Search Results
Found 13 results
510(k) Data Aggregation
(203 days)
For the qualitative determination of IgG antibodies to Helicobactor pylori antigen in human sera by indirect enzyme immunoasay. The H.pylori IgG assay may be used as an aid in the diagnosis of H. pylori infection in adult patients with gastrointestinal symptoms. The test can be performed either manually or in conjunction with the MAGO® PLUS Automated EIA Processor.
The & H. pylori IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG antibodies to H. pylori antigen in human serum.
This document describes the performance characteristics of the Is-H. pylori IgG ELISA Kit, an in vitro diagnostic (IVD) device used for the qualitative determination of IgG antibodies to Helicobacter pylori antigen in human sera.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the reported performance metrics, demonstrating the device's accuracy and precision.
| Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Sensitivity (Biopsy GT) | High sensitivity (e.g., >85-90%) | 94.9% (95% CI: 89.3-98.1%) |
| Specificity (Biopsy GT) | High specificity (e.g., >85-90%) | 90.4% (95% CI: 83.8-94.9%) |
| Overall Agreement (Biopsy GT) | High overall agreement (e.g., >85-90%) | 92.6% (95% CI: 88.5-95.6%) |
| Relative Agreement (vs. Predicate ELISA) | High agreement (e.g., >95%) | 99.6% |
| Precision (Intra-assay CV%) | Low variability (e.g., <20% for positive, <40% for negative, these are approximate common ranges for ELISA tests) | Ranged from 2.16% to 37.22% for Site #1, 3.40% to 20.01% for Site #2, 2.08% to 19.99% for Site #3 (Manual) and 2.97% to 62.36% for MAGO Plus |
| Precision (Inter-assay CV%) | Low variability (e.g., <25% for positive, <50% for negative) | Ranged from 3.24% to 24.91% for Site #1, 4.93% to 18.56% for Site #2, 5.42% to 16.60% for Site #3 (Manual) and 5.19% to 57.40% for MAGO Plus |
| Correlation (Manual vs. MAGO Plus) | Strong positive correlation (e.g., r > 0.95) | Pearson Correlation Coefficient: 0.993 |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Sensitivity and Specificity Study (Section A): 249 patients.
- 121 sera characterized as positive for H. pylori.
- 128 sera characterized as negative for H. pylori.
- Sample Size for Relative Agreement Study (Section B): 249 patients (same samples as in Section A).
- Data Provenance: The sera were frozen retrospective samples from patients. The country of origin is not explicitly stated in the provided text.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The ground truth for characterizing H. pylori infection in the test set (Section A) was established using "biopsy with culture, stain and CLO results for H.pylori". While these are considered reference standard clinical tests, the document does not specify the number of experts (e.g., pathologists, microbiologists) who interpreted these results or their specific qualifications (e.g., years of experience as a pathologist). The interpretation of such tests inherently involves expert judgment, but the level of detail is not provided.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method (like 2+1 or 3+1) for resolving discrepancies in the biopsy, culture, stain, or CLO results that formed the ground truth. It states that "Based on the results of this testing, the patient sera were characterized." This implies that a consensus or a defined decision rule based on these multiple tests was used to establish the positive or negative status for each patient, but the specific adjudication process is not detailed. Equivocal results from the device under test were excluded from calculations, but this is a different kind of exclusion from an adjudication of ground truth.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
No MRMC comparative effectiveness study was mentioned. This device is an immunoassay kit, not an AI-based image analysis or interpretation tool that would typically involve human readers and AI assistance for interpretation. Therefore, this section is not applicable.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
For the sensitivity and specificity study (Section A), the device (Is-H. pylori IgG Test Kit) was tested on sera, and its results were compared to the biopsy-based ground truth. This represents standalone performance, as the interpretation of the ELISA results would be direct based on the established cut-off, without a human "reader" making a subjective judgment in the same way a radiologist reads an image.
The section on "Correlation of Manual and MAGO Plus Results" (Section D) and "MAGO Plus Precision" (Section E) studies the performance of the device when used with an automated EIA processor. This also represents standalone performance of the device in either manual or automated modes, where the device directly produces a result.
7. The Type of Ground Truth Used
The primary ground truth used for the sensitivity and specificity study (Section A) was based on biopsy with culture, stain, and CLO results for H. pylori. These are objective laboratory and histopathological methods considered gold standards for H. pylori infection diagnosis.
For the "Relative Agreement Versus Another ELISA" study (Section B), the "ground truth" was the results from "another commercially available kit for H. pylori IgG antibodies." However, the notice explicitly states: "Please be advised that 'relative' refers to the comparison of the assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison's accuracy to predict disease." This clarifies that the predicate ELISA was used for relative performance comparison, not as a true gold standard for disease presence.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning. This device is an immunoassay kit (ELISA), which is a biochemical test, not an algorithm that requires a separate training set. The various studies described (Sensitivity/Specificity, Relative Agreement, Precision) are all validation studies for the device's performance.
9. How the Ground Truth for the Training Set Was Established
As this is an immunoassay and not an AI/ML algorithm requiring a training set, this question is not applicable. The device itself produces a result based on a biochemical reaction and predefined interpretation rules.
Ask a specific question about this device
(143 days)
For the qualitative and semi-quantitative determination of IgGi antibodies to Mumps virus m human serum of adults over cighteen vears of age by indirect enzyme immunoassay as an aid in the diaunosis at Mumps infection. The evaluation of paired sera, to determine a significant increase in Mumps IgG antihody itel can also aid in the diagnosis of acute infection by seroconversion through testing acute and convalescent sera. The test can be performed either manually or in conjunction with the MAGO® Plus Automated EIA processor. Performance characteristics have not been established on children
The & Mumps IgG ELISA Kit is an Enzyme-Linked ImmunoSorbent Assay (E.I.S.A) (6) the detection of IgG antibodies to Mumps antigen in human serum.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Mumps IgG ELISA Kit:
Acceptance Criteria and Device Performance for Is-Mumps IgG Test Kit
| Acceptance Criteria Category | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Accuracy (vs. Predicate) | High relative sensitivity (close to 100%) and acceptable relative specificity and overall agreement when compared to a commercially available predicate Mumps IgG ELISA kit. This implies the device should perform comparably to a legally marketed equivalent device in identifying positive and negative samples for Mumps IgG antibodies. (No explicit numerical targets are given, but the comparison aims to show "substantial equivalence"). | Relative Sensitivity: 100% (160/160) (95% Cl: 97.7% - 100%) Relative Specificity: 76.9% (10/13) (95% Cl: 46.2% - 95.0%) Overall Agreement: 98.3% (170/173) (95% Cl: 95.0% - 99.5%) |
| Reproducibility (Manual) | Acceptable intra-assay and interassay coefficients of variation (CV%) for both positive and negative sera across multiple runs and sites. (No explicit numerical targets are given for CV%, but the tables demonstrate the observed variability). | Intra-Assay/Interassay CV% (Site #1):- Positive Sera (A, B, C, D): CV% generally in the range of 4-9% (Interassay CV% 6.36%-8.70%).- Negative Sera (E, F): CV% generally in the range of 21-40% (Interassay CV% 32.18%-39.03%). Intra-Assay/Interassay CV% (Site #2):- Positive Sera (A, B, C, D): CV% generally in the range of 4-10% (Interassay CV% 8.92%-9.63%).- Negative Sera (E, F): CV% generally in the range of 4-18% (Interassay CV% 13.02%-18.06%). Intra-Assay/Interassay CV% (Site #3):- Positive Sera (A, B, C, D): CV% generally in the range of 2-10% (Interassay CV% 4.06%-8.04%).- Negative Sera (E, F): CV% generally in the range of 8-35% (Interassay CV% 17.49%-35.19%). |
| Correlation (Manual vs. MAGO Plus) | Strong correlation between manual assay results and results obtained using the MAGO® Plus Automated EIA Processor. (Implicitly, a high Pearson Correlation Coefficient is expected). | Pearson Correlation Coefficient: 0.954 (indicating good correlation, as stated). |
| Reproducibility (MAGO Plus) | Acceptable intra-assay and interassay coefficients of variation (CV%) for both positive and negative sera when performed on the MAGO® Plus. (No explicit numerical targets are given for CV%). | Intra-Assay/Interassay CV% (MAGO Plus - Site #2):- Positive Sera (A, B, C, D): CV% generally in the range of 6-13% (Interassay CV% 8.84%-11.50%).- Negative Sera (E, F): CV% generally in the range of 0-28% (Interassay CV% 21.19%-23.79%). |
Study Details:
-
Sample sizes used for the test set and the data provenance:
- Accuracy (Comparison Study): 173 fresh sera samples.
- Provenance: "a clinical commercial laboratory, located in the Mid-Atlantic area" (United States, retrospective).
- Reproducibility (Manual & MAGO Plus): 6 sera (4 positive, 2 negative) for each site, assayed 10 times in 3 runs. This means for each Reproducibility section (Manual, MAGO Plus), a total of (6 sera * 10 replicates * 3 runs) + (interassay variability) were evaluated at each site. This refers to the number of measurements/replicates rather than unique patient samples like in the accuracy study.
- Provenance: Not explicitly stated for the source of these 6 sera, but the testing was done at "the manufacturer, a research and development laboratory, and a clinical laboratory."
- Accuracy (Comparison Study): 173 fresh sera samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The ground truth for the comparison study was established by a "commercially available kit for Mumps IgG antibodies" (the predicate device), not by human experts. The study clarifies: "There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison's accuracy to predict disease."
-
Adjudication method for the test set:
- Not applicable as the ground truth was established by another ELISA test, not by human interpretation requiring adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No, an MRMC study was not done. This device is an in-vitro diagnostic (IVD) assay, not a medical imaging or interpretation device that would typically involve multiple human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the performance characteristics (accuracy against a predicate, reproducibility) are reported for the device as a standalone test system, both in manual operation and when automated with the MAGO® Plus. Human "readers" are not part of the performance evaluation beyond performing the lab procedures.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Comparative Device Results: The primary "ground truth" for the accuracy study was the results from a predicate, commercially available Mumps IgG ELISA kit.
- IFA Reference: The footnotes in Table 1 mentioned "313 sera were positive by IFA" and "373 sera were positive by II/A", suggesting that some samples may have also been referenced against Immunofluorescence Assay (IFA), but the main comparison was with the predicate ELISA. The report explicitly states it was not correlated with disease presence/absence.
-
The sample size for the training set:
- The document does not explicitly mention a separate "training set" for the device itself. For IVDs, the development process typically involves internal validation and optimization, but the performance data presented here are usually from a "test set" demonstrating the final product's characteristics.
-
How the ground truth for the training set was established:
- Not applicable, as a distinct training set with established ground truth is not detailed in the provided information. If there were a training phase, it would likely involve similar methods as the test set (e.g., comparison to a reference method, known positive/negative samples).
Ask a specific question about this device
(69 days)
The TPO IgG kit is an Enzyme-Linked Immunosorbent Assay (ELISA) For the qualitative detection and quantitation of antibodies against the thyroid peroxidase antigen in serum as an aid in the diagnosis of thyroid autoimmune disease. The test can be performed either manually or in conjunction with the Mago Plus automated EIA processor.
The & TPO IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG to thyroid peroxidase in human serum.
Acceptance Criteria and Study for Is-anti-TPO IgG ELISA Test Kit
This document describes the acceptance criteria and the studies performed to demonstrate the safety and effectiveness of the Is-anti-TPO IgG ELISA Kit.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Relative Sensitivity | Not explicitly stated, but compared to a commercially available ELISA kit. | 100% (71/71) when equivocal results are excluded. |
| Relative Specificity | Not explicitly stated, but compared to a commercially available ELISA kit. | 97.1% (167/172) when equivocal results are excluded. |
| Overall Agreement | Not explicitly stated, but compared to a commercially available ELISA kit. | 97.9% (238/243) when equivocal results are excluded. |
| Clinical Specificity (Normals) | Not explicitly stated, but assessed in normal patient group. | 91.4% (160/175) with a 95% CI of 86.3-95.1%. |
| Clinical Sensitivity (Grave's Disease) | Not explicitly stated, but assessed in Grave's Disease patient group. | 64.3% (9/14) with a 95% CI of 35.1-87.2%. |
| Clinical Sensitivity (Hashimoto's Disease) | Not explicitly stated, but assessed in Hashimoto's Disease patient group. | 75.0% (12/16) with a 95% CI of 47.6-92.7%. |
| Intra-Assay Precision (CV%) | Desired to be low for consistent results within a single assay run. | Site #1 (Manual): Range from 4.96% to 14.37% for positive sera, and 36.15% to 316.23% for negative sera (higher CV% for very low values is expected).Site #2 (Manual): Range from 6.31% to 19.84% for positive sera, and 6.35% to 15.90% for negative sera.MAGO Plus: Range from 8.42% to 20.61% for all sera. |
| Interassay Precision (CV%) | Desired to be low for consistent results across different assay runs. | Site #1 (Manual): Range from 5.89% to 11.04% for positive sera, and 41.92% to 161.76% for negative sera.Site #2 (Manual): Range from 8.29% to 21.47% for positive sera, and 43.97% to 55.73% for negative sera.MAGO Plus: Range from 11.43% to 48.32% for all sera. |
| Correlation of Manual and MAGO Plus Results | Demonstrate equivalence between manual and automated methods. | Pearson Correlation Coefficient of 0.977 (indicated good correlation). |
2. Sample Size and Data Provenance (Test Set)
- Relative Sensitivity and Specificity Study: 255 frozen retrospective sera.
- Clinical Sensitivity and Specificity Study: 207 frozen retrospective, characterized sera.
- Precision Study: 4 positive sera and 2 negative sera, each assayed 10 times in 3 different runs at 2 different sites. For MAGO Plus precision, 6 sera were assayed 10 times each in 3 different runs.
- Correlation of Manual and MAGO Plus Results: 104 serum samples.
Data Provenance: All data appears to be retrospective. The country of origin of the data is not explicitly stated in the provided text, but given the context of a US regulatory submission, it is likely the data was collected in the USA or a region with comparable clinical practices.
3. Number of Experts and Qualifications for Ground Truth (Test Set)
The document does not explicitly state the number of experts used to establish the ground truth for the test set, nor their specific qualifications (e.g., radiologist with X years of experience).
- For the "Relative Sensitivity and Specificity" study, a "commercially available ELISA kit for TPO antibodies" was used as the comparator, implying its results served as the reference for ground truth in that comparison.
- For the "Clinical Sensitivity and Specificity" study, "characterized sera" were used. This suggests that the ground truth for these samples (e.g., normal, Grave's Disease, Hashimoto's Disease) was previously established through clinical diagnosis, but the method and expert involvement are not detailed.
4. Adjudication Method (Test Set)
The document does not describe an adjudication method for the test set, such as 2+1 or 3+1. The ground truth appears to be based on results from a comparative device or pre-characterized clinical diagnoses.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The device is an in-vitro diagnostic (ELISA kit), not an imaging or AI-assisted diagnostic tool that typically involves human readers. Therefore, the concept of human readers improving with or without AI assistance is not applicable here.
6. Standalone Performance Study
Yes, a standalone performance study was conducted. The "Relative Sensitivity and Specificity" and "Clinical Sensitivity and Specificity" sections, along with the "Precision" studies, evaluate the performance of the "Is-anti-TPO IgG Test Kit" as a standalone device. The correlation study with the MAGO Plus demonstrates the performance of the kit both manually and when integrated with an automated processor, but these are still evaluations of the algorithm/kit itself.
7. Type of Ground Truth Used
- Relative Sensitivity and Specificity Study: The "ground truth" was established by the results of a commercially available ELISA kit for TPO antibodies. This is a form of comparative reference standard.
- Clinical Sensitivity and Specificity Study: The "ground truth" was based on characterized clinical diagnoses (Normals, Grave's Disease, Hashimoto's Disease). This can be categorized as clinical diagnosis/outcomes data, although the specific methods of characterization are not detailed.
8. Sample Size for the Training Set
The document does not mention a "training set" or "validation set" in the context of device development. The studies described are performance evaluations of the final device. This type of in-vitro diagnostic development typically focuses on analytical and clinical performance studies rather than machine learning model training.
9. How Ground Truth for Training Set Was Established
As no training set is described, the method for establishing its ground truth is not applicable or provided.
Ask a specific question about this device
(63 days)
For the qualitative presumptive detection of IgM antibodies to Borrelia burgdorferi in human serum. This ELISA should only be used for patients with signs and symptoms that are consistent with Lyme disease. Equivocal or positive results must be supplemented by testing with a standardized Western blot procedure. Positive supplemental results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease. The test can be performed either manually or in conjunction with the MAGO TM PLUS Automated EIA Processor.
The Is-anti-Borrelia burgdorferi IgM Test Kit can be used during the acute phase (0-4 weeks of symptoms onset) of B. burgdorferi infection. After this early period, infected patients are usually found to develop IgG antibodies. A positive IgM test alone is not recommended for use in determining active disease in persons with illness of longer than one month.
The & Borrelia burgdorferi IgM ELISA Kit is an enzyme-finked immunosorbent assay (ELISA) for the detection of IgM to Borrelia burgdorferi antigen in human serum.
Here's a breakdown of the acceptance criteria and study information for the Borrelia burgdorferi IgM ELISA Kit, based on the provided text:
Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in a quantitative or numerical format (e.g., "Sensitivity must be >X%"). Instead, it presents performance data for comparison against other methods or expected clinical behavior. For example, it provides agreement percentages with characterized serum panels and precision values.
Therefore, the table below will summarize the provided performance characteristics rather than predefined acceptance criteria. The implied acceptance is that these performance characteristics are adequate for the intended use and comparable to similar devices (as suggested by the substantial equivalence to "Zeus Lyme Elisa").
| Performance Metric | Reported Device Performance |
|---|---|
| Clinical Sensitivity (CDC Panel) | 40.4% agreement (equivocals considered positive) overall |
| > 1 Yr from onset | 12.5% agreement |
| 3 - 12 Months from onset | 45.0% agreement |
| 1 - 2 Months from onset | 33.3% agreement |
| < 1 Month from onset | 20.0% agreement |
| Clinical Specificity (CDC Panel) | 100.0% agreement in negatives (5/5) |
| Clinical Sensitivity (Wisconsin Panel) | 81.9% agreement (equivocals considered positive) overall |
| > 1 Yr from onset | 66.7% agreement |
| 3 - 12 Months from onset | 84.6% agreement |
| 1 - 2 Months from onset | 69.2% agreement |
| < 1 Month from onset | 86.1% agreement |
| Prospective Study (EIA Pos. or Equiv.) | 2.89% (5/173) |
| Prospective Study (EIA Pos. or Equiv. and Western Blot Pos.) | 0.58% (1/173) |
| Prospective Study (% Western Blot Pos. among EIA Pos. or Equiv.) | 20% (1/5) |
| Intra-assay Precision (CV) | Range: 9.89% - 17.23% (across sites, positive and negative sera) |
| Inter-assay Precision (CV) | Range: 11.23% - 30.41% (across sites, positive and negative sera) |
| Cross-Reactivity | Mostly 0 positive/equivocal results in various cross-reactive panels (1 equivocal in RPR+, 1 equivocal/1 positive in EBV+) |
| Manual vs. MAGO Plus Correlation | Pearson Correlation Coefficient: 0.973 |
| MAGO Plus Intra-assay Precision (CV) | Range: 6.70% - 35.14% (across runs, positive and negative sera) |
| MAGO Plus Inter-assay Precision (CV) | Range: 10.72% - 23.79% (across runs, positive and negative sera) |
Study Information
-
Sample size used for the test set and the data provenance:
- Clinical Sensitivity and Specificity (CDC Panel):
- Test set size: 47 sera (42 Lyme disease, 5 normal).
- Data provenance: Characterized sera obtained from the CDC (Centers for Disease Control and Prevention). This suggests the data may be retrospective, well-characterized samples collected for reference.
- Clinical Sensitivity (Wisconsin Panel):
- Test set size: 72 sera with a clinical diagnosis of Lyme disease.
- Data provenance: Characterized sera obtained from a clinical lab in Wisconsin. Likely retrospective.
- Prospective Sample Study:
- Test set size: 173 prospective sera.
- Data provenance: Sera from patients of various ages and gender from an endemic area, submitted to a clinical laboratory for B. burgdorferi antibody testing. This indicates prospective collection.
- Precision Studies (Manual and MAGO Plus):
- 6 sera (4 positive, 2 negative) tested 10 times each in 3 different runs at 3 different sites (for manual precision).
- 6 sera tested 10 times each in 3 different runs (for MAGO Plus precision).
- Data provenance: Not explicitly stated, but likely laboratory-prepared or reference sera.
- Specificity with Potentially Cross-Reactive Sera:
- N varies for each condition, totaling around 71 sera.
- Data provenance: Sera with specific laboratory results that may cross-react or interfere.
- Correlation of Manual and MAGO Plus Results:
- Test set size: 294 sera.
- Data provenance: Not explicitly detailed, but likely a mix of clinical samples.
- Clinical Sensitivity and Specificity (CDC Panel):
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number of experts or their qualifications for establishing the ground truth.
- For the CDC panel, it states "characterized sera obtained from the CDC," implying a robust characterization process, but details on experts are absent.
- For the Wisconsin panel, it mentions "characterized sera obtained from a clinical lab" with a "clinical diagnosis of Lyme disease," again, without specific expert details.
- For the prospective study, "Western blot method" was used as a supplemental second-step, which is a standardized diagnostic procedure, but expert interpretation of these results beyond the standard protocol is not described.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not describe an adjudication method for establishing the primary ground truth of the test sets. Ground truth appears to be based on pre-characterized panels (CDC, Wisconsin) or a combination of clinical context and Western Blot results (prospective study).
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done, as this is a serological assay (ELISA) and not an AI-assisted diagnostic device requiring human reader interpretation in the context described by MRMC.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the performance characteristics (clinical sensitivity/specificity, precision, cross-reactivity) are all standalone performance metrics of the ELISA kit itself. The "MAGO PLUS Automated EIA Processor" represents an automated method, and its correlation and precision studies are also standalone evaluations of the device's performance when automated.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- CDC Panel: "Clinical diagnosis of Lyme disease" for positives, and "normal sera" for negatives. This suggests a combination of clinical findings, potentially confirmed by other reference tests at the CDC, forming a characterized reference standard.
- Wisconsin Panel: "Clinical diagnosis of Lyme disease." This is primarily a clinical diagnosis standard.
- Prospective Study: A two-step diagnostic algorithm where equivocal or positive ELISA results were supplemented by a standardized Western blot method. "Positive supplemental results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease." This forms a hybrid ground truth involving the primary test, a confirmatory test, and clinical context.
-
The sample size for the training set:
- The document does not describe a separate "training set" in the context of machine learning or AI. This is a traditional diagnostic kit, and the performance studies evaluate the kit directly using characterized and prospective samples. There is no mention of an algorithm being "trained."
-
How the ground truth for the training set was established:
- As there is no mention of a training set for an algorithm, this question is not applicable.
Ask a specific question about this device
(63 days)
For the qualitative presumptive detection of total (IgG/IgM) antibodies to Borrelia burgdorferi in human serum. This ELISA should only be used for patients with signs and symptoms that are consistent with Lyme disease. Equivocal or positive results must be supplemented by testing with a standardized Western blot procedure. Positive supplemental results are supportive evidence of exposure to B. burgdorferi and can be used to support a clinical diagnosis of Lyme disease. The test can be performed either manually or in conjunction with the MAGO TM PLUS Automated EIA Processor.
The & Borrelia burgdorferi IgG/IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG/IgM to Borrelia burgdorferi antigen in human serum.
The provided 510(k) summary for the "Borrelia burgdorferi IgG/IgM ELISA Kit" details performance characteristics but does not explicitly state specific acceptance criteria in terms of numerical thresholds for sensitivity, specificity, or precision that the device had to meet for clearance. Instead, it presents the device's performance results across various studies and implies that these results were considered acceptable by the FDA for substantial equivalence to a predicate device.
However, based on the presented data, we can infer the reported device performance.
1. Table of Acceptance Criteria and Reported Device Performance
As specific acceptance criteria were not explicitly stated, the reported performance is presented as demonstrated.
| Performance Metric | Inferred Acceptance Criteria (Implicit) | Reported Device Performance (Is-EIA) |
|---|---|---|
| Clinical Sensitivity (CDC Panel) | Sufficient agreement with characterized sera, particularly for later stages of infection. The predicate device's performance likely served as the benchmark. | 70.2% overall agreement (equivocal treated as positive). Specifically: 100% agreement for >1 Yr, 60% for 3-12 Months, 55.6% for 1-2 Months, 60% for <1 Month. |
| Clinical Sensitivity (Wisconsin Lab) | Sufficient agreement with characterized Lyme sera, particular for later stages of infection. | 80.6% overall agreement (equivocal treated as positive). Specifically: 100% agreement for >1 Yr, 100% for 3-12 Months, 76.9% for 1-2 Months, 74.4% for <1 Month. |
| Clinical Specificity (CDC Panel) | 100% agreement for negative samples. | 100% agreement for 5 known negative samples. |
| Prospective Study (WB Positive) | Demonstrated ability to identify Western Blot positive cases among EIA positive/equivocal results, performing comparably or better than existing methods. | 78.6% of Is-EIA positive/equivocal results were confirmed by Western Blot (11/14), compared to 38.9% for another commercial EIA (14/36). |
| Precision (Intra-assay CV) | Consistency (low coefficient of variation) across multiple runs at different sites. Numerical thresholds for CV are not stated but generally, lower CVs are desirable. | Generally low CVs for positive samples (e.g., Site 1: 5.00% to 10.55%; Site 2: 3.51% to 8.80%; Site 3: 3.64% to 7.01%). Higher CVs for negative (e.g., Site 1: 14.78% to 22.30%). |
| Precision (Inter-assay CV) | Consistency (low coefficient of variation) across different runs and sites. | Generally low CVs for positive samples (e.g., Site 1: 9.62% to 12.52%; Site 2: 8.26% to 11.64%; Site 3: 5.21% to 6.79%). Higher CVs for negative (e.g., Site 1: 19.20% to 20.68%). |
| Cross-Reactivity | Minimal false positives or equivocal results with potentially cross-reactive sera. | Few positive or equivocal results observed with various cross-reactive conditions (e.g., RPR+, ds-DNA+, Lipemic+, EBV+, CMV+ etc.). Specific counts provided in Table 7. |
| Automated vs. Manual Correlation | High correlation coefficient between automated and manual testing methods. | Pearson Correlation Coefficient of 0.991 between manual and MAGO Plus results. |
| Automated Precision (Intra/Inter CV) | Consistency (low coefficient of variation) when performed on the automated platform. | Similar to manual precision, generally low CVs for positive samples (e.g., Intra-assay: 4.39% to 11.53%; Inter-assay: 7.05% to 10.66%). Higher CVs for negative (e.g., Intra-assay: 15.06% to 37.16%; Inter-assay: 26.90% to 35.59%). |
2. Sample Sizes and Data Provenance
-
Clinical Sensitivity and Specificity (Test Set 1 - CDC Panel):
- Sample Size: 47 sera (8 >1 Yr, 20 for 3-12 Months, 9 for 1-2 Months, 5 for <1 Month, 5 Negatives).
- Data Provenance: Obtained from the CDC (Centers for Disease Control and Prevention), tested by Diamedix Corp. Retrospective.
-
Clinical Sensitivity and Specificity (Test Set 2 - Wisconsin Lab Panel):
- Sample Size: 72 sera (3 >1 Yr, 13 for 3-12 Months, 13 for 1-2 Months, 43 for <1 Month).
- Data Provenance: Obtained from a clinical laboratory in Wisconsin, assayed by Diamedix. Retrospective.
-
Prospective Sample Study (Test Set 3):
- Sample Size: 173 prospective sera.
- Data Provenance: From patients of various ages and gender from an endemic area, submitted to a clinical laboratory for B. burgdorferi antibody testing. Prospective.
-
Precision Studies (Manual & Automated):
- Sample Size: 6 sera (4 positive, 2 negative) for each site's manual precision (tested 10 times in 3 runs).
- Sample Size: 6 sera (4 positive, 2 negative) for automated precision (tested 10 times in 3 runs).
- Data Provenance: Not explicitly stated, but likely laboratory-prepared control samples for an in-house study.
-
Specificity with Potentially Cross-Reactive Sera (Test Set 4):
- Sample Size: Total not explicitly given, but individual groups are: 20 RPR+, 15 ds-DNA+, 5 RF+, 5 Lipemic+, 5 Bilirubin+, 4 Elevated ESR, 5 Elevated CRP, 7 EBV+, 6 CMV+, 4 Rocky MT Spotted Fever (total 76 unique samples, assuming no overlap).
- Data Provenance: Not explicitly stated, likely sourced from clinical laboratories or specialized panels.
-
Correlation of Manual and MAGO Plus Results:
- Sample Size: 296 sera.
- Data Provenance: Not explicitly stated, but implies concurrent testing of patient samples or a diverse set of samples.
3. Number of Experts and Qualifications for Ground Truth
The document does not explicitly state the number of experts or their qualifications for establishing ground truth.
- For the CDC Panel and Wisconsin Lab Panel: The sera are described as "characterized sera" or "sera with a clinical diagnosis of Lyme disease." This implies that the diagnosis was established by clinicians or epidemiologists based on clinical presentation and potentially other laboratory tests, but expert qualifications are not detailed.
- For the Prospective Study: The ground truth for positive and equivocal EIA results was established by "testing with a Western Blot method." The interpretation of Western Blot results usually involves adherence to standardized criteria, which are often overseen by qualified laboratory personnel, but no specific number or qualifications of experts are given.
4. Adjudication Method for the Test Set
- Clinical Sensitivity and Specificity (CDC & Wisconsin Panels): The samples were "characterized." No explicit adjudication method (e.g., 2+1, 3+1) is described for resolving discrepancies in these panels. However, a key point is mentioned: "Note that equivocal samples were considered positive for the above calculations due to the fact that all equivocal samples would be tested by immunoblotting in a 2-step system." This reflects a specific decision logic for the primary analysis.
- Prospective Sample Study: Positive and equivocal results from both EIAs were "supplemented by testing with a Western Blot method." This acts as the confirmatory or adjudicating test for initial EIA findings. No human adjudicators for the Western Blot results are mentioned; it's implied that the Western Blot itself serves as the definitive test in a two-step algorithm.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This device is an ELISA kit, which is a laboratory assay for detecting antibodies, not an imaging or diagnostic AI algorithm that humans interact with to make decisions. Therefore, the concept of "human readers improving with AI vs without AI assistance" is not applicable to this type of device.
6. Standalone Performance Study
- Yes, a standalone performance study was done. The entire clinical sensitivity and specificity evaluation, prospective sample study, precision study, and cross-reactivity study demonstrate the performance of the "Borrelia burgdorferi IgG/IgM ELISA Kit" as a standalone diagnostic assay. The results presented are exclusively based on the device's output. The "correlation of Manual and MAGO Plus Results" section also shows the device's performance in both manual and automated standalone modes.
7. Type of Ground Truth Used
- Expert Consensus/Clinical Diagnosis/Reference Method (with further confirmation):
- Clinical Sensitivity/Specificity Panels (CDC & Wisconsin): "Characterized sera" with a "clinical diagnosis of Lyme disease" and "normal sera." This implies a ground truth established by clinical diagnosis and possibly backed by other diagnostic markers.
- Prospective Study: The primary ground truth for confirmation for an EIA result was a standardized Western Blot procedure. This is a commonly accepted reference method for Lyme disease serology in a two-tier testing algorithm.
- Specificity with Potentially Cross-Reactive Sera: Ground truth was based on the "laboratory results" for specific conditions (e.g., RPR+, ds-DNA+), meaning a confirmed diagnosis of those conditions or positive results on tests for those conditions.
8. Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning or AI development. As this is an ELISA kit (a chemical/biological assay), the concept of a separate training set as used in AI development is not directly applicable.
However, if "training set" is implicitly referring to the samples used during the development and optimization of the assay before formal clinical validation, that information is not provided in this 510(k) summary. The presented data falls under validation and verification studies.
9. How the Ground Truth for the Training Set Was Established
As no explicit "training set" distinct from the validation/test sets is mentioned for an AI/machine learning context, the method for establishing ground truth for such a set is not detailed. The ground truth for the validation sets (as described in point 7) was established through clinical characterization and standardized Western Blot testing.
Ask a specific question about this device
(59 days)
The TG IgG kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection and quantitation of antibodies against the Thyroglobulin (TG) antigen in serum as an aid in the diagnosis of thyroid autoimmune disease. The test can be performed either manually or in conjunction with the Mago Plus automated ETA processor.
The TG IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG to Thyroglobulin in human serum.
Here's a breakdown of the acceptance criteria and the study details for the anti-TG IgG ELISA Kit, based on the provided 510k summary:
1. Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" as a set of pre-defined thresholds that the device must meet in a table. Instead, it presents performance characteristics from studies that demonstrate its capabilities. I will infer the implied acceptance criteria from the reported values, particularly in comparison to a commercially available ELISA kit mentioned in the "Relative Sensitivity and Specificity" section.
| Performance Characteristic | Implied Acceptance Criteria (Inferred) | Reported Device Performance (anti-TG IgG ELISA Kit) |
|---|---|---|
| Relative Sensitivity | High sensitivity to detect positive cases (implied: comparable to |
| a benchmark kit, with a lower bound within the 95% CI that indicates acceptable detection) | 83.3% (95% CI: 67.2-93.0) |
| Relative Specificity | High specificity to correctly identify negative cases (implied: comparable to a benchmark kit, with a lower bound within the 95% CI that indicates acceptable rejection of negatives) | 97.2% (95% CI: 93.5-99.1) |
| Overall Agreement | High overall agreement with a benchmark kit (implied: overall performance nearing 90% or above, with a lower bound within the 95% CI that indicates acceptable consistency) | 94.8% (95% CI: 90.9-97.4) |
| Clinical Specificity (Normals) | High specificity in a normal population. Specific numerical criteria are not specified, but a high percentage is expected to avoid false positives in healthy individuals. | 97.7% (95% CI: 94.3-99.4) |
| Clinical Sensitivity (Grave's Disease) | Moderate to high sensitivity in Grave's Disease patients; while not an explicit threshold, the reported value (38.5%) should be considered clinically acceptable for an aid in diagnosis based on the device's role. | 38.5% (95% CI: 13.9-68.4) |
| Clinical Sensitivity (Hashimoto's Disease) | Moderate to high sensitivity in Hashimoto's Disease patients; similar to Grave's, the reported value (64.3%) should be acceptable for diagnostic aid. | 64.3% (95% CI: 35.1-87.2) |
| Precision (Intra-assay & Inter-assay CV%) | Low Coefficient of Variation (CV%) for both positive and negative samples across different runs and sites. Although no explicit threshold is given, typically CVs below 15-20% are considered acceptable, especially for quantitative assays. | Max Intra-assay CV%: 61.85% (F, NEG, Site 1, Run 1) Max Inter-assay CV%: 98.47% (F, NEG, Site 1) (Values for positive samples generally much lower, e.g., ~10-15%) |
| Correlation (Manual vs. MAGO Plus) | High correlation between manual and automated methods. The Pearson Correlation Coefficient (r) should be close to 1. | 0.988 (Pearson Correlation Coefficient) |
2. Sample Size and Data Provenance:
- Relative Sensitivity and Specificity:
- Test Set Sample Size: 227 (Frozen retrospective sera). The table provided specifies N=213 for calculation after excluding equivocal results.
- Data Provenance: Retrospective sera. Country of origin is not specified but is implied to be relevant to the market the device is seeking approval for, likely the US.
- Clinical Sensitivity and Specificity:
- Test Set Sample Size: 207 (Frozen retrospective, clinically characterized sera). The table breaks this down into:
- Normals: 176 (175 used for calculation)
- Grave's Disease: 12 (13 used for calculation)
- Hashimoto's Disease: 16 (14 used for calculation)
- Data Provenance: Retrospective sera. Clinically characterized. Country of origin not specified.
- Test Set Sample Size: 207 (Frozen retrospective, clinically characterized sera). The table breaks this down into:
- Precision:
- Test Set Sample Size: 6 sera (4 positive, 2 negative) tested 10 times in 3 different runs at 2 different sites.
- Data Provenance: Not specified; likely laboratory samples.
- Correlation of Manual and MAGO Plus Results:
- Test Set Sample Size: 91 serum samples.
- Data Provenance: Not specified; likely laboratory samples.
- MAGO Plus Precision:
- Test Set Sample Size: 6 sera (implied same a precision study data set).
- Data Provenance: Not specified; likely laboratory samples.
3. Number of Experts and Qualifications for Ground Truth:
- The document does not specify the number of experts used to establish the ground truth or their qualifications for any of the studies.
- For the "Relative Sensitivity and Specificity" study, the comparison is made to "a commercially available ELISA KIT." The "ground truth" here is essentially the results of this comparator kit.
- For the "Clinical Sensitivity and Specificity" study, the "ground truth" for patient groups (Normals, Grave's Disease, Hashimoto's Disease) is established through "clinically characterized sera." How this clinical characterization was performed (e.g., by a panel of endocrinologists, by a single specialist for each patient, based on a gold standard diagnostic) is not detailed.
4. Adjudication Method for the Test Set:
- The document does not explicitly describe an adjudication method for the test set in the context of expert review.
- For the "Relative Sensitivity and Specificity" study, "Equivocal results were excluded from calculations." This acts as a form of exclusion, but not a multi-expert adjudication to reconcile discrepant results.
- For "Clinical Sensitivity and Specificity," "Equivocal results were excluded from calculations" as well.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) ELISA kit, not an imaging or interpretation device that would involve multiple human readers. The studies focus on the analytical and clinical performance of the assay itself.
6. Standalone Performance Study:
- Yes, standalone performance was done for the algorithm (the ELISA kit). The primary studies (Relative Sensitivity and Specificity, Clinical Sensitivity and Specificity, Precision) all evaluate the performance of the anti-TG IgG ELISA Kit in isolation. The "Correlation of Manual and MAGO Plus Results" and "MAGO Plus Precision" evaluate its performance when integrated with an automated system, but the core analytical performance is still inherent to the kit itself.
7. Type of Ground Truth Used:
- For Relative Sensitivity and Specificity: The "ground truth" was established by comparison to a "commercially available ELISA KIT" designed for detecting anti-TG IgG antibodies. This is a comparator device ground truth.
- For Clinical Sensitivity and Specificity: The "ground truth" was based on clinical characterization of the patient sera (categorized as Normals, Grave's Disease, Hashimoto's Disease). This can be considered clinical diagnosis/outcomes data ground truth, although the specific diagnostic criteria and methods for this characterization are not elaborated upon.
- For Precision, Correlation, and MAGO Plus Precision: The ground truth is the inherent concentration or presence/absence of the analyte in the tested sera, with consistency and correlation being the metrics of interest against established values or the manual method.
8. Sample Size for the Training Set:
- The document does not specify a separate "training set" for the ELISA kit. ELISA kits are typically developed and optimized through iterative R&D processes, but the performance studies described here are for validation rather than training a machine learning model. If any optimization involved specific sample sets, it is not detailed as a formal "training set" in this summary.
9. How Ground Truth for the Training Set Was Established:
- As a formal "training set" is not described, the method for establishing its ground truth is not applicable/not provided in this summary.
Ask a specific question about this device
(142 days)
For the qualitative determination of IgM antibodies in human serum to Epstein Barr (recombinant) early antigen diffuse (EA-D) antigen. The & EA-D IgM assay should be used in conjunction with other Epstein-Barr serologies (VCA IgM, VCA IgG, EBNA-1 IgG, EBNA-1 IgM, EA-D IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis. The test can be performed either manually or in conjunction with the MAGO PLUS The Automated EIA Processor.
The & EA-D IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgM antibodies to Epstein-Barr Early antigen diffuse in human serum.
Here's an analysis of the provided text regarding the acceptance criteria and study for the EA-D IgM ELISA Kit:
The document does not explicitly state pre-defined acceptance criteria for the device's performance. Instead, it presents the results of its performance characteristics study. The acceptance for market clearance seems to be based on "substantial equivalence" to legally marketed predicate devices, implying that the observed performance falls within an acceptable range compared to those predicates, even if specific numerical targets aren't listed here.
However, we can infer the "reported device performance" from the study results.
1. A table of acceptance criteria and the reported device performance
As stated, explicit acceptance criteria are not presented in the document. The table below outlines the reported performance from the study, which would have been evaluated against a benchmark for substantial equivalence.
| Performance Metric | Reported Device Performance | Comments |
|---|---|---|
| Clinical Performance (vs. Serological Status) | ||
| Relative Specificity (Convalescent) | 94.9% (95% CI: 88.5-98.3%) | Based on 93/98 sera |
| Relative Sensitivity (Current Infection) | 45.9% (95% CI: 29.5-63.1%) | Based on 17/37 sera |
| Relative Specificity (Seronegative) | 96.7% (95% CI: 82.8-99.9%) | Based on 29/30 sera |
| Overall Agreement | 84.2% (95% CI: 78.7-89.8%) | Based on 139/165 sera (equivocal excluded) |
| Precision (Internal Site #1) | Coefficient of Variation (CV%) | |
| Intra-Assay CV% (Positive samples) | 4.53% - 12.81% | Range across 4 positive sera and 3 runs |
| Intra-Assay CV% (Negative samples) | 35.23% - 46.08% | Range across 2 negative sera and 3 runs |
| Inter-Assay CV% (Positive samples) | 5.35% - 11.20% | Range across 4 positive sera |
| Inter-Assay CV% (Negative samples) | 34.16% - 42.74% | Range across 2 negative sera |
| Inter-Site Precision | Coefficient of Variation (CV%) | |
| Inter-Site CV% (Positive samples) | 9.10% - 17.32% | Range across 4 positive sera (n=90) |
| Inter-Site CV% (Negative samples) | 30.28% - 31.41% | Range across 2 negative sera (n=90) |
| Correlation (Manual vs. MAGO Plus) | ||
| Pearson's Correlation Coefficient | 0.975 | Between manual and MAGO Plus results (128 samples) |
| Specificity (Cross-Reactivity) | Some expected | Non-reactive for anti-EA-D IgM in VZV, CMV, HSV positive sera; suggests some cross-reactivity may occur with these analytes. |
| MAGO Plus Precision (Site #2) | Coefficient of Variation (CV%) | |
| Intra-Assay CV% (Positive samples) | 3.76% - 16.10% | Range across 4 positive sera and 3 runs |
| Intra-Assay CV% (Negative samples) | 0.00% - 37.16% | Range across 2 negative sera and 3 runs |
| Inter-Assay CV% (Positive samples) | 4.35% - 11.76% | Range across 4 positive sera |
| Inter-Assay CV% (Negative samples) | 20.86% - 30.51% | Range across 2 negative sera |
2. Sample sized used for the test set and the data provenance
- Test Set Sample Size: 176 patient sera were used for the clinical performance study.
- Data Provenance: The data used was from retrospective sera from patients. The country of origin is not explicitly stated, but given the applicant and reporting of a 510(k) to the FDA, it is highly likely to be from the USA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The ground truth for the test set was established by characterizing the 176 frozen retrospective sera using "commercially available kits for VCA IgM, VCA IgG, EBNA IgG and heterophile antibodies."
- Number of Experts: Not applicable, as the "ground truth" (or reference standard) was established using other commercial serological assays, not by expert interpretation.
- Qualifications of Experts: Not applicable.
4. Adjudication method for the test set
The concept of an "adjudication method" as typically applied in scenarios involving multiple human readers (e.g., 2+1, 3+1 for discrepancies) is not relevant here. The ground truth for the test set was determined by the results of other commercial serological assays, which are considered objective measurements, not subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study Done: No, an MRMC comparative effectiveness study was not done.
- Effect Size of Human Readers Improvement with/without AI: Not applicable, as this is an in vitro diagnostic device for antibody detection, not an imaging analysis or interpretive AI system that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the primary clinical performance study (Section A) evaluates the device in standalone mode. The results of the Is-EBV-EA-D IgM Test Kit were directly compared to the classification of patient sera based on other commercial EBV serologies. While a technician performs the test, the result (positive, negative, equivocal) is generated by the kit itself, making it a standalone assessment of the device's diagnostic capability. The section on manual vs. MAGO Plus correlation also demonstrates standalone performance for both methods.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the clinical performance study was established using a combination of commercially available serological assays for Epstein-Barr Virus (EBV) antibodies. This allowed for the classification of sera into "convalescent," "current infection," and "seronegative" categories based on established serological patterns for EBV infection. It's a form of clinical reference standard based on established serology.
- Specifically:
- Convalescent (past infection): Positive for VCA IgG and/or EBNA IgG antibodies; negative for VCA IgM and heterophile antibody.
- Seronegative: Negative for VCA IgG, VCA IgM, EBNA IgG, and heterophile antibody.
- Current (recent) infection: Positive for VCA IgM and/or heterophile antibody; negative for EBNA IgG.
8. The sample size for the training set
The document does not explicitly describe a separate "training set" for the development of the EA-D IgM ELISA Kit. The provided studies focus on the performance evaluation of the developed device. For in vitro diagnostic kits, the "training" (development and optimization) phase is usually internal and not detailed in 510(k) summaries in the same way as machine learning models. Therefore, the sample size for a training set (in the ML sense) is not reported.
9. How the ground truth for the training set was established
As there is no distinct "training set" described in the context of machine learning model development, the method for establishing its ground truth is not applicable/not reported in this 510(k) summary. The development of an ELISA kit involves biochemical optimization and formulation, rather than algorithmic training on labeled data in the way an AI/ML model would.
Ask a specific question about this device
(142 days)
For the qualitative determination of IgM antibodies in human serum to Epsien Barr Virus (recombinant) Nuclear antigen (EBNA-1) antigen. The 78 EBNA-1 IgM assay should be used in conjunction with other Epstein-Barr serologies (VCA IgM, VCA IgG, EA-D IgG, EBNA-1 IgG and heterophile) as an aid in the diagnosis of infectious mononucleosis. The test can be performed either manually or in conjunction with the MAGO PLUS TM Automated EIA Processor.
The & EBNA-1 IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgM antibodies to Epstein-Barr Nuclear antigen-1 in human serum.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Clinical Performance | Relative Specificity (Convalescent): High accuracy in identifying individuals who have had a past EBV infection (negative for current infection markers). | 97.0% (95% CI: 91.6-99.4) |
| Relative Sensitivity (Current Infection): High accuracy in identifying individuals with a current (recent) EBV infection (positive for current infection markers). | 59.5% (95% CI: 42.1-75.3) | |
| Relative Specificity (Seronegative): High accuracy in identifying individuals who have never been infected with EBV (negative for all EBV markers). | 100.0% (95% CI: 89.1-100.0) | |
| Overall Agreement: General concordance with the established EBV serological status across all characterized sera. | 89.4% (95% CI: 84.8-94.0) | |
| Precision | Intra-Assay and Interassay Precision (Site #1, 2, 3): Consistent results within a single run and across multiple runs at individual sites for positive and negative sera, indicating reliability and reproducibility of the assay. (Implicitly, CV% values should be within an acceptable range, although specific numerical acceptance criteria are not explicitly stated. The provided CV% values are reported.) | Site #1: Interassay CV% range for positive sera: 4.85-10.00%; negative sera: 16.74-44.75%. Site #2: Interassay CV% range for positive sera: 5.75-6.51%; negative sera: 15.67-34.34%. Site #3: Interassay CV% range for positive sera: 4.43-22.09%; negative sera: 23.91-62.45%. |
| Inter-Site Precision: Consistent results across different testing sites, indicating robustness of the assay across various laboratory environments. (Implicitly, CV% values should be within an acceptable range.) | Inter-site CV% range for positive sera: 6.80-14.33%; negative sera: 10.37-39.85%. | |
| Cross-Reactivity | Specificity with Potentially Cross-Reactive Sera: Minimal false positives when tested against samples known to be positive for other common viral infections (Varicella Zoster, Cytomegalovirus, Herpes Simplex Virus IgM). The expectation is that the EBNA-1 IgM Test Kit should not react with these other antibodies. (Implicitly, a low number of cross-reactive results is desired.) | Some cross-reactivity expected: 2/5 anti-CMV IgM positive sera were non-reactive for anti-EBNA-1 IgM, 4/4 anti-VZV IgM positive sera were non-reactive, and 4/4 anti-HSV positive sera were non-reactive. This phrasing suggests some reactivity with CMV was observed, but the non-reactive counts imply a level of specificity. |
| Automation Equivalence | Correlation of Manual and MAGO Plus Results: Strong positive correlation between results obtained manually and those obtained using the MAGO Plus™ Automated EIA Processor, demonstrating that the automated system provides equivalent results to the manual method. (Implicitly, high Pearson Correlation Coefficient and close agreement.) | Pearson Correlation Coefficient of 0.985 with a linear regression of Y = 0.0186 + 0.9826 X. |
| MAGO Plus Precision: Consistent results (intra-assay and interassay precision) when the assay is performed on the MAGO Plus Automated EIA Processor. (Implicitly, CV% values should be within an acceptable range for automated use). | Interassay CV% range for positive sera: 8.11-10.15%; negative sera: 6.88-66.58%. |
Study Details
-
Test Set Sample Size and Data Provenance:
- Sample Size: 176 sera.
- Data Provenance: Frozen retrospective sera from patients.
- Country of Origin: Not specified.
-
Experts Used for Ground Truth and Qualifications:
- The text does not specify the number of experts used to establish the ground truth or their qualifications.
- The "ground truth" was established by characterizing the sera using commercially available kits for VCA IgM, VCA IgG, EBNA-1 IgG, and heterophile antibodies, and then classifying them into "convalescent," "current infection," or "seronegative" based on the results of this testing.
-
Adjudication Method for the Test Set:
- The text does not describe an adjudication method for the test set. The characterization of sera appears to be based on the results of multiple commercial assays rather than a human consensus process.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic test, and the study evaluated its performance against established serological markers, not against human readers with and without AI assistance.
-
Standalone (Algorithm Only) Performance:
- Yes, a standalone performance study was done. The entire clinical performance section evaluates the device (the EBNA-1 IgM ELISA Kit) directly against the characterized sera without human-in-the-loop assistance for the interpretation of the EBNA-1 IgM ELISA results. Human interpretation was part of the "ground truth" establishment using other assays, but the device's output (positive, negative, equivocal) was directly compared.
-
Type of Ground Truth Used:
- Expert Consensus (indirectly) / Reference Comparator Assay Results: The ground truth was established by classifying patient sera based on the results of multiple commercially available EBV serology kits (VCA IgM, VCA IgG, EBNA-1 IgG, and heterophile antibodies). This is a form of reference standard based on accepted laboratory diagnostics.
-
Training Set Sample Size:
- The document does not specify a training set sample size. This is typical for traditional ELISA kits, where optimization and assay development are performed, but a distinct "training set" for an algorithm in the modern AI sense is not applicable. The device's parameters (e.g., cut-offs) would have been established during its development phase.
-
How Ground Truth for Training Set Was Established:
- As there is no distinct "training set" in the AI sense for this device, the concept of establishing ground truth for it is not directly applicable. The development of the assay (e.g., determining optimal concentrations, incubation times, and cut-off values) would have involved extensive experimentation and standardization using characterized samples, similar to how the ground truth for the test set was established using comparator assays.
Ask a specific question about this device
(142 days)
For the qualitative determination of IgM antibodies in human serum to Epstein Barr (recombinant) Viral Capsid antigen (EBV-VCA) antigen. The EBV-VCA IgM assay should be used in conjunction with other Epstein-Barr serologies (EBV-VCA IgG, EBNA-1 IgG, EA-D IgG, EA-D IgM, EBNA-1 IgM and heterophile) as an aid in the diagnosis of infectious monomucleosis. The test can be performed either manually or in conjunction with the MAGO PLUS ™ Automated EIA Processor.
The & EBV-VCA IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgM antibodies to Epstein-Barr Viral Capsid antigen in human serum.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
Device Name: EBV-VCA IgM ELISA Kit (also referred to as Is-EBV-VCA IgM Test Kit and 7 VCA IgM ELISA)
| Acceptance Criteria Category | Specific Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|---|
| Clinical Performance | Relative Specificity (Convalescent) | High specificity (specific threshold not explicitly stated but implied by high performance) | 97.0% (95% CI: 91.6-99.4) |
| Relative Sensitivity (Current Infection) | High sensitivity (specific threshold not explicitly stated but implied by high performance) | 95.0% (95% CI: 83.1-99.4) | |
| Relative Specificity (Seronegative) | High specificity (specific threshold not explicitly stated but implied by high performance) | 100% (95% CI: 89.1-100) | |
| Overall Agreement (with predicate/characterized sera) | High agreement (specific threshold not explicitly stated but implied by high performance) | 97.1% (95% CI: 93.4-99.1) | |
| Precision | Intra-assay Precision (CV%) | Low variability (threshold not explicitly stated) | Site #1: Positive sera CV% = 6.07-9.92; Negative sera CV% = 8.51-49.37 (higher for negatives, especially low index ones)Site #2: Positive sera CV% = 3.95-6.35; Negative sera CV% = 7.41-17.42Site #3: Positive sera CV% = 4.15-6.22; Negative sera CV% = 7.16-12.82MAGO Plus: Positive sera CV% = 3.53-9.27; Negative sera CV% = 0.00-22.59 (0.00 for one low negative, likely due to rounding or being below detection) |
| Interassay Precision (CV%) | Low variability (threshold not explicitly stated) | Site #1: Positive sera CV% = 12.05-13.69; Negative sera CV% = 22.75-39.31Site #2: Positive sera CV% = 6.53-7.34; Negative sera CV% = 9.38-14.80Site #3: Positive sera CV% = 5.08-5.80; Negative sera CV% = 12.70-13.26MAGO Plus: Positive sera CV% = 4.55-8.64; Negative sera CV% = 17.76-0.00 (0.00 for one low negative, likely due to rounding or being below detection) | |
| Inter-Site Precision (CV%) | Low variability (threshold not explicitly stated) | Positive sera CV% = 11.13-13.43; Negative sera CV% = 20.69-28.77 (NC has 60.11%, likely due to very low index values making CV% highly sensitive) | |
| Cross-Reactivity | Specificity with Potentially Cross-Reactive Sera | Minimal cross-reactivity expected | Some cross-reactivity is suggested: 1/4 anti-VZV IgM positive sera were reactive for anti-VCA IgM; 2/5 anti-CMV IgM positive sera were reactive for anti-VCA IgM. 4/4 anti-HSV positive sera were non-reactive. Results indicate that some specific cross-reactivity should be expected with VZV and CMV. |
| Method Correlation | Manual and MAGO Plus Results Correlation | Strong correlation (specific threshold not explicitly stated but implied by high performance) | Pearson Correlation Coefficient = 0.990 (indicating strong positive correlation) |
Study Details Proving Acceptance Criteria
-
Sample Size used for the test set and the data provenance:
- Clinical Sensitivity and Specificity: 176 frozen retrospective sera. The provenance (country of origin) is not explicitly stated, but it's referred to as "sera from one hundred and seventy-six patients" which suggests human samples. The study explicitly states it was a "pre-selected, retrospective, population."
- Precision:
- Intra-assay and Interassay Precision: Four positive and two negative sera (6 unique sera) were assayed 10 times each, in 3 different runs. This was done at 3 different sites.
- Inter-Site Precision: The document mentions n=90 for serum samples A-F (presumably 90 total measurements across sites for each serum type, or 3 sites * 3 runs * 10 replicates = 90 total for each serum type), and n=36 for CAL (Calibrator), n=18 for LPC (Low Positive Control), n=18 for NC (Negative Control).
- Specificity with Potentially Cross-Reactive Sera: 13 sera.
- Correlation of Manual and MAGO Plus Results: 120 serum samples.
- MAGO Plus Precision: Six sera were assayed 10 times each in 3 different runs (similar to the manual precision study).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For the "Clinical Sensitivity and Specificity Using Characterized Sera" section, the sera were "characterized using commercially available kits for VCA IgM, VCA IgG, EBNA IgG and heterophile antibodies." This implies an established laboratory testing method rather than human expert opinion directly establishing ground truth for individual cases. No human experts are explicitly mentioned for ground truth establishment.
-
Adjudication method for the test set:
- No adjudication method (e.g., 2+1, 3+1) is mentioned. The ground truth was established by comparing to results from "commercially available kits" for other EBV serologies.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This is not an MRMC study. The device is an ELISA kit, which is an in vitro diagnostic (IVD) test, not an AI-assisted diagnostic tool for human readers. There is no mention of human readers or AI assistance in this context. The comparison is between the new EBV-VCA IgM ELISA Kit and existing commercial kits/methods.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, for the EBV-VCA IgM ELISA Kit, the performance data provided (sensitivity, specificity, precision, cross-reactivity) represents the standalone performance of the diagnostic assay. This assay is a laboratory test, and its results are read and interpreted, but the performance characteristics apply to the kit itself.
- The "Correlation of Manual and MAGO Plus Results" section compares the standalone performance of the kit when processed manually versus using an automated EIA processor (MAGO Plus). Both are "standalone" in the sense of the chemical assay's performance.
-
The type of ground truth used:
- Clinical Performance: "EBV Serological Status" derived from other "commercially available kits for VCA IgM, VCA IgG, EBNA IgG and heterophile antibodies." This is a lab result-based ground truth (or predicate method comparison) rather than pathology, clinical outcome, or expert consensus on a subjective measure. The document explicitly states: "There was not an attempt to correlate the assay's results with disease presence or absence."
- Precision: Ground truth for positive/negative samples was pre-established (e.g., "four positive and two negative sera").
- Cross-Reactivity: Sera "reactive for IgM antibodies to varicella zoster, cytomegalovirus and herpes simplex virus by EIA". This is again a lab result-based ground truth to check for specific antibody interference.
- Correlation: Comparison between the manual and MAGO Plus results on the same samples.
-
The sample size for the training set:
- The document does not explicitly describe a "training set" in the context of machine learning or AI. This is an in vitro diagnostic (IVD) assay. The development of such assays typically involves research and development where reagents and protocols are optimized. The "Characterized Sera" mentioned in the clinical performance section serves as the test set for the final validation of the device.
-
How the ground truth for the training set was established:
- As no "training set" in the AI sense is explicitly mentioned, this question is not fully applicable. For the development and optimization of the ELISA kit itself, ground truth for sample characterization would have been established through well-characterized reference materials, clinical samples with known serological status using predicate methods, or samples from a disease registry. The "Characterized Sera" in the performance study were characterized using "commercially available kits for VCA IgM, VCA IgG, EBNA IgG and heterophile antibodies," which indicates the method used to establish the "ground truth" for those specific samples relative to established serological definitions.
Ask a specific question about this device
(27 days)
The & PR-3 IgG kit is an Enzyme-Linked Immunosorbent Assay (ELISA) For the qualitative detection and semi-quantitation of antibodies against the Proteinase-3 antigen in serum as an aid in the diagnosis of Wegeners granulomatosis. The test can be performed either manually or in conjunction with the Mago Plus automated EIA processor.
The & PR-3 IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG to Proteinase-3 in human serum.
Here's a breakdown of the acceptance criteria and the study details for the Columbia Bioscience, Inc. & PR-3 IgG ELISA Kit based on the provided 510(k) summary:
Acceptance Criteria and Device Performance
The acceptance criteria are implied by the results presented in the performance studies, showing that the device achieved these values. Explicit pre-defined acceptance thresholds are not directly stated in the summary, but the reported performance serves as the basis for substantial equivalence.
Table 1: Acceptance Criteria (Implied) and Reported Device Performance
| Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Relative Sensitivity | High (e.g., >90%) | 100.0% (95% CI: 93.7-100%) |
| Relative Specificity | High (e.g., >95%) | 97.9% (95% CI: 94.7-99.4) |
| Overall Agreement | High (e.g., >95%) | 98.4% (95% CI: 95.9-99.6) |
| Clinical Specificity (Normals) | High (e.g., >95%) | 98.3% (95% CI: 95.0-99.6) |
| Clinical Sensitivity (Wegener's Granulomatosis) | High (e.g., >90%) | 95.0% (95% CI: 83.1-91.4) |
| Clinical Sensitivity (Microscopic Polyangiitis) | Acceptable (Specific target not clear, but 55.3% is reported) | 55.3% (95% CI: 38.3-71.4) |
| Correlation (Manual vs. MAGO Plus) | Strong positive correlation | Pearson Correlation Coefficient: 0.983 |
| Precision (Intra-assay CV%) | Low variability (e.g., <20%) | Site 1: POS sera: 5.94-15.86%, NEG sera: 15.17-57.92% Site 2: POS sera: 5.27-29.87%, NEG sera: 12.95-31.44% |
| Precision (Inter-assay CV%) | Low variability (e.g., <20%) | Site 1: POS sera: 8.60-22.56%, NEG sera: 38.89-50.84% Site 2: POS sera: 7.04-29.87%, NEG sera: 17.26-31.44% |
| MAGO Plus Precision (Intra-assay CV%) | Low variability (e.g., <20%) | POS sera: 7.59-23.20%, NEG sera: 24.00-78.26% |
| MAGO Plus Precision (Inter-assay CV%) | Low variability (e.g., <20%) | POS sera: 10.61-23.50%, NEG sera: 44.80-44.89% |
Note on Acceptance Criteria: For diagnostic kits like ELISA, typical acceptance criteria for sensitivity and specificity are often high (e.g., >90% or >95%), and for precision, Coefficient of Variation (CV%) values are expected to be low (e.g., <20% for positive samples, although negative samples can have higher CV% due to values close to zero). The provided document doesn't explicitly state the pre-defined acceptance criteria but presents the achieved performance as a demonstration of suitability.
Study Details
This submission describes standalone performance studies for the & PR-3 IgG ELISA Kit. It does not appear to involve a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it explicitly mention human readers or AI assistance. The device is an in-vitro diagnostic (IVD) kit for antibody detection.
2. Sample Size and Data Provenance
-
Test Set (Relative Sensitivity and Specificity):
- Sample Size: 256 frozen retrospective sera.
- Data Provenance: Not explicitly stated, but the mention of "frozen retrospective sera" suggests samples collected from various sources over time. No country of origin is specified.
- Retrospective/Prospective: Retrospective.
-
Test Set (Clinical Sensitivity and Specificity):
- Sample Size: 256 frozen retrospective, characterized sera.
- Data Provenance: Not explicitly stated.
- Retrospective/Prospective: Retrospective.
-
Test Set (Precision):
- Sample Size: Six sera (four positive, two negative) assayed ten times each in three different runs. This represents 180 individual tests per site (6 samples * 10 reps * 3 runs).
- Data Provenance: Not specified.
- Retrospective/Prospective: Prospective (lab-based testing of selected sera).
-
Test Set (Correlation of Manual and MAGO Plus Results):
- Sample Size: 90 serum samples.
- Data Provenance: Not specified.
- Retrospective/Prospective: Not specified, but likely prospective testing on a set of existing samples.
-
Test Set (MAGO Plus Precision):
- Sample Size: Six sera assayed ten times each in three different runs. This represents 180 individual tests.
- Data Provenance: Not specified.
- Retrospective/Prospective: Prospective (lab-based testing of selected sera).
3. Number of Experts and Qualifications for Ground Truth
- Relative Sensitivity and Specificity: The "ground truth" for this section was another commercially available ELISA for anti-PR-3 antibodies. No human experts were involved in establishing this "ground truth" as it compares one test kit against another.
- Clinical Sensitivity and Specificity: The "ground truth" was established by "characterized sera" which were classified into "Normals," "Wegener's Granulomatosis," and "Microscopic Polyangiitis" patient groups. The summary does not specify the number of experts, their qualifications, or the method by which these characterizations were made. It's likely these were clinically diagnosed and confirmed cases based on established medical criteria, but details are absent.
4. Adjudication Method for the Test Set
Not applicable. The concept of adjudication by human experts typically applies to image interpretation or clinical diagnosis scenarios where multiple experts review and reconcile differences. For this ELISA kit, the "ground truth" was either another commercial ELISA kit or pre-characterized patient groups.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not conducted. This refers to comparing human reader performance with and without AI assistance, which is not relevant to this IVD (ELISA) kit.
6. Standalone Performance Study (Algorithm Only)
Yes, a standalone study was performed. The entire performance evaluation focuses on the performance of the & PR-3 IgG ELISA Kit itself, either manually or in conjunction with the MAGO Plus automated EIA processor. There is no human-in-the-loop component being evaluated in the reported performance metrics.
7. Type of Ground Truth Used
- Relative Sensitivity and Specificity: The ground truth was based on the results of a "similar assay" (another commercially available ELISA).
- Clinical Sensitivity and Specificity: The ground truth was based on "characterized sera" representing patient groups (Normals, Wegener's Granulomatosis, Microscopic Polyangiitis). This implies a form of clinical diagnosis/classification rather than direct pathology or outcome data, although pathology would likely be part of the diagnosis for these conditions.
8. Sample Size for the Training Set
The document does not describe a "training set" in the context of an algorithm or AI model development. This ELISA kit is a biochemical assay, not a machine learning algorithm. The samples described are for performance validation (test set).
9. How Ground Truth for the Training Set Was Established
Not applicable, as there is no training set described for an algorithm.
Ask a specific question about this device
Page 1 of 2