510(k) Data Aggregation

    K Number
    K173590

    Validate with FDA (Live)

    Date Cleared
    2020-03-30

    (860 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    System 83 Plus™ Washer/Disinfector is designed for the simultaneous high level disinfection of up to two flexible submersible endoscopes that are used in the gastrointestinal and/or pulmonary tracts. Flexible endoscopes that are precleaned and manually cleaned, and then exposed to the washing /disinfection cycle of the System 83 Plu s™, may be high level disinfected when the disinfection cycle corresponds to the System 83 Plus™ validated contact conditions of the high level disinfectant.

    Note: The System 83 Plus™ device includes two models.

    · The System 83 Plus™ 2 has one processing chamber that can process 1 to 2 flexible endoscopes at a time .

    · The System 83 Plus™ 9 has two processing chambers that can process 1 to 2 flexible endoscopes at a time in each independently operated processing chamber for a total of up to 4 endoscopes simultaneously.

    Device Description

    The System 83 Plus™ is an automated, computer controlled, electro-mechanical system intended to wash and high-level disinfect one or two submersible flexible endoscopes (per bay) utilizing a detergent and FDA cleared high-level disinfectant validated by CUI. The System 83 Plus™ device family includes two models. The System 83 Plus™ 2; a model with one processing chamber, which can process one to two flexible endoscopes at a time. The System 83 Plus™ 9: a model with two processing chambers which can asynchronously reprocess up to 2 endoscopes per bay (total of up to 4 endoscopes).

    The System 83 Plus™ utilizes a processing chamber and the immersion method to perform washing, disinfection, rinsing, and alcohol flush of an endoscope to render a high level disinfected endoscope. The System 83 Plus™ is capable of automated detergent dispensing and transfer of disinfectant solution between the reservoir and processing chamber. Following disinfection, the endoscopes and channels are automatically rinsed with potable water that is filtered through a water filtration system that contains a 0.1 micron bacteria filter and the channels are then flushed with air. A semi-automated air/alcohol flush must be completed at the end of the cycle. At completion, the operator prints out the endoscope reprocessing information with a printer.

    Built-in sensors detect fluid levels, fluid temperature, fluid flow, and the operating states of the components within System 83 Plus™.

    AI/ML Overview

    The provided text is a compilation of FDA documents related to the 510(k) premarket notification for the "System 83 Plus Washer/Disinfector." It describes the device, its intended use, comparison to a predicate device, and a summary of non-clinical testing performed to demonstrate its safety and effectiveness.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance TestingAcceptance CriteriaReported Device Performance
    Simulated use testing (High-level disinfection)≥6 Log reduction of M.terrae at all inoculated sitesPass
    In-use testing (High-level disinfection)<1 CFU at all processed test sitesPass
    Alcohol and detergent line disinfection≥6 Log reduction of M.terraePass
    Toxicological evaluation of residues and rinsing validationReactivity grade of 2 or lessPass
    Channel volume flushingSatisfy endoscope manufacturer's manual flushing requirementsPass
    Water filtration system validation<10 CFU per 100 mLPass
    In-line Disc filter validation≥99% efficient at removing particles ≥250µmPass
    Electrical safety testingUL 60601-1:2003 standard for safetyComplies
    Electromagnetic compatibility testing (EMC)IEC 60601-1-2:2001 + A1:2004 standard for EMCComplies
    Software verification and validation testingNot explicitly numeric, but implies meeting FDA guidance for software in medical devicesConducted and documentation provided as recommended by FDA guidance

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific numerical sample sizes (e.g., number of endoscopes, number of runs) for each performance test. It uses terms like "representative worst case endoscopes" and "worst case simulated use conditions" for several tests.

    • Sample Size: Not explicitly stated as a number for individual tests.
    • Data Provenance: The document implies the tests were conducted by or on behalf of Custom Ultrasonics, Inc. (the submitter) as part of their 510(k) submission. There is no information about the country of origin of the data collectors or whether the studies were retrospective or prospective, though 510(k) testing is typically prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    The document does not provide any information about:

    • The number of experts used.
    • The qualifications of those experts.
    • How ground truth was established for the performance tests.

    For this type of device (washer/disinfector), "ground truth" for disinfection performance would typically be established through microbiological assays (e.g., plating, colony counting) to determine bacterial reduction, rather than expert human interpretation in the sense of a medical image analysis.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method. Given the nature of the testing (microbiological, chemical, engineering performance), adjudication by multiple human experts is unlikely to be relevant in the same way it would be for, say, a diagnostic AI device analyzing medical images. The acceptance criteria are objective and measurable (e.g., log reduction, CFU count).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI devices that assist human readers in interpreting medical images or data. The System 83 Plus Washer/Disinfector is an automated reprocessing device, not a diagnostic AI system involving human interpretation of complex medical cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The performance testing described ("Simulated use testing," "In-use testing," etc.) represents a "standalone" or "device-only" performance evaluation. The device itself performs the high-level disinfection process without human intervention during the active cycle. The tests evaluate the device's ability to meet the specified disinfection and other performance criteria independently.

    7. The Type of Ground Truth Used

    The ground truth for the performance testing is based on objective, quantitative measurements related to disinfection efficacy and device performance. Specifically:

    • Microbiological assays: For disinfection testing (simulated use, in-use, alcohol/detergent line), the ground truth is established by measuring the reduction of microbial load (e.g., M. terrae counts, CFU counts).
    • Chemical analysis: For toxicological evaluation of residues, ground truth is based on reactivity grades consistent with ISO standards.
    • Physical measurements/compliance: For channel volume flushing, water filtration, and disc filter efficiency, ground truth is based on meeting specified flow rates, filtration efficiency percentages, or industry standards.
    • Engineering standards compliance: For electrical safety and EMC, ground truth is compliance with recognized industry standards (UL 60601-1, IEC 60601-1-2).

    8. The Sample Size for the Training Set

    The document does not mention a training set. This is because the System 83 Plus Washer/Disinfector is an electro-mechanical device with software logic, not an AI/machine learning device that requires a "training set" in the conventional sense (i.e., for learning from data to make predictions). The software verification and validation refer to traditional software development and testing, not AI model training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set for an AI model, this question is not applicable. The device's "ground truth" is based on established engineering principles, microbiological science, and regulatory standards for medical device performance and safety.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162120

    Validate with FDA (Live)

    Date Cleared
    2017-04-28

    (270 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scope-Assist Flushing Sink is indicated for flushing the channels of flexible endoscopes according to the endoscope manufacturer's instructions.

    Device Description

    The Scope-Assist Flushing Sink is designed to mechanically assist the channel flushing portion of an endoscope's manual cleaning process. The sink facilitates compliance with endoscope manufacturer's cleaning requirements by:
    • Providing a means of flushing flexible endoscope channels with detergent solution
    • Providing a means of flushing endoscope channels with water
    • Providing an environment for soaking and manual wiping/brushing of flexible endoscope exterior surfaces and accessories
    • Providing an environment to perform wet leak testing of endoscopes
    The Scope-Assist Flushing Sink is designed:
    • To be used where the facility performs its endoscope cleaning functions, such as in soiled utility rooms, central processing decontamination areas, or endoscope cleaning (reprocessing) areas.
    • To be used with Tergal 800 Detergent
    The Scope-Assist Flushing Sink is not designed:
    ● To be used at bedside, in critical care units, or in sterile environments
    ● To be used as a terminal process or as part of the manual high level disinfection process

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "Scope-Assist Flushing Sink" device. It is a device intended to mechanically assist the channel flushing portion of an endoscope's manual cleaning process. The information provided is for regulatory approval and does not contain a study of the type you're asking for (e.g., related to AI or diagnostic accuracy improvements). Therefore, many of your requested items regarding acceptance criteria and study details cannot be answered based on this document.

    Here's what can be extracted and what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Stated Goal)Reported Device Performance (Summary from document)
    Flushing performance meets or exceeds endoscope manufacturer's requirements in specified flushing times.Device is able to meet or exceed the endoscope manufacturer's flushing requirements in the specified flushing times. (Specific quantitative results not provided)
    Detergent injection delivers correct amount of detergent consistently and reliably.Device is able to consistently and reliably deliver the correct amount of detergent. (Specific quantitative results not provided)
    Tergal detergent is compatible with materials in the flow path of the sink.Use of Tergal detergent is compatible with the materials in the flow path of the sink. (Specific test results not provided)
    Electrical safety according to IEC 61010-1.Electrical safety testing according to IEC 61010-1 was conducted and demonstrated substantial equivalence. (Specific parameters/results not provided)
    Electromagnetic compatibility according to IEC 60601-1-2.Electromagnetic compatibility testing according to IEC 60601-1-2 was conducted and demonstrated substantial equivalence. (Specific parameters/results not provided)
    Amount of fluid delivered is sufficient to meet or exceed endoscope manufacturer's requirements for volume.Amount sufficient to meet or exceed the endoscope manufacturer's requirements for volume of fluid delivery through endoscope channels.
    Length of fluid circulation time is sufficient to deliver fluid volumes that meet or exceed endoscope manufacturer's requirements.Variable, the length of the fluid circulation time is sufficient to deliver fluid volumes that meet or exceed the endoscope manufacturer's requirements.

    2. Sample size used for the test set and the data provenance:

    This document describes performance testing for a medical device (a flushing sink), not a diagnostic or AI algorithm. Therefore, the concept of a "test set" in the context of data (like medical images or patient records) does not apply here. The testing would have involved physical devices and potentially endoscopes, but no sample size or data provenance is specified in terms of clinical or image data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This is a physical device, and ground truth in the context of expert review for data is not part of this submission. The "ground truth" would be established by engineering specifications and industry standards for flushing efficacy.

    4. Adjudication method for the test set:

    Not applicable. No expert adjudication process for a data test set is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, What was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a physical medical device, not an AI or diagnostic tool that would involve human readers or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    For the performance tests described (flushing, detergent injection, compatibility), the "ground truth" would likely be derived from:

    • Endoscope manufacturer's instructions/specifications: for required flushing volumes, times, and cleaning protocols.
    • Engineering specifications and standards: for detergent delivery accuracy, material compatibility.
    • International standards (IEC 61010-1, IEC 60601-1-2): for electrical safety and electromagnetic compatibility.

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is a physical medical device, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122172

    Validate with FDA (Live)

    Device Name
    SYSTEM 83 PLUS
    Manufacturer
    Date Cleared
    2013-01-25

    (186 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    System 83 Plus Endoscope Washer/Disinfector is designed for the simultaneous reprocessing of up to two flexible submentible endoscopes that are used in the gastrointestinal and/or pulmonary tracts. Flexible scopes that are submention oneoooped to the washing/disinfection cycle of the System 83 Plus may be high level disinfected predicalled and the ruspood to the labeled contact conditions for the germicide as in the predicated device. Note: System 83 Plus device includes two models.

    Note. System 83 Plus 2 is a device with one processing chamber which can process 1 to 2 flexible endoscopes at a time. The System 83 Plus 9 is two 'System 83 Plus 2' units put together. It has two processing chambers which can process 1 to 2 flexible endoscopes in each independently operated processing chamber.

    Device Description

    System 83 Plus Endoscope Washer/Disinfector is designed for the simultaneous reprocessing of up to two flexible submentible endoscopes that are used in the gastrointestinal and/or pulmonary tracts. Flexible scopes that are submention oneoooped to the washing/disinfection cycle of the System 83 Plus may be high level disinfected predicalled and the ruspood to the labeled contact conditions for the germicide as in the predicated device. Note: System 83 Plus device includes two models.

    Note. System 83 Plus 2 is a device with one processing chamber which can process 1 to 2 flexible endoscopes at a time. The System 83 Plus 9 is two 'System 83 Plus 2' units put together. It has two processing chambers which can process 1 to 2 flexible endoscopes in each independently operated processing chamber.

    AI/ML Overview

    This document is a 510(k) clearance letter for a medical device (System 83 Plus Endoscope Washer/Disinfector). It does not contain a detailed study description with acceptance criteria and performance results directly within its text.

    The primary purpose of a 510(k) clearance is to establish substantial equivalence to a legally marketed predicate device, not to provide a full clinical study report with detailed performance metrics against pre-defined acceptance criteria as commonly found in de novo classifications or PMA applications. While the submission to the FDA for a 510(k) would include such data, it is not typically part of the publicly available clearance letter itself.

    Therefore,Based on the provided document, I cannot extract the information required to populate the table and answer the questions regarding acceptance criteria and study details. The document is an FDA 510(k) clearance letter, which confirms that the device is substantially equivalent to a predicate device, but does not provide the specific performance data, acceptance criteria, or study design information.

    To obtain this information, one would typically need to review the original 510(k) submission (if made publicly available by the FDA or the manufacturer) or other technical documentation from Custom Ultrasonics, Incorporated.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061430

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-01-12

    (234 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MinFlex Washer-Disinfector is intended for washing and high-level disinfecting one or two submersible flexible endoscopes that do not contact normally sterile areas of the body and that feature either no internal channel (i.e., a probe) or one internal channel. The types of flexible endoscopes for which the MinFlex Washer-Disinfector is labeled include those that are used to examine, diagnose, and treat diseases of the pulmonary tract, urinary tract, uterine cavity, and ear, nose, and throat (ENT) cavities.

    Device Description

    System 83 Plus MiniFlex Washer -- Disinfector

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device called the "System 83 Plus MiniFlex Washer-Disinfector." It does not contain information about the acceptance criteria or a study proving the device meets acceptance criteria in the context of an AI/ML medical device.

    The provided text describes the FDA's decision to clear the device for marketing based on its substantial equivalence to a legally marketed predicate device. The information requested (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details, ground truth establishment) is typically found in a clinical study report or a risk management document for an AI/ML device, which this document is not.

    Therefore, I cannot fulfill the request using the provided input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K983017

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-09-22

    (25 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The System 83 Plus Endoscope Washer/Disinfector is designed for the simultaneous reprocessing of up to two flexible endoscopes used in the gastrointestinal and/or pulmonary tracts. These flexible scopes may be high level disinfected when the washing/disinfection cycle of the System 83 Plus is used with the labeled contact conditions for the germicide as in the predicated device.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA to Custom Ultrasonics, Inc. regarding their "System 83 Plus" device. This document does not contain information about acceptance criteria for a device, a study proving it meets such criteria, or any details about AI/ML algorithm performance.

    The document primarily focuses on:

    • The FDA's determination of substantial equivalence for the "System 83 Plus" to a legally marketed predicate device.
    • Instructions for marketing the device under general controls provisions.
    • Contact information for compliance and regulatory questions.
    • The Indications for Use statement for the System 83 Plus, which describes its function as an endoscope washer/disinfector for reprocessing up to two pre-cleaned submersible endoscopes. It states that the device provides high-level disinfection when the washing/disinfection cycle meets the labeled contact conditions for the germicide, similar to the predicate device.

    Therefore, I cannot provide the requested information, such as:

    1. A table of acceptance criteria and reported device performance.
    2. Sample sizes, data provenance.
    3. Number and qualifications of experts.
    4. Adjudication method.
    5. MRMC comparative effectiveness study details.
    6. Standalone performance details.
    7. Type of ground truth used.
    8. Training set sample size.
    9. Ground truth establishment for the training set.

    These details are typically found in a device's design validation studies, clinical trial reports, or specific performance testing documentation, which are not part of this FDA clearance letter.