Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K161537

    Validate with FDA (Live)

    Date Cleared
    2016-10-13

    (132 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This TENS device is to be used for temporary relief of pain associated with sore or aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.

    Device Description

    The Counter Pain Therapy Device, Model P.T.S.-IV (K161537) is identical to the previous cleared version of the device with the same name (K132993). The purpose of this 510(k) submission is to seek different (expanded) Indications for Use. The only changes are in the 'Indication for Use Statement'.

    The Counter Pain Therapy Device, Model P.T.S.-IV features two channels and seven (7) preset programs. The output of both channels can be adjusted individually for intensity in sixteen (16) steps and the treatment time is selectable from 20, 30 or 40 minutes. All features are well indicated by large icons on a backlit LCD screen for easy reading.

    The device is powered by three (3) rechargeable Lithium Batteries which are permanently built into the main unit and can be recharged with the charging unit supplied for the device.

    The Pain Therapy Device consists of the handheld main unit (controller), two pairs of self-adhesive electrodes, a plug-in charging unit (adaptor 120VAC to 5VDC) and two pairs of electrode cables.

    AI/ML Overview

    This document describes the regulatory clearance (K161537) for the OTC TENS Device, Model PTS-IV. However, it explicitly states that no clinical tests were performed or relied upon to determine substantial equivalence for this specific submission because the device itself is identical to a previously cleared version (K132993), and the only change made was an expansion of the Indications for Use. Therefore, the information requested regarding acceptance criteria and a study proving the device meets those criteria, particularly in the context of clinical performance, cannot be fully provided from this document.

    The document primarily focuses on non-clinical testing and substantial equivalence to predicate devices based on design, technological characteristics, and safety standards.

    Here's a breakdown of the information that can be extracted or inferred from the provided text, and where the requested details are not available:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: The document refers to meeting the requirements of recognized consensus standards for safety and performance. These standards are the acceptance criteria for non-clinical aspects.
    • Reported Device Performance: The performance is stated in terms of compliance with these standards.
    Acceptance Criteria (Non-Clinical)Reported Device Performance
    Safety & Essential Performance: IEC 60601-1:1988 + A1:1991 + A2:1995 "Medical electrical Equipment – Part 1: General requirements for basic safety and essential performance".Device was tested to and met the requirements in K132993. Non-clinical testing validated the design and assured conformance.
    Safety of Nerve and Muscle Stimulators: IEC 60601-2-10: 1987 (first Edition + amendment 1 (2001) for use in conjunction with IEC 60601-1 (1988). Amendments 1 (1991) and 2 (1995) "Medical electrical equipment – Part 2: "Particular Requirements for safety of nerve and muscle stimulators".Device was tested to and met the requirements in K132993. Non-clinical testing validated the design and assured conformance.
    Electromagnetic Compatibility: IEC 60601-1-2: 2001 + A1: 2004 & IEC60601-2-10: 1987 + A1: 2001 Clause 36 Medical Electrical Equipment, Part1-2: "General requirements for Basic Safety and Essential Performance" Collateral Standard: Electromagnetic Compatibility.Device was tested to and met the requirements in K132993. Performance testing established that, with respect to EMC and electrical safety in its intended operational environment, the device conforms to all applicable requirements.
    Biocompatibility: Electrodes are body contacting parts.Electrodes were tested for biocompatibility as part of their separate FDA 510(k) clearance.
    Software Performance: Software performs within specifications and is safe for intended use.Test results concluded software performs within specifications and is safe. Test protocol sufficiently verifies main functional operation.
    Cleaning Effectiveness: Cleaning instructions are sufficient.Validation of the manual cleaning method as per instructions. Testing concluded the device can be cleaned by the described methods.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Not provided for clinical performance. The document explicitly states: "Counter did not conduct, nor rely upon, clinical tests to determine substantial equivalence of the Pain Therapy Device Model P.T.S.-IV. vs. the predicates."
    • For non-clinical testing, the "sample size" would refer to the number of devices tested to the standards. This specific number is not provided, only that "testing" was performed. The data provenance would be from internal company testing and third-party lab testing for standards compliance (e.g., electrical safety).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical test set with human data and expert-established ground truth was used for this specific 510(k) submission. The regulatory decision was based on substantial equivalence to predicate devices and non-clinical engineering and safety testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a Transcutaneous Electrical Nerve Stimulator (TENS device), not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The device is a TENS unit, not an algorithm. Its performance is inherent in its electrical output and mechanical operation, as demonstrated by compliance with engineering standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the non-clinical tests, the "ground truth" was established by the requirements of the international consensus standards (e.g., IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2) which define the acceptable limits and performance criteria for safety and electrical characteristics.
    • For biocompatibility, the ground truth would be chemical/biological analyses demonstrating safety as per relevant biocompatibility standards for medical devices, which were part of the electrode's separate FDA clearance.
    • For software, the ground truth was the expected functional specifications and safety requirements as defined internally and verified through testing.

    8. The sample size for the training set

    • Not applicable. This device does not use machine learning or AI that requires a "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no training set for this type of device.

    In summary:

    This 510(k) cleared the OTC TENS Device, Model PTS-IV, by demonstrating substantial equivalence to a previously cleared version of the same device (K132993) and to another predicate device (K121757) for its expanded indications for use. The clearance was based entirely on non-clinical testing which confirmed the device's compliance with established safety and performance standards (IEC 60601 series). No clinical studies or human performance data were used for this submission, as the physical device itself had already been vetted in a prior submission, and the current submission only sought to expand its marketing claims (Indications for Use) based on existing equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150277

    Validate with FDA (Live)

    Date Cleared
    2016-02-12

    (372 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back due to strain from exercise or normal household work activities (Choose Mode A, B, or C)

    To be used for temporary relief of pain associated with sore and aching muscles in the upper extremities (Arms) due to strain from exercise or normal household work activities (Choose Mode A, B or C)

    To be used for temporary relief of pain associated with sore and aching muscles in the lower extremities (Legs) due to strain from exercise or normal household work activities (Choose Mode A, B or C)

    To be used for symptomatic relief and management of chronic, intractable pain associated with Arthritis (Choose Mode A)

    To stimulate healthy muscles in order to improve and facilitate muscle performance (Choose Mode B)

    Device Description

    The Pain Therapy System, Model PTS-II is a selectable dual channel TENS and EMS device operated by DC 3.0V (AAA*2 batteries). It is made up of one main unit, two electrode cables, and two pairs of electrode pads.

    There are 3 selectable, pre-programmed output waveforms (modes) to choose from. The two channels share one knob to adjust, the intensity are adjustable from 0 to 5. Running time can be selected among 10, 20 or 40min and recycled. There is a dial with on/off on the side that enables the device to be turned on or off, and it can be used to adjust the output intensity level, turn clockwise to increase intensity and turn anticlockwise to decrease the intensity. In the front of the device, there are 2 buttons and 6 indicator lights, in which one button is the mode selection button, and the other is for time selection. There are three modes A. B. C. respectively indicated by three indicator lights, which are on the right of the mode selection button. The device has 3 timing levels, namely 10, 20, 40 min, and indicated by three indicator lights, which are on the right of the timing selection button.

    AI/ML Overview

    The provided document is a 510(k) premarket notification from the FDA for a device called "Pain Therapy System, Model PTS-II". It describes the device, its intended use, and a comparison to a predicate device to establish substantial equivalence.

    Crucially, this document is for a Transcutaneous Electrical Nerve Stimulator (TENS) device, not an AI/ML medical device. Therefore, the concepts of "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of an AI/ML device (e.g., performance metrics like accuracy, sensitivity, specificity, or reader studies) are not applicable here. The substantial equivalence for this TENS device is established through non-clinical testing against recognized consensus standards and a comparison of technological characteristics with a legally marketed predicate device.

    The questions you've asked (about sample sizes, ground truth experts, MRMC studies, standalone performance, training sets, etc.) are highly relevant to the evaluation of AI/ML medical devices. However, they do not apply to the traditional medical device evaluation process exemplified by this 510(k) submission for a TENS device.

    Here's a breakdown of what is relevant from the document regarding its "acceptance criteria" (which are more accurately "conformance criteria" in this context) and "proof":

    1. A table of acceptance criteria and the reported device performance:

    Instead of acceptance criteria for AI performance, the document lists technological characteristics of the proposed device and compares them to the predicate device. The "acceptance criteria" here are essentially that the proposed device's characteristics are sufficiently similar to the predicate device, or any differences do not raise new questions of safety or effectiveness.

    ParameterSubject Device (Pain Therapy System, Model PTS-II) K150277Predicate Device (OTC TENS, Model T1040 aka Aurawave) K124055
    General
    Power Source3.0 v (2*1.5vAAA)4.5 v (3*1.5vAAA batteries)
    Method of Line Current IsolationUse resistance to isolateUse transformer to isolate
    Patient Leakage Current (Normal Condition & Single Fault Condition)N/A (implies within acceptable limits, not explicitly stated values)N/A (implies within acceptable limits, not explicitly stated values)
    Average DC current through electrodes when device is on but not connected to patient0μA0μA
    Numbers of Output Modes3 modes (A,B,C)10 modes (Mode1-6, AUTO1-4)
    Number of Output Channels2 Synchronous1 Synchronous
    Method of ChannelParallel connectionN/A
    Regulated current or regulated voltage?Regulated voltageRegulated voltage
    Waveform Parameters (Mode B for Subject Device, assumed comparable mode for Predicate)
    Waveform (e.g. positive-going, reverse, biphasic)Positive-going, Reverse and BiphasicPositive-going, Reverse and Biphasic
    Shape (e.g., spike, rectangular, square wave)square wavesquare wave
    Maximum Output Voltage (Volts) (+/-15%)88vp @500Ω (Mode B has max)102vp @2kΩ106vp @10kΩ40.7vp @500Ω105.1vp @2kΩ154.1vp @10kΩ
    Maximum Output Current (+/-15%)176mA @500Ω (Mode B has max)51.0mA @2kΩ10.6mA @10kΩ81.4mA @500Ω52.5.0mA @2kΩ15.4mA @10kΩ
    Duration of primary phase00
    Pulse Duration170 μs200 μs
    Frequency1-136 Hz1-250 Hz
    For multiphase waveforms only: Symmetrical phases, Pulse DurationSymmetrical phases, N/AN/A
    Net charge (micro coulombs μC) per pulse) (If zero, state method of achieving zero net charge.)1.63μC @500Ω1.9μC @500Ω
    Maximum Phase Charge (μC)29.9μC @500Ω16.9μC @500Ω
    Maximum Current Density (mA/cm2, r.m.s.)8.31mA/cm² @500Ω3.84mA/cm² @500Ω
    Maximum Average Current (average absolute value), mA2.22mA1.28mA
    Maximum Average Power Density (using smallest electrode conductive surface area)$0.115mW/cm^2 @500\Omega$$0.039mW/cm^2 @500\Omega$

    The "reported device performance" is essentially the demonstrated compliance with the listed consensus standards and the presentation of these technological parameters. The conclusion (Page 8) states: "bench testing and safety report documentation demonstrate that the submitted device could maintain the same safety and effectiveness as that of predicate device."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. This isn't a study involving human subjects or data sets in the AI/ML sense. The "testing" refers to bench testing and compliance with electrical safety and performance standards.
    • The device is manufactured by "Counter Scientific Development (GZ) Ltd" in Guangzhou, China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth in the AI/ML context is not relevant for this type of device submission. The "ground truth" here is the adherence to established engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No adjudication in the AI/ML context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. No MRMC study. This is a physical device, not an AI assistance tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. Not an algorithm. The "standalone performance" is the device's ability to meet its operational specifications and safety standards on its own.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. The "ground truth" for this device's evaluation is compliance with recognized consensus standards (ANSI/AAMI ES60601-1, IEC 60601-2-10, IEC 60601-1-2) which define safety and essential performance parameters, and the comparison of its technical specifications to a predicate device.

    8. The sample size for the training set

    • Not applicable. No training set for an AI/ML model.

    9. How the ground truth for the training set was established

    • Not applicable. No training set.

    In summary, the provided document details a 510(k) submission for a non-AI medical device (TENS). The evaluation methodology relies on comparison to a predicate device and compliance with established electrical and safety standards, rather than the AI/ML-specific testing and evaluation criteria you've outlined.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151279

    Validate with FDA (Live)

    Date Cleared
    2015-12-16

    (216 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back due to strain from exercise or normal household work activities choose Manual modes A, B, C, D, F, E or Auto 4.

    To be used for temporary relief of pain associated with sore and aching muscles in the upper extremities (arm) due to strain from exercise or normal household work activities choose Manual modes A, B, C, D, F, E or Auto 1 or Auto 3..

    To be used for temporary relief of pain associated with sore and aching muscles in the lower extremities (leg) due to strain from exercise or normal household work activities choose Manual modes A, B, E or Auto 1 or Auto 4.

    To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, choose Manual modes A and C or Auto 1, Auto 2, Auto 3, or Auto 4.

    To be used to stimulate healthy muscles in order to improve and facilitate muscle performance choose Manual Mode A or Mode B.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a letter from the FDA regarding a 510(k) premarket notification for a Transcutaneous Electrical Nerve Stimulator (TENS) device. It confirms the substantial equivalence of the device to legally marketed predicate devices and includes the "Indications for Use" for the device.

    However, the document does not contain any information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or comparative effectiveness studies. These types of details are typically found in the accompanying technical documentation or clinical study reports submitted as part of the 510(k) application, which are not present here.

    Therefore, I cannot provide the requested information from this document. The prompt asks to "describe the acceptance criteria and the study that proves the device meets the acceptance criteria" based on the input, but this input does not contain that specific information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132993

    Validate with FDA (Live)

    Date Cleared
    2014-04-24

    (212 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This TENS Device is to be used for temporary relief of pain associated with sore or aching muscles of lower back, arms, or legs due to strain from exercise or normal household and work activities.

    Device Description

    The Counter Pain Therapy Device, Model P.T.S.-IV features two channels and seven (7) preset programs. The output of both channels can be adjusted individually for intensity in sixteen (16) steps and the treatment time is selectable from 20, 30 or 40 minutes. All features are well indicated by large icons on a backlit LCD screen for easy reading. The device is powered by three (3) rechargeable Lithium Batteries which are permanently built into the main unit and can be recharged with the charging unit supplied for the device. The Pain Therapy Device consists of the handheld main unit (controller), two pairs of self-adhesive electrodes, a plug-in charging unit (adaptor 120VAC to 5VDC) and two pairs of electrode cables.

    AI/ML Overview

    The Counter Pain Therapy Device, Model P.T.S.-IV is a Transcutaneous Electrical Nerve Stimulator (TENS) device intended for the temporary relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.

    The device's acceptance criteria and performance were established through non-clinical testing for substantial equivalence to a predicate device, the OMRON Electro Therapy Pain Relief, Model PM3030 (K110068).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria StandardReported Device Performance
    Electrical SafetyIEC 60601-1:1988 + A1:1991 + A2:1995 "Medical electrical Equipment - Part 1: General requirements for basic safety and essential performance".The device has been tested to and meets the requirements of IEC 60601-1. Performance testing established that, with respect to electrical safety in its intended operational environment, the device conforms to all applicable requirements of IEC 60601-1.
    Particular Safety for Nerve & MuscleIEC 60601-2-10: 1987 (first Edition + amendment 1 (2001) for use in conjunction with IEC 60601-1 (1988), Amendments 1 (1991) and 2 (1995) "Medical electrical equipment - Part 2: "Particular Requirements for safety of nerve and muscle stimulators".The device has been tested to and meets the requirements of IEC 60601-2-10.
    Electromagnetic Compatibility (EMC)IEC 60601-1-2: 2001 + A1: 2004 & IEC60601-2-10: 1987 + A1: 2001 Clause 36 Medical Electrical Equipment, Part1-2: "General requirements for Basic Safety and Essential Performance" Collateral Standard: Electromagnetic Compatibility. (Also cited as IEC 60601-1-2 in other sections)The device has been tested to and meets the requirements of IEC 60601-1-2. Performance testing established that, with respect to EMC in its intended operational environment, the device conforms to all applicable requirements of IEC 60601-1-2.
    BiocompatibilityNot explicitly stated as a separate standard, but addressed.The electrodes (the only body contacting parts) have been tested for biocompatibility as part of their separate FDA 510(k) clearance.
    Software PerformanceNot explicitly stated as a separate standard, but addressed internally.Based upon test results, it was concluded that the software performs within specifications and is safe for the stated intended use. A permanent hazard analysis is implemented in the software development process, and the test protocol sufficiently verifies the software's main functional operation due to its clear architecture.
    Cleaning EffectivenessNot explicitly stated as a separate standard, but addressed internally.The cleaning instructions in the Instruction Manual have been tested and validated to be sufficient. All testing concluded that the device can be cleaned by the methods described in the Instruction Manual.
    Substantial Equivalence (Overall)Performance and safety compared to predicate device (OMRON Electro Therapy Pain Relief, Model PM3030 - K110068).Bench testing and safety report documentation demonstrated that the submitted device could maintain the same safety and effectiveness as the predicate device. Differences (7 program modes vs 3, 2 channels vs 1) do not affect intended use, raise new questions of safety or effectiveness, or alter the fundamental scientific technology. Therefore, the device is substantially equivalent to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    No specific "test set" in the context of clinical data (e.g., patient data) was used. The substantial equivalence determination was based on non-clinical testing, primarily engineering and performance verification against recognized consensus standards.

    • Sample size: Not applicable in terms of patient data. Testing involved engineering samples of the device.
    • Data Provenance: The testing was conducted by or for Counter Scientific Development (GZ) Ltd. in China, to demonstrate compliance with international standards for medical electrical equipment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. Ground truth, in this context, pertains to clinical outcomes or expert consensus on medical conditions, which were not part of this non-clinical submission. The "truth" in this submission was adherence to engineering and safety standards, evaluated through standard test methodologies, not expert medical assessment of patient data.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was conducted. The submission explicitly states: "Counter did not conduct, nor rely upon, clinical tests to determine substantial equivalence of the Pain Therapy Device Model P.T.S.-IV. vs. the predicate."

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable, as this is a physical medical device (TENS unit), not an algorithm or AI system for diagnostic or treatment decisions. The "software" mentioned refers to embedded firmware controlling the device's functions, not a standalone diagnostic algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this submission was compliance with recognized consensus standards for electrical safety, electromagnetic compatibility, and specific requirements for nerve and muscle stimulators (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2). Additionally, design requirements and specifications from the manufacturer formed the basis for verifying software performance and cleaning efficacy.

    8. Sample Size for the Training Set

    Not applicable, as there was no artificial intelligence or machine learning component requiring a training set. This was a non-clinical device submission based on engineering and safety testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set in the context of machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1