Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K252480

    Validate with FDA (Live)

    Device Name
    CE Deliver (DLV)
    Manufacturer
    Date Cleared
    2025-12-19

    (134 days)

    Product Code
    Regulation Number
    876.1300
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K192662

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-14

    (142 days)

    Product Code
    Regulation Number
    876.1300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CapsoCam ® Plus video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

    Device Description

    CapsoCam® Plus (SV-3) capsule is a single-use. ingestible video capsule that acquires and stores video images in on-board memory while moving through the gastrointestinal tract, propelled by natural peristalsis. The patient retrieves the provided retrieval kit and returns it to the physician who downloads and reviews the images on a computer. The capsule is typically excreted within 3 to 30 hours after swallowing.

    CapsoCam® Plus (SV-3) capsule endoscope is a single-use ingestible capsule system for diagnostic visualization of the adult small bowel. The system consists of the following accessories:

    • CapsoCam® Plus (SV-3) capsules, which are intended for visualization of the small bowel mucosa in adults and used as a tool in the detection of abnormalities of the small bowel.
    • CapsoView (CVV) software program, which is used to download and view CapsoCam® Plus ● (SV-3) capsule images and to generate capsule endoscopy reports.
    • CapsoAccess (CDAS) Capsule Data Access System, which accesses data from the ● CapsoCam® Plus (SV-3) capsules.
    • CapsoRetrieve (CVR1) Capsule Retrieval Kit, which is used for the collection, storage, and ● transportation of the excreted CapsoCam® Plus (SV-3) capsules.
    • CapsoCloud (CLD), which is a cloud-based software application used to manage procedures ● and to review and analyze images from the CapsoCam® Plus (SV-3) capsule.
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the CapsoCam® Plus (SV-3) Capsule Endoscopy System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" with numerical targets for clinical performance. Instead, it focuses on demonstrating substantial equivalence to a predicate device (CapsoCam® Plus (SV-3) System, K183192).

    The key performance aspect tested for the new CapsoCloud software application was image comparison and Image Substantial Equivalence Analysis between the proposed and predicate software.

    Acceptance Criterion (Implied)Reported Device Performance
    Image Substantial EquivalenceDemonstrated substantial equivalence through testing of the CapsoCloud software application and in-vivo image comparison.
    Safety and EffectivenessThe material and technological characteristics do not raise different questions of safety or effectiveness compared to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "in-vivo testing for the image comparison from the proposed and the predicate software."

    • Sample Size: Not explicitly stated.
    • Data Provenance: "in-vivo testing" suggests data from live subjects, but details like country of origin or whether it was retrospective or prospective are not provided.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document.

    4. Adjudication Method

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not conducted or reported in this document. The focus was on demonstrating substantial equivalence to a predicate device, specifically regarding the addition of a new cloud-based software. There is no mention of human reader performance improvement with or without AI assistance.

    6. Standalone (Algorithm Only) Performance

    A standalone performance study focused purely on the algorithm's diagnostic capabilities without human interaction was not explicitly described for this submission. The validation focused on the CapsoCloud software and image equivalence to the predicate system, which implies the system as a whole maintains its performance characteristics when the cloud software is introduced.

    7. Type of Ground Truth Used

    The ground truth for the "image comparison from the proposed and the predicate software" would be rooted in the images generated by the predicate device's software. The goal was to show that the new software (CapsoCloud) processes and presents images that are substantially equivalent to those processed by the predicate's CapsoView software. The document doesn't detail how the ground truth for actual abnormalities or diagnoses in these images was established, as the primary comparison was between software versions, not necessarily clinical diagnostic accuracy.

    8. Sample Size for the Training Set

    This information is not provided in the document.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183192

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-04-19

    (151 days)

    Product Code
    Regulation Number
    876.1300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CapsoCam® Plus video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

    Device Description

    CapsoCam Plus (SV-3) capsule is a single-use, ingestible video capsule that acquires and stores video images in on-board memory while moving through the gastrointestinal tract, propelled by natural peristalsis. The patient retrieves the capsule using the provided retrieval kit and returns it to the physician who downloads and reviews the images on a computer. The capsule is typically excreted within 3 to 30 hours after swallowing.

    CapsoCam Plus (SV-3) capsule endoscope is a single-use ingestible capsule system for diagnostic visualization of the adult small bowel. The overall system consists of an ingestible Capsule CapsoCam (SV-3), the CapsoRetrieve® (CVR1) Capsule Retrieval Kit, the CapsoAccess® Capsule Data Access System (CDAS) and the CapsoView® (CVV) software. The capsule contains a panoramic color digital video camera, two silver-oxide watch batteries, white-LED light sources, a laser diode for data download and system-control and nonvolatile flash-memory data-storage electronics.

    AI/ML Overview

    The provided text is a 510(k) summary for the CapsoCam Plus (SV-3) video capsule system. This document outlines the device's indications for use, its description, and its substantial equivalence to a previously cleared predicate device. It also briefly mentions non-clinical and clinical testing performed.

    However, the document does not contain specific acceptance criteria, study details demonstrating performance against criteria, sample sizes for test or training sets, ground truth establishment methods, expert qualifications, adjudication methods, or information about AI assistance or standalone algorithm performance.

    The clinical summary states that the testing demonstrated:

    • Images are captured throughout an entire small bowel exam, with complete small bowel exam defined as the ability of the capsule to reach the cecum while still recording images.
    • Images are correctly downloaded using the CapsoAccess® capsule data access system (CDAS3).
    • Small bowel images of diagnostic quality are able to be reviewed using the CapsoView® CVV Software and determined to be effective for diagnostic purposes by the study investigator.

    It concludes that "the results of this clinical testing demonstrate that it meets the primary endpoints per the study protocol and the performance of the CapsoCam Plus (SV-3) Capsule Endoscope System does not raise new issues of safety or effectiveness when compared to the predicate device (K161773)."

    The purpose of this 510(k) submission appears to be related to minor changes or reaffirmation of substantial equivalence to an already cleared device (K161773). Therefore, the detailed performance study information, including specific acceptance criteria and quantitative results, would have been part of the original K161773 submission, not this K183192 summary.

    Therefore, based solely on the provided text, I cannot provide the requested information in detail because it is not present. The document focuses on showing substantial equivalence rather than detailing a de novo performance study with specific metrics and acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151635

    Validate with FDA (Live)

    Device Name
    CapsoCam (SV-1)
    Manufacturer
    Date Cleared
    2016-02-09

    (237 days)

    Product Code
    Regulation Number
    876.1300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CapsoCam SV-1 video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

    Device Description

    CapsoCam® SV-1 is a single-use, ingestible video capsule that acquires and stores video images in on-board memory while moving through the gastrointestinal tract, propelled by natural peristalsis. The patient retrieves the capsule using the provided retrieval kit, and returns it to the physician who downloads and reviews the images on a personal computer (PC). The capsule is typically excreted within 3 to 30 hours after swallowing.

    The CapsoCam® SV-1 video capsule is a panoramic imaging system which includes a panoramic lens system and a multi-array CMOS image sensor. The CapsoCam® has four side-facing cameras arrayed about the circumference of the capsule that together image a full 360° circumference and capture high-resolution color images of the mucosa.

    The CapsoCam® SV-1 acquires video images while moving through the patient's gastrointestinal tract. The frame rate is 5 frames per second (FPS) for each of 4 cameras (20 FPS total) the first two hours, and then decreases to 3 FPS per camera (12 FPS total). The initial frame rate is higher because the capsule moves more quickly in the duodenum than in the jejunum or ileum.

    After excretion, the capsule is retrieved by the patient using CapsoRetrive, a capsule retrieval system, including a strainer which is placed in the patient's toilet to retrieve the capsule while passing fecal matter and rinse water. The patient then returns the capsule to the clinic, where a technician disinfects the capsule and retrieves the data using the CapsoAccess, Capsule Data Access System (CDAS). The CDAS is comprised of a capsule carrier for holding the capsule opener for cutting off one end of the capsule, and a docking system for forming an electrical contact to the capsule printed circuit board (PCB) through the open end. The data is transferred to the workstation over the link shown schematically in Figure 4 below, from the capsule to the docking system by a serial peripheral interface (SPI) and to the workstation by a universal serial bus (USB).

    The reviewing workstation is a personal computer (PC) with the Capso Vieweimage review software installed. The CapsoView@(CVV) software displays the video and employs a variety of image-enhancement, video navigation, and image analysis features to facilitate the physician's diagnosis. The software allows the physician to efficiently annotate individual frames from the video and compile a procedure report.

    AI/ML Overview

    1. Acceptance Criteria and Device Performance:

    The document outlines acceptance criteria based on two co-primary endpoints and several secondary endpoints.

    MetricAcceptance Criteria (Lower 95% Confidence Limit)Reported Device Performance (CapsoCam SV-1)
    Co-Primary Endpoint 1: Completed Exams (vs. PillCam SB2)Not less than -15% difference in proportion-0.028% (meaning CapsoCam SV-1 was non-inferior)
    Co-Primary Endpoint 2: Positive Percent Agreement (PPA)Not explicitly stated as a numerical threshold, but implied to demonstrate similar performance to predicate.69.05% (53.97%, 80.93%)
    Co-Primary Endpoint 2: Negative Percent Agreement (NPA)Not explicitly stated as a numerical threshold, but implied to demonstrate similar performance to predicate.81.94% (71.52%, 89.13%)
    Secondary Endpoints (Primary Diagnosis Concordance by Type)Implied to demonstrate similar performance to predicate, confidence intervals provided for each category.Normal: 81.43% (70.77, 88.81)
    Vascular: 66.67% (45.37, 82.81)
    Mass/Polyp: 0.00% (0.00, 48.99)
    Blood: 33.33% (6.15, 79.23)
    Ulcerative: 64.29% (38.76, 83.66)
    Other: 50.00% (9.45, 90.55)
    Battery LifeNot explicitly stated as an acceptance criterion in the provided text, but data gathered.Mean: 16.5 Hrs, Median: 16 Hrs, Standard Deviation: 2.39 Hrs

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 121 subjects (age 18 to 85)
    • Data Provenance:
      • Country of origin: USA (Multi Center Pivotal Study)
      • Retrospective or Prospective: Prospective (clinical trial was performed "on the CapsoCam® Capsule Endoscope System per Protocol Number SV-1- Comparison of Capso Vision SV-1 to PillCam® SB2")

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Number of Experts: Three (3) independent readers per set of images.
    • Qualifications of Experts: The document states "independent readers, who were blinded to the subject clinical information or suspected diagnosis." Specific medical qualifications (e.g., "radiologist with 10 years of experience") are not provided in the submitted text. They are likely gastroenterologists or other medical professionals specializing in capsule endoscopy review.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: If any of the three (3) readers in a group disagreed with the Primary Diagnosis or complete exam, the findings were discussed by members of the group until a 2/3 majority or consensus agreement for Primary Diagnosis was reached.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Yes, a comparative effectiveness study was done. The study compared the CapsoCam® SV-1 to the predicate device, PillCam® SB2.
    • Effect Size of Human Readers' Improvement with AI vs. Without AI Assistance:
      • The provided text does not contain information on human readers improving with AI assistance vs. without AI assistance. The study design involved humans reading images from two different capsule devices (CapsoCam SV-1 and PillCam SB2) in a blinded fashion. It does not describe an AI assistance component for human readers. This was a direct comparison between two devices through human interpretation.

    6. Standalone (Algorithm Only) Performance:

    • No, a standalone (algorithm only) performance study was not described. The study evaluates the "CapsoCam® Capsule Endoscope System," which includes the capsule and the CapsoView® software. Performance metrics are based on the interpretations of human readers utilizing the device's output, making it a human-in-the-loop assessment rather than a standalone algorithm evaluation. The software primarily focuses on image display, enhancement, navigation, and report generation for human review.

    7. Type of Ground Truth Used:

    • Type of Ground Truth: Expert Consensus. The ground truth ("Primary Diagnosis" and "completed exam" status) for each case was established by a 2/3 majority or consensus agreement among a group of three independent readers comparing the findings from both CapsoCam and PillCam. The document refers to "the most clinically significant Primary Diagnosis."

    8. Sample Size for the Training Set:

    • Not explicitly provided. The document describes a "Multi Center Pivotal Study" (test set) but does not detail the sample size or specific methodology for a separate training set. The descriptions of training sets are typically associated with machine learning models, and this document focuses on clinical performance comparison of a diagnostic device through human review.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable / Not provided. As the document does not describe a distinct training set for a machine learning model, the method for establishing its ground truth is not mentioned. The document primarily focuses on the validation study (test set) for regulatory approval.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1