Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K170423

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-25

    (254 days)

    Product Code
    Regulation Number
    880.6300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DentureID Microchip is intended to enable access to secure patient identification and device information when used with complete dentures, partial dentures and other removable oral appliances.

    Device Description

    DentureID Microchip enables access to secure patent identification and device information when used with removable oral appliances. The microchip is permanently embedded into a denture and is a digital link to owner contact information and information about the denture in the event that a repair or replacement is needed. All of the information is controlled on the secure DentureID.com website by the denture owner and dental professional. The DentureID Microchip can be read by any NFC compatible Android Smart Phone (after downloading the DentureID.com App from Google Play) and holding the phone against the microchip. The information on the DentureID.com website may be modified at any time by the patient or dental professional by entering a username and password. The information that is on the website will appear on the smart phone when reading the DentureID Microchip. DentureID Microchips are classified as RFID ISO 14443 which use NFC (near-field-communication). DentureID Microchips are designed to be read by ISO 14443 NFC-compliant smart phones with DentureID App installed. DentureID Microchips are encased in a medical-grade epoxy resin. The size is 1.5mm X 6 mm in diameter. They are inserted into the buccal flange of a denture and completely covered with self-cure repair resin. The DentureID Microchip does not directly contact the patient.

    AI/ML Overview

    This document describes the non-clinical performance testing of the DentureID Microchip.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test DescriptionStandardAcceptance Criteria (Implied by Result)Reported Device Performance
    Biocompatibility
    CytotoxicityISO 10993-5:2009Non-cytotoxic (score '0')The test article scored '0' at 24, 48, and 72 ± 4 hours and is considered non-cytotoxic under the conditions of this test.
    IrritationISO 10993-10:2010Mean test and control scores of extract dermal observations less than 1.0The differences in the mean test and control scores of the extract dermal observations were less than 1.0, indicating that the requirements of the ISO Intracutaneous Reactivity Test have been met by the test article.
    SensitizationISO 10993-10:2010No sensitization response greater than '0' for test article extractsNone of the negative control animals challenged with the control vehicles were observed with a sensitization response greater than '0'. None of the animals challenged with the test article extracts were observed with a sensitization response greater than '0'. The normal saline extract of the test material had a sensitization response of '0' under valid test conditions. The sesame oil extract of the test material had a sensitization response of '0' under valid test conditions. Under the conditions of this protocol, the test article did not elicit a sensitization response.
    Chemical Characterization of Materials (Leachate)Proprietary standards of Nelson Laboratories and ChemTech Ford LaboratoriesPotential toxicity from leaching of harmful chemicals is not significantThe incremental substances found in the study are presented and compared to toxicity standards for the respective materials. As a result of the study, we feel that the potential toxicity from leaching of harmful chemicals from the DentureID RFID tag is not significant.
    Information Security ProceduresNot specifiedPassedPassed
    Software ValidationNot specifiedPassedPassed
    Migration Testing of Implanted TransponderNot specifiedPassedPassed
    Performance Testing of Implanted TransponderNot specifiedPassedPassed
    Electromagnetic Compatibility (EMC)
    Radiated EmissionsEN 55011:2009PassPass
    Electrostatic Discharge ImmunityEN 61000-4-2:2008PassPass
    Radiated Electromagnetic Field ImmunityEN 61000-4-3:2010PassPass
    Magnetic Field ImmunityEN 61000-4-8:2009PassPass
    Data Integrity (during EMC)Not specifiedNo loss or corruption of data, latency, or throughputNo loss or corruption of the data, latency or through-put, which was coordinated with the electromagnetic compatibility (EMC) performance of the microchip, scanner and wireless data link.
    Magnetic Resonance Imaging Compatibility (MRI)
    Magnetic Field Interactions at 3-TeslaASTM F2052-15No additional risk or hazard to a patient in the 3-Tesla MRI environment or less with regard to torque (qualitative measurement of 0)The qualitatively measured torque at 3-Tesla for the DentureID was 0, no torque. As such, this device will not present an additional risk or hazard to a patient in the 3-Tesla MRI environment or less with regard to torque.
    MRI-related heating, 1.5-Tesla and 3-TeslaASTM F2182-11aMaximum temperature rise matches background (e.g., 1.5°C at 1.5T, 1.9°C at 3T)1.5-Tesla demonstrated a maximum of 1.5°C temperature rise and the 3-T system demonstrated a 1.9°C rise. Both of these temperature rises matched the maximum background temperature rise. In conclusion, MRI 1.5 and 3 do not induce significant heating to DentureID Microchip.
    Artifacts at 3-TeslaASTM F2119-07 (Reapproved 2013)Localized signal voids corresponding to device size and shape; maximum artifact size around 10-mm relative to device size.The artifacts that appeared on the MR images were shown as localized signal voids (i.e., signal loss) that corresponded to the size and shape of this device. The gradient echo pulse sequence produced larger artifacts than the T1-weighted, spin echo pulse sequence for the device. The maximum artifact size (i.e., as seen on the gradient echo pulse sequence) extends approximately 10-mm relative to the size and shape of this device.
    Effects of MRI at 1.5-Tesla and 3-Tesla on FunctionNot specified (Internal investigation protocol)100% pre and post-exposure performanceDentureID Microchips performed 100% pre and post exposures.
    Simulated Wear (Denture Cleaning)Not specified (Internal protocol based on 5-year simulated wear with cleaning medium/brush)No impact on device performance; clear self-cure dental acrylic not affectedThe clear self-cure dental acrylic placed over the DentureID Microchip was not affected by the scrubbing action. Therefore, cleaning dentures will not impact the performance of DentureID Microchip.

    Study Information:

    The provided document describes non-clinical performance testing to demonstrate the substantial equivalence of the DentureID Microchip to its predicate devices. It does not detail a clinical study involving human readers or a training set in the context of typical AI device evaluation.

    Here's a breakdown of the available information:

    2. Sample Sizes and Data Provenance for Test Set:

    • Cytotoxicity, Irritation, Sensitization: These tests involve biological samples (e.g., cell cultures, animals). The document doesn't specify the exact number of samples/animals used but refers to the standards (ISO 10993-5, ISO 10993-10) which define such sample sizes. The data provenance is implied to be laboratory testing in facilities adhering to these standards, likely in the US given the FDA submission. These are retrospective tests conducted on device materials.
    • Chemical Characterization: Not specified, but involved proprietary standards of Nelson Laboratories and ChemTech Ford Laboratories. Retrospective laboratory testing.
    • EMC Testing: Not specified for the number of devices tested, but it's laboratory testing of the DentureID Microchips. Retrospective.
    • MRI Compatibility: Not specified for the number of devices tested, but it's laboratory testing of DentureID Microchips to ASTM standards. Retrospective.
    • Simulated Wear: Not specified for the number of DentureID Microchips installed in dentures, but it's a non-clinical performance test over a 5-year simulation. Retrospective.

    3. Number of Experts and their Qualifications for Ground Truth:

    • This information is not applicable in the context of this non-clinical performance testing. The "ground truth" for these tests is established by adhering to recognized international standards (ISO, ASTM, EN) and laboratory protocols, with results interpreted by qualified laboratory personnel (e.g., toxicologists, engineers) who execute these specific tests.

    4. Adjudication Method for the Test Set:

    • Not applicable. These are objective, quantitative and qualitative laboratory tests against defined scientific and engineering standards, not subjective interpretations requiring adjudication by experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was NOT done. This document focuses on the technical and safety performance of the device itself (hardware and its interaction with a smartphone app), not on the diagnostic or interpretative ability of human readers or AI in a clinical setting. Therefore, there is no effect size of how much human readers improve with AI vs. without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance):

    • Yes, in essence, standalone performance testing was done for the core function of the microchip. The various tests (Biocompatibility, EMC, MRI Compatibility, Simulated Wear, Migration Testing, Performance Testing of Implanted Transponder) evaluate the physical and functional characteristics of the DentureID Microchip itself, independent of a human interpreting its output. The "algorithm" here is the RFID chip's ability to respond with its serial number and the smartphone app's ability to retrieve information from a database. "Performance Testing of Implanted Transponder" is a direct measure of this standalone function.

    7. Type of Ground Truth Used:

    • The ground truth for this non-clinical testing is based on:
      • Standardized test methods and protocols: Defined by international standards (ISO, ASTM, EN) and proprietary laboratory standards.
      • Objective measurements and observations: E.g., cytotoxicity scores, temperature rises, torque measurements, pass/fail criteria for EMC.
      • Chemical analysis: For leachate testing.

    8. Sample Size for the Training Set:

    • Not applicable. This device (DentureID Microchip) is a passive RFID transponder and its associated database/app. It does not employ machine learning or AI models that require a "training set" in the conventional sense for image analysis or diagnostic tasks. Its function is to communicate a serial number, which then links to pre-entered data on a website.

    9. How Ground Truth for Training Set Was Established:

    • Not applicable, as there is no training set for an AI/ML model for this device. The information stored on the DentureID.com website, which is retrieved by the microchip's serial number, is entered by a dental lab or dentist. This user-provided information serves as the "ground truth" for the data it's designed to return, but it's not a training set for an algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170375

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-02

    (237 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nobilplast Denture Resin is intended for the fabrication of partial or full removable dentures, as well as occlusal splints and night quards.

    Device Description

    Nobilplast Denture Resin is a polyamide (nylon) resin available in pink and white colors. The resin granules are packaged in injectable aluminum cartridges or sold in a bulk container where users can fill empty aluminum cartridges that are sold separately.

    AI/ML Overview

    The FDA 510(k) summary for Nobilplast Denture Resin is a submission for substantial equivalence based on comparison to legally marketed predicate devices. The document does not describe a clinical study in the typical sense (e.g., a multi-reader, multi-case study with human-in-the-loop performance or standalone algorithm performance). Instead, it relies on bench testing and a comparison of material properties and indications for use to existing predicate devices.

    Therefore, many of the requested details about acceptance criteria, study design, ground truth establishment, and expert involvement are not applicable in this context. The document focuses on demonstrating that the new device is as safe and effective as the predicate devices based on material characteristics and standard tests.

    Here is an attempt to address the prompts based on the provided document, noting where information is not applicable (N/A) for this type of submission:


    Acceptance Criteria and Device Performance for Nobilplast Denture Resin

    The acceptance criteria for Nobilplast Denture Resin are implicitly based on demonstrating that its material properties and performance characteristics are substantially equivalent to its primary predicate device, FLEXSTAR V, and other reference devices (TCS Unbreakable, DENTSOLL Rezen NF, and Valplast). This is achieved through a series of bench tests and a comparison of intended use and material composition.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission based on direct comparison rather than a device with specific quantifiable "performance" in a diagnostic or assistive capacity, the acceptance criteria are framed as "similarity" or "comparability" to established predicate devices. The "reported device performance" is how Nobilplast Denture Resin's measured properties compare to those of the predicate.

    PropertyAcceptance Criteria (Implicit for Substantial Equivalence to Predicates)Reported Nobilplast Denture Resin Performance
    Intended UseMust be substantially similar to predicate devices for fabrication of partial/full removable dentures, occlusal splints, and night guards.Intended for the fabrication of partial or full removable dentures, as well as occlusal splints and night guards. (Minor differences from reference devices are explained as not practical for the material or application).
    Type and ClassMust align with predicate devices (Type III, Class 1 denture material).Type III, Class 1.
    Chemical FamilyMust be Polyamide, similar to predicate devices.Polyamide.
    CAS NumberShould be identical or closely related to the primary predicate (FLEXSTAR V: CAS# 25587-80-8).CAS# 25587-80-8.
    Material used to investMust be Dental Stone, similar to predicate devices.Dental Stone.
    Storage TemperatureMust be within a similar range to predicate devices (e.g., 60°F – 90°F).60°F – 90°F.
    DosageMust be consistent (1 cartridge).1 cartridge.
    Processing Time/TemperatureMust be comparable to predicate devices (e.g., primary predicate: 11 minutes @ 550°F (290°C)).11 minutes @ 550°F (290°C) (furnace heats one cartridge at a time).
    Injection TimeMust be comparable to predicate devices (e.g., primary predicate: 3 minutes).3 minutes (using manual press).
    Cool TimeMust be comparable to predicate devices (30 minutes).30 minutes.
    Physical PropertiesMust demonstrate expected properties (semi-flexible, rigid thick sections, flexible thin sections for clasps) consistent with the material type and intended use, similar to predicate devices.Partial denture is semi-flexible: Thick sections are rigid to hold denture teeth into place while thinner areas, such as clasps, are slightly flexible to accommodate undercuts on abutment teeth. (Identical to description for FLEXSTAR V, DENTSOLL Rezen NF, and Valplast).
    Bench Test PerformanceMust meet standard industry specifications (ASTM, ISO) for material properties (Density, Melting Point, Hardness, Tensile, Flexural, Impact, Sorption & Solubility, DSC, TGA, Melt Flow Rate).Subjected to various bench tests (Density per ASTM D792, Melting Point per ISO 11357, Hardness Shore per ISO 868, Tensile per ISO 527, Ultimate Flexural and Flexural Modulus per ISO 178, Charpy Impact ISO 179/1eU, Sorption & Solubility per ISO 20795, DSC Analysis per ASTM D3418, TGA Analysis per ASTM E1311, Melt Flow Rate Analysis, Izod Impact Strength per ISO 180 or ASTM D256). Specific numerical results are not provided in this summary but are implied to be within acceptable ranges to demonstrate equivalence.
    BiocompatibilityMust demonstrate no biocompatibility issues.Testing results indicate no biocompatibility issues exist for Nobilplast Denture Resin.
    Shelf LifeMust demonstrate a reasonable shelf-life, comparable to predicate.5-year shelf-life demonstrated by real-time packaging study on the nearly identical product: FlexStar V.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a "sample size" in terms of cases or images for testing an AI device. Instead, it refers to the number of samples used for various bench tests of the material properties. These are standard material science tests, and the sample sizes would be defined by the specific ASTM/ISO standards referenced (e.g., a certain number of specimens for tensile strength, impact, etc.). The document does not provide these detailed sample counts.
    • Data Provenance: The data provenance is from laboratory bench testing of the Nobilplast Denture Resin material, as well as biological testing. This is not clinical data derived from patients or images. It is prospective testing performed on the manufactured material. The country of origin for the data is not explicitly stated, but the company is based in the USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    N/A. This is a material testing submission, not an AI or diagnostic device submission requiring expert human interpretation of data for ground truth. The "ground truth" here is the measured physical and chemical properties of the material, verified through standardized testing protocols (ASTM, ISO).

    4. Adjudication Method for the Test Set

    N/A. No human adjudication is involved for material properties testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    N/A. This type of study is for diagnostic or assistive devices involving human readers interpreting cases. Nobilplast Denture Resin is a material for fabricating dentures.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    N/A. This is not an algorithm or software device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Standardized Bench Testing: Direct measurement of material properties (e.g., density, melting point, hardness, tensile strength, flexural properties, impact strength, sorption & solubility) according to internationally recognized ASTM and ISO standards.
    • Biocompatibility Testing: Evaluation of biological response as suggested by FDA Guidance (Cytotoxicity, Irritation, Sensitization tests).
    • Comparison to Predicate Devices: The established and legally marketed performance and safety profiles of the predicate devices (FLEXSTAR V, TCS Unbreakable, DENTSOLL Rezen NF, Valplast) serve as a de facto "ground truth" for substantial equivalence.

    8. The Sample Size for the Training Set

    N/A. There is no training set as this is not a machine learning or AI device.

    9. How the Ground Truth for the Training Set Was Established

    N/A. Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072851

    Validate with FDA (Live)

    Date Cleared
    2008-02-20

    (139 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K072402

    Validate with FDA (Live)

    Device Name
    FLEX-STAR V
    Date Cleared
    2007-10-19

    (53 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CMP's Flex-Star V is used for the fabrication of partial or full removable dentures, as well as occusal splints and night guards. It is to be used only by professional dental practitioners who make or repair dentures for patients

    Device Description

    Flex-Star V is used for the fabrication of partial or full removable dentures, as well as occusal splints and night guards.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and study information:

    Based on the provided document, which is a 510(k) premarket notification letter from the FDA, the device in question is a "Denture Relining, Repairing, or Rebasing Resin" called "Flex-Star V."

    This type of FDA correspondence, specifically a 510(k) clearance, does not typically contain detailed information about specific acceptance criteria or the study data that "proves" the device meets those criteria in the way a clinical trial report would. Instead, a 510(k) submission generally relies on demonstrating "substantial equivalence" to a legally marketed predicate device. This means showing that the new device has the same intended use and technological characteristics as a predicate device, or different technological characteristics that do not raise new questions of safety and effectiveness.

    Therefore, the requested information elements related to detailed study design, sample sizes, expert qualifications, and specific performance metrics cannot be found within this FDA clearance letter.

    However, I can extract what limited information is available and explain why other requested details are absent:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Implied)Reported Device Performance (Implied by Substantial Equivalence)
    Safe for dental useEquivalent safety profile to predicate devices.
    Effective for its intended use (fabrication of partial/full removable dentures, occlusal splints, night guards)Equivalent effectiveness for its intended use compared to predicate devices.
    Biocompatible (not explicitly stated but a general requirement for dental materials)Biocompatibility equivalent to predicate devices.
    Meets predicate device specifications (e.g., strength, durability, aesthetics)Performance characteristics (including physical and chemical properties) are substantially equivalent to predicate devices.

    Explanation: In a 510(k), specific numerical performance criteria and their corresponding results are generally not published in the clearance letter itself. Instead, the FDA reviews the manufacturer's submission which would have contained this data, demonstrating that the Flex-Star V performs as safely and effectively as a legally marketed predicate device. The letter states the device is "substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices."

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not specified in this document.
    • Data Provenance (country of origin, retrospective/prospective): Not specified in this document.

    Explanation: The letter does not detail the specific studies performed for the 510(k) submission, let alone the sample sizes or provenance of the data used in those studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    Explanation: This information would be specific to clinical or performance testing, which is not detailed in this FDA clearance letter.

    4. Adjudication method for the test set:

    • Adjudication Method: Not specified.

    Explanation: Similar to the above, this pertains to study design not present in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, this device is a dental resin, not an AI-assisted diagnostic tool. Therefore, an MRMC study related to human readers improving with AI is irrelevant and was not conducted.

    Explanation: This category doesn't apply to a denture repair/rebase resin.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance Study: No, this device is a material, not an algorithm.

    Explanation: This category doesn't apply to a denture repair/rebase resin.

    7. The type of ground truth used:

    • Type of Ground Truth: For a dental material, "ground truth" would typically relate to established physical, chemical, and biological performance standards (e.g., strength, water absorption, biocompatibility, bond strength to other materials, wear resistance) rather than expert consensus on images or pathology. This would be established through laboratory testing (e.g., ISO standards for dental materials) and potentially non-clinical in vitro/in vivo studies.

    Explanation: The relevant "ground truth" for a dental resin would be its adherence to material science specifications and biocompatibility, not clinical outcomes in the same way a diagnostic device would. These would be assessed through lab tests.

    8. The sample size for the training set:

    • Sample Size for Training Set: Not applicable in the context of this type of device and 510(k) submission.

    Explanation: "Training set" implies machine learning or AI, which is not relevant for a dental resin.

    9. How the ground truth for the training set was established:

    • How Ground Truth for Training Set was Established: Not applicable.

    Explanation: This question is not applicable to the device described.


    In summary: The provided document is an FDA 510(k) clearance letter for a medical device (a dental resin). It confirms that the device is "substantially equivalent" to predicate devices, allowing it to be marketed. However, it does not contain the detailed study protocols, acceptance criteria, or performance data that would be found in a comprehensive study report for a diagnostic AI or imaging device. The questions posed are largely designed for diagnostic/AI devices, not for a material such as a dental resin.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033020

    Validate with FDA (Live)

    Date Cleared
    2003-12-19

    (84 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CMP' IMPAK denture relining, repairing/rebasing liquid resin is indicated for relining a denture surface which contains tissue, for repairing a fractured denture, or forming a new denture base.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a dental device. It does not contain information about acceptance criteria, device performance, study designs, sample sizes, expert qualifications, or ground truth establishment. This type of information is typically found in the 510(k) summary or in the actual premarket notification submission, but not in the clearance letter itself.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1