510(k) Data Aggregation

    K Number
    K250723

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-25

    (46 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreedomFlow™ Orbital Circumferential Atherectomy System is indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries. The therapy is intended for patients who are acceptable candidates for percutaneous transluminal atherectomy.

    Device Description

    The FreedomFlow™ Orbital Circumferential Atherectomy System is a flexible over-the-wire rotational device used to ablate atherosclerotic plaque from arterial blood vessels within the body. The FreedomFlow™ Orbital Circumferential Atherectomy System is used together with a compatible introducer sheath and 0.014-inch diameter x 300 cm (minimum length) atherectomy guidewire. The driveshaft is introduced into the patient's vasculature by traditional minimally invasive techniques. The FreedomFlow™ User Handle is available in model numbers that are listed below with vessel size ranges.

    The FreedomFlow Orbital Circumferential Atherectomy System - Electric (FreedomFlow™) includes an integrated driveshaft with multiple abrasive spheres on a rotating driveshaft. The abrasive spheres are eccentrically mounted onto the driveshaft so that when the driveshaft is rotated, they move outward due to centrifugal force. These abrasive spheres are spaced along the driveshaft in a spiral configuration to optimize plaque modification within a vessel while still maintaining flexibility for treating tortuous arterial anatomy. The User Handle includes two rotational speeds: low speed at 50,000 revolutions per minute (RPM) and high speed at 76,000 RPM.

    The FreedomFlow Orbital Circumferential Atherectomy System - Electric (FreedomFlow™) is powered by Cardio Flow Power Supply H7001, which is a hospital-grade portable, reusable component. H7001 provides DC power to rotate the FreedomFlow™ driveshaft. H7001 also provides DC power to a saline pump integrated into the FreedomFlow™ User Handle. During operation the saline pump delivers saline to the distal tip of the driveshaft.

    The FreedomFlow™ orbital atherectomy User Handle is supplied single patient use, sterile. The package contents include the following items.

    • FreedomFlow™ orbital atherectomy User Handle with integrated electric motor and saline pump
    • Saline infusion tubing set
    AI/ML Overview

    I'm sorry, but your request cannot be fulfilled. The provided document is an FDA 510(k) clearance letter for a medical device (FreedomFlow™ Orbital Circumferential Atherectomy System), not for an AI/ML-driven medical device.

    The document describes the physical characteristics of an atherectomy system, its intended use, and bench testing to demonstrate its mechanical and performance characteristics (like plaque removal efficiency, tensile strength, rotational speed, etc.). It does not involve any AI/ML components, nor does it describe a study involving human readers, ground truth establishment by experts, or MRMC studies, which are typical for validating AI/ML-driven medical devices.

    Therefore, I cannot extract the information required to describe acceptance criteria and a study proving an AI/ML device meets those criteria from this non-AI/ML medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233483

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-15

    (111 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FreedomFlow™ Orbital Circumferential Atherectomy System is indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries. The therapy is intended for patients who are acceptable candidates for percutaneous transluminal atherectomy.

    Device Description

    The FreedomFlow™ Orbital Circumferential Atherectomy System (H6004 5Fr 3-Sphere) is a flexible over-the-wire rotational device used to ablate atherosclerotic plaque from peripheral arterial blood vessels within the body. The FreedomFlow™ System consists of three components: User Handle with integrated driveshaft, Tubing Set, and Power Supply. The User Handle with the integrated driveshaft is sterile, single use, and disposable. The User Handle provides the operator interface to control driveshaft rotation (with two speed options) and translation within the vessel. The User Handle also incorporates a guidewire mechanical clamp. The User Handle controls the saline fluid flow down the driveshaft catheter. The User Handle utilizes firmware and hardware to perform these functions. The Tubing Set is sterile, single use, and disposable. The Tubing Set connects a user provided sterile saline supply to the User Handle. The Tubing Set is connected by the operator to the User Handle. The Power Supply is a medical grade AC powered transformer, supplying DC electrical energy to the User Handle. The power supply is reusable.

    AI/ML Overview

    The provided text describes a medical device, the FreedomFlow™ Orbital Circumferential Atherectomy System (H6004 5Fr 3-Sphere), and its substantial equivalence determination by the FDA. However, the document does not contain the level of detail requested regarding acceptance criteria and the specifics of a study proving a device meets these criteria in the context of an AI/human-in-the-loop system.

    The "study" mentioned in this document refers to performance testing for a mechanical medical device, specifically:

    • Dimensional Analysis Testing
    • Orbit Characterization Testing (effectiveness of sanding technology, lumen diameter vs. speed vs. time, plaque removal efficiency)
    • Simulated Life Test
    • Torque Stall and Loaded Start Test
    • Joint Tensile Test
    • Joint Torque Test
    • Particulate Analysis Comparison Test
    • Electric System Human Cadaver Pre-Clinical Test

    There is no indication that this device incorporates AI or sophisticated software that would necessitate human-in-the-loop performance studies, expert ground truth adjudication, or metrics like sensitivity, specificity, or AUC, which are typically associated with AI-powered diagnostic or assistive devices.

    Therefore, I cannot provide the requested information in the format of the questions because the document describes a mechanical atherectomy system, not a device whose performance is measured in the ways implied by the prompt's questions (e.g., AI assistance, expert readers, ground truth establishment for AI training/testing).

    If the prompt intended for me to infer or create a hypothetical scenario for an AI-powered atherectomy system, I would need additional context or a different document. Based only on the provided text, the questions are not applicable to the described device and its testing.