Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K233049

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-11-21

    (57 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TD 200® Automated TEE Probe Disinfector with TD-12® High Level Disinfectant with QwikCheck™ Chemical Indicators is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system can use TD-12® disinfectant, which is designed to be used only with the TD 200® disinfector. The disinfectant be reused in the system.

    TD-12® disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 200® disinfector for high-level disinfection of TEE ultrasound probes.

    TD-12® high level disinfectant and TD 200® disinfector is intended for use by qualified individuals trained in its use.

    TD-12® disinfectant should be used with the following contact conditions in TD 200® disinfector:

    High-Level Disinfectant: TD-12®
    Time: 3 minutes
    Temperature: 38°C
    Minimum Recommended Concentration: 1750 ppm peracetic acid

    QwikCheck™ Chemical Indicators are used prior to the disinfection cycle to ensure that the disinfectant used is above the minimum recommended concentration (MRC) of 1750ppm peracetic acid.

    Device Description

    The TD 200 disinfector provides high-level disinfection of transesophageal (TEE) ultrasound probes when used according to the operating instructions and when used with TD-12 disinfectant. The TD 200 disinfector is for use only with TD-12 disinfectant. The TD-12 is for use only in the TD 200 disinfector. Thus, the TD 200 disinfector and TD-12 disinfectant represent a dedicated system. Each soiled TEE probe is bedside cleaned and manually cleaned according to the TEE probe manufacturer's instructions before insertion into the TD 200 disinfector. A fresh, unopened bottle of TD-12 disinfectant is loaded into the TD 200 disinfector. The TD 200 disinfector heats the TD-12 disinfectant to at least 38°C, soaks the TEE probe at least three (3) minutes, and then thoroughly rinses the disinfectant off the TEE probe before the cycle is complete. The TD 200 disinfector prints a verification report indicating a successful disinfection cycle, the time and the average temperature during the disinfection. The TEE probe is then removed from the TD 200 disinfector and dried according to the TEE probe manufacturer's instructions. The TD 200 disinfector is ready for a new cycle immediately after the preceding cvcle is completed. A chemical indicator is used to ensure that the solution is above the MRC.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the TD 200® Automated TEE Probe Disinfector with TD-12® High-Level Disinfectant and QwikCheck™ Chemical Indicators. This device is intended for the high-level disinfection of Transesophageal (TEE) ultrasound probes. The submission primarily focuses on demonstrating substantial equivalence to a predicate device (TD 200/TD-12 disinfector/disinfectant system, K192228), with the main change being a new supplier for the QwikCheck Chemical Indicator.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the demonstration of "high-level disinfection" and the equivalence to the predicate device. The performance is assessed through various non-clinical tests.

    | Acceptance Criteria (Implicit from Indication and Predicate) | Reported Device Performance (TD 200 with TD-12) |
    | High-Level Disinfection of TEE Probes | Automated Simulated Use Studies: Measurably complete kill of microorganisms after TEE probe processing. |
    | QwikCheck Chemical Indicator accuracy for MRC | The QwikCheck Chemical Indicator, with its starch and iodide agents, produces a color change within 30 seconds. A color similar to the FAIL block indicates ≤ 1750 ppm, while a color similar to the PASS block indicates > 1750 ppm peracetic acid. This indicates it properly discriminates between concentrations above and below the MRC. |
    | Safety and Compatibility (Electrical, Biocompatibility, Material, TEE Probe) | Electrical Safety: Passed testing according to UL 61010-1, 3rd Edition. Biocompatibility: TD-12 is not a sensitizer, irritant, and non-cytotoxic. Residue levels on TEE probes are deemed not likely to have toxic effects. Material Compatibility: No observable effects from exposure of TD 200 materials to TD-12 disinfectant. TEE Probe Compatibility: No cosmetic or functional deterioration observed in GE, Philips, and Siemens TEE probes after processing. TD-12 probe compatibility testing was performed using a use-suspension of 3% w/v (3000ppm) PAA TD-12 at 38°C. |
    | Software Validation | All qualification activities defined in the TD 200 Automated TEE Probe Disinfector with TD-12 High Level Disinfectant Verification and Validation Plan were successfully completed, and documented evidence indicates that the TD 200 consistently performs as intended. All data collected met established acceptance criteria. |

    2. Sample Size Used for the Test Set and Data Provenance

    • TD 200/TD-12 Disinfection Performance:

      • Automated Simulated Use Studies: The exact sample size (number of runs or probes) is not specified, but the results state "In all cases there was a measurably complete kill of microorganisms." This implies a sufficient number of tests were conducted to support the claim.
      • Clinical In-Use Testing: The exact sample size (number of soiled TEE probes) is not specified. The results state "In all cases, there was a complete kill of microorganisms after TEE probe processing." This implies a sufficient number of tests were conducted.
      • Data Provenance: The document does not specify the country of origin for the data or whether it was retrospective or prospective. It appears to be prospective testing conducted by the manufacturer.
    • QwikCheck Chemical Indicator:

      • Test Set Description: The summary describes its mechanism (starch and iodide agents, color change within 30 seconds) and how it visually indicates ppm levels (similar to FAIL block for ≤1750 ppm, similar to PASS block for >1750 ppm). No specific sample size for a test set proving its accuracy for this submission is provided, as the primary difference for this component is the change in supplier, implying the performance characteristics are already established and documented as part of that product's existing regulatory clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not suggest the use of human experts to establish ground truth for the disinfection or chemical indicator performance. The "ground truth" for disinfection is based on objective microbiological enumeration (complete kill of microorganisms), and for the chemical indicator, it's based on objective chemical concentration levels and colorimetric responses.

    4. Adjudication Method for the Test Set

    Not applicable, as there is no indication of human reader adjudication for these studies. The assessment of disinfection effectiveness and chemical indicator performance relies on objective laboratory methods.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The device is an automated disinfectant system and a chemical indicator, not an AI diagnostic tool requiring human interpretation comparison.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the automated disinfection system and the chemical indicator as standalone devices.

    • TD 200/TD-12 Disinfector: Yes, the performance tests (automated simulated use studies, clinical in-use testing) assess the standalone performance of the TD 200 disinfector with TD-12 disinfectant in achieving high-level disinfection.
    • QwikCheck Chemical Indicator: Yes, the description of its function and visual interpretation demonstrates its standalone capability to indicate disinfectant concentration.

    7. The Type of Ground Truth Used

    • For Disinfection Performance (TD 200/TD-12): The ground truth used is objective microbiological enumeration (i.e., "complete kill of microorganisms").
    • For QwikCheck Chemical Indicator: The ground truth is objective chemical concentration levels (1750 ppm peracetic acid) and its associated colorimetric response.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set" in the conventional sense. The studies described are for verification and validation of the device's physical and chemical performance.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230381

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-06-07

    (114 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ethos automated ultrasound probe cleaner disinfector is designed to provide cleaning and high-level disinfection of surface and endocavity ultrasound probes. The system uses AquaCide cleaner/disinfectant, which is designed to be used only with the Ethos automated cleaner disinfector. The disinfectant bottles cannot be reused in the system.

    AquaCide cleaner/disinfectant is intended for use as a single use cleaner and high-level disinfectant, used exclusively in the Ethos automated cleaner disinfector for cleaning and high-level disinfection of surface and endocavity ultrasound probes.

    AquaCide cleaner/disinfectant should be used with the following contact conditions in the Ethos automated cleaner disinfector:

    High-level Disinfectant: AquaCide
    Time: 3 minutes
    Temperature: 47 degrees C
    Minimum Recommended Concentration: 1750 ppm peracetic acid

    The QwikCheck Chemical Indicator is for use in Ethos to determine whether the concentration of peracetic acid, the active ingredient in AquaCide, is above or below the Minimum Recommended Concentration (MRC) of 1750 ppm.

    The Ethos cleaner disinfector with AquaCide and QwikCheck are intended for use by qualified individuals trained in its use.

    Device Description

    The Ethos cleaner disinfector provides cleaning and high-level disinfection of surface and endocavity ultrasound probes when used according to the operating instructions and when used with AquaCide cleaner/disinfectant. The Ethos cleaner disinfector is for use only with AquaCide cleaner/disinfectant. The AquaCide cleaner/disinfectant is for use only in the Ethos cleaner disinfector. Thus, the Ethos cleaner disinfector, AquaCide cleaner/disinfectant represent a dedicated system. Each soiled ultrasound probe has the condom/cover removed and is bedside cleaned according to the ultrasound probe manufacturer's instructions before insertion into the Ethos cleaner disinfector. If a condom/cover is not used the user must manually clean the probe. A fresh, unopened bottle of granular PAA AquaCide cleaner/disinfectant is loaded into the Ethos brings in water, mixes and heats the AquaCide solution to a minimum of 47 degrees. While that is occurring, the Ethos brings in water and the ultrasound probe is pre-rinsed. After temperature is achieved and MRC is confirmed, the Ethos bring the AquaCide solution to the probe and cleans and disinfects the ultrasound probe for at least three minutes. Then the Ethos thoroughly rinses the AquaCide off the ultrasound probe before the cycle is complete. The ultrasound probe is then removed from the Ethos and dried according to the ultrasound probe manufacturer's instructions. The Ethos is ready for a new cycle immediately after the preceding cycle is completed. The Ethos cleaner disinfector incorporates a method for validating the PAA solution through an automatic chemical indicator to ensure each dose of PAA is at or above the MRC. The Ethos cleaner disinfector prints a verification report indicating a successful cleaning and disinfection cycles as well as the time and the average temperature during the cleaning/disinfection. The ultrasound probe is then removed from the Ethos cleaner disinfector and dried according to the ultrasound probe manufacturer's instructions. The Ethos cleaner disinfector is ready for a new cycle immediately after the preceding cycle is completed.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device (Ethos Automated Ultrasound Probe Cleaner Disinfector). It outlines the device's indications for use, comparison to predicate devices, and a summary of non-clinical studies performed to demonstrate its safety and effectiveness.

    However, the document does not describe a study involving human readers or AI assistance, but rather studies demonstrating the efficacy of a cleaning and high-level disinfection device for ultrasound probes. Therefore, many of the requested criteria related to AI/human reader studies (e.g., MRMC studies, effect size, number of experts for ground truth, adjudication methods, training/test set sample sizes for AI models) are not applicable to this document.

    The document primarily focuses on chemical and microbiological testing to validate the disinfection capabilities of the device and its associated chemicals.

    Here's an attempt to extract the relevant information based on the provided text, while noting the inapplicable sections:


    Device Under Review: Ethos Automated Ultrasound Probe Cleaner Disinfector with AquaCide Cleaner/High-Level Disinfectant and QwikCheck Chemical Indicator (K230381)

    Purpose of the Device: To provide cleaning and high-level disinfection of surface and endocavity ultrasound probes.

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, "performance" relates to its ability to clean and disinfect effectively, and the acceptance criteria are typically microbiological kill rates and chemical stability/concentration.

    Test TitleAcceptance Criteria (What success looks like)Reported Device Performance (Result)
    Biocompatibility Skin Irritation TestISO 10993-10: device extract can be a slight irritant (the acceptance criteria is the result here)Slight irritant
    Biocompatibility Sensitization TestISO 10993-10: device extract can be a slight sensitizing (the acceptance criteria is the result here)Non-sensitizing
    Biocompatibility Cytotoxicity TestISO 10993-5: device extract is not cytotoxicNot cytotoxic
    Toxicology AssessmentFDA Submissions for Liquid Chemical Sterilants/High Level Disinfectants: evidence/justifications provided device is non-toxicLow risk of Toxic
    Bacillus subtilis Bench Test (Sporicidal)(AOAC) Official Method 966.04 for 5.0 hrs at 45°C: No growthNo growth
    Clostridium sporogenes Bench Test (Sporicidal)(AOAC) Official Method 966.04 for 5.0 hrs at 45°C: No growthNo growth
    Mycobacterium terrae Bench Test (Tuberculocidal)(AOAC) Official Method 965.12 for 3.0 mins at 45°C: At least 6log reductionNo growth (implies >6 log reduction, as typically "no growth" in such tests means the detection limit was not met after inoculation)
    Staphylococcus aureus Bench Test (Bactericidal)(AOAC) Official Method 955.15 for 3.0 mins at 45°C: At least 6log reductionNo growth (implies >6 log reduction)
    Salmonella enterica Bench Test (Bactericidal)(AOAC) Official Method 955.14 for 3.0 mins at 45°C: At least 6log reductionNo growth (implies >6 log reduction)
    Pseudomonas aeruginosa Bench Test (Bactericidal)(AOAC) Official Method 964.02 for 3.0 mins at 45°C: At least 6log reductionNo growth (implies >6 log reduction)
    Trichophyton interdigitale Bench Test (Fungicidal)(AOAC) Official Method 955.17 for 3.0 mins at 45°C: At least 6log reductionNo growth (implies >6 log reduction)
    Herpes Simplex Virus Type 1 Bench Test (Virucidal)ASTM E1053-20 for 3.0 mins at 45°C: Greater than 6log reductionGreater than 6log reduction
    Human Influenza Virus A (H1N1) Bench Test (Virucidal)ASTM E1053-20 for 3.0 mins at 45°C: Greater than 6log reductionGreater than 6log reduction
    Adenovirus Type 1 Bench Test (Virucidal)ASTM E1053-20 for 3.0 mins at 45°C: Greater than 6log reductionGreater than 6log reduction
    Mycobacterium Terrae Simulated Use TestPerformed in Ethos per FDA Submissions for Liquid Chemical Sterilants/High Level Disinfectants for 3.0 mins at 45°C: Greater than 6log reduction for all probes and all lotsAll lots and all probes had greater than a 6log reduction
    AquaCide Storage Stability TestFDA Submissions for Liquid Chemical Sterilants/High Level Disinfectants: Solubilized to meet MRC of 1750 ppm PAA at 18 months18 month shelf life (meets specifications at 18 months)
    Clinical In-Use TestingMeasurable complete kill of microorganisms after ultrasound probe processingMeasurable complete kill of microorganisms after ultrasound probe processing

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • Bench Tests (Microbiological Efficacy): Not explicitly stated as numerical counts of "samples." Performance is typically demonstrated on multiple replicates of inoculated carriers/surfaces or cultures. The phrases "all lots" and "all probes" are used in some tests (e.g., Mycobacterium Terrae Simulated Use Test), suggesting multiple manufacturing lots of AquaCide and multiple types of probes were tested.
      • Biocompatibility/Toxicology: Not explicitly stated as numerical counts. These are typically performed on extracts from the device materials or reprocessed items.
      • Clinical In-Use Testing: "Soiled surface and endocavity probes" were used, but the specific number (sample size) is not provided.
    • Data Provenance: The document does not specify the country of origin for the data; it focuses on adherence to FDA guidelines and recognized standards (e.g., ISO, AOAC, ASTM). The studies are generally retrospective in the sense that they are conducted in a lab setting to support a premarket submission, not in real-world clinical trials. All studies are "non-clinical" as described in the summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable: This document describes device efficacy testing (microbiological kill, chemical properties, biocompatibility) rather than creating ground truth for an AI diagnostic model by human experts. The "ground truth" for the disinfection efficacy is established through standardized microbiological methods and chemical analysis, performed by trained laboratory personnel.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable: Adjudication methods like 2+1 are typically for reconciling discrepancies in expert opinion when establishing ground truth for diagnostic image sets. This is not relevant to the type of device testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: No MRMC study was performed as this is not an AI diagnostic device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is an automated cleaning and disinfection device, not an algorithm. Its "standalone performance" is its ability to disinfect as measured by the non-clinical studies.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for disinfection efficacy is established through:
      • Microbiological Standards: Demonstrated reduction/kill of specific microorganisms (bacteria, fungi, viruses, spores) on inoculated samples, measured by established laboratory methods (e.g., AOAC, ASTM).
      • Chemical Analysis: Verification of disinfectant concentration (e.g., 1750 ppm peracetic acid) and stability over time.
      • Biocompatibility Standards: Assessment against ISO 10993 for irritation, sensitization, and cytotoxicity.

    8. The sample size for the training set

    • Not Applicable: This is not an AI device that requires a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable: This is not an AI device that requires "ground truth for a training set."
    Ask a Question

    Ask a specific question about this device

    K Number
    K160921

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-11-03

    (213 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TD 100® disinfector is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system can use TD-50 or TD-80 disinfectants, which are designed to be used only with the TD 100® disinfector. The disinfectant bottles cannot be reused in the system.

    TD-50 disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 100@ disinfector for high-level disinfection of TEE ultrasound probes.

    TD-50 high level disinfectant and TD 1000 disinfector system are intended for use by qualified individuals trained in its use.

    TD-50 disinfectant should be used with the following contact conditions in TD 1008 disinfector:

    High-Level Disinfectant: TD-5®
    Time: 5 minutes
    Temperature: 38°C - 40°C
    Minimum Recommended Concentration: 1.7% glutaraldehyde

    TD-80 disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 1000 disinfector for high-level disinfection of TEE ultrasound probes.

    TD-8@ high level disinfectant and TD 1000 disinfector system are intended for use by qualified individuals trained in its usc.

    TD-80 disinfectant should be used with the following contact conditions in TD 1008 disinfector:

    High-Level Disinfectant: TD-8®
    Time: 5 minutes
    Temperature: 38°C - 40°C
    Minimum Recommended Concentration: 0.3% ortho-phthalaldehyde

    Device Description

    The TD 100® disinfector provides high level disinfection of transesophageal (TEE) ultrasound probes when used according to the operating instructions, and when used with TD-50 or TD-80 disinfectants. The TD 1000 disinfector is for use only with TD-50 or TD-8® disinfectants. The TD-5º and TD-8º disinfectants are for use only in the TD 100 disinfector. Thus, the TD 100® disinfector with TD-50 and TD-80 disinfectants represents a dedicated system. Each soiled TEE probe is bedside cleaned and manually cleaned according to the TEE probe manufacturer's instructions before insertion into the TD 100 disinfector. A fresh, unopened bottle of TD-50 or TD-80 disinfectant is loaded into the TD 100% disinfector. The TD 100% disinfector heats the TD-5% or TD-8% disinfectant to 38°-40°C, soaks the TEE probe for at least five minutes, and then thoroughly rinses the disinfectant off the TEE probe before the cycle is complete. The TD 100 disinfector prints a verification report indicating a successful disinfection cycle and the time and average temperature during the disinfection. The TEE probe is then removed from the TD 100® disinfector and dried according to the TEE probe manufacturer's instructions. The TD 100® disinfector is ready for a new cycle immediately after the preceding cycle is completed. Because a fresh bottle of TD-5 or TD-8 disinfectant is used with each cycle, no monitoring of the disinfectant's potency is required, nor is there any requirement for daily testing of the disinfectant solution.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (TD-100 Disinfector with TD-5 and TD-88 High Level Disinfectants). It describes the device, its indications for use, and makes comparisons to predicate devices. However, the document does not contain the specific information requested in the prompt regarding an AI device, acceptance criteria tables with performance, sample sizes for test/training sets, expert ground truth, adjudication methods, or MRMC studies.

    Instead, the document focuses on the disinfector's ability to achieve high-level disinfection, material compatibility, and stability in comparison to other disinfectants. The "acceptance criteria" presented are related to the effectiveness of the disinfectants against various microorganisms and operational parameters of the disinfection process (time, temperature, concentration).

    Here's a breakdown of what is and is not in the provided text, in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • What is present: The document includes a table (Table C.1c) comparing the TD-8 and TD-5 disinfectants' performance against various organisms (Sporicidal, Tuberculocidal, Bactericidal, Fungicidal, Virucidal, Simulated Use, Clinical In-Use) with predicate devices. The "performance" here is typically the contact time and temperature required to achieve the desired effect (e.g., kill the organism). It does not present a typical "acceptance criteria" table with a defined threshold and then the device's reported achievement against that threshold in the way an AI performance summary would.
    • Missing: Specific quantitative acceptance criteria (e.g., "sensitivity must be >X%"). The table shows what the disinfectants achieve in terms of kill time/temperature.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Missing from the text: No sample sizes are provided for any of the microbial efficacy tests (e.g., number of replicates, number of devices tested).
    • Missing from the text: No data provenance (country of origin, retrospective/prospective) is mentioned for the microbial efficacy studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Not Applicable / Missing: This concept (experts establishing ground truth for a test set) is relevant for AI/diagnostic devices that interpret data (e.g., images). For a disinfectant device, the "ground truth" is established through laboratory microbiological testing (e.g., demonstrating reduction in colony-forming units). The document does not describe the laboratory personnel or their qualifications.

    4. Adjudication Method for the Test Set:

    • Not Applicable / Missing: Adjudication methods (like 2+1, 3+1) are for human review of data, typically in diagnostic studies. This is not mentioned for the disinfectant efficacy testing.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    • Not Applicable: This type of study is for evaluating the impact of AI on human reader performance in diagnostic tasks. It is not relevant for a high-level disinfectant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable: This pertains to AI algorithm performance. The device is a disinfector, not an algorithm. The "standalone" performance here would be the device's ability to disinfect, which is what the microbial efficacy tests address.

    7. The Type of Ground Truth Used:

    • What is present: The "ground truth" for this device's performance is microbiological efficacy (kill of specified microorganisms) demonstrated through laboratory testing (Sporicidal, Tuberculocidal, Bactericidal, Fungicidal, Virucidal, Simulated Use, Clinical In-Use). The "Clinical In-Use" section mentions "measurably complete kill of microorganisms" from "Wild-type from patients."
    • Missing details: The specific methodologies for measuring kill (e.g., log reduction) are not detailed in this summary.

    8. The Sample Size for the Training Set:

    • Not Applicable / Missing: This concept is for AI/machine learning models. The device is a physical disinfector with chemical agents. There is no "training set" in this context. The development of the disinfectant formulation and system design would involve R&D and testing, but not a "training set" in the AI sense.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable / Missing: As above, there is no "training set" for this type of device.

    Summary of what can be extracted, interpreted, and what is missing based on the prompt's requirements:

    1. A table of acceptance criteria and the reported device performance

    TestOrganismAcceptance Criteria (Inferred from Predicates & General HLD Standards)Reported Device Performance (TD-8)
    SporicidalBacillus subtilisAchieve kill/inactivation32.0 hrs at 37°C
    SporicidalClostridium sporogenesAchieve kill/inactivation32.0 hrs at 37°C
    TuberculocidalMycobacterium terraeAchieve kill/inactivation (typically 5-10 min)5.0 min at 37°C
    BactericidalStaphylococcus aureusAchieve kill/inactivation (typically 5 min)5.0 min at 37°C
    BactericidalSalmonella entericaAchieve kill/inactivation (typically 5 min)5.0 min at 37°C
    BactericidalPseudomonas aeruginosaAchieve kill/inactivation (typically 5 min)5.0 min at 37°C
    FungicidalTrichophyton mentagrophytesAchieve kill/inactivation (typically 5 min)5.0 min at 37°C
    VirucidalPoliovirus Type 1Achieve kill/inactivation (typically 5 min)5.0 min at 37°C
    VirucidalHerpes Simplex Virus Type 1Achieve kill/inactivation (typically 5 min)5.0 min at 25°C
    VirucidalHuman Influenza Virus AAchieve kill/inactivation (typically 5 min)5.0 min at 25°C
    VirucidalAdenovirus Type 1Achieve kill/inactivation (typically 5 min)5.0 min at 25°C
    Simulated UseMycobacterium terraeAchieve kill/inactivation5.0 min at 37°C
    Clinical In-UseWild-type from patientsComplete kill of microorganismsMeasurably complete kill
    • Note: The "Acceptance Criteria" column is inferred from the common requirements for high-level disinfectants and the performance of the predicate devices. The document explicitly states the "Minimum Recommended Concentration" (MRC) for TD-5 (1.7% glutaraldehyde) and TD-8 (0.3% ortho-phthalaldehyde) and the contact conditions (Time: 5 min, Temperature: 38°C - 40°C) as part of its claim.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    * Sample Size: Not specified in the provided text.
    * Data Provenance: Not specified in the provided text. The studies appear to be laboratory-based efficacy tests ("Simulated Use," "Clinical In-Use").

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    * Not applicable as this is a disinfectant device, not an AI/diagnostic device involving expert interpretation. The "ground truth" is derived from direct microbial efficacy testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    * Not applicable for this type of device and study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    * Not applicable for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    * Not applicable in the AI sense. The microbial efficacy tests (Table C.1c) demonstrate the standalone performance of the disinfectant system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    * Microbiological efficacy (kill/inactivation of various microorganisms) determined through laboratory testing and "Clinical In-Use" studies showing "measurably complete kill of microorganisms" from probes after routine clinical use.

    8. The sample size for the training set
    * Not applicable for this type of device.

    9. How the ground truth for the training set was established
    * Not applicable for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051305

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-08-18

    (91 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TD-100 disinfector is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system uses the TD-5 disinfectant, which is designed to be used only with the TD-100. The disinfectant bottles cannot be reused in the system.

    TD-5 is intended for use as a single use high-level disinfectant to be used exclusively in the TD-100 for the high-level disinfection of TEE ultrasound probes.

    The TD-5 High Level Disinfectant and TD-100 disinfector system is intended for use by qualified individuals trained in its use.

    Device Description

    The TD-100 disinfector provides high-level disinfection of transesophageal (TEE) ultrasound probes when used according to the operating instructions, and when used with TD-5 disinfectant. Each soiled TEE probe is pre-cleaned manually before insertion into the TD-100 disinfector. A fresh, unopened bottle of TD-5 disinfectant is loaded into the TD-100 disinfector. The TD-100 disinfector heats the TD-5 disinfectant to 38-40C, soaks the TEE probe at least fiveminutes, and then thoroughly rinses the disinfectant off the TEE probe before the cycle is complete. The TEE probe is then removed from the TD-100 disinfector and dried according to the TEE probe manufacturer's instructions. The TD-100 disinfector is ready for a new cycle immediately after the preceding cycle is completed. Because a fresh bottle of TD-5 disinfectant is used with each cycle no monitoring of the disinfectant's potency required, nor is there any requirement for daily testing of the disinfectant solution.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the CS Medical TD-100 Transesophageal Probe Disinfector and TD-5 High-level Disinfectant, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Target Performance)Reported Device Performance (Achieved)
    High-Level Disinfection:Simulated Use Testing: A 6-Log₁₀ reduction of M. terrae was achieved when disinfecting TEE probes dosed with artificial soil containing M. terrae.
    Residual Disinfectant Toxicity:Biocompatibility: Analysis indicated that the level of residue on the TEE probes following processing in the TD-100 using TD-5 is not likely to have toxic effects on humans. (Implicitly, the acceptance criterion is that there are no toxic levels of residue, and this was met).
    Material Compatibility:Material Compatibility: No observable adverse effects were seen on the materials used to construct the TD-100 after extended exposure to TD-5 High-level Disinfectant. (Implicitly, the acceptance criterion is no material degradation, and this was met).
    Disinfectant Stability:Stability: TD-5 High-level Disinfectant was tested and shown to be stable for a shelf life of one year. The concentration of the disinfectant at the end of 18 months of testing was well above the 1.7% minimum recommended concentration. (Implicitly, the acceptance criterion is stability for at least one year and maintaining minimum concentration, and this was met with exceedance on concentration over a longer period).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state a numerical sample size for either the simulated use testing or the in-use clinical testing.

    • Simulated Use Testing: It states, "In all cases a 6-Log₁₀ reduction of M. terrae was achieved." This implies multiple tests were conducted, but the specific number is not provided.
    • In Use Testing: It states, "TEE Probes used in a clinical setting were examined. The probes were bioburden post cleaning and a new population of probes was examined for bioburden after disinfection. In all cases, there was a complete kill of microbes after disinfection." Again, "in all cases" implies multiple probes, but no specific number is given.

    Data Provenance: The document does not explicitly state the country of origin for the data. The study appears to be retrospective for the "In Use Testing" as it refers to "probes used in a clinical setting" that were then examined. The "Simulated Use Testing" likely involved prospective testing in a lab environment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The efficacy testing revolves around microbial reduction and absence of growth, which would typically involve laboratory analysis by microbiologists rather than interpretation by clinical experts like radiologists.

    4. Adjudication Method for the Test Set:

    Not applicable. The efficacy testing involves quantitative microbial reduction (Log reductions) and qualitative assessment of microbial kill, which are objective laboratory measurements, not subjective evaluations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This type of study was not conducted. The device is a high-level disinfectant system, not an AI-powered diagnostic tool that assists human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This classification doesn't directly apply because the device is a physical disinfectant system, not a software algorithm. However, the "Efficacy Testing" sections (Simulated Use and In Use Testing) demonstrate the standalone performance of the device system itself (TD-100 plus TD-5) in achieving high-level disinfection without human intervention during the disinfection cycle (after manual pre-cleaning and probe loading).

    7. Type of Ground Truth Used:

    • Microbiological Culture/Counts: For efficacy testing, the ground truth was established by microbiological methods measuring the reduction of specific microorganisms (M. terrae for simulated use) and the complete kill of microbes (for in-use testing). This is a direct measurement of disinfection effectiveness.
    • Chemical Analysis: For biocompatibility (residue) and material compatibility, the ground truth would be chemical analysis of residue levels and visual/physical inspection for material degradation.

    8. Sample Size for the Training Set:

    Not applicable. This device is not an AI/ML product developed with a training set. The "development" would involve chemical formulation and engineering design, not data training.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1