X2 ECG Cable and Lead Wire Systems are not stand alone devices but are accessories to the host monitoring devices.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device called the "Multi-Link X2 ECG Cable and Lead Wire System." This document aims to demonstrate that the new device is substantially equivalent to legally marketed predicate devices. It covers the device's indications for use, technological characteristics, and performance data, primarily focusing on compliance with established medical device standards.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a quantitative manner for specific diagnostic performance metrics (e.g., sensitivity, specificity, AUC). Instead, the device performance is demonstrated through adherence to recognized medical device standards. The "Performance Data" section lists the standards the device was tested against and states that the tests were performed to support substantial equivalence.

    Performance CharacteristicStandardReported Device Performance
    Biocompatibility
    CytotoxicityAAMI ANSI ISO 10993-5:2009/ (R2014)The lead wire is considered to be a skin contact with prolonged duration. Tests for Cytotoxicity, Sensitization, and Irritation were performed (implied successful by submission).
    SensitizationAAMI ANSI ISO 10993-10:2010/ (R2014)The lead wire is considered to be a skin contact with prolonged duration. Tests for Cytotoxicity, Sensitization, and Irritation were performed (implied successful by submission).
    IrritationAAMI ANSI ISO 10993-10:2010/ (R2014)The lead wire is considered to be a skin contact with prolonged duration. Tests for Cytotoxicity, Sensitization, and Irritation were performed (implied successful by submission).
    General BiocompatibilityAAMI ANSI ISO 10993-1:2009/ (R) 2013Biological evaluation performed within a risk management process (implied successful by submission).
    Electrical Performance
    Compatibility Testing with Bedside Monitors60601-2-27:2011Tests performed to ensure compatibility (implied successful by submission).
    Lifecycle and Contact Resistance (Instrument Connectors)EC53:2013 (Section 5.3.5 and 5.3.7)Tests performed (implied successful by submission).
    Grabber and Snap Lifecycle (Long SPUL and Direct Connect Lead wires)EC53:2013 (Section 5.3.5 and 5.3.7)Tests performed (implied successful by submission).
    Multi-Link Yoke and Long Lead wiresEC53:2013 (Section 5.3.5, 5.3.6 and 5.3.7)Tests performed (implied successful by submission).
    Inspection of Air Clearance60601-1:2012 and 60601-2-27:2011Tests performed (implied successful by submission).
    Defibrillation Protection and Energy Reduction60601-1Tests performed (implied successful by submission).
    Leakage Current Test60601-1:2012Tests performed (implied successful by submission).

    The document states, "Successful verification test results ensured that the proposed device does not raise any different question of safety and effectiveness." This indicates that the device met the requirements of the listed standards, and this compliance serves as the "acceptance criteria" for demonstrating substantial equivalence for this type of device (ECG cables and lead wires).

    For a device like an ECG cable and lead wire system, performance is primarily assessed through its ability to safely and effectively transmit ECG signals, which is verified through compliance with electrical safety, biocompatibility, and functional standards, rather than diagnostic accuracy metrics typically associated with AI-driven diagnostic devices.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a "test set" in the context of clinical data, as it describes a hardware device that transmits signals rather than a diagnostic algorithm. The testing described refers to engineering verification and validation of the device's physical and electrical characteristics against recognized safety and performance standards. Therefore, information about "sample size" for a clinical test set or "data provenance" (country of origin, retrospective/prospective) is not applicable here. The testing would involve a sufficient number of device units or components to demonstrate compliance with the specified standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable to the presented device and study. The "ground truth" for an ECG cable and lead wire system is its adherence to electrical and material safety standards, which is assessed through laboratory testing and engineering analysis, not through expert clinical consensus on diagnostic findings.

    4. Adjudication Method for the Test Set:

    This information is not applicable. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, often with AI systems. The studies here involve objective measurements against technical standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is an ECG cable and lead wire system, which are hardware accessories for patient monitoring. It does not incorporate AI technology, nor is it a diagnostic tool that assists human readers in interpreting results. Therefore, an MRMC comparative effectiveness study with AI assistance is not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is a hardware component; it is not an algorithm and does not operate in a standalone algorithmic capacity.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device is compliance with established international and national medical device standards for electrical safety, biocompatibility, and functional performance (e.g., "60601-2-27:2011," "EC53:2013," "AAMI ANSI ISO 10993"). This is determined through objective engineering tests and material analyses.

    8. The Sample Size for the Training Set:

    This information is not applicable. The device is an ECG cable and lead wire system, which undergoes engineering verification and validation. It is not an AI algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable, as there is no "training set" for this hardware device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153554

    Validate with FDA (Live)

    Date Cleared
    2016-03-21

    (98 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Genesis™ Low Temperature Sterilization Container System is a device intended to enclose another medical device that is to be sterilized by a health care provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed medical device until used for a maximum of 180 days.

    Containers are suitable for various STERRAD and V-PRO low temperature sterilization modalities when used as described in the instructions for use.

    Reusable baskets and accessory items (pins, dividers, mats, etc.) are intended to organize and secure enclosed medical devices during sterilization and storage of the container.

    Data cards are used to record information regarding a specific sterilization process load. Filter media allows ingress and egress of sterilant while providing a microbial barrier. Tamper evident arrows provide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains a modality-specific external process indicator that serves as a visual indication that the system has been exposed to a specific sterilization cycle parameter. Data cards, filters and tamper evident arrows are single use only.

    Device Description

    The Genesis™ Low Temperature Container System is an assortment of rigid, reusable, stackable containers that are used to enclose other medical devices and maintain sterility of these devices until used. The container system is comprised of a lid, bottom, filter, tamper evident arrows, and data cards.

    The container system houses baskets of varying depths and organizing accessory devices that are used to organize and to secure surgical instrumentation and/or other medical devices.

    AI/ML Overview

    The provided document describes the Genesis™ Low Temperature Reusable Rigid Sterilization Container System, detailing its indications for use and performance data to support its substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria / TestStandard / FDA GuidanceReported Device Performance
    Sterilization Efficacy - Lethality (STERRAD NX Standard, Advanced; STERRAD 100 NX Standard, Flex, Express; V-PRO Non-Lumen, Lumen, Flexible)ANSI AAMI ST 77 2013Testing demonstrated a 12 log reduction and a sterility assurance level (SAL) of 10^-6 using biological (BI) overkill method. This means the device effectively sterilizes the enclosed medical instruments to a high degree of confidence.
    Aerosol Challenge (STERRAD NX Advanced, STERRAD 100NX Standard, VPRO 60 Lumen)ANSI AAMI ST 77 2013Containers subjected to an aerosol challenge test with an exposure of 1.0 - 5.0 x 10^6 CFU. Agar vessels placed in each container sample demonstrated no growth of the indicator organism following the incubation period. This indicates the container effectively acts as a microbial barrier and prevents contamination.
    Cycled Aerosol / Limits of Reuse (STERRAD 100NX Standard, VPRO 1-Plus/60 Lumen)ANSI AAMI ST 77 2013After 100 simulated full use cycles or the equivalent followed by an aerosol challenge test with an exposure of 1.0 - 5.0 x 10^6 CFU, agar vessels placed in each container sample demonstrated no growth of the indicator organism following the incubation period. This demonstrates the container maintains its microbial barrier effectiveness even after repeated use.
    Simulated Use (STERRAD NX Advanced, STERRAD 100NX Flex, VPRO maX Flexible, VPRO 60 Flexible)ANSI AAMI ST 77 2013Testing demonstrated no recovery of viable challenge organism (all sterile results) from the lumens on the test device after processing within a worst-case Genesis Container as part of a worst-case cycle simulated use container validation load. This confirms the device's ability to sterilize challenging configurations, such as lumens, under realistic usage scenarios.
    Shelf Life / Package Integrity (STERRAD NX Advanced, STERRAD 100NX Standard, VPRO 60 Lumen) - For maintaining sterility for a maximum of 180 days.ANSI AAMI ST 77 2013All sample devices remained sterile (exhibit no growth) following exposure to a full sterilization cycle and weekly rotation events at a 180-day storage duration. This validates the 180-day sterility maintenance claim.
    Filter Material Properties (Bacterial Filtration Efficiency, Trapezoidal Tear Strength, and Burst Strength) before and after sterilization (STERRAD NX Advanced, STERRAD 100NX Standard, VPRO 60 Lumen)Filter properties were reportedAll filter properties were found to be acceptable. This indicates the filter media maintains its physical and functional integrity after sterilization, which is crucial for maintaining the microbial barrier.
    Biocompatibility (for accessories used in the system, processed in low-temperature sterilization modalities)ISO 10993-4 2009Blood hemolysis testing demonstrated that any accessories used in the Genesis™ Low Temperature Sterilization Container System are acceptable for use after processing in the low-temperature sterilization modalities. This confirms that the materials interacting with the medical devices after sterilization are safe for patient contact.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document provides details of the tests performed but does not explicitly state the exact sample sizes for each test set. For instance:

    • Aerosol Challenge: "Containers subjected to an aerosol challenge test with an exposure of 1.0 - 5.0 x 10^6 CFU. Agar vessels placed in each container sample..." implies multiple container samples were used, but the specific number is not given.
    • Cycled Aerosol / Limits of Reuse: "After 100 simulated full use cycles or the equivalent followed by an aerosol challenge test with an exposure of 1.0 - 5.0 x 10^6 CFU. Agar vessels placed in each container sample..." again implies multiple samples over 100 cycles, but the number of distinct items tested is not specified.
    • Simulated Use: "no recovery of viable challenge organism...from the lumens on the test device after processing within a worst case Genesis Container as part of a worst case cycle simulated use container validation load." This suggests a focus on "worst-case" scenarios, likely involving a limited number of "test devices" and container configurations, but the exact count isn't in the provided text.
    • Shelf Life / Package Integrity: "All sample devices remained sterile...following exposure to a full sterilization cycle and weekly rotation events at a 180 day storage duration" implies multiple "sample devices" but the number is not provided.
    • Biocompatibility: "Blood hemolysis testing demonstrated that any accessories used..." does not specify the number of accessories or tests performed.

    The data provenance is not specified regarding country of origin; however, the context of an FDA 510(k) submission implies testing conducted under relevant regulatory standards. The studies described are prospective in nature, as they involve performing specific tests (e.g., sterilization cycles, aerosol challenges, simulated use, shelf-life studies) to validate performance criteria for the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The document describes non-clinical performance tests, not clinical studies involving human experts for ground truth establishment. The "ground truth" for these tests is based on objective scientific measurements (e.g., presence/absence of microbial growth, sterility assurance levels, material property changes) as defined by established standards like ANSI AAMI ST 77 2013 and ISO 10993-4 2009. Therefore, there were no human experts establishing ground truth in the way one might for diagnostic imaging.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is a non-clinical device performance study based on objective measurements against defined standards, not an assessment requiring human interpretation or adjudication for ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. This document pertains to the performance validation of a sterilization container system, which is a physical medical device, not an AI or diagnostic imaging system. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable.

    6. Standalone (Algorithm Only) Performance:

    No. This is a physical medical device, not an algorithm.

    7. Type of Ground Truth Used:

    The ground truth used in these studies is objective, measurable criteria defined by recognized scientific standards and validated laboratory methods. This includes:

    • Microbiological Negativity: Absence of microbial growth (biological indicators, indicator organisms in aerosol challenge) to establish sterility.
    • Sterility Assurance Level (SAL): Quantitative measure (10^-6) of sterilization effectiveness.
    • Physical and Mechanical Integrity: Maintaining filter properties (BFE, tear strength, burst strength), container integrity.
    • Biocompatibility: Demonstrating no harmful interactions with blood (hemolysis testing).

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device undergoing performance validation, not a machine learning model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151036

    Validate with FDA (Live)

    Date Cleared
    2016-02-01

    (287 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments are designed to transmit cutting, clamping, grasping, suturing and dissecting force to working tips in minimally invasive surgeries such as plastic and general laparoscopic procedures.

    Device Description

    Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments consists of a handle, shaft and insert. The insert contains an end-effector (jaw pattern). There are several device models that encompass various lengths, diameters and jaw patterns based on the surgeon's needs.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments:

    Based on the provided 510(k) Summary (K151036), this document describes a medical device modification (design change from single-piece to take-apart) rather than an AI/algorithm-based product. Therefore, many of the requested points related to AI/algorithm studies (like human-in-the-loop performance, training sets, etc.) are not applicable to this submission.

    The acceptance criteria and supporting study are focused on ensuring the modified physical device maintains its safety and effectiveness compared to its predicate.


    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard / Test / FDA Guidance)Reported Device Performance
    Instrument strength (resistance to breakage/flexing)Strength TestsPASS
    Inadvertent disassembly (during use)Pull Strength TestPASS
    Ease of disassembly/reassembly (user convenience)Force TestsPASS
    Handle reusabilityReliability TestPASS
    Handle compatibility with existing shafts and insertsHandle ability to connect to worst case shafts and insertsPASS
    Device cleanability and sterilizabilityAAMI TIR12, AAMI TIR30, ANSI AAMI ST79, ANSI AAMI ST81, ISO 11138, ISO 17664, ISO 17665PASS
    Biocompatibility of device materialsISO 10993PASS

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify the exact sample sizes (e.g., number of instruments tested) for each non-clinical performance test. It only states that the tests were conducted and the device "PASSED" the criteria.

    Data provenance: Not explicitly stated, but these are typically retrospective non-clinical engineering and laboratory tests conducted by the manufacturer, CareFusion 2200 Inc. The country of origin of the data would likely be the location where CareFusion performs its product testing, which can be global, but typically for FDA submissions, the testing adheres to international (ISO) and US (AAMI, ANSI) standards.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This question is not applicable. The "ground truth" for physical device performance tests like strength, pull, force, reliability, cleaning validation, and biocompatibility is established by engineering specifications, recognized national and international standards (e.g., AAMI, ANSI, ISO), and regulatory requirements, not by expert medical opinion in the same way an AI model's diagnostic accuracy might be judged. These tests are objective measurements against defined thresholds.


    4. Adjudication Method for the Test Set

    This question is not applicable. Adjudication methods (like 2+1, 3+1) are typically used for subjective assessments or when there's ambiguity in classifying clinical data for ground truth establishment. For the objective non-clinical performance tests mentioned, the results are typically quantitative measurements that either meet or fail pre-defined acceptance limits based on standards and engineering requirements.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "N/A - No clinical tests were conducted for this submission." MRMC studies are typically for evaluating diagnostic accuracy or clinical effectiveness, often involving human readers and AI assistance, neither of which is present in this submission.


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone study (referring to an algorithm's performance) was not done. This submission is for a physical surgical instrument, not an AI or algorithm-based device.


    7. Type of Ground Truth Used

    The "ground truth" for the performance tests undertaken are:

    • Engineering Specifications and Design Requirements: For parameters like instrument strength, inadvertent disassembly, ease of disassembly/reassembly, handle reusability, and handle compatibility. These are defined by the manufacturer based on expected operational loads and user interaction.
    • National and International Standards: For aspects like cleanability and sterilizability (e.g., AAMI TIR12, AAMI TIR30, ANSI AAMI ST79, ANSI AAMI ST81, ISO 11138, ISO 17664, ISO 17665) and biocompatibility (ISO 10993). These standards provide objective criteria and methodologies for testing.

    8. Sample Size for the Training Set

    This question is not applicable. There is no AI/algorithm, so there is no training set for an algorithm.


    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no AI/algorithm and thus no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142529

    Validate with FDA (Live)

    Date Cleared
    2015-01-08

    (121 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Genesis Reusable Rigid Sterilization Container System is a device intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It allows sterilization of the enclosed medical device and maintains sterility of the enclosed device until used for a maximum of 180 days.

    Containers are suitable for dynamic air re-vacuum) steam sterilization, immediate use pre-vacuum steam sterilization and 100% ethylene oxide sterilization when used as described in the instructions for use.

    Reusable baskets and accessory items (pins, dividers, mats, etc.) are intended to organize and secure enclosed medical devices during sterilization and storage of the container.

    Data cards are used to recording a specific sterilization process load. Filter media allows ingress and egress of sterilant while providing a microbial barrier. Tamper evide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains a modality-specific external process indication indication that the system has been exposed to a specific sterilization cycle parameter. Data cards, filters and tamper evident arrows are single use only.

    Device Description

    The Genesis Reusable Rigid Container System is an assortment of rigid, reusable, stackable containers that are used to sterilize other medical devices and maintain sterility of these devices until used. The container system is comprised of a lid, bottom, filter, tamper evident arrows, and data cards. The container system houses baskets of varying depths and organizing accessory devices that are used to organize and to secure surgical instrumentation and/or other medical devices.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (Genesis™ Reusable Rigid Sterilization Container System). It outlines the device's indications for use, comparison to a predicate device, and a summary of non-clinical performance tests.

    Here's an analysis of the acceptance criteria and study information provided, focusing on the available details:


    Description of Acceptance Criteria and Device Performance

    The device under review is the Genesis™ Reusable Rigid Sterilization Container System. The primary function of this system is to enclose other medical devices for sterilization and maintain their sterility for up to 180 days. The acceptance criteria for this device revolve around demonstrating its efficacy in sterilization and maintaining sterility under various conditions.

    The study presented focuses on non-clinical performance testing to validate the sterilization efficacy and dry time of the container system for specific sterilization modalities and contents.

    1. Table of Acceptance Criteria and the Reported Device Performance:

    Characteristic / Acceptance CriteriaReported Device Performance
    Sterilization Efficacy: Pre-vacuum Steam
    - Standard/Test/FDA Guidance: ST 77 2013 Containment Devices for Reusable Medical Device SterilizationTesting demonstrated a 12 log reduction and a sterility assurance level (SAL) of 10-6 using biological (BI) overkill method.
    Dry Time for Dynamic Air Removal (Pre-vacuum steam modality)
    - Standard/Test/FDA Guidance: ST 77 2013 Containment Devices for Reusable Medical Device SterilizationTesting demonstrated validated method of drying by absence of visible moisture.

    Additional Performance Parameters and Conditions (from "Sterilization Parameters for the Genesis™ Reusable Rigid Container System by Modality" and "Accessories and Device Challenges by Sterilization Modality and Container Style" tables):

    Sterilization MethodCycle Parameters / Contents / ConfigurationsApplicable Containers/Accessories / Performance
    Dynamic Air Removal (Pre-Vacuum) Steam (Regular & Immediate Use)- Exposure Temp: 270°F (132°C)- Pre-conditioning Pulses: 3- Exposure Time: 4 Minutes- Dry Time Cycle: 30 minutes (Regular)- Minimum Cool Time: 60 minutes (Regular)- Stack Height: Max 3- Total System Weight: 25 lbs.- Lumen Devices: 2.68 mm x 450 mm (Qty 16), 1.37 mm x 242 mm (Qty 10), 3 mm x 400 mm (Qty 2)Demonstrated efficacy for Solid and Perforated Bottom Containers, Stacking/Non-Stacking Baskets, various Lumen Devices (specified quantities), Devices with conjoined surfaces, Silicone Support Bars/Mats, Laparoscopic Racks, Optional Protective Plates. Materials include Metals, Polymers, Composites. (Performance is "Yes" for all listed components/configurations under these steam methods).
    100% Ethylene Oxide- EO Sterilant Conc: 100% EO, 725 mg/L- Pre-Conditioning Time: 30 mins- Temperature: 130°F (55°C)- Exposure Time: 60 mins- Relative Humidity: 50-80%- Aeration: 8 hours @ 109.4°F (43°C)- No stacking- Total System Weight: 15 lbs.- Lumen Devices: 3.0 mm x 400 mm (Qty 5)Demonstrated efficacy for Perforated Bottom Containers (except CD2-10BDL and DINCD2-8B), Non-Stacking Baskets, Lumen Devices (specified quantity), Devices with conjoined surfaces, Silicone Support Bars, Laparoscopic Racks. Materials include Stainless Steel, Aluminum, Silicone, Radel. (Performance is "Yes" for listed components/configurations under EO).

    The document explicitly states: "The Genesis Reusable Rigid Container System has been validated to meet the established performance criteria." This indicates that for all the criteria mentioned in the tables, the device successfully demonstrated compliance through the stated non-clinical tests.

    2. Sample size(s) used for the test set and the data provenance:

    • The document mentions "Testing" for sterilization efficacy and dry time but does not specify the exact sample sizes (number of containers, number of cycles, number of biological indicators, etc.) for the test sets.
    • Data provenance is not explicitly stated regarding country of origin or whether the tests were retrospective or prospective. Given it's a 510(k) submission for a medical device by a US-based company (CareFusion 2200 Inc, Vernon Hills, IL), and referring to an FDA guidance document (ST 77 2013), it is prospective testing conducted in a controlled laboratory setting, likely within the US.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This document describes non-clinical performance testing of a sterilization container, not a diagnostic or AI-driven medical imaging device. Therefore, the concept of "experts establishing ground truth for a test set" in the context of clinical interpretation (e.g., radiologists) is not applicable.
    • The "ground truth" for sterilization efficacy is established by scientific principles and validated biological indicators (BI) and sterility assurance levels (SAL) as per recognized standards (ST 77 2013). The validation of dry time is based on the "absence of visible moisture." These are objective measurements rather than subjective expert consensus.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • As this is non-clinical performance testing, adjudication methods typically used for subjective clinical assessments are not applicable. The results are based on objective laboratory measurements and adherence to specified performance standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, this is not applicable. The document is for a physical medical device (sterilization container), not an AI algorithm or a diagnostic tool involving human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, this is not applicable. This document pertains to a physical sterilization container.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the performance of the sterilization container is based on objective measurements of sterilization efficacy and dryness as defined by recognized industry standards (ST 77 2013).
      • For sterilization efficacy: achieving a 12 log reduction and a Sterility Assurance Level (SAL) of 10^-6 using biological indicators (overkill method).
      • For dry time: absence of visible moisture.

    8. The sample size for the training set:

    • Not applicable. This relates to a physical device evaluated through non-clinical performance testing, not an AI model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As indicated in point 8, there is no training set for this type of device evaluation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140761

    Validate with FDA (Live)

    Date Cleared
    2014-04-21

    (26 days)

    Product Code
    Regulation Number
    884.1300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kahn Uterine Trigger Cannula, Jarcho Self-Retaining Uterine Cannula, and Neal Fallopian Cannula are indicated to test the patency (lack of obstruction) of the fallopian tubes by pressurizing the uterus and filling them with carbon dioxide gas.

    Device Description

    The V. Mueller® Uterine Cannula devices are constructed with a stopcock, finger ring control, an elongate shaft and a distal end with a rubber acorn.

    AI/ML Overview

    The provided text describes the 510(k) summary for Uterine Cannulas, focusing on demonstrating substantial equivalence to a predicate device due to a "material modification". This submission is for a medical device, not an AI/ML powered device, so many of the requested categories (e.g., sample size for test/training sets, number of experts, MRMC studies, standalone algorithm performance) are not applicable.

    However, I can extract the relevant acceptance criteria and study information provided for this medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard/Test/FDA Guidance)Reported Device Performance (Results Summary)
    BiocompatibilityISO 10993-1:2009 / AC 2010PASS
    FunctionalityPlating Thickness AnalysisPASS

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. This submission concerns a physical medical device (Uterine Cannula) and its material modification, not an AI/ML algorithm requiring a test set of data. The performance evaluations are based on non-clinical tests (biocompatibility and functionality).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth as typically understood for AI/ML device evaluations (e.g., expert consensus on medical images) is not relevant here. The "ground truth" for this device's performance is established by adherence to recognized standards for biocompatibility and engineering analysis for functionality.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is not an AI/ML device requiring adjudication of results from a test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is a physical medical device, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device does not involve an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is based on established international standards for biocompatibility (ISO 10993-1:2009 / AC 2010) and engineering analysis (Plating Thickness Analysis) for functionality.

    8. The Sample Size for the Training Set

    Not applicable. As this is not an AI/ML device, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1